Philips Converter test kit: troubleshooting and error codes
The ServiceForge knowledge base holds 63 documented symptoms for the Philips Converter test kit, with 72 ranked causes, 141 diagnostic checks with test points and expected values, and 8 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
ENF1 circuit breaker tripped high voltage
ENF1 breaker trips, possibly with mains fuses also tripped
Ranked causes
- 1. kV unit malfunction (kV-control, kV-power, converter, HV transformer) (common)
- 2. External parts failure (HV cable or tube) (occasional)
Measure DC voltage on PCB kV-power EQ100
- Test point
- E2 – C1 at both converters
- Expected
- < 10VDC, < 10
- Tools
- DMM
ENF1, ENF2, or ENF3 protection relay tripped high voltage
One or more ENF (mains fuse protection) relays have activated. Mains fuses may also have tripped.
Ranked causes
- 1. Rotor control unit fault (EY or EYA) (common)
- 2. Other electrical fault (occasional)
Verify mains power is completely disconnected
- Test point
- Mains supply
- Expected
- Power off at room mains switch or earth leakage current breaker
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ENF2 circuit breaker tripped high voltage
ENF2 breaker trips, possibly with ENF1 and mains fuses also tripped
Ranked causes
- 1. Internal generator units malfunction (Optimus RAD or R/F) (unknown)
- 2. External units malfunction connected to breaker ENF2 (unknown)
Resonance frequency mismatch between converters high voltage
Optimus 65/80 with dual converters: individual resonance frequencies are not matched
Ranked causes
- 1. HV transformer in one converter out of tolerance (unknown)
- 2. Resonance capacitor(s) mismatched between converters (unknown)
Measure resonance frequency of converter 1 primary coil
- Test point
- Converter 1 primary coil terminals
- Expected
- 17.160 kHz to 18.970 kHz, ±5%; 17.160–18.970 kHz
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resonance frequency out of range high voltage
Measured resonance frequency of primary coils or individual coils falls outside specified tolerance band
Ranked causes
- 1. HV transformer fault (unknown)
- 2. Resonance capacitor(s) out of tolerance (unknown)
Measure resonance frequency of primary coils
- Test point
- Primary coil terminals
- Expected
- 17.160 kHz to 18.970 kHz, ±5% on inductance and capacitors; 17.160–18.970 kHz
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube arcs during reconditioning high voltage
X-ray tube arcs (electrical discharge) at a specific kV step during the reconditioning procedure
Ranked causes
- 1. Tube contamination or surface defects requiring reconditioning (common)
- 2. Tube failure after multiple reconditioning attempts (rare)
Monitor tube arcing during exposure sequence
- Test point
- EZ130 X3 AV_HT (high voltage monitor), ground at EZ130 X6
- Expected
- No arc during exposure
- Tools
- Oscilloscope probe (20kV/V), large focus only
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube or high-voltage cable defect detected after high-voltage application high voltage
Defective tube or HV cable may not be evident from battery test, visual check, or audible check but becomes apparent when high voltage is applied. Procedure uses 1ms exposures to verify tube and cable integrity before full-power operation.
Ranked causes
- 1. Tube defective (unknown)
- 2. HV cable defective (unknown)
Generator OFF and test cables established at backpanel service measuring points
- Test point
- EZX74 (CTRL_X_C/ cathode of diode) and EZX77 (X_ACT_S/ anode of diode)
- Expected
- Test cable with diode connected; resistor cable at mAs short-circuit plug EG100
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TV image does not appear or appears too slowly after fluoroscopy start
TV image visibility delayed beyond acceptable response time after fluoroscopy is switched ON.
Ranked causes
- 1. Gain factor too low (occasional)
- 2. TV_pos_limit set too high (occasional)
Measure current Gain factor setting
- Test point
- XRGSCOPE: OPTIMUS > Adjustments > Dose rate control > Amplification gain > CONT
- Expected
- Current Gain factor value (range 0.5 to 2.0)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TV image flashes or overshoots when fluoroscopy is switched on
Visible overshoot or flashing of TV image upon fluoroscopy activation, indicating excessive gain or rapid response.
Ranked causes
- 1. TV_pos_limit set too low (occasional)
- 2. Gain factor too high (rare)
Measure overshoot on oscilloscope
- Test point
- Oscilloscope Y1 (EN X C/) and Y2 (DR DFP) during fluoroscopy ON transition
- Expected
- Overshoot ≤ 20% (final video level 100%, overshoot peak ≤ 120%)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anode disk acceleration sounds during start-up
Audible noise from tube housing during generator start-up resembles acceleration even when disk was stationary before power-ON
Ranked causes
- 1. Normal brake sequence operation after mains restoration (common)
Perform audible check at tube housing during start-up
- Test point
- tube housing
- Expected
- Brake sequence sounds present for approximately 40 seconds
Anode speed decreases rapidly after generator shutdown
Anode speed audibly decreases within 2-3 minutes after switching off generator with PREP held and free run engaged
Ranked causes
- 1. Tube vacuum lost or defective (common)
Observe anode speed decrease after generator shutdown with PREP held
- Test point
- Tube housing
- Expected
- Anode speed remains high for several minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Asymmetrical primary current waveform (Optimus 65/80) high voltage
Oscilloscope displays non-symmetrical waveform of primary current instead of expected symmetrical pattern
Ranked causes
- 1. HV transformer defective (common)
Inspect waveform symmetry on oscilloscope
- Test point
- Primary current transformer secondary, both probes
- Expected
- Symmetrical waveform about center line
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Area Dose Calculator default adjustment radiation
Verify that the Area Dose Calculator displays accurate dose values at specified kV settings compared to measured doses
Ranked causes
- 1. Default yield curve requires correction when measured dose deviates from displayed value (occasional)
Place lead plate and measuring cell in central radiation beam at 1 m distance (SMD)
- Test point
- Central radiation beam, table surface
- Expected
- Lead plate reduces scatter approximately 10%
- Tools
- Lead plate, calibrated dosemeter (DALI with cell 77334 or PMX3)
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Area Dose Calculator filter correction radiation
Correct filter attenuation factors (2 mm Al, 1 mm Al + 0.1 mm Cu, 1 mm Al + 0.2 mm Cu) for accurate dose product calculations
Ranked causes
- 1. Filter correction factors deviate from defaults and require update (occasional)
Verify test setup and acceptance of dose values without filter from yield measurement or remeasurement
- Test point
- Measurement records
- Expected
- Dose values at kV checkpoints 50, 70, 100, 150 kV from section 1.2 measurements
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Area Dose Calculator reference files and defaults
Overview of default yield and filter correction reference files stored on floppy disk
Ranked causes
- 1. Need to restore or verify default calibration files (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Area Dose Calculator specific yield correction radiation
Correct the specific yield curve for X-ray tube output at multiple kV checkpoints to ensure accurate area dose calculations
Ranked causes
- 1. Measured specific yield values deviate from default curve and require update (occasional)
Verify test setup: 1 m SMD, free cassette, kV-mAs-s technique, 10 mAs, 0.1 s, 10×10 cm collimation, no filter
- Test point
- Generator setup
- Expected
- Settings match section 1.2 prerequisites exactly
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose rate alignment for continuous fluoroscopy radiation
Adjust amplification factor so TV image becomes visible within 400ms after fluoroscopy switch-on and overshoot does not exceed 20% in quick mode or slightly higher in normal mode.
Ranked causes
- 1. Initial DR TV control circuit parameters not optimized (common)
Verify switch-on time and overshoot with software settings
- Test point
- T1 (transition time in quick mode)
- Expected
- T1 ≦ 400ms in quick mode or ≦ 500ms in normal mode, ≤ specified
- Tools
- XRGSCOPE software
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Function Unit kV alignment high voltage
Alignment of kV controller Factor Duty Cycle to ensure actual kV value attained within 2ms without kV break-in or overshoot during exposure
Ranked causes
- 1. Factor Duty Cycle requires alignment after FRU replacement (common)
Warm up tube before starting alignment
- Test point
- Tube heating status
- Expected
- Tube warmed up
9 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maximum patient entrance dose rate radiation
Entrance Dose rate Limiter (EDL) adjustment to comply with national regulations (e.g., 10R/min = 1.4567mGy/s in USA) by reducing maximum power output
Ranked causes
- 1. Dose rate exceeds regulatory limit (unknown)
Cover I.I. with lead for protection
- Test point
- Image intensifier
- Expected
- Lead covering in place
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ohmmeter diode test polarity verification
Verify ohmmeter diode test feature polarity configuration before proceeding with converter part measurements
Ranked causes
- 1. Incorrect meter polarity configuration (rare)
Verify diode test anode polarity
- Test point
- Ohmmeter diode test mode
- Expected
- Anode symbol on meter = positive voltage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PCB kV-power IGBT transistor precaution caution
Before working on IGBT gate connections, take precautions against electrostatic discharge and mobile phone interference.
Ranked causes
- 1. IGBT gate damage from ESD or RF interference (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Contacts welded at breakers and contactors high voltage
Contacts at breakers ENF1, ENF2 or contactor ENK1 are welded together, preventing normal operation
Ranked causes
- 1. Contacts welded at breakers ENF1, ENF2 or contactor ENK1 (occasional)
Switch OFF breakers ENF1 and ENF2
- Test point
- Breakers ENF1, ENF2
- Expected
- Both breakers in OFF position
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter chassis cracked or bent caution
Visual inspection reveals cracks visible at white IGBT body through converter cage grid, or kV-power PCB Q100 bent towards cover plate
Ranked causes
- 1. Physical damage to IGBT body or kV-power PCB Q100 (unknown)
Visual inspection of IGBT body through cage grid
- Test point
- White IGBT body
- Expected
- No cracks visible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter DC discharge time excessive high voltage
DC voltage remains on converter after power-off; discharge time > 11 minutes instead of normal 40 seconds to <60VDC when ENK2 discharging auxiliary contact is open.
Ranked causes
- 1. ENK2 main contactor contact welded (occasional)
Measure DC voltage at auxiliary contacts 41-42 of ENK2
- Test point
- ENK2 auxiliary contacts 41-42
- Expected
- 0V
- Tools
- Voltmeter, DC capable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Converter visual defects caution
Cracks visible on white IGBT body through converter cage grid or kV-power PCB Q100 bent towards cover plate
Ranked causes
- 1. Physical damage to IGBT or PCB (common)
Visual inspection of converter for physical damage
- Test point
- White IGBT body viewed through converter cage grid
- Expected
- No cracks visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cracks in ceramic resistor bodies caution
Ceramic bodies of damping resistors show visible cracks, requiring spring resoldering.
Ranked causes
- 1. Thermal or mechanical stress on resistor (unknown)
Inspect ceramic resistor bodies for cracks
- Test point
- Damping resistors R1, R2, R3 ceramic bodies
- Expected
- No cracks visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Duty cycle factor misalignment caution
Error codes 02HG (warning) or 02HH (error) indicate incorrect duty cycle factor setting
Ranked causes
- 1. Duty cycle factor out of adjustment (occasional)
Verify filament adaptation
- Test point
- Filament conditioning and adaptation status
- Expected
- Filament properly adapted before duty cycle realignment
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ENF1 and/or ENF2 and/or ENF3 tripped high voltage
One or more emergency fuses have tripped; mains fuses might have additionally tripped after a failure event.
Ranked causes
- 1. Rotor control unit EY (high speed) or EYA (low speed) failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ENF1, ENF2, or ENF3 fuse tripped high voltage
One or more of the emergency shutdown fuses (ENF1, ENF2, ENF3) have tripped, possibly with mains fuses also tripped, indicating a generator fault condition
Ranked causes
- 1. Rotor control unit EY (high speed) or EYA (low speed) failure (common)
Verify which fuses have tripped
- Test point
- ENF1, ENF2, ENF3, mains fuses
- Expected
- One or more tripped
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error messages 02HI and 02HJ appear during tube check
Error codes appear because mains lines are disconnected from converters during test
Ranked causes
- 1. Converter DC voltage out of range due to disconnected mains (common)
Verify mains lines disconnected from converters during tube check
- Test point
- converter primary power connections
- Expected
- Mains lines not connected; converter DC = 0V
Fuses tripped, cause undetermined high voltage
ENF1, ENF2, or ENF3 fuses tripped but the specific cause is not the rotor control unit; generator is in fault condition
Ranked causes
- 1. Unknown fault condition (unknown)
Verify generator is de-energized
- Test point
- Mains power
- Expected
- Generator OFF and completely disconnected from mains (room switch or earth leakage breaker open)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator failure with fuses tripped (non-rotor case) high voltage
ENF1, ENF2, and/or ENF3 tripped in cases other than rotor control failure. Generator must remain in this condition for diagnosis.
Ranked causes
- 1. Unknown failure requiring qualified technician diagnosis (unknown)
Ensure generator is switched OFF
- Test point
- Main power switch
- Expected
- OFF position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IGBT emitter-collector resistance out of specification caution
IGBT emitter to collector resistance measurements outside nominal range when converter fully assembled
Ranked causes
- 1. Defective IGBT transistor (common)
IGBT E1 to C1 resistance measurement
- Test point
- Q100 kV-power PCB E1 to C1
- Expected
- 21.5 - 22kΩ with ohm setting PLUS/MINUS probe on diode symbol, 21.5 - 22kΩ
- Tools
- Ohmmeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IGBT gate-emitter resistance out of specification caution
IGBT gate to emitter resistance measurements outside nominal range
Ranked causes
- 1. Defective IGBT transistor (common)
IGBT G1 to E1 gate-emitter resistance
- Test point
- Q100 kV-power PCB G1 to E1
- Expected
- 3.8kΩ ±10%, ±10%
- Tools
- Ohmmeter
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
kV values not symmetrical high voltage
Oscilloscope measurement shows asymmetric anode and cathode voltage behavior during exposure
Ranked causes
- 1. HV transformer defective (common)
Measure anode voltage with oscilloscope
- Test point
- EZ130 X4 AV_AN (anode high voltage), ground EZ130 X6
- Expected
- Symmetric waveform matching chart
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
kV-driver test high voltage
Functional test of kV-driver subsystem; error codes 02HI and 02HJ expected when mains not connected
Ranked causes
- 1. kV-driver malfunction or calibration needed (unknown)
Run kV-driver diagnostic test
- Test point
- Converter control interface
- Expected
- Test successful or specific error codes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
kV-power PCB Q100 bent towards cover plate caution
PCB Q100 is bent and in contact with or pressing towards the cover plate
Ranked causes
- 1. kV-power PCB Q100 deformed (unknown)
Visual inspection of kV-power PCB Q100 alignment
- Test point
- kV-power PCB Q100 position relative to cover plate
- Expected
- PCB is not bent and does not contact cover plate
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
mA-control fatal error after test cable connection caution
LED on mA-control EZ119 displays FATAL error condition after test cable with resistor is connected and generator has been running for some time.
Ranked causes
- 1. Test cable resistor causes mA-control error state (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Open springs at damping resistors high voltage
Springs at damping resistors R1, R2, R3 are open; may occur when all breakers are ON at service arrival, indicating prior short-circuit in converter or mains rectifier/filter
Ranked causes
- 1. Short-circuit in converter or mains rectifier/filter (common)
Visual inspection of springs at damping resistors R1, R2, R3
- Test point
- R1, R2, R3 springs
- Expected
- Springs closed, no open connections
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Overvoltage diode conduction failure caution
Overvoltage protection diodes V500/V501/V502/V503 not conducting at specified voltage or showing incorrect reverse leakage
Ranked causes
- 1. Defective overvoltage diode (common)
V500 diode cathode to anode forward voltage test
- Test point
- V500 cathode (heat sink plate) to anode
- Expected
- 310mV ±10%, ±10%
- Tools
- Diode test meter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power cable harness connector overheating caution
Connectors at resonance capacitors C3:2 and C13:2 show signs of thermal degradation or abnormal appearance
Ranked causes
- 1. Corroded or thermally damaged connector (occasional)
Visual inspection of connector at C3:2
- Test point
- Resonance capacitor C3:2 connector
- Expected
- Connector appearance (2): normal with no thermal damage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary current measurement out of specification high voltage
Oscilloscope waveform of primary current does not match expected range during 40kV, 50mA, 10ms generator operation
Ranked causes
- 1. HV transformer asymmetrical waveform (Optimus 65/80) (occasional)
Turn off generator and discharge converter DC
- Test point
- Generator power
- Expected
- Generator off, wait 1 minute
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rectifier diode fault check
Testing rectifier diodes for opens or shorts using ohmmeter diode test feature
Ranked causes
- 1. Rectifier diode failure (unknown)
Rectifier diode test with ohmmeter diode function
- Test point
- L1 to minus terminal
- Expected
- 450mV ±10% in forward direction, ±10%
- Tools
- Ohmmeter with diode test feature
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Rectifier diode malfunction high voltage
Mains rectifier diodes in EQV5 not conducting or blocking as expected, measured via ohmmeter diode test mode
Ranked causes
- 1. Rectifier diode forward/reverse resistance out of tolerance (unknown)
Verify ohmmeter polarity for diode test
- Test point
- Meter anode/cathode symbols
- Expected
- Anode symbol = positive (PLUS), cathode symbol = negative (MINUS)
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rectifier failure high voltage
Mains rectifier diodes not conducting or conducting incorrectly at specified test points
Ranked causes
- 1. Defective rectifier diode (common)
Diode test on L1 to minus terminal
- Test point
- L1 / – on Rectifier EQV5
- Expected
- 450mV ±10% with PLUS probe on diode symbol
- Tools
- Ohmmeter with diode test feature
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resistor group out of specification caution
Parallel resistor groups R506/507/508 or R509/510/511 not meeting combined or individual resistance values
Ranked causes
- 1. Defective or out-of-tolerance resistor in group (occasional)
Measure R506/507/508 group total resistance
- Test point
- R506/507/508 parallel combination
- Expected
- 39Ω ±5%, ±5%
- Tools
- Ohmmeter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resonance capacitor short or open circuit caution
Resonance capacitors C3:1 or C13:1 not functioning correctly; ohmmeter shows short-circuit or open connection
Ranked causes
- 1. Defective resonance capacitor (short or open) (occasional)
Disconnect primary lines from C3:1 and C13:1
- Test point
- Resonance capacitors C3:1 and C13:1
- Expected
- Primary connections removed for testing
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Short-circuit in converter or mains rectifier/filter high voltage
Electrical short-circuit detected in converter or its mains rectifier/filter circuit
Ranked causes
- 1. Component failure or insulation breakdown in converter or rectifier/filter (unknown)
Switch OFF all breakers
- Test point
- ENF1, ENF2 breakers
- Expected
- All breakers OFF
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Short-circuit on converter mains lines high voltage
Converter mains lines show short-circuit resistance when measured between phases.
Ranked causes
- 1. Damaged mains rectifier/filter circuit or converter wiring fault (occasional)
Remove converter mains lines from contactor ENK1
- Test point
- Contactor ENK1 terminals :1, :3, :5
- Expected
- Mains lines disconnected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube arcing detected patient safety
Oscilloscope shows kV break-in (dropout) within less than 10 microseconds during exposure
Ranked causes
- 1. Tube age or conditioning issue (occasional)
Monitor oscilloscope trace for kV dropout during exposure
- Test point
- EZ130 X4 (anode) or EZ130 X5 (cathode) with oscilloscope
- Expected
- Stable kV waveform with no break-in
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube arcing during kV adjustment patient safety
Tube arcing occurs during alignment exposure setting
Ranked causes
- 1. Tube not sufficiently warmed up (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube braking does not occur after generator shutdown
Anode does not automatically brake or brakes incorrectly after switching off generator with PREP engaged
Ranked causes
- 1. Rotor control system malfunction (high-speed mode) (unknown)
Verify tube braking after generator shutdown
- Test point
- Tube anode rotation
- Expected
- Automatic braking (high-speed control) or manual braking via PREP release (low-speed control)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube does not adapt after reconditioning high voltage
Tube remains defective and cannot be conditioned through the standard procedure even after multiple attempts
Ranked causes
- 1. Tube failure; reconditioning unsuccessful after three attempts (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube does not brake after mains breaker trip caution
Anode disk continues rotating at high speed after mains power loss during rotation; CPU does not complete brake sequence
Ranked causes
- 1. Brake sequence interrupted by mains loss during rotor operation (common)
Observe rotor behavior during generator start-up
- Test point
- tube housing
- Expected
- Anode disk decelerates and stops within 40 seconds of power-ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube does not brake after mains power loss patient safety
Anode disk continues rotating when mains breakers are tripped; CPU does not complete brake process
Ranked causes
- 1. Mains breaker tripped while tube rotating (high-speed rotor control) (common)
Verify anode disk rotation status before power-ON
- Test point
- Tube housing
- Expected
- Anode disk not rotating
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visible cracks on converter IGBT body caution
Cracks visible on the white IGBT body when looking through the converter cage grid
Ranked causes
- 1. IGBT body damaged (unknown)
Visual inspection of IGBT body
- Test point
- White IGBT body viewed through converter cage grid
- Expected
- No cracks visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveform (1) shows lower amplitude or different kV behavior than (2) and (3) high voltage
Converter test waveform analysis shows (1) has reduced amplitude or abnormal kV behavior compared to reference waveforms (2) and (3)
Ranked causes
- 1. X-ray tube defective (common)
Compare waveform (1) amplitude and kV behavior to waveforms (2) and (3)
- Test point
- Converter test waveforms (1), (2), (3)
- Expected
- Waveforms (1) and (2) should look almost the same with symmetry and amplitude
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveform (3) discharge ramp slower than expected high voltage
Converter test waveform (3) shows slower discharging ramp compared to normal, indicating possible cable capacitance issue
Ranked causes
- 1. HV cable capacitance present or incorrect cable installed (occasional)
Measure discharging ramp steepness of waveform (3)
- Test point
- Converter test waveform (3)
- Expected
- Steeper (faster) discharging ramp; tank 3 nF, 20 m cable typically 1.5 nF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveform (3) shows asymmetric amplitude high voltage
Converter test waveform (3) displays asymmetric amplitude pattern
Ranked causes
- 1. HV tank defective (common)
Observe waveform (3) for amplitude symmetry
- Test point
- Converter test waveform (3)
- Expected
- Waveform (3) should show symmetric amplitude (steeper discharging ramp than (1) and (2))
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveforms (2) and (3) differ in amplitude and symmetry; (3) is symmetric high voltage
Converter test shows waveforms (2) and (3) have different amplitude and symmetry, with (3) appearing symmetric
Ranked causes
- 1. HV cable(s) defective (common)
Compare waveforms (2) and (3) for amplitude and symmetry differences
- Test point
- Converter test waveforms (2) and (3)
- Expected
- Waveforms (2) and (3) should match in amplitude and symmetry
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Welded contacts at breakers and contactors
- X-ray tube shows visible damage
- X-ray tube visible damage assessment
3 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
02HG | Duty cycle factor misalignment warning in background | Duty cycle factor out of adjustment | warning |
02HH | Duty cycle factor misalignment error | Duty cycle factor out of adjustment | error |
02HI | Converter DC voltage out of range (0V) | Mains lines not connected to converter during test | warning |
02HJ | Converter DC voltage out of range (0V) | Mains lines not connected to converter during test | warning |
02HW | Asymmetry warning in background | kV anode and cathode values are not symmetrical | warning |
02HX | Asymmetry error | kV anode and cathode values are not symmetrical | error |
02WG | Tube arcing warning in background | kV break-in detected within < 10µs during exposure | warning |
02WH | Tube arcing error | kV break-in detected within < 10µs during exposure | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI