Philips EarlyVue VS30: troubleshooting and error codes
The ServiceForge knowledge base holds 67 documented symptoms for the Philips EarlyVue VS30, with 139 ranked causes, 79 diagnostic checks with test points and expected values, and 150 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Monitor does not power on with AC power high voltage
Device turns on with battery but not when connected to AC mains
Ranked causes
- 1. AC power cord unplugged (common)
- 2. AC power cord broken (common)
- 3. Fuse is blown (occasional)
- 4. Power supply cable malfunctioned (occasional)
- 5. Power supply malfunctioned (occasional)
- 6. AC power connector malfunctioned (occasional)
- 7. Main board malfunctioned (rare)
Verify AC power cord connection
- Test point
- AC outlet
- Expected
- Power cord plugged in
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements not displayed patient safety
Temperature probe readings do not appear on monitor screen
Ranked causes
- 1. Probe disconnected from monitor (common)
- 2. Probe misplaced (common)
- 3. Predictive temperature out of range (occasional)
- 4. Probe malfunctioned (occasional)
- 5. Front end malfunction (rare)
- 6. Temperature module malfunctioned (rare)
- 7. Main board malfunctioned (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on with battery patient safety
Device turns on with AC power but fails to start on battery power
Ranked causes
- 1. Battery not charged (common)
- 2. Battery or connector loose (common)
- 3. Battery end of life (occasional)
- 4. Battery interface board no connection (rare)
- 5. Main board malfunctioned (rare)
Verify battery charge level
- Test point
- Battery status indicator
- Expected
- Battery fully charged
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements not displayed patient safety
Blood pressure readings do not appear on monitor screen
Ranked causes
- 1. Wrong cuff size or incorrect cuff placement (common)
- 2. Initial pressure setting incorrect (common)
- 3. Air tube bent, twisted, or leaking (occasional)
- 4. External problem (occasional)
- 5. NBP module error (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call function does not signal alarm conditions patient safety
Nurse call system does not trigger or signal when alarm conditions occur
Ranked causes
- 1. Minimum Nurse Call Alarm Priority set too high (common)
- 2. Alarm audio turned off (common)
- 3. Phono connector loose or broken (occasional)
- 4. IO board malfunctioned (occasional)
- 5. Main board malfunctioned (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not print caution
Recorder does not produce output or prints are blank
Ranked causes
- 1. Recorder out of paper (common)
- 2. Recorder door is open (common)
- 3. System does not recognize recorder (occasional)
- 4. Recorder module malfunctioned (occasional)
- 5. Main board malfunctioned (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements not displayed patient safety
Oxygen saturation readings do not appear on monitor screen
Ranked causes
- 1. SpO2 module in learning phase (common)
- 2. SpO2 sensor malfunctioned (common)
- 3. SpO2 board faulty connection to front end assembly (occasional)
- 4. SpO2 board malfunctioned (occasional)
- 5. Main board malfunctioned (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not sound patient safety
Monitor does not produce alarm sounds even when alarms are triggered
Ranked causes
- 1. Monitor in Audio Pause or Audio Off mode (common)
- 2. Speaker connection faulty (occasional)
- 3. Speaker malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 (Carbon Dioxide) patient safety
CO2 module calibration at initial use and periodic intervals; module replacement after 20,000 hours
Ranked causes
- 1. First calibration due at 1,200 hours of use or one year, whichever sooner (common)
- 2. Recurring calibration due annually or after 4,000 hours of use (common)
- 3. Calibration required after repairs or parts replacement (occasional)
- 4. CO2 module end-of-life reached at 20,000 hours (occasional)
Check equipment operating hours
- Test point
- Usage counter
- Expected
- Less than 1,200 hours and less than one year since last calibration for initial; less than 4,000 hours and less than one year for recurring
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 measurements not displayed patient safety
Capnography readings do not appear on monitor screen
Ranked causes
- 1. CO2 module initializing (common)
- 2. CO2 sampling line bent or twisted (common)
- 3. CO2 module malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP cuff does not inflate patient safety
Non-invasive blood pressure measurement cuff fails to inflate during measurement
Ranked causes
- 1. Air tube or cuff bent or twisted (common)
- 2. Air leak in air tube or cuff (common)
- 3. NBP module malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements unreliable patient safety
Blood pressure readings fluctuate or are inconsistent
Ranked causes
- 1. Wrong cuff size or incorrect cuff placement (common)
- 2. Initial pressure setting incorrect (common)
- 3. External problem (occasional)
- 4. NBP module needs calibration (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements unreliable patient safety
Temperature readings are inconsistent or fluctuate
Ranked causes
- 1. Probe malfunctioned (common)
- 2. Module not calibrated (common)
- 3. Front end malfunctioned (occasional)
- 4. Temperature module malfunctioned (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temporal temperature measurement too low patient safety
Temporal temperature reading is below expected value
Ranked causes
- 1. Measurement area too moist from perspiration (common)
- 2. Multiple measurements in rapid succession (common)
- 3. Probe moved in curved motion down side of face (occasional)
- 4. Lens dirty (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Enclosure leakage current safety test high voltage
Annual safety verification of enclosure leakage current per IEC 60601-1 standard
Ranked causes
- 1. Scheduled annual safety test per IEC 60601-1 (common)
- 2. Safety test required after opening monitor (front and back separation) (common)
- 3. Safety test required after impact or damage (occasional)
Measure enclosure leakage current
- Test point
- Equipment enclosure
- Expected
- Within IEC 60601-1 limit
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ground integrity safety test high voltage
Annual verification of ground integrity per IEC 60601-1 standard
Ranked causes
- 1. Scheduled annual safety test per IEC 60601-1 (common)
- 2. Safety test required after opening monitor (common)
- 3. Safety test required after impact or damage (occasional)
Measure ground resistance
- Test point
- Equipment ground point
- Expected
- Within IEC 60601-1 specification
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Patient leakage current safety test high voltage
Annual verification of patient leakage current per IEC 60601-1 standard
Ranked causes
- 1. Scheduled annual safety test per IEC 60601-1 (common)
- 2. Safety test required after opening monitor (common)
- 3. Safety test required after impact or damage (occasional)
Measure patient leakage current
- Test point
- Patient connection points
- Expected
- Within IEC 60601-1 limit
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 measurements unreliable patient safety
Capnography readings are inconsistent or fluctuate
Ranked causes
- 1. CO2 sampling line bent or twisted (common)
- 2. Problem external to monitor (occasional)
- 3. CO2 board not calibrated (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays random or distorted graphics with white background
Display shows corrupted or garbled graphics with white background fill
Ranked causes
- 1. LCD signal cable not attached correctly (common)
- 2. LCD malfunctioned (occasional)
- 3. Main board malfunctioned (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to touch screen caution
Touch screen input does not register or monitor is unresponsive to touch
Ranked causes
- 1. Touch screen is locked (common)
- 2. Touch screen not calibrated (common)
- 3. Touch screen malfunctioned (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor screen is blank while powered on
Device is powered on but display shows no content or is completely dark
Ranked causes
- 1. Monitor in Standby mode (common)
- 2. LCD backlight connection faulty (occasional)
- 3. Main board malfunctioned (rare)
Verify Standby mode is not active
- Test point
- Display screen and status indicators
- Expected
- Display powered on and not in Standby
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper jamming caution
Paper is stuck or jamming in recorder mechanism
Ranked causes
- 1. Paper not installed correctly (common)
- 2. Wrong type of paper installed (common)
- 3. Recorder mechanical problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP initial pressure too high patient safety
Initial inflation pressure reading exceeds expected value
Ranked causes
- 1. Cuff too tight (common)
- 2. Initial inflation pressure setting incorrect (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker Malfunc or AUDIO FAILED message displayed
Monitor displays speaker malfunction message or audio failed error
Ranked causes
- 1. One or both speakers malfunctioned (occasional)
- 2. Faulty connection to speaker (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements unreliable patient safety
Oxygen saturation readings are inconsistent or fluctuate
Ranked causes
- 1. External problem (e.g., excessive patient movement) (common)
- 2. Perfusion is low (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto Set Alarm Limits button unavailable
Auto Set Alarm Limits button cannot be selected or is grayed out
Ranked causes
- 1. Function only available in Interval Setting Menu for NBP, CO2, and SpO2 (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Charge Failure message displayed caution
Monitor displays battery charging failure message
Ranked causes
- 1. Battery overcharged or charging current too high (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Overcharged message displayed caution
Monitor shows battery overcharge warning
Ranked causes
- 1. Battery overcharged or charging current too high (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Overheat message displayed and shutdown patient safety
Monitor displays battery overheat warning and shuts down automatically
Ranked causes
- 1. Battery exceeded 60°C temperature (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 gas measurement
Check and calibrate the Microstream CO2 gas measurement system. Required after first 1,200 hours or 1 year, then every 4,000 hours or annually. Also perform when equipment is repaired or parts are replaced.
Ranked causes
- 1. Calibration check required per maintenance schedule (common)
Verify CO2 module status
- Test point
- CO2 numeric pane menu
- Expected
- CO2 Hardware set to On
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 module static pressure
CO2 module calibration with 5% CO2 gas concentration; calibration value display
Ranked causes
- 1. CO2 module requires calibration (common)
CO2 module calibration status
- Test point
- CO2 Test menu
- Expected
- Last Calibration Date and Hours to Calibration reset
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP (Non-invasive Blood Pressure) patient safety
Periodic calibration of blood pressure measurement system to maintain accuracy
Ranked causes
- 1. Scheduled preventive maintenance interval reached (common)
Perform NBP calibration
- Test point
- NBP module
- Expected
- Calibration completed successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP accuracy test patient safety
Periodic performance verification of blood pressure measurement accuracy
Ranked causes
- 1. Scheduled performance verification interval or suspected measurement error (common)
Perform NBP accuracy performance test
- Test point
- NBP measurement system
- Expected
- Measurement accuracy within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP module pressure calibration
NBP module calibration procedure with manual manometer; sets calibration points
Ranked causes
- 1. NBP module calibration adjustment required (occasional)
NBP calibration start
- Test point
- NBP Test menu, NBP Calibration submenu
- Expected
- Monitor inflates expansion chamber; message 'Starting NBP Calibration...' appears
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP static pressure accuracy
NBP accuracy test using manometer and expansion chamber at 280 mmHg and 150 mmHg
Ranked causes
- 1. NBP module accuracy verification needed (common)
NBP static pressure test at high pressure
- Test point
- NBP connector with manometer and 250 ml expansion chamber
- Expected
- 280 mmHg ± 3 mmHg, ± 3 mmHg
- Tools
- Reference manometer (accuracy 0.2% of reading), hand pump, valve, expansion chamber 250 ml ± 10%
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse call relay
Periodic verification of nurse call relay function where used in facility protocols
Ranked causes
- 1. Scheduled performance verification when relay is part of facility protocols (common)
Test nurse call relay function
- Test point
- Nurse call relay output
- Expected
- Relay activates and deactivates correctly per protocol
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Predictive temperature accuracy patient safety
Periodic verification of predictive temperature measurement performance
Ranked causes
- 1. Scheduled performance verification or suspected measurement error (common)
Perform predictive temperature accuracy test
- Test point
- Temperature measurement system
- Expected
- Temperature measurement within specification tolerances
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SpO2 (Oxygen Saturation) patient safety
Periodic performance verification of pulse oximetry measurement
Ranked causes
- 1. Scheduled performance verification or suspected measurement error (common)
Perform SpO2 performance test
- Test point
- Pulse oximeter module
- Expected
- SpO2 measurement within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temporal Temperature patient safety
Annual calibration of temporal temperature measurement system using Exergen calibration module
Ranked causes
- 1. Scheduled preventive maintenance interval reached (common)
Perform Temporal Temperature calibration with Exergen module
- Test point
- Temporal temperature sensor
- Expected
- Calibration completed successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen
Touch screen requires calibration after LCD replacement; five cross-hair targets must be touched in sequence
Ranked causes
- 1. LCD module replaced (common)
Open Diagnostics Menu
- Test point
- Diagnostics Menu
- Expected
- Menu accessible
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot access fuses high voltage
Fuse box difficult or impossible to open during replacement procedure
Ranked causes
- 1. Excessive force or wrong tool used on fuse box (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 module replacement reminder
CO2 module has a service life of 20,000 hours and must be replaced after this time.
Ranked causes
- 1. Module reaches end of service life at 20,000 hours (common)
Monitor module operating hours
- Test point
- CO2 Test menu operating hours display
- Expected
- Module operating hours displayed in CO2 Test menu
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display test pattern verification
Visual inspection of LCD display using pattern sequence including vertical bars, solid colors, gray rectangles, and VCOM test pattern; each pattern displays for 5 seconds
Ranked causes
- 1. LCD display not functioning or defective (unknown)
Observe display test pattern sequence
- Test point
- LCD screen
- Expected
- Vertical bars (black/white), vertical bars inverse, solid white, solid red, solid green, solid blue, gray rectangles, VCOM pattern; each for 5 seconds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High Battery Temp charge stopped message displayed caution
Monitor displays battery temperature warning and stops charging
Ranked causes
- 1. Battery temperature exceeds 50°C (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED charging and alarm indicators test
Visual verification that charging LED (right side) and alarm LED (top center) illuminate in correct color sequences and patterns
Ranked causes
- 1. Charging or alarm LED not functioning or defective (unknown)
Verify charging LED sequence
- Test point
- Charging LED (right side)
- Expected
- Solid yellow (5 sec), flashing yellow (5 sec), solid green (5 sec), flashing green (5 sec)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display not functioning
Monitor is powered but screen shows no image or incorrect display output
Ranked causes
- 1. Display module failure or video controller fault (unknown)
Verify monitor display is functioning
- Test point
- Display screen
- Expected
- Image displayed, readable content
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on at all high voltage
Device fails to power on with either AC or battery power
Ranked causes
- 1. LCD, cables, battery, power supply, or main board malfunctioned (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor freezes during startup patient safety
Device powers on, opening screen and color bar start, then display freezes
Ranked causes
- 1. Main board malfunctioned (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor not receiving power
Monitor fails to turn on or power indicator not lit
Ranked causes
- 1. Power switch off, battery uncharged, or AC cord disconnected (common)
Verify power switch is in ON position
- Test point
- Power switch
- Expected
- ON
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP alarm not sounding or error message not appearing
When NBP hose is connected without cuff and NBP button is pressed, no error message appears and no alarm tone sounds
Ranked causes
- 1. NBP alarms or module malfunction (unknown)
Verify all alarms are enabled on monitor
- Test point
- Monitor settings
- Expected
- Monitor not in Audio Pause or Audio Off mode
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
NBP error message does not appear during NBP alarm test
When NBP hose is connected without cuff and NBP button is pressed, expected error message and alarm tone do not occur.
Ranked causes
- 1. NBP module malfunction or alarm system disabled (unknown)
Verify monitor alarm settings
- Test point
- Monitor alarm controls
- Expected
- All alarms enabled, not in Audio Pause or Audio Off mode
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call relay does not function as expected
Nurse call alarm output relay (phone jack connector) not operating correctly; tip and ring contacts not responding to alarm condition as expected
Ranked causes
- 1. Nurse call relay malfunction (unknown)
Measure relay contact resistance on tip (normally open) during alarm condition
- Test point
- Phone jack connector tip
- Expected
- Closed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Only one charging LED color displayed caution
Charging LED shows only one color instead of expected indicator pattern
Ranked causes
- 1. Membrane switch assembly malfunctioned (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient data cleared unexpectedly patient safety
All patient data is cleared when enabling Demo mode; monitor displays DEMO indicator on main screen
Ranked causes
- 1. Demo Mode checkbox enabled (common)
Verify Demo mode status on main screen
- Test point
- Patient information bar
- Expected
- DEMO indicator in white block with red text visible when enabled; absent when disabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder test strip print
Functional verification of integrated recorder module (if installed); test print pattern produced by print head and motor mechanism
Ranked causes
- 1. Faulty recorder print head or motor (unknown)
Print test strip from recorder
- Test point
- Recorder mechanism
- Expected
- Clear test pattern printed on paper; message Recorder is idle appears when complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace Battery Immediately message displayed patient safety
Monitor displays critical battery replacement message
Ranked causes
- 1. Battery cycle count exceeds 400 or capacity below 75% (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace Battery message displayed
Monitor displays Replace Battery message on screen
Ranked causes
- 1. Battery cycle count exceeds 300 or capacity below 80% (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resident self test - memory and module verification
Automated memory test and front-end test of all available measurement modules (recorder, SpO2, NBP, temperature, CO2); each component test displays In Progress, then Pass/Fail/Not Supported
Ranked causes
- 1. Module hardware failure or memory defect (unknown)
Run memory test
- Test point
- System RAM/ROM
- Expected
- Passed
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker audio test
Auditory verification of speaker functionality; monitor plays three tones at increasing volume for each speaker (left and right if dual-speaker system)
Ranked causes
- 1. Speaker or audio amplifier defective (unknown)
Listen for three tones at increasing volume
- Test point
- Speaker output
- Expected
- Three audible tones, each louder than previous; sequence repeated for each speaker
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurement out of expected range
SpO2 sensor connected and on finger, but displayed value is not between 95% and 100%
Ranked causes
- 1. Sensor or monitor measurement error (unknown)
Connect adult SpO2 sensor to monitor
- Test point
- SpO2 connector on monitor
- Expected
- Red LED in sensor is lit
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- SpO2 value outside expected range during SpO2 sensor test
- Temporal temperature measurement too high
- Temporal temperature probe displays ERR
- Test: NBP pneumatic leakage
- Unsupported Battery or slow charge message displayed
- Upgrade Masimo firmware
- View Masimo firmware version
7 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1025 | System Error (1025) | Could not load Resource file | error |
1026 | System Error (1026) | Could not load data from flash | error |
1027 | System Error (1027) | Could not load system configuration data | error |
1028 | System Error (1028) | Could not load profile | error |
1031 | System Error (1031) | Monitoring and watchdog thread could not initialize | error |
1032 | System Error | Could not open audio files | error |
1043 | System Error | Thread blocked | error |
1044 | System Error | Front end software error detected | error |
1045 | System Error | Printing software error detected | error |
1047 | System Error | Monitoring software error detected | error |
1048 | System Error | Exception detected in alarming software | error |
1051 | Cannot initialize NBP module | NBP module fails to initialize during startup or diagnostic test | error |
1053 | Loss of monitoring signal | Monitoring signal lost or system reset occurred | info |
1055 | Cannot initialize LCD display | LCD display fails to initialize during startup or diagnostic test | error |
257 | System Error | Battery charger power failure | error |
261 | System Error | Battery is overcharged | error |
264 | System Error | Battery is over temperature | error |
266 | System Error | Unrecognized battery detected | error |
267 (1) | System Error | Battery cycle count >300 or capacity ≤80% of design capacity | warning |
267 (2) | System Error | Battery cycle count >400 or capacity ≤75% of design capacity | error |
401 | System Error | Could not read main board EEPROM | error |
402 | System Error | Could not write main board EEPROM | error |
403 | System Error | Main board EEPROM checksum error | error |
818 (101) | NBP Equipment Malfunction | Task overrun detected | error |
818 (102) | NBP Equipment Malfunction | Interrupt stack overwritten | critical |
818 (103) | NBP Equipment Malfunction | User stack overwritten | critical |
818 (104) | NBP Equipment Malfunction | Data validity check failed | error |
818 (106) | NBP Equipment Malfunction | Unable to write init flag to non-volatile memory | error |
818 (107) | NBP Equipment Malfunction | Unable to verify init flag written to non-volatile memory | error |
818 (109) | NBP Equipment Malfunction | Unable to write NBP configuration data to non-volatile memory | error |
818 (110) | NBP Equipment Malfunction | Data validity issue: environmental data verification failed | error |
818 (111) | NBP Equipment Malfunction | Calibration data checksum error | error |
818 (112) | NBP Equipment Malfunction | Unable to write calibration data to non-volatile memory | error |
818 (113) | NBP Equipment Malfunction | Unable to verify calibration data in non-volatile memory | error |
818 (114) | NBP Equipment Malfunction | Error log checksum error | error |
818 (115) | NBP Equipment Malfunction | Unable to write error log to non-volatile memory | warning |
818 (116) | NBP Equipment Malfunction | Unable to verify error log in non-volatile memory | warning |
818 (117) | NBP Equipment Malfunction | Inflation cycle count and runtime data checksum error | warning |
818 (118) | NBP Equipment Malfunction | Unable to write inflation cycle and runtime data | warning |
818 (119) | NBP Equipment Malfunction | Unable to verify inflation cycle and runtime data | warning |
818 (120) | NBP Equipment Malfunction | New message sent while previous message incomplete | error |
818 (121) | NBP Equipment Malfunction | Pressure value from second pressure sensor not available | error |
818 (122) | NBP Equipment Malfunction | 5-volt supply voltage too high or low | error |
818 (123) | NBP Equipment Malfunction | Error from the 5-volt regulator | error |
818 (124) | NBP Equipment Malfunction | 12-volt supply voltage too high or low | error |
818 (128) | NBP Equipment Malfunction | Offset of pressure sensor 1 out of range | error |
818 (129) | NBP Equipment Malfunction | Offset of pressure sensor 2 out of range | error |
818 (130) | NBP Equipment Malfunction | Response of pressure sensor 2 to test signal out of range | error |
818 (131) | NBP Equipment Malfunction | Difference between two pressure channels too large | error |
818 (132) | NBP Equipment Malfunction | One of the two valves activated for too long | error |
818 (133) | NBP Equipment Malfunction - CRC mismatch in calibration data | CRC error detected in calibration data on startup | error |
818 (134) | NBP Equipment Malfunction - CRC mismatch in error log data | CRC error detected in error log data on startup | error |
818 (135) | NBP Equipment Malfunction - CRC mismatch in inflation cycle and runtime data | CRC error in inflation cycle count and runtime data | error |
818 (136) | NBP Equipment Malfunction - Pneumatic safety switch malfunction | Pneumatic safety switch does not respond correctly | error |
818 (137) | NBP Equipment Malfunction - Valve 2 defective or blocked | Valve 2 fails to open or close properly | error |
818 (141) | NBP Equipment Malfunction - Inflation time exceeded | Inflation time for selected patient type exceeds normal duration | warning |
818 (199) | NBP Equipment Malfunction - No data from NBP block | No communication with NBP block module | error |
818 (25) | NBP Equipment Malfunction | NBP hardware malfunction due to no data | error |
819 | NBP Equipment Malfunction - Not Calibrated | NBP module has not been calibrated | error |
819 (127) | NBP Equipment Malfunction - NBP module not calibrated | NBP module calibration not performed or invalid | error |
820 (138) | NBP Hose Blocked | Blockage or pressure drop detected in pneumatic system | error |
821 (139) | NBP Air Leak | Air leak or pressure loss detected in pneumatic system | error |
833 (1) | SpO2 Equipment Malfunction - ROM checksum error | SpO2 board ROM fails checksum verification on startup | error |
833 (10) | SpO2 Equipment Malfunction - Host communication error | SpO2 board fails to communicate with main processor | error |
833 (100) | SpO2 Equipment Malfunction - Masimo board not enabled | Masimo SpO2 board is installed but not enabled in firmware | error |
833 (101) | SpO2 Equipment Malfunction - Masimo DSP Raw Queue Overrun | Masimo DSP raw data queue overflows during operation | error |
833 (102) | SpO2 Equipment Malfunction - Masimo DSP hardware failure | Masimo DSP processor fails initialization or self-test | error |
833 (103) | SpO2 Equipment Malfunction - Masimo MCU failure | Masimo microcontroller unit fails initialization or test | error |
833 (104) | SpO2 Equipment Malfunction - Masimo watchdog failure | Masimo watchdog timer detects processor hang or failure | critical |
833 (105) | SpO2 Equipment Malfunction - Masimo invalid board type | Masimo board type identifier does not match expected value | error |
833 (106) | SpO2 Equipment Malfunction - Masimo invalid master control state | Masimo master control state machine detects invalid state | error |
833 (107) | SpO2 Equipment Malfunction - Masimo SRAM transfer failure | Masimo SRAM memory transfer fails during initialization | error |
833 (108) | SpO2 Equipment Malfunction - Masimo SRAM task queue overrun | Masimo SRAM task queue overflows during operation | error |
833 (109) | SpO2 Equipment Malfunction - Masimo database failure | Masimo SpO2 board database initialization or access fails | error |
833 (110) | SpO2 Equipment Malfunction - Masimo invalid flash memory device | Masimo flash memory device not recognized or fails test | error |
833 (2) | SpO2 Equipment Malfunction - RAM test error | SpO2 board RAM fails self-test on startup | error |
833 (20) | SpO2 Equipment Malfunction - Hardware defect | SpO2 board detects internal hardware fault | error |
833 (4) | SpO2 Equipment Malfunction - Analog signal path self test failed | SpO2 analog signal path fails self-test | error |
833 (40) | SpO2 Equipment Malfunction - Non-volatile memory corrupted | SpO2 board non-volatile memory checksum or integrity failure | error |
833 (6) | SpO2 Equipment Malfunction - Self test timeout | SpO2 board self-test does not complete in expected time | error |
833 (8) | SpO2 Equipment Malfunction - Internal DSP communication error | Communication fails between SpO2 board and DSP | error |
834 | SpO2 Sensor Malfunction | SpO2 sensor does not respond or connection lost | error |
836 | SpO2 Communication Error | Communication lost with SpO2 module during operation | info |
837 | SpO2 Error - Front end will reset | No data received from SpO2 module | error |
838 (100) | SpHb Equipment Malfunction - not enabled | SpHb module installed but not enabled | error |
839 (100) | RRa Equipment Malfunction - not enabled | RRa module installed but not enabled | error |
850 | CO2 Error - Front end will reset | No data received from CO2 module | error |
851 | CO2 Equipment Malfunction - Communication error | CO2 module communication fails | error |
851 (1) | CO2 Equipment Malfunction - Calibration error | CO2 module calibration verification fails | error |
851 (16) | CO2 Equipment Malfunction - Main board error | CO2 module main board fault detected | error |
851 (17) | CO2 Equipment Malfunction - Module malfunction | General CO2 module malfunction detected | error |
851 (18) | CO2 Equipment Malfunction - Scrubber pump error | CO2 scrubber or pump mechanism fails | error |
851 (19) | CO2 Equipment Malfunction - Sensor malfunction | CO2 sensor fails to respond or produce valid data | error |
851 (2) | CO2 Equipment Malfunction - Sampling line malfunction | CO2 sampling line blockage or disconnection detected | warning |
851 (20) | CO2 Equipment Malfunction - 15V voltage out of range | 15 V supply voltage to CO2 module out of specification | error |
851 (3) | CO2 Equipment Malfunction - Occlusion in input line | Blockage detected in CO2 input line | warning |
866 (21) | Temp Probe Error - Operating above 43.3°C | Temperature probe operating in conditions above 43.3°C (112°F) | error |
866 (22) | Temp Probe Error - Excessive heater energy | Excessive heater energy detected or slow temperature response | error |
866 (32) | Temp Probe Error - Thermistor pulled away from tip | Thermistor separation or heater failure detected | error |
866 (33) | Temp Probe Error - Probe not responsive | Probe does not respond to commands or not calibrated | error |
866 (34) | Temperature probe not characterized or calibrated | Probe initialization or measurement attempt | warning |
866 (45) | Temperature measurement below allowable range | Measurement reading significantly below low ambient limits | warning |
866 (46) | Temperature measurement above allowable range | Measurement reading significantly above high ambient or patient limits | warning |
866 (52) | Temperature probe not characterized or calibrated | Probe initialization or characterization check | warning |
866 (63) | Temperature probe well missing or improperly installed | Probe well detection during initialization | error |
866 (70) | Error reading probe EEPROM | Probe EEPROM read operation during startup or measurement | error |
866 (75) | Probe EEPROM read error or probe not factory tested | Probe EEPROM read or factory test validation | error |
866 (78) | Probe non-volatile error detection mechanism triggered | Probe error detection during operation | error |
866 (79) | Temperature module non-volatile memory error detected | Module error detection mechanism during operation | error |
866 (80) | Probe heater stuck on | Probe heater turnoff command but heater still on | error |
866 (81) | Probe heater stuck off | Probe heater turnon command but heater still off | error |
867 (101) | Temperature Module Malfunction (101) | Checksum error in module | error |
867 (102) | Temperature Module Malfunction (102) | Calibration error in module | error |
867 (103) | Temperature Module Malfunction (103) | Ambient temperature below minimum operating limit | warning |
867 (104) | Temperature Module Malfunction (104) | Ambient temperature above maximum operating limit | warning |
867 (201) | Temporal Temperature Module Malfunction (201) | Temporal temperature module checksum error | error |
867 (202) | Temporal Temperature Module Malfunction (202) | Internal firmware error in temporal module | error |
867 (203) | Temporal Temperature Module Malfunction (203) | Temporal module below allowable ambient temperature | warning |
867 (204) | Temporal Temperature Module Malfunction (204) | Temporal module above allowable ambient temperature | warning |
867 (24) | Ambient temperature too high | Ambient temperature measurement exceeds safe operating range | warning |
867 (25) | Ambient temperature too low | Ambient temperature measurement below safe operating range | warning |
867 (27) | Battery or power supply voltage exceeds maximum | Supply voltage monitoring detects overvoltage condition | error |
867 (28) | Battery or power supply voltage below minimum | Supply voltage monitoring detects undervoltage condition | error |
867 (47) | Internal calibration resistor damaged | Calibration resistor validation during operation | error |
867 (48) | Internal calibration resistor damaged | Calibration resistor validation during operation | error |
867 (49) | Circuit validation resistor damaged | Circuit validation resistor check during operation | error |
867 (50) | Circuit validation resistor damaged | Circuit validation resistor check during operation | error |
867 (51) | A/D measurement timeout | Analog to digital conversion does not complete in time | error |
867 (59) | Battery or power supply voltage below maximum | Supply voltage monitoring detects low voltage condition | error |
867 (60) | Battery or power supply voltage exceeds minimum | Supply voltage monitoring detects high voltage condition | error |
867 (61) | Reference voltage circuit under voltage or unstable | Reference voltage regulation monitoring during operation | error |
867 (62) | Temperature module not calibrated | Module calibration check during startup | error |
867 (65) | Error saving to module EEPROM | EEPROM write operation during calibration or operation | error |
867 (66) | Module non-volatile calibration memory error detected | Calibration memory error detection mechanism during operation | error |
867 (67) | Error reading or writing module EEPROM | EEPROM read/write operation during operation | error |
867 (68) | Error reading or writing module EEPROM | EEPROM read/write operation during operation | error |
867 (69) | Error reading or writing module EEPROM | EEPROM read/write operation during operation | error |
867 (74) | Temperature Module Malfunction (74) | Internal error during EEPROM initialization | error |
867 (75) | Temperature Module Malfunction (75) | Cannot read probe EEPROM correctly or probe untested | error |
867 (82) | Temperature Module Malfunction (82) | Signal HTR_Q on, HTRC off, but voltage present | error |
867 (83) | Temperature Module Malfunction (83) | Signal HTR_Q tri-stated with HTRC enabled and heater power present | error |
867 (84) | Temperature Module Malfunction (84) | Q&C on but heater voltage insufficient | error |
867 (85) | Temperature Module Malfunction (85) | Heater fail-safe did not turn off as expected | error |
868 | Temperature Error | No data from temperature module | error |
881 | Speaker Malfunction | Speaker failure | warning |
882 | Wireless Module Malfunction (882) | Wireless module malfunction | error |
883 | System Error (883) | Internal state error for power management detected | error |
887 | System Error (887) | Backup image used to boot due to latest software image failure | error |
888 | System Error | System could not load font files | error |
889 | System Error (889) | System failed to mount one of the file systems | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.