Philips BuckyDiagnost: troubleshooting and error codes
The ServiceForge knowledge base holds 82 documented symptoms for the Philips BuckyDiagnost, with 136 ranked causes, 148 diagnostic checks with test points and expected values, and 112 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power on high voltage
Yellow LED on EN 100 is off, relay EN K2 does not switch over
Ranked causes
- 1. Mains supply failure (common)
- 2. Circuit breaker EN F1 not released (common)
- 3. Circuit breaker EN F2 not released (common)
- 4. Low-voltage supply failure (occasional)
- 5. Filament circuit failure (occasional)
- 6. Tube extension failure (occasional)
- 7. Rotor control failure (occasional)
- 8. External current consumer issue (occasional)
- 9. EN 100 board failure (rare)
- 10. Emergency-OFF circuit open (occasional)
- 11. Operating desk not connected (occasional)
- 12. Safety circuit open (occasional)
Verify mains supply present
- Test point
- System input
- Expected
- Mains voltage present
5 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator selftest fails
Red LED on Z rack is ON continuously; Central Unit, Voltage Control, Current Control, Basic Interface, or Rotor Control module not responding
Ranked causes
- 1. Central Unit EZ139 fails (occasional)
- 2. Voltage Control EZ130 fails (occasional)
- 3. Current Control EZ119 fails (occasional)
- 4. Basic Interface EZ150 fails (occasional)
- 5. Rotor Control EY fails (occasional)
Verify Central Unit EZ139 CAN communication
- Test point
- EZ139 module
- Expected
- CAN communication established
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator selftest fails with red LED continuously ON high voltage
During power-on, one red LED on Z rack should flash briefly; continuous red LED indicates selftest failure
Ranked causes
- 1. Central unit EZ139 fails (unknown)
- 2. Voltage control EZ130 fails (unknown)
- 3. Current control EZ119 fails (unknown)
- 4. Basic interface EZ150 fails (unknown)
- 5. Rotor control EY fails (unknown)
Observe status LEDs on Z rack during startup
- Test point
- Z rack status LEDs
- Expected
- All LEDs flash, then one red LED flashes briefly
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not power on high voltage
Yellow LED on EN 100 does not illuminate; system fails to start selftest sequence
Ranked causes
- 1. Mains supply failure (common)
- 2. Circuit breaker failure (common)
- 3. Low-voltage supply failure (occasional)
- 4. Safety circuit open (occasional)
- 5. Emergency-OFF circuit open (occasional)
Verify mains power supply
- Test point
- System mains input
- Expected
- Mains voltage present
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tomography run fails radiation
Tomography procedure does not complete successfully; system may not execute the programmed tomography sequence
Ranked causes
- 1. Ceiling Suspension (CS) assembly malfunction (unknown)
- 2. CS longitudinal Drive failure (unknown)
- 3. CS alpha Drive failure (unknown)
- 4. Table tomo Assemblies malfunction (unknown)
- 5. Tomo Controller failure (unknown)
Verify tomography run using VT100 program on service PC
- Test point
- Service PC display
- Expected
- All steps during tomography run are visible in VT100 output
- Tools
- Service PC, VT100 program
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bucky unit programming and cassette sizing
Setup and verification of bucky controller settings, display units, receptor offsets, and cassette size calibration via XSCOPE software
Ranked causes
- 1. Incorrect display unit setting (metric/imperial mismatch) (common)
- 2. Receptor offset out of specification (common)
- 3. Cassette size settings mismatch (common)
- 4. Cassette sizing calibration not performed after new bucky installation (occasional)
Verify VA2-board display unit setting
- Test point
- VA2-board switch
- Expected
- Set to inch or metric
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bucky unit cassette sizing radiation
Ensure cassette size display and positioning are correct after bucky installation or when cassette size detection fails
Ranked causes
- 1. Damaged cassettes or incorrect cassette size display (common)
- 2. Incorrect bucky unit programming or settings (common)
- 3. Analogue outputs not correct for inserted cassettes (occasional)
Inspect cassette condition and display
- Test point
- Control handle display
- Expected
- Correct cassette format displayed; no physical damage to cassettes
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cassette sizing and positioning
Verify bucky cassette detection, size display, and positioning tolerance after installation or programming changes
Ranked causes
- 1. Cassette not detected or incorrect size displayed (common)
- 2. Incorrect programming or offset settings (common)
- 3. Cassette positioning outside acceptable tolerance (occasional)
Insert cassette centered and verify size display
- Test point
- control handle display
- Expected
- correct format
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ceiling suspension does not track table or wallstand movement
Ceiling suspension fails to follow bucky table or wallstand position changes; SID not displayed correctly on control handle
Ranked causes
- 1. SID sensor voltage supply missing or faulty (common)
- 2. SID sensor current output zero or at maximum (common)
- 3. Height sensor electronic part defective (occasional)
Verify 24V supply presence
- Test point
- UE2 X1:10 via UZ X21:10/X22:1
- Expected
- 24VDC
- Tools
- Voltmeter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Selftest fails on BuckyDiagnost system high voltage
System does not complete selftest sequence; red/yellow/green LEDs indicate incomplete initialization during power-on
Ranked causes
- 1. CAN bus power supply 15V missing or insufficient (common)
- 2. CAN bus cable disconnected or wires damaged (common)
- 3. SZ1 board power supply 5V or 24V fails (occasional)
Measure 15V CAN bus power supply
- Test point
- SZ1 X2:9 and EZ X43:9 to GND
- Expected
- 15V
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tracking function not working
Ceiling suspension does not follow table or wallstand movement; SID display may not update correctly on control handle.
Ranked causes
- 1. Table height sensor malfunction or cable fault (common)
- 2. Wallstand height sensor malfunction or cable fault (common)
- 3. Ceiling suspension height sensor malfunction or cable fault (common)
Measure table height sensor current
- Test point
- SZ1 X4
- Expected
- 5–18 mA (5 mA at lowest position, 18 mA at uppermost), 5–18 mA range
- Tools
- Service PC with VT100, analog display
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wallstand height tracking not functioning
Ceiling suspension fails to track wallstand bucky up/down movement; height sensor output suspect
Ranked causes
- 1. Wallstand height sensor current output zero or shorted (common)
- 2. Wallstand height sensor cable damaged (occasional)
- 3. Control board VN1 defective or disconnected (occasional)
Measure wallstand height sensor current
- Test point
- SZ1 X5 at VN1
- Expected
- 5–18 mA (5 mA at lowest, 18 mA at uppermost), 5–18 mA
- Tools
- Current meter or service PC with VT100
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky carriage and cassette selftest fails caution
Bucky carriage does not move to tomo centre position; cassette carriage does not move to both sides and center position; tomography blocked after 2 test runs
Ranked causes
- 1. Bucky or cassette carriage movement restricted (common)
- 2. Special circuit failure (occasional)
Verify bucky carriage movement
- Test point
- Bucky carriage
- Expected
- Moves freely to tomo centre position
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky/Tomo unit selftest fails patient safety
Bucky carriage does not move to tomo centre position or cassette carriage fails to move side-to-side and centre
Ranked causes
- 1. Restricted movement of tray (common)
- 2. Special circuit failure (occasional)
Inspect bucky carriage and cassette carriage for obstructions
- Test point
- Carriage movement paths
- Expected
- Free movement, no mechanical obstruction
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette format not displayed after tray insertion
Control handle display does not show format when cassette is inserted and tray pulled out slightly
Ranked causes
- 1. Cassette present switch malfunction (common)
- 2. Horizontal encoder plate misaligned (occasional)
Insert cassette and pull tray out 1.5-2.5 mm, observe display
- Test point
- Control handle display
- Expected
- Format still displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette tray levers bent or stiff caution
Levers on cassette tray are visibly bent or do not move easily and smoothly
Ranked causes
- 1. Levers mechanically damaged or misaligned (common)
- 2. Loose screws causing stiffness (occasional)
Visually inspect cassette tray levers for bending
- Test point
- Cassette tray levers
- Expected
- not bent
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control desk displays failure during selftest
Control desk C fails to display normal startup sequence; displays error instead
Ranked causes
- 1. Mains supply failure (common)
- 2. Circuit breaker failure (common)
Verify mains power to control desk
- Test point
- Control desk power input
- Expected
- Mains voltage present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control desk selftest fails
Control desk displays a failure message; main display does not indicate selftest running or displays are not illuminated
Ranked causes
- 1. Mains supply failure (common)
- 2. Circuit breaker tripped (common)
Verify mains supply to control desk
- Test point
- Control desk power input
- Expected
- Mains voltage present
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Encoder signal levels out of specification
Measured voltage or track signals do not meet expected values for cassette size and position.
Ranked causes
- 1. Cassette positioned incorrectly or incompletely (common)
- 2. Encoder disk or plate misalignment (occasional)
Measure voltage at VA2 X2:5/6
- Test point
- VA2 X2:5/6
- Expected
- Within ±0.3V of table value for cassette size, ±0.3V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Selftest fails on BuckyDiagnost CS Tomo Unit
LEDs on drive boards indicate selftest running but does not complete successfully
Ranked causes
- 1. CAN bus failure (common)
- 2. CAN bus wiring disconnected or damaged (occasional)
Measure 15V power supply voltage on CAN bus
- Test point
- SZ1 X2:9 and EZ X43:9 to GND
- Expected
- 15V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Selftest fails on BuckyDiagnost TH2
Red LED 1 indicates selftest in progress but does not complete; system does not reach ready state
Ranked causes
- 1. Power supply 5V or 24V fails (common)
- 2. SZ1 board failure (occasional)
Measure power supply voltages on SZ1 board
- Test point
- SZ1 board X6 (24V) and internal 5V rail
- Expected
- 24V ±tolerance; 5V ±tolerance
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor track output voltage below threshold
Sensor track voltage does not reach required high level when 43 cm cassette is inserted in vertical and horizontal directions
Ranked causes
- 1. Sensor malfunction (common)
- 2. Incorrect distance between sensor and encoder plate (occasional)
Measure voltage on every track of every sensor with largest 43 cm cassette
- Test point
- X20/X21/X22 pins 3,4,5,6 (pink, white, grey, brown) to Gnd
- Expected
- All 4 tracks are "H", high-level >13.0 V, >13.0 V
- Tools
- Voltmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table height tracking not functioning
Ceiling suspension does not follow table up/down movement; SID not displayed on control handle
Ranked causes
- 1. Height sensor current output zero (common)
- 2. Height sensor stuck at maximum current (occasional)
Measure table height sensor current
- Test point
- SZ1 X4
- Expected
- 5–18 mA (5 mA at lowest, 18 mA at uppermost), 5–18 mA
- Tools
- Current meter or service PC with VT100
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wallstand height potentiometer misaligned or disengaged caution
Wallstand bucky position not accurately tracked; potentiometer has lost engagement with toothed belt or is stuck
Ranked causes
- 1. Height potentiometer disengaged from toothed belt (occasional)
- 2. Height potentiometer mechanical wear or internal stop interference (rare)
Verify potentiometer engagement during full range motion
- Test point
- Bucky frame toothed belt and potentiometer assembly
- Expected
- Potentiometer moves smoothly and is engaged throughout full up/down range
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky controller board malfunction
SZ1 board fails to initialize properly; W2 battery switch may not be in ON position on replacement board
Ranked causes
- 1. W2 battery switch not set to ON on replacement SZ1 board (occasional)
Inspect W2 battery switch on replacement SZ1 board
- Test point
- W2 switch on SZ1 board
- Expected
- W2 in ON position before board installation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
BuckyDiagnost TH2 selftest LED sequence incomplete high voltage
Red LED 1 does not turn off after selftest completes; yellow LED 2 does not flash; green LED 3 does not turn on
Ranked causes
- 1. SZ1 board 5V or 24V power supply failure (common)
Measure 5V and 24V supplies on SZ1 board with oscillograph
- Test point
- SZ1 X6 power connector
- Expected
- 5V stable and 24V stable with no ripple
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bucky exposure delay radiation
Set exposure delay between grid magnet dropout and X-ray release to prevent coincidence with grid reversal point during short-time exposures. Involves measuring delay time and performing test exposures.
Ranked causes
- 1. Exposure delay requires adjustment for different generator types (common)
Measure delay time with oscilloscope
- Test point
- VA1 K2:3 and K2:4
- Expected
- 120ms to 150ms after grid magnet dropout (varies by generator type)
- Tools
- Oscilloscope
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Bucky exposure delay time radiation
Exposure must occur 120–150 ms after grid magnet dropout to prevent coincidence with grid reversal point during short-time exposures. Adjustment via potentiometer VA1 P1; verification by oscilloscope measurement and test film exposures.
Ranked causes
- 1. Delay time requires adjustment per generator type (common)
Measure delay time with oscilloscope
- Test point
- VA1–K2:3 and K2:4
- Expected
- Between 21 (XGRDEX) and 20 (RQXGEX) signals
- Tools
- Oscilloscope
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: cassette present switch adjustment
Spring plate movement verification when cassette tray is inserted
Ranked causes
- 1. Spring plate out of adjustment (unknown)
Measure spring plate movement when cassette tray is inserted
- Test point
- spring plate
- Expected
- 0.8 ... 1.2 mm, 0.8 to 1.2 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cassette sizing and offset
Verify bucky unit cassette detection, size display, and receptor offset settings match actual cassette dimensions and positioning after installation or recalibration.
Ranked causes
- 1. Incorrect cassette size displayed or offset not calibrated (common)
Inspect cassettes for damage
- Test point
- Physical cassette examination
- Expected
- Cassettes undamaged and in working condition
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: exposure delay time radiation
Exposure must occur 120–150 ms after grid magnet dropout to prevent release coinciding with grid reversal during short exposures. Adjustment required due to different generator delay times.
Ranked causes
- 1. Generator-dependent exposure delay time variance (common)
Measure delay time between XGRDEX and RQXGEX signals with oscilloscope
- Test point
- VA1–K2:3 and K2:4
- Expected
- 120–150 ms after grid magnet dropout
- Tools
- Oscilloscope
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Horizontal encoder plate distance adjustment caution
Distance between horizontal encoder plate and sensor must be set to specification using leveling shims
Ranked causes
- 1. Encoder plate distance out of specification (unknown)
Measure distance between horizontal encoder plate and sensor
- Test point
- horizontal encoder plate to sensor
- Expected
- 1.0 mm ± 0.5 mm, ±0.5 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Template alignment for 5" x 7" cassette
Test procedure to verify proper template centering and alignment for 5" x 7" cassette size in Bucky unit
Ranked causes
- 1. Template misalignment or incorrect positioning (unknown)
Insert template for 5" x 7" cassette size and verify centering
- Test point
- Template alignment mark
- Expected
- Template centered with respect to mark
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Template alignment for 9.5" x 9.5" cassette
Test procedure to verify proper template centering and alignment for 9.5" x 9.5" cassette size in Bucky unit
Ranked causes
- 1. Template misalignment or incorrect positioning (unknown)
Insert template for 9.5" x 9.5" cassette size and verify centering
- Test point
- Template alignment mark
- Expected
- Template centered with respect to mark
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tomo center position switch
Position the ceiling suspension at system center and adjust the switching cam to touch only the respective switch
Ranked causes
- 1. Switch plate requires positioning with respect to switching cam (common)
Position ceiling suspension at system center
- Test point
- Catch pins and catch plates
- Expected
- Pins lock into catch plates
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tomo drive centering caution
Verify toothed belt runs parallel to rails along full longitudinal range; adjust motor unit position if necessary
Ranked causes
- 1. Tomo motor unit misalignment causes belt to run non-parallel to rails (occasional)
Move ceiling suspension through full longitudinal range
- Test point
- Toothed belt relative to rails 1
- Expected
- Belt runs parallel to rails throughout movement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tomo drive centering check
Verify toothed belt runs parallel to rails across full longitudinal range
Ranked causes
- 1. Tomo motor unit misaligned in transverse rail (common)
Inspect belt tracking throughout longitudinal range
- Test point
- Toothed belt and rails 1
- Expected
- Belt runs parallel to rails
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tomo force adjustment
Verify and adjust tomo suspension pulling force to specified ranges with and without servo
Ranked causes
- 1. Toothed belt tension incorrect (common)
Measure pulling force with system OFF
- Test point
- Control handle
- Expected
- 30 ... 40N, 30-40N
- Tools
- Spring balance
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tomo safety end position switch
Position ceiling suspension at tomo safety end position and adjust switch plate to ensure cam contacts only the respective switch
Ranked causes
- 1. Switch plate requires positioning to cam at safety end position (common)
Position ceiling suspension at tomo safety end position
- Test point
- Switch and cam contact
- Expected
- Switch 1 touches cam 2
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Verify cassette availability before recalibration
Check that cassettes are functional and of correct sizes prior to performing bucky recalibration procedures
Ranked causes
- 1. Cassettes damaged or bucky not functioning correctly (unknown)
Verify analogue outputs are correct for inserted cassettes
- Test point
- Bucky analogue outputs
- Expected
- correct for inserted cassettes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette format not displayed during Z-direction movement
Control handle display does not maintain format display when tray is moved in Z-direction with cassette inserted
Ranked causes
- 1. Horizontal encoder-plate misaligned or damaged (common)
Move tray in Z-direction with cassette inserted and format displayed
- Test point
- Control handle display
- Expected
- Format still displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette size indicator misalignment caution
9.5 inch cassette size lines not visible at top of alignment triangle when levers touch cassette
Ranked causes
- 1. Levers not properly positioned for 9.5 inch cassette (common)
Insert 9.5 inch cassette with template and observe alignment pattern
- Test point
- Cassette tray alignment triangles
- Expected
- The 9.5" lines shall be visible at the top of the triangle when levers touch cassette
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette size not recognized correctly
System fails to recognize inserted cassette format; test of 10" x 12" cassette size fails
Ranked causes
- 1. Bumper distance incorrect (common)
Verify bumper spacing distance
- Test point
- Distance between bumpers
- Expected
- 144 ... 145 mm, 144-145 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette size not recognized during insertion
Cassette format is not detected when cassette tray is inserted into bucky frame
Ranked causes
- 1. Switch point misaligned - cassette size recognition switch not triggering at correct position (common)
Verify cassette present switch is closed
- Test point
- cassette present switch
- Expected
- closed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette size not recognized horizontally
System fails to recognize cassette size when inserted horizontally; displayed value on control handle does not show correct size
Ranked causes
- 1. Cassette positioned outside tolerance range (common)
Insert 9.5" x 12" cassette horizontally
- Test point
- Bucky unit cassette slot
- Expected
- Cassette recognizes as 9.5" x 12" on control handle, ±2 mm out of center
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette size not recognized in eccentric position
Cassette not detected when placed off-center in Bucky unit
Ranked causes
- 1. Non-12 inch cassette in eccentric position (common)
Test cassette detection in eccentric position with all available site cassettes
- Test point
- control handle display
- Expected
- 12 inch cassettes detected; others may not register
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette size not recognized vertically
9.5 x 9.5 inch cassette size not displayed correctly on control handle when inserted vertically
Ranked causes
- 1. Cassette inserted outside vertical tolerance range (common)
Insert 9.5 x 9.5 inch vertical cassette and verify size recognition
- Test point
- control handle display
- Expected
- 9.5 x 9.5 inch size recognized, ±4 mm out of center
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette size not recognized with template
Template fails to detect cassette sizes in both directions; sizes should be 9.5×12", 10×12", 9.5×9.5", 7×5"
Ranked causes
- 1. Cassette not positioned within tolerance in eccentric position (common)
Cassette size recognition with template in both directions
- Test point
- Template
- Expected
- 9.5×12", 10×12", 9.5×9.5", 7×5" recognized
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette support plate conversion from left to right version caution
Reconfiguring the cassette support plate to operate on the opposite side, including cam rotation and axis relocation
Ranked causes
- 1. Cassette support plate requires directional conversion (unknown)
Verify torsion springs are correctly positioned after axis relocation
- Test point
- Torsion springs
- Expected
- Correct position
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator selftest fails caution
Red LEDs on LA2 and LA4 boards are not OFF; shutters do not open/close; ruler does not roll up or stop correctly
Ranked causes
- 2. Power supply 5V/24V failure (occasional)
Verify red LEDs on LA2 and LA4 boards
- Test point
- LA2 and LA4 boards
- Expected
- Red LEDs OFF after selftest
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control handle selftest fails
Control handle displays do not illuminate; main display does not indicate selftest running; red LED under cover not flashing
Ranked causes
- 2. Circuit breaker tripped (common)
Verify mains supply to control handle
- Test point
- Control handle power input
- Expected
- Mains voltage present
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exposure not possible radiation
X-ray exposure function fails to initiate.
Ranked causes
- 1. Exposure subsystem fault (unknown)
Verify exposure subsystem status
- Test point
- Status LEDs and CAN bus controllers
- Expected
- Exposure ready for operation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator delay time verification radiation
Verify that the bucky delay time matches the specific generator type's delay specification. Different generator models require different delay ranges (3–60 ms).
Ranked causes
- 1. Generator type mismatch with bucky delay setting (occasional)
Verify generator type installed on system
- Test point
- Generator identification label
- Expected
- One of: OPTIMUS, SCP, MEDIO CP/CPH, MCM, MM 80/85, MM 100/150, SM 40/70, SM 80/100, MEDIO 51
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator ready sequence not completing
Red status LED does not light off; relay EN K1 does not switch on; READY lamp on control desk display does not light
Ranked causes
- 1. External safety switches not ready (door contact, thermal contact, or tube overloaded) (common)
Verify all external safety switches in ready state
- Test point
- Unit ready; door contact closed; thermal tube contact closed; tube not overloaded
- Expected
- All switches ready
Grid release switch adjustment caution
Verify and adjust the grid release switch position and magnet brake distance for proper operation
Ranked causes
- 1. Grid release switch requires adjustment or verification (unknown)
Measure distance between holder and magnet brake after powering on
- Test point
- Between holder (1) and magnet brake (2)
- Expected
- ≤12 mm, ≤12 mm
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grid release switch requires adjustment
Magnet brake distance and switch position must be verified and adjusted for proper operation
Ranked causes
- 1. Misaligned holder or incorrect magnet brake distance (unknown)
Verify holder parallelism with respect to magnet brake
- Test point
- holder and magnet brake
- Expected
- parallel alignment
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Horizontal size recognition tolerance failure
Cassette size not recognized within acceptable horizontal centering tolerance; displayed value on control handle incorrect
Ranked causes
- 1. Cassette positioned outside ±2 mm horizontal tolerance (common)
Insert 9.5×12" cassette horizontally
- Test point
- Cassette position center
- Expected
- Cassette recognized; positioning display within ±2 mm, ±2 mm out of center
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect measurement display for film size on VA2
Film size display reads incorrectly due to inch/metric programming setting being in wrong position. 11x14" displays as 12x14", 10x12" as 9.5x12", or 9.5x12" as 10x12"
Ranked causes
- 1. S1.1 switch set to incorrect position (inch/metric) (common)
Verify current S1.1 switch position
- Test point
- S1.1 switch on VA2
- Expected
- OFF for inch mode or ON for metric mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Selftest fails or does not complete
BuckyDiagnost system does not run through selftest after power-on; status LEDs not flashing after successful selftest.
Ranked causes
- 1. Controller communication failure or firmware issue (unknown)
Selftest failure or yellow LED flashing continuously
System completes power-on but status LEDs flash yellow indicating busy CAN bus communication or controllers not responding.
Ranked causes
- 1. CAN bus communication fault or controller unresponsive (common)
Observe status LED behavior during and after selftest
- Test point
- Bucky controller board status LEDs (red, yellow, green)
- Expected
- LEDs flashing during selftest; steady after selftest completes
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Sensor contamination or damage
- Sensor supply voltage out of range
- Sensor-encoder plate distance incorrect
- System does not indicate ready or exposure cannot be released
- System ready indication fails
- Test exposure for delay time validation
- Tomo collimator angle not defined
- Tomo collimator height not defined
- Tomo collimator height too high
- Tomo collimator height too low
- Tomo collimator sync position not defined
- Tomo servo relay stuck
- Tomo table top move failure
- Tomography not possible
- Torsion springs replacement
- Torsion springs require replacement
- Tracking not possible
- Unrecognized cassette format detection
- Verify cassette centric position
- Vertical short/long sensor replacement
- Vertical size recognition tolerance failure
- Zero voltage output at X15/X16 connectors
22 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0200 | Tomo node state unknown or Tomo drive internal error | Tomography run initialization or operation | error |
0201 | Tomo drive CAN event | CAN bus communication event | warning |
0202 | Tomo test relay stuck | Test relay K4 on SZ1 remains activated | error |
0203 | Tomo enable X relay active | Relay K3 on SZ1 activation during starting phase | warning |
0204 | Tomo test relay dead | Test relay K4 fails to switch | error |
0205 | Tomo register device changed | Different RGDV selected during tomography run | warning |
0206 | Unknown error code 0206 | Unknown | unknown |
0207 | Tomo safety relay opened | Safety relay K1 on SZ1 opens unexpectedly | critical |
0208 | Tomo controller timeout | Tomography sequence exceeds time limit | error |
0209 | Tomo master/Alpha/Tray drive down | Tomo drive unit cannot communicate via CAN bus | critical |
0210 | Tomo state machine crash | Tomography control software state error | critical |
0211 | Tomo not activated | Tomography drive not energized before run | error |
0212 | Tomo not prepared | Tray/CS/Alpha fails to reach start position | error |
0213 | Tomo X-ray stopped during run | X-ray generator stops during tomography | error |
0214 | Tomo exposure button 2 not pressed | Exposure release switch RQ_SN_X not held throughout exposure | info |
0215 | Tomo safety relay stuck | Safety relay K1 state not recognized | error |
0216 | Tomo run abort reason unknown | Tomography sequence aborted without error code | warning |
0217 | Tomo run aborted | Tomography sequence terminated by system or operator | warning |
0218 | Tomo controller crash | Tomography controller internal failure | critical |
0219 | Tomo servo relay stuck (UE2 K1 not functioning) | Relay UE2 K1 recognized by bucky controller but relay function not responding | error |
0220 | Tomo table top move failure | Tomography table top does not move to expected position | error |
0250 | Tomo collimation system transfer not defined | Position switch or height sensor caused unexpected function during tomo run | warning |
0251 | Tomo collimation system height not defined | Height sensor signal absent or out of expected range during tomo run | warning |
0252 | Tomo collimation system height too low | Height sensor reading below minimum threshold during tomo run | warning |
0253 | Tomo collimation system height too high | Height sensor reading above maximum threshold during tomo run | warning |
0254 | Tomo collimation system beta angle not defined | Switch UFB S1 signal absent or threshold not met during tomo run | warning |
0255 | Tomo tbl hght too low | Tomography table height sensor or position switch fault during tomo run | warning |
0256 | Tomo tbl hght too high | Tomography table height too high during tomo run | warning |
0257 | Tomo cs beta not def | Tomography cs beta parameter not defined | warning |
0258 | Tomo cs beta not def | Tomography cs beta parameter not defined | warning |
0300 | Error log init | Bucky controller firmware changed, battery not connected at W2, or SZ1 defective | error |
0301 | Servicedata init | Service data initialization failure | error |
0302 | Servicedata special | Service data special parameter issue | error |
0303 | Servicedata param | Service data parameter issue | error |
0304 | Servicedata unconfig | Service data unconfigured | error |
0400 | CAN node disconnect | CAN node becomes disconnected from the bus | error |
0401 | CAN node start fail | CAN node fails to start | error |
0402 | CAN node guarding fail | CAN node guard mechanism failure | error |
0403 | CAN node unknown | Unknown CAN node detected | warning |
0404 | CAN node prepare fail | CAN node preparation failed | error |
0405 | CAN node rx queue fail | CAN receive queue fails or overflows | error |
0406 | CAN node CAN overrun | CAN bus overrun condition detected | error |
0407 | CAN node CAN bus off | CAN bus enters off state | error |
0408 | CAN node err status bit | Error status bit set on CAN node | error |
0409 | CAN node tx queue over | CAN transmit queue overflows | error |
0410 | CAN local node discon | Local CAN node becomes disconnected | error |
0411 | CAN node stopped | CAN node operation stopped | error |
0412 | CAN node guard again | CAN node guarding recurs | error |
0413 | CAN node connect fail | CAN node fails to connect | error |
0414 | CAN slave node discon | CAN slave node becomes disconnected | error |
0415 | CAN node started | CAN node started successfully | info |
0416 | CAN msg def fail | CAN message definition error | error |
0700 | System invalid register device | Invalid generator configuration | error |
0701 | System register device not 2 rooms | Generator and bucky controller register device mismatch | error |
0702 | System SID invalid | SID calculation fails | error |
0703 | System raster not released | Grid start and timer sequence failure | error |
0800 | Collimator filter invalid | Bucky controller receives unknown filter info from collimator | error |
0801 | Collimator fieldsize invalid | Bucky controller receives unknown fieldsize info from collimator | error |
0802 | Collimator ruler error | Bucky controller receives unknown ruler value from collimator | error |
0803 | Collimator no local APRs | All local APRs on bucky controller are invalid | error |
0900 | CS long invalid | Longitudinal position sensing failure. | error |
0901 | CS trans invalid | Transverse position sensing failure. | error |
0902 | CS hght def invalid | Height definition position sensing failure. | error |
0903 | CS tomo hght mismatch | Tomography height adjustment mismatch detected. | error |
0904 | CS hght invalid | Height position sensing failure. | error |
0905 | CS hght sens short cct | Height sensor short circuit detected. | error |
0906 | CS hght sens open cct | Height sensor open circuit detected. | error |
0907 | Table hght invalid | Table height position sensing failure. | error |
0908 | Table hght sens short | Table height sensor short circuit detected. | error |
0909 | Table hght sens open | Table height sensor open circuit detected. | error |
0910 | CS alpha invalid | CS rotation alpha position sensing failure. | error |
0911 | CS beta invalid | CS rotation beta position sensing failure. | error |
0912 | WS hght invalid | Wallstand height position sensing failure. | error |
0913 | WS hght sens short cct | Wallstand height sensor short circuit detected. | error |
0914 | WS hght sens open cct | Wallstand height sensor open circuit detected. | error |
0915 | WS hght mismatch | Wallstand height position mismatch on tilted bucky unit. | error |
0916 | WS tilt invalid | Wallstand tilt position sensing failure. | error |
1000 | Bst watchdog | Bucky controller self-test failure on watchdog timer | critical |
1001 | Bst ram | Power-up RAM test failure | critical |
1002 | Bst eprom low byte | EPROM low byte self-test failure | critical |
1003 | Bst eprom high byte | EPROM high byte self-test failure | critical |
1004 | Bst uart | UART self-test failure, V24 port fails | critical |
1005 | Bst CAN controller | CAN controller self-test failure | critical |
1006 | Bst corrupt error log | Error log corruption detected during self-test | critical |
1007 | Bst servicedata upgrade | Service data upgrade detected during self-test | warning |
1008 | Bst servicedata downgr. | Service data downgrade detected during self-test | warning |
1009 | Bst two should be set | Configuration error in self-test | error |
1010 | Bst room two CAN master | Room 2 CAN master configuration error detected | error |
1011 | Bst old world also set | Old world configuration error detected | error |
1012 | Bst info level cancel | Logging level configuration error | info |
1013 | Bst dev info level | Development info level configuration error | info |
1014 | Bst CAN master override | CAN master override condition detected | unknown |
1015 | Bst not CAN master | System not configured as CAN master | unknown |
1016 | Bst two rooms with URF | Invalid configuration: two rooms with URF | unknown |
1017 | Bst OWI with URF | Invalid OWI with URF configuration detected | unknown |
1018 | Bst master with URF | Invalid master with URF configuration detected | unknown |
1019 | Bst selftest failed | Overall self-test failure on bucky controller | critical |
1100 | CH general - Unknown control handle connected | System detects unrecognized control handle | error |
1101 | CH test relay stuck | Relay on control handle self-test fails | critical |
1102 | CH test watchdog | Watchdog timer test fails | info |
1103 | CH test key stuck | Control handle key detected as stuck | critical |
1104 | CH test RAM | RAM self-test fails on control handle | critical |
1105 | CH test EPROM | EPROM self-test fails on control handle | critical |
1106 | CH test CAN | CAN controller self-test fails | critical |
1200 | Bucky controller processor error | Watchdog reset by software | error |
1300 | Tracking timeout at calibration | Calibration sequence timeout | error |
1301 | Tracking timeout at velocity | Velocity measurement timeout | error |
1302 | Tracking timeout at position | Position measurement timeout | error |
1303 | Tracking height inconsistent | Height measurement inconsistency detected | error |
1304 | Tracking reached position inconsistency | Position reached but inconsistent | error |
1305 | Tracking CSD error message | CSD communication error | error |
Cassette format not recognized | Cassette size does not match any entry in the centric position table | Encoder signals do not correspond to any standard cassette format | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.