Philips HeartStart MRx: troubleshooting and error codes
The ServiceForge knowledge base holds 143 documented symptoms for the Philips HeartStart MRx, with 263 ranked causes, 153 diagnostic checks with test points and expected values, and 52 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Status log messages indicate component failures patient safety
Device logs INOP messages and communications failures for various modules including NBP, SpO2, ECG, Processor, Fan, Printer, and Therapy components
Ranked causes
- 1. Bluetooth card malfunction (occasional)
- 2. Charge button mechanical failure (occasional)
- 3. CO2 module equipment malfunction (occasional)
- 4. Communication error with modules (common)
- 5. CPR sensor communication or selftest failure (occasional)
- 6. ECG equipment malfunction (occasional)
- 7. ECG cable failure (occasional)
- 8. Fan assembly failure (occasional)
- 9. NBP module selftest or communication failure (occasional)
- 10. Shock or charge button failure (occasional)
21 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Won't charge in manual defib mode high voltage
Device cannot charge when charge button pressed on paddles or device in manual defibrillation mode
Ranked causes
- 1. Paddles not connected properly (common)
- 2. Paddles defective (common)
- 3. Internal connection problem (occasional)
- 4. Therapy port defective (occasional)
- 5. Therapy PCA defective (occasional)
- 6. Processor PCA failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 measurement not displayed patient safety
'Failure to display CO2 measurement' or no CO2 value shown.
Ranked causes
- 1. FilterLine issue or internal connection failure (common)
- 2. Internal CO2 tubing broken or disconnected (occasional)
- 3. CO2 module failure (occasional)
- 4. Processor PCA failure (rare)
- 5. Therapy PCA failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Distorted or buzzing audio
Audio output is distorted or produces buzzing/grinding noise
Ranked causes
- 1. Damaged speaker cover or label (common)
- 2. Debris between speaker and label (common)
- 3. Speaker hardware loose (occasional)
- 4. Speaker damage or failure (occasional)
- 5. Processor PCA failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No transmission device detected
Status log shows 'No Transmission Device Detected' when device is not responding or unreachable
Ranked causes
- 1. Bluetooth device not turned on (common)
- 2. Bluetooth device out of range (common)
- 3. Bluetooth device not authorized to pair (occasional)
- 4. Bluetooth pairing information lost (occasional)
- 5. Bluetooth card is corrupt (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No transmission devices detected via RS-232
Status log shows 'No Transmission Devices Detected' message during RS-232 setup
Ranked causes
- 1. Cell phone not connected properly to serial cable/RS-232 port (common)
- 2. Cell phone RS-232 port not configured correctly (occasional)
- 3. Cell phone RS-232 hardware incompatible (occasional)
- 4. Serial cable is defective or incompatible (occasional)
- 5. Cell phone not set up properly as modem (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor ECG signal quality - monitoring electrodes patient safety
Noisy trace, wandering baseline, or other poor signal quality from ECG monitoring electrodes.
Ranked causes
- 1. Poor skin preparation or electrode contact (common)
- 2. Bad ECG cable or leads (common)
- 3. Radio frequency interference (RFI) (occasional)
- 4. Problem with measurement module panel (occasional)
- 5. Problem with Processor PCA (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmission failed – connection failed or interrupted
Status log shows 'Transmission Failed. Connection Failed' or 'Connection Interrupted' message
Ranked causes
- 1. Wrong Bluetooth Profile selected (common)
- 2. Network down or cellular signal insufficient (common)
- 3. Bluetooth device out of range (common)
- 4. Incorrect dial string in Phone Profile settings (occasional)
- 5. Data transfer service unavailable on phone (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Won't discharge shock in manual defib mode high voltage
Device cannot deliver shock when shock button pressed on paddles or device in manual mode
Ranked causes
- 1. Faulty paddles (common)
- 2. Internal connection problem (occasional)
- 3. Therapy port defective (occasional)
- 4. Therapy PCA defective (occasional)
- 5. Processor PCA failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
'Leads Off' message - ECG lead open circuit patient safety
'Leads Off' error displayed even though ECG cable and leads replaced and properly connected, indicating open circuit in leads wiring.
Ranked causes
- 1. Poor skin preparation or electrode contact (common)
- 2. Bad ECG cable or leads (common)
- 3. Problem with measurement module panel (occasional)
- 4. Problem with Processor PCA (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 reading inaccurate
CO2 measurement readings are incorrect or unreliable
Ranked causes
- 1. FilterLine blocked or damaged (common)
- 2. CO2 module needs calibration (common)
- 3. CO2 module failure (occasional)
- 4. Internal tubing connection problem (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR sensor unplugged patient safety
Compression sensor or cables not properly connected, sensor not detected
Ranked causes
- 1. Compression sensor unplugged (common)
- 2. Pads/CPR cable not connected to therapy port (common)
- 3. Pads/CPR cable faulty (occasional)
- 4. Compression sensor faulty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display completely dim or dark
Display screen completely dim, unresponsive, all light or all dark; no display output
Ranked causes
- 1. Backlight failure (common)
- 2. Display failure (common)
- 3. Power supply failure (occasional)
- 4. Processor PCA failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Faint or absent printing
Paper advances but printing is faint, absent, or poor quality
Ranked causes
- 1. Door improperly latched (common)
- 2. Dirty printhead (common)
- 3. Operating temperature out of range (occasional)
- 4. Printer failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Bluetooth devices detected
Status log shows 'No Bluetooth Devices Detected' when HeartStart MRx cannot discover Bluetooth device
Ranked causes
- 1. Bluetooth device turned off (common)
- 2. Bluetooth device out of range (common)
- 3. Bluetooth device not authorized to pair (occasional)
- 4. Bluetooth card is corrupt (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Data Card Present message
System displays message indicating no data card is detected even when card is inserted
Ranked causes
- 1. Insufficient time allowed for data card recognition (common)
- 2. Data card failure (occasional)
- 3. Data card not seated properly due to bent pins (occasional)
- 4. Processor PCA failure (occasional)
Verify data card is fully inserted
- Test point
- Data card slot
- Expected
- Card fully seated in slot
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No network connectivity icon present
Network connectivity icon is missing from clinical views, indicating broken network connection
Ranked causes
- 1. Physical wired connection broken (common)
- 2. Physical wireless connection broken (common)
- 3. IntelliVue Network option not enabled (occasional)
- 4. Network settings or configuration incorrect (occasional)
Physical wired and wireless connections
- Test point
- Physical connectors
- Expected
- Properly seated and secured
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor ECG signal quality - paddles/pads patient safety
Noisy trace, wandering baseline, or other poor signal quality from therapy paddles or pads.
Ranked causes
- 1. Poor skin preparation or pads/paddles contact (common)
- 2. Radio frequency interference (RFI) (occasional)
- 3. Problem with internal cables (occasional)
- 4. Problem with Therapy port (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Radio Unplugged INOP
Radio Unplugged or Check Device message displayed, indicating loose or disconnected radio/AC module or RS-232 cable
Ranked causes
- 1. Loose RS-232 connection (common)
- 2. Loose radio/AC module connection (common)
- 3. Damaged RS-232 cable (occasional)
- 4. Faulty RS-232 transceiver (rare)
RS-232 cable connection tightness
- Test point
- RS-232 screwlocks at connection
- Expected
- Screwlocks tight, cable fully seated
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Won't charge using device charge button high voltage
Device won't charge when charge button on HeartStart MRx pressed in manual mode
Ranked causes
- 1. Therapy cable issue (common)
- 2. Front panel button failure (occasional)
- 3. Therapy PCA defective (occasional)
- 4. Processor PCA failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
'Pads Off' message - therapy port open circuit patient safety
'Pads Off' error displayed even though pads cable replaced and properly connected, indicating open circuit in therapy port wiring.
Ranked causes
- 1. Cable connection fault from Therapy port to Therapy PCA (common)
- 2. Defective Therapy port or cable to Power PCA (occasional)
- 3. Defective Power PCA (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth connection lost during transmission
Status log shows 'Bluetooth Connection Lost' message during 12-Lead and Event Summary transmission or testing
Ranked causes
- 1. Network is down (occasional)
- 2. Bluetooth device out of range (common)
- 3. Bluetooth connection not established (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth device pairing failed
Status log shows 'Bluetooth Device Pairing Failed' when attempting to pair device with HeartStart MRx
Ranked causes
- 1. Wrong pass key entered on Bluetooth device (common)
- 2. Pairing process timed out (occasional)
- 3. Bluetooth card is corrupt (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 exhaust tube blocked
CO2 exhaust tube or outlet port blocked, preventing measurement sample collection
Ranked causes
- 1. FilterLine problem (common)
- 2. Exhaust tubing kinked or blocked (common)
- 3. Internal tubing or CO2 module failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device disarms during shock high voltage
Device charges OK but disarms when shock button pressed or spontaneously without button press
Ranked causes
- 1. Pads or paddles losing contact or defective (common)
- 2. Therapy PCA defective (occasional)
- 3. Processor PCA failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid password during transmission
Status log shows 'Invalid Password' message with user/pw failure sub-codes for 12LXmit Network or Server
Ranked causes
- 1. Wrong Bluetooth Profile selected (common)
- 2. Incorrect PPP User Name or Password in Phone settings (common)
- 3. Server User Name or Password is incorrect (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Central Monitoring INOP or no network connectivity icon patient safety
No Central Monitoring message displayed or network icon missing, indicating no connection to IntelliVue Central Station
Ranked causes
- 1. No network connection - general (common)
- 2. No network connection - wired only (occasional)
- 3. No network connection - wireless only (occasional)
Equipment Label configuration
- Test point
- Main screen display or Device Info
- Expected
- Equipment Label present, unique, not blank
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No CO2 waveform
CO2 module does not start when FilterLine connected or no CO2 waveform displayed
Ranked causes
- 1. FilterLine damaged (common)
- 2. Intake receptacle failure (occasional)
- 3. CO2 module or sensor failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper not moving through printer
Printer paper does not advance through printer mechanism
Ranked causes
- 1. Paper improperly loaded or jammed (common)
- 2. Paper is wet (common)
- 3. Printer failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Radio Malfunction INOP patient safety
Radio Malfunction or Check Device message displayed, indicating radio/AC module or Therapy PCA failure
Ranked causes
- 1. Radio/AC module malfunction (common)
- 2. Therapy PCA malfunction (occasional)
- 3. Incompatible Therapy PCA hardware version (rare)
Radio/AC module physical connection
- Test point
- Battery compartment B
- Expected
- Module seated with audible click
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 no response - no value displayed patient safety
No SpO2 value on screen, no pleth bar, no numeric reading.
Ranked causes
- 1. Bad SpO2 sensor (common)
- 2. Bad internal connection (occasional)
- 3. SpO2 PCA failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmission failed – no dial tone
Status log shows 'Transmission Failed. No dial tone' message
Ranked causes
- 1. Cell phone service unavailable or signal insufficient (common)
- 2. Bluetooth modem connection not secure (occasional)
- 3. Connection does not provide dial tones (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth disconnect failed
Status log shows 'Disconnect Failed' for Bluetooth FTP Device after transmission
Ranked causes
- 1. Bluetooth device stack incompatibility (occasional)
- 2. Bluetooth device moved out of range after transmission (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 module accuracy patient safety
CO2 measurement readings do not fall within acceptable tolerance when exposed to calibration gas
Ranked causes
- 1. CO2 measurement module requires calibration (common)
- 2. CO2 measurement module defective after recalibration failure (occasional)
Warm up monitor with FilterLine connected
- Test point
- CO2 Inlet port
- Expected
- Operating for at least 20 minutes with FilterLine connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 module flow rate
Test and adjust flow rate to 50±7.5 ml/min; calibrate if out of tolerance
Ranked causes
- 1. Flow rate out of tolerance (not 50±7.5 ml/min) (occasional)
- 2. Flow rate cannot be adjusted to tolerance; defective CO2 module (rare)
Disconnect tubing from flowmeter inlet
- Test point
- flowmeter inlet connection
- Expected
- tubing disconnected
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 module flow rate outside tolerance patient safety
Gas flow rate does not meet 50 ± 7.5 ml/min specification
Ranked causes
- 1. Flow rate requires adjustment or tubing obstruction (common)
- 2. CO2 module defective (rare)
Measure CO2 flow rate
- Test point
- Flowmeter connected to FilterLine outlet
- Expected
- Flow rate 50 ± 7.5 ml/min, ±7.5 ml/min
- Tools
- Flowmeter
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control buttons not responding
Front panel buttons (Print, Sync, Charge, Shock, softkeys) not responding or responding incorrectly
Ranked causes
- 1. Processor PCA failure (common)
- 2. Display assembly failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Controls do not respond to key presses patient safety
One or more controls (e.g., select lead, softkeys) do not respond to user input. Keypress not detected or processed.
Ranked causes
- 1. Bad connection between keys and detection circuits (common)
- 2. Failure in keys (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPR sensor malfunction patient safety
Compression sensor power-on self-test failed, sensor not functioning
Ranked causes
- 1. Hardware version not at level B2 (common)
- 2. Compression sensor defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display half dim or uneven brightness
Left or right half of display is dim or display brightness is not uniform
Ranked causes
- 1. Backlight cable disconnected (common)
- 2. Backlight bulb failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid password during RS-232 transmission
Status log shows 'Invalid Password' message during RS-232 12-Lead transmission setup
Ranked causes
- 1. Incorrect PPP User Name or Password in Serial Profile Phone settings (common)
- 2. Server User Name or Password is incorrect (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP reading inaccurate patient safety
Blood pressure measurement values are inaccurate or incorrect.
Ranked causes
- 1. NBP module needs calibration (common)
- 2. NBP module failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No audio output
No audio or sound output from device speaker at all
Ranked causes
- 1. Speaker failure (common)
- 2. Processor PCA failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No dial tone during RS-232 transmission
Status log shows 'No Dial tone' message during RS-232 12-Lead transmission
Ranked causes
- 1. Cell phone not connected properly to serial cable (common)
- 2. Cell phone service unavailable (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paddle position error during operational check
PCI LEDs light up indicating paddles not properly seated in holders during Operational Check Setup
Ranked causes
- 1. Paddles not seated properly in holders (common)
- 2. Paddle electrode surfaces dirty or corroded (occasional)
Verify paddles are secured in pockets
- Test point
- Paddle holders
- Expected
- PCI LEDs do not light; paddles firmly seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Shock delivery problem high voltage
Device delivers incorrect energy, no energy, or only partial energy in shock
Ranked causes
- 1. Therapy PCA defective (common)
- 2. Processor PCA failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 noisy or intermittent signal patient safety
SpO2 signal is noisy, intermittent, or unreliable.
Ranked causes
- 1. Bad SpO2 sensor (common)
- 2. Processor PCA failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 reads wrong value patient safety
SpO2 reading is obviously incorrect or inaccurate.
Ranked causes
- 1. Bad SpO2 sensor (common)
- 2. SpO2 PCA failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor disconnected patient safety
'SpO2 unplugged. Turn off SpO2?' message. Sensor is disconnected or not responding.
Ranked causes
- 1. SpO2 sensor disconnected or faulty (common)
- 2. Incompatible software after SpO2 repair (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Therapy knob not responding
Therapy knob fails to respond during operational check or service mode controls test
Ranked causes
- 1. Therapy knob not responding (common)
- 2. Therapy switch defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmission failed – error configuring device
Status log shows 'Transmission Failed. Error Configuring Transmission Device' message
Ranked causes
- 1. Wrong Bluetooth Profile selected (common)
- 2. Incorrect configuration string in Phone Profile settings (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmission failed with directory or permission error
Status log shows 'Transmission Failed' with 'Host Directory Failure' for FTP device when directory cannot be found or created
Ranked causes
- 1. Destination directory missing or cannot be created on Bluetooth device (occasional)
- 2. User settings in Bluetooth device do not allow transaction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmission failed with timeout or server error
Status log shows 'Transmission Failed' with timeout, request error, or HTTP client error conditions
Ranked causes
- 1. Network connection timeout or unavailable (occasional)
- 2. Server configuration or version mismatch (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio voice prompt not heard during operational check caution
Operator does not hear expected Shock Delivered or No Shock Delivered announcement after Shock test
Ranked causes
- 1. Audio system defective or volume too low (occasional)
Listen for voice prompt after Shock test completes
- Test point
- Device speaker
- Expected
- Audible announcement of Shock Delivered or No Shock Delivered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery compartment calibration recommended
Operational Check displays 'Cal Recommended' prompt for Battery Compartment A, indicating battery calibration is required
Ranked causes
- 1. Battery capacity or calibration drift (common)
Respond to calibration prompt during Operational Check
- Test point
- Battery Compartment A test result
- Expected
- Cal Recommended or Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery failure detected during operational check
Fail/BF result indicates low or defective battery
Ranked causes
- 1. Battery low or defective (common)
Check battery charge level
- Test point
- Battery compartment
- Expected
- Battery charged to at least 20%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth communication failure
Status log shows 'Bluetooth Communication Failure' indicating incompatible or corrupt Bluetooth card
Ranked causes
- 1. Bluetooth card is incompatible or corrupt (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth malfunction reported
Status log shows 'Bluetooth Malfunction' message with 'No/Bad Bluetooth Card' sub-message
Ranked causes
- 1. Bluetooth card not installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 module caution
CO2 calibration procedure using calibration gases, regulator, flowmeter and associated equipment
Ranked causes
- 1. CO2 module requires calibration verification (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 module 10% accuracy check
Test measurement accuracy with 10% calibration gas; displayed reading must fall within ±7% tolerance of expected value
Ranked causes
- 1. 10% reading outside ±7% tolerance (occasional)
Disconnect 5% gas and connect 10% calibration gas
- Test point
- CO2 Inlet port, calibration gas regulator
- Expected
- 10% gas connected in place of 5%
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: CO2 module 5% accuracy check
- Calibration: CO2 module ambient pressure
- Calibration: CO2 module auto-calibration
- Calibration: CO2 module leakage inlet
- Calibration: CO2 module leakage outlet
- Calibration: CO2 module noise check
- Calibration: CO2 module pump strength
- Calibration: CO2 module verification
- Calibration: Internal Assemblies - Rear Case
- Calibration: NBP module
- Calibration: Performance Verification after External Repairs
- Calibration: Performance Verification after Internal Repairs
- Calibration: Performance Verification after Printer Replacement
- Calibration: Performance Verification after Top Assembly Repair
- Charge button does not respond during operational check
- Clock battery replacement needed
- CO2 calibration overdue
- CO2 equipment malfunction
- CO2 inlet tubing leakage detected
- CO2 module calibration overdue
- CO2 module noise exceeds threshold
- CO2 module operating hours exceed service life
- CO2 module pump vacuum insufficient
- CO2 occlusion - FilterLine blocked
- CO2 outlet tubing leakage detected
- CO2 overrange - reading exceeds measurement range
- CO2 service required
- Data Card Full message
- Data Card Full message persists after erasing data
- Does not copy data to data card
- Flat line on ECG display - no waveform, no 'Leads Off' message
- Flat line on paddles/pads - no waveform, no 'Pads Off' message
- Incompatible Data Card message
- Instrument Telemetry Radio shows malfunction
- Internal Memory Failure INOP appears every time device is turned on
- Internal Memory Failure message with beep
- Leads ECG test cable connection issue
- Leads ECG test cable disconnection needed
- NBP (blood pressure) measurement malfunction
- NBP calibration overdue
- NBP cuff will not deflate
- NBP cycle counter exceeds service limit
- NBP measurement failed
- NBP module calibration overdue
- NBP module calibration status out of range
- NBP service required
- Network Connection Rejection - Assigned Central is not available
- Network Connection Rejection - Central cannot identify this bed
- Network Connection Rejection - No Central assigned to this bed
- Network Connection Rejection - No Central check software revision
- Network Connection Rejection - No Central duplicate monitor label
- No Bluetooth devices configured
- No Bluetooth profiles configured
- No Instrument Telemetry Radio detected
- Operational check critical failure
- Operational check failed with service or component replacement required
- Operational check fails with battery failure
- Operational check fails with cable problem
- Operational check fails with compression sensor issue
- Operational check fails with critical failure
- Operational check fails with non-critical failure
- Operational check fails with non-therapy component issue
- Operational check fails with shock/pacing/ECG prevention issue
- Operational check non-critical failure
- Operational check passed but battery low
- Paddle PCI LEDs continuously lit
- Plastic shield and service loop routing in rear case
- Press tabs into place during assembly
- Printer malfunction
- Radio Weak Signal INOP
- Ready For Use indicator shows blinking red X with periodic chirp
- Ready For Use indicator shows solid red X with periodic chirp
- Ready For Use indicator shows solid red X without periodic chirp
- Replace front case assembly
- Replace front panel buttons
- Shock button does not respond during operational check
- SpO2 equipment malfunction
- SpO2 sensor malfunction
- Transmission failed – cannot reach server
- Transmission failed – server unknown
- Transmission failed – settings configured incorrectly
- Transmission settings not configured
- Weekly shock test fails to deliver correct energy
83 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Assigned Central is not available | Assigned IIC is not functional or powered off | HeartStart MRx cannot connect to configured IIC | error |
Bad URL Format - 12LXmit Server | Transmission settings configured incorrectly | Invalid server URL | error |
Calibration failed. Check that the pressure applied is corre | Calibration point entry failed due to incorrect pressure application | Wrong pressure applied when entering calibration point (0 or 250 mmHg) | error |
Central cannot identify this bed | IIC unable to identify HeartStart MRx due to unexpected errors | Unexpected communication or configuration error | error |
Charge Button Failure | Charge button not responding correctly | User pressed No Button Response instead of Charge during operational check | error |
Configure Failed - 12LXmit Device | Error configuring transmission device | Configuration error | error |
Confirm retries exceeded (MRx Aborted Assoc) | Confirmation timeout retries exceeded during network association | No response to confirmation requests, keeps timing out | warning |
Connect Failed - 12LXmit Dialing | Connection failed during dialing | Cannot establish connection | error |
Connection Failed | Network connection rejected by IIC | IIC rejects connection attempt from HeartStart MRx | error |
Critical Failure Detected | Critical device failure (software version prior to B.05) | Device startup critical failure | critical |
Device Error. Service Required | Device error requiring service (software version B.05 and greater) | Device startup error | critical |
Device serial number not entered | Serial number not configured | Device startup - no serial number | warning |
DNS query failure - 12LXmit Server | Server unknown - DNS resolution failed | DNS timeout or failure | error |
Fail/BF | Battery failure during Operational Check | Automated Tests detect low battery condition | warning |
Fail/C | Critical failure during Operational Check | Automated tests detect failure impacting defibrillation, pacing, or ECG acquisition | critical |
Fail/CX | Cable failure detected | Automated operational check detects cable problem | critical |
Fail/D | Non-therapy component failure detected | Automated operational check detects component failure not affecting therapy | error |
Fail/DX | Device problem preventing therapy delivery | Automated operational check detects failure affecting shock, pacing or ECG acquisition | critical |
Fail/NC | Non-critical failure during Operational Check | Automated tests detect failure not impacting life-saving functionality | warning |
Fail/S | Compression sensor failure detected | Automated operational check detects compression sensor failure | warning |
HTTP client error - 12L Transmit | Web server rejected transmission data | 12-lead transmission HTTP error | error |
IIC Association Abort received | IIC sent association abort signal to HeartStart MRx | IIC terminates established connection | warning |
IIC Life Ticks timed out (MRx Aborted Assoc) | Heartbeat timeout between HeartStart MRx and IIC | No life tick received within timeout period | warning |
Incompatible Printer | Incorrect printer configuration detected | 50mm printer installed but software expects 75mm printer | error |
Internal Memory Failure | Internal memory card error | Memory card malfunction or corruption | critical |
Modem Connection Lost - 12LXmit Network | Connection interrupted during transmission | Network connection dropped | error |
NBP Overpressure | NBP module opens relief valve when cuff pressure exceeds 300 mmHg | Cuff pressure exceeds 300 mmHg | warning |
NBP Safety Timeout | NBP module times out if pressure remains > 10 mmHg for 3 minutes | Cuff kept pressurized for more than 3 minutes | warning |
No Central assigned to this bed | No IIC configured to monitor the HeartStart MRx with assigned equipment label | Device attempts network connection with unassigned equipment label | error |
No Central – check software revision | Software revisions incompatible between HeartStart MRx and IIC | Version mismatch detected during connection attempt | error |
No Central – duplicate monitor label | Another device with same equipment label already connected to IIC | Second device with duplicate label attempts to connect | error |
No Central Monitoring | No connection to IntelliVue Central Station or network connectivity icon missing | Network connection is broken or not established | error |
No Data From Bed | IntelliVue Central Station not receiving data from HeartStart MRx | Network connection failure or radio module malfunction | error |
No Instrument Telemetry Radio detected | Radio/AC module not detected by Instrument Telemetry service screen | Service screen cannot communicate with radio module | error |
No Transmission Device Detected - Rosetta-Lt | Rosetta-Lt transmission device not detected | No transmission device found at startup | error |
Operational Check Passed | All tests completed successfully | All device components pass during Operational Check | info |
Operational Check Passed. Check Battery A (or B) | Tests pass but battery low | Battery charge low but not below operational threshold | warning |
Pacer Equip Malfunction | Pacing system failure | Pacing hardware failure detected, incorrect current or measurement | critical |
Partial transmission - 12L Transmit | 12-lead transmission incomplete | Phone disconnected during transmission | warning |
Pass | All operational checks passed | Automated operational checks complete successfully | info |
Printer Malfunction | Printer hardware or communication failure | Printer is faulty or communication problem detected | warning |
Radio Malfunction | Radio/AC module or Therapy PCA hardware failure | Radio module detects hardware failure or communication error | critical |
Radio Unplugged | Radio/AC module disconnected or RS-232 connection loose | RS-232 cable or radio module connection fails | error |
Radio Weak Signal | Wireless radio signal strength below acceptable level (no INOP tone) | RSSI value below -84 dBm | warning |
Request Timeout - 12LXmit Server | 12-lead transmission failed due to server connection timeout | Server connection has timed out | error |
Shock Button Failure | Shock button not responding correctly | User pressed No Button Response instead of Shock during operational check | error |
Shock Equip Malfunction | Defibrillator shock delivery system failure | Device unable to deliver correct energy or no energy delivered during shock | critical |
Single confirmation from IIC timed out | Confirmation timeout for single transaction from IIC | No confirmation received in expected time frame | warning |
Transmission Failed - PPP Attach Timeout | 12-lead transmission failed due to network timeout | PPP Attach Timeout on 12LXmit Network | error |
Transmission Failed - Rosetta-Lt | Rosetta-Lt transmission failed | Transmission error | error |
Transmission Timeout - Rosetta-Lt | Rosetta-Lt transmission timeout | Transmission timed out | error |
Unreachable - 12LXmit Server | Cannot reach transmission server | Server unavailable or connection lost | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.