Philips EarlyVue VS30: troubleshooting and error codes

The ServiceForge knowledge base holds 67 documented symptoms for the Philips EarlyVue VS30, with 139 ranked causes, 79 diagnostic checks with test points and expected values, and 150 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Monitor does not power on with AC power high voltage

Device turns on with battery but not when connected to AC mains

Ranked causes

  1. 1. AC power cord unplugged (common)
  2. 2. AC power cord broken (common)
  3. 3. Fuse is blown (occasional)
  4. 4. Power supply cable malfunctioned (occasional)
  5. 5. Power supply malfunctioned (occasional)
  6. 6. AC power connector malfunctioned (occasional)
  7. 7. Main board malfunctioned (rare)
First check

Verify AC power cord connection

Test point
AC outlet
Expected
Power cord plugged in

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurements not displayed patient safety

Temperature probe readings do not appear on monitor screen

Ranked causes

  1. 1. Probe disconnected from monitor (common)
  2. 2. Probe misplaced (common)
  3. 3. Predictive temperature out of range (occasional)
  4. 4. Probe malfunctioned (occasional)
  5. 5. Front end malfunction (rare)
  6. 6. Temperature module malfunctioned (rare)
  7. 7. Main board malfunctioned (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not power on with battery patient safety

Device turns on with AC power but fails to start on battery power

Ranked causes

  1. 1. Battery not charged (common)
  2. 2. Battery or connector loose (common)
  3. 3. Battery end of life (occasional)
  4. 4. Battery interface board no connection (rare)
  5. 5. Main board malfunctioned (rare)
First check

Verify battery charge level

Test point
Battery status indicator
Expected
Battery fully charged

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP measurements not displayed patient safety

Blood pressure readings do not appear on monitor screen

Ranked causes

  1. 1. Wrong cuff size or incorrect cuff placement (common)
  2. 2. Initial pressure setting incorrect (common)
  3. 3. Air tube bent, twisted, or leaking (occasional)
  4. 4. External problem (occasional)
  5. 5. NBP module error (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call function does not signal alarm conditions patient safety

Nurse call system does not trigger or signal when alarm conditions occur

Ranked causes

  1. 1. Minimum Nurse Call Alarm Priority set too high (common)
  2. 2. Alarm audio turned off (common)
  3. 3. Phono connector loose or broken (occasional)
  4. 4. IO board malfunctioned (occasional)
  5. 5. Main board malfunctioned (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder does not print caution

Recorder does not produce output or prints are blank

Ranked causes

  1. 1. Recorder out of paper (common)
  2. 2. Recorder door is open (common)
  3. 3. System does not recognize recorder (occasional)
  4. 4. Recorder module malfunctioned (occasional)
  5. 5. Main board malfunctioned (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurements not displayed patient safety

Oxygen saturation readings do not appear on monitor screen

Ranked causes

  1. 1. SpO2 module in learning phase (common)
  2. 2. SpO2 sensor malfunctioned (common)
  3. 3. SpO2 board faulty connection to front end assembly (occasional)
  4. 4. SpO2 board malfunctioned (occasional)
  5. 5. Main board malfunctioned (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarms do not sound patient safety

Monitor does not produce alarm sounds even when alarms are triggered

Ranked causes

  1. 1. Monitor in Audio Pause or Audio Off mode (common)
  2. 2. Speaker connection faulty (occasional)
  3. 3. Speaker malfunctioned (occasional)
  4. 4. Main board malfunctioned (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 (Carbon Dioxide) patient safety

CO2 module calibration at initial use and periodic intervals; module replacement after 20,000 hours

Ranked causes

  1. 1. First calibration due at 1,200 hours of use or one year, whichever sooner (common)
  2. 2. Recurring calibration due annually or after 4,000 hours of use (common)
  3. 3. Calibration required after repairs or parts replacement (occasional)
  4. 4. CO2 module end-of-life reached at 20,000 hours (occasional)
First check

Check equipment operating hours

Test point
Usage counter
Expected
Less than 1,200 hours and less than one year since last calibration for initial; less than 4,000 hours and less than one year for recurring

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 measurements not displayed patient safety

Capnography readings do not appear on monitor screen

Ranked causes

  1. 1. CO2 module initializing (common)
  2. 2. CO2 sampling line bent or twisted (common)
  3. 3. CO2 module malfunctioned (occasional)
  4. 4. Main board malfunctioned (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP cuff does not inflate patient safety

Non-invasive blood pressure measurement cuff fails to inflate during measurement

Ranked causes

  1. 1. Air tube or cuff bent or twisted (common)
  2. 2. Air leak in air tube or cuff (common)
  3. 3. NBP module malfunctioned (occasional)
  4. 4. Main board malfunctioned (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP measurements unreliable patient safety

Blood pressure readings fluctuate or are inconsistent

Ranked causes

  1. 1. Wrong cuff size or incorrect cuff placement (common)
  2. 2. Initial pressure setting incorrect (common)
  3. 3. External problem (occasional)
  4. 4. NBP module needs calibration (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurements unreliable patient safety

Temperature readings are inconsistent or fluctuate

Ranked causes

  1. 1. Probe malfunctioned (common)
  2. 2. Module not calibrated (common)
  3. 3. Front end malfunctioned (occasional)
  4. 4. Temperature module malfunctioned (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temporal temperature measurement too low patient safety

Temporal temperature reading is below expected value

Ranked causes

  1. 1. Measurement area too moist from perspiration (common)
  2. 2. Multiple measurements in rapid succession (common)
  3. 3. Probe moved in curved motion down side of face (occasional)
  4. 4. Lens dirty (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Enclosure leakage current safety test high voltage

Annual safety verification of enclosure leakage current per IEC 60601-1 standard

Ranked causes

  1. 1. Scheduled annual safety test per IEC 60601-1 (common)
  2. 2. Safety test required after opening monitor (front and back separation) (common)
  3. 3. Safety test required after impact or damage (occasional)
First check

Measure enclosure leakage current

Test point
Equipment enclosure
Expected
Within IEC 60601-1 limit

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Ground integrity safety test high voltage

Annual verification of ground integrity per IEC 60601-1 standard

Ranked causes

  1. 1. Scheduled annual safety test per IEC 60601-1 (common)
  2. 2. Safety test required after opening monitor (common)
  3. 3. Safety test required after impact or damage (occasional)
First check

Measure ground resistance

Test point
Equipment ground point
Expected
Within IEC 60601-1 specification

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Patient leakage current safety test high voltage

Annual verification of patient leakage current per IEC 60601-1 standard

Ranked causes

  1. 1. Scheduled annual safety test per IEC 60601-1 (common)
  2. 2. Safety test required after opening monitor (common)
  3. 3. Safety test required after impact or damage (occasional)
First check

Measure patient leakage current

Test point
Patient connection points
Expected
Within IEC 60601-1 limit

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 measurements unreliable patient safety

Capnography readings are inconsistent or fluctuate

Ranked causes

  1. 1. CO2 sampling line bent or twisted (common)
  2. 2. Problem external to monitor (occasional)
  3. 3. CO2 board not calibrated (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor displays random or distorted graphics with white background

Display shows corrupted or garbled graphics with white background fill

Ranked causes

  1. 1. LCD signal cable not attached correctly (common)
  2. 2. LCD malfunctioned (occasional)
  3. 3. Main board malfunctioned (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not respond to touch screen caution

Touch screen input does not register or monitor is unresponsive to touch

Ranked causes

  1. 1. Touch screen is locked (common)
  2. 2. Touch screen not calibrated (common)
  3. 3. Touch screen malfunctioned (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor screen is blank while powered on

Device is powered on but display shows no content or is completely dark

Ranked causes

  1. 1. Monitor in Standby mode (common)
  2. 2. LCD backlight connection faulty (occasional)
  3. 3. Main board malfunctioned (rare)
First check

Verify Standby mode is not active

Test point
Display screen and status indicators
Expected
Display powered on and not in Standby

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder paper jamming caution

Paper is stuck or jamming in recorder mechanism

Ranked causes

  1. 1. Paper not installed correctly (common)
  2. 2. Wrong type of paper installed (common)
  3. 3. Recorder mechanical problem (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP initial pressure too high patient safety

Initial inflation pressure reading exceeds expected value

Ranked causes

  1. 1. Cuff too tight (common)
  2. 2. Initial inflation pressure setting incorrect (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Speaker Malfunc or AUDIO FAILED message displayed

Monitor displays speaker malfunction message or audio failed error

Ranked causes

  1. 1. One or both speakers malfunctioned (occasional)
  2. 2. Faulty connection to speaker (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurements unreliable patient safety

Oxygen saturation readings are inconsistent or fluctuate

Ranked causes

  1. 1. External problem (e.g., excessive patient movement) (common)
  2. 2. Perfusion is low (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auto Set Alarm Limits button unavailable

Auto Set Alarm Limits button cannot be selected or is grayed out

Ranked causes

  1. 1. Function only available in Interval Setting Menu for NBP, CO2, and SpO2 (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Charge Failure message displayed caution

Monitor displays battery charging failure message

Ranked causes

  1. 1. Battery overcharged or charging current too high (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Overcharged message displayed caution

Monitor shows battery overcharge warning

Ranked causes

  1. 1. Battery overcharged or charging current too high (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Overheat message displayed and shutdown patient safety

Monitor displays battery overheat warning and shuts down automatically

Ranked causes

  1. 1. Battery exceeded 60°C temperature (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 gas measurement

Check and calibrate the Microstream CO2 gas measurement system. Required after first 1,200 hours or 1 year, then every 4,000 hours or annually. Also perform when equipment is repaired or parts are replaced.

Ranked causes

  1. 1. Calibration check required per maintenance schedule (common)
First check

Verify CO2 module status

Test point
CO2 numeric pane menu
Expected
CO2 Hardware set to On

7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 module static pressure

CO2 module calibration with 5% CO2 gas concentration; calibration value display

Ranked causes

  1. 1. CO2 module requires calibration (common)
First check

CO2 module calibration status

Test point
CO2 Test menu
Expected
Last Calibration Date and Hours to Calibration reset

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NBP (Non-invasive Blood Pressure) patient safety

Periodic calibration of blood pressure measurement system to maintain accuracy

Ranked causes

  1. 1. Scheduled preventive maintenance interval reached (common)
First check

Perform NBP calibration

Test point
NBP module
Expected
Calibration completed successfully

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NBP accuracy test patient safety

Periodic performance verification of blood pressure measurement accuracy

Ranked causes

  1. 1. Scheduled performance verification interval or suspected measurement error (common)
First check

Perform NBP accuracy performance test

Test point
NBP measurement system
Expected
Measurement accuracy within specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: NBP module pressure calibration

NBP module calibration procedure with manual manometer; sets calibration points

Ranked causes

  1. 1. NBP module calibration adjustment required (occasional)
First check

NBP calibration start

Test point
NBP Test menu, NBP Calibration submenu
Expected
Monitor inflates expansion chamber; message 'Starting NBP Calibration...' appears

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: NBP static pressure accuracy

NBP accuracy test using manometer and expansion chamber at 280 mmHg and 150 mmHg

Ranked causes

  1. 1. NBP module accuracy verification needed (common)
First check

NBP static pressure test at high pressure

Test point
NBP connector with manometer and 250 ml expansion chamber
Expected
280 mmHg ± 3 mmHg, ± 3 mmHg
Tools
Reference manometer (accuracy 0.2% of reading), hand pump, valve, expansion chamber 250 ml ± 10%

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Nurse call relay

Periodic verification of nurse call relay function where used in facility protocols

Ranked causes

  1. 1. Scheduled performance verification when relay is part of facility protocols (common)
First check

Test nurse call relay function

Test point
Nurse call relay output
Expected
Relay activates and deactivates correctly per protocol

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Predictive temperature accuracy patient safety

Periodic verification of predictive temperature measurement performance

Ranked causes

  1. 1. Scheduled performance verification or suspected measurement error (common)
First check

Perform predictive temperature accuracy test

Test point
Temperature measurement system
Expected
Temperature measurement within specification tolerances

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SpO2 (Oxygen Saturation) patient safety

Periodic performance verification of pulse oximetry measurement

Ranked causes

  1. 1. Scheduled performance verification or suspected measurement error (common)
First check

Perform SpO2 performance test

Test point
Pulse oximeter module
Expected
SpO2 measurement within specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temporal Temperature patient safety

Annual calibration of temporal temperature measurement system using Exergen calibration module

Ranked causes

  1. 1. Scheduled preventive maintenance interval reached (common)
First check

Perform Temporal Temperature calibration with Exergen module

Test point
Temporal temperature sensor
Expected
Calibration completed successfully

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch screen

Touch screen requires calibration after LCD replacement; five cross-hair targets must be touched in sequence

Ranked causes

  1. 1. LCD module replaced (common)
First check

Open Diagnostics Menu

Test point
Diagnostics Menu
Expected
Menu accessible

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot access fuses high voltage

Fuse box difficult or impossible to open during replacement procedure

Ranked causes

  1. 1. Excessive force or wrong tool used on fuse box (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 module replacement reminder

CO2 module has a service life of 20,000 hours and must be replaced after this time.

Ranked causes

  1. 1. Module reaches end of service life at 20,000 hours (common)
First check

Monitor module operating hours

Test point
CO2 Test menu operating hours display
Expected
Module operating hours displayed in CO2 Test menu

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display test pattern verification

Visual inspection of LCD display using pattern sequence including vertical bars, solid colors, gray rectangles, and VCOM test pattern; each pattern displays for 5 seconds

Ranked causes

  1. 1. LCD display not functioning or defective (unknown)
First check

Observe display test pattern sequence

Test point
LCD screen
Expected
Vertical bars (black/white), vertical bars inverse, solid white, solid red, solid green, solid blue, gray rectangles, VCOM pattern; each for 5 seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High Battery Temp charge stopped message displayed caution

Monitor displays battery temperature warning and stops charging

Ranked causes

  1. 1. Battery temperature exceeds 50°C (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

LED charging and alarm indicators test

Visual verification that charging LED (right side) and alarm LED (top center) illuminate in correct color sequences and patterns

Ranked causes

  1. 1. Charging or alarm LED not functioning or defective (unknown)
First check

Verify charging LED sequence

Test point
Charging LED (right side)
Expected
Solid yellow (5 sec), flashing yellow (5 sec), solid green (5 sec), flashing green (5 sec)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor display not functioning

Monitor is powered but screen shows no image or incorrect display output

Ranked causes

  1. 1. Display module failure or video controller fault (unknown)
First check

Verify monitor display is functioning

Test point
Display screen
Expected
Image displayed, readable content

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not power on at all high voltage

Device fails to power on with either AC or battery power

Ranked causes

  1. 1. LCD, cables, battery, power supply, or main board malfunctioned (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor freezes during startup patient safety

Device powers on, opening screen and color bar start, then display freezes

Ranked causes

  1. 1. Main board malfunctioned (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor not receiving power

Monitor fails to turn on or power indicator not lit

Ranked causes

  1. 1. Power switch off, battery uncharged, or AC cord disconnected (common)
First check

Verify power switch is in ON position

Test point
Power switch
Expected
ON

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP alarm not sounding or error message not appearing

When NBP hose is connected without cuff and NBP button is pressed, no error message appears and no alarm tone sounds

Ranked causes

  1. 1. NBP alarms or module malfunction (unknown)
First check

Verify all alarms are enabled on monitor

Test point
Monitor settings
Expected
Monitor not in Audio Pause or Audio Off mode

3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

NBP error message does not appear during NBP alarm test

When NBP hose is connected without cuff and NBP button is pressed, expected error message and alarm tone do not occur.

Ranked causes

  1. 1. NBP module malfunction or alarm system disabled (unknown)
First check

Verify monitor alarm settings

Test point
Monitor alarm controls
Expected
All alarms enabled, not in Audio Pause or Audio Off mode

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call relay does not function as expected

Nurse call alarm output relay (phone jack connector) not operating correctly; tip and ring contacts not responding to alarm condition as expected

Ranked causes

  1. 1. Nurse call relay malfunction (unknown)
First check

Measure relay contact resistance on tip (normally open) during alarm condition

Test point
Phone jack connector tip
Expected
Closed

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Only one charging LED color displayed caution

Charging LED shows only one color instead of expected indicator pattern

Ranked causes

  1. 1. Membrane switch assembly malfunctioned (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient data cleared unexpectedly patient safety

All patient data is cleared when enabling Demo mode; monitor displays DEMO indicator on main screen

Ranked causes

  1. 1. Demo Mode checkbox enabled (common)
First check

Verify Demo mode status on main screen

Test point
Patient information bar
Expected
DEMO indicator in white block with red text visible when enabled; absent when disabled

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder test strip print

Functional verification of integrated recorder module (if installed); test print pattern produced by print head and motor mechanism

Ranked causes

  1. 1. Faulty recorder print head or motor (unknown)
First check

Print test strip from recorder

Test point
Recorder mechanism
Expected
Clear test pattern printed on paper; message Recorder is idle appears when complete

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Replace Battery Immediately message displayed patient safety

Monitor displays critical battery replacement message

Ranked causes

  1. 1. Battery cycle count exceeds 400 or capacity below 75% (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Replace Battery message displayed

Monitor displays Replace Battery message on screen

Ranked causes

  1. 1. Battery cycle count exceeds 300 or capacity below 80% (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Resident self test - memory and module verification

Automated memory test and front-end test of all available measurement modules (recorder, SpO2, NBP, temperature, CO2); each component test displays In Progress, then Pass/Fail/Not Supported

Ranked causes

  1. 1. Module hardware failure or memory defect (unknown)
First check

Run memory test

Test point
System RAM/ROM
Expected
Passed

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Speaker audio test

Auditory verification of speaker functionality; monitor plays three tones at increasing volume for each speaker (left and right if dual-speaker system)

Ranked causes

  1. 1. Speaker or audio amplifier defective (unknown)
First check

Listen for three tones at increasing volume

Test point
Speaker output
Expected
Three audible tones, each louder than previous; sequence repeated for each speaker

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurement out of expected range

SpO2 sensor connected and on finger, but displayed value is not between 95% and 100%

Ranked causes

  1. 1. Sensor or monitor measurement error (unknown)
First check

Connect adult SpO2 sensor to monitor

Test point
SpO2 connector on monitor
Expected
Red LED in sensor is lit

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • SpO2 value outside expected range during SpO2 sensor test
  • Temporal temperature measurement too high
  • Temporal temperature probe displays ERR
  • Test: NBP pneumatic leakage
  • Unsupported Battery or slow charge message displayed
  • Upgrade Masimo firmware
  • View Masimo firmware version

7 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1025System Error (1025)Could not load Resource fileerror
1026System Error (1026)Could not load data from flasherror
1027System Error (1027)Could not load system configuration dataerror
1028System Error (1028)Could not load profileerror
1031System Error (1031)Monitoring and watchdog thread could not initializeerror
1032System ErrorCould not open audio fileserror
1043System ErrorThread blockederror
1044System ErrorFront end software error detectederror
1045System ErrorPrinting software error detectederror
1047System ErrorMonitoring software error detectederror
1048System ErrorException detected in alarming softwareerror
1051Cannot initialize NBP moduleNBP module fails to initialize during startup or diagnostic testerror
1053Loss of monitoring signalMonitoring signal lost or system reset occurredinfo
1055Cannot initialize LCD displayLCD display fails to initialize during startup or diagnostic testerror
257System ErrorBattery charger power failureerror
261System ErrorBattery is overchargederror
264System ErrorBattery is over temperatureerror
266System ErrorUnrecognized battery detectederror
267 (1)System ErrorBattery cycle count >300 or capacity ≤80% of design capacitywarning
267 (2)System ErrorBattery cycle count >400 or capacity ≤75% of design capacityerror
401System ErrorCould not read main board EEPROMerror
402System ErrorCould not write main board EEPROMerror
403System ErrorMain board EEPROM checksum errorerror
818 (101)NBP Equipment MalfunctionTask overrun detectederror
818 (102)NBP Equipment MalfunctionInterrupt stack overwrittencritical
818 (103)NBP Equipment MalfunctionUser stack overwrittencritical
818 (104)NBP Equipment MalfunctionData validity check failederror
818 (106)NBP Equipment MalfunctionUnable to write init flag to non-volatile memoryerror
818 (107)NBP Equipment MalfunctionUnable to verify init flag written to non-volatile memoryerror
818 (109)NBP Equipment MalfunctionUnable to write NBP configuration data to non-volatile memoryerror
818 (110)NBP Equipment MalfunctionData validity issue: environmental data verification failederror
818 (111)NBP Equipment MalfunctionCalibration data checksum errorerror
818 (112)NBP Equipment MalfunctionUnable to write calibration data to non-volatile memoryerror
818 (113)NBP Equipment MalfunctionUnable to verify calibration data in non-volatile memoryerror
818 (114)NBP Equipment MalfunctionError log checksum errorerror
818 (115)NBP Equipment MalfunctionUnable to write error log to non-volatile memorywarning
818 (116)NBP Equipment MalfunctionUnable to verify error log in non-volatile memorywarning
818 (117)NBP Equipment MalfunctionInflation cycle count and runtime data checksum errorwarning
818 (118)NBP Equipment MalfunctionUnable to write inflation cycle and runtime datawarning
818 (119)NBP Equipment MalfunctionUnable to verify inflation cycle and runtime datawarning
818 (120)NBP Equipment MalfunctionNew message sent while previous message incompleteerror
818 (121)NBP Equipment MalfunctionPressure value from second pressure sensor not availableerror
818 (122)NBP Equipment Malfunction5-volt supply voltage too high or lowerror
818 (123)NBP Equipment MalfunctionError from the 5-volt regulatorerror
818 (124)NBP Equipment Malfunction12-volt supply voltage too high or lowerror
818 (128)NBP Equipment MalfunctionOffset of pressure sensor 1 out of rangeerror
818 (129)NBP Equipment MalfunctionOffset of pressure sensor 2 out of rangeerror
818 (130)NBP Equipment MalfunctionResponse of pressure sensor 2 to test signal out of rangeerror
818 (131)NBP Equipment MalfunctionDifference between two pressure channels too largeerror
818 (132)NBP Equipment MalfunctionOne of the two valves activated for too longerror
818 (133)NBP Equipment Malfunction - CRC mismatch in calibration dataCRC error detected in calibration data on startuperror
818 (134)NBP Equipment Malfunction - CRC mismatch in error log dataCRC error detected in error log data on startuperror
818 (135)NBP Equipment Malfunction - CRC mismatch in inflation cycle and runtime dataCRC error in inflation cycle count and runtime dataerror
818 (136)NBP Equipment Malfunction - Pneumatic safety switch malfunctionPneumatic safety switch does not respond correctlyerror
818 (137)NBP Equipment Malfunction - Valve 2 defective or blockedValve 2 fails to open or close properlyerror
818 (141)NBP Equipment Malfunction - Inflation time exceededInflation time for selected patient type exceeds normal durationwarning
818 (199)NBP Equipment Malfunction - No data from NBP blockNo communication with NBP block moduleerror
818 (25)NBP Equipment MalfunctionNBP hardware malfunction due to no dataerror
819NBP Equipment Malfunction - Not CalibratedNBP module has not been calibratederror
819 (127)NBP Equipment Malfunction - NBP module not calibratedNBP module calibration not performed or invaliderror
820 (138)NBP Hose BlockedBlockage or pressure drop detected in pneumatic systemerror
821 (139)NBP Air LeakAir leak or pressure loss detected in pneumatic systemerror
833 (1)SpO2 Equipment Malfunction - ROM checksum errorSpO2 board ROM fails checksum verification on startuperror
833 (10)SpO2 Equipment Malfunction - Host communication errorSpO2 board fails to communicate with main processorerror
833 (100)SpO2 Equipment Malfunction - Masimo board not enabledMasimo SpO2 board is installed but not enabled in firmwareerror
833 (101)SpO2 Equipment Malfunction - Masimo DSP Raw Queue OverrunMasimo DSP raw data queue overflows during operationerror
833 (102)SpO2 Equipment Malfunction - Masimo DSP hardware failureMasimo DSP processor fails initialization or self-testerror
833 (103)SpO2 Equipment Malfunction - Masimo MCU failureMasimo microcontroller unit fails initialization or testerror
833 (104)SpO2 Equipment Malfunction - Masimo watchdog failureMasimo watchdog timer detects processor hang or failurecritical
833 (105)SpO2 Equipment Malfunction - Masimo invalid board typeMasimo board type identifier does not match expected valueerror
833 (106)SpO2 Equipment Malfunction - Masimo invalid master control stateMasimo master control state machine detects invalid stateerror
833 (107)SpO2 Equipment Malfunction - Masimo SRAM transfer failureMasimo SRAM memory transfer fails during initializationerror
833 (108)SpO2 Equipment Malfunction - Masimo SRAM task queue overrunMasimo SRAM task queue overflows during operationerror
833 (109)SpO2 Equipment Malfunction - Masimo database failureMasimo SpO2 board database initialization or access failserror
833 (110)SpO2 Equipment Malfunction - Masimo invalid flash memory deviceMasimo flash memory device not recognized or fails testerror
833 (2)SpO2 Equipment Malfunction - RAM test errorSpO2 board RAM fails self-test on startuperror
833 (20)SpO2 Equipment Malfunction - Hardware defectSpO2 board detects internal hardware faulterror
833 (4)SpO2 Equipment Malfunction - Analog signal path self test failedSpO2 analog signal path fails self-testerror
833 (40)SpO2 Equipment Malfunction - Non-volatile memory corruptedSpO2 board non-volatile memory checksum or integrity failureerror
833 (6)SpO2 Equipment Malfunction - Self test timeoutSpO2 board self-test does not complete in expected timeerror
833 (8)SpO2 Equipment Malfunction - Internal DSP communication errorCommunication fails between SpO2 board and DSPerror
834SpO2 Sensor MalfunctionSpO2 sensor does not respond or connection losterror
836SpO2 Communication ErrorCommunication lost with SpO2 module during operationinfo
837SpO2 Error - Front end will resetNo data received from SpO2 moduleerror
838 (100)SpHb Equipment Malfunction - not enabledSpHb module installed but not enablederror
839 (100)RRa Equipment Malfunction - not enabledRRa module installed but not enablederror
850CO2 Error - Front end will resetNo data received from CO2 moduleerror
851CO2 Equipment Malfunction - Communication errorCO2 module communication failserror
851 (1)CO2 Equipment Malfunction - Calibration errorCO2 module calibration verification failserror
851 (16)CO2 Equipment Malfunction - Main board errorCO2 module main board fault detectederror
851 (17)CO2 Equipment Malfunction - Module malfunctionGeneral CO2 module malfunction detectederror
851 (18)CO2 Equipment Malfunction - Scrubber pump errorCO2 scrubber or pump mechanism failserror
851 (19)CO2 Equipment Malfunction - Sensor malfunctionCO2 sensor fails to respond or produce valid dataerror
851 (2)CO2 Equipment Malfunction - Sampling line malfunctionCO2 sampling line blockage or disconnection detectedwarning
851 (20)CO2 Equipment Malfunction - 15V voltage out of range15 V supply voltage to CO2 module out of specificationerror
851 (3)CO2 Equipment Malfunction - Occlusion in input lineBlockage detected in CO2 input linewarning
866 (21)Temp Probe Error - Operating above 43.3°CTemperature probe operating in conditions above 43.3°C (112°F)error
866 (22)Temp Probe Error - Excessive heater energyExcessive heater energy detected or slow temperature responseerror
866 (32)Temp Probe Error - Thermistor pulled away from tipThermistor separation or heater failure detectederror
866 (33)Temp Probe Error - Probe not responsiveProbe does not respond to commands or not calibratederror
866 (34)Temperature probe not characterized or calibratedProbe initialization or measurement attemptwarning
866 (45)Temperature measurement below allowable rangeMeasurement reading significantly below low ambient limitswarning
866 (46)Temperature measurement above allowable rangeMeasurement reading significantly above high ambient or patient limitswarning
866 (52)Temperature probe not characterized or calibratedProbe initialization or characterization checkwarning
866 (63)Temperature probe well missing or improperly installedProbe well detection during initializationerror
866 (70)Error reading probe EEPROMProbe EEPROM read operation during startup or measurementerror
866 (75)Probe EEPROM read error or probe not factory testedProbe EEPROM read or factory test validationerror
866 (78)Probe non-volatile error detection mechanism triggeredProbe error detection during operationerror
866 (79)Temperature module non-volatile memory error detectedModule error detection mechanism during operationerror
866 (80)Probe heater stuck onProbe heater turnoff command but heater still onerror
866 (81)Probe heater stuck offProbe heater turnon command but heater still offerror
867 (101)Temperature Module Malfunction (101)Checksum error in moduleerror
867 (102)Temperature Module Malfunction (102)Calibration error in moduleerror
867 (103)Temperature Module Malfunction (103)Ambient temperature below minimum operating limitwarning
867 (104)Temperature Module Malfunction (104)Ambient temperature above maximum operating limitwarning
867 (201)Temporal Temperature Module Malfunction (201)Temporal temperature module checksum errorerror
867 (202)Temporal Temperature Module Malfunction (202)Internal firmware error in temporal moduleerror
867 (203)Temporal Temperature Module Malfunction (203)Temporal module below allowable ambient temperaturewarning
867 (204)Temporal Temperature Module Malfunction (204)Temporal module above allowable ambient temperaturewarning
867 (24)Ambient temperature too highAmbient temperature measurement exceeds safe operating rangewarning
867 (25)Ambient temperature too lowAmbient temperature measurement below safe operating rangewarning
867 (27)Battery or power supply voltage exceeds maximumSupply voltage monitoring detects overvoltage conditionerror
867 (28)Battery or power supply voltage below minimumSupply voltage monitoring detects undervoltage conditionerror
867 (47)Internal calibration resistor damagedCalibration resistor validation during operationerror
867 (48)Internal calibration resistor damagedCalibration resistor validation during operationerror
867 (49)Circuit validation resistor damagedCircuit validation resistor check during operationerror
867 (50)Circuit validation resistor damagedCircuit validation resistor check during operationerror
867 (51)A/D measurement timeoutAnalog to digital conversion does not complete in timeerror
867 (59)Battery or power supply voltage below maximumSupply voltage monitoring detects low voltage conditionerror
867 (60)Battery or power supply voltage exceeds minimumSupply voltage monitoring detects high voltage conditionerror
867 (61)Reference voltage circuit under voltage or unstableReference voltage regulation monitoring during operationerror
867 (62)Temperature module not calibratedModule calibration check during startuperror
867 (65)Error saving to module EEPROMEEPROM write operation during calibration or operationerror
867 (66)Module non-volatile calibration memory error detectedCalibration memory error detection mechanism during operationerror
867 (67)Error reading or writing module EEPROMEEPROM read/write operation during operationerror
867 (68)Error reading or writing module EEPROMEEPROM read/write operation during operationerror
867 (69)Error reading or writing module EEPROMEEPROM read/write operation during operationerror
867 (74)Temperature Module Malfunction (74)Internal error during EEPROM initializationerror
867 (75)Temperature Module Malfunction (75)Cannot read probe EEPROM correctly or probe untestederror
867 (82)Temperature Module Malfunction (82)Signal HTR_Q on, HTRC off, but voltage presenterror
867 (83)Temperature Module Malfunction (83)Signal HTR_Q tri-stated with HTRC enabled and heater power presenterror
867 (84)Temperature Module Malfunction (84)Q&C on but heater voltage insufficienterror
867 (85)Temperature Module Malfunction (85)Heater fail-safe did not turn off as expectederror
868Temperature ErrorNo data from temperature moduleerror
881Speaker MalfunctionSpeaker failurewarning
882Wireless Module Malfunction (882)Wireless module malfunctionerror
883System Error (883)Internal state error for power management detectederror
887System Error (887)Backup image used to boot due to latest software image failureerror
888System ErrorSystem could not load font fileserror
889System Error (889)System failed to mount one of the file systemserror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.