Philips Avalon Fetal Monitor FM20: troubleshooting and error codes
The ServiceForge knowledge base holds 51 documented symptoms for the Philips Avalon Fetal Monitor FM20, with 118 ranked causes, 14 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
INOP Check Paper issued
INOP Check Paper message displayed
Ranked causes
- 1. Drawer open with paper on sensor (common)
- 2. Paper jam occurred (occasional)
- 3. Paper sensor dirty (occasional)
- 4. Paper sensor defective (rare)
- 5. Rubber roller dirty (occasional)
- 6. Paper not approved by Philips (occasional)
- 7. Inadequate contrast of paper marks (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is blank or brightness reduced
No image visible on display or display appears very dim
Ranked causes
- 1. Display Adapter Board cable not connected (common)
- 2. Backlight tubes defective (occasional)
- 3. Back-light inverter defective (occasional)
- 4. Display adapter board defective (occasional)
- 5. LCD flat panel defective (occasional)
- 6. Main CPU Board defective (rare)
Verify Display Adapter cable connection
- Test point
- Cable between Display assembly and Recorder Adapter Board
- Expected
- Cable properly connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data
Data is not being transmitted to an external device connected via MIB/RS232 or USB port
Ranked causes
- 1. MIB/RS232 port not configured for data export (common)
- 2. Wrong data export protocol driver configured (common)
- 3. Cable not connected correctly or defective (common)
- 5. Terminal concentrator does not support dynamic speed negotiation (rare)
- 6. MIB/RS232 board in wrong slot (occasional)
- 7. MIB/RS232 board or connector board defective (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No paper transport
Paper does not advance through recorder
Ranked causes
- 1. Poor internal connection (occasional)
- 2. Paper jam occurred (occasional)
- 3. Motor cable disconnected (occasional)
- 4. Motor defective (rare)
- 5. Drawer is open (common)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen not functioning
Touchscreen does not respond to touch input or is disabled
Ranked causes
- 1. Touchscreen functionality temporarily disabled (common)
- 2. Touch screen cable not connected (common)
- 3. Touch controller defective (occasional)
- 4. Touch sensor defective (occasional)
- 5. Main CPU Board defective (rare)
Check for padlock symbol on Main Setup key
- Test point
- Main Setup button display
- Expected
- No padlock symbol
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer Finder LED does not light after plugging in
Transducer Finder LED does not illuminate and transducer does not work, with or without INOP Bus Master Malfunc message
Ranked causes
- 1. Defective transducer cable (common)
- 2. Defective connector block (common)
- 3. Transducer or connector block is defective (common)
- 4. Bus Master Board is defective (occasional)
- 5. No power to Bus Master Board (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Green power LED does not light, no test tone heard from speaker
Ranked causes
- 1. No AC mains connection (common)
- 2. Power supply defective (occasional)
- 3. Power On/Off switch defective (occasional)
- 4. Software configuration interrupted (rare)
- 5. Main CPU Board defective (rare)
Verify power cord connection
- Test point
- Power cord and AC mains socket
- Expected
- Power cord plugged in and undamaged
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse not functioning
Keyboard or Mouse attached directly to monitor not functioning
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. USB/PS/2 I/O board not properly plugged in (occasional)
- 4. USB/PS/2 I/O board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No recorder key available and INOP FetRec Equip Malf issued
Recorder key missing from screen display, INOP FetRec Equip Malf message shown
Ranked causes
- 1. Recorder not calibrated (common)
- 2. EEPROM on Recorder Adapter Board defective (rare)
- 3. Recorder Controller on Main CPU Board defective (rare)
- 4. Recorder cable disconnected (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper Empty warning issued but paper not out
Paper empty warning displayed at bottom of screen though paper is loaded
Ranked causes
- 1. Drawer is open (common)
- 2. Paper jam occurred (occasional)
- 3. Paper sensor dirty (occasional)
- 4. Paper sensor defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no alarm sound issued patient safety
Alarm triggered but no audio output
Ranked causes
- 1. Volume set to 0 (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on Main CPU Board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarm does not activate Nurse Call patient safety
Monitor generates an alarm but Nurse Call system is not activated
Ranked causes
- 1. Incorrect configuration present (Relay latency, Relay trigger) (common)
- 2. Cable connection defective to monitor or nurse call system (common)
- 3. Flexible Nurse Call Relay I/O board is defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No test sound issued
Speaker does not produce test tone during boot or INOP Speaker Malfunct message appears
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Main CPU Board defective (rare)
Verify speaker cable connections
- Test point
- Speaker cable connections
- Expected
- Cables properly connected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper remains blank during normal recorder operation
Recorder running normally but paper output is blank
Ranked causes
- 1. Thermal Printhead disconnected (occasional)
- 2. Thermal Printhead defective (occasional)
- 3. Paper loaded wrong side up (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter INOP displayed patient safety
Equipment malfunction message displayed for transducer, sensor, patient module, or board (FHR1/2/3, DFHR1/2/3, MECG, Toco, IUP, NBP, FetRec, SpO2, Bus Master)
Ranked causes
- 1. Transducer, sensor, patient module, or board connection issue (common)
- 2. Bus master firmware incompatible with software (occasional)
- 3. Bus master board defective (occasional)
Swap component with known good replacement
- Test point
- Parameter measurement after swap
- Expected
- INOP disappears and parameter measures normally
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor print quality
Printout has poor or unclear quality
Ranked causes
- 1. Heat setting needs adjusting (occasional)
- 2. Thermal Printhead dirty (occasional)
- 3. Thermal Printhead failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Settings Malfunct INOP
Settings malfunction INOP message displayed
Ranked causes
- 1. Problem during cloning process (occasional)
- 2. Memory space for settings corrupted (occasional)
- 3. Main CPU Board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker Malfunct INOP
Speaker Malfunct INOP message displayed
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on Main CPU Board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer connected with Bus Master Malfunc message
Transducer is connected but INOP Bus Master Malfunc message is displayed continuously
Ranked causes
- 1. Main CPU board is defective (occasional)
- 2. Transducer is defective (common)
- 3. Interrupted transducer upgrade or software malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green LED blinks continuously
Green power LED blinks indicating cyclic reboots; monitor does not complete boot phase
Ranked causes
- 1. Hardware failure (common)
- 2. Software fault (occasional)
Connect Support Tool and search for defective devices
- Test point
- Monitor connection via crossover cable
- Expected
- Device detected by Support Tool
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor temperature too high caution
Check Monitor Temp INOP message displayed
Ranked causes
- 1. Environmental temperature too high (common)
- 2. Main Board defective (occasional)
Verify ambient operating environment
- Test point
- Room temperature and ventilation
- Expected
- Within manufacturer operating specifications
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor voltage abnormal caution
Check Monitor Func INOP message displayed indicating problem with voltages
Ranked causes
- 1. Defective I/O board (common)
- 2. Main Board defective (occasional)
Remove and reseat I/O boards one at a time
- Test point
- Each I/O board slot
- Expected
- INOP clears when defective board identified
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No video on remote display
Connected remote display is not showing any video output
Ranked causes
- 1. Video cable is not properly connected (common)
- 2. Remote display has no power (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
None of the connected USB devices are functioning
Multiple USB-connected devices attached to the monitor are not working
Ranked causes
- 1. USB port in monitor is defective (occasional)
- 2. Invalid combination of connected devices or devices are defective (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper not feeding properly
Paper does not advance correctly through recorder
Ranked causes
- 1. Paper incorrectly loaded (common)
- 2. Rubber roller dirty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trace not printed correctly with reference to paper gridlines
Printed trace misaligned with paper grid reference lines
Ranked causes
- 1. Offset needs adjusting (common)
- 2. Wrong paper scale being used (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tympanic Thermometer does not function
Tympanic Thermometer is not working or displaying data
Ranked causes
- 1. No connection between base station and thermometer, or base station and monitor (common)
- 2. Thermometer is defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
All transducers do not work with Bus Master Malfunc message
All transducers (US, Toco, IUP, ECG) are not functioning and INOP Bus Master Malfunc is displayed
Ranked causes
- 1. Bus Master Board is defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad or distorted printout within first 1 cm of trace
Poor print quality at beginning of recording, first 1 cm distorted
Ranked causes
- 1. Paper drawer not fully closed before recording (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Empty INOP patient safety
Batt Empty INOP tone, battery LED flashes, alarms cannot be paused or switched off
Ranked causes
- 1. Estimated remaining battery time ≤10 minutes (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Incompat INOP
Batt Incompat INOP tone, incorrect battery installed
Ranked causes
- 1. Incorrect battery type installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Low INOP
Batt Low INOP tone, battery-powered operating time remaining less than 20 minutes
Ranked causes
- 1. Battery charge low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Malfunct INOP patient safety
Battery Malfunct INOP tone, battery LED flashes, alarms cannot be paused or switched off if not connected to AC power
Ranked causes
- 1. Monitor cannot determine battery status or communication problem between battery and battery board (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery symbol not displayed
Battery symbol missing from display
Ranked causes
- 1. Monitor not configured for battery operation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger Malfunct INOP
Charger Malfunct INOP tone, battery LED may flash, problem with battery charger in monitor
Ranked causes
- 1. Problem with battery charger in monitor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Batt Temp INOP caution
Check Batt Temp INOP tone, battery temperature too high
Ranked causes
- 1. Battery temperature too high (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FHR transducer cannot be identified patient safety
When multiple ultrasound transducers are attached, cannot determine which FHR transducer is producing INOP using Finder LED
Ranked causes
- 1. Transducer Finder LED defect (occasional)
Activate Finder LED on suspected transducer
- Test point
- Touchscreen parameter selection; transducer light output
- Expected
- Finder LED illuminates on correct transducer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green LED lights but does not turn off
Green power LED stays on continuously during normal operation
Ranked causes
- 1. Main CPU Board defective (common)
Load new software
- Test point
- Monitor software
- Expected
- Green LED turns off after boot
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP after software upgrade
INOP message occurs following software upgrade, possible incomplete or unsuccessful upgrade
Ranked causes
- 1. Incomplete or unsuccessful software upgrade (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP after software upgrade or option change with NST Report
INOP displayed after software upgrade or option change, monitor settings need updating for NST Report feature
Ranked causes
- 1. NST Report settings not configured (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP FetRec Chk Config issued
INOP FetRec Chk Config message displayed
Ranked causes
- 1. Recorder Speed and/or Scale Type not configured (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP occurs during normal operation
INOP message displayed, possible monitor software problem
Ranked causes
- 1. Monitor software problem (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Printhead Overheat issued caution
INOP Printhead Overheat message displayed, printhead temperature too high
Ranked causes
- 1. Printhead temperature too high (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Wrong Paper Scale issued
INOP Wrong Paper Scale message displayed, wrong scale paper loaded
Ranked causes
- 1. Wrong paper scale loaded (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal.Comm.Malf INOP
Internal communication malfunction INOP message displayed
Ranked causes
- 1. Main CPU Board defective (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse Call Relay INOP message displayed
Monitor displays 'Check Nurse Call Relay' error message
Ranked causes
- 1. Flexible Nurse Call Relay I/O board is defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature reading is unusually high patient safety
Thermometer displays an abnormally elevated temperature value
Ranked causes
- 1. Probe cover is damaged or not attached correctly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature reading is unusually low patient safety
Thermometer displays an abnormally low temperature value
Ranked causes
- 1. Probe, probe cover, or ear canal is obstructed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch calibration incorrect
Touchscreen position invalid or not calibrated
Ranked causes
- 1. Touch not calibrated (common)
Perform touch calibration
- Test point
- Calibration procedure in Service Mode
- Expected
- Calibration completed successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer appears not to work with Finder LED briefly lighting
Transducer Finder LED lights briefly but either INOP Bus Master Malfunc is displayed or transducer is not recognized with no parameters displayed
Ranked causes
- 1. Transducer is defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer belt button is broken or damaged caution
Transducer knob or belt button is mechanically damaged
Ranked causes
- 1. Mechanical damage to belt button (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.