Philips BiPAP Vision: troubleshooting and error codes
The ServiceForge knowledge base holds 80 documented symptoms for the Philips BiPAP Vision, with 155 ranked causes, 133 diagnostic checks with test points and expected values, and 172 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Exhalation Port Test: Low flow result
Circuit displays lower than normal leak rate at exhalation port; vents blocked or circuit unsound.
Ranked causes
- 1. Exhalation port vents blocked (common)
- 2. Circuit not sound (common)
- 3. Patient circuit (PC) defective (occasional)
- 4. Ambient flow module (AFM) defective (occasional)
- 5. Pressure relief valve (PRV) defective (occasional)
- 6. Intake low flow resistor (ILFR) defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not power on high voltage
Device does not turn on when power switch is activated; no display or indicator lights
Ranked causes
- 1. Power cord damaged or disconnected (common)
- 2. Blown fuses (common)
- 3. PSS (Power Supply System) failure (occasional)
- 4. Main power switch non-functional (occasional)
- 5. AC Inlet failure (rare)
- 6. Transformer failure (rare)
Inspect power cord
- Test point
- Power cord connection to unit
- Expected
- Cord fully inserted with no damage
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator inoperative (Vent Inop) patient safety
Unit displays Vent Inop indicator; system stops therapy delivery; LED lit on subsystem; hardware or communications failure detected
Ranked causes
- 1. Hardware failure in subsystem (common)
- 2. Spurious interrupt (MCU E01) (occasional)
- 3. Communications error (CRC/NAK/timeout) (occasional)
- 4. ROM CRC error (E0A) (rare)
- 5. RAM hardware error (E14) (rare)
- 6. Watchdog timer failure (rare)
Observe Vent Inop indicator and subsystem LEDs
- Test point
- MC, DC, PC LED indicators
- Expected
- One subsystem LED lit indicating hardware fault
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation port test shows low flow
Circuit displays lower than normal leak rate at exhalation port during self test
Ranked causes
- 1. Exhalation port vents blocked (common)
- 2. Defective patient circuit (PC) (common)
- 3. Defective air flow management (AFM) (common)
- 4. Defective pressure relief valve (PRV) (common)
- 5. Defective inline flow resistor (ILFR) (common)
Verify exhalation port vents are clear
- Test point
- Exhalation port
- Expected
- Vents not blocked and circuit sound
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation port test shows occluded port
Leak rate less than predicted during exhalation port test
Ranked causes
- 1. Exhalation port blocked (common)
- 2. Defective patient circuit (PC) (common)
- 3. Defective air flow management (AFM) (common)
- 4. Defective pressure relief valve (PRV) (common)
- 5. Defective inline flow resistor (ILFR) (common)
Verify exhalation port is not obstructed
- Test point
- Exhalation port
- Expected
- Port clear and unblocked
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation Port Test: Occluded exhalation port result
Leak rate was less than predicted; exhalation port blocked.
Ranked causes
- 1. Exhalation port blocked (common)
- 2. Patient circuit (PC) defective (occasional)
- 3. Ambient flow module (AFM) defective (occasional)
- 4. Pressure relief valve (PRV) defective (occasional)
- 5. Intake low flow resistor (ILFR) defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation port test shows excessive flow
Circuit displays higher than normal leak rate at exhalation port during self test
Ranked causes
- 1. Internal circuit improperly assembled (common)
- 2. Defective patient circuit (PC) (common)
- 3. Defective pressure relief valve (PRV) (common)
- 4. Defective inline flow resistor (ILFR) (common)
Inspect internal circuit assembly
- Test point
- Internal circuit connections
- Expected
- All connections properly seated and assembly correct
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation Port Test: Excessive flow result
Circuit displays higher than normal leak rate at exhalation port; internal circuit improperly assembled.
Ranked causes
- 1. Internal circuit improperly assembled (common)
- 2. Patient circuit (PC) defective (occasional)
- 3. Pressure relief valve (PRV) defective (occasional)
- 4. Intake low flow resistor (ILFR) defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Outlet air temperature too warm caution
Air exhaust from unit is excessively hot during normal operation
Ranked causes
- 1. High ambient room temperature (common)
- 2. Inlet filter blocked or dirty (common)
- 3. Blower/Valve malfunction (occasional)
- 4. PC (Pneumatics Controller) failure (rare)
Check ambient temperature
- Test point
- Room temperature near unit
- Expected
- Below 25°C (77°F)
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
No display, no LEDs lit, no audible output, no airflow
Ranked causes
- 1. RTC failure on MC (rare)
- 2. NVRAM CRC error on MC (rare)
- 3. Backup battery failure (occasional)
- 4. Reference voltage failure on MC (rare)
Measure input AC voltage at power receptacle
- Test point
- Unit AC input
- Expected
- 110-240 VAC depending on unit configuration
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower speed exceeds maximum RPM caution
Blower motor running at > 16500 RPM; speed control circuit or sensor failure
Ranked causes
- 1. Blower speed control circuit failure (common)
- 2. Blower speed sensor malfunction (occasional)
- 3. Blower motor mechanical fault (occasional)
Measure blower control voltage at PC board
- Test point
- Blower motor control output on PC
- Expected
- 0-5V or 0-10V depending on design
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation port test shows intermittent excessive flow
Leak rate intermittently high during exhalation port test
Ranked causes
- 1. Internal circuit not properly sealed or occluded (common)
- 2. Defective pressure relief valve (PRV) (common)
- 3. Defective patient circuit (PC) (common)
Verify circuit occlusion and seal integrity
- Test point
- Internal and external circuit
- Expected
- Properly sealed and occluded
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation port test shows pressure regulation error
Leak test pressures cannot be attained during exhalation port test
Ranked causes
- 1. Internal pressure line disconnected or obstructed (common)
- 2. Defective air flow management (AFM) (common)
- 3. Defective patient circuit (PC) (common)
Verify internal pressure line connection and patency
- Test point
- Internal pressure line
- Expected
- Connected and unobstructed
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation port test shows proximal line disconnected
Proximal pressure line is disconnected during exhalation port test
Ranked causes
- 1. Proximal pressure line internal connection loose or disconnected (common)
- 2. Proximal pressure line external connection loose or disconnected (common)
- 3. Proximal pressure line obstructed (occasional)
Verify proximal pressure line internal connection
- Test point
- Internal proximal line connector
- Expected
- Connected and secure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation Port Test: Intermittent excessive flow result
Leak rate was intermittently high during test; internal or external circuit not properly sealed.
Ranked causes
- 1. Internal or external circuit not properly sealed (common)
- 2. Pressure relief valve (PRV) defective (occasional)
- 3. Patient circuit (PC) defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation Port Test: Pressure regulation error result
Leak test pressures cannot be attained; internal pressure line disconnected or obstructed.
Ranked causes
- 1. Internal pressure line disconnected or obstructed (common)
- 2. Ambient flow module (AFM) defective (occasional)
- 3. Patient circuit (PC) defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid blower calibration data caution
Blower calibration failed during installation; speed or pressure data not stored
Ranked causes
- 1. Blower calibration procedure not completed correctly (common)
- 2. Blower pressure transducer fault (occasional)
- 3. PC memory or EEPROM failure (rare)
Re-run blower calibration from menu
- Test point
- Unit calibration interface or service port
- Expected
- Calibration completes without error; data stored
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen module cable disconnected during operation patient safety
OM detection error; oxygen module became unplugged or connection failed in use
Ranked causes
- 1. OM cable loose or disconnected (common)
- 2. OM cable connector corroded or damaged (occasional)
- 3. Oxygen module internal failure (occasional)
Visually inspect OM cable connection at unit and module
- Test point
- OM cable connectors at PC board and oxygen module
- Expected
- Connectors fully seated; no bent pins or corrosion visible
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure regulation failure patient safety
Cannot regulate to 2 cm H2O in Pressure Out or Flow Limit Control mode
Ranked causes
- 1. ILFR valve stuck or failed (common)
- 2. PRV valve stuck or failed (common)
- 3. Blower output insufficient (common)
Measure outlet pressure with ILFR valve commanded open
- Test point
- Unit outlet pressure sensor
- Expected
- Reaches 2 cm H2O setpoint or higher
- Tools
- Pressure gauge
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not start up high voltage
PC fails startup self-tests with voltage reference errors on MUX or ADC reads
Ranked causes
- 1. –12V voltage reference out of range (common)
- 2. –5V voltage reference out of range (common)
- 3. Reference voltage (positive) out of range (common)
Measure –12V reference at MUX
- Test point
- MUX input
- Expected
- –10.12 to –14.04 V, −14.04 to −10.12
- Tools
- DMM
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator Inoperative icon illuminated with audible alarm patient safety
Red wrench icon displayed and alarm sounds; ventilator shuts down immediately
Ranked causes
- 1. ILFR (Inspiratory Limiter/Flow Regulator) failure (occasional)
- 2. PRV (Pressure Relief Valve) failure (occasional)
- 3. Blower failure (occasional)
Refer to Vent Inop Troubleshooting Flow Charts
- Test point
- System status on display
- Expected
- Follow diagnostic flowchart on pages 6-14 and 6-15
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm battery voltage low caution
Low battery indicator displayed; backup battery depleted or charging failed
Ranked causes
- 1. Battery charge depleted (common)
- 2. Battery charging circuit failure (occasional)
Measure battery voltage with multimeter
- Test point
- Battery terminals or DC output
- Expected
- Nominal battery voltage (12V or 24V per design)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm not sounding patient safety
Alarm mute or no sound output during alarm condition; audio circuit fault
Ranked causes
- 1. Audible alarm current error on DC (common)
- 2. Alarm driver circuit fault (occasional)
Verify alarm speaker electrical continuity
- Test point
- Speaker connector pins on DC board
- Expected
- < 10 ohms resistance between pins
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backlight not illuminating caution
Display visible only in bright light; backlight LED or voltage circuit failure
Ranked causes
- 1. Backlight voltage error on DC (common)
- 2. Backlight LED failure (occasional)
Measure backlight supply voltage at DC connector
- Test point
- DC backlight power output
- Expected
- Specified backlight voltage (typically 12-24V)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PC/Laptop setup for BiPAP Vision testing
Configure serial communication between PC/Laptop and BiPAP Vision unit for troubleshooting and testing via Hyperterminal or Terminal emulation
Ranked causes
- 1. PC/Laptop communication not established (unknown)
- 2. Incorrect test cable or connection (unknown)
Verify correct test cable for unit serial number
- Test point
- Unit serial number
- Expected
- S/N 105999 or lower uses 582161; S/N 106000 or higher uses 1004823; upgraded hybrids use 1004823 with 1004699
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Vent indicator LED malfunction patient safety
LED not lit when alarm should display; indicator driver circuit failure
Ranked causes
- 1. Check Vent LED current error on DC (common)
- 2. LED driver IC failure (occasional)
Measure voltage across Check Vent LED
- Test point
- LED anode/cathode on DC board
- Expected
- ~2V across LED when activated
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display blank when powered on
Unit powers on as evidenced by indicator lights but LCD screen remains blank with no display output
Ranked causes
- 1. LCD display failure (common)
- 2. DC (Display Controller) board failure (occasional)
Verify power to LCD display
- Test point
- LCD power connector
- Expected
- Power present at LCD connector
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display voltage error caution
Display dim, flickering, or no image; display module may not power
Ranked causes
- 1. DC display voltage out of range (common)
- 2. Display module failure (occasional)
Measure display supply voltage on DC board
- Test point
- DC power output connector
- Expected
- Nominal supply voltage (typically 3.3V or 5V per schematic)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuses failing repeatedly high voltage
Fuses blow repeatedly after replacement indicating ongoing overcurrent condition
Ranked causes
- 1. Transformer short or overload (common)
- 2. PSS (Power Supply System) internal short (occasional)
Install new fuses and disconnect transformer
- Test point
- Fuse holders after transformer disconnected from PSS
- Expected
- Fuses remain intact without blowing time
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent On/Off condition high voltage
Device cycles between power on and power off unpredictably during operation
Ranked causes
- 1. Power cord loose or damaged (common)
- 2. Main power switch intermittently failing (occasional)
Verify power cord insertion and condition
- Test point
- Power cord at AC inlet
- Expected
- Cord fully seated with no visible damage or intermittent contact
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad key stuck or not responding caution
Key held down unresponsively; key press registered continuously for >30 seconds
Ranked causes
- 1. Physical key stuck (common)
- 2. Keypad scanning circuit malfunction (occasional)
Verify key is not mechanically stuck
- Test point
- Physical key buttons on front panel
- Expected
- All keys press and release freely
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noise from ventilator
Audible noise or rattling sounds coming from the unit during operation
Ranked causes
- 1. Blower bearing wear or obstruction (occasional)
- 2. Transformer vibration or mounting loose (occasional)
Locate source of noise
- Test point
- Unit during operation
- Expected
- Normal quiet operation or identified rattle/whine
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotary encoder does not adjust selection
Turning rotary encoder knob does not change displayed values or menu selections
Ranked causes
- 1. Rotary encoder mechanical or electrical failure (common)
- 2. DC (Display Controller) board failure (occasional)
Test rotary encoder function
- Test point
- Rotary encoder during operation
- Expected
- Turning encoder changes displayed value
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch pad not responding to selection
Pressing touch pad buttons does not register input or change settings
Ranked causes
- 1. Touch pad sensor failure (common)
- 2. DC (Display Controller) board failure (occasional)
Test touch pad response
- Test point
- Touch pad buttons
- Expected
- Button press registers on display
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent Inop indicator LED malfunction patient safety
LED not illuminating during Vent Inop fault condition
Ranked causes
- 1. Vent Inop LED current error on DC (common)
- 2. LED driver circuit failure (occasional)
Measure voltage across Vent Inop LED
- Test point
- Vent Inop LED on DC board
- Expected
- ~2V when fault condition present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower failure during operation
Blower overcurrent, undervoltage, or shutdown detected during operation
Ranked causes
- 1. Blower motor bad or drive circuit failure (common)
Verify blower operation and current draw
- Test point
- Blower motor connector
- Expected
- Motor spins at commanded speed without stalling
- Tools
- Ammeter, tachometer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower voltage reference out of range at startup
Blower drive voltage exceeds 100 mV during start-up self-test
Ranked causes
- 1. Blower voltage reference bad at startup (common)
Measure blower drive voltage at startup
- Test point
- Blower drive output on MUX
- Expected
- ≤ 100 mV, 0 to 100 mV
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Blower and Valves
Calibration procedure for blower and valves after major component replacement to ensure proper pressure and flow control.
Ranked causes
- 1. Major component replacement requires blower and valve recalibration (common)
Calibration: Blower/Valve Calibration
Calibrates the Blower and Valves of the unit after major component replacement.
Ranked causes
- 1. Major component replacement requires blower and valve recalibration (common)
Calibration: Earth Leakage Current (Normal Pole, No Earth, L2) caution
Measure leakage current on normal pole configuration without earth and with L2 active
Ranked causes
- 1. Insulation breakdown or component leakage (rare)
Measure normal pole, no earth, L2 leakage current
- Test point
- Earth leakage measurement points
- Expected
- Less than 300 microamps, < 300 microamps
- Tools
- Leakage current tester
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Earth Leakage Current (Reverse Pole, No Earth, L2) caution
Measure leakage current on reverse pole configuration without earth and with L2 active
Ranked causes
- 1. Insulation breakdown or component leakage (rare)
Measure reverse pole, no earth, L2 leakage current
- Test point
- Earth leakage measurement points
- Expected
- Less than 300 microamps, < 300 microamps
- Tools
- Leakage current tester
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Earth Resistance caution
Measure earth resistance to verify electrical safety ground integrity
Ranked causes
- 1. Poor ground connection or corroded earth terminal (occasional)
Measure earth resistance with 25 amp test current
- Test point
- Earth/ground terminal
- Expected
- Less than 0.10 Ohms, < 0.10 Ohms
- Tools
- Earth resistance tester
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation Port
System characterizes circuit leak rate; learned value stored in memory for accurate leak, minute ventilation, and tidal volume display. Without successful test, displays total leak instead of patient leak only.
Ranked causes
- 1. Exhalation port requires characterization after service or circuit changes (common)
Install circuit with exhalation device and pressure measurement
- Test point
- Unit outlet, pressure connection
- Expected
- 6' smooth inner lumen tubing, whisper swivel, pressure pick-off port connected; outlet occluded
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation Port Test patient safety
System learns intentional exhalation port leak over pressure range to store in memory for accurate patient leak, minute ventilation, and tidal volume display. Test must complete successfully to show 'Pt. Leak'; failure shows 'Tot. Leak' instead.
Ranked causes
- 1. Exhalation port leak characterization required for accurate display and alarms (common)
Verify equipment setup complete
- Test point
- Unit outlet and pressure connections
- Expected
- 6' tubing installed with whisper swivel and pressure pick-off port; pressure line connected; outlet occluded
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse Call/Remote Alarm Option 1 (Remote Alarm)
Verify Nurse Call/Remote Alarm output connector impedance changes on patient alarm activation
Ranked causes
- 1. MC board jumper configuration incorrect or relay malfunction (occasional)
Measure continuity across Nurse Call/Remote Alarm output connector with unit off
- Test point
- Nurse Call/Remote Alarm connector High and Low
- Expected
- Approximately 48.6K to 53.7K Ohms, 48.6K to 53.7K
- Tools
- DVM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse Call/Remote Alarm Option 2 (Normally Closed)
Verify Nurse Call/Remote Alarm output is normally closed and opens on patient alarm
Ranked causes
- 1. MC board jumper configuration incorrect or relay malfunction (occasional)
Measure continuity across Nurse Call/Remote Alarm output connector with unit off
- Test point
- Nurse Call/Remote Alarm connector High and Low
- Expected
- Closed circuit
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse Call/Remote Alarm Option 3 (Normally Open)
Verify Nurse Call/Remote Alarm output is normally open and closes on patient alarm
Ranked causes
- 1. MC board jumper configuration incorrect or relay malfunction (occasional)
Measure continuity across Nurse Call/Remote Alarm output connector with unit off
- Test point
- Nurse Call/Remote Alarm connector High and Low
- Expected
- Open circuit
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse Call/Remote Alarm output – Option 1 (Remote Alarm)
Verification of Remote Alarm function with JP1 and JP2 set at 2,3: resistance ~48.6–53.7 kΩ when unit off; open circuit when alarm active.
Ranked causes
- 1. Nurse Call relay or circuit malfunction (unknown)
Measure continuity at Nurse Call connector (unit off)
- Test point
- Nurse Call/Remote Alarm High and Low
- Expected
- 48.6K to 53.7K Ohms kΩ, 48.6–53.7 kΩ
- Tools
- Fluke DVM or equivalent
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse Call/Remote Alarm output – Option 2 (Normally Closed)
Verification of Nurse Call normally closed function with JP1 at 2,3 and JP2 at 1,2: closed circuit when unit off; open circuit when alarm active.
Ranked causes
- 1. Nurse Call relay or output circuit malfunction (unknown)
Measure continuity at Nurse Call connector (unit off)
- Test point
- Nurse Call/Remote Alarm High and Low
- Expected
- Closed circuit
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse Call/Remote Alarm output – Option 3 (Normally Open)
Verification of Nurse Call normally open function with JP1 and JP2 at 1,2: open circuit when unit off; closed circuit when alarm active.
Ranked causes
- 1. Nurse Call relay or output circuit malfunction (unknown)
Measure continuity at Nurse Call connector (unit off)
- Test point
- Nurse Call/Remote Alarm High and Low
- Expected
- Open circuit
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen alarm at no flow condition patient safety
Verification that unit generates 'no O2 Flow' alarm within 15 breaths or less when O2 cylinder is closed and disconnected.
Ranked causes
- 1. O2 flow alarm circuit malfunction (unknown)
Verify O2 Flow alarm activates after disconnect
- Test point
- Vision alarm display and audio
- Expected
- Alarm activates within 15 breaths
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Delivery at 100% patient safety
Verify oxygen analyzer reading is between 90-109% when oxygen set to 100%
Ranked causes
- 1. IPAP pressure drift affecting tidal volume (common)
Set O2 to 100%, measure oxygen analyzer after stabilization
- Test point
- Oxygen Analyzer
- Expected
- 90% to 109% (inclusive), 90 to 109%
- Tools
- Oxygen Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen delivery at 100% setpoint patient safety
Verification that oxygen analyzer reads at least 90% (up to 109%) at 100% O2 setting after stabilization, with tidal volume maintained.
Ranked causes
- 1. O2 delivery system malfunction at maximum setpoint (unknown)
Set O2 to 100% and measure analyzer reading
- Test point
- Vision O2 parameter and Oxygen Analyzer display
- Expected
- 90% to 109%, 90–109%
- Tools
- Oxygen Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen delivery at 18–24% setpoint patient safety
Verification that oxygen analyzer stabilizes and reads within 18–24% when O2 module connected at initial 21% setting with test circuit and cylinder at 50 psi.
Ranked causes
- 1. Oxygen module or analyzer malfunction (unknown)
Connect O2 line and set regulator pressure
- Test point
- O2 cylinder valve and regulator
- Expected
- 50 psi
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Delivery at 21% patient safety
Verify oxygen analyzer reading is within 18-24% when oxygen module is set to room air equivalent
Ranked causes
- 1. Oxygen analyzer not stabilized or calibration drift (common)
Connect O2 line and set cylinder to 50 psi, measure oxygen level
- Test point
- Oxygen Analyzer
- Expected
- 18% to 24%, 18 to 24%
- Tools
- Oxygen Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Delivery at 30% patient safety
Verify oxygen analyzer reading is between 27-33% when oxygen module set to 30%
Ranked causes
- 1. IPAP pressure drift or oxygen delivery variance (common)
Set O2 to 30% via Parameters menu, measure oxygen level
- Test point
- Oxygen Analyzer
- Expected
- 27% to 33% (inclusive), 27 to 33%
- Tools
- Oxygen Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen delivery at 30% setpoint patient safety
Verification that oxygen analyzer reads 27–33% after adjusting O2 parameter to 30% and stabilization, with tidal volume maintained in range.
Ranked causes
- 1. O2 parameter setting or delivery system malfunction (unknown)
Set O2 parameter and measure analyzer reading
- Test point
- Vision O2 parameter and Oxygen Analyzer display
- Expected
- 27% to 33%, 27–33%
- Tools
- Oxygen Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Delivery at 40% patient safety
Verify oxygen analyzer reading is between 36-44% when oxygen set to 40%
Ranked causes
- 1. IPAP pressure drift affecting tidal volume (common)
Set O2 to 40%, measure oxygen analyzer after stabilization
- Test point
- Oxygen Analyzer
- Expected
- 36% to 44% (inclusive), 36 to 44%
- Tools
- Oxygen Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen delivery at 40% setpoint patient safety
Verification that oxygen analyzer reads 36–44% after setting O2 to 40%, alarm is cleared, and no alarms remain after 8 breath cycles, with tidal volume in range.
Ranked causes
- 1. O2 delivery or alarm reset malfunction (unknown)
Verify no alarms after 8 breath cycles at 40% O2
- Test point
- Vision alarm display
- Expected
- No alarms
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Delivery at 60% patient safety
Verify oxygen analyzer reading is between 54-66% when oxygen set to 60%
Ranked causes
- 1. IPAP pressure drift affecting tidal volume (common)
Set O2 to 60%, measure oxygen analyzer after stabilization
- Test point
- Oxygen Analyzer
- Expected
- 54% to 66% (inclusive), 54 to 66%
- Tools
- Oxygen Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Oxygen delivery at 60% setpoint
- Calibration: Oxygen Delivery at 80%
- Calibration: Oxygen delivery at 80% setpoint
- Calibration: Oxygen Flow Module zero voltage
- Calibration: Oxygen Module Alarm at No Flow
- Calibration: PAV/T mode activation and display verification
- Calibration: Run-In Cycle Procedure
- Calibration: Tidal Volume
- Check Ventilator icon illuminated with audible alarm
- Circulation fan not operational
- EPROM upgrade or PAV/T mode installation
- Exhalation Port Test: Proximal pressure line disconnected result
- ILFR valve voltage reference out of range at startup
- Loss of AC power detection and response
- Main Power indicator blinks sporadically
- O2 valve voltage reference out of range at startup
- Oxygen baffle installation
- PRV valve voltage reference out of range
- Touch pad adhesive failure during removal
- Touch pad requires replacement
20 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
100 | RTC failure on the MC | Real Time Clock chip malfunction detected during power-up self-test | critical |
101 | MC NVRAM CRC error | Checksum of stored configuration data does not match calculated value | critical |
102 | Backup battery failure on MC | Battery voltage low or not present; backup power lost | warning |
103 | Vref failure on MC | Reference voltage not stable or out of tolerance | critical |
11D1 | Keypad error during startup | Key pressed during start-up tests or held down too long (>10 seconds) | warning |
11D7 | Invalid start-up test data | Invalid test case data in startup self-test sequence | error |
11D8 | Invalid BIST test data | Invalid Built-In Self Test case data detected during startup | error |
1201 | ADC timeout (Analog-to-Digital converter) | Internal ADC conversion incomplete after 173 usec delay during read of ILFR/PRV/O2/ANA/Vref | error |
1202 | Blower speed test fail | Invalid data in blower test status signal during operational test | error |
1203 | Error controlling operational tests | Invalid data in test case primary or secondary control signals | error |
1204 | MUX voltage error | Invalid data in MUX channel parameter or out-of-range reading | error |
1205 | Other voltage error | Invalid data in ADC channel parameter or unexpected reading | error |
1206 | IADC failed (internal ADC) | Internal ADC did not complete conversion sequence before background sensors read | error |
1207 | MC did not communicate within startup time | MC did not request status from PC within 30 seconds of PC power-up | critical |
1208 | 12-volt voltage reference test failed (operational) | 12V signal from MUX measured < 11V or > 13V during runtime test | error |
120A | Bulk supply voltage reference test failed (operational) | Bulk voltage measured < 20V or > 38.88V during operational test | error |
120B | -12-volt voltage reference test failed (operational) | -12V signal measured < -13V or > -11V during operational test | error |
120C | -5-volt voltage reference test failed (operational) | -5V signal measured < -5.425V or > -4.547V during operational test | error |
120D | Reference voltage test failed (operational) | Voltage reference measured < 3.749V or > 4.445V during operational test | error |
120E | QSM failed (Queued Serial Module) | QSM communication or initialization test failed | error |
120F | Error in rise rate processing | Rise rate information from MC-to-PC message value outside range 0-4 | warning |
1210 | Blower voltage bad in safe state | Blower drive voltage measured > 15mV when unit in safe state (should be 0V) | error |
1211 | PRV voltage bad in safe state | PRV (pressure relief valve) drive voltage > 15mV during safe state | error |
1212 | ILFR voltage bad in safe state | ILFR (inspiratory flow rate?) drive voltage > 15mV during safe state | error |
1213 | O2 valve voltage bad in safe state | Oxygen valve drive voltage > 15mV during safe state | error |
1214 | Failure controlling startup tests | Invalid data in startup test case parameter | error |
1215 | 12-volt voltage reference test failed (startup) | 12V supply measured < 10.12V or > 14.04V during power-up self-test | warning |
1216 | ANA ground voltage reference test failed (startup) | ANA ground signal measured < 0mV or > 500mV during power-up test | warning |
1217 | Bulk supply voltage reference test failed (startup) | Bulk voltage measured < 18.4V or > 38.88V during power-on self-test | warning |
1218 | –12V voltage reference test failed | –12V signal from MUX is outside acceptable range (–14.04 to –10.12 V) | critical |
1219 | –5V voltage reference test failed | –5V signal from MUX is outside acceptable range (–5.833 to –4.467 V) | critical |
121A | Reference voltage test failed | Voltage reference signal from MUX outside range (3.749 to 4.445 V) | critical |
121B | Blower voltage reference bad during startup | Blower drive voltage exceeds 100 mV | critical |
121C | ILFR voltage reference bad during startup | ILFR drive voltage from internal ADC exceeds 100 mV | critical |
121D | O2 valve voltage reference bad during startup | Oxygen module drive voltage from internal ADC exceeds 100 mV | critical |
121E | PRV voltage reference bad during operation | PRV drive voltage from internal ADC exceeds 100 mV | error |
121F | Blower voltage reference bad on second try after startup | Blower drive voltage exceeds 40 mV on second try immediately after first try | warning |
1220 | ILFR voltage reference bad on second try after startup | ILFR drive voltage exceeds 40 mV on second try immediately after first try | warning |
1221 | O2 valve voltage reference bad on second try after startup | O2 drive voltage exceeds 40 mV on second try immediately after first try | warning |
1222 | PRV voltage reference bad on second try after startup | PRV drive voltage exceeds 40 mV on second try immediately after first try | warning |
1225 | Internal ADC calibration divide by zero error | Analog reference voltage equals analog ground during ADC calibration | critical |
1226 | Bad O2 temperature sensor | O2 module temperature outside range (40–160°F) for 2500 counts or conversion not complete | error |
1227 | Blower out of calibration or stuck | Blower speed drifts ±750 RPM from desired setting for 12 seconds | error |
1228 | Bad pressure setpoint calculated | Low-pass filter of pressure setpoint in rise rate calculation out of range | error |
1229 | Patient pressure sensor drift error | Low-pass filter of pressure change average exceeds 10 cm H2O when flow < 5 LPM | error |
122A | Corrupted resistance table | Invalid data detected in resistance table during startup | critical |
122B | Invalid mode selected | Selected mode not supported in software during startup | critical |
122C | Blower failure | Blower overcurrent, undervoltage, or shutdown detected | critical |
122D | Bad backup unit outlet pressure sensor | Data cannot be read from backup unit outlet pressure sensor | critical |
122E | Unit outlet pressure difference too large | Unit outlet and backup unit outlet sensor readings differ excessively | error |
122F | Stuck backup unit outlet pressure sensor | Bad backup unit outlet pressure sensor or ADC reading | error |
1230 | Voltage failure | Voltage reference test failed during operation | critical |
1300 | Bad AFM calibration data CRC | AFM calibration data all zeroes or CRC check fails | critical |
1301 | Bad O2 module calibration data CRC | OM calibration all zeroes or CRC check fails | critical |
1302 | Bad PC calibration data CRC | PC calibration data all zeroes or CRC check fails | critical |
1304 | QSPI timeout during EEPROM read | QSPI does not finish within 100 microseconds after EEPROM selection for read | critical |
1305 | No AFM detected | AFM detection signal supplied by hardware exceeds 300 mV (cutoff for AFM presence) | critical |
1308 | Bad pressure setpoint message from MC | Master controller commands pressure greater than 40 cm H2O in message | error |
1309 | Bad rise rate message from MC | Rise rate setpoint in MC message exceeds 4 | error |
130A | Bad IPAP setpoint message from MC | IPAP setpoint in MC message exceeds 40 cm H2O (non-PAV) or 50 cm H2O (PAV) | error |
130B | Bad O2 concentration setpoint message from MC | O2 concentration in MC message < 21% or > 100% | error |
130C | Invalid MC message structure | Byte count in message does not match expected byte count | error |
1312 | AFM detect conversion timeout | Read of ADC not completed in time for AFM detection read | critical |
1313 | OM detect conversion timeout | Read of ADC not completed in time for OM detection read | critical |
1316 | Stuck absolute atmospheric pressure sensor | Atmospheric pressure reading < 20 in. Hg or > 40 in. Hg for 2.5 seconds | error |
1317 | Stuck unit outlet pressure sensor | Pressure reading < –5 cm H2O or > 70 cm H2O for 2.5 seconds | error |
1318 | Stuck patient pressure sensor | Pressure reading < –5 cm H2O or > 50 cm H2O for 2.5 seconds | error |
1319 | Stuck total flow sensor | Flow reading < –200 LPM or > 300 LPM for 2.5 seconds | error |
131A | Stuck O2 flow sensor | O2 flow reading > 120 LPM for 2.5 seconds | error |
131B | Bad air temperature sensor | Air temperature outside range (40–160°F) for 2500 counts or conversion incomplete | error |
131C | ATM detected bad unit outlet pressure sensor | (High ATM − Low ATM) × 100 ≥ 5 × High ATM | error |
131D | Bad PC calibration data 2 CRC | PC calibration data 2 CRC fails after filling with default or calculated values | critical |
131E | Bad PC calibration data 3 CRC | PC calibration data 3 CRC fails after filling with default values | critical |
131F | Error in drift test case | Invalid data detected in drift test case during calibration | error |
1320 | Bad AFM drift calibration data CRC | AFM calibration data 1 CRC fails after filling with default or calculated values | error |
1321 | Bad PC calibration data 4 CRC | PC calibration data 4 CRC fails after filling with default or calculated values | critical |
1322 | Sensor drift failure | Unit outlet or pressure sensor out of range (Tolerance ± 2 cm H2O around nominal) | error |
1323 | Bad calibration data 6 CRC | Drift calibration data could not be updated | critical |
1324 | Bad OM drift calibration data CRC | OM calibration data 1 CRC fails after filling with default or calculated values | error |
1325 | Total flow sensor drift | Total flow sensor drifted beyond acceptable tolerance | warning |
1326 | Oxygen flow sensor drift | Oxygen flow sensor drifted beyond acceptable tolerance | warning |
200 | DC display voltage error | Display supply voltage outside acceptable range | error |
201 | DC audible alarm current error | Alarm speaker current draw abnormal; open circuit or short detected | warning |
202 | DC Check Vent indicator current error | Check Vent LED current not detected or excessive | warning |
203 | DC Vent Inop indicator current error | Vent Inop LED current abnormal or not present | warning |
204 | DC backlight voltage error | Backlight supply voltage out of range or unstable | error |
205 | DC alarm battery low | Battery voltage below operational threshold | warning |
206 | Keypad error | Key held down continuously for 30+ seconds; no release detected | critical |
207 | Illegal operational test requested | Requested test does not exist in software | warning |
209 | +12V out of range | ADC voltage test on +12V rail fails | critical |
20A | –12V out of range | ADC voltage test on –12V rail fails | critical |
20B | +24V out of range | ADC voltage test on +24V supply fails | critical |
20C | Reference voltage out of range | ADC voltage test on reference voltage fails | critical |
20D | ADC could not return +12V value | ADC conversion of +12V reference fails during test | critical |
20E | ADC could not return –12V value | ADC conversion of –12V reference fails during test | critical |
20F | ADC could not return +24V value | ADC conversion of +24V reference fails during test | critical |
210 | ADC could not return reference voltage value | ADC conversion of reference voltage fails during test | critical |
211 | Power fail without AC fail | Loss of bulk supply without loss of AC input, caused by PS1 trip | critical |
300 | Circulation fan not operational | Fan motor not running or speed below minimum threshold | error |
301 | Invalid byte count in DCS message | Byte count from DCS not within acceptable range or equals zero | error |
303 | Blower speed exceeds 16500 RPM | Blower speed feedback exceeds safe maximum during operation | critical |
304 | Divide by zero in sum total breaths calculation | Sum of breath counts yields zero in divide operation | error |
305 | Divide by zero in baseline error calculation | Clock cycle times per breath equals zero | error |
306 | OM detection error (oxygen module disconnected) | Oxygen module cable disconnected during operation after successful initialization | critical |
306 | Divide by zero in BPM/Min calculation | Sum total time equals zero in BPM calculation | error |
307 | Divide by zero in Ti/Ttot calculation | Clock cycles time per breath equals zero | error |
30A | MC scan cannot be performed | Response task does not complete processing before 10 msec interrupt occurs again | error |
30B | Invalid command for CPAP mode | MC receives CPAP-incompatible command from DC intended for S/T or PAV/T modes | error |
30C | Invalid command for S/T mode | MC receives S/T-incompatible command from DC intended for CPAP or PAV/T modes | error |
30D | Invalid command for PAV/T mode | MC receives PAV/T-incompatible command from DC intended for CPAP or S/T modes | error |
30E | No dispatch function for command | Command ID from DC not recognized by MC | error |
310 | Invalid mode update | Mode update message contains invalid data | error |
311 | Illegal alarm module detected | Hardware returns module type other than built-in Module B | warning |
312 | MC scan overruns 3 times in 1 hour | Response task overrun occurs 3 times within a 1-hour period | error |
313 | PAV/T security failure | PAV/T software implementation not accessible or corrupted | critical |
601 | Spurious interrupt | MCU received unexpected interrupt signal; stack or vector corrupted | error |
602 | Unassigned interrupt | Interrupt occurred for which no handler is defined | error |
603 | Bus interrupt | Illegal bus cycle or address error detected | error |
604 | Illegal instruction executed | CPU fetched and attempted to execute invalid opcode | critical |
605 | Breakpoint error | Software breakpoint trap executed but not expected by debugger | warning |
606 | Divide by zero error | CPU attempted integer division by zero; trap generated | error |
607 | Uninitialized interrupt | Interrupt occurred but handler not initialized or null | error |
608 | Software interrupt executed | Software deliberately triggered interrupt; handler not found or error state | error |
609 | Unused interrupt executed | Interrupt vector reserved or unused triggered an event | warning |
60A | ROM CRC error (EPROM corrupted) | Calculated checksum of ROM does not match stored CRC | critical |
60C | Bad MC state | Invalid value detected on MC state signal line | error |
60E | Watchdog timer failure | Watchdog circuitry malfunction; software continues after long watchdog delay | critical |
614 | Walking RAM test error (RAM hardware fault) | Pattern written to RAM not read back correctly from one or more locations | critical |
616 | Watchdog reset after watchdog alarm PAL test | Power-on reset occurred after watchdog test completed successfully | error |
617 | SCI failed (serial communication interface) | SCI register not ready for character output after delay loop iterations complete | error |
700 | CRC error on third send to DC-MC | Calculated CRC does not match transmitted CRC on third attempt to send to DC | error |
701 | NAK received on third send to DC-MC | Negative acknowledgment message received on third attempt to send data to DC | error |
702 | Tack timeout on third send to DC-MC | No acknowledgment received within 750 microseconds on third send attempt | error |
703 | Trply timeout on third send to DC-MC | Slave did not begin responding within 1 millisecond of last byte from MC | error |
705 | Packet does not fit in message buffer | Message buffer overflow; packet size exceeds available space in "Move Packet To Message" | error |
706 | Invalid timer value | Timer not initialized before call or invalid data in timer parameter | warning |
707 | Invalid delay value | Timer delay exceeds one minute limit or timer not initialized | warning |
708 | Invalid delay (microsecond) | Microsecond delay requested exceeds 85 usec or timer not initialized | warning |
709 | GPT not initialized | Call to initiate timers made but GPT (general purpose timer) not initialized | error |
70A | Invalid usec ticks | Timer microsecond parameter too large or timer not initialized | warning |
70B | Invalid priority level | PCsed priority level value contains invalid data | warning |
70C | CRC error on third send to PC | Calculated CRC does not match transmitted CRC when sending to PC on third attempt | error |
70D | NAK received on third send to PC | Negative acknowledgment received when sending to PC on third attempt | error |
70E | Tack timeout on third send to PC | No acknowledgment within 750 usec when sending to PC on third attempt | error |
70F | Trply timeout on third send to PC | PC did not start responding within 1 msec of last byte sent from MC | error |
AC Power Failure | Power to the device is lost while running | AC power lost while power switch in Start position | critical |
Apnea | No spontaneous triggered breath detected within set apnea interval | Apnea monitor detects absence of spontaneous trigger within selected time interval | warning |
B00 | Bad ICB message from MC | Invalid status in self-test, byte count > 125, or unrecognized command ID in ICB message | critical |
B01 | Bad download sequence from MC | DC receives download message when not expecting it | critical |
B02 | Bad mode | Invalid data in test mode signals or mode-specific messages received in wrong mode | error |
B03 | Error decoding hard key table | ICB message corresponding to selected hard key not valid | error |
B04 | Bad screen state | Soft key selection received for mode when impossible (illegal keystroke) | error |
B09 | Simultaneous failure of audible alarm and system error LED | Simultaneous failure detected by hardware as indicated in status | critical |
B0A | MC did not start communication within start-up time | MC did not send status request within 30 seconds of DCS startup | critical |
B0B | Bad font type | Invalid data on screen font type signal during display initialization | warning |
B0C | Bad video memory address | Calculated screen pixel address exceeds screen bounds (off screen) | error |
B0D | DC queue overflow | No room in display queue for incoming MC message (background not running often enough) | warning |
B0E | Bad graph size | X or Y length of graph <= 0 (invalid memory data) | error |
B0F | No graph structure available | Attempt to initialize a fourth graph when only 3 are supported | warning |
B10 | Spurious keypad interrupts | More than 10 keypad interrupts in a row with no key depressed | warning |
Check Vent Error 301 | Ventilator operation error that may occur after PC/MC board upgrade | When upgraded unit is first turned on after installing repair kits 1004713 or 1000356 | warning |
Check Ventilator | System error has occurred but ventilator continues to operate | System error detected during operation | warning |
Disconnect | Mask has been removed or excessive leak detected | Mask removed or becomes dislodged causing excessive leak | warning |
Hi P | High pressure alarm | Proximal airway pressure exceeds high pressure setting for more than 0.5 seconds | warning |
Hi Rate | High total breathing rate alarm | Total breathing rate (machine + spontaneous) exceeds Hi Rate alarm setting | warning |
LoMinVent | Low minute ventilation alarm | Minute ventilation falls 50% or 5 LPM below limit for one minute period | warning |
Lo P | Low pressure alarm | Proximal airway pressure remains below low pressure setting for duration set by Low Pressure Delay control | warning |
Lo Rate | Low total breathing rate alarm | Total breathing rate (machine + spontaneous) below Lo Rate alarm setting | warning |
O2 Flow | Oxygen supply lost or insufficient | Oxygen supply pressure or flow is lost during operation | warning |
P Regulation | Loss of pressure regulation alarm | Measured proximal pressure differs more than ±5 cm H2O from set pressure for greater than 5 seconds | warning |
ProxLine Disc | Proximal pressure line disconnect | Proximal pressure measures less than 1.0 cm H2O for greater than 1.0 second | warning |
System Failure | System level failure that impairs performance | AC power failure, temperature overload fuse activation, or no bulk supply voltage | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.