Philips Avalon FM20/FM30: troubleshooting and error codes
The ServiceForge knowledge base holds 92 documented symptoms for the Philips Avalon FM20/FM30, with 175 ranked causes, 138 diagnostic checks with test points and expected values, and 21 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
INOP Check Paper
INOP Check Paper message issued; drawer open with paper on sensor or paper jam or sensor issue
Ranked causes
- 1. Drawer open with paper on sensor (common)
- 2. Paper jam occurred (occasional)
- 3. Paper sensor is dirty (occasional)
- 4. Paper sensor defective (rare)
- 5. Rubber roller dirty (occasional)
- 6. Paper not approved by Philips (rare)
- 7. Inadequate paper mark contrast (rare)
- 8. Paper changed while recorder in operation (rare)
Verify drawer closure and paper position
- Test point
- Paper drawer and paper sensor
- Expected
- Drawer fully closed and latched; paper not contacting sensor
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data; MIB/RS232
External device not receiving data via MIB/RS232 interface
Ranked causes
- 1. MIB/RS232 port not configured for data export (common)
- 2. Wrong data export protocol driver configured (common)
- 3. Cable not connected correctly or defective (common)
- 4. External device does not support data export protocol version (occasional)
- 5. Terminal concentrator does not support dynamic speed negotiation (occasional)
- 6. MIB/RS232 board in wrong slot (occasional)
- 7. MIB/RS232 board or connector board defective (rare)
Verify MIB/RS232 port configuration
- Test point
- Configuration Mode MIB/RS232 settings
- Expected
- Port configured with correct protocol for data export
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data; LAN/RS232
External device such as surveillance system (OB TraceVue/IntelliSpace Perinatal) not receiving data from monitor
Ranked causes
- 1. LAN/RS232 port not configured for data export (common)
- 2. Cable not connected correctly or defective (common)
- 3. External device does not support data export protocol version (occasional)
- 4. Terminal concentrator does not support dynamic speed negotiation (occasional)
- 5. LAN/RS232 board in wrong slot (occasional)
- 6. LAN/RS232 board defective (rare)
Verify LAN/RS232 port configuration
- Test point
- Configuration Mode LAN/RS232 settings
- Expected
- Port configured with correct protocol and parameters for target device
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen not functioning
Touch input on display does not respond to user input
Ranked causes
- 1. Touchscreen functionality has been temporarily disabled (common)
- 2. Touch screen cable not connected (occasional)
- 3. Touch not calibrated (occasional)
- 4. Touch controller defective (rare)
- 5. Touch sensor defective (rare)
- 6. Main CPU Board defective (rare)
Check if touchscreen is disabled via padlock symbol
- Test point
- Main Setup key display
- Expected
- No padlock symbol or padlock disabled
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is blank or brightness is reduced
LCD display does not show image or has reduced brightness during operation
Ranked causes
- 1. Display Adapter Board cable not connected (common)
- 2. Backlight tubes defective (occasional)
- 3. Back-light inverter defective (occasional)
- 4. LCD flat panel defective (occasional)
- 5. Main CPU Board defective (rare)
Verify Display cable connection
- Test point
- Display assembly to Recorder Adapter Board connector
- Expected
- Cable firmly connected
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No paper transport
Paper does not move through recorder mechanism during operation
Ranked causes
- 1. Poor internal connector connection (common)
- 2. Motor cable disconnected (common)
- 3. Motor defective (occasional)
- 4. Paper jam occurred (occasional)
- 5. Paper drawer is open (common)
Check all internal connectors
- Test point
- Internal connector pins and seats
- Expected
- All connectors fully seated and no visible corrosion
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer Finder LED does not light and transducer not working
After plugging in transducer, Finder LED does not light up briefly twice as expected, and transducer appears not to function. May display 'INOP Bus Master Malfunc' message.
Ranked causes
- 1. Defective transducer cable (occasional)
- 2. Defective connector block (occasional)
- 3. Transducer or connector block is defective (occasional)
- 4. Bus Master Board is defective (occasional)
- 5. No power to Bus Master Board (rare)
Visually inspect transducer cable
- Test point
- Transducer cable and connector
- Expected
- No visible damage, cracks, or corrosion on cable or connectors
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Green LED does not light up and no test tone is heard after power switch is turned on
Ranked causes
- 1. No AC mains connection (common)
- 2. Power supply defective (occasional)
- 3. Power On/Off switch defective (occasional)
- 4. Aborted/interrupted or inconsistent software configuration (occasional)
- 5. Main CPU Board defective (rare)
Verify AC power cord connection
- Test point
- AC mains socket and monitor power inlet
- Expected
- Power cord firmly seated and not damaged
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP FetRec Equip Malfunct; no recorder key available
INOP FetRec Equip Malfunct message appears and recorder key not available on display screen
Ranked causes
- 1. Recorder not calibrated (common)
- 2. EEPROM on Recorder Adapter Board defective (rare)
- 3. Recorder Controller on Main CPU Board defective (rare)
- 4. Recorder cable disconnected (common)
Verify recorder cable connection
- Test point
- Recorder cable connectors at both ends
- Expected
- Cable fully inserted and latched at both connections
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or Mouse not functioning
Keyboard/Mouse attached directly to monitor are not responsive or do not work
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse are defective (occasional)
- 3. USB/PS/2 I/O board not properly plugged in (occasional)
- 4. USB/PS/2 I/O board is defective (rare)
Verify keyboard and mouse connections
- Test point
- Monitor USB/PS/2 ports
- Expected
- All cables firmly connected to appropriate ports
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse not functioning
Attached USB/PS2 keyboard or mouse does not respond to input
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. USB/PS2 I/O board not properly plugged in (occasional)
- 4. USB/PS2 I/O board defective (rare)
Verify keyboard/mouse cable connections
- Test point
- Connector ports on monitor rear panel
- Expected
- Cables fully inserted and seated
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper empty warning issued but paper loaded
Paper empty warning appears in status line at bottom of screen but paper is not actually out
Ranked causes
- 1. Paper drawer is open (common)
- 2. Paper jam occurred (occasional)
- 3. Paper sensor is dirty (occasional)
- 4. Paper sensor defective (rare)
Verify paper drawer position
- Test point
- Paper drawer
- Expected
- Drawer fully closed and latched
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no sound issued patient safety
Visual alarm occurs but no audible alarm tone is generated
Ranked causes
- 1. Volume set to 0 (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on Main CPU Board defective (rare)
Check alarm volume setting
- Test point
- Audio settings menu
- Expected
- Alarm volume greater than 0
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery malfunct INOP
INOP tone sounds and battery LED flashes; monitor cannot determine battery status or communication between battery and battery board fails
Ranked causes
- 1. Battery faulty (common)
- 2. Battery board defective (occasional)
- 3. Main board defective (rare)
Test battery in different monitor or charger
- Test point
- Battery
- Expected
- Battery functions in known-good system
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger malfunct INOP
INOP tone sounds; battery LED may flash indicating problem with battery charger in monitor
Ranked causes
- 1. Battery charger defective (common)
- 2. Battery board defective (occasional)
- 3. Main board defective (rare)
Power cycle the monitor
- Test point
- Monitor power button
- Expected
- INOP clears after power off and on cycle
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No test sound issued or INOP Speaker Malfunct message
Monitor does not issue test sound during boot or displays Speaker Malfunct INOP
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Main CPU Board defective (rare)
Verify speaker cable connection
- Test point
- Speaker connector
- Expected
- Speaker cable firmly connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse Call Relay INOP message patient safety
System displays 'INOP message Check Nurse Call Relay' when monitor alarms but nurse call system not activated
Ranked causes
- 1. Flexible Nurse Call Relay I/O board defective (occasional)
- 2. Incorrect configuration present (Relay latency, Relay trigger) (common)
- 3. Cable connection defect (occasional)
Verify monitor configuration settings
- Test point
- Main Setup menu
- Expected
- Relay latency and trigger parameters set per system requirements
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor print quality
Printed output has reduced contrast, faint marks, or unclear trace definition
Ranked causes
- 1. Heat setting needs adjusting (common)
- 2. Thermal printhead dirty (occasional)
- 3. Thermal printhead failure (rare)
Verify and adjust printhead heat setting
- Test point
- Recorder heat adjustment control
- Expected
- Heat setting produces clear, readable marks without smudging
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper blank
Recorder operates normally but no data marks appear on paper
Ranked causes
- 1. Thermal printhead disconnected (common)
- 2. Thermal printhead defective (occasional)
- 3. Wrong side of paper facing up (occasional)
Verify thermal printhead connection
- Test point
- Printhead connector
- Expected
- Connector fully seated and latched
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Settings Malfunct INOP
INOP message indicating problem with monitor settings storage or cloning process
Ranked causes
- 1. Problem during cloning process (common)
- 2. Settings memory space corrupted (occasional)
- 3. Main CPU Board defective (rare)
Reclone configuration file via Support Tool
- Test point
- Support Tool connection and settings file
- Expected
- Configuration file successfully cloned and loaded
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker malfunction
Speaker cable disconnected or speaker defective; INOP Message Malfunct displayed or no alarm sound issued
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on Main CPU Board defective (rare)
Verify speaker cable connection
- Test point
- Speaker connector on main board
- Expected
- Cable fully inserted and seated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Monitor Temp INOP caution
INOP message displayed indicating internal monitor temperature is too high
Ranked causes
- 1. Environment temperature too high or ventilation blocked (common)
- 2. Main Board defective (occasional)
Verify operating environment temperature and ventilation
- Test point
- Monitor placement and ambient temperature
- Expected
- Adequate airflow around monitor; ambient temperature within specifications
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Settings INOP
INOP message displayed during normal operation indicating possible monitor software problem
Ranked causes
- 1. Monitor or patient settings unexpected or corrupted (common)
- 2. Software problem unresolved after default reload (occasional)
Check monitor and patient settings
- Test point
- Setup menu
- Expected
- Settings as expected for current configuration
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check TI Config INOP
INOP message after software upgrade or option change indicating NST Report settings not configured
Ranked causes
- 1. NST Report Guideline not set or missing (common)
- 2. No NST Report Guideline available (occasional)
Check NST Report Guideline setting
- Test point
- Setup menu NST Report Guideline option
- Expected
- Guideline set to appropriate value, not 'none'
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green LED blinks indicating cyclic reboots
Green power LED blinks repetitively indicating the monitor is rebooting continuously
Ranked causes
- 1. Hardware failure (common)
- 2. Software fault (occasional)
Connect Support Tool and detect devices
- Test point
- Support Tool serial connection
- Expected
- Device detected or no device detected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Individual parameter INOP displayed
INOP message for specific parameter measurement (FHR, MECG, NBP, etc.) indicating equipment malfunction
Ranked causes
- 1. Transducer, sensor, patient module, or measurement board defective (common)
- 2. Bus Master firmware incompatible with new software (occasional)
Determine if INOP is FHR1/FHR2/FHR3 related (for multi-transducer systems)
- Test point
- INOP message display
- Expected
- Identify which transducer has malfunction using blue Finder LED
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor loses settings after power loss exceeding one minute
When AC and battery power are lost for more than one minute, the monitor may load User Defaults or retain settings depending on Automat. Default configuration setting
Ranked causes
- 1. Automat. Default set to Yes (unknown)
- 2. Automat. Default set to No but power restored after 48 hours (unknown)
Confirm duration of power loss
- Test point
- Monitor power status
- Expected
- Less than one minute or more than one minute
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No USB devices functioning
None of the connected USB devices operate or are recognized by monitor
Ranked causes
- 1. USB port in monitor defective (common)
- 2. Invalid device combination or defective devices (occasional)
Test each USB device individually on known-good port
- Test point
- Individual USB devices
- Expected
- At least one device functions on known-good USB port
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video on remote display
Connected remote display shows no video output from monitor
Ranked causes
- 1. Video cable not properly connected (common)
- 2. Remote display has no power (common)
Verify video cable connections
- Test point
- Video cable connectors at monitor and display
- Expected
- Cable securely seated at both ends
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noninvasive blood pressure measurement out of accuracy specification
NBP reading differs from reference manometer by more than 3 mmHg or fails leakage/linearity/valve tests
Ranked causes
- 1. NBP measurement requires calibration (common)
- 2. Leakage in pressure system or valve (occasional)
Accuracy test at 280 mmHg
- Test point
- NBP channel 1 and 2 readings
- Expected
- reading below 50 in Service Mode with static pressure applied mmHg
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper not feeding properly
Paper does not advance smoothly through recorder mechanism during operation
Ranked causes
- 1. Paper incorrectly loaded (common)
- 2. Rubber roller dirty (occasional)
Inspect paper loading and alignment
- Test point
- Paper in drawer
- Expected
- Paper centered, aligned properly with feed mechanism, not twisted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature reading questionable; unusually low or high
Tympanic thermometer reading is uncertain, unreliable, or inconsistent; temperature shows as unusually low or high without apparent cause
Ranked causes
- 1. Wrong site mode configured (common)
- 2. Thermometer requires calibration (occasional)
Verify site mode configuration
- Test point
- Thermometer configuration menu
- Expected
- Site mode setting matches measurement method (rectal, oral, tympanic, etc.)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trace not printed correctly with reference to gridlines
Printed trace is offset or misaligned relative to paper gridlines
Ranked causes
- 1. Offset needs adjusting (common)
- 2. Wrong paper scale being used (occasional)
Verify paper scale configuration
- Test point
- Paper pack label and recorder configuration menu
- Expected
- Paper scale label matches configured scale setting exactly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tympanic thermometer does not function
Tympanic thermometer attached to monitor does not operate or respond
Ranked causes
- 1. No connection between thermometer and base station, or base station and monitor (common)
- 2. Thermometer defective (occasional)
Verify base station to monitor connection
- Test point
- Interface cable connection between base station and host monitor
- Expected
- Cable fully connected and secure at both ends
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tympanic thermometer system error other than 12
Tympanic thermometer display shows system error code other than 12
Ranked causes
- 1. Probe cover damaged or not attached correctly (common)
- 2. Temperature probe defective (occasional)
Reset thermometer with probe cover
- Test point
- Probe cover installation
- Expected
- System error clears after installing fresh probe cover
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accessory integrity compromised caution
Accessories (power cords, cables, connectors) show damage or defects
Ranked causes
- 1. Wear, damage, or degradation of accessory materials (occasional)
Visually inspect all detachable power cords and accessory cables
- Test point
- Power cords, patient leads, adapter cables, connectors
- Expected
- No visible cracks, discoloration, or exposed conductors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
All transducers (US, Toco, IUP, ECG) do not work
No transducers of any type function. 'INOP Bus Master Malfunc' message displayed.
Ranked causes
- 1. Bus Master Board is defective (common)
Verify no transducer types function
- Test point
- All transducer inputs (US, Toco, IUP, ECG)
- Expected
- No transducers light Finder LED or produce output
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Applied Part Leakage Current exceeds limit high voltage
Safety test failure: leakage current from applied part to earth exceeds 5000 μA (BF type) under single fault condition with mains on applied part
Ranked causes
- 1. System installation non-compliance or protective measures inadequate (common)
Measure Applied Part Leakage Current from applied part to earth with mains on applied part
- Test point
- Applied part to earth connection
- Expected
- X1 <= 5000 μA (BF)
- Tools
- Safety test device with measuring circuit per IEC 62353
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Applied part leakage current out of spec patient safety
Leakage current from applied part to earth exceeds acceptable limit when mains voltage is applied to applied part
Ranked causes
- 1. Insulation failure in applied part circuit (occasional)
Measure applied part leakage current with mains on applied part (264 V)
- Test point
- Applied part connector (fetal sensor or SpO2, one at a time)
- Expected
- X1 ≤ 50 μA, ≤ 50 μA
- Tools
- Safety Analyzer with test adapter cable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad or distorted printout within first 1 cm
Printed trace shows distortion, poor quality, or missing data in first 1 cm of paper
Ranked causes
- 1. Paper drawer not fully closed before recording (common)
Verify paper drawer closure
- Test point
- Paper drawer latch
- Expected
- Drawer fully closed and latched with audible click
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner not reading information correctly
Barcode information does not appear in the patient admission field after scanning
Ranked causes
- 1. Barcode scanner window is dirty or obstructed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Empty INOP and battery LED flashes patient safety
INOP message and flashing battery LED indicating battery has less than 10 minutes remaining operating time
Ranked causes
- 1. Battery charge depleted; insufficient operating time remaining (common)
Verify battery is charging via M8023A external power supply
- Test point
- Battery LED and M8023A connection
- Expected
- Battery LED yellow (charging) or green (>90% charged) after connection to power supply
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Incompat INOP patient safety
INOP message indicating battery installed is not compatible with this monitor model
Ranked causes
- 1. Incorrect or incompatible battery installed (common)
Verify battery model number
- Test point
- Battery housing or M8023A power supply display
- Expected
- Battery model M4605A installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Low INOP patient safety
INOP message indicating battery has less than 20 minutes of operating time remaining
Ranked causes
- 1. Battery charge insufficient; less than 20 minutes operating time remaining (common)
Initiate battery recharge
- Test point
- M8023A power supply connection and battery LED
- Expected
- Battery LED turns yellow (charging in progress)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery requires maintenance symbol displayed
Battery icon shows maintenance-required indicator on screen; time to empty and time to full display increasing inaccuracy
Ranked causes
- 1. Battery conditioning overdue (common)
Verify battery charging state
- Test point
- Battery LED
- Expected
- Green (>90% charged)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Beeper sounds when power is lost on battery-equipped models
When power is lost (AC source or battery unavailable), an audible beeper alerts the operator; tone can be silenced by pressing On/Standby button
Ranked causes
- 1. Power loss from AC or battery (common)
Verify power source status
- Test point
- AC power and battery
- Expected
- No power available from AC source or battery
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Batt Temp INOP
INOP tone sounds indicating battery temperature is too high
Ranked causes
- 1. Monitor exposed to excessive heat (common)
Verify monitor placement and ambient temperature
- Test point
- Monitor environment
- Expected
- Monitor in temperature range per specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Monitor Func INOP
INOP message displayed indicating problem with monitor voltages
Ranked causes
- 1. Problem with 5 V voltages in the monitor (common)
Isolate defective I/O board
- Test point
- I/O board connections to main board
- Expected
- INOP clears when correct I/O board is reinserted or remains if main board defective
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration mode changes not saved patient safety
Settings changes made in Configuration Mode may be lost if not properly stored
Ranked causes
- 1. Settings not saved individually per Settings Block or Profile (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG electrode contact with conductive parts patient safety
Unsafe ECG measurement if electrodes contact conductive parts or earth
Ranked causes
- 1. ECG electrodes contacting conductive parts or earth (occasional)
Inspect ECG electrode placement
- Test point
- ECG electrodes
- Expected
- Electrodes in contact with skin only, not touching metal or conductive parts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment label not configured
Unique equipment label required for OB TraceVue/IntelliSpace Perinatal system integration not set
Ranked causes
- 1. Equipment label not entered in system (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment leakage current out of spec high voltage
Leakage current from exposed metal parts exceeds acceptable limit during normal operation
Ranked causes
- 1. Insulation degradation of internal wiring or components (occasional)
Measure equipment leakage current with S1 closed (normal condition)
- Test point
- Exposed metal parts (top cover, interface slot)
- Expected
- X1 ≤ 100 μA, ≤ 100 μA
- Tools
- Safety Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment leakage current out of spec under single fault condition high voltage
Leakage current from exposed metal parts exceeds limit when protective earth is disconnected
Ranked causes
- 1. Failed component or compromised insulation under fault condition (occasional)
Measure equipment leakage current with S1 open (single fault condition)
- Test point
- Exposed metal parts (top cover, interface slot)
- Expected
- X1 ≤ 100 μA, ≤ 100 μA
- Tools
- Safety Analyzer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment mountings not correctly installed or insecure patient safety
Mounting hardware or fixtures are loose, improperly installed, or not secure
Ranked causes
- 1. Improper installation or loose fasteners (unknown)
Inspect all equipment mountings
- Test point
- All mounting hardware and fixtures
- Expected
- All mountings correctly installed and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive leakage current from multiple interconnected devices high voltage
Leakage current may exceed safety limits when several monitoring devices are interconnected
Ranked causes
- 1. Multiple monitoring devices interconnected (occasional)
Count interconnected monitoring devices
- Test point
- patient cable connections
- Expected
- Minimal number of interconnected devices
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exterior housing cracked or damaged caution
Visual cracks or damage visible on the equipment exterior housings
Ranked causes
- 1. Physical impact or handling damage (unknown)
Inspect all exterior housings for cracks and damage
- Test point
- Entire exterior surface
- Expected
- No cracks, splits, or damage visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External cables damaged or cracked high voltage
Cables show splits, cracks, or signs of twisting during visual inspection
Ranked causes
- 1. Physical damage, wear, or improper handling (unknown)
Inspect all external cables for damage
- Test point
- All external cables
- Expected
- No splits, cracks, or twisting visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External cables damaged or kinked patient safety
Splits, cracks, or twisting visible on external cables
Ranked causes
- 1. Physical damage or improper handling (unknown)
Inspect all external cables for splits, cracks, and twisting
- Test point
- Full length of each external cable
- Expected
- No splits, cracks, or signs of twisting
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fetal recorder printout quality is faint or degraded
Traces on printout become faint or unclear, indicating need for recorder maintenance
Ranked causes
- 1. Recorder rubber roller, thermal printhead, or paper sensor contaminated (common)
Visual inspection of fetal recorder printout quality
- Test point
- recorder output
- Expected
- clear, legible traces
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foreign bodies or loose parts in device caution
Loose parts or foreign bodies remain inside device after servicing or repair
Ranked causes
- 1. Incomplete inspection or negligent assembly after service (occasional)
Visually inspect interior for loose parts or foreign bodies
- Test point
- All internal cavities and circuit board areas
- Expected
- No loose parts or foreign bodies present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Fuse specification not met
- Green LED stays on continuously
- INOP FetRec Chk Config
- INOP Printhead Overheat
- INOP Wrong Paper Scale
- Internal.Comm.Malf INOP
- Line frequency setting incorrect
- Mechanical integrity compromised
- Monitor does not meet operational requirements on receipt
- Monitor in shipment mode after delivery
- Monitor malfunction during defibrillation
- Monitor malfunction during electrosurgery
- Monitor malfunction during MRI
- Monitor returns to previous operating mode after electrical transient or burst
- Monitor use in flammable anesthetic environment
- Mountings not correctly installed or insecure
- Nurse call relay operation verification
- Paper scale setting not appropriate for installed paper
- Physical damage visible on receipt
- Post-installation system leakage current test not performed
- Protective earth resistance out of spec
- Recorder speed setting not appropriate
- SpO2 measurement accuracy verification
- System configuration validation before installation
- Temperature reading unusually high
- Temperature reading unusually low
- Toco+ transducer adapter cable socket damaged
- Toco+ transducer or patient module adapter cable socket damaged
- Transducer appears not to work
- Transducer belt button broken or damaged
- Tympanic thermometer display blank
- Tympanic thermometer system error 12
32 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Batt Empty INOP | Battery empty; less than 10 minutes operating time remaining | Battery charge monitor detects battery depleted to critical level | critical |
Batt Incompat INOP | Battery incompatible with this monitor | Battery installed is not the correct type for this monitor model | error |
Batt Low INOP | Battery low; less than 20 minutes operating time remaining | Battery charge monitor detects battery charge below 20 minutes operating time threshold | warning |
Bus Master Malfunc | Bus master board communication or function malfunction | Bus master board fails to communicate with other components or firmware incompatible after software upgrade | error |
Check Monitor Func | Monitor voltage check failure; problem detected with internal 5 V power supply | Monitor self-test detects voltage regulation failure | warning |
Check Monitor Temp | Monitor temperature check failure; internal temperature too high | Monitor thermal sensor detects temperature above normal operating range | warning |
Check Settings | Monitor settings check failure; unexpected settings detected or software issue after upgrade | Monitor detects unexpected settings configuration or incomplete software upgrade | warning |
Check TI Config | NST Report configuration check failure after software upgrade or option change | NST Report Guideline setting not configured after software upgrade or option change | warning |
FetRec Equip Malf | Fetal recorder equipment malfunction | Recorder module or related electronics fails or disconnects | error |
FHR1 Equip Malf | Fetal heart rate measurement equipment malfunction on channel 1 | Ultrasound transducer or related electronics on FHR1 channel fails or disconnects | error |
FHR2 Equip Malf | Fetal heart rate measurement equipment malfunction on channel 2 | Ultrasound transducer or related electronics on FHR2 channel fails or disconnects | error |
FHR3 Equip Malf | Fetal heart rate measurement equipment malfunction on channel 3 | Ultrasound transducer or related electronics on FHR3 channel fails or disconnects | error |
INOP NBP Equip Malf | NBP equipment malfunction in monitoring mode | Occurs after failed NBP calibration or accuracy/leakage test failure | error |
Internal.Comm.Malf | Internal communication malfunction on main CPU board | Main CPU board detects communication failure with internal subsystems | error |
IUP Equip Malf | Intrauterine pressure measurement equipment malfunction | Intrauterine pressure sensor or related electronics fails or disconnects | error |
MECG Equip Malf | Maternal ECG measurement equipment malfunction | ECG sensor or related electronics fails or disconnects | error |
NBP Equip Malf | Non-invasive blood pressure measurement equipment malfunction | Blood pressure sensor or related electronics fails or disconnects | error |
Settings Malfunct | Settings storage or cloning malfunction | Settings cloning process fails or settings memory space corrupted | error |
SpO₂ Equip Malf | Pulse oximetry equipment malfunction | Pulse oximetry sensor or related electronics fails or disconnects | error |
SpO₂ Sensor Malf | Pulse oximetry sensor malfunction | Pulse oximetry sensor fails, disconnects, or is incompatible | error |
Toco Equip Malf | Tocometry (contraction) measurement equipment malfunction | Tocometry sensor or related electronics fails or disconnects | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.