Orex PcCR 1417ACL4: troubleshooting and error codes

The ServiceForge knowledge base holds 3 documented symptoms for the Orex PcCR 1417ACL4, with 4 ranked causes, 9 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Sensor voltage measurement

Verification of sensor output voltage during PM assembly inspection

Ranked causes

  1. 1. Sensor off state voltage out of specification (unknown)
  2. 2. Other sensor voltage measurement incorrect (unknown)
First check

Measure voltage on sensor (not J511)

Test point
pin 2 and pin 3
Expected
Sensor on: 5v; Sensor off: 0v

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Erase lamp sensor voltage abnormal

Test point voltage does not change appropriately when lamps operate or fail; expected transition from 0V (dark) to 3V (light) not observed

Ranked causes

  1. 1. Erase lamp failure (common)
First check

Measure voltage at erase lamp sensor test point in dark condition

Test point
Test Point
Expected
0v V

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rollers Sensor replacement

Rollers Sensor needs replacement; verify correct position and LED indicator function after installation

Ranked causes

  1. 1. Rollers Sensor requires replacement (unknown)
First check

Turn on system and complete initialization

Test point
System power
Expected
Initialization complete

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI