Philips BuckyDiagnost: troubleshooting and error codes

The ServiceForge knowledge base holds 82 documented symptoms for the Philips BuckyDiagnost, with 136 ranked causes, 148 diagnostic checks with test points and expected values, and 112 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit does not power on high voltage

Yellow LED on EN 100 is off, relay EN K2 does not switch over

Ranked causes

  1. 1. Mains supply failure (common)
  2. 2. Circuit breaker EN F1 not released (common)
  3. 3. Circuit breaker EN F2 not released (common)
  4. 4. Low-voltage supply failure (occasional)
  5. 5. Filament circuit failure (occasional)
  6. 6. Tube extension failure (occasional)
  7. 7. Rotor control failure (occasional)
  8. 8. External current consumer issue (occasional)
  9. 9. EN 100 board failure (rare)
  10. 10. Emergency-OFF circuit open (occasional)
  11. 11. Operating desk not connected (occasional)
  12. 12. Safety circuit open (occasional)
First check

Verify mains supply present

Test point
System input
Expected
Mains voltage present

5 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator selftest fails

Red LED on Z rack is ON continuously; Central Unit, Voltage Control, Current Control, Basic Interface, or Rotor Control module not responding

Ranked causes

  1. 1. Central Unit EZ139 fails (occasional)
  2. 2. Voltage Control EZ130 fails (occasional)
  3. 3. Current Control EZ119 fails (occasional)
  4. 4. Basic Interface EZ150 fails (occasional)
  5. 5. Rotor Control EY fails (occasional)
First check

Verify Central Unit EZ139 CAN communication

Test point
EZ139 module
Expected
CAN communication established

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator selftest fails with red LED continuously ON high voltage

During power-on, one red LED on Z rack should flash briefly; continuous red LED indicates selftest failure

Ranked causes

  1. 1. Central unit EZ139 fails (unknown)
  2. 2. Voltage control EZ130 fails (unknown)
  3. 3. Current control EZ119 fails (unknown)
  4. 4. Basic interface EZ150 fails (unknown)
  5. 5. Rotor control EY fails (unknown)
First check

Observe status LEDs on Z rack during startup

Test point
Z rack status LEDs
Expected
All LEDs flash, then one red LED flashes briefly

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not power on high voltage

Yellow LED on EN 100 does not illuminate; system fails to start selftest sequence

Ranked causes

  1. 1. Mains supply failure (common)
  2. 2. Circuit breaker failure (common)
  3. 3. Low-voltage supply failure (occasional)
  4. 4. Safety circuit open (occasional)
  5. 5. Emergency-OFF circuit open (occasional)
First check

Verify mains power supply

Test point
System mains input
Expected
Mains voltage present

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tomography run fails radiation

Tomography procedure does not complete successfully; system may not execute the programmed tomography sequence

Ranked causes

  1. 1. Ceiling Suspension (CS) assembly malfunction (unknown)
  2. 2. CS longitudinal Drive failure (unknown)
  3. 3. CS alpha Drive failure (unknown)
  4. 4. Table tomo Assemblies malfunction (unknown)
  5. 5. Tomo Controller failure (unknown)
First check

Verify tomography run using VT100 program on service PC

Test point
Service PC display
Expected
All steps during tomography run are visible in VT100 output
Tools
Service PC, VT100 program

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bucky unit programming and cassette sizing

Setup and verification of bucky controller settings, display units, receptor offsets, and cassette size calibration via XSCOPE software

Ranked causes

  1. 1. Incorrect display unit setting (metric/imperial mismatch) (common)
  2. 2. Receptor offset out of specification (common)
  3. 3. Cassette size settings mismatch (common)
  4. 4. Cassette sizing calibration not performed after new bucky installation (occasional)
First check

Verify VA2-board display unit setting

Test point
VA2-board switch
Expected
Set to inch or metric

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bucky unit cassette sizing radiation

Ensure cassette size display and positioning are correct after bucky installation or when cassette size detection fails

Ranked causes

  1. 1. Damaged cassettes or incorrect cassette size display (common)
  2. 2. Incorrect bucky unit programming or settings (common)
  3. 3. Analogue outputs not correct for inserted cassettes (occasional)
First check

Inspect cassette condition and display

Test point
Control handle display
Expected
Correct cassette format displayed; no physical damage to cassettes

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Cassette sizing and positioning

Verify bucky cassette detection, size display, and positioning tolerance after installation or programming changes

Ranked causes

  1. 1. Cassette not detected or incorrect size displayed (common)
  2. 2. Incorrect programming or offset settings (common)
  3. 3. Cassette positioning outside acceptable tolerance (occasional)
First check

Insert cassette centered and verify size display

Test point
control handle display
Expected
correct format

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ceiling suspension does not track table or wallstand movement

Ceiling suspension fails to follow bucky table or wallstand position changes; SID not displayed correctly on control handle

Ranked causes

  1. 1. SID sensor voltage supply missing or faulty (common)
  2. 2. SID sensor current output zero or at maximum (common)
  3. 3. Height sensor electronic part defective (occasional)
First check

Verify 24V supply presence

Test point
UE2 X1:10 via UZ X21:10/X22:1
Expected
24VDC
Tools
Voltmeter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Selftest fails on BuckyDiagnost system high voltage

System does not complete selftest sequence; red/yellow/green LEDs indicate incomplete initialization during power-on

Ranked causes

  1. 1. CAN bus power supply 15V missing or insufficient (common)
  2. 2. CAN bus cable disconnected or wires damaged (common)
  3. 3. SZ1 board power supply 5V or 24V fails (occasional)
First check

Measure 15V CAN bus power supply

Test point
SZ1 X2:9 and EZ X43:9 to GND
Expected
15V

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tracking function not working

Ceiling suspension does not follow table or wallstand movement; SID display may not update correctly on control handle.

Ranked causes

  1. 1. Table height sensor malfunction or cable fault (common)
  2. 2. Wallstand height sensor malfunction or cable fault (common)
  3. 3. Ceiling suspension height sensor malfunction or cable fault (common)
First check

Measure table height sensor current

Test point
SZ1 X4
Expected
5–18 mA (5 mA at lowest position, 18 mA at uppermost), 5–18 mA range
Tools
Service PC with VT100, analog display

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wallstand height tracking not functioning

Ceiling suspension fails to track wallstand bucky up/down movement; height sensor output suspect

Ranked causes

  1. 1. Wallstand height sensor current output zero or shorted (common)
  2. 2. Wallstand height sensor cable damaged (occasional)
  3. 3. Control board VN1 defective or disconnected (occasional)
First check

Measure wallstand height sensor current

Test point
SZ1 X5 at VN1
Expected
5–18 mA (5 mA at lowest, 18 mA at uppermost), 5–18 mA
Tools
Current meter or service PC with VT100

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bucky carriage and cassette selftest fails caution

Bucky carriage does not move to tomo centre position; cassette carriage does not move to both sides and center position; tomography blocked after 2 test runs

Ranked causes

  1. 1. Bucky or cassette carriage movement restricted (common)
  2. 2. Special circuit failure (occasional)
First check

Verify bucky carriage movement

Test point
Bucky carriage
Expected
Moves freely to tomo centre position

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bucky/Tomo unit selftest fails patient safety

Bucky carriage does not move to tomo centre position or cassette carriage fails to move side-to-side and centre

Ranked causes

  1. 1. Restricted movement of tray (common)
  2. 2. Special circuit failure (occasional)
First check

Inspect bucky carriage and cassette carriage for obstructions

Test point
Carriage movement paths
Expected
Free movement, no mechanical obstruction

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette format not displayed after tray insertion

Control handle display does not show format when cassette is inserted and tray pulled out slightly

Ranked causes

  1. 1. Cassette present switch malfunction (common)
  2. 2. Horizontal encoder plate misaligned (occasional)
First check

Insert cassette and pull tray out 1.5-2.5 mm, observe display

Test point
Control handle display
Expected
Format still displayed

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette tray levers bent or stiff caution

Levers on cassette tray are visibly bent or do not move easily and smoothly

Ranked causes

  1. 1. Levers mechanically damaged or misaligned (common)
  2. 2. Loose screws causing stiffness (occasional)
First check

Visually inspect cassette tray levers for bending

Test point
Cassette tray levers
Expected
not bent

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control desk displays failure during selftest

Control desk C fails to display normal startup sequence; displays error instead

Ranked causes

  1. 1. Mains supply failure (common)
  2. 2. Circuit breaker failure (common)
First check

Verify mains power to control desk

Test point
Control desk power input
Expected
Mains voltage present

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control desk selftest fails

Control desk displays a failure message; main display does not indicate selftest running or displays are not illuminated

Ranked causes

  1. 1. Mains supply failure (common)
  2. 2. Circuit breaker tripped (common)
First check

Verify mains supply to control desk

Test point
Control desk power input
Expected
Mains voltage present

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Encoder signal levels out of specification

Measured voltage or track signals do not meet expected values for cassette size and position.

Ranked causes

  1. 1. Cassette positioned incorrectly or incompletely (common)
  2. 2. Encoder disk or plate misalignment (occasional)
First check

Measure voltage at VA2 X2:5/6

Test point
VA2 X2:5/6
Expected
Within ±0.3V of table value for cassette size, ±0.3V

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Selftest fails on BuckyDiagnost CS Tomo Unit

LEDs on drive boards indicate selftest running but does not complete successfully

Ranked causes

  1. 1. CAN bus failure (common)
  2. 2. CAN bus wiring disconnected or damaged (occasional)
First check

Measure 15V power supply voltage on CAN bus

Test point
SZ1 X2:9 and EZ X43:9 to GND
Expected
15V

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Selftest fails on BuckyDiagnost TH2

Red LED 1 indicates selftest in progress but does not complete; system does not reach ready state

Ranked causes

  1. 1. Power supply 5V or 24V fails (common)
  2. 2. SZ1 board failure (occasional)
First check

Measure power supply voltages on SZ1 board

Test point
SZ1 board X6 (24V) and internal 5V rail
Expected
24V ±tolerance; 5V ±tolerance

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor track output voltage below threshold

Sensor track voltage does not reach required high level when 43 cm cassette is inserted in vertical and horizontal directions

Ranked causes

  1. 1. Sensor malfunction (common)
  2. 2. Incorrect distance between sensor and encoder plate (occasional)
First check

Measure voltage on every track of every sensor with largest 43 cm cassette

Test point
X20/X21/X22 pins 3,4,5,6 (pink, white, grey, brown) to Gnd
Expected
All 4 tracks are "H", high-level >13.0 V, >13.0 V
Tools
Voltmeter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table height tracking not functioning

Ceiling suspension does not follow table up/down movement; SID not displayed on control handle

Ranked causes

  1. 1. Height sensor current output zero (common)
  2. 2. Height sensor stuck at maximum current (occasional)
First check

Measure table height sensor current

Test point
SZ1 X4
Expected
5–18 mA (5 mA at lowest, 18 mA at uppermost), 5–18 mA
Tools
Current meter or service PC with VT100

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wallstand height potentiometer misaligned or disengaged caution

Wallstand bucky position not accurately tracked; potentiometer has lost engagement with toothed belt or is stuck

Ranked causes

  1. 1. Height potentiometer disengaged from toothed belt (occasional)
  2. 2. Height potentiometer mechanical wear or internal stop interference (rare)
First check

Verify potentiometer engagement during full range motion

Test point
Bucky frame toothed belt and potentiometer assembly
Expected
Potentiometer moves smoothly and is engaged throughout full up/down range

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bucky controller board malfunction

SZ1 board fails to initialize properly; W2 battery switch may not be in ON position on replacement board

Ranked causes

  1. 1. W2 battery switch not set to ON on replacement SZ1 board (occasional)
First check

Inspect W2 battery switch on replacement SZ1 board

Test point
W2 switch on SZ1 board
Expected
W2 in ON position before board installation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

BuckyDiagnost TH2 selftest LED sequence incomplete high voltage

Red LED 1 does not turn off after selftest completes; yellow LED 2 does not flash; green LED 3 does not turn on

Ranked causes

  1. 1. SZ1 board 5V or 24V power supply failure (common)
First check

Measure 5V and 24V supplies on SZ1 board with oscillograph

Test point
SZ1 X6 power connector
Expected
5V stable and 24V stable with no ripple

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bucky exposure delay radiation

Set exposure delay between grid magnet dropout and X-ray release to prevent coincidence with grid reversal point during short-time exposures. Involves measuring delay time and performing test exposures.

Ranked causes

  1. 1. Exposure delay requires adjustment for different generator types (common)
First check

Measure delay time with oscilloscope

Test point
VA1 K2:3 and K2:4
Expected
120ms to 150ms after grid magnet dropout (varies by generator type)
Tools
Oscilloscope

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Bucky exposure delay time radiation

Exposure must occur 120–150 ms after grid magnet dropout to prevent coincidence with grid reversal point during short-time exposures. Adjustment via potentiometer VA1 P1; verification by oscilloscope measurement and test film exposures.

Ranked causes

  1. 1. Delay time requires adjustment per generator type (common)
First check

Measure delay time with oscilloscope

Test point
VA1–K2:3 and K2:4
Expected
Between 21 (XGRDEX) and 20 (RQXGEX) signals
Tools
Oscilloscope

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: cassette present switch adjustment

Spring plate movement verification when cassette tray is inserted

Ranked causes

  1. 1. Spring plate out of adjustment (unknown)
First check

Measure spring plate movement when cassette tray is inserted

Test point
spring plate
Expected
0.8 ... 1.2 mm, 0.8 to 1.2 mm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Cassette sizing and offset

Verify bucky unit cassette detection, size display, and receptor offset settings match actual cassette dimensions and positioning after installation or recalibration.

Ranked causes

  1. 1. Incorrect cassette size displayed or offset not calibrated (common)
First check

Inspect cassettes for damage

Test point
Physical cassette examination
Expected
Cassettes undamaged and in working condition

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: exposure delay time radiation

Exposure must occur 120–150 ms after grid magnet dropout to prevent release coinciding with grid reversal during short exposures. Adjustment required due to different generator delay times.

Ranked causes

  1. 1. Generator-dependent exposure delay time variance (common)
First check

Measure delay time between XGRDEX and RQXGEX signals with oscilloscope

Test point
VA1–K2:3 and K2:4
Expected
120–150 ms after grid magnet dropout
Tools
Oscilloscope

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Horizontal encoder plate distance adjustment caution

Distance between horizontal encoder plate and sensor must be set to specification using leveling shims

Ranked causes

  1. 1. Encoder plate distance out of specification (unknown)
First check

Measure distance between horizontal encoder plate and sensor

Test point
horizontal encoder plate to sensor
Expected
1.0 mm ± 0.5 mm, ±0.5 mm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Template alignment for 5" x 7" cassette

Test procedure to verify proper template centering and alignment for 5" x 7" cassette size in Bucky unit

Ranked causes

  1. 1. Template misalignment or incorrect positioning (unknown)
First check

Insert template for 5" x 7" cassette size and verify centering

Test point
Template alignment mark
Expected
Template centered with respect to mark

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Template alignment for 9.5" x 9.5" cassette

Test procedure to verify proper template centering and alignment for 9.5" x 9.5" cassette size in Bucky unit

Ranked causes

  1. 1. Template misalignment or incorrect positioning (unknown)
First check

Insert template for 9.5" x 9.5" cassette size and verify centering

Test point
Template alignment mark
Expected
Template centered with respect to mark

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tomo center position switch

Position the ceiling suspension at system center and adjust the switching cam to touch only the respective switch

Ranked causes

  1. 1. Switch plate requires positioning with respect to switching cam (common)
First check

Position ceiling suspension at system center

Test point
Catch pins and catch plates
Expected
Pins lock into catch plates

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tomo drive centering caution

Verify toothed belt runs parallel to rails along full longitudinal range; adjust motor unit position if necessary

Ranked causes

  1. 1. Tomo motor unit misalignment causes belt to run non-parallel to rails (occasional)
First check

Move ceiling suspension through full longitudinal range

Test point
Toothed belt relative to rails 1
Expected
Belt runs parallel to rails throughout movement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tomo drive centering check

Verify toothed belt runs parallel to rails across full longitudinal range

Ranked causes

  1. 1. Tomo motor unit misaligned in transverse rail (common)
First check

Inspect belt tracking throughout longitudinal range

Test point
Toothed belt and rails 1
Expected
Belt runs parallel to rails

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tomo force adjustment

Verify and adjust tomo suspension pulling force to specified ranges with and without servo

Ranked causes

  1. 1. Toothed belt tension incorrect (common)
First check

Measure pulling force with system OFF

Test point
Control handle
Expected
30 ... 40N, 30-40N
Tools
Spring balance

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tomo safety end position switch

Position ceiling suspension at tomo safety end position and adjust switch plate to ensure cam contacts only the respective switch

Ranked causes

  1. 1. Switch plate requires positioning to cam at safety end position (common)
First check

Position ceiling suspension at tomo safety end position

Test point
Switch and cam contact
Expected
Switch 1 touches cam 2

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Verify cassette availability before recalibration

Check that cassettes are functional and of correct sizes prior to performing bucky recalibration procedures

Ranked causes

  1. 1. Cassettes damaged or bucky not functioning correctly (unknown)
First check

Verify analogue outputs are correct for inserted cassettes

Test point
Bucky analogue outputs
Expected
correct for inserted cassettes

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette format not displayed during Z-direction movement

Control handle display does not maintain format display when tray is moved in Z-direction with cassette inserted

Ranked causes

  1. 1. Horizontal encoder-plate misaligned or damaged (common)
First check

Move tray in Z-direction with cassette inserted and format displayed

Test point
Control handle display
Expected
Format still displayed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette size indicator misalignment caution

9.5 inch cassette size lines not visible at top of alignment triangle when levers touch cassette

Ranked causes

  1. 1. Levers not properly positioned for 9.5 inch cassette (common)
First check

Insert 9.5 inch cassette with template and observe alignment pattern

Test point
Cassette tray alignment triangles
Expected
The 9.5" lines shall be visible at the top of the triangle when levers touch cassette

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette size not recognized correctly

System fails to recognize inserted cassette format; test of 10" x 12" cassette size fails

Ranked causes

  1. 1. Bumper distance incorrect (common)
First check

Verify bumper spacing distance

Test point
Distance between bumpers
Expected
144 ... 145 mm, 144-145 mm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette size not recognized during insertion

Cassette format is not detected when cassette tray is inserted into bucky frame

Ranked causes

  1. 1. Switch point misaligned - cassette size recognition switch not triggering at correct position (common)
First check

Verify cassette present switch is closed

Test point
cassette present switch
Expected
closed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette size not recognized horizontally

System fails to recognize cassette size when inserted horizontally; displayed value on control handle does not show correct size

Ranked causes

  1. 1. Cassette positioned outside tolerance range (common)
First check

Insert 9.5" x 12" cassette horizontally

Test point
Bucky unit cassette slot
Expected
Cassette recognizes as 9.5" x 12" on control handle, ±2 mm out of center

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette size not recognized in eccentric position

Cassette not detected when placed off-center in Bucky unit

Ranked causes

  1. 1. Non-12 inch cassette in eccentric position (common)
First check

Test cassette detection in eccentric position with all available site cassettes

Test point
control handle display
Expected
12 inch cassettes detected; others may not register

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette size not recognized vertically

9.5 x 9.5 inch cassette size not displayed correctly on control handle when inserted vertically

Ranked causes

  1. 1. Cassette inserted outside vertical tolerance range (common)
First check

Insert 9.5 x 9.5 inch vertical cassette and verify size recognition

Test point
control handle display
Expected
9.5 x 9.5 inch size recognized, ±4 mm out of center

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette size not recognized with template

Template fails to detect cassette sizes in both directions; sizes should be 9.5×12", 10×12", 9.5×9.5", 7×5"

Ranked causes

  1. 1. Cassette not positioned within tolerance in eccentric position (common)
First check

Cassette size recognition with template in both directions

Test point
Template
Expected
9.5×12", 10×12", 9.5×9.5", 7×5" recognized

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cassette support plate conversion from left to right version caution

Reconfiguring the cassette support plate to operate on the opposite side, including cam rotation and axis relocation

Ranked causes

  1. 1. Cassette support plate requires directional conversion (unknown)
First check

Verify torsion springs are correctly positioned after axis relocation

Test point
Torsion springs
Expected
Correct position

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator selftest fails caution

Red LEDs on LA2 and LA4 boards are not OFF; shutters do not open/close; ruler does not roll up or stop correctly

Ranked causes

  1. 2. Power supply 5V/24V failure (occasional)
First check

Verify red LEDs on LA2 and LA4 boards

Test point
LA2 and LA4 boards
Expected
Red LEDs OFF after selftest

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control handle selftest fails

Control handle displays do not illuminate; main display does not indicate selftest running; red LED under cover not flashing

Ranked causes

  1. 2. Circuit breaker tripped (common)
First check

Verify mains supply to control handle

Test point
Control handle power input
Expected
Mains voltage present

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exposure not possible radiation

X-ray exposure function fails to initiate.

Ranked causes

  1. 1. Exposure subsystem fault (unknown)
First check

Verify exposure subsystem status

Test point
Status LEDs and CAN bus controllers
Expected
Exposure ready for operation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator delay time verification radiation

Verify that the bucky delay time matches the specific generator type's delay specification. Different generator models require different delay ranges (3–60 ms).

Ranked causes

  1. 1. Generator type mismatch with bucky delay setting (occasional)
First check

Verify generator type installed on system

Test point
Generator identification label
Expected
One of: OPTIMUS, SCP, MEDIO CP/CPH, MCM, MM 80/85, MM 100/150, SM 40/70, SM 80/100, MEDIO 51

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator ready sequence not completing

Red status LED does not light off; relay EN K1 does not switch on; READY lamp on control desk display does not light

Ranked causes

  1. 1. External safety switches not ready (door contact, thermal contact, or tube overloaded) (common)
First check

Verify all external safety switches in ready state

Test point
Unit ready; door contact closed; thermal tube contact closed; tube not overloaded
Expected
All switches ready

Grid release switch adjustment caution

Verify and adjust the grid release switch position and magnet brake distance for proper operation

Ranked causes

  1. 1. Grid release switch requires adjustment or verification (unknown)
First check

Measure distance between holder and magnet brake after powering on

Test point
Between holder (1) and magnet brake (2)
Expected
≤12 mm, ≤12 mm

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Grid release switch requires adjustment

Magnet brake distance and switch position must be verified and adjusted for proper operation

Ranked causes

  1. 1. Misaligned holder or incorrect magnet brake distance (unknown)
First check

Verify holder parallelism with respect to magnet brake

Test point
holder and magnet brake
Expected
parallel alignment

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Horizontal size recognition tolerance failure

Cassette size not recognized within acceptable horizontal centering tolerance; displayed value on control handle incorrect

Ranked causes

  1. 1. Cassette positioned outside ±2 mm horizontal tolerance (common)
First check

Insert 9.5×12" cassette horizontally

Test point
Cassette position center
Expected
Cassette recognized; positioning display within ±2 mm, ±2 mm out of center

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect measurement display for film size on VA2

Film size display reads incorrectly due to inch/metric programming setting being in wrong position. 11x14" displays as 12x14", 10x12" as 9.5x12", or 9.5x12" as 10x12"

Ranked causes

  1. 1. S1.1 switch set to incorrect position (inch/metric) (common)
First check

Verify current S1.1 switch position

Test point
S1.1 switch on VA2
Expected
OFF for inch mode or ON for metric mode

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Selftest fails or does not complete

BuckyDiagnost system does not run through selftest after power-on; status LEDs not flashing after successful selftest.

Ranked causes

  1. 1. Controller communication failure or firmware issue (unknown)

Selftest failure or yellow LED flashing continuously

System completes power-on but status LEDs flash yellow indicating busy CAN bus communication or controllers not responding.

