Philips Compano: troubleshooting and error codes
The ServiceForge knowledge base holds 24 documented symptoms for the Philips Compano, with 26 ranked causes, 96 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Sensitivity adjustment out of tolerance radiation
S value, pixel values, or density readings fall outside acceptable ranges after exposure and readout
Ranked causes
- 1. Incorrect dose measurement or entry (common)
- 2. Improper shading and polygon settings (occasional)
- 3. Reader software or calibration data corruption (occasional)
Verify measured dose value entry
- Test point
- CALCULATION / dose entry field
- Expected
- 1 mR ± 10% (range 0.5–10 mR acceptable, 1–2 mR recommended), ±10%
- Tools
- Calibrated dosimeter
9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: sensitivity adjustment radiation
Adjust CR reader sensitivity by exposing imaging plate to measured dose and verifying output values after readout
Ranked causes
- 1. Sensitivity calibration required (common)
Record baseline HVCENT and HVDATA values
- Test point
- DATA MANAGEMENT > DISPLAY CURRENT DATA
- Expected
- HVCENT and HVDATA values recorded (e.g. 474)
15 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensitivity and shading correction radiation
Ensure scanner produces correct pixel values and density across uniform exposure. Sensitivity check compares S value and pixel values to specification; if out of range, automatic correction must be run.
Ranked causes
- 1. Sensitivity out of specification after exposure (unknown)
Wait for scanner stabilization after exposure
- Test point
- Reader cassette entry timing
- Expected
- 10 minutes elapsed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensitivity and shading correction check radiation
Verification that scanner sensitivity S value and pixel values are within specification after uniform exposure
Ranked causes
- 1. Initial sensitivity check before correction adjustment (common)
Scanner cleaning precondition
- Test point
- Scanning unit
- Expected
- Unit cleaned
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensitivity correction with double dose radiation
Alternative sensitivity check using doubled exposure (2mR / 17.4 µGy) when standard dosage is not available.
Ranked causes
- 1. Double dose used for sensitivity check (occasional)
Measure S value with double dose exposure
- Test point
- System display EasyVision
- Expected
- 85 … 115
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Film annotation character format incorrect
Annotation characters printed on film are blurred, chipped, or otherwise abnormal; check format matches Console input.
Ranked causes
- 1. Annotation printing defect or font configuration error (unknown)
Verify film annotation characters match Console input format
- Test point
- printed film annotation area
- Expected
- characters properly printed and match Console input
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Output image format incorrect
White clear portion of output film exceeds acceptable width; check film width against standard.
Ranked causes
- 1. Improper image output configuration or printer settings (unknown)
Measure white clear portion width on output film at 100% image output
- Test point
- output film edge
- Expected
- does not exceed 2 mm in width
Output image irregularities present
Output film or monitor display shows peculiar irregularities; both must be free from abnormalities.
Ranked causes
- 1. Reader or console image processing defect (unknown)
Visually inspect output film for irregularities
- Test point
- output film surface
- Expected
- free from peculiar irregularities
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Pixel value verification – main scan direction
Verify pixel uniformity across IP in main scan direction (left to right edge) falls within specified tolerance
Ranked causes
- 1. Pixel value out of specification after calibration (occasional)
Measure pixel value across main scan direction (left to right)
- Test point
- IP main scan axis
- Expected
- Pixel value 486–536 (center 511 ±5%), ±5%
Pixel value verification – subscan direction
Verify pixel uniformity across IP in subscan direction (top to bottom edge) falls within specified tolerance
Ranked causes
- 1. Pixel value out of specification in subscan direction (occasional)
Measure pixel value across subscan direction (top to bottom)
- Test point
- IP subscan axis
- Expected
- Pixel value 460–562 (center 511 ±10%), ±10%
Reader CPU board may be damaged if power turned off during update caution
Risk of CPU board damage if Reader power is turned OFF while panel LEDs are illuminated during software and configuration update
Ranked causes
- 1. Reader power turned off during LED illumination phase (common)
Monitor Reader panel LEDs during initialization
- Test point
- Reader panel indicators
- Expected
- LEDs remain illuminated throughout update cycle
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reader LEDs do not illuminate during initialization
LEDs on Reader panel should illuminate with short beeps during Reader software and configuration data update after Console application starts
Ranked causes
- 1. Software reload with existing configuration data (occasional)
Verify Reader power is ON
- Test point
- Reader power switch
- Expected
- Reader powered on
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reader LEDs not illuminating during software update
During Console application startup after selecting Programs > Fuji Film > FCR, the LEDs on the Reader panel should illuminate and short beeps generated, but they do not appear
Ranked causes
- 1. Existing configuration data being reused from previous startup (occasional)
Sensitivity adjustment radiation
Calibration procedure to adjust reader sensitivity after IP exposure to known dose
Ranked causes
- 1. Reader sensitivity requires periodic adjustment (common)
Record initial HVCENT and HVDATA values
- Test point
- DATA MANAGEMENT - DISPLAY SCN DATA and DISPLAY CURRENT DATA
- Expected
- e.g. 474
9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity adjustment required radiation
Calibration procedure to adjust reader sensitivity using measured dose exposure and IP cassette
Ranked causes
- 1. Sensitivity calibration (common)
Record baseline HVCENT value
- Test point
- DATA MANAGEMENT, DISPLAY SCN DATA ETC
- Expected
- e.g. 474
9 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity adjustment verification radiation
After exposure and readout, HVCENT and QVDATA values are checked to confirm calibration changes took effect. Extreme differences indicate adjustment failure.
Ranked causes
- 1. Sensitivity adjustment not properly applied to reader (occasional)
Compare HVCENT value before and after IP readout
- Test point
- DISPLAY CURRENT DATA menu
- Expected
- Value should differ a little bit from the previous one
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity and shading correction adjustment needed radiation
S value and pixel values out of specification after uniform IP exposure; density or pixel values deviate from expected range
Ranked causes
- 1. Sensitivity settings require adjustment after scanner maintenance or calibration drift (common)
Verify scanner cleaning completed
- Test point
- Scanning unit
- Expected
- Cleaned before procedure
10 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity and shading correction verification radiation
S value and pixel/density values outside acceptable range after uniform IP exposure and scanning
Ranked causes
- 1. Sensitivity and shading calibration requires adjustment (unknown)
Ensure scanner unit cleaned and scanner data copied to floppy diskette via PC-MUTL service program
- Test point
- Scanning unit, floppy diskette
- Expected
- Unit clean, data backup present
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity check with double dose exposure radiation
Verification of S value range when using double radiation dose for testing
Ranked causes
- 1. Alternative sensitivity check procedure using higher dose (occasional)
Perform uniform exposure with double dose and measure S value
- Test point
- IP cassette, EasyVision or film output
- Expected
- S value 85...115 S units, 85–115 range
- Tools
- Dosimeter, X-ray tube
Sensitivity setting verification radiation
Procedure to verify that sensitivity adjustment has been applied correctly by re-checking values and confirming calibration stability after adjustment.
Ranked causes
- 1. Sensitivity adjustment requires verification to confirm proper calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity setting verification failed radiation
HVCENT or QVDATA values show extreme differences after IP read-out, or pixel/S/density values out of tolerance
Ranked causes
- 1. Reader drift or miscalibration causing sensitivity error (occasional)
Power cycle reader completely
- Test point
- Main power switch
- Expected
- Reader boots and reaches normal operating state
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity values outside tolerance
After calibration, measured HVCENT shows extreme deviation from baseline, indicating potential reader malfunction or environmental factor affecting sensitivity.
Ranked causes
- 1. Reader firmware or parameter initialization error (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify density tolerance radiation
Check that density measurement falls within acceptable tolerance range based on processing settings
Ranked causes
- 1. Density out of tolerance for current processing mode (occasional)
Measure density with no processing (curve 'a')
- Test point
- EasyVision output - no processing mode
- Expected
- 1.2...1.5...1.8, ±0.3
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify S value tolerance radiation
Check that S value (sensitivity measurement) falls within acceptable tolerance range
Ranked causes
- 1. S value out of tolerance range (occasional)
Measure S value
- Test point
- Reader output - main scan direction
- Expected
- 180...200...220, ±10%
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI