Philips EasyVision RAD 4.2: troubleshooting and error codes
The ServiceForge knowledge base holds 5 documented symptoms for the Philips EasyVision RAD 4.2, with 10 ranked causes, 11 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Density verification
Verify measured density values are within acceptable tolerance of target density during calibration
Ranked causes
- 1. Printer calibration drift (occasional)
- 2. Densitometer malfunction or misalignment (occasional)
- 3. Incorrect calibration data entry or calculation (occasional)
Measure density value at grey level
- Test point
- Target density (DT)
- Expected
- Measured density (DM) within 5% of DT plus 0.02 offset
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Divergence calculation and density verification
Measured optical density values fall outside acceptable tolerance range during calibration verification
Ranked causes
- 1. Printer own calibration out of tolerance (common)
- 2. Densitometer not operating correctly (occasional)
- 3. Incorrect calibration values or calculations (occasional)
Calculate tolerance range for target density
- Test point
- Target density (DT) value
- Expected
- Range = DT ± (5% DT + 0.02)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor brightness and contrast
Monitor brightness and contrast settings require adjustment after system configuration and before clinical use. Procedure uses test images to set optimal display parameters.
Ranked causes
- 1. Factory settings not configured for optimal display (common)
- 2. Dual monitor brightness perception mismatch (occasional)
Monitor warm-up time
- Test point
- System power-on
- Expected
- At least 30 minutes elapsed with no screensave operation
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pixel size calibration setup
Configuration of hard copy pixel size calibration parameters for imaging workstation
Ranked causes
- 1. Printer selection mismatch (unknown)
Verify correct printer is selected
- Test point
- Printer selection setting
- Expected
- Same printer as used in density calibration
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test image access
Test images for printer calibration must be loaded from protected area into local database for monitor adjustment procedure.
Ranked causes
- 1. Test images not loaded in local database (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.