Philips BuckyDiagnost IsoRad Floor: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the Philips BuckyDiagnost IsoRad Floor, with 13 ranked causes, 42 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
X-ray field size and limitation inaccurate radiation
Indicated field size does not match actual exposed field size; collimator field limitation exceeds specification
Ranked causes
- 1. Collimator calibration drift or mechanical wear (unknown)
- 2. PBL (Positive Beam Limiting) system not operational (unknown)
Measure film actual length vs exposed length for 18x24 cm cassette at SID 100 cm
- Test point
- 18x24 cm film L1 and L2 measurements
- Expected
- L1 - L2 ≤3 cm; indicated vs exposed difference ≤2 cm, ±0 to 3 cm
10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BLD light field misaligned with X-ray field radiation
X-ray field size and location does not match BLD light field for large or small focus
Ranked causes
- 1. Collimator or light field system mechanical misalignment (unknown)
Measure total misalignment of BLD light field edges with X-ray field edges for large focus
- Test point
- film edges L1, L2, W1, W2
- Expected
- L1 + L2 ≤2.0% SID; W1 + W2 ≤2.0% SID, ≤2.0% SID
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cassette loader radiation
Calibrate cassette loader size detection for small and large cassettes
Ranked causes
- 1. Cassette loader requires automatic calibration (common)
Select cassette sizes and loader
- Test point
- Size selection boxes and loader selection
- Expected
- Small and large cassette sizes selected; Table or Wallstand loader selected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ceiling Suspension height radiation
Adjust the minimum and maximum height positions of the ceiling suspension system
Ranked causes
- 1. Height adjustment required (common)
Move ceiling suspension to minimum or maximum height
- Test point
- Suspension position
- Expected
- Clear of table, pointing towards table
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ceiling Suspension SID positions radiation
Calibrate source-to-image distances for ceiling suspension at fixed positions
Ranked causes
- 1. SID positions require calibration (common)
Position x-ray tube pointing toward target
- Test point
- Tube angle and direction
- Expected
- Tube at +90 or -90 degrees pointing toward wallstand or D77 stand/FOCH
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ceiling Suspension tracking offset radiation
Adjust ceiling suspension to wallstand tracking offset for laser alignment
Ranked causes
- 1. Tracking offset requires adjustment (common)
Verify x-ray tube pointing toward wallstand
- Test point
- Tube angle and laser position
- Expected
- X-ray tube pointing toward wallstand; laser hitting same point
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Horizontal wallstand tracking position radiation
Set tracking position for horizontal tilted wallstand
Ranked causes
- 1. Wallstand tracking position requires configuration (common)
Center suspension over horizontal wallstand
- Test point
- Suspension position and control handle
- Expected
- Suspension centered over tilted wallstand
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Program APR data radiation
Configure and store Automatic Posterior-Anterior Registrations for bucky controller
Ranked causes
- 1. APR programs require configuration for registration devices (common)
Select registration device
- Test point
- Input window device selection
- Expected
- Registration device selected and highlighted
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tomography table height radiation
Adjust minimum and maximum height positions for tomography table operation
Ranked causes
- 1. Tomography table height adjustment required (common)
Table adjustment must be completed first
- Test point
- System configuration
- Expected
- Table heights already calibrated
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tomography working position radiation
Set and validate the tomography working position for ceiling suspension
Ranked causes
- 1. Tomography working position requires configuration (common)
Move ceiling suspension to tomography working position
- Test point
- Suspension position and control handle
- Expected
- Suspension in desired tomography working position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Wallstand height radiation
Adjust the minimum and maximum height positions of the wallstand
Ranked causes
- 1. Wallstand height adjustment required (common)
Ceiling suspension adjustment must be completed first
- Test point
- System configuration
- Expected
- Ceiling suspension heights already set
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray field center misaligned with image receptor radiation
X-ray field center does not coincide with image receptor center
Ranked causes
- 1. Tube centering mechanism misaligned or worn (unknown)
Measure displacement between X-ray field center and image receptor center at SID 100 cm
- Test point
- film center and X-ray field center intersection points
- Expected
- ≤20 mm, ±0 to 20 mm
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.