Ranked causes

  1. 1. CAN bus communication fault or controller unresponsive (common)
First check

Observe status LED behavior during and after selftest

Test point
Bucky controller board status LEDs (red, yellow, green)
Expected
LEDs flashing during selftest; steady after selftest completes

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Sensor contamination or damage
  • Sensor supply voltage out of range
  • Sensor-encoder plate distance incorrect
  • System does not indicate ready or exposure cannot be released
  • System ready indication fails
  • Test exposure for delay time validation
  • Tomo collimator angle not defined
  • Tomo collimator height not defined
  • Tomo collimator height too high
  • Tomo collimator height too low
  • Tomo collimator sync position not defined
  • Tomo servo relay stuck
  • Tomo table top move failure
  • Tomography not possible
  • Torsion springs replacement
  • Torsion springs require replacement
  • Tracking not possible
  • Unrecognized cassette format detection
  • Verify cassette centric position
  • Vertical short/long sensor replacement
  • Vertical size recognition tolerance failure
  • Zero voltage output at X15/X16 connectors

22 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0200Tomo node state unknown or Tomo drive internal errorTomography run initialization or operationerror
0201Tomo drive CAN eventCAN bus communication eventwarning
0202Tomo test relay stuckTest relay K4 on SZ1 remains activatederror
0203Tomo enable X relay activeRelay K3 on SZ1 activation during starting phasewarning
0204Tomo test relay deadTest relay K4 fails to switcherror
0205Tomo register device changedDifferent RGDV selected during tomography runwarning
0206Unknown error code 0206Unknownunknown
0207Tomo safety relay openedSafety relay K1 on SZ1 opens unexpectedlycritical
0208Tomo controller timeoutTomography sequence exceeds time limiterror
0209Tomo master/Alpha/Tray drive downTomo drive unit cannot communicate via CAN buscritical
0210Tomo state machine crashTomography control software state errorcritical
0211Tomo not activatedTomography drive not energized before runerror
0212Tomo not preparedTray/CS/Alpha fails to reach start positionerror
0213Tomo X-ray stopped during runX-ray generator stops during tomographyerror
0214Tomo exposure button 2 not pressedExposure release switch RQ_SN_X not held throughout exposureinfo
0215Tomo safety relay stuckSafety relay K1 state not recognizederror
0216Tomo run abort reason unknownTomography sequence aborted without error codewarning
0217Tomo run abortedTomography sequence terminated by system or operatorwarning
0218Tomo controller crashTomography controller internal failurecritical
0219Tomo servo relay stuck (UE2 K1 not functioning)Relay UE2 K1 recognized by bucky controller but relay function not respondingerror
0220Tomo table top move failureTomography table top does not move to expected positionerror
0250Tomo collimation system transfer not definedPosition switch or height sensor caused unexpected function during tomo runwarning
0251Tomo collimation system height not definedHeight sensor signal absent or out of expected range during tomo runwarning
0252Tomo collimation system height too lowHeight sensor reading below minimum threshold during tomo runwarning
0253Tomo collimation system height too highHeight sensor reading above maximum threshold during tomo runwarning
0254Tomo collimation system beta angle not definedSwitch UFB S1 signal absent or threshold not met during tomo runwarning
0255Tomo tbl hght too lowTomography table height sensor or position switch fault during tomo runwarning
0256Tomo tbl hght too highTomography table height too high during tomo runwarning
0257Tomo cs beta not defTomography cs beta parameter not definedwarning
0258Tomo cs beta not defTomography cs beta parameter not definedwarning
0300Error log initBucky controller firmware changed, battery not connected at W2, or SZ1 defectiveerror
0301Servicedata initService data initialization failureerror
0302Servicedata specialService data special parameter issueerror
0303Servicedata paramService data parameter issueerror
0304Servicedata unconfigService data unconfigurederror
0400CAN node disconnectCAN node becomes disconnected from the buserror
0401CAN node start failCAN node fails to starterror
0402CAN node guarding failCAN node guard mechanism failureerror
0403CAN node unknownUnknown CAN node detectedwarning
0404CAN node prepare failCAN node preparation failederror
0405CAN node rx queue failCAN receive queue fails or overflowserror
0406CAN node CAN overrunCAN bus overrun condition detectederror
0407CAN node CAN bus offCAN bus enters off stateerror
0408CAN node err status bitError status bit set on CAN nodeerror
0409CAN node tx queue overCAN transmit queue overflowserror
0410CAN local node disconLocal CAN node becomes disconnectederror
0411CAN node stoppedCAN node operation stoppederror
0412CAN node guard againCAN node guarding recurserror
0413CAN node connect failCAN node fails to connecterror
0414CAN slave node disconCAN slave node becomes disconnectederror
0415CAN node startedCAN node started successfullyinfo
0416CAN msg def failCAN message definition errorerror
0700System invalid register deviceInvalid generator configurationerror
0701System register device not 2 roomsGenerator and bucky controller register device mismatcherror
0702System SID invalidSID calculation failserror
0703System raster not releasedGrid start and timer sequence failureerror
0800Collimator filter invalidBucky controller receives unknown filter info from collimatorerror
0801Collimator fieldsize invalidBucky controller receives unknown fieldsize info from collimatorerror
0802Collimator ruler errorBucky controller receives unknown ruler value from collimatorerror
0803Collimator no local APRsAll local APRs on bucky controller are invaliderror
0900CS long invalidLongitudinal position sensing failure.error
0901CS trans invalidTransverse position sensing failure.error
0902CS hght def invalidHeight definition position sensing failure.error
0903CS tomo hght mismatchTomography height adjustment mismatch detected.error
0904CS hght invalidHeight position sensing failure.error
0905CS hght sens short cctHeight sensor short circuit detected.error
0906CS hght sens open cctHeight sensor open circuit detected.error
0907Table hght invalidTable height position sensing failure.error
0908Table hght sens shortTable height sensor short circuit detected.error
0909Table hght sens openTable height sensor open circuit detected.error
0910CS alpha invalidCS rotation alpha position sensing failure.error
0911CS beta invalidCS rotation beta position sensing failure.error
0912WS hght invalidWallstand height position sensing failure.error
0913WS hght sens short cctWallstand height sensor short circuit detected.error
0914WS hght sens open cctWallstand height sensor open circuit detected.error
0915WS hght mismatchWallstand height position mismatch on tilted bucky unit.error
0916WS tilt invalidWallstand tilt position sensing failure.error
1000Bst watchdogBucky controller self-test failure on watchdog timercritical
1001Bst ramPower-up RAM test failurecritical
1002Bst eprom low byteEPROM low byte self-test failurecritical
1003Bst eprom high byteEPROM high byte self-test failurecritical
1004Bst uartUART self-test failure, V24 port failscritical
1005Bst CAN controllerCAN controller self-test failurecritical
1006Bst corrupt error logError log corruption detected during self-testcritical
1007Bst servicedata upgradeService data upgrade detected during self-testwarning
1008Bst servicedata downgr.Service data downgrade detected during self-testwarning
1009Bst two should be setConfiguration error in self-testerror
1010Bst room two CAN masterRoom 2 CAN master configuration error detectederror
1011Bst old world also setOld world configuration error detectederror
1012Bst info level cancelLogging level configuration errorinfo
1013Bst dev info levelDevelopment info level configuration errorinfo
1014Bst CAN master overrideCAN master override condition detectedunknown
1015Bst not CAN masterSystem not configured as CAN masterunknown
1016Bst two rooms with URFInvalid configuration: two rooms with URFunknown
1017Bst OWI with URFInvalid OWI with URF configuration detectedunknown
1018Bst master with URFInvalid master with URF configuration detectedunknown
1019Bst selftest failedOverall self-test failure on bucky controllercritical
1100CH general - Unknown control handle connectedSystem detects unrecognized control handleerror
1101CH test relay stuckRelay on control handle self-test failscritical
1102CH test watchdogWatchdog timer test failsinfo
1103CH test key stuckControl handle key detected as stuckcritical
1104CH test RAMRAM self-test fails on control handlecritical
1105CH test EPROMEPROM self-test fails on control handlecritical
1106CH test CANCAN controller self-test failscritical
1200Bucky controller processor errorWatchdog reset by softwareerror
1300Tracking timeout at calibrationCalibration sequence timeouterror
1301Tracking timeout at velocityVelocity measurement timeouterror
1302Tracking timeout at positionPosition measurement timeouterror
1303Tracking height inconsistentHeight measurement inconsistency detectederror
1304Tracking reached position inconsistencyPosition reached but inconsistenterror
1305Tracking CSD error messageCSD communication errorerror
Cassette format not recognizedCassette size does not match any entry in the centric position tableEncoder signals do not correspond to any standard cassette formatwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI