Philips FM-2: troubleshooting and error codes

The ServiceForge knowledge base holds 7 documented symptoms for the Philips FM-2, with 10 ranked causes, 11 diagnostic checks with test points and expected values, and 36 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit does not power on or display inactive

Display completely dark or very faint; no visible screen output after power applied

Ranked causes

  1. 1. Display backlight fluorescent tube or high voltage driver failed (common)
  2. 2. Internal battery exhausted or defective (occasional)
  3. 3. Monitor power supply or internal electronics failure (occasional)
First check

Examine display in dim ambient light

Test point
Display face
Expected
Faint outline visible if backlight failed only

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor fails power-on self-test patient safety

Error screen displayed during startup; all monitoring functions disabled; error number logged in error log

Ranked causes

  1. 1. Program checksum failure during RAM load (unknown)
  2. 2. Watchdog timer failure (unknown)
First check

Press On/Off button and observe startup sequence

Test point
LCD screen and audio output
Expected
LCD illuminates blank; two-tone jingle plays; startup screen with Agilent logo, bootcode version, application version displayed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery depleted or faulty caution

Monitor battery does not hold charge, fails self-test, or does not power unit during cordless operation

Ranked causes

  1. 1. Battery depleted or not charged (common)
First check

Verify battery charge state before testing

Test point
Battery indicator or charge status display
Expected
Battery fully charged or at acceptable level per manual

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor and accessory damage patient safety

Visual inspection for damage to monitor body, cables, line cords, transducers, instruments that may affect patient safety or performance

Ranked causes

  1. 1. Physical damage from handling, impact, or wear (occasional)
First check

Visual inspection of monitor enclosure

Test point
Monitor body exterior
Expected
No cracks, dents, burns, discoloration, or physical damage

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor displays error screen and sounds tone patient safety

Display shows error condition, audible error tone sounds, and all monitoring functions stop

Ranked causes

  1. 1. Self-detectable error condition (unknown)
First check

Observe error screen display

Test point
Monitor display
Expected
Error screen displayed

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Monitor shuts down without warning on undetected error patient safety

Operational error not self-detected by monitor; unit shuts down with no audible or visual warnings or indicators

Ranked causes

  1. 1. Undetected operational error (unknown)

Monitoring display degraded or unreliable patient safety

Display shows artifacts, gaps, or poor quality; FHR(s) or TOCO values appear unreasonable after successful power-up and self-test

Ranked causes

  1. 1. Defective or worn transducer (common)
First check

Examine display after power-up and self-test

Test point
Display screen
Expected
Clear, artifact-free display with reasonable FHR and TOCO values

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1Monitor memory corruptedMemory integrity check failurecritical
10Voltage out of rangePower supply voltage outside acceptable limitserror
11Watchdog timer not functioningWatchdog timer hardware failurecritical
12System received too many resetsExcessive reset events detectedcritical
13Recorder software incompatible with monitorVersion mismatch between monitor and recordererror
14Flash memory type incompatible with softwareFlash memory type mismatcherror
15Phase lock loop lost lock (non-critical)PLL synchronization lossinfo
16Failed to allocate requested memoryMemory allocation failurecritical
17Problem with codec detectedCodec hardware or driver failureerror
18High half-word of flash failed to programFlash programming errorcritical
19Low half-word of flash failed to programFlash programming errorcritical
1xxAlgorithm errorsFetal monitoring algorithm failureerror
2Invalid internal software stateSoftware state machine failureerror
20High half-word of flash failed to verifyFlash verification errorcritical
21Low half-word of flash failed to verifyFlash verification errorcritical
2xxPrinter errorsPrinter system failureerror
3Unable to acquire software resourceQueue or semaphore allocation failureerror
30Error in communications microcontrollerCommunications MCU failureerror
3xxFatal operating system errorsOS system failurecritical
4Stack overflowStack memory exhaustedcritical
4xxFatal processor exceptionsCPU exception or faultcritical
5Bad parameter value passed to functionFunction parameter validation failureerror
50Corrupt patient record (trend file)Patient trend file integrity check failureerror
51Corrupt file systemFile system structure failureerror
52Attempt to read deleted fileRead operation on deleted fileerror
53File system wrapped memoryFile system memory wrap-arounderror
54Attempt to read corrupt fileRead operation on corrupted fileerror
55Failed to create fileFile creation failureerror
6Software in unexpected stateState machine errorerror
7Software assertion failed (coding error)Internal assertion check failureerror
70File corrupted - FM-2 viewer communication errorCommunication error during FM-2 viewer data transfererror
71Invalid patient ID - FM-2 viewer communication errorPatient ID validation failure during data transfererror
72Invalid sequence number - FM-2 viewer communication errorSequence number mismatch during data transfererror
73Remote error - FM-2 viewer communication errorError message from FM-2 viewer systemerror
8Task watchdog timer check-in failureTask did not respond to watchdog timererror
9Software CRC in memory invalidCRC verification failurecritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.