Philips Avalon FM40/FM50: troubleshooting and error codes

The ServiceForge knowledge base holds 44 documented symptoms for the Philips Avalon FM40/FM50, with 106 ranked causes, 29 diagnostic checks with test points and expected values, and 27 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

External device not receiving data

LAN/RS232 connected device (e.g., OB TraceVue surveillance system) does not receive data from monitor

Ranked causes

  1. 1. LAN/RS232 port not configured for data export (common)
  2. 2. Cable not connected correctly or defective (common)
  3. 3. External device does not support data export protocol version (occasional)
  4. 4. Terminal concentrator does not support dynamic speed negotiation (occasional)
  5. 5. LAN/RS232 board in wrong slot (occasional)
  6. 6. LAN/RS232 board defective (rare)
  7. 7. Wrong cable used for RS232 (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display is blank or brightness reduced

Screen shows no image or very dim display

Ranked causes

  1. 1. Display flex cable not connected properly (common)
  2. 2. Backlight inverter cable not connected properly (common)
  3. 3. Backlight tubes defective (occasional)
  4. 4. Backlight inverter defective (occasional)
  5. 5. LCD flat panel defective (rare)
  6. 6. API Board defective (rare)
First check

Inspect display flex cable

Test point
Display assembly to API Board connection
Expected
Firmly connected at both ends

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP CHECK PAPER message issued

Check paper error message displayed on recorder or console

Ranked causes

  1. 1. Drawer open with paper on sensor (common)
  2. 2. Paper jam (common)
  3. 3. Paper sensor dirty (occasional)
  4. 4. Paper sensor defective (occasional)
  5. 5. Rubber roller slipping (occasional)
  6. 6. Non-approved paper being used (common)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Green LED does not light up

No power indication and no test tone heard from speaker

Ranked causes

  1. 1. No AC mains connection (common)
  2. 2. Power supply defective (occasional)
  3. 3. Switch board cable disconnected (common)
  4. 4. Switch board defective (occasional)
  5. 5. API Board defective (occasional)
First check

Verify AC power cord connection

Test point
Power cord connector
Expected
Firmly seated in AC outlet

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transducer Finder LED does not light after plugging in

Transducer Finder LED does not illuminate after insertion, transducer appears not to work, INOP BUS MASTER MALFUNC may display

Ranked causes

  1. 1. Defective transducer cable (common)
  2. 2. Defective connector block (common)
  3. 3. Transducer or connector block is defective (occasional)
  4. 4. API Board is defective (occasional)
  5. 5. No power to API Board (rare)
First check

Visually inspect transducer cable connector

Test point
transducer cable connector
Expected
no damage visible

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard/Mouse not functioning

Keyboard or mouse attached directly to monitor does not respond

Ranked causes

  1. 1. Keyboard/Mouse not connected properly (common)
  2. 2. Keyboard/Mouse defective (occasional)
  3. 3. PS/2 I/O board not properly plugged in (occasional)
  4. 4. PS/2 I/O board defective (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No paper transport in recorder

Fetal recorder paper does not advance during operation

Ranked causes

  1. 1. Drawer is open or not properly closed (common)
  2. 2. Paper jam (common)
  3. 3. Stepper motor cable disconnected (occasional)
  4. 4. Stepper motor defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No recorder key available and INOP FetRec EQUIP MALF issued

Recorder control key does not appear on screen; equipment malfunction message displayed

Ranked causes

  1. 1. Recorder has not been calibrated (common)
  2. 2. EEPROM on API Board defective (rare)
  3. 3. Recorder Controller on API Board defective (rare)
  4. 4. Recorder cable disconnected (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper empty warning issued but paper is not depleted

Fetal recorder paper empty warning displayed in status line while paper remains in drawer

Ranked causes

  1. 1. Drawer is open (common)
  2. 2. Paper jam (common)
  3. 3. Paper sensor disconnected (occasional)
  4. 4. Paper sensor defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Speaker malfunction

INOP message 'Speaker Malfunct.' is displayed; alarm occurs but no sound is issued

Ranked causes

  1. 1. Speaker cable disconnected (common)
  2. 2. Speaker defective (common)
  3. 3. Sound amplifier on API Board defective (occasional)
  4. 4. Volume set to 0 (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen not functioning

Screen does not respond to touch input

Ranked causes

  1. 1. Touchscreen functionality disabled (common)
  2. 2. Touchscreen cable not connected (occasional)
  3. 3. Touch controller defective (occasional)
  4. 4. Touch Sensor defective (occasional)
First check

Check for padlock symbol on Main Screen

Test point
Main Screen key
Expected
No padlock icon visible

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Individual Parameter INOP patient safety

Single parameter shows INOP (FHR, NBP, SpO2, TOCO, etc) indicating component malfunction

Ranked causes

  1. 1. Transducer, sensor, patient module or board defective (common)
  2. 2. Bus master firmware incompatible after software upgrade (occasional)
  3. 3. API Board defective (occasional)
First check

Swap suspected component

Test point
Component slot/connector
Expected
INOP disappears and parameter measures normally

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No test sound or Speaker Malfunction INOP

Speaker does not produce test tone during boot or INOP message displayed

Ranked causes

  1. 1. Speaker cable disconnected (common)
  2. 2. Speaker defective (occasional)
  3. 3. API Board defective (rare)
First check

Inspect speaker cable

Test point
Speaker connectors
Expected
Firmly connected

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor print quality

Fetal recorder printout is of poor quality with missing or faint marks

Ranked causes

  1. 1. Heat setting needs adjusting (common)
  2. 2. Thermal Printhead dirty (common)
  3. 3. Thermal Printhead failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder paper remains blank

Fetal recorder appears to operate normally but paper prints no signal or marks

Ranked causes

  1. 1. Thermal Printhead disconnected (occasional)
  2. 2. Thermal Printhead defective (occasional)
  3. 3. Wrong side of paper facing up (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Remote touch display not responding

Touch screen operation of connected remote display via MIB/RS232 is not possible

Ranked causes

  1. 1. Cable not properly connected (common)
  2. 2. MIB/RS232 port not configured (common)
  3. 3. Factory Defaults were loaded (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Settings malfunction

INOP message indicating problem during configuration cloning process or corrupted settings memory

Ranked causes

  1. 1. Problem during cloning process (occasional)
  2. 2. Memory space for settings corrupted (occasional)
  3. 3. Main CPU Board defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Monitor Temperature INOP caution

INOP message indicates internal temperature exceeds safe operating limit

Ranked causes

  1. 1. Environment temperature too high (common)
  2. 2. Main Board defective (occasional)
First check

Verify operating environment temperature

Test point
Monitor location
Expected
Within manufacturer specifications

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Settings INOP

INOP occurs during normal operation indicating possible software problem or settings corruption

Ranked causes

  1. 1. Monitor software problem (occasional)
  2. 2. Incomplete or unsuccessful software upgrade (occasional)
First check

Verify monitor and patient settings

Test point
Configuration Mode
Expected
Settings consistent with expected values

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Green LED does not blink after power On/Off

Power switch activated but green LED remains unresponsive

Ranked causes

  1. 1. Switch board defective (common)
  2. 2. Defective processor on API Board (occasional)
First check

Activate power On/Off switch

Test point
Green LED
Expected
LED blinks during boot

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP BUS MASTER MALFUNC displayed, transducer Finder LED lights briefly

Transducer appears not to work; Finder LED lights briefly after plugging in transducer but INOP BUS MASTER MALFUNC message appears or no parameters displayed on screen

Ranked causes

  1. 1. Transducer is defective (common)
  2. 2. API Board is defective (occasional)
First check

Observe Finder LED and system messages after plugging in transducer

Test point
transducer connector
Expected
Finder LED lights briefly and parameters appear on screen

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No video on remote display

Remote display connected via MIB/RS232 interface shows no video output

Ranked causes

  1. 1. Video cable not properly connected (common)
  2. 2. Remote display has no power (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper not feeding properly

Fetal recorder paper advance is sluggish or does not move smoothly

Ranked causes

  1. 1. Paper incorrectly loaded (common)
  2. 2. Rubber roller dirty (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Trace not aligned with gridlines

Fetal recorder output trace is offset from paper gridlines

Ranked causes

  1. 1. Offset needs adjusting (common)
  2. 2. Wrong paper scale being used (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unknown transducer connected message displayed

Monitor displays 'Unknown transducer connected' message at bottom of screen, either after software upgrade or without upgrade

Ranked causes

  1. 1. Transducer in boot mode after software upgrade (common)
  2. 2. Transducer is defective (occasional)
First check

Check if message appeared after software upgrade

Test point
transducer history
Expected
message appeared after upgrade = boot mode issue

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

All transducers not working

US, Toco, IUP and ECG transducers all non-functional, INOP BUS MASTER MALFUNC displayed

Ranked causes

  1. 1. API Board is defective (common)
First check

Test multiple transducers in different sockets

Test point
all transducer sockets
Expected
at least one transducer lights Finder LED

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

API Board LED off

LED indicator on API Board is not illuminated

Ranked causes

  1. 1. Power supply issue (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

API Board LED on steadily

LED indicator on API Board is lit continuously rather than blinking

Ranked causes

  1. 1. API Board malfunction (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bad or distorted printout within first 1 cm

Fetal recorder output shows distorted or poor quality marks at start of trace

Ranked causes

  1. 1. Paper drawer not fully closed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Fetal recorder maintenance

Preventive maintenance of recorder platen, thermal printhead and paper sensor to maintain print quality

Ranked causes

  1. 1. Scheduled preventive maintenance (common)
First check

Inspect recorder platen, thermal printhead and paper sensor

Test point
Fetal recorder assembly
Expected
No visible dust or debris; printhead clean and functional

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Noninvasive blood pressure measurement patient safety

Performance test for noninvasive blood pressure measurement system per service requirements

Ranked causes

  1. 1. Scheduled preventive maintenance (common)
First check

Perform full performance testing of blood pressure parameter

Test point
Noninvasive blood pressure measurement system
Expected
Per patient simulator or specified tools test results

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Protective earth leakage current high voltage

Measurement and verification of protective earth leakage current in multiple socket outlet configuration

Ranked causes

  1. 1. Scheduled safety compliance test (common)
First check

Measure protective earth leakage current with all components connected and powered

Test point
Multiple portable socket outlet system
Expected
Sys3 <= 100μA, <= 100

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Transducer belt button replacement caution

Replace belt button on Philips Avalon FM40/FM50 fetal monitor transducer. Procedure uses supplied tool to remove old button and install new one with controlled torque application.

Ranked causes

  1. 1. Belt button broken or requires replacement (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Transducer cable replacement caution

Exchange transducer cable on Philips Avalon FM40/FM50 fetal monitor. Procedure involves removing screw covers, top cover, disconnecting old cable and fitting new one with gasket replacement.

Ranked causes

  1. 1. Transducer cable requires replacement (unknown)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Internal Voltage INOP

INOP message indicates voltage problem in monitor

Ranked causes

  1. 1. Problem with 5V voltages in monitor (unknown)
First check

Remove and test I/O boards sequentially

Test point
Monitor startup
Expected
System operates without voltage INOP

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check TI Config INOP

INOP occurs after software upgrade or option change, NST Report settings require update

Ranked causes

  1. 1. NST Report settings not configured for software feature (common)
First check

Verify Setup NST Report Guideline setting

Test point
Setup menu
Expected
Guideline not set to 'none'

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cyclic reboot loop

Monitor attempts boot, power fails, LED blinks, display briefly appears then goes dark, cycle repeats

Ranked causes

  1. 1. Hardware failure from short circuit, likely NBP board defect (common)
First check

Observe boot sequence behavior

Test point
Green LED and display
Expected
Continuous boot cycle without failure

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inadequate contrast or no paper marks on printout

Fetal recorder output shows poor contrast or missing marks on paper

Ranked causes

  1. 1. Non-approved paper being used (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP FetRec CHK CONF message issued

Recorder configuration error indicated; recorder speed or scale type not configured

Ranked causes

  1. 1. Recorder Speed and/or Scale Type not configured (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP PRINTHEAD OVERHEAT message issued caution

Thermal printhead temperature exceeds safe operating limit

Ranked causes

  1. 1. Printhead too hot (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP WRONG PAPER SCALE message issued

Wrong paper scale loaded; international scale loaded instead of US scale or vice versa

Ranked causes

  1. 1. Wrong scale paper loaded (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal communication malfunction

INOP message displayed indicating internal communication error on main CPU board

Ranked causes

  1. 1. Main CPU Board defective (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch position invalid

Touchscreen responds but cursor position does not match point of contact

Ranked causes

  1. 1. Touch calibration problem (common)
First check

Measure touch offset

Test point
Cursor position vs contact point
Expected
Offset <= 5 mm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transducer belt button broken or damaged patient safety

Transducer belt button or knob is broken, cracked or mechanically damaged

Ranked causes

  1. 1. Mechanical damage to belt button (common)
First check

Visually inspect belt button and knob

Test point
belt button assembly
Expected
no cracks, breaks or damage visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
BUS MASTER MALFUNCBus master section of API Board malfunctionBus master firmware incompatible or API Board defective after software upgradeerror
CheckInternVoltageInternal voltage problem detectedAbnormal 5V or other internal power supply voltage detected during operationwarning
Check Monitor FuncMonitor function check requiredInternal monitor function test failurewarning
Check Monitor TempMonitor temperature warningInternal temperature exceeds safe operating limitwarning
Check SettingsMonitor settings validation failureSettings corruption during operation or after software upgradewarning
Check TI ConfigNST Report configuration issueNST Report Guideline not configured after software upgrade or option changeinfo
DFHR1 EQUIP MALFFetal Heart Rate Channel 1 equipment malfunctionDefective FHR1 transducer, sensor, or associated electronicserror
DFHR2 EQUIP MALFFetal Heart Rate Channel 2 equipment malfunctionDefective FHR2 transducer, sensor, or associated electronicserror
DFHR3 EQUIP MALFFetal Heart Rate Channel 3 equipment malfunctionDefective FHR3 transducer, sensor, or associated electronicserror
FetRec EQUIP MALFFetal Recorder equipment malfunctionDefective fetal recorder or associated electronicserror
FHR1 EQUIP MALFFetal Heart Rate Channel 1 equipment malfunctionDefective FHR1 transducer or associated electronicserror
FHR2 EQUIP MALFFetal Heart Rate Channel 2 equipment malfunctionDefective FHR2 transducer or associated electronicserror
FHR3 EQUIP MALFFetal Heart Rate Channel 3 equipment malfunctionDefective FHR3 transducer or associated electronicserror
INOP CHECK PAPERCheck paper error on fetal recorderPaper sensing error detectedwarning
INOP FetRec CHK CONFFetal recorder configuration check errorRecorder speed or scale type not configuredwarning
INOP FetRec EQUIP MALFFetal recorder equipment malfunctionRecorder not calibrated or controller hardware failureerror
INOP PRINTHEAD OVERHEATThermal printhead overheatingThermal printhead temperature exceeds safe limitwarning
INOP WRONG PAPER SCALEWrong paper scale loaded in recorderPaper scale does not match configuration settingwarning
Internal.Comm.Malf.Internal communication malfunctionMain CPU Board communication errorcritical
IUP EQUIP MALFIntrauterine Pressure equipment malfunctionDefective IUP sensor or associated electronicserror
MECG EQUIP MALFMaternal ECG equipment malfunctionDefective maternal ECG sensor or associated electronicserror
NBP EQUIP MALFNon-invasive Blood Pressure equipment malfunctionDefective NBP sensor or associated electronicserror
Settings Malfunc.Settings malfunctionConfiguration cloning error or corrupted settings memoryerror
Speaker Malfunct.Speaker malfunctionSpeaker cable disconnected, speaker defective, or sound amplifier failurewarning
SpO2 EQUIP MALFOxygen saturation equipment malfunctionDefective SpO2 sensor or associated electronicserror
SpO2 SENSOR MALFOxygen saturation sensor malfunctionDefective SpO2 sensorerror
TOCO EQUIP MALFTocography equipment malfunctionDefective TOCO sensor or associated electronicserror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI