Philips HeartStart MRx: troubleshooting and error codes

The ServiceForge knowledge base holds 143 documented symptoms for the Philips HeartStart MRx, with 263 ranked causes, 153 diagnostic checks with test points and expected values, and 52 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Status log messages indicate component failures patient safety

Device logs INOP messages and communications failures for various modules including NBP, SpO2, ECG, Processor, Fan, Printer, and Therapy components

Ranked causes

  1. 1. Bluetooth card malfunction (occasional)
  2. 2. Charge button mechanical failure (occasional)
  3. 3. CO2 module equipment malfunction (occasional)
  4. 4. Communication error with modules (common)
  5. 5. CPR sensor communication or selftest failure (occasional)
  6. 6. ECG equipment malfunction (occasional)
  7. 7. ECG cable failure (occasional)
  8. 8. Fan assembly failure (occasional)
  9. 9. NBP module selftest or communication failure (occasional)
  10. 10. Shock or charge button failure (occasional)

21 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Won't charge in manual defib mode high voltage

Device cannot charge when charge button pressed on paddles or device in manual defibrillation mode

Ranked causes

  1. 1. Paddles not connected properly (common)
  2. 2. Paddles defective (common)
  3. 3. Internal connection problem (occasional)
  4. 4. Therapy port defective (occasional)
  5. 5. Therapy PCA defective (occasional)
  6. 6. Processor PCA failure (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 measurement not displayed patient safety

'Failure to display CO2 measurement' or no CO2 value shown.

Ranked causes

  1. 1. FilterLine issue or internal connection failure (common)
  2. 2. Internal CO2 tubing broken or disconnected (occasional)
  3. 3. CO2 module failure (occasional)
  4. 4. Processor PCA failure (rare)
  5. 5. Therapy PCA failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Distorted or buzzing audio

Audio output is distorted or produces buzzing/grinding noise

Ranked causes

  1. 1. Damaged speaker cover or label (common)
  2. 2. Debris between speaker and label (common)
  3. 3. Speaker hardware loose (occasional)
  4. 4. Speaker damage or failure (occasional)
  5. 5. Processor PCA failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No transmission device detected

Status log shows 'No Transmission Device Detected' when device is not responding or unreachable

Ranked causes

  1. 1. Bluetooth device not turned on (common)
  2. 2. Bluetooth device out of range (common)
  3. 3. Bluetooth device not authorized to pair (occasional)
  4. 4. Bluetooth pairing information lost (occasional)
  5. 5. Bluetooth card is corrupt (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No transmission devices detected via RS-232

Status log shows 'No Transmission Devices Detected' message during RS-232 setup

Ranked causes

  1. 1. Cell phone not connected properly to serial cable/RS-232 port (common)
  2. 2. Cell phone RS-232 port not configured correctly (occasional)
  3. 3. Cell phone RS-232 hardware incompatible (occasional)
  4. 4. Serial cable is defective or incompatible (occasional)
  5. 5. Cell phone not set up properly as modem (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor ECG signal quality - monitoring electrodes patient safety

Noisy trace, wandering baseline, or other poor signal quality from ECG monitoring electrodes.

Ranked causes

  1. 1. Poor skin preparation or electrode contact (common)
  2. 2. Bad ECG cable or leads (common)
  3. 3. Radio frequency interference (RFI) (occasional)
  4. 4. Problem with measurement module panel (occasional)
  5. 5. Problem with Processor PCA (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transmission failed – connection failed or interrupted

Status log shows 'Transmission Failed. Connection Failed' or 'Connection Interrupted' message

Ranked causes

  1. 1. Wrong Bluetooth Profile selected (common)
  2. 2. Network down or cellular signal insufficient (common)
  3. 3. Bluetooth device out of range (common)
  4. 4. Incorrect dial string in Phone Profile settings (occasional)
  5. 5. Data transfer service unavailable on phone (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Won't discharge shock in manual defib mode high voltage

Device cannot deliver shock when shock button pressed on paddles or device in manual mode

Ranked causes

  1. 1. Faulty paddles (common)
  2. 2. Internal connection problem (occasional)
  3. 3. Therapy port defective (occasional)
  4. 4. Therapy PCA defective (occasional)
  5. 5. Processor PCA failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

'Leads Off' message - ECG lead open circuit patient safety

'Leads Off' error displayed even though ECG cable and leads replaced and properly connected, indicating open circuit in leads wiring.

Ranked causes

  1. 1. Poor skin preparation or electrode contact (common)
  2. 2. Bad ECG cable or leads (common)
  3. 3. Problem with measurement module panel (occasional)
  4. 4. Problem with Processor PCA (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 reading inaccurate

CO2 measurement readings are incorrect or unreliable

Ranked causes

  1. 1. FilterLine blocked or damaged (common)
  2. 2. CO2 module needs calibration (common)
  3. 3. CO2 module failure (occasional)
  4. 4. Internal tubing connection problem (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPR sensor unplugged patient safety

Compression sensor or cables not properly connected, sensor not detected

Ranked causes

  1. 1. Compression sensor unplugged (common)
  2. 2. Pads/CPR cable not connected to therapy port (common)
  3. 3. Pads/CPR cable faulty (occasional)
  4. 4. Compression sensor faulty (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display completely dim or dark

Display screen completely dim, unresponsive, all light or all dark; no display output

Ranked causes

  1. 1. Backlight failure (common)
  2. 2. Display failure (common)
  3. 3. Power supply failure (occasional)
  4. 4. Processor PCA failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Faint or absent printing

Paper advances but printing is faint, absent, or poor quality

Ranked causes

  1. 1. Door improperly latched (common)
  2. 2. Dirty printhead (common)
  3. 3. Operating temperature out of range (occasional)
  4. 4. Printer failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Bluetooth devices detected

Status log shows 'No Bluetooth Devices Detected' when HeartStart MRx cannot discover Bluetooth device

Ranked causes

  1. 1. Bluetooth device turned off (common)
  2. 2. Bluetooth device out of range (common)
  3. 3. Bluetooth device not authorized to pair (occasional)
  4. 4. Bluetooth card is corrupt (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Data Card Present message

System displays message indicating no data card is detected even when card is inserted

Ranked causes

  1. 1. Insufficient time allowed for data card recognition (common)
  2. 2. Data card failure (occasional)
  3. 3. Data card not seated properly due to bent pins (occasional)
  4. 4. Processor PCA failure (occasional)
First check

Verify data card is fully inserted

Test point
Data card slot
Expected
Card fully seated in slot

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No network connectivity icon present

Network connectivity icon is missing from clinical views, indicating broken network connection

Ranked causes

  1. 1. Physical wired connection broken (common)
  2. 2. Physical wireless connection broken (common)
  3. 3. IntelliVue Network option not enabled (occasional)
  4. 4. Network settings or configuration incorrect (occasional)
First check

Physical wired and wireless connections

Test point
Physical connectors
Expected
Properly seated and secured

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor ECG signal quality - paddles/pads patient safety

Noisy trace, wandering baseline, or other poor signal quality from therapy paddles or pads.

Ranked causes

  1. 1. Poor skin preparation or pads/paddles contact (common)
  2. 2. Radio frequency interference (RFI) (occasional)
  3. 3. Problem with internal cables (occasional)
  4. 4. Problem with Therapy port (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Radio Unplugged INOP

Radio Unplugged or Check Device message displayed, indicating loose or disconnected radio/AC module or RS-232 cable

Ranked causes

  1. 1. Loose RS-232 connection (common)
  2. 2. Loose radio/AC module connection (common)
  3. 3. Damaged RS-232 cable (occasional)
  4. 4. Faulty RS-232 transceiver (rare)
First check

RS-232 cable connection tightness

Test point
RS-232 screwlocks at connection
Expected
Screwlocks tight, cable fully seated

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Won't charge using device charge button high voltage

Device won't charge when charge button on HeartStart MRx pressed in manual mode

Ranked causes

  1. 1. Therapy cable issue (common)
  2. 2. Front panel button failure (occasional)
  3. 3. Therapy PCA defective (occasional)
  4. 4. Processor PCA failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

'Pads Off' message - therapy port open circuit patient safety

'Pads Off' error displayed even though pads cable replaced and properly connected, indicating open circuit in therapy port wiring.

Ranked causes

  1. 1. Cable connection fault from Therapy port to Therapy PCA (common)
  2. 2. Defective Therapy port or cable to Power PCA (occasional)
  3. 3. Defective Power PCA (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth connection lost during transmission

Status log shows 'Bluetooth Connection Lost' message during 12-Lead and Event Summary transmission or testing

Ranked causes

  1. 1. Network is down (occasional)
  2. 2. Bluetooth device out of range (common)
  3. 3. Bluetooth connection not established (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth device pairing failed

Status log shows 'Bluetooth Device Pairing Failed' when attempting to pair device with HeartStart MRx

Ranked causes

  1. 1. Wrong pass key entered on Bluetooth device (common)
  2. 2. Pairing process timed out (occasional)
  3. 3. Bluetooth card is corrupt (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 exhaust tube blocked

CO2 exhaust tube or outlet port blocked, preventing measurement sample collection

Ranked causes

  1. 1. FilterLine problem (common)
  2. 2. Exhaust tubing kinked or blocked (common)
  3. 3. Internal tubing or CO2 module failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device disarms during shock high voltage

Device charges OK but disarms when shock button pressed or spontaneously without button press

Ranked causes

  1. 1. Pads or paddles losing contact or defective (common)
  2. 2. Therapy PCA defective (occasional)
  3. 3. Processor PCA failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid password during transmission

Status log shows 'Invalid Password' message with user/pw failure sub-codes for 12LXmit Network or Server

Ranked causes

  1. 1. Wrong Bluetooth Profile selected (common)
  2. 2. Incorrect PPP User Name or Password in Phone settings (common)
  3. 3. Server User Name or Password is incorrect (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Central Monitoring INOP or no network connectivity icon patient safety

No Central Monitoring message displayed or network icon missing, indicating no connection to IntelliVue Central Station

Ranked causes

  1. 1. No network connection - general (common)
  2. 2. No network connection - wired only (occasional)
  3. 3. No network connection - wireless only (occasional)
First check

Equipment Label configuration

Test point
Main screen display or Device Info
Expected
Equipment Label present, unique, not blank

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No CO2 waveform

CO2 module does not start when FilterLine connected or no CO2 waveform displayed

Ranked causes

  1. 1. FilterLine damaged (common)
  2. 2. Intake receptacle failure (occasional)
  3. 3. CO2 module or sensor failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper not moving through printer

Printer paper does not advance through printer mechanism

Ranked causes

  1. 1. Paper improperly loaded or jammed (common)
  2. 2. Paper is wet (common)
  3. 3. Printer failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Radio Malfunction INOP patient safety

Radio Malfunction or Check Device message displayed, indicating radio/AC module or Therapy PCA failure

Ranked causes

  1. 1. Radio/AC module malfunction (common)
  2. 2. Therapy PCA malfunction (occasional)
  3. 3. Incompatible Therapy PCA hardware version (rare)
First check

Radio/AC module physical connection

Test point
Battery compartment B
Expected
Module seated with audible click

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 no response - no value displayed patient safety

No SpO2 value on screen, no pleth bar, no numeric reading.

Ranked causes

  1. 1. Bad SpO2 sensor (common)
  2. 2. Bad internal connection (occasional)
  3. 3. SpO2 PCA failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transmission failed – no dial tone

Status log shows 'Transmission Failed. No dial tone' message

Ranked causes

  1. 1. Cell phone service unavailable or signal insufficient (common)
  2. 2. Bluetooth modem connection not secure (occasional)
  3. 3. Connection does not provide dial tones (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth disconnect failed

Status log shows 'Disconnect Failed' for Bluetooth FTP Device after transmission

Ranked causes

  1. 1. Bluetooth device stack incompatibility (occasional)
  2. 2. Bluetooth device moved out of range after transmission (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 module accuracy patient safety

CO2 measurement readings do not fall within acceptable tolerance when exposed to calibration gas

Ranked causes

  1. 1. CO2 measurement module requires calibration (common)
  2. 2. CO2 measurement module defective after recalibration failure (occasional)
First check

Warm up monitor with FilterLine connected

Test point
CO2 Inlet port
Expected
Operating for at least 20 minutes with FilterLine connected

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 module flow rate

Test and adjust flow rate to 50±7.5 ml/min; calibrate if out of tolerance

Ranked causes

  1. 1. Flow rate out of tolerance (not 50±7.5 ml/min) (occasional)
  2. 2. Flow rate cannot be adjusted to tolerance; defective CO2 module (rare)
First check

Disconnect tubing from flowmeter inlet

Test point
flowmeter inlet connection
Expected
tubing disconnected

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 module flow rate outside tolerance patient safety

Gas flow rate does not meet 50 ± 7.5 ml/min specification

Ranked causes

  1. 1. Flow rate requires adjustment or tubing obstruction (common)
  2. 2. CO2 module defective (rare)
First check

Measure CO2 flow rate

Test point
Flowmeter connected to FilterLine outlet
Expected
Flow rate 50 ± 7.5 ml/min, ±7.5 ml/min
Tools
Flowmeter

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control buttons not responding

Front panel buttons (Print, Sync, Charge, Shock, softkeys) not responding or responding incorrectly

Ranked causes

  1. 1. Processor PCA failure (common)
  2. 2. Display assembly failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controls do not respond to key presses patient safety

One or more controls (e.g., select lead, softkeys) do not respond to user input. Keypress not detected or processed.

Ranked causes

  1. 1. Bad connection between keys and detection circuits (common)
  2. 2. Failure in keys (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CPR sensor malfunction patient safety

Compression sensor power-on self-test failed, sensor not functioning

Ranked causes

  1. 1. Hardware version not at level B2 (common)
  2. 2. Compression sensor defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display half dim or uneven brightness

Left or right half of display is dim or display brightness is not uniform

Ranked causes

  1. 1. Backlight cable disconnected (common)
  2. 2. Backlight bulb failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid password during RS-232 transmission

Status log shows 'Invalid Password' message during RS-232 12-Lead transmission setup

Ranked causes

  1. 1. Incorrect PPP User Name or Password in Serial Profile Phone settings (common)
  2. 2. Server User Name or Password is incorrect (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP reading inaccurate patient safety

Blood pressure measurement values are inaccurate or incorrect.

Ranked causes

  1. 1. NBP module needs calibration (common)
  2. 2. NBP module failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No audio output

No audio or sound output from device speaker at all

Ranked causes

  1. 1. Speaker failure (common)
  2. 2. Processor PCA failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No dial tone during RS-232 transmission

Status log shows 'No Dial tone' message during RS-232 12-Lead transmission

Ranked causes

  1. 1. Cell phone not connected properly to serial cable (common)
  2. 2. Cell phone service unavailable (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paddle position error during operational check

PCI LEDs light up indicating paddles not properly seated in holders during Operational Check Setup

Ranked causes

  1. 1. Paddles not seated properly in holders (common)
  2. 2. Paddle electrode surfaces dirty or corroded (occasional)
First check

Verify paddles are secured in pockets

Test point
Paddle holders
Expected
PCI LEDs do not light; paddles firmly seated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Shock delivery problem high voltage

Device delivers incorrect energy, no energy, or only partial energy in shock

Ranked causes

  1. 1. Therapy PCA defective (common)
  2. 2. Processor PCA failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 noisy or intermittent signal patient safety

SpO2 signal is noisy, intermittent, or unreliable.

Ranked causes

  1. 1. Bad SpO2 sensor (common)
  2. 2. Processor PCA failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 reads wrong value patient safety

SpO2 reading is obviously incorrect or inaccurate.

Ranked causes

  1. 1. Bad SpO2 sensor (common)
  2. 2. SpO2 PCA failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 sensor disconnected patient safety

'SpO2 unplugged. Turn off SpO2?' message. Sensor is disconnected or not responding.

Ranked causes

  1. 1. SpO2 sensor disconnected or faulty (common)
  2. 2. Incompatible software after SpO2 repair (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy knob not responding

Therapy knob fails to respond during operational check or service mode controls test

Ranked causes

  1. 1. Therapy knob not responding (common)
  2. 2. Therapy switch defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transmission failed – error configuring device

Status log shows 'Transmission Failed. Error Configuring Transmission Device' message

Ranked causes

  1. 1. Wrong Bluetooth Profile selected (common)
  2. 2. Incorrect configuration string in Phone Profile settings (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transmission failed with directory or permission error

Status log shows 'Transmission Failed' with 'Host Directory Failure' for FTP device when directory cannot be found or created

Ranked causes

  1. 1. Destination directory missing or cannot be created on Bluetooth device (occasional)
  2. 2. User settings in Bluetooth device do not allow transaction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transmission failed with timeout or server error

Status log shows 'Transmission Failed' with timeout, request error, or HTTP client error conditions

Ranked causes

  1. 1. Network connection timeout or unavailable (occasional)
  2. 2. Server configuration or version mismatch (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio voice prompt not heard during operational check caution

Operator does not hear expected Shock Delivered or No Shock Delivered announcement after Shock test

Ranked causes

  1. 1. Audio system defective or volume too low (occasional)
First check

Listen for voice prompt after Shock test completes

Test point
Device speaker
Expected
Audible announcement of Shock Delivered or No Shock Delivered

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery compartment calibration recommended

Operational Check displays 'Cal Recommended' prompt for Battery Compartment A, indicating battery calibration is required

Ranked causes

  1. 1. Battery capacity or calibration drift (common)
First check

Respond to calibration prompt during Operational Check

Test point
Battery Compartment A test result
Expected
Cal Recommended or Pass

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery failure detected during operational check

Fail/BF result indicates low or defective battery

Ranked causes

  1. 1. Battery low or defective (common)
First check

Check battery charge level

Test point
Battery compartment
Expected
Battery charged to at least 20%

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth communication failure

Status log shows 'Bluetooth Communication Failure' indicating incompatible or corrupt Bluetooth card

Ranked causes

  1. 1. Bluetooth card is incompatible or corrupt (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth malfunction reported

Status log shows 'Bluetooth Malfunction' message with 'No/Bad Bluetooth Card' sub-message

Ranked causes

  1. 1. Bluetooth card not installed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 module caution

CO2 calibration procedure using calibration gases, regulator, flowmeter and associated equipment

Ranked causes

  1. 1. CO2 module requires calibration verification (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 module 10% accuracy check

Test measurement accuracy with 10% calibration gas; displayed reading must fall within ±7% tolerance of expected value

Ranked causes

  1. 1. 10% reading outside ±7% tolerance (occasional)
First check

Disconnect 5% gas and connect 10% calibration gas

Test point
CO2 Inlet port, calibration gas regulator
Expected
10% gas connected in place of 5%

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Calibration: CO2 module 5% accuracy check
  • Calibration: CO2 module ambient pressure
  • Calibration: CO2 module auto-calibration
  • Calibration: CO2 module leakage inlet
  • Calibration: CO2 module leakage outlet
  • Calibration: CO2 module noise check
  • Calibration: CO2 module pump strength
  • Calibration: CO2 module verification
  • Calibration: Internal Assemblies - Rear Case
  • Calibration: NBP module
  • Calibration: Performance Verification after External Repairs
  • Calibration: Performance Verification after Internal Repairs
  • Calibration: Performance Verification after Printer Replacement
  • Calibration: Performance Verification after Top Assembly Repair
  • Charge button does not respond during operational check
  • Clock battery replacement needed
  • CO2 calibration overdue
  • CO2 equipment malfunction
  • CO2 inlet tubing leakage detected
  • CO2 module calibration overdue
  • CO2 module noise exceeds threshold
  • CO2 module operating hours exceed service life
  • CO2 module pump vacuum insufficient
  • CO2 occlusion - FilterLine blocked
  • CO2 outlet tubing leakage detected
  • CO2 overrange - reading exceeds measurement range
  • CO2 service required
  • Data Card Full message
  • Data Card Full message persists after erasing data
  • Does not copy data to data card
  • Flat line on ECG display - no waveform, no 'Leads Off' message
  • Flat line on paddles/pads - no waveform, no 'Pads Off' message
  • Incompatible Data Card message
  • Instrument Telemetry Radio shows malfunction
  • Internal Memory Failure INOP appears every time device is turned on
  • Internal Memory Failure message with beep
  • Leads ECG test cable connection issue
  • Leads ECG test cable disconnection needed
  • NBP (blood pressure) measurement malfunction
  • NBP calibration overdue
  • NBP cuff will not deflate
  • NBP cycle counter exceeds service limit
  • NBP measurement failed
  • NBP module calibration overdue
  • NBP module calibration status out of range
  • NBP service required
  • Network Connection Rejection - Assigned Central is not available
  • Network Connection Rejection - Central cannot identify this bed
  • Network Connection Rejection - No Central assigned to this bed
  • Network Connection Rejection - No Central check software revision
  • Network Connection Rejection - No Central duplicate monitor label
  • No Bluetooth devices configured
  • No Bluetooth profiles configured
  • No Instrument Telemetry Radio detected
  • Operational check critical failure
  • Operational check failed with service or component replacement required
  • Operational check fails with battery failure
  • Operational check fails with cable problem
  • Operational check fails with compression sensor issue
  • Operational check fails with critical failure
  • Operational check fails with non-critical failure
  • Operational check fails with non-therapy component issue
  • Operational check fails with shock/pacing/ECG prevention issue
  • Operational check non-critical failure
  • Operational check passed but battery low
  • Paddle PCI LEDs continuously lit
  • Plastic shield and service loop routing in rear case
  • Press tabs into place during assembly
  • Printer malfunction
  • Radio Weak Signal INOP
  • Ready For Use indicator shows blinking red X with periodic chirp
  • Ready For Use indicator shows solid red X with periodic chirp
  • Ready For Use indicator shows solid red X without periodic chirp
  • Replace front case assembly
  • Replace front panel buttons
  • Shock button does not respond during operational check
  • SpO2 equipment malfunction
  • SpO2 sensor malfunction
  • Transmission failed – cannot reach server
  • Transmission failed – server unknown
  • Transmission failed – settings configured incorrectly
  • Transmission settings not configured
  • Weekly shock test fails to deliver correct energy

83 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Assigned Central is not availableAssigned IIC is not functional or powered offHeartStart MRx cannot connect to configured IICerror
Bad URL Format - 12LXmit ServerTransmission settings configured incorrectlyInvalid server URLerror
Calibration failed. Check that the pressure applied is correCalibration point entry failed due to incorrect pressure applicationWrong pressure applied when entering calibration point (0 or 250 mmHg)error
Central cannot identify this bedIIC unable to identify HeartStart MRx due to unexpected errorsUnexpected communication or configuration errorerror
Charge Button FailureCharge button not responding correctlyUser pressed No Button Response instead of Charge during operational checkerror
Configure Failed - 12LXmit DeviceError configuring transmission deviceConfiguration errorerror
Confirm retries exceeded (MRx Aborted Assoc)Confirmation timeout retries exceeded during network associationNo response to confirmation requests, keeps timing outwarning
Connect Failed - 12LXmit DialingConnection failed during dialingCannot establish connectionerror
Connection FailedNetwork connection rejected by IICIIC rejects connection attempt from HeartStart MRxerror
Critical Failure DetectedCritical device failure (software version prior to B.05)Device startup critical failurecritical
Device Error. Service RequiredDevice error requiring service (software version B.05 and greater)Device startup errorcritical
Device serial number not enteredSerial number not configuredDevice startup - no serial numberwarning
DNS query failure - 12LXmit ServerServer unknown - DNS resolution failedDNS timeout or failureerror
Fail/BFBattery failure during Operational CheckAutomated Tests detect low battery conditionwarning
Fail/CCritical failure during Operational CheckAutomated tests detect failure impacting defibrillation, pacing, or ECG acquisitioncritical
Fail/CXCable failure detectedAutomated operational check detects cable problemcritical
Fail/DNon-therapy component failure detectedAutomated operational check detects component failure not affecting therapyerror
Fail/DXDevice problem preventing therapy deliveryAutomated operational check detects failure affecting shock, pacing or ECG acquisitioncritical
Fail/NCNon-critical failure during Operational CheckAutomated tests detect failure not impacting life-saving functionalitywarning
Fail/SCompression sensor failure detectedAutomated operational check detects compression sensor failurewarning
HTTP client error - 12L TransmitWeb server rejected transmission data12-lead transmission HTTP errorerror
IIC Association Abort receivedIIC sent association abort signal to HeartStart MRxIIC terminates established connectionwarning
IIC Life Ticks timed out (MRx Aborted Assoc)Heartbeat timeout between HeartStart MRx and IICNo life tick received within timeout periodwarning
Incompatible PrinterIncorrect printer configuration detected50mm printer installed but software expects 75mm printererror
Internal Memory FailureInternal memory card errorMemory card malfunction or corruptioncritical
Modem Connection Lost - 12LXmit NetworkConnection interrupted during transmissionNetwork connection droppederror
NBP OverpressureNBP module opens relief valve when cuff pressure exceeds 300 mmHgCuff pressure exceeds 300 mmHgwarning
NBP Safety TimeoutNBP module times out if pressure remains > 10 mmHg for 3 minutesCuff kept pressurized for more than 3 minuteswarning
No Central assigned to this bedNo IIC configured to monitor the HeartStart MRx with assigned equipment labelDevice attempts network connection with unassigned equipment labelerror
No Central – check software revisionSoftware revisions incompatible between HeartStart MRx and IICVersion mismatch detected during connection attempterror
No Central – duplicate monitor labelAnother device with same equipment label already connected to IICSecond device with duplicate label attempts to connecterror
No Central MonitoringNo connection to IntelliVue Central Station or network connectivity icon missingNetwork connection is broken or not establishederror
No Data From BedIntelliVue Central Station not receiving data from HeartStart MRxNetwork connection failure or radio module malfunctionerror
No Instrument Telemetry Radio detectedRadio/AC module not detected by Instrument Telemetry service screenService screen cannot communicate with radio moduleerror
No Transmission Device Detected - Rosetta-LtRosetta-Lt transmission device not detectedNo transmission device found at startuperror
Operational Check PassedAll tests completed successfullyAll device components pass during Operational Checkinfo
Operational Check Passed. Check Battery A (or B)Tests pass but battery lowBattery charge low but not below operational thresholdwarning
Pacer Equip MalfunctionPacing system failurePacing hardware failure detected, incorrect current or measurementcritical
Partial transmission - 12L Transmit12-lead transmission incompletePhone disconnected during transmissionwarning
PassAll operational checks passedAutomated operational checks complete successfullyinfo
Printer MalfunctionPrinter hardware or communication failurePrinter is faulty or communication problem detectedwarning
Radio MalfunctionRadio/AC module or Therapy PCA hardware failureRadio module detects hardware failure or communication errorcritical
Radio UnpluggedRadio/AC module disconnected or RS-232 connection looseRS-232 cable or radio module connection failserror
Radio Weak SignalWireless radio signal strength below acceptable level (no INOP tone)RSSI value below -84 dBmwarning
Request Timeout - 12LXmit Server12-lead transmission failed due to server connection timeoutServer connection has timed outerror
Shock Button FailureShock button not responding correctlyUser pressed No Button Response instead of Shock during operational checkerror
Shock Equip MalfunctionDefibrillator shock delivery system failureDevice unable to deliver correct energy or no energy delivered during shockcritical
Single confirmation from IIC timed outConfirmation timeout for single transaction from IICNo confirmation received in expected time framewarning
Transmission Failed - PPP Attach Timeout12-lead transmission failed due to network timeoutPPP Attach Timeout on 12LXmit Networkerror
Transmission Failed - Rosetta-LtRosetta-Lt transmission failedTransmission errorerror
Transmission Timeout - Rosetta-LtRosetta-Lt transmission timeoutTransmission timed outerror
Unreachable - 12LXmit ServerCannot reach transmission serverServer unavailable or connection losterror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.