Philips Bodyguard: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Philips Bodyguard, with 20 ranked causes, 3 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
TSM continuous reset loop
TSM repeatedly resets and does not remain stable
Ranked causes
- 1. Defective PUB-Ethernet Switch connection or loose UTP connector (common)
- 2. Ethernet cables connected with no TSM attached, functioning as antenna (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
02HH error when starting fluoroscopy high voltage
Error logging shows '02HH: kV actual value out of range' from generator when starting fluoroscopy; works after a while.
Ranked causes
- 1. Timing problem with EN_X signal activation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIST error after Geometry software upgrade
BIST (Built-In Self-Test) error occurs following upgrade from R10.1.x or lower to R10.2.x or higher
Ranked causes
- 1. Defective TSO cables or incorrect jumper installation (common)
Verify jumper configuration
- Test point
- Jumpers per FCO 72200009
- Expected
- All jumpers installed per FCO 72200009 specification
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Black screen on TSM after APC selection with stack overflow
TSM turns black with wait message after APC selection. System returns after few seconds. Event log shows stack overflow error C00000FD
Ranked causes
- 1. APC list has only one entry or identical entries, or missing unique names causing stack overflow (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fast stand movement stops unexpectedly caution
When sliding Geo TSO knob to FullSpeed position, stand stops after small movement instead of using maximum speed.
Ranked causes
- 1. TSO speed position drops from 100 to 0% at FullSpeed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Geometry Module and Xper Module crash after APC selection
Geometry Module and Xper Module crash after selecting APC sequence. Xper Module does not respond and Geometry module resets
Ranked causes
- 1. Same APC position selected twice or single position in preset list causing infinite loop (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
GUI error with red cross status icon
System displays GUI error resulting in red cross on status icon. Technical event log shows Acquisition error 20483 and FieldService error 20484. LEDs H5 and H6 on SIB not lit
Ranked causes
- 1. PC Box Power supply failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard does not respond caution
Keyboard becomes unresponsive; switching system off/on does not restore function; requires hospital power OFF to unlock.
Ranked causes
- 1. Firmware issue in KVM receiver (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard error on PUB at power-on
System intermittently fails to start correctly after power-on; PUB does not start-up with displayed 'Keyboard error' message; Reset/restart clears problem.
Ranked causes
- 1. Incorrect BIOS version in PUB 4522 129 20291 (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Spare part PUB does not start up after installation
New spare part PUB starts with Touchscreen Alignment and SetID but fails to continue with SW download and does not start-up. No network connection available with Xper Module.
Ranked causes
- 1. Windows CE version 4.1.2 incompatible with lower versions (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tableclamp does not fit on new modules caution
Current tableclamp does not fit on new Operator Controls modules ordered as spare parts for Allura Xper FD20 systems.
Ranked causes
- 1. Supplier change for Operator Controls modules (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSM black screen after new PUB installation
No display on TSM screen after installing new PUB, only black screen shown
Ranked causes
- 1. PUB downloading from host, download not yet completed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSM display white screen at startup
Touchscreen Monitor (TSM) shows white screen at power-on; no software interface buttons or graphics visible; software does not initialize
Ranked causes
- 1. LVDS transmitter not properly seated (common)
Verify LVDS transmitter seating in PUB
- Test point
- LVDS transmitter connector in PUB
- Expected
- Transmitter fully seated and locked in connector
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSM displays black screen after new PUB installation
After installing new PUB (Portable User Interface Box), TSM screen shows black with no display elements
Ranked causes
- 1. PUB downloading UI file from host; download not complete (common)
Monitor Technical event log for download status
- Test point
- GUI event log filter
- Expected
- No 'Downloading UI file' message, or message cleared after completion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSM not available or not working correctly
Xper Module not available or not working correctly
Ranked causes
- 1. No power connected to HUB or faulty HUB (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSM screen blank after software download caution
TSM screen is blank (black or blueish). Download Board Software procedure shows 100% progress almost immediately but download does not complete correctly
Ranked causes
- 1. Corrupted flash memory card settings or old software version on flash memory card inside PUB (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSM submenu buttons not displayed
When pressing Acquisition, Processing, APC or QA buttons, TSM screen shows no selection buttons and stays empty
Ranked causes
- 1. Missing license for menu option (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
TSM white screen from start-up
White screen displayed on TSM from start-up with no software interface or buttons visible
Ranked causes
- 1. LVDS transmitter not inserted properly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Xper Geometry modules incomplete
Returned Xper Geometry modules have missing components with comments about incomplete delivery or option not installed.
Ranked causes
- 1. Buttons not mounted for customization (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI