Philips IntelliVue MP20: troubleshooting and error codes
The ServiceForge knowledge base holds 83 documented symptoms for the Philips IntelliVue MP20, with 164 ranked causes, 30 diagnostic checks with test points and expected values, and 20 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Integrated display is blank or brightness reduced
Screen is blank or has reduced brightness, possibly due to overheating from high room temperature or nearby instruments
Ranked causes
- 1. Display overheating from room temperature or nearby instruments (common)
- 2. Flat Panel Adapter flex not connected (occasional)
- 3. Backlight Inverter Cable not connected (occasional)
- 4. Backlight tubes defective (occasional)
- 5. Backlight Inverter board defective (rare)
- 6. Flat Panel Adapter flex defective (rare)
- 7. LCD Flat panel defective (rare)
- 8. Main board defective (rare)
Verify room temperature and monitor environment
- Test point
- ambient environment
- Expected
- not exceeding 40°C or below 5°C
- Tools
- thermometer
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Touch Screen not functioning
External touchscreen device does not respond to user input
Ranked causes
- 1. Touchscreen functionality temporarily disabled (common)
- 2. Touchscreen functionality permanently disabled (occasional)
- 3. External Touch cable not connected (occasional)
- 4. External Touch driver configuration issue (occasional)
- 5. MIB Board defective (rare)
- 6. External touch defective (rare)
- 7. Main board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power On LED and Error LED remain off after power button pressed caution
Both green Power On LED and red Error LED do not illuminate when the power button is pressed
Ranked causes
- 1. Remote devices interfering with boot (common)
- 2. Power Switch Micro Controller hung (common)
- 3. Carrier Power on LED PCA not connected (occasional)
- 4. I/O Board defective (occasional)
- 5. Touch controller, flex adapter, MSL board, recorder PCA, battery PCA, or IIT module defective (occasional)
- 6. Carrier Power on LED PCA defective (rare)
- 7. Main Board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch Screen not functioning
Touchscreen does not respond to user input
Ranked causes
- 1. Touchscreen functionality temporarily disabled (common)
- 2. Touchscreen functionality permanently disabled (occasional)
- 3. HIF PCA not connected (occasional)
- 4. Touch panel cable not connected (occasional)
- 5. HIF PCA or touch sensor defective (rare)
- 6. Main board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC Power LED does not light up caution
Green AC power indicator does not illuminate when monitor is powered with AC connection
Ranked causes
- 1. AC mains not powered or power cord disconnected (common)
- 2. Carrier Power on LED PCA not connected (occasional)
- 3. Carrier Power on LED PCA defective (occasional)
- 4. Main Board defective (rare)
- 5. Power supply defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Display is blank
Slave external monitor shows no image
Ranked causes
- 1. Video cable to external display not connected or defective (common)
- 2. External display has no power (common)
- 3. External display is defective (occasional)
- 4. System Interface board defective (rare)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED stays on continuously caution
Red Error LED remains illuminated without blinking during operation
Ranked causes
- 1. External connected device defective (common)
- 2. I/O Board defective (occasional)
- 3. Touch controller, flex adapter, MSL board, recorder PCA, battery PCA, or IIT module defective (occasional)
- 4. Measurement Server Mount defective (rare)
- 5. Main Board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote input device not functioning
Mouse or keyboard attached to Remote Extension Device does not respond
Ranked causes
- 1. Remote Extension Device not connected to monitor (common)
- 2. Input Device not connected properly (occasional)
- 3. Input Device defective (occasional)
- 4. Remote Extension Device defective (rare)
- 5. Remote Device I/O board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpeedPoint not functioning
SpeedPoint attached to Remote Extension Box does not respond
Ranked causes
- 1. Remote Extension Device not connected to monitor (common)
- 2. SpeedPoint not connected properly (occasional)
- 3. SpeedPoint defective (occasional)
- 4. Remote Extension Device defective (rare)
- 5. Remote Device I/O board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not sound patient safety
Alarm occurs but no alarm sound is issued from speaker
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to zero (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier on main board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm on screen but no LED or sound on alarm device patient safety
Screen displays alarm but Remote Alarm Device produces no LED or audio indication
Ranked causes
- 1. Cabling not connected (common)
- 2. Cabling defective (occasional)
- 3. I/O board defective (occasional)
- 4. Remote Alarm Device defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse on Remote Extension Box not functioning
Keyboard or mouse connected to Remote Extension Box does not respond
Ranked causes
- 1. Remote Extension Box not connected or input device disconnected (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. Remote Extension Device defective (rare)
- 4. Remote Device I/O board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor in docking station has AC Power but network connection could not be established
Monitor powered via docking station but LAN connection fails
Ranked causes
- 1. Docking interface board not installed in monitor (common)
- 2. LAN cable not connected or defective (common)
- 3. Flex cable in docking station not connected properly or defective (occasional)
- 4. Docking Station main board defective (rare)
Verify docking interface board installation in monitor
- Test point
- Monitor internal docking connector
- Expected
- Docking interface board present and seated
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequent wireless dropouts and network disconnects patient safety
Monitor frequently loses connection to wireless network
Ranked causes
- 1. Excessive interference from other radio equipment or microwave (common)
- 2. System capacity exceeded in coverage area (occasional)
- 3. Access point configuration problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message MsmtSrv not Supp issued caution
INOP message displayed: MsmtSrv not Supp
Ranked causes
- 1. Wrong software revision compatibility (common)
- 2. Too many measurement servers connected (occasional)
- 3. Unsupported type of measurement server (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Unsupported LAN issued
INOP message issued with network icon displayed
Ranked causes
- 1. Network infrastructure failure (common)
- 2. Monitor connected to wrong network (occasional)
- 3. IP address conflict after infrastructure re-installation (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse not functioning
Keyboard or mouse attached directly to monitor does not respond
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. PS/2 I/O Board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement Server configuration not supported caution
Prompt message: Measurement Server Configuration not supported
Ranked causes
- 1. Unsupported Measurement Server Extension connected (common)
- 2. Measurement Server Extension defective (occasional)
- 3. Measurement Server defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not boot or constantly reboots but functions properly outside docking station caution
Monitor fails to boot or reboots repeatedly when docked; operates normally when undocked
Ranked causes
- 1. Flex cable in docking station not connected properly or defective (common)
- 2. Docking Connector soiled or blocked (common)
- 3. Docking Station main board defective (occasional)
Check flex cable connection in docking station
- Test point
- Flex cable connector
- Expected
- Cable connected and secured
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC LED on Docking Station does not light up caution
AC indicator LED on docking station is off
Ranked causes
- 1. AC connection not ok (common)
- 2. Docking Station or Docking System Interface board defective (occasional)
Verify AC mains power and power cord connection
- Test point
- AC mains inlet
- Expected
- Power present and cord properly seated
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC LED on Docking Station lights up but AC LED on monitor does not caution
Docking station AC LED illuminated but monitor AC LED remains off
Ranked causes
- 1. Monitor not docked correctly (common)
- 2. Docking Station defective (occasional)
Verify monitor docking lever position
- Test point
- Docking lever
- Expected
- Right-hand position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED does not light patient safety
Alarm occurs but no LED indicator lights up
Ranked causes
- 1. Environmental lighting too bright (common)
- 2. Main board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs illuminate but no alarm sound patient safety
Remote Alarm Device shows LED indicators but produces no audible alert
Ranked causes
- 1. Speaker defective (common)
- 2. Remote device I/F defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bed not being monitored at central patient safety
Monitor connected to LAN with IP address but bed not monitored at central station
Ranked causes
- 1. Monitor not assigned to sector (common)
- 2. Duplicate Equipment Label on network (common)
Measurement Server LEDs normal but does not start caution
Measurement Server LEDs are normal but device does not power up
Ranked causes
- 1. Communication lines broken in MSL cable or connector (common)
- 2. MSL connector board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Navigation Point not functioning
Navigation Point device attached directly to monitor does not respond
Ranked causes
- 1. Navigation Point not connected properly (common)
- 2. Navigation Point defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to PIC with no error message patient safety
No prompt or error message on monitor but no connection to Philips Information Center
Ranked causes
- 1. Hardware defect in LAN connection (common)
- 2. Configuration problem (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Network icon (IntelliVue 802.11 adapter) patient safety
No network icon or icon flashing with IntelliVue 802.11 Bedside Adapter
Ranked causes
- 1. Communication problem with 802.11 adapter or low RSSI (common)
- 2. Configuration problem in WLAN settings (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No test sound issued during boot
Speaker does not produce expected test tone during system startup
Ranked causes
- 1. Speaker defective (common)
- 2. Main Board defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Other Bed Overview not available
Other Bed Overview functionality not working
Ranked causes
- 1. Configuration problem (common)
- 2. Wireless configuration not supporting feature (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor recorder print quality
Printhead produces faint, streaky or low quality output
Ranked causes
- 1. Printhead dirty (common)
- 2. Printhead failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED blinks repeatedly (cyclic reboots) caution
Red Error LED flashes repeatedly indicating recurring system reboots
Ranked causes
- 1. Hardware failure detected (common)
- 2. Software fault in operating system (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
802.11 Bedside Adapter LEDs off and remain off caution
LEDs on IntelliVue 802.11 Bedside Adapter are not illuminating
Ranked causes
- 1. Communication problem or cable disconnected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs do not illuminate
Alarm indicator LEDs remain off during boot sequence or operation
Ranked causes
- 1. Main Board defective (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sound issued but no LEDs light up
Audio alarm is heard but Remote Alarm Device LED indicators remain off
Ranked causes
- 1. LED failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery deeply discharged or not communicating caution
Indicator shows CHARGE BATT1 or CHARGE BATT2 with INOP tone
Ranked causes
- 1. Battery deeply discharged or communication failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Barometric pressure check
Verify measured ambient pressure matches reference barometric pressure for accurate CO2 measurement
Ranked causes
- 1. Measured ambient pressure differs from reference (unknown)
Compare reference barometric pressure with measured ambient pressure on monitor
- Test point
- monitor display, reference source (airport, weather station)
- Expected
- difference less than 12 mmHg, <12 mmHg
Calibration: CO2 calibration verification 10%
Verify CO2 sensor accuracy using 10% calibration gas as final verification
Ranked causes
- 1. Measured CO2 value differs from calculated value beyond tolerance (unknown)
Compare measured CO2 value to calculated value using 10% calibration gas
- Test point
- monitor display with 10% calibration gas applied
- Expected
- ±(0.07 × calculated value), ±(0.07 × calculated value)
Calibration: CO2 gas calibration check 5%
Verify CO2 sensor accuracy using 5% calibration gas
Ranked causes
- 1. Measured CO2 value differs from calculated value by more than 2.6 mmHg (unknown)
Compare measured CO2 value to calculated value using 5% calibration gas
- Test point
- monitor display with calibration gas applied
- Expected
- 5% of ambient pressure ±2 mmHg; difference less than 2.6 mmHg, <2.6 mmHg
Calibration: CO2 pump and IR source operating time
Verify CO2 pump and IR source operating hours are within safe limits (15,000 hours continuous use) and reset counters after component replacement
Ranked causes
- 1. CO2 pump or IR source operating time exceeds 15,000 hours (common)
Display Setup CO2 menu and observe operating time counters
- Test point
- Reset PumpOpTime and Reset IRSourceTime selections
- Expected
- less than 15,000 hours of continuous use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow rate check
Verify CO2 sensor gas flow rate is within tolerance at 50.0 ml/min
Ranked causes
- 1. Gas flow rate differs from expected 50.0 ml/min (unknown)
Measure gas flow rate through sensor
- Test point
- flowmeter
- Expected
- 50.0 ml/min ±7.5 ml/min, <7.5 ml/min difference
Calibration: Leakage check parts 1 and 2
Verify CO2 system leakage does not exceed acceptable limits on both measurement points
Ranked causes
- 1. Leakage exceeds acceptable rate (unknown)
Measure leakage rate on part 1 using flowmeter
- Test point
- flowmeter
- Expected
- less than 4.0 ml/min, <4.0 ml/min
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Microstream CO2
Microstream carbon dioxide module calibration procedure during preventive maintenance
Ranked causes
- 1. Preventive maintenance requirement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP
Non-invasive blood pressure module calibration procedure during preventive maintenance
Ranked causes
- 1. Preventive maintenance requirement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Noise level check
Verify CO2 sensor noise index does not exceed acceptable threshold
Ranked causes
- 1. Noise index exceeds 3.0 (unknown)
Observe noise index displayed on monitor
- Test point
- monitor display
- Expected
- noise index less than 3.0, <3.0
Calibration: Pump occlusion pressure check
Verify CO2 pump can generate adequate pressure during occlusion condition
Ranked causes
- 1. Pump pressure during occlusion below 120 mmHg (unknown)
Measure pressure difference between cell and ambient during occlusion
- Test point
- monitor display during occlusion
- Expected
- greater than 120 mmHg, >120 mmHg
High battery temperature caution
Temperature indicator shows CHECK BATT TEMP with INOP tone
Ranked causes
- 1. Ventilation openings blocked or monitor exposed to heat (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message CHECK INPUT DEVICES
Monitor displays INOP message indicating input device problem
Ranked causes
- 1. Navigation Point or other input device defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Installation verification: display via video output
Test sequence required when monitor is installed with external display connected to video output
Ranked causes
- 1. Installation of monitor with video output display (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Installation verification: Docking Station
Test sequence required when monitor is installed on a Docking Station
Ranked causes
- 1. Installation of monitor with Docking Station (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Installation verification: IntelliVue 802.11 Bedside Adapter
Test sequence required when monitor is installed with IntelliVue 802.11 Bedside Adapter wireless module
Ranked causes
- 1. Installation of monitor with 802.11 Bedside Adapter (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Installation verification: IntelliVue Instrument Telemetry
Test sequence required when monitor is installed with IntelliVue Instrument Telemetry (IIT) module
Ranked causes
- 1. Installation of monitor with IIT module (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mainstream CO2 accuracy check
Verify M2501A mainstream CO2 sensor measurement accuracy using 5% calibration gas at sensor operating temperature
Ranked causes
- 1. Sensor measurement exceeds tolerance (rare)
Attach M2501A CO2 sensor to patient monitor with airway adapter
- Test point
- sensor attachment point
- Expected
- sensor firmly attached, disconnected from patient circuit
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement Server does not start (no LEDs) caution
Measurement Server does not start up with no LED activity and no INOP or prompt
Ranked causes
- 1. No power to Measurement Server (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cannot be properly docked to the docking station caution
Monitor will not dock correctly with docking station
Ranked causes
- 1. Docking interface board not correctly installed in the monitor (common)
Verify docking interface board installation
- Test point
- Monitor internal docking connector
- Expected
- Docking interface board present and seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor label not set patient safety
Prompt message issued: no central assigned to this bed
Ranked causes
- 1. Monitor label not set in monitor (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MSL Power High INOP issued caution
INOP message: MSL Power High; if persists >15 min shows INOP MSL Power Off
Ranked causes
- 1. Attached devices drawing too much power (common)
Navigation Point Knob Rotation or Selection not functioning
Navigation Point knob rotation or selection control does not respond
Ranked causes
- 1. Navigation Point defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity in multi-access point coverage area patient safety
Some parts of coverage area with multiple access points have no connectivity
Ranked causes
- 1. Configuration problem in access points (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG OUT signal to defibrillator high voltage
ECG OUT signal missing to defibrillator interface
Ranked causes
- 1. Main board defective (occasional)
Test ECG OUT signal with known good defibrillator and cables
- Test point
- Defib connector
- Expected
- ECG signal present with marker pulse
- Tools
- Known good defibrillator, ECG cables, MMS
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- No IP address assignment
- No LAN connection
- No marker pulse displayed on monitor
- No network icon or icon flashes (IIT adapter)
- Other Bed alarms not appearing
- Recorder door detected as open
- Recorder shows out of paper
- Remote Extension Device buttons not functioning
- Repair verification: Docking Station
- Repair verification: general IntelliVue system repairs
- Repair verification: IIT module
- Repair verification: impact damage
- Repair verification: M3015A module
- Repair verification: unit opened
- Sidestream CO2 accuracy check
- Touch position invalid
- Touch position invalid (external)
- Wireless adapter defective
- Wireless Adapter firmware revision wrong
- Wireless Adapter has no power
- Wireless monitor losing connection to access point
- Wireless monitor out of range of access point
- Wireless overview or printing not working
23 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
BATTERIES EMPTY / BATT 1/BATT 2 EMPTY | Estimated remaining battery-powered operating time less than 10 minutes | Battery capacity monitoring indicates critical low level | critical |
BATTERIES INCOMPAT / BATT 1/BATT 2 INCOMPAT | Indicated batteries cannot be used with this monitor | Battery type or protocol mismatch detected | error |
BATTERIES LOW / BATT 1/BATT 2 LOW | Estimated battery-powered operating time remaining is less than 20 minutes | Battery capacity monitoring indicates low level | warning |
BATTERIES MALF / BATT 1/BATT 2 MALFUNCTION | Monitor cannot determine battery status or communication problem between battery and board | Battery status unknown or communication failure detected | error |
CHARGER MALFUNCT | Problem with battery charger in monitor | Charger circuit failure detected | error |
Check Alarm Lamps | Alarm lamp indicator circuit failure | Alarm LED indicator circuit malfunction | error |
CHECKINTERNVOLTAGE / CHECK MONITOR FUNC | Problem with too low voltages (5V, 12V) in the monitor | Internal voltage monitoring detects insufficient voltage | error |
CHECK MONITOR TEMP | The temperature inside the monitor is too high | Internal temperature sensor detects excessive heat | warning |
CHECK NURSE CALL RELAY | Nurse Call Relay board failure | Nurse Call Relay I/O board malfunction | error |
CO2 EQUIP MALF | CO2 measurement equipment malfunction | CO2 parameter module or MMS device failure | error |
CO2 OCCLUSION | CO2 occlusion detected or pump operation time reset | Resetting the PumpOpTime counter in Setup CO2 menu | warning |
ECG EQUIP MALF | ECG measurement equipment malfunction | ECG parameter module or MMS device failure | error |
INOP BAD SERVER LINK | Unexpected data detected on MSL or incompatible software | Invalid data on serial link or incompatible MMS software revision | error |
MSL Power Off INOP | Attached devices drawing too much power from monitor | Multiple MMS devices or high power device connected to monitor | error |
MSL Power Overload INOP | Short circuit within MSL system | Electrical short detected in Multi-Measurement Server system | error |
SERVERLINK MALF | Hardware for communicating with Multi-Measurement Server is faulty | Communication hardware failure with periodic beep every two seconds | error |
SETTINGS MALFUNCTION | Problem during configuration cloning or corrupted settings memory | Configuration cloning process fails or settings memory corrupted | error |
SPEAKER MALFUNCTION | Audio output system failure | Speaker, cable, or sound amplifier circuit fault | warning |
SpO2 EQUIP MALF | SpO2 measurement equipment malfunction | SpO2 parameter module or MMS device failure | error |
SpO2 TRANSDUC MALF | SpO2 transducer malfunction | SpO2 transducer sensor failure | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.