Philips EPIQ: troubleshooting and error codes
The ServiceForge knowledge base holds 10 documented symptoms for the Philips EPIQ, with 15 ranked causes, 22 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power supply LEDs indicate status
Power supply LED indicators on rear of power supply show AC, ATX, 24V, and battery status without removal
Ranked causes
- 1. AC power loss or supply failure (common)
- 2. ATX voltage rails missing (common)
- 3. 24V supply fault or system in conservation mode (occasional)
- 4. Battery system unavailable or low charge (occasional)
Observe AC present LED
- Test point
- Power supply rear panel AC present LED
- Expected
- Green
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System airflow temperature exceeds threshold caution
Monitored temperature locations in E-box exceed factory-set thresholds; system records error and shuts down
Ranked causes
- 1. Airflow blockage or fan failure (common)
- 2. High ambient temperature or insufficient ventilation (occasional)
Verify system power is off and locate temperature sensor on ACB ARGON PCB
- Test point
- ACB ARGON location on board
- Expected
- Accessible for inspection
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not boot
Control panel logic circuits or cabling prevent system power-up; testing required to isolate cause
Ranked causes
- 1. Control panel logic circuit or cable failure (common)
- 2. VPPM (power supply) fault (occasional)
Verify system power is off and access Backplane
- Test point
- System Backplane
- Expected
- System powered off, Backplane accessible
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery status indicates bad batteries
Battery status utility test shows one or both batteries failing condition test
Ranked causes
- 1. One or both batteries failed condition test (unknown)
Run battery status utility test
- Test point
- System batteries
- Expected
- Both batteries pass individual and combined tests
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Export system test results or diagnostic logs
Need to save system test results or error logs to external media for analysis
Ranked causes
- 1. Logs need to be transferred to external storage (unknown)
Verify USB media is recognized
- Test point
- USB connector
- Expected
- USB drive detected and port available in export destination list
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Field generator stand does not hold position caution
Field generator arm falls or shifts from desired position despite being positioned
Ranked causes
- 1. Spring arm tension insufficient (common)
Visually inspect field generator position stability
- Test point
- Field generator arm
- Expected
- Field generator remains in set position without drift
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Playback user control activations for troubleshooting
Replay recorded control activations to reproduce or analyze problem behavior
Ranked causes
- 1. Need to review user control sequence to identify problem (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Simulate transducer for system testing without physical transducer
Enable system operation and control verification when transducers are not connected or need to be tested offline
Ranked causes
- 1. Need to test system controls and monitor display without transducer hardware (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error displayed or suspected system problems
Operator reports system malfunction or error message appears on monitor during operation
Ranked causes
- 1. System operational status not verified (unknown)
Perform comprehensive system test
- Test point
- System Test tab
- Expected
- All subtests pass; pass-fail result displayed on monitor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test individual PCB or module hardware
Need to isolate and test specific circuit boards or modules for hardware failures
Ranked causes
- 1. Individual hardware component suspected of failure (unknown)
Locate specific hardware test
- Test point
- Hardware Tests list
- Expected
- Test list expandable with arrow; desired PCB/module visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI