Philips CRX-400: troubleshooting and error codes
The ServiceForge knowledge base holds 92 documented symptoms for the Philips CRX-400, with 210 ranked causes, 238 diagnostic checks with test points and expected values, and 17 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Anode over current fault high voltage
ANODE O.C. (DS42) LED lit on System Control Board (A7A1); ANALOG FAULT (DS16) LED also lit
Ranked causes
- 1. Anode Arc Fault occurring spontaneously (occasional)
- 2. Inverter jumper configuration incorrect (occasional)
- 3. X-ray tube requires seasoning (rare)
- 4. High voltage cable fault between inverter and multiplier (occasional)
- 5. X-ray tube failure (rare)
- 6. Anode feedback cable L8763 open circuit (occasional)
- 7. Anode Feedback and Buffer Board failure (rare)
Determine if fault occurs spontaneously without exposure
- Test point
- Error logger
- Expected
- Check for Anode Arc Fault presence
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cathode over-current fault high voltage
HVG-ANA: C-0.C. error message displayed; CATHODE O.C. indicator (DS43) and ANALOG FAULT indicator (DS16) lit on System Control Board
Ranked causes
- 1. X-ray tube arc or cathode fault (common)
- 2. High voltage cable connection fault (occasional)
- 3. High voltage transformer secondary wiring fault (occasional)
- 4. X-ray tube failure (occasional)
- 5. Cable W11 open circuit (occasional)
- 6. Cathode Feedback and Buffer board jumper open (occasional)
- 7. Cathode Feedback and Buffer board fault (rare)
Verify fault message reporting
- Test point
- DS43 indicator (CATHODE O.C.) on System Control Board A7
- Expected
- Indicator lit
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequency maximum fault high voltage
System triggers frequency maximum fault condition
Ranked causes
- 1. Low input voltage (480 VAC below 432 VAC) (common)
- 2. Cable L8762 (W6/W7) disconnected or loose (occasional)
- 3. High preheat current (exceeds 1V/20mA) (occasional)
- 4. XSC+ board fault (occasional)
- 5. VCO pulse fault (occasional)
- 6. System Control Board voltage fault (occasional)
- 7. W2 (L7210) cable open circuit (occasional)
Measure 480 VAC input supply voltage
- Test point
- 480 VAC supply
- Expected
- >=432 VAC
- Tools
- DMM
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator cathode arc fault high voltage
ANALOG FAULT (DS16) and GENERATOR CATHODE ARC (DS46) LEDs lit on System Control Board; fault message logged
Ranked causes
- 1. Cable W11 (L8763) open or poor continuity (occasional)
- 2. System control board failure (occasional)
- 3. Cathode Multiplier assembly failure (occasional)
- 4. X-ray tube arc condition (occasional)
- 5. X-ray tube failure (occasional)
- 6. High tension cable failure (Anode or Cathode) (occasional)
Verify fault occurs without exposure
- Test point
- Error logger
- Expected
- No Generator Cathode Arc Fault logged after reset
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interlock fault message displayed high voltage
PANEL INTLK (DS7) LED lit on System Control Board (A7), indicating interlock switch malfunction or defeat condition
Ranked causes
- 1. Interlock switch in defeat position (common)
- 2. Low voltage at JB6 pin 22 (occasional)
- 3. Short circuit on System Control Board at A7A1JB6-22 (occasional)
- 4. Open circuit on System Control Board at A7A1JB6-22 (occasional)
- 5. Open or short on cable L8760 (W2) (rare)
- 6. Open or short on cable L7209 (W3) (rare)
Verify interlock fault message and LED status
- Test point
- DS7 LED on System Control Board (A7)
- Expected
- PANEL INTLK LED lit
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interlock fault reported high voltage
System reports interlock fault; PANEL INTLK LED (DS7) lit on System Control Board (A7)
Ranked causes
- 1. Interlock switch in defeat position or not properly seated (common)
- 2. Low voltage at JB6 pin 22 on System Control Board (occasional)
- 3. Short circuit on System Control Board at A7A1JB6-22 to ground (occasional)
- 4. Open circuit on System Control Board at A7A1JB6-22 to ground (occasional)
- 6. Short circuit in L8760 (W2) connector (rare)
- 8. Short circuit in L7209 (W3) connector (rare)
Verify PANEL INTLK LED status
- Test point
- DS7 on System Control Board (A7)
- Expected
- LED lit
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator anode arc fault reported radiation
GEN ANODE ARC (DS45) and ANALOG FAULT (DS16) LEDs lit on System Control Board; error logger shows Generator Anode Arc Fault message
Ranked causes
- 1. Cable W11 (L8763) continuity fault or connection issue (common)
- 2. Anode Feedback and Buffer Board logic fault (low voltage) (occasional)
- 3. Anode Multiplier assembly fault (high voltage) (occasional)
- 4. X-ray tube arc or high tension cable fault (occasional)
- 5. Generator internal fault or tube degradation (rare)
Verify fault recurrence after system reset
- Test point
- Error logger
- Expected
- No Generator Anode Arc Fault message
3 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interlock fault message reported high voltage
PANEL INTLK (DS7) LED lit on System Control Board (A7) indicating interlock switch failure or defeat condition.
Ranked causes
- 1. Interlock switch in defeat position or pulled out (common)
- 2. Low voltage at JB6 pin 22 on System Control Board (occasional)
- 3. Short circuit on System Control Board at A7A1JB6-22 (rare)
- 4. Open circuit in cable W2 (L8760) between Inverter and Control Chassis (occasional)
- 5. Open circuit in cable W3 (L7209) between System Control and A.C. Input Chassis (occasional)
Verify PANEL INTLK (DS7) LED status
- Test point
- LED DS7 on System Control Board (A7)
- Expected
- Lit
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inverter power fault in filament output module high voltage
High-voltage X-ray filament inverter module power failure with diagnostic LED indicators
Ranked causes
- 1. Cable W8 connection issue (common)
- 2. Fuse defect (occasional)
- 3. Voltage supply issue (occasional)
- 4. Cable malfunction (occasional)
- 5. AC diagnostic board voltage issue (rare)
Verify LED status
- Test point
- FDBK (DS6) on Filament Power Control PCB and FIL FAULT (DS5) on System Control Board
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube anode arc fault reported high voltage
System reports anode arc fault; TUBE ANODE ARC LED (DS39) and ANALOG FAULT LED (DS16) lit on System Control Board
Ranked causes
- 1. Cable W11 (L8763) open or poor continuity (occasional)
- 2. Anode Feedback and Buffer Board fault at pin 8 of A1A2J2 (occasional)
- 3. X-ray tube arc condition (common)
- 4. High tension cables W7 (L8762) or W6 (L8762) defective (occasional)
- 5. X-ray tube failure (rare)
Measure continuity of cable W11 (L8763)
- Test point
- Pin 8 end to pin 8 other end of W11
- Expected
- less than 1 ohm, <1
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube cathode arc fault reported radiation
System reports tube cathode arc fault; TUBE CATHODE ARC LED (DS44) and ANALOG FAULT LED (DS16) lit on System Control Board (A7A1)
Ranked causes
- 1. Cathode arc occurring during exposure (common)
- 2. Cable W11 (L8763) continuity fault (occasional)
- 3. Anode feedback signal loss (pin 8 A1A2J2 out of range) (occasional)
- 4. Anode or cathode high tension cable failure (occasional)
- 5. X-ray tube internal fault (rare)
Determine if fault occurs without exposure
- Test point
- Error logger
- Expected
- No anode arc fault when no exposure is made
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
+15V supply missing in inverter high voltage
No +15V switched supply to inverter chassis; LED DS8 (LCL PWR) lit on System Control Board
Ranked causes
- 1. Cable W2 (L7210) or W3 (L7209) disconnected or short circuit (common)
- 2. Anode Multiplier connection issue (occasional)
- 3. Cathode Multiplier connection issue (occasional)
- 4. Inverter Module #1, #2, or #3 short or disconnected (occasional)
Verify fault message and check LCL PWR LED status
- Test point
- LED DS8 on System Control Board
- Expected
- DS8 lit confirms +15V supply missing
12 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Anode over voltage fault high voltage
ANODE OVP (DS40) LED lit on System Control Board (A7A1); fault message reported
Ranked causes
- 1. Anode arc fault occurring without exposure (occasional)
- 2. Anode voltage feedback signal exceeds 8.6 volts (86 kV) (common)
- 3. Cable W11 (L8763) open or high resistance between feedback board connectors (occasional)
- 4. Anode Feedback and Buffer Board (A1A2) voltage fault at JB2-25 (rare)
Verify fault message and LED status
- Test point
- ANODE OVP (DS40) on System Control Board (A7A1)
- Expected
- LED lit
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bias voltage fault in AC chassis high voltage
Bias fault message displayed; BIAS VOLTAGE LED (DS10) lit on AC Diagnostics Board indicating power supply bias voltage failure in AC input chassis
Ranked causes
- 1. Cable L7209 short circuit or open connection (common)
- 2. Cable L7210/L8760 connection or short to inverter chassis (occasional)
- 3. Supply fault in AC chassis power distribution (occasional)
- 4. Local power fault in filament module (occasional)
Verify proper fault reporting; check status of LEDs
- Test point
- AC Diagnostics Board A5A3A2 LED DS10 and System Control Board A7A1 LED DS4
- Expected
- DS10 lit (BIAS VOLTAGE); DS4 lit (AC FAULT); DS12 not lit (+/-24V UNSW OFF)
- Tools
- Visual inspection, multimeter
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cathode over-voltage fault message high voltage
CATHODE OVP LED (DS41) lit on System Control Board (A7); error logger reports cathode over-voltage
Ranked causes
- 1. Voltage regulation error at TP13 (common)
- 2. Cathode kV feedback (CAKVFB) signal not tracking output voltage (common)
- 3. Cable W11 (L8763) open or short circuit (occasional)
- 4. Cathode Feedback and Buffer PWB fault (occasional)
Verify fault occurs without exposure
- Test point
- Error logger
- Expected
- Cathode Over-voltage fault logged after reset without exposure
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament short detected high voltage
Fault message displayed with LCL PWR/TEST indicator lit on Filament Power PCB and FIL FAULT indicator lit on System Control Board
Ranked causes
- 1. Cathode HV module malfunction (common)
- 2. XSC board malfunction (occasional)
- 3. System Control board malfunction (occasional)
- 4. Punctured X-ray tube insert (rare)
Check FIL PROG signal voltage
- Test point
- A4A2JB2 pins 25 and 26
- Expected
- voltage level compared to 1 VDC threshold
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator cathode arc fault reported radiation
ANALOG FAULT (DS16) and GENERATOR CATHODE ARC (DS46) LEDs lit on System Control Board; error logger shows Generator Cathode Arc Fault message
Ranked causes
- 1. Cable W11 (L8763) continuity failure or connector issues (common)
- 2. System control board failure (occasional)
- 3. Cathode Multiplier assembly failure (occasional)
- 4. X-ray tube arc or high tension cable issue (occasional)
Verify fault occurs when no exposure is made
- Test point
- Error logger
- Expected
- No Generator Cathode Arc Fault present after reset
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Local power fault in filament module high voltage
System detects and reports local power fault in filament module power supply. Multiple LED status indicators show fault condition.
Ranked causes
- 1. Supply fault in power unit (HVG-SYS, HVG-INV1 bias, HVG-ACIN) (common)
- 2. Open circuit in filament power control board harness (A4A2JB2 to A7A1JB4) (occasional)
- 3. Open circuit on system control board internal connections (A7A1JB3 to A7A1JB4) (occasional)
- 4. Open circuit in filament control cable W2 (L7210) (occasional)
Check LED status indicators for power fault confirmation
- Test point
- +15 UNSW (DS2) on AC input chassis (A5); +15 UNSW (DS1) on system control board (A&); +15V UNSW (DS1) on filament power
- Expected
- LED indicators lit showing power fault condition
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rail short fault in AC chassis high voltage
Overcurrent fault detected in AC power distribution chassis; system reports RAIL SHORT fault on AC Diagnostics PWB and AC FAULT on System Control Board
Ranked causes
- 1. Open circuit or short circuit in AC power cable W9 (common)
- 2. Short or open circuit in Inverter #1 power distribution (common)
- 3. Short or open circuit in Inverter #2 power distribution (common)
- 4. Short or open circuit in Inverter #3 power distribution (common)
Verify fault reporting
- Test point
- RAIL SHORT LED (DS4) on AC Diagnostics PWB (A5A3A2) and AC FAULT LED (DS4) on System Control Board (A7)
- Expected
- Both LEDs lit
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC input fuse blown high voltage
Unit will not power on; fuse appears darkened or open when removed; 480VAC power not reaching internal circuits
Ranked causes
- 1. Input overcurrent event (common)
- 2. Shorted input load in unit (occasional)
- 3. Chronically undersized or failing fuse (rare)
Inspect fuse cartridge or element visually
- Test point
- AC input fuse on power input board
- Expected
- Fuse element intact and not darkened/discolored; wire or element unbroken
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC input missing phase C high voltage
System detects missing phase C in AC power input
Ranked causes
- 1. Bad fuse in AC input circuit (unknown)
- 2. Slip rings defective or worn (unknown)
- 3. AC input connector or wiring fault (unknown)
Actual Length of Exposure error reported radiation
ALE error logged to Error Logger during X-ray exposure
Ranked causes
- 1. ALE signal fault on XSC+ PCB (common)
- 2. ALE signal fault on High Voltage Generator (common)
- 3. ALE signal fault on System Control Board (occasional)
Check Error Logger for ALE fault presence
- Test point
- Error Logger
- Expected
- No ALE faults present
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AKVFDBK and CKVFDBK voltage at fault time radiation
Anode kilovoltage feedback (AKVFDBK) or C-arm kilovoltage feedback (CKVFDBK) voltages measured after EXGATE goes low at end of exposure
Ranked causes
- 1. System Control board defective; both voltages zero at fault (common)
- 2. Defective multiplier (one feedback voltage non-zero) (common)
- 3. Low discharge-current X-ray tube (Z-beam); both voltages non-zero (occasional)
Measure AKVFDBK at A7A1-TP15 after EXGATE goes low
- Test point
- A7A1-TP15
- Expected
- Well below 1.5 VDC before ALE Blanking goes HIGH
- Tools
- Digital multimeter, oscilloscope for timing observation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cathode multiplier over current high voltage
Fault message appears; CATHODE MULT O.C. indicator (DS20) on System Control Board (A7) is active
Ranked causes
- 1. Cathode Multiplier assembly failure (common)
- 2. Cable W11 (L8763) continuity failure (occasional)
- 3. System Control PW board failure (rare)
Verify fault occurs only during exposure
- Test point
- CATHODE MULT O.C. indicator (DS20)
- Expected
- Indicator off when no exposure is made
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cathode multiplier over current fault high voltage
Fault message reported; CATHODE MULT O.C. indicator (DS20) on System Control Board (A7) illuminates
Ranked causes
- 1. Cathode Multiplier assembly failure (common)
- 2. Cable W11 (L8763) continuity loss or damage (occasional)
- 3. System Control PW board failure (rare)
Verify fault message is reported correctly
- Test point
- CATHODE MULT O.C. indicator (DS20) on System Control Board (A7)
- Expected
- Indicator illuminated when fault occurs
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error: Cathode multiplier over current (E-CMOC) high voltage
System reports cathode multiplier overcurrent condition via DS20 indicator
Ranked causes
- 1. Defective Cathode Multiplier module (common)
- 2. Defective cable L8763 between control board and multiplier (occasional)
- 3. System Control board (A7) failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament rail fault displayed radiation
LED DS7 (FIL RAIL FAULT) illuminated on AC Diag PWB; AC FAULT LED DS4 may illuminate on System Control Board indicating X-ray tube filament power or control fault
Ranked causes
- 1. Three-phase supply voltage too low (common)
- 2. Filament power board indicator or circuitry fault (occasional)
- 3. F3 or F4 fuse open (occasional)
Measure three-phase voltage at fuse holder
- Expected
- 432 VAC minimum (480 VAC nominal), minimum
- Tools
- DMM AC voltage
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequency minimum fault reported high voltage
FMIN LED (DS13) lit on System Control Board (A7) indicates frequency minimum fault condition
Ranked causes
- 1. SW1 (7-10) switch incorrectly positioned (common)
- 2. Inverter enable signals incorrect (occasional)
- 3. XSC board failure (rare)
Verify SW1 switch position
- Test point
- SW1 (7-10) on System Control Board
- Expected
- CLOSED/ON position
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator cathode arc fault recurs high voltage
Fault message persists or reappears in error logger during normal operation after initial reset
Ranked causes
- 1. Tube arc condition (common)
- 2. X-ray tube degradation or failure (occasional)
- 3. High tension cable degradation (rare)
Perform exposure at reduced protocol settings
- Test point
- ELTP settings
- Expected
- No fault recurrence in error logger
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Negative bias fault reported high voltage
System reports negative bias error; check status of NEG BIAS 1, 2, 3 indicators on System Control Bd (A7)
Ranked causes
- 1. Defective inverter module voltage supplies (unknown)
- 2. Jumper JP2C defective or disconnected (unknown)
- 3. Defective ribbon cable connection (unknown)
Verify fault message proper reporting
- Test point
- NEG BIAS 1 (DS27), NEG BIAS 2 (DS29), NEG BIAS 3 (DS31) indicators
- Expected
- Indicators lit as appropriate
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Phase fault in 480VAC power supply high voltage
Unit displays phase fault error; system unable to deliver therapy; AC power distribution failure detected at input
Ranked causes
- 1. Customer facility power is out of specification (common)
- 2. Blown input fuse to unit (common)
- 3. AC chassis internal power distribution fault (occasional)
Measure three-phase AC voltage at unit input fuse holders
- Test point
- Fuse holders on AC power input panel
- Expected
- 432 VAC minimum phase-to-phase (480V ±10%/-12%), +10%/-12%
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulation error high voltage
X-ray high-voltage output regulation failure; TKVFB feedback signal anomaly detected
Ranked causes
- 1. SEQ LED status or TKVFB signal issue (common)
- 2. High voltage cable integrity issue (occasional)
- 3. TKVFB voltage out of range (occasional)
Verify SEQ LED status on System Control Board A7A1
- Expected
- SEQ LED illuminated
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Shutdown fault detected high voltage
System shutdown fault signal is active. Power supply or shutdown control circuit malfunction.
Ranked causes
- 1. Shutdown signal voltage out of specification at test point (common)
- 2. Actual Length of Exposure (ALE) error or Illegal KV fault (occasional)
- 3. System Control Board failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC voltage on diagnostic bias power pins abnormal radiation
Voltage between pins 17 and 18 of connector JB1 on AC Diagnostic and Bias Power board (A5A2A1) measured during exposure is out of specification
Ranked causes
- 1. AC chassis fault (unknown)
- 2. AC Diagnostic Board fault (unknown)
Measure voltage between A5A2A1JB1 pins 17 and 18 during exposure
- Test point
- A5A2A1JB1 pins 17 and 18
- Expected
- greater than or less than 20 VDC
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AKVFDBK or CKVFDBK not zero at fault time high voltage
Anode or cathode kV feedback voltages remain above zero when they should be zero
Ranked causes
- 1. Defective multiplier (anode or cathode) (occasional)
- 2. Low discharge-current x-ray tube (Z-beam) (rare)
Measure AKVFDBK at A7A1-TP15 after EXGATE goes low
- Test point
- A7A1-TP15
- Expected
- Well below 1.5 VDC before ALE Blanking goes high, <1.5
- Tools
- Voltmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable continuity fault between control boards
Intermittent or open connection between System Control board and XSC+ PCB
Ranked causes
- 1. Cable L7097 open or intermittent at A7A1JB1 (occasional)
- 2. Cable L7096 open or intermittent at A7A1JB2 (occasional)
Test cable L7097 continuity at A7A1JB1
- Test point
- Cable L7097, connector A7A1JB1
- Expected
- Continuity confirmed across all pins
- Tools
- Multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cathode over-voltage fault reported high voltage
CATHODE OVP LED (DS41) lit on System Control Board (A7); fault message logged in error logger
Ranked causes
- 1. Cable W11 (L8763) open or shorted between chassis (unknown)
- 2. Cathode Feedback and Buffer PWB defective (unknown)
Verify fault occurs only during exposure attempt or persistently
- Test point
- Error logger
- Expected
- Fault absent when no exposure is made
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC rail voltage imbalance high voltage
Voltages across DC rail capacitors not balanced within specification, indicating potential component failure or circuit imbalance.
Ranked causes
- 1. Open bleeder resistor on one rail (common)
- 2. Failed DC rail capacitor on one side (occasional)
Measure line-to-line resistance of Main DC rail cable L8761
- Test point
- black to red wire
- Expected
- greater than 2 kilohms
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament feedback fault high voltage
Filament feedback circuit fault indicated by error logger; may be accompanied by other generator faults
Ranked causes
- 1. XSC+ module malfunction (common)
- 2. Other generator faults present (occasional)
Access Error Logger and check for other generator faults
- Test point
- Error Logger display
- Expected
- List of any faults present
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inverter interlock fault high voltage
INV INTLK indicator (DS34) lit on System Control Board (A7)
Ranked causes
- 1. INTLK signal low or open circuit (unknown)
- 2. INTLK signal high or shorted (unknown)
Verify fault message reporting
- Test point
- INV INTLK (DS34) indicator
- Expected
- Lit on System Control Board (A7)
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
KV balance fault high voltage
KV BAL LED (DS18) lit on System Control Board (A7A2); system reports kV balance discrepancy between anode and cathode multipliers
Ranked causes
- 1. KV feedback signal difference exceeds 5% tolerance (common)
- 2. High Tension Transformer (HTT) output imbalance (occasional)
Measure KV feedback voltage across anode multiplier feedback board
- Test point
- A1A2TP5
- Expected
- In scale with selected kV (1V per 10kV by protocol); difference <2.5% of selected kV
- Tools
- Oscilloscope
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
kV/mA error detected radiation
Heat calculator error indicates kV or mA measurement out of tolerance
Ranked causes
- 1. kV feedback out of tolerance (unknown)
- 2. Anode mA out of tolerance (unknown)
Run sx diagnostic
- Test point
- Diagnostic system
- Expected
- Diagnostic completes without errors
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low voltage at JB6 pin 22 high voltage
Voltage measurement at JB6 pin 22 on System Control Board is below 1.5 volts
Ranked causes
- 1. Defective interlock circuit on System Control Board (common)
- 2. Short circuit to ground on control board (occasional)
Measure voltage at JB6 pin 22
- Test point
- JB6 pin 22 on System Control Board
- Expected
- 1.5 volts or higher V
- Tools
- DMM
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multiplier capacitor discharge fault high voltage
AKVFDBK or CKVFDBK voltage remains non-zero at time of fault after exposure
Ranked causes
- 1. Defective multiplier (A or C kV) (occasional)
- 2. Low discharge-current X-ray tube (Z-beam) (rare)
Measure AKVFDBK at A7A1-TP15 after EXGATE goes low at exposure end
- Test point
- A7A1-TP15
- Expected
- 0 VDC before ALE Blanking goes High
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overcurrent in filament chassis high voltage
O.C. LED (DS5) lit in Filament Chassis (A4A2) and FIL FAULT LED (DS5) lit in Control Chassis (A7)
Ranked causes
- 1. Filament Power/Control PWB Assembly voltage out of specification (common)
- 2. XSC board failure (occasional)
Measure voltage between JB2 pin 25 and JB2-26 on Filament Power/Control PWB Assembly
- Test point
- JB2 pin 25 to JB2-26
- Expected
- 6 volts in standby V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage out of range on AC Diagnostic and Bias Power board radiation
During exposure, voltage on pins 17 and 18 of connector JB1 (A5A2A1) measures incorrect value
Ranked causes
- 1. AC chassis fault (unknown)
- 2. AC Diagnostic Board fault (unknown)
Measure voltage between pins 17 and 18 of JB1 connector on A5A2A1 board during exposure
- Test point
- A5A2A1JB1 pins 17 and 18
- Expected
- greater than 20 VDC or less than 20 VDC
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC input rails below 200V high voltage
System displays fault message indicating rails voltage is low, suggesting AC input or inverter malfunction
Ranked causes
- 1. Bad AC input (ACIN) or inverter (unknown)
ALE goes HIGH after ALE Blanking high voltage
ALE (Anode Load Enable) signal transitions high after ALE Blanking signal goes high, indicating timing violation
Ranked causes
- 1. ALE timing violation (unknown)
Monitor ALE and ALE Blanking signals simultaneously
- Test point
- System control signals
- Expected
- ALE stays LOW while ALE Blanking is HIGH
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anode Multiplier Over Current fault radiation
Fault message reported; ANODE MULTIPLIER O.C. indicator (DS19) on System Control Board (A7) is active
Ranked causes
- 2. Cable W11 (L8763) continuity fault (occasional)
Verify fault occurrence during normal operation
- Test point
- System operation
- Expected
- Fault occurs only during exposure attempts
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anode/Cathode Arcs fault reported radiation
Anode or Cathode arcing faults logged to Error Logger during X-ray exposure
Ranked causes
- 1. Anode/Cathode arcing during exposure (occasional)
Check Error Logger for Anode/Cathode Arc faults
- Test point
- Error Logger
- Expected
- No arc faults present
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Both AKVFDBK and CKVFDBK non-zero at fault time high voltage
Both anode and cathode feedback voltages elevated simultaneously when they should be zero
Ranked causes
- 1. Low discharge-current x-ray tube (Z-beam) (rare)
Verify both AKVFDBK and CKVFDBK are non-zero
- Test point
- A7A1-TP15 and A7A1-TP14
- Expected
- Both should be zero; if both non-zero at fault time indicates possible Z-beam tube
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Jumper and switch settings verification
Verify correct jumper and switch configuration
Ranked causes
- 1. Incorrect jumper or switch settings (unknown)
Verify jumper and switch settings
- Test point
- Jumper and switch configuration on board
- Expected
- Per Page 3 specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System voltage feedback scaling radiation
Verify tube kilovoltage feedback signal scaling (1 V per 20 kV) is correctly calibrated across diagnostic test points
Ranked causes
- 1. Feedback scaling reference needed for troubleshooting (unknown)
Verify TKVFDBK scaling factor
- Test point
- A7A1-TP16
- Expected
- 1 V = 20 kV, per 20 kV
- Tools
- Digital multimeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: X-ray tube seasoning for cathode arc reduction radiation
Procedure to condition cathode after detecting tube arc faults during troubleshooting
Ranked causes
- 1. X-ray tube arc detected (common)
Set reduced power protocol
- Test point
- ELTP settings
- Expected
- Lower kV and mA values configured
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cathode multiplier diagnostics - cable interchange procedure high voltage
Test procedure to isolate cable faults by physically interchanging anode and cathode multiplier cables to determine if error moves
Ranked causes
- 1. Cable routing or connector fault (occasional)
Rerun exposure after cable interchange
- Test point
- System Control Board indicator
- Expected
- Error moves to anode circuit (indicates cable problem) or remains on cathode (indicates multiplier problem)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC rail voltage below minimum high voltage
One or both sides of DC rail capacitor voltage falls below 200 VDC, indicating insufficient power supply output.
Ranked causes
- 1. Faulty AC chassis (common)
Measure voltage across DC rail capacitors
- Test point
- DC rail caps
- Expected
- 200 VDC minimum on each side
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error persists after initial remedies high voltage
Fault continues after checking multiplier assembly and cables
Ranked causes
- 1. System Control PW board defective (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error recurs after waveform check high voltage
System repeats error condition despite correct waveforms observed
Ranked causes
- 1. Defective System Control board (common)
Observe waveforms
- Test point
- System output
- Expected
- As specified in preceding steps
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error recurs despite correct waveforms high voltage
System error repeats even when voltage waveforms appear normal on oscilloscope
Ranked causes
- 1. System Control board defective (common)
Inspect voltage waveforms on oscilloscope
- Test point
- System control circuit
- Expected
- Waveforms normal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error recurs during X-ray exposure radiation
Waveforms indicate recurring system error during operation
Ranked causes
- 1. System Control board failure (common)
Measure TKVFDBK at test point A7A1-TP16
- Test point
- A7A1-TP16
- Expected
- >30kV (1.5 V), >1.5 V
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Filament circuit AC bias supply fault
- High voltage feedback out of range
- High Voltage Generator fault
- High voltage interlock fault
- High voltage on AC Diagnostic and Bias Power board during exposure
- HV interlock fault reported
- Illegal kV Fault error reported
- KV balance fault diagnostic procedure
- Line voltage sag
- Line voltage sag detected
- Low AC input voltage
- Low DC rail voltage
- Low voltage on AC Diagnostic and Bias Power board during exposure
- Main DC rail cable insulation degradation
- Open filament fault
- Rail voltage fault
- Rail voltage fault reported
- RELAY FAULT
- Relay fault reported
- Sequence error reported
- Sequence fault - missing XPREP at XGATE
- Shoot thru in filament module
- Shutdown indicator LED inactive when expected
- System requires component replacement
- Test point signal voltage abnormal
- TKVFDBK voltage below 30 kV
- TKVFDBK voltage exceeds 30 kV
- TKVFDBK voltage out of range
- Tube anode arc fault after exposure
- Unit does not start up
- X-ray timeout error
- X-ray tube arcing
32 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1012 | Uncommanded X-ray exposure event | Unintended radiation emission detected | critical |
ACIN: MISSING PHASE C | AC input phase C is missing or not detected | System power-up or phase loss during operation | error |
ACIN: RAILS <200V | AC input rails voltage fault | AC input voltage drops below 200V | error |
ANA: CATHODE OVER VOLTAGE - BAD MULTI/CNTRL | Cathode over-voltage condition detected; multifunction or control board suspected | High voltage at cathode filament supply exceeds safe limits | critical |
CATHODE MULT O.C. | Cathode multiplier over current fault | Over-current condition in cathode multiplier circuit or cable | error |
CATHODE OVER VOLTAGE | Cathode voltage has exceeded safe operating limits | Detected during X-ray tube operation or high-voltage supply monitoring | error |
FIL | Open filament fault | Filament circuit open condition detected | error |
FIL: OPEN FILAMENT | Filament circuit is open | X-ray tube filament connection failure or component fault | error |
FIL RAIL FAULT | Filament rail power or control fault on X-ray tube circuit | Self-test detects fault in filament rail power or control; LED DS7 illuminates on AC Diag PWB | error |
FMIN | Frequency minimum fault | Inverter enable signal or system control issue detected | error |
GEN ANODE ARC | Generator anode arc fault detected | Arc detected in anode circuit or related control board fault | error |
Generator Cathode Arc Fault | Cathode arc detected in X-ray generator circuit | Arc condition in cathode circuit or cable continuity loss | error |
HVG-ANA: C-0.C. | Cathode over-current fault in high voltage generator analog circuit | Cathode current exceeds safe limit during X-ray generation or at generator reset | error |
HV INTERLOCK FAULT | High voltage interlock fault detected | Open circuit detected in HV interlock wiring chain (W2 cable or multiplier assembly connections) | error |
kV/mA ERROR | Heat calculator kV or mA measurement out of specification | Measured kV or mA exceeds tolerance limits during system operation | error |
LCL PWR: MISSING +15V - BAD INV/ACIN | Local power missing +15 volts, inverter or AC input fault | System control board detects absence of +15V supply | critical |
PHASE FAULT | Three-phase AC input voltage out of specification or imbalanced | Automatic detection of 480VAC phase-to-phase voltage < 432VAC or phase imbalance at unit power input | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community