Philips Avalon CL: troubleshooting and error codes
The ServiceForge knowledge base holds 53 documented symptoms for the Philips Avalon CL, with 85 ranked causes, 47 diagnostic checks with test points and expected values, and 51 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
CL transducer out of range patient safety
Numeric display replaced by blank space; CL transducer LED turns cyan and intermittent INOP issued
Ranked causes
- 1. Distance between base station and transducer exceeds operating range (common)
- 2. CL transducer battery depleted (common)
- 3. Base station improperly mounted in metal enclosure (occasional)
- 4. Signal obstruction by building structure (occasional)
- 5. Interference from other Avalon CL or CTS systems on same channel (occasional)
Measure distance and verify line of sight between base station and CL transducer
- Test point
- Physical separation
- Expected
- Distance less than 100 m/300 ft with unobstructed line of sight
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Base station LED does not light when On/Standby button pressed
On/Standby button LED does not illuminate when pressed on base station
Ranked causes
- 1. System needs reset (common)
- 2. Interface cable not firmly connected to fetal monitor (common)
- 3. Fetal monitor is powered off (occasional)
- 4. Interface cable damaged or has defective connections (occasional)
Perform system reset
- Test point
- Base station
- Expected
- LED illuminates
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CL transducer connection failure
CL transducer parameters not activated; descending tone from base station when transducer removed
Ranked causes
- 1. Maximum CL transducers already active (occasional)
- 2. Frequency band mismatch between CL transducer and base station (occasional)
- 3. Antenna disconnected from PCA board (rare)
- 4. RF interference detected (occasional)
Check Tele Info window at monitor for active CL transducers
- Test point
- Monitor display
- Expected
- Fewer than 4 active CL transducers or transducer can activate
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CL Pods cannot be charged
CL Pods fail to charge at cableless charging contacts on base station despite being placed in charging dock.
Ranked causes
- 1. Battery defect (common)
- 2. Cableless charging contacts defect (occasional)
- 3. Base station interface defect (occasional)
Swap CL Pod with known good pod in same charging slot
- Test point
- CL base station charging dock
- Expected
- Known good pod charges normally
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CL Pods not recognized at monitor
Monitor does not display or respond to CL Pods despite pods being present and powered at base station.
Ranked causes
- 1. CL Pods not assigned to base station (common)
- 2. CL Pods out of communication range (common)
- 3. CL Pods improperly charged or battery depleted (common)
Verify CL Pod assignment in monitor settings
- Test point
- Monitor configuration menu
- Expected
- CL Pods listed as assigned to active base station
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Tele symbol displayed on fetal monitor patient safety
The Tele symbol does not appear on the fetal monitor screen when Avalon CL base station is connected
Ranked causes
- 1. Base station not properly connected to fetal monitor (common)
- 2. Fetal monitor cannot recognize base station (occasional)
- 3. Base station in standby mode (common)
Verify interface cable connection
- Test point
- Connection between base station and fetal monitor
- Expected
- Cable firmly connected to both devices
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter measurements not possible patient safety
System fails to perform or display parameter measurements; mode not shown on recorder trace or displays
Ranked causes
- 1. System not connected correctly to fetal monitor (common)
- 2. Interface cable not properly connected (common)
- 3. Fetal monitor recorder switched off (common)
Verify correct interface cable type
- Test point
- Interface cable between base station and fetal monitor
- Expected
- Cable matches technical specifications for this system
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer malfunction patient safety
INOP tone and numeric replaced by question mark on fetal monitor display
Ranked causes
- 1. Transducer not properly assigned to base station (common)
- 2. Transducer out of range or signal blocked (common)
- 3. Interference from nearby wireless device (occasional)
Broadband RF interference causing intermittent noise caution
All telemetry channels show periodic signal degradation with quiet periods alternating with noisy periods; system recovers when noise source is off.
Ranked causes
- 1. Broadband RF noise source active in building (occasional)
- 2. Electrical equipment interference from same frequency band (occasional)
Run RSSI scan during noisy period
- Test point
- Monitor OBR menu RSSI trace
- Expected
- RSSI values elevated on all channels during interference event
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charging contacts dirty or contact spring failure caution
Transducer not assigned to base station or behaves abnormally when docked; charging and communication contacts blocked by dirt or ultrasound gel; springs do not deflect properly
Ranked causes
- 1. Dirt, ultrasound gel, or debris on charging contacts (common)
- 2. Charging contact spring mechanism worn or damaged (occasional)
Visual inspection of charging contacts
- Test point
- Docking contacts on base station and transducer
- Expected
- Contacts clean and clear of visible dirt, gel, or debris
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CL transducer cannot be charged; LED remains off when docked
CL transducer LED does not illuminate and transducer does not charge when placed in dock at base station charging slot.
Ranked causes
- 1. Battery malfunction in transducer (common)
- 2. Charging contacts defect on transducer or base station (occasional)
Swap transducer to different charging slot
- Test point
- CL base station charging slots
- Expected
- LED illuminates in alternate slot
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CL transducer hardware malfunction
One of INOPs cl US Malfunction, cl Toco Batt Malf, or cl ECG/IUP Batt Malf issued at monitor; LED behavior abnormal
Ranked causes
- 1. CL transducer hardware failure (occasional)
- 2. CL transducer firmware not updated after base station firmware update (occasional)
Visually inspect CL transducer for physical damage or LED malfunction
- Test point
- CL transducer housing and LED
- Expected
- No physical damage; LED responds to power and user interaction
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fetal monitor does not recognize connected base station patient safety
No Tele symbol displayed on fetal monitor screen; system connection not established
Ranked causes
- 1. Incorrect interface cable connection or missing cable (common)
- 2. Fetal monitor lacking required software version for base station (occasional)
Verify interface cable is connected
- Test point
- Interface cable connector at monitor and base station
- Expected
- Cable firmly seated on both ends, correct cable model for interface type
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod battery malfunction
INOP tone indicating battery system failure in NBP Pod
Ranked causes
- 1. NBP Pod not properly assigned or out of range (common)
- 2. Wireless interference from nearby device (occasional)
NBP Pod disconnected patient safety
INOP tone indicating lost base station connection
Ranked causes
- 1. Pod not assigned to base station or out of range (common)
- 2. Wireless signal blocked or interference (occasional)
RF interference from on-channel sources caution
Specific telemetry channel shows signal degradation or dropout due to narrow-band RF interference from another device on same frequency.
Ranked causes
- 1. Another telemetry unit operating on same frequency (occasional)
- 2. External RF source interference (radio, paging, ambulance) (occasional)
Run OBR channel scan to identify occupied frequencies
- Test point
- Monitor OBR menu in Service Mode
- Expected
- Scan results show quiet and occupied channels
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod battery malfunction
INOP tone indicating SpO2 Pod battery system failure
Ranked causes
- 1. Pod not assigned or out of range (common)
- 2. Wireless signal interference (occasional)
SpO2 Pod disconnected patient safety
INOP tone indicating lost wireless connection to base station
Ranked causes
- 1. Pod not assigned or out of range (common)
- 2. Signal blocked or interference present (occasional)
System is completely inoperative
Base station and associated components are not functioning
Ranked causes
- 1. Hardware or communications failure (unknown)
- 2. RF-related problem (unknown)
Reset the system
- Test point
- Base station
- Expected
- System powers on and LED illuminates
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weak or intermittent RF signals or signal loss
Cableless transducer signals drop out, degrade, or disappear during monitoring, causing parameter measurements to be lost or transducer unable to reach ready state.
Ranked causes
- 1. Low carrier signal strength below expected level (common)
- 2. RF interference from external sources (occasional)
Observe transducer LED during walk-away test
- Test point
- CL transducer LED
- Expected
- LED off in normal mode; lights cyan and emits beep when out of range
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Avalon CL top cover malfunction caution
Top cover of Avalon CL base station not functioning properly
Ranked causes
- 1. Top cover board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Base station in standby mode patient safety
Tele symbol displays but base station not actively monitoring; transducers not yet assigned
Ranked causes
- 1. User has not powered on base station (common)
Verify base station power state
- Test point
- Power indicator LED on base station
- Expected
- LED illuminated continuously (not blinking)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Base station malfunction detected
INOP tone issued; base station malfunction reported
Ranked causes
- 1. Base station hardware failure (occasional)
Review base station diagnostic procedure
- Test point
- Base station
- Expected
- Diagnostic results indicate pass or fail
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Base station software incompatibility
INOP tone issued; base station not supported by monitor firmware
Ranked causes
- 1. Outdated base station software revision (occasional)
Verify base station firmware version matches monitor requirements
- Test point
- Base station software version
- Expected
- Current software revision installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CL Pods not activated when picked up from base station
CL Pods fail to activate or become operational when removed from charging dock despite being charged.
Ranked causes
- 1. Missing or expired CL2 software license (common)
Verify CL2 software license at monitor
- Test point
- Fetal monitor software menu
- Expected
- CL2 option listed and licensed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod battery cycle limit reached
INOP tone indicating only 50 charge/discharge cycles remain
Ranked causes
- 1. Battery nearing end of service life (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod battery empty
INOP tone indicating low battery time below 30 minutes on blood pressure transducer
Ranked causes
- 1. NBP Pod battery depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod battery end of life
INOP tone indicating battery no longer accepts charge
Ranked causes
- 1. Battery expired after maximum cycles (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod battery incompatible
INOP tone with incompatible battery message
Ranked causes
- 1. Wrong battery type installed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod battery temperature high
INOP tone indicating critically high battery temperature
Ranked causes
- 1. NBP Pod exposed to heat source or covered (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod not in cradle
INOP tone indicating NBP Pod removed from docking cradle
Ranked causes
- 1. Pod removed from cradle (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP Pod temperature too high patient safety
INOP tone indicating battery temperature exceeds safe limit
Ranked causes
- 1. Pod exposed to heat or covered during use (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No transducers or pods active on base station patient safety
Base station connected and powered but no monitoring active; all transducers or pods still docked or unassigned
Ranked causes
- 1. Transducers not assigned to monitoring channels (common)
Verify at least one transducer docked and assigned
- Test point
- Base station docking slots and transducer assignment status
- Expected
- Minimum one CL transducer or CL Pod assigned and active
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OBR channel configuration invalid
INOP tone with configuration error message on monitor
Ranked causes
- 1. Radio channel configuration missing or incorrect (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
OBR configuration incompatibility
INOP Check OBR Config issued at monitor indicating general incompatibility or configuration mismatch
Ranked causes
- 1. OBR configuration incomplete or mismatched with base station type (occasional)
Navigate to OBR configuration setting on fetal monitor
- Test point
- Monitor OBR configuration menu
- Expected
- OBR configuration setting accessible and correct for base station type
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power switch or On/Standby button malfunction after top cover service caution
After opening top cover or completing repair, On/Standby button fails to respond to user input
Ranked causes
- 1. Power switch connection loose after top cover service (common)
Inspect power switch connector
- Test point
- Power switch connector on main board
- Expected
- Connector firmly seated on PCB pins, no loose wires
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod battery cycle limit
INOP tone indicating 50 charge cycles remain before end of life
Ranked causes
- 1. Battery approaching service life limit (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod battery end of life
INOP tone indicating battery no longer holds charge
Ranked causes
- 1. Battery expired after maximum service cycles (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod battery incompatible
INOP tone with incompatible battery warning
Ranked causes
- 1. Wrong battery type in SpO2 Pod (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod battery low
INOP tone indicating battery below 30 minutes remaining
Ranked causes
- 1. SpO2 Pod battery depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod battery temperature critical
INOP tone indicating dangerously high battery temperature
Ranked causes
- 1. SpO2 Pod exposed to heat or covered (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod not in cradle
INOP tone indicating SpO2 Pod not seated in charging cradle
Ranked causes
- 1. Pod removed from cradle (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Pod temperature too high patient safety
INOP tone indicating battery temperature exceeds safe limit during use
Ranked causes
- 1. Pod exposed to heat source or covered (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR incompatible firmware
INOP tone indicating SRR board or firmware incompatible with base station software
Ranked causes
- 1. SRR firmware version mismatch (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR interference on monitor
Short range radio connection experiences interference from another device; INOP tone heard
Ranked causes
- 1. Wireless interference from external device (common)
Verify no other wireless devices operating on same frequency
- Test point
- Operating environment
- Expected
- No SRR interference detected on alternative channel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR invalid channel
INOP tone indicating radio channel configuration error
Ranked causes
- 1. Radio channel or channel mask incorrectly configured (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR malfunction
INOP tone indicating Short Range Radio module failure
Ranked causes
- 1. SRR module hardware failure or connectivity issue (occasional)
Transducer battery empty patient safety
INOP message and tone indicating critically low battery time below 15 minutes
Ranked causes
- 1. Battery fully depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer battery low
INOP message on monitor indicating low battery remaining time on wireless transducer
Ranked causes
- 1. Battery charge depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer battery malfunction
INOP tone with message indicating battery system failure detected
Ranked causes
- 1. Battery charger circuit failure or incompatible battery (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer disconnected patient safety
INOP tone indicating lost connection to base station
Ranked causes
- 1. Wireless connection lost or blocked (common)
Transducer out of range from base station patient safety
Transducer moved beyond communication range; artificial QRS sound replaces ultrasound signal; DECG and MECG waveforms not displayed
Ranked causes
- 1. Transducer distance exceeds wireless communication range (common)
Verify transducer within communication range
- Test point
- Transducer position relative to base station
- Expected
- Transducer signal present (US sound heard, DECG/MECG waveforms displayed)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transducer temperature critical patient safety
INOP tone with temperature warning; charging interrupted if docked
Ranked causes
- 1. Transducer or battery overheated (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Check OBR Config | OBR channel configuration invalid or missing | INOP tone and message during startup or operation | error |
cl ECG/IUP Batt Low | ECG/IUP transducer battery low | Remaining monitoring time below 30 minutes | warning |
clECG/IUP Batt Malf | ECG/IUP transducer battery malfunction | INOP tone and message | error |
clECG/IUP Chk Temp | ECG/IUP transducer temperature critical | INOP tone and message | warning |
cl ECG/IUP Disconn | ECG/IUP transducer disconnected from base station | INOP tone and message | error |
!!cl ECG/IUP Empty | ECG/IUP transducer battery critically low with escalation | Remaining monitoring time below 15 minutes | error |
!!!clECG/IUP Empty | ECG/IUP transducer battery critically low with high escalation | Remaining monitoring time below 15 minutes | critical |
cl ECG/IUP Empty | ECG/IUP transducer battery critically low | Remaining monitoring time below 15 minutes | error |
cl ECG/IUP Malf | ECG/IUP transducer malfunction | INOP tone and message | error |
cl NBP Batt Empty | NBP Pod battery empty or critically low | Remaining battery time below 30 minutes | warning |
!!!cl NBP Batt Empty | NBP Pod battery empty with high escalation | Remaining battery time below 30 minutes | critical |
!!cl NBP Batt Empty | NBP Pod battery empty with escalation | Remaining battery time below 30 minutes with escalation indication | warning |
cl NBP Batt Incomp | NBP Pod battery incompatible | INOP tone and message | error |
cl NBP Batt Low | NBP Pod battery low | Remaining battery time below 2 hours | warning |
cl NBP Batt Malf | NBP Pod battery malfunction | INOP tone and message | error |
cl NBP Batt Temp | NBP Pod battery temperature critically high | INOP tone and message | warning |
cl NBP Check Batt | NBP Pod battery nearing end of life | INOP tone when only 50 charge/discharge cycles remain | warning |
cl NBP Disconnect | NBP Pod disconnected from base station | INOP tone and message | error |
cl NBP No Cradle | NBP Pod not in cradle | INOP tone and message | info |
cl NBP Remove | NBP Pod temperature too high | INOP tone when battery temperature exceeds limit | error |
cl NBP Serv Batt | NBP Pod battery end of service life | INOP tone when battery reaches end of life | error |
!!!cl SpO₂ Batt Empty | SpO2 Pod battery empty with high escalation | Battery critically low with high escalation | critical |
!!cl SpO₂ Batt Empty | SpO2 Pod battery empty with escalation | Battery critically low with escalation indication | error |
cl SpO₂ Batt Incmp | SpO2 Pod battery incompatible | INOP tone and message | error |
cl SpO₂ Batt Low | SpO2 Pod battery low | Remaining battery time below 30 minutes or 2 hours | warning |
cl SpO₂ Batt Malf | SpO2 Pod battery malfunction | INOP tone and message | error |
cl SpO₂ Batt Temp | SpO2 Pod battery temperature critically high | INOP tone when temperature exceeds safe limit | warning |
cl SpO₂ Check Batt | SpO2 Pod battery nearing end of life | INOP tone when 50 charge/discharge cycles remain | warning |
cl SpO₂ Disconnect | SpO2 Pod disconnected from base station | INOP tone and message | error |
cl SpO₂ No Cradle | SpO2 Pod not in cradle | INOP tone and message | info |
cl SpO₂ Remove | SpO2 Pod temperature too high | INOP tone when battery temperature exceeds limit | error |
cl SpO₂ Serv Batt | SpO2 Pod battery end of service life | INOP tone when battery reaches end of life | error |
!!!clToco Batt Empty | Tocodynamometer battery critically low with high escalation | Remaining monitoring time below 15 minutes | critical |
cl Toco Batt Empty | Tocodynamometer battery critically low | Remaining monitoring time below 15 minutes | error |
!!clToco Batt Empty | Tocodynamometer battery critically low with escalation | Remaining monitoring time below 15 minutes | error |
cl Toco Batt Low | Tocodynamometer battery low | Remaining monitoring time below 30 minutes | warning |
cl Toco Batt Malf | Tocodynamometer battery malfunction | INOP tone and message | error |
cl Toco Chk Temp | Tocodynamometer temperature critical | INOP tone and message | warning |
cl Toco Disconnect | Tocodynamometer disconnected from base station | INOP tone and message | error |
cl Toco Malf | Tocodynamometer malfunction | INOP tone and message | error |
cl US Batt Empty | Transducer ultrasound battery critically low | Remaining monitoring time below 15 minutes | error |
!!cl US Batt Empty | Transducer ultrasound battery critically low with escalation | Remaining monitoring time below 15 minutes with additional indication | error |
!!!cl US Batt Empty | Transducer ultrasound battery critically low with high escalation | Remaining monitoring time below 15 minutes with additional indication | critical |
cl US Batt Low | Transducer ultrasound battery low | Remaining monitoring time below 30 minutes | warning |
cl US Batt Malf | Transducer ultrasound battery malfunction | INOP tone and message | error |
cl US Check Temp | Transducer ultrasound temperature critical | INOP tone and message when temperature exceeds safe limit | warning |
cl US Disconnect | Transducer ultrasound disconnected from base station | INOP tone and message | error |
cl US Malfunction | Transducer ultrasound malfunction | INOP tone with numeric replaced by question mark on monitor | error |
SRR Incompatible | SRR board or firmware incompatible with base station | INOP tone when incompatible SRR module detected at startup | error |
SRR Invalid Chan | SRR invalid channel configuration | INOP tone and message | error |
SRR Malfunction | SRR module malfunction | INOP tone and message | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.