Philips HeartStart XLT: troubleshooting and error codes
The ServiceForge knowledge base holds 127 documented symptoms for the Philips HeartStart XLT, with 250 ranked causes, 240 diagnostic checks with test points and expected values, and 30 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit unresponsive to power commands patient safety
Unit does not power up or respond when powered on after connection to power source
Ranked causes
- 1. Power source voltage low (common)
- 2. Failure in Power Module (occasional)
- 3. Short in unit's 12 volt wiring (occasional)
- 4. Short on Power PCA (occasional)
- 5. Short on Control PCA (occasional)
- 6. Failure on Power PCA during power-up attempt (occasional)
- 7. Failure on Control PCA during power-up attempt (occasional)
- 8. Main fuse open (common)
- 9. Open in battery wiring (occasional)
- 10. Short or other failure in printer or display (occasional)
- 11. Short or other failure in system (rare)
Connect Power Module to power source only and observe Power indicator
- Test point
- Power Module indicator
- Expected
- Power indicator turns RED or GREEN
4 more checks with pass and fail actions, 17 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Extended Self Test failure
One or more test results show Fail instead of Pass on Extended Self Test printout
Ranked causes
- 1. ROM test failure (rare)
- 2. RAM test failure (rare)
- 3. System integrity failure (rare)
- 4. Defib capacitor charge/discharge failure (occasional)
- 5. Front End DSP communication failure (occasional)
- 6. Timebase clock discrepancy (rare)
Verify test load is connected to pads cable
- Test point
- pads cable connector
- Expected
- test load connected
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzing noise during audio patient safety
Buzzing or unwanted noise heard when audio tones or voice prompts are active
Ranked causes
- 1. Damage to speaker label (common)
- 2. Debris between speaker and speaker label (common)
- 3. Speaker hardware loose (occasional)
- 4. Speaker failure (occasional)
- 5. Control PCA failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Won't charge in AED mode high voltage
Unit does not charge defibrillation capacitor when in AED mode
Ranked causes
- 1. Pads connector or internal wiring failure (occasional)
- 2. Power PCA failure (occasional)
- 3. Control PCA failure (occasional)
- 4. Interface PCA failure (occasional)
- 5. CHARGE key failure (occasional)
Try unit in Manual Mode
- Test point
- Mode selector
- Expected
- Unit charges normally in Manual Mode
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Won't charge in manual mode high voltage
Unit does not charge defibrillation capacitor when in manual mode
Ranked causes
- 1. Pads connector or internal wiring failure (occasional)
- 2. Power PCA failure (occasional)
- 3. Control PCA failure (occasional)
- 4. Interface PCA failure (occasional)
- 5. CHARGE key failure (occasional)
Try unit in AED Mode
- Test point
- Mode selector
- Expected
- Unit charges normally in AED Mode
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leads off or ECG electrode contact loss patient safety
ECG monitoring electrodes not applied or not making proper electrical contact with patient
Ranked causes
- 1. Monitoring electrodes not properly applied (common)
- 2. ECG cable not connected (common)
- 3. Internal cable from ECG connector to Interface PCA disconnected (rare)
- 4. Failure on Interface PCA (rare)
Inspect electrode placement on patient
- Test point
- 3- or 5-lead monitoring electrodes
- Expected
- Electrodes firmly adhered to skin
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery life caution
Battery depletes quickly during use
Ranked causes
- 1. Frequent use without adequate charging (common)
- 2. Battery failure (occasional)
- 3. Power module failure (occasional)
- 4. Control PCA, Power PCA, or Battery PCA failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Shock Delivered patient safety
Shock not delivered due to patient impedance out of range or pad application issue.
Ranked causes
- 1. Patient impedance out of range (occasional)
- 2. Pads not applied properly (common)
- 3. Defective pads (occasional)
- 4. Defective pads cable (rare)
Verify pads properly applied to patient
- Test point
- Defib pads
- Expected
- Pads fully adhered with good skin contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ECG trace - constant baseline noise
Constant noise on ECG baseline indicating possible configuration or interference issue.
Ranked causes
- 1. Incorrect power line frequency or filter settings (common)
- 2. Nearby source of constant interference (occasional)
- 3. ECG front end failure (rare)
- 4. Signal processing circuit failure (rare)
Verify power line frequency setting
- Test point
- Configuration menu
- Expected
- Power line frequency set to 50 or 60 Hz matching local mains
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper moves but printing is faint or absent
Printer paper advances normally but no printing or very faint printing appears on paper
Ranked causes
- 1. Door improperly latched (occasional)
- 2. Dirty printhead (common)
- 3. Printer failure (occasional)
- 4. Control PCA failure (occasional)
Verify printer door is properly latched
- Test point
- Printer door latch
- Expected
- Door fully closed and latched
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor SpO2 signal quality
SpO2 monitoring shows no response, no value on screen, or no pleth bar
Ranked causes
- 1. Bad internal connection (common)
- 2. SpO2 PCA failure (occasional)
- 3. Control PCA failure (occasional)
- 4. Interface PCA failure (occasional)
Inspect and reseat flex cables between SpO2 connector and PCA modules
- Test point
- SpO2 connector to SpO2 PCA and SpO2 PCA to Interface PCA
- Expected
- Cables fully seated
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer test failure caution
Printer does not produce proper test output or shows stuck printhead elements
Ranked causes
- 1. Printhead elements stuck (occasional)
- 2. Paper speed incorrect (occasional)
- 3. Character/symbol quality degraded (occasional)
- 4. Stray marks in test output (rare)
Run Printer Test from Diagnostic menu
- Test point
- printer output
- Expected
- test patterns with diamonds, vertical bars, diagonal lines as shown in Figure 2-4
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio distorted patient safety
Audio tones or voice prompts are distorted or unclear
Ranked causes
- 1. Damage to speaker label (common)
- 2. Speaker failure (occasional)
- 3. Control PCA failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backlight doesn't light in dark ambient conditions
Display backlight fails to illuminate when ambient light is low
Ranked causes
- 1. Display failure (occasional)
- 2. Control PCA failure (occasional)
- 3. Ambient light sensor failure (occasional)
Run Display Test in Diagnostic Mode and test BACKLIGHT ON and BACKLIGHT OFF softkeys
- Test point
- Diagnostic mode Display Test backlight controls
- Expected
- Backlight responds to softkey commands
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Charges but disarms when press shock high voltage
Unit charges successfully but disarms discharge when SHOCK key is pressed
Ranked causes
- 1. Patient impedance sensed as too high or too low during energy delivery (common)
- 2. Power PCA failure (occasional)
- 3. Control PCA failure (occasional)
Inspect pads for proper contact and cable for damage
- Test point
- Pads on patient and pads cable connectors
- Expected
- Pads fully adhered, cable intact and properly connected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data Card not recognized caution
Data card is not recognized when plugged into unit
Ranked causes
- 1. Unit power already on during card insertion (common)
- 2. Card full or unsupported type (occasional)
- 3. Card removed while unit power on (data corrupted) (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display all white or all black
LCD display shows no response with either all white or all black screen
Ranked causes
- 1. Contrast misadjusted (common)
- 2. Control PCA failure (occasional)
- 3. Display failure (occasional)
Adjust display contrast control
- Test point
- Contrast adjustment control
- Expected
- Display shows readable text and waveforms
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display blank but backlight works
LCD display is blank with no visible content even though backlight illuminates
Ranked causes
- 1. Bad connections involving Display Data Cable (common)
- 2. Display failure (occasional)
- 3. Control PCA failure (occasional)
Inspect and reseat Display Data Cable connections
- Test point
- Display data cable connectors at display and Control PCA
- Expected
- Cables fully seated and making contact
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Done indicator does not illuminate caution
Done indicator does not light even after extended charging (>14.5 hours)
Ranked causes
- 1. System/Shift Test resets charging clock (common)
- 2. Battery aged, worn out, or failed (occasional)
- 3. Power Module failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
FE Extended Self Test failure
Front End (FE) failure detected during Extended Self Test.
