Philips EverFlo: troubleshooting and error codes
The ServiceForge knowledge base holds 49 documented symptoms for the Philips EverFlo, with 67 ranked causes, 82 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Low-pressure reading below 20 psig
Abnormal low-pressure reading measured at 5 LPM flow rate
Ranked causes
- 1. System leaks (unknown)
- 2. Tubing damage or blockage (unknown)
- 3. Fitting damage or leaks (unknown)
- 4. Dirty filters (unknown)
- 5. Compressor system pressure loss (unknown)
Measure oxygen percentage at 5 LPM
- Test point
- Output at 5 LPM
- Expected
- Normal oxygen percentage
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Impeded oxygen flow patient safety
Yellow LED illuminates continuously, Red LED blinking, audible alarm beeping periodically
Ranked causes
- 1. Canister to Compressor tubing failure (common)
- 2. Compressor failure (occasional)
- 3. Valve Solenoid failure (occasional)
- 4. Sieve Canister leaking (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low pressure patient safety
Device producing low oxygen pressure output
Ranked causes
- 1. Canister to Compressor tubing failure (common)
- 2. Compressor failure (common)
- 3. Valve Solenoid failure (occasional)
- 4. Sieve Canister leaking (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High pressure patient safety
Device producing excessive oxygen pressure
Ranked causes
- 1. Pressure relief valve Assembly failure (common)
- 2. Compressor failure (occasional)
- 3. Valve Solenoid failure (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate oxygen readings patient safety
OPI sensor displays incorrect oxygen concentration values
Ranked causes
- 1. Leaks in OPI sensor lines (common)
- 2. Main PCA failure (occasional)
- 3. Compressor failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System oxygen verification (OPI units)
Verify main PCA oxygen concentration output after 10 minute warm-up at 5.0 LPM to confirm proper oxygen purity.
Ranked causes
- 1. Oxygen concentration below 82% (red LED) (unknown)
- 2. Yellow or red LED illumination after 10 minutes (unknown)
Plug unit into power outlet
- Test point
- Power outlet
- Expected
- Unit receives 120 VAC/60 Hz
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluctuations in oxygen pressure or flow ball patient safety
Oxygen pressure or flow ball shows intermittent variations during operation
Ranked causes
- 1. Pressure Regulator failure (common)
- 2. Valve Solenoid failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low oxygen output patient safety
Device producing lower than normal oxygen concentration or flow
Ranked causes
- 1. Sieve bed failure (common)
- 2. Compressor failure (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not turn on
Device does not power up when power switch is activated
Ranked causes
- 1. No power to the unit (common)
- 2. LED failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm continuously sounding, all 3 LEDs illuminated patient safety
Audible alarm sounding continuously with all three LEDs lit
Ranked causes
- 1. System malfunction detected (unknown)
Audible alarm continuously sounding, no LEDs illuminated
Audible alarm sounding continuously with no LED indicators lit
Ranked causes
- 1. Device not plugged in or power failure (common)
Audible alarm does not sound during startup
No alarm sound for two seconds during initial power-on self-test
Ranked causes
- 1. Speaker failure or alarm circuit fault (occasional)
Verify audible alarm sounds for two seconds during startup
- Test point
- Speaker output
- Expected
- Audible alarm sounds for two seconds
Audible alarm sounding continuously, all 3 LEDs illuminated patient safety
Continuous audible alarm with all three LEDs (Green, Yellow, Red) illuminated
Ranked causes
- 1. System malfunction detected (unknown)
Audible alarm sounding continuously, no LEDs illuminated caution
Continuous audible alarm with all LEDs dark.
Ranked causes
- 2. Device turned on but not operating (common)
Calibration: Compressor flow output verification
Verify compressor output pressure and flow rate at 20 psi using test fixture with pressure gauge and flow meter.
Ranked causes
- 1. Compressor output below specification (unknown)
Verify test fixture assembly per Option 1 or 2 instructions
- Test point
- All connections and hose clamps
- Expected
- Secure fittings, no leaks
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: EverFlo system final test
Performance verification test after repairs or between patient usages to confirm concentrator operation meets specifications
Ranked causes
- 1. Routine system verification required (unknown)
Verify unit warm-up time and cabinet installation
- Test point
- System startup and cabinet assembly
- Expected
- Unit running for minimum 1 hour with front and back cabinets installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System Final Test parameters
Verification of outlet pressure range and oxygen concentration thresholds during post-repair testing
Ranked causes
- 1. Calibration reference values required for acceptance (unknown)
Outlet pressure acceptance range
- Test point
- Outlet barb with pressure gauge at 5 LPM flow
- Expected
- 5.0-7.0 psig, 5.0-7.0
- Tools
- Pressure gauge, test flow meter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System pressure verification
Procedure to verify system pressure operation of compressor and sieve canister assemblies is within normal ranges for the unit model
Ranked causes
- 1. Routine calibration and verification (common)
Model 1020000
- Test point
- System pressure
- Expected
- 15-27 psi (units without purge valves)
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Compressor requires replacement
Compressor has failed and needs to be replaced; unit may not produce oxygen or produce insufficient oxygen flow
Ranked causes
- 1. Compressor failure (unknown)
Identify compressor type before replacement
- Test point
- Thomas label on compressor
- Expected
- 0.38 Compressor shows 1098014 and 2450AE38 numbers; 0.44 Compressor shows 1020070 and 2450AE44 numbers
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Constant Audible Alarm with no LEDs caution
Alarm sounds continuously without any LED indicators illuminated
Ranked causes
- 1. Fuse Failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous red LED (pressure relief valve activated) patient safety
Red LED illuminates constantly indicating pressure relief valve activation
Ranked causes
- 1. Low pressure cycle failure (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cooling fan replacement required caution
Cooling fan needs to be removed and reinstalled, or concentrator overheats due to incorrect fan orientation
Ranked causes
- 1. Fan blade installed concave side facing away from compressor (common)
Visually inspect fan blade orientation
- Test point
- Cooling fan in front cabinet
- Expected
- Concave side of blade faces compressor
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluctuating pressure
Difference in peak pressure greater than 1 psig during operation
Ranked causes
- 1. Sieve canister degradation (unknown)
Measure oxygen percentage at 5 LPM
- Test point
- Output at 5 LPM
- Expected
- Normal oxygen percentage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green LED continuously illuminated, other LEDs off, alarm silent
Green LED on continuously, other LEDs off, audible alarm silent
Ranked causes
- 1. Device operating normally (common)
Green LED does not stay on after startup
Green LED turns off or fails to remain illuminated after initial startup sequence completes
Ranked causes
- 1. LED or driver circuit failure (occasional)
Verify green LED remains illuminated after startup
- Test point
- Green LED
- Expected
- Green LED stays on
Green LED illuminated continuously, other LEDs off, alarm silent
Green LED lights continuously, red and yellow LEDs are off, and audible alarm is silent.
Ranked causes
- 1. Device operating normally (common)
Green LED illuminates continuously, other LEDs off, audible alarm silent
Green LED continuously illuminated, Red and Yellow LEDs off, no audible alarm
Ranked causes
- 1. Device operating normally (common)
High-pressure reading above 25 psig
Abnormal high-pressure reading measured at 5 LPM flow rate at 60 Hz
Ranked causes
- 1. Improper valve cycling (unknown)
Measure oxygen percentage at 5 LPM
- Test point
- Output at 5 LPM
- Expected
- Normal oxygen percentage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate oxygen flow patient safety
Flow meter does not display correct oxygen flow rate
Ranked causes
- 1. Flow Meter failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LEDs do not light up during startup
When power switch is moved to ON position, LEDs fail to illuminate and audible alarm does not sound
Ranked causes
- 1. Power supply failure or improper connection (common)
Verify power cord connected to proper power source
- Test point
- Power outlet
- Expected
- Power cord seated and outlet functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Long Life Filter restricted
Filter may be clogged and restricting airflow through the concentrator
Ranked causes
- 1. Long-Life Filter restricted or clogged (unknown)
Power up concentrator and set flow rate
- Test point
- Flow setting
- Expected
- 5 LPM
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Long-Life Filter restriction detection
Oxygen concentration drops after filter removal, indicating filter blockage requiring replacement
Ranked causes
- 1. Long-Life Filter restricted or clogged (unknown)
Power up concentrator and set flow to 5 LPM
- Test point
- Flow control
- Expected
- 5 LPM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of power with low pressure patient safety
Device shuts down unexpectedly and produces low pressure output
Ranked causes
- 1. Compressor thermal switch failure (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Oxygen percentage indicator not functioning
OPI display does not show oxygen concentration reading
Ranked causes
- 1. Main PCA failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Outlet pressure not stabilized at startup
Outlet/regulated pressure does not stabilize after unit is powered on and runs
Ranked causes
- 1. Insufficient warm-up time or pressure regulation fault (common)
Allow unit to run for stabilization time
- Test point
- Unit operation
- Expected
- Unit running continuously for at least two minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Outlet pressure out of specification
Pressure gauge reading at outlet does not fall between 5.0 and 7.0 psig when measured at flowmeter set to 5 LPM (1 LPM for pediatric units)
Ranked causes
- 1. Regulator pressure adjustment out of specification (unknown)
Power up unit and stabilize
- Test point
- Unit operation
- Expected
- Unit runs for at least two minutes
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen output below specification patient safety
Oxygen purity reading at 5 LPM does not meet model-specific minimum requirements when measured with calibrated analyzer
Ranked causes
- 1. Pressure regulator or sieve canister assembly malfunction (unknown)
Power up concentrator and allow warm-up
- Test point
- Unit outlet
- Expected
- Running state after 15 minutes minimum
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Peak pressure variation during cycling
Measured peak pressure across four pump cycles varies by more than one psig
Ranked causes
- 1. Pump or pressure regulator out of specification (unknown)
Install pressure gauge and "T" fitting in pump discharge line
- Test point
- Sieve canister outlet
- Expected
- Gauge connected and ready to measure
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Power loss alarm does not sound patient safety
When device is disconnected from power source after warm-up, no continuous audible alarm is heard during minimum 7 minutes
Ranked causes
- 1. Main PCA failure or alarm circuit defect (common)
Connect device to proper power source
- Test point
- Power input
- Expected
- Device connected and operational
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red or yellow LED does not turn off after startup
Red and yellow LEDs remain illuminated after unit startup instead of turning off as expected
Ranked causes
- 1. Internal component failure or startup incomplete (occasional)
Verify red and yellow LEDs turn off after startup sequence
- Test point
- LED status
- Expected
- Red and yellow LED lights go off
Sieve canister assembly needs replacement
Sieve canister assembly removal and installation procedure for EverFlo oxygen concentrators with guidance on purge valve compatibility and integrated valve designs.
Ranked causes
- 1. Scheduled maintenance or canister degradation (unknown)
18 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Start-up alarms inoperable caution
Alarm system does not function during device startup
Ranked causes
- 1. Piezoelectric speaker failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Final Test after repairs
Performance verification procedure following any repairs to the EverFlo concentrator; results recorded on Testing Data Sheet
Ranked causes
- 1. Post-repair verification required (unknown)
Run-in and document unit parameters
- Test point
- Unit label and hour meter
- Expected
- Minimum 1 hour runtime; serial/model and voltage/frequency recorded
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System pressure out of normal range
Measured system pressure falls outside the documented normal operating ranges for the unit model when tested with pressure gauge connected to Pressure Regulator
Ranked causes
- 1. Compressor or Sieve Canister Assembly malfunction (unknown)
Run unit with front and rear cabinets in place
- Test point
- Unit operation
- Expected
- Unit operates normally
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not start after power switch turned on
Power switch moved to ON but unit fails to run
Ranked causes
- 1. Power not reaching unit (common)
Unit starts running after power switch moved to ON
- Test point
- Unit operation
- Expected
- Unit runs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit is inoperable
Device does not function or respond to control inputs
Ranked causes
- 1. Main PCA failure (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Valve Solenoid replacement required caution
Solenoid valve requires replacement or has failed; three manufacturer types exist (SMC air-piloted original, ASCO air-piloted, ASCO oxygen-piloted current)
Ranked causes
- 1. Valve Solenoid Assembly worn or defective (unknown)
Identify solenoid valve manufacturer and type
- Test point
- Valve body markings
- Expected
- SMC, ASCO air-piloted, or ASCO oxygen-piloted
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warm cabinet temperature caution
Cabinet or housing is excessively hot
Ranked causes
- 1. Cooling Fan failure (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Yellow LED blinking, Red LED off, audible alarm beeping periodically patient safety
Yellow LED blinking while Red LED is off and audible alarm beeps periodically
Ranked causes
- 1. High oxygen flow condition detected (unknown)