Ranked causes
- 1. FE failure on Control PCA (rare)
- 2. FE failure on Power PCA (rare)
- 3. FE failure on Interface PCA (rare)
Replace Control PCA and rerun Extended Self Test
- Test point
- Control PCA module
- Expected
- Extended Self Test passes without FE failure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leads Off message despite cable replacement and proper connection
Leads Off error persists after ECG cable replaced and properly connected to simulator.
Ranked causes
- 1. Bad connection from ECG Connector to Interface PCA (common)
- 2. Defective ECG Connector or cable to Interface PCA (occasional)
- 3. Defective Interface PCA (occasional)
Reconnect ECG Connector to Interface PCA cable
- Test point
- Cable connection at Interface PCA
- Expected
- Cable fully seated with no Leads Off message
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery warning tone
Repeating tone alternating between two frequencies beginning 1 minute before shutdown due to low battery charge.
Ranked causes
- 1. Low battery charge (common)
- 2. Power Module fault (occasional)
- 3. General service needed (rare)
Check battery charge level and status
- Test point
- Battery compartment and charge indicator
- Expected
- Battery fully charged or Power Module connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main Key unresponsive (OFF, 1, 2, 3) caution
One or more main keys (OFF, 1, 2, 3) fail to respond
Ranked causes
- 1. Key failure (common)
- 2. Interface PCA failure (occasional)
- 3. Control PCA failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ECG trace - intermittent bursts
Intermittent bursts of noise or random spikes on ECG trace.
Ranked causes
- 1. Time-varying interference source (common)
- 2. ECG front end failure (occasional)
- 3. Signal processing circuit failure (rare)
Identify and relocate away from interference
- Test point
- Patient cable routing and unit location
- Expected
- Cables routed away from radio transmitters, pagers, walkie-talkies, X-ray equipment
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Other keys unresponsive (display, printer, manual/pacing) caution
Keys other than main keys fail to respond
Ranked causes
- 1. Key failure (common)
- 2. Interface PCA failure (occasional)
- 3. Control PCA failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pads Off message despite cable replacement and proper connection
Pads Off error persists after cable replaced and properly connected to simulator.
Ranked causes
- 1. Bad connection from Patient Connector to Power PCA (common)
- 2. Defective Patient Connector or cable to Power PCA (occasional)
- 3. Defective Power PCA (occasional)
Reconnect Patient Connector to Power PCA cable
- Test point
- Cable connection at Power PCA
- Expected
- Cable fully seated with no Pads Off message
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor ECG signal quality patient safety
Multifunction pads not making proper contact with patient; signal shows artifacts or is too weak to detect
Ranked causes
- 1. Pads not properly applied or making poor contact (common)
- 2. Multifunction pads outdated or dried-out (common)
- 3. Radio frequency interference (RFI) (occasional)
Visually inspect pad application and patient skin contact
- Test point
- Pads on patient chest
- Expected
- Pads fully adhered to clean, dry skin
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor Leads ECG signal quality patient safety
Weak or poor quality ECG signal from leads monitoring electrodes.
Ranked causes
- 1. Monitoring electrodes not making proper contact with patient (common)
- 2. Electrodes outdated or dried out (common)
- 3. Radio frequency interference (RFI) (occasional)
Verify electrode placement and skin contact
- Test point
- Patient body and electrode pads
- Expected
- Electrodes properly applied with good skin contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
QRS beeper inaudible or irregular
QRS beeper does not sound or beeps do not occur with each QRS complex
Ranked causes
- 1. QRS beeper is configured to Off (common)
- 2. Volume set too low (common)
- 3. QRS complex amplitude too small to detect (occasional)
Verify QRS beeper configuration in menu settings
- Test point
- Unit menu display
- Expected
- QRS beeper set to On
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service Unit message during Shift/System Check
Service Unit message appears during Shift/System Check indicating maintenance required.
Ranked causes
- 1. Data Card failure (occasional)
- 2. Lithium battery failure (occasional)
- 3. Control PCA failure (occasional)
Inspect and replace Data Card
- Test point
- Data Card slot
- Expected
- New empty M3510A/M5510A Data Card inserted
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not respond to power-on patient safety
No sounds, no display changes, no LED activity when power button pressed after battery insertion
Ranked causes
- 1. Discharged or defective battery (common)
- 2. Defective Data Card (occasional)
- 3. Defective Control PCA (occasional)
Verify battery charge level and condition
- Test point
- Battery M3516A/M5516A
- Expected
- Fully charged, undamaged
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unsupported Battery message displayed caution
Display shows 'Unsupported Battery' message on screen
Ranked causes
- 1. Not an M3516A battery installed (common)
- 2. Battery damage (occasional)
- 3. Battery PCA failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Won't discharge high voltage
Unit charges but does not deliver shock when SHOCK key is pressed
Ranked causes
- 1. Control PCA failure (occasional)
- 2. Interface PCA failure (occasional)
- 3. SHOCK key failure (occasional)
Run Controls Test in diagnostic mode and test SHOCK key
- Test point
- Diagnostic mode Controls Test
- Expected
- SHOCK key responds to test
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Attach Pads message patient safety
Multifunction defib electrode pads are not making proper contact with patient.
Ranked causes
- 1. Pads not applied to patient (common)
- 2. Pads worn or defective (occasional)
Verify pad contact and placement
- Test point
- Defib pads on patient chest
- Expected
- Pads fully adhered to patient skin per package directions
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio test failure
Tones or voice prompts not produced or incorrect during Audio Test
Ranked causes
- 1. Audio amplifier or speaker malfunction (occasional)
- 2. Voice prompt encoder malfunction (occasional)
Run Audio Test and select Message Alert
- Test point
- unit speakers
- Expected
- repeating series of 3 short tones followed by pause
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fails Capacity Test caution
Battery does not pass Battery Capacity Test
Ranked causes
- 1. Battery aged, worn out, or failed (common)
- 2. Control PCA failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Black line running along paper
Printer output shows continuous black line running along length of paper
Ranked causes
- 1. Dots (printhead elements) stuck on (occasional)
- 2. Control PCA failure (occasional)
Print test pattern and observe output
- Test point
- Printer output
- Expected
- No continuous black line on paper
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charges but disarms spontaneously high voltage
Unit charges successfully but disarms discharge without operator pressing SHOCK key
Ranked causes
- 1. Unit sensed Pads Off or Cable Off (common)
- 2. DISARM key failure (intermittent) (occasional)
Inspect pads and cable connections
- Test point
- Pads on patient and pads cable connectors
- Expected
- Pads fully adhered, cable firmly connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charges slowly
Defibrillator charges in 4-5 seconds instead of normal 2-3 seconds
Ranked causes
- 1. Battery too old or not fully charged (common)
- 2. Unit senses unsupported battery (occasional)
Verify battery is fully charged and recent date
- Test point
- Battery status and date code
- Expected
- Full charge, recent manufacture date
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Printer message
Printer paper absent, jammed, or printer door not closed properly.
Ranked causes
- 1. Printer paper absent or jammed (common)
- 2. Printer door not closed properly (occasional)
Check printer paper status
- Test point
- Paper compartment
- Expected
- Paper loaded and not jammed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Contrast does not adjust for ambient temperature
Display contrast does not automatically adjust when ambient temperature changes
Ranked causes
- 1. Display failure (occasional)
- 2. Control PCA failure (occasional)
Verify automatic contrast adjustment feature in menu settings
- Test point
- Unit menu display settings
- Expected
- Automatic contrast adjustment enabled and functioning
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data Card Full message
Incident duration exceeds 2 hours, causing Data Card to reach capacity.
Ranked causes
- 1. Incident exceeds 2 hours duration (occasional)
- 2. Empty Data Card not inserted for incident (occasional)
Verify Data Card capacity before incident
- Test point
- Data Card
- Expected
- Empty M3510A/M5510A Data Card inserted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data Card test failure caution
Data Card fails test in Diagnostic Mode
Ranked causes
- 1. Card full or unsupported type (common)
- 2. Control PCA failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defib Disarmed message patient safety
Defibrillator disarmed due to compromised pad connection or mode change during charge.
Ranked causes
- 1. Pad connection compromised (common)
- 2. Mode changed from Manual to AED while charged (occasional)
Verify pad contact with patient
- Test point
- Defib pads
- Expected
- Pads in proper contact with patient skin
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display test failure
Display does not cycle through colors or backlight does not respond
Ranked causes
- 1. Display backlight malfunction (occasional)
- 2. Ambient light sensor blocked or faulty (occasional)
Run Display Test from Diagnostic menu
- Test point
- display
- Expected
- completely dark, then completely light, then black vertical bar scrolls left to right
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG amplifier noise verification patient safety
Measuring noise floor of Pads and Leads ECG amplifiers with simulator off or disconnected to verify noise performance
Ranked causes
- 1. Periodic verification of amplifier noise performance required during performance verification (common)
- 2. Electromagnetic interference from external sources (occasional)
Measure Leads input noise floor
- Test point
- Leads ECG input (Lead II), Diagnostic noise reading
- Expected
- 0 + 30 uV, +30
- Tools
- DMM or diagnostic display, simulator off
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG controls do not respond
One or more ECG control functions do not respond (e.g., select lead or ECG size).
Ranked causes
- 1. Failure in keypress detection/processing (common)
- 2. Failure in keys/connections (occasional)
Replace Control PCA and test ECG controls
- Test point
- Control PCA module and keypad
- Expected
- ECG control functions respond to key presses
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG test display issues
ECG Test displays incorrect Lead selection, AC Line Filter setting, or Front End status
Ranked causes
- 1. Incorrect AC Line Filter setting for region (common)
- 2. PCI setting enabled (rare)
Verify AC Line Filter setting matches local power frequency
- Test point
- ECG Test display AC Line Filter field
- Expected
- 50 Hz in 50 Hz areas; 60 Hz in 60 Hz areas
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG trace distorted patient safety
Distorted ECG waveform on display.
Ranked causes
- 1. ECG front end failure (occasional)
- 2. Signal processing circuit failure (rare)
Replace ECG front end PCA
- Test point
- Power PCA (pads) or Interface PCA (leads)
- Expected
- ECG trace clear and undistorted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flat line ECG - no waveform, no Leads Off or Pads Off message patient safety
No ECG waveform displayed with no Leads Off or Pads Off error message.
Ranked causes
- 1. Short in internal ECG wiring or front end (occasional)
- 2. Signal processing circuit failure (rare)
Replace ECG front end PCA
- Test point
- Power PCA (pads) or Interface PCA (leads)
- Expected
- ECG waveform appears on display
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio output patient safety
No audio tones or voice prompts heard
Ranked causes
- 1. Speaker failure (common)
- 2. Control PCA failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pads cable disconnected from defibrillator patient safety
Pads cable connector not connected to defibrillator unit
Ranked causes
- 1. Pads cable connector not locked in place (common)
- 2. Failure on Power PCA (rare)
Inspect pads cable connector
- Test point
- Pads cable connector at defibrillator
- Expected
- Connector fully seated and locked
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pads not making proper contact with patient patient safety
Defibrillation pads applied but not making adequate electrical contact with patient skin
Ranked causes
- 1. Pads not properly applied to patient skin (common)
- 2. Failure on Power PCA (rare)
Inspect pads application to patient
- Test point
- Defibrillation pads on patient chest
- Expected
- Pads firmly adhered with full skin contact
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper moves but print quality poor or dots missing
Printer paper advances but printing is incomplete with missing dots or poor quality output
Ranked causes
- 1. Dirty printhead (common)
- 2. Printer failure (occasional)
Inspect printhead for dirt or debris
- Test point
- Printer printhead
- Expected
- Printhead clean and all elements clear
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper moves then stops
Printer paper advances initially but stops during printing operation
Ranked causes
- 1. Door improperly latched (common)
- 2. Paper improperly loaded or jammed (common)
Verify printer door is properly latched
- Test point
- Printer door latch
- Expected
- Door fully closed and latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper won't move
Printer paper does not advance during printing operation
Ranked causes
- 1. Paper improperly loaded or jammed, or paper is wet (common)
- 2. Printer failure (occasional)
Inspect paper loading and jam condition
- Test point
- Printer paper path
- Expected
- Paper properly loaded, no jams, paper is dry
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Extended Self Test failure
SpO2 failure detected on SpO2 PCA or Control PCA during Extended Self Test.
Ranked causes
- 1. SpO2 failure on SpO2 PCA (rare)
- 2. SpO2 failure on Control PCA (rare)
Replace SpO2 PCA and rerun Extended Self Test
- Test point
- SpO2 PCA module
- Expected
- Extended Self Test passes without SpO2 failure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 noisy signal
Excessive noise in SpO2 reading due to patient movement, electrical interference, or optical interference.
Ranked causes
- 1. Excessive patient movement (common)
- 2. Electrical or optical interference (common)
Observe patient movement and sensor placement
- Test point
- Sensor site
- Expected
- Minimal movement at application site
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor fail
SpO2 cable not connected to device or cable/sensor is broken.
Ranked causes
- 1. SpO2 cable not connected (common)
- 2. SpO2 cable or sensor broken (occasional)
Verify SpO2 cable connection
- Test point
- M3500B/M5500B SpO2 connector
- Expected
- Cable firmly seated at connector
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails display test (ambient light readings)
Display diagnostic test fails on ambient light sensor readings
Ranked causes
- 1. Control PCA failure (occasional)
- 2. Interface PCA failure (occasional)
Run Display Test in diagnostic mode focusing on ambient light channel
- Test point
- Diagnostic mode Display Test ambient light sensor
- Expected
- Ambient light readings respond to light level changes
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Unit fails display test (backlight)
- Unit fails display test in diagnostic mode (general)
- Unit fails printer test in diagnostic mode
- Unit powers up with error message or blank screen
- Waveforms or text distorted on printout
- White line running along paper
- Attach Leads message
- Battery capacity test failure
- Charging indicator does not illuminate
- CODEC Extended Self Test failure
- Configuration lost or reset
- Data Card disabled or not recognized
- Data Card Extended Self Test failure
- Data Card Interrupted message
- Data Card Not In Service message
- Defib Extended Self Test failure
- Defibrillator Disarm test
- Defibrillator disarm test failure
- Defibrillator energy delivery verification (AC Power)
- Defibrillator energy delivery verification (Battery Power)
- Defibrillator subsystem failure
- Defibrillator test with AC power module
- Defibrillator test with battery power
- Earth Leakage Current test (AC power module)
- Earth leakage exceeds limit
- ECG amplifier gain verification
- ECG data acquisition system failure
- Incompatible Data Card message
- IRDA Extended Self Test failure
- Key Inactive message
- Loud buzzing or grinding noise from printer
- Low battery warning
- Monitor failure or subsystem error
- No Data Card Present tone
- No defibrillation pads detected
- Noisy ECG trace - low frequency periodic
- Normal power on sequence tone
- Pacer Extended Self Test failure
- Pacer output below specification
- Pacing function verification
- Pads cable not properly attached
- Patient Contact Impedance (PCI) function verification
- Patient lead leakage exceeds limit
- Patient Lead Leakage test - Defibrillator pads
- Patient Lead Leakage test - ECG cable
- Patient Lead Leakage test - SPO2 cable
- Printer paper jam or wet paper in printer
- RAM Extended Self Test failure
- ROM Extended Self Test failure
- SHOCK button not pressed within 30 seconds of charge
- SpO2 cable not connected
- SpO2 circuitry failure
- SpO2 sensor ambient light interference
- SpO2 signal too low for accurate reading
- SpO2 signal too weak or absent
- Stop Pacer message
- System Extended Self Test failure
- System Failure - Cycle Power
- System failure continuous beep tone
- System failure requiring service
- Timebase Extended Self Test failure
- Tones present but no voice prompt (AED Mode)
- Unit fails ECG test in diagnostic mode (leads)
- Unit fails ECG test in diagnostic mode (pads)
- Unit has minimal power response
- Unsupported battery detected
- Voice prompt present but no tones
67 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
00000 - 00400 | Defibrillator failure in charging circuits | Charging circuit malfunction detected during operation or self-test | critical |
01000 | Defibrillator failure in biphasic processor | Biphasic waveform processing error detected | critical |
02000 | Leads front end failure | ECG signal acquisition circuitry failure detected | error |
04000 | Pacing system failure | Pacing output or control circuitry failure detected | error |
08000 | Processor error | CPU or processor malfunction detected | error |
10000 - 1FFFF | System monitor failure | Monitoring subsystem fault detected | warning |
10001 | Processor synchronization error | Processor clock or timing synchronization fault | warning |
10004 | Synchronization timeout | Processor failed to synchronize within timeout period | warning |
20000 - 2FFFF | Front end failure | Signal acquisition front end circuit failure | error |
30000 - 3FFFF | Pacing system failure | Pacer circuitry or output failure | error |
40000 - 4FFFF | Monitor processor failure | Monitoring processor circuit failure | error |
50000 - 5FFFF | SpO2 system problem | SpO2 sensor or measurement circuit failure | warning |
60000 - 6FFFF | Advisory system failure | Advisory message generation circuit failure | warning |
80000 - 8FFFF | IRDA (infrared) communication failure | Infrared communication interface malfunction | warning |
90000 - 90002 | Self-test failure in RAM/ROM or gate array | Memory or gate array self-test failed | error |
90003 | Self-test failure in data card circuits | Data card interface circuit self-test failed | error |
90004 and 90005 | Self-test failure in codec or time base | Audio codec or timing circuit self-test failed | error |
90006 | Self-test failure in SpO2 circuitry | SpO2 measurement circuit self-test failed | warning |
90007 and 90008 | Self-test failure in pacer or defibrillator | Pacer or defibrillator circuit self-test failed or operator touched unit during test | error |
90009 | Self-test failure in front end | Signal acquisition front end self-test failed | error |
9000A | Lithium backup battery failure | Backup battery voltage out of specification or failed | warning |
A0000 - A7FFF | Data card failure | Data card read/write or interface failure | error |
A8000 - AFFFF | Data archival error | Data storage or retrieval from card failed | warning |
B0000 - BFFFF | Audio system failure | Audio amplifier or speaker circuitry failure | warning |
F0000 - FFFFF | RAM or ROM memory failure | Main system memory self-test failure | critical |
F0002 | Promo mode failure | Demonstration or promotional mode circuitry failure | warning |
F0003 | Keyscanner failure | Keyboard scanning circuit malfunction | error |
F0004 | System communication error | Internal system bus or communication protocol failure | critical |
F0007 | System communication error | Transient or intermittent communication protocol failure | warning |
System Log errors | Hardware or firmware errors detected and logged by system | Error conditions detected during unit operation | unknown |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.