On Site Gas Systems POGS 33C: troubleshooting and error codes
The ServiceForge knowledge base holds 114 documented symptoms for the On Site Gas Systems POGS 33C, with 194 ranked causes, 327 diagnostic checks with test points and expected values, and 8 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
System will not reach stand-by pressure
System fails to achieve or maintain target standby pressure. Continues cycling without reaching set pressure.
Ranked causes
- 1. Too much O2 flow (common)
- 2. Medical air system output too high (common)
- 3. System leaks (common)
- 4. Pressure switch setting incorrect (occasional)
- 5. Process valve failure (rare)
- 6. Check valve malfunction (rare)
- 7. Cycle timing incorrect (rare)
- 8. Sieve material blockage or moisture damage (rare)
- 9. Timer motor faulty or broken (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low air tank pressure (psig)
Pressure reading below normal operating range. Insufficient tank pressure for operation.
Ranked causes
- 1. Manual drain valve left open (common)
- 2. Medical air ports open (common)
- 3. Inlet air leaking (occasional)
- 4. Auto drain stuck open (occasional)
- 5. System leaks (occasional)
- 6. Pressure gauge malfunctioning (rare)
- 7. Process valve failure (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not shut off caution
Compressor continues running; pressure does not cut off at 110 psig
Ranked causes
- 1. Compressor inlet/outlet reversed (occasional)
- 2. Tank pressure above 110 psig (common)
- 3. System leak (occasional)
- 4. Pressure switch incorrectly set (occasional)
- 5. Drain line solenoid valve stuck open (rare)
- 6. Blue regulator open or cracked (rare)
Verify inlet/outlet connection
- Test point
- Generator inlet hose and compressor inlet fitting
- Expected
- Generator inlet hose NOT connected to compressor inlet
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Operate/Calibrate mode verification
Analyzer reading does not match expected values when set to Operate (90–96% O2) or Calibrate (20.9% O2 in Air) mode
Ranked causes
- 1. Orifice clogged, restricting flow (common)
- 2. Sensor cable disconnected (occasional)
- 3. Fault codes displayed on analyzer (occasional)
- 4. Sensor expired (occasional)
- 5. O-ring damaged or sensor not securely threaded (occasional)
Verify arrow position on mode selector
- Test point
- Mode selector arrow
- Expected
- Arrow coincides with O2 and Air settings
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No lights on system
Display and indicator lights not functioning. No visual feedback from equipment.
Ranked causes
- 1. Circuit breaker tripped (common)
- 2. House power problem (common)
- 3. Power switch off (common)
- 4. Molex connector loose or bad contact (occasional)
- 5. LED burned out or not connected (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 input no output or reduced output caution
Microboost system produces no O2 output or flow rate below specification
Ranked causes
- 1. O2 generator valve closed (common)
- 2. System leaks (occasional)
- 3. Regulator set too low (occasional)
- 4. Fill valves closed or non-fill valves leaking (occasional)
- 5. Incorrect cylinder size or flow rate configuration (rare)
Check O2 generator valve position
- Test point
- O2 generator valve
- Expected
- Valve at 12:00 (open) position
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen pressure too low
Oxygen output pressure insufficient for system demand
Ranked causes
- 1. Insufficient time for pressure build-up (common)
- 2. Oxygen outlet valve not fully closed (common)
- 3. Manual condensate drain valve open (common)
- 4. Feed air pressure too low (occasional)
- 5. System leaks (occasional)
Generator pressure build time
- Test point
- system startup
- Expected
- 5-10 minutes for pressure to build
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not run
Generator on line but compressor not starting; unit remains on standby
Ranked causes
- 1. Line pressure already at 110 psig (common)
- 2. Pressure switch not properly set (common)
- 3. Heaters tripped or wrong size (occasional)
- 4. Starter box button stuck (occasional)
- 5. Motor overheating (occasional)
Check air tank pressure
- Test point
- Air tank pressure gage
- Expected
- < 110 psig
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO Flow alarm
Carbon monoxide analyzer flow rate is abnormal or blocked
Ranked causes
- 1. Process valve malfunction (occasional)
- 2. Analyzer filter clogged (occasional)
- 3. Sample valve in wrong position (occasional)
- 4. Faulty dryer - no flow post dryer (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor not operating
Feed Air Compressor does not run or generate air supply
Ranked causes
- 1. Compressor power switch in OFF position (common)
- 2. Compressor in standby mode (common)
- 3. Incorrect power source for compressor (occasional)
- 4. Compressor circuit breaker tripped (common)
Compressor power switch position
- Test point
- Feed Air Compressor Power Switch
- Expected
- ON position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator not operating
Generator does not run or start
Ranked causes
- 1. Power switch in OFF position (common)
- 2. Circuit breaker tripped (common)
- 3. Incorrect power source (occasional)
- 4. Generator in standby mode (common)
Power switch position
- Test point
- Generator Power Switch
- Expected
- ON position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator runs continuously (does not switch to standby)
Generator continues running instead of entering standby mode when oxygen pressure reaches target
Ranked causes
- 1. Oxygen being actively used (common)
- 2. Altitude higher than 5000 ft above sea level (occasional)
- 3. Oxygen outlet valve not closed (common)
- 4. Manual condensate drain valve open (common)
Oxygen demand in system
- Test point
- oxygen outlet valves and flow meters
- Expected
- no oxygen flow if system not in use
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
HV booster no O2 input or no output high voltage
HV booster produces no output or insufficient O2 flow
Ranked causes
- 1. O2 generator valve not open (common)
- 2. System leaks (occasional)
- 3. Regulator set too low (occasional)
- 4. Fill valves closed or non-fill valves open (occasional)
Check O2 generator valve open
- Test point
- O2 generator valve
- Expected
- Valve open
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
HV Booster will not start high voltage
High Voltage Booster unit fails to start or stalls quickly
Ranked causes
- 1. Power switch in OFF position (common)
- 2. Incorrect power source for booster (occasional)
- 3. Input pressure too low (common)
- 4. All needle valves closed (common)
HV Booster power switch position
- Test point
- booster power switch
- Expected
- HAND or AUTO position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No O2 output
HV Booster not producing oxygen output
Ranked causes
- 1. O2 generator valve closed (common)
- 2. Output valves closed (common)
- 3. System leaks (occasional)
- 4. Regulator set too low (occasional)
Verify O2 generator valve is open
- Test point
- O2 generator valve
- Expected
- Valve in open position
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
HV booster system no power high voltage
HV booster does not turn on or respond to power input
Ranked causes
- 1. Input power not available (common)
- 2. Manual start switch not in correct position (common)
- 3. Breaker tripped (occasional)
Check power input
- Test point
- power switch and outlet
- Expected
- Power available and switch on
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Flow fault detected by CO monitor caution
CO monitor detects insufficient flow through sensor; audible and visual alarm; prevents accurate CO monitoring
Ranked causes
- 1. Feed Air Compressor off or not building correct pressure (common)
- 2. Dirty or clogged Medical Air HEPA filter (common)
- 3. Clogged particulate filter inside Med-Air 2000 (common)
Recognize Low Flow fault alarm
- Test point
- Med-Air 2000 monitor audible/visual alarm
- Expected
- Audible and visual alarm indicating Low Flow fault
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microboost system no power caution
Microboost system does not turn on or respond to power input
Ranked causes
- 1. Input power not available (common)
- 2. Power switch not engaged (common)
- 3. Power reset button popped out (occasional)
Check power switch and outlet
- Test point
- power supply input
- Expected
- Power available at outlet and switch engaged
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No green light caution
Power indicator light not illuminated; unit not receiving power or breaker trip
Ranked causes
- 1. Breaker tripped (common)
- 2. House power not available (common)
- 3. Loose connections (occasional)
Check circuit breaker status
- Test point
- Blue button circuit breaker
- Expected
- Breaker not tripped
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
No power to the HV Booster unit
Ranked causes
- 1. Input power not supplied (common)
- 2. Manual start switch not engaged (common)
- 3. Circuit breaker tripped (occasional)
Verify power switch is on and outlet is powered
- Test point
- Power switch and outlet
- Expected
- Switch engaged, outlet live
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Carbon Monoxide (CO) and Dew Point Monitor patient safety
Pre-calibrated from factory; CO sensor requires calibration every 3 months using 20.9% zero gas and 20 PPM CO; dew point sensor replaced every 12 months.
Ranked causes
- 1. Routine maintenance calibration (common)
- 2. Dew point sensor end-of-life replacement (common)
Verify CO and Dew Point Monitor display and alarm function
- Test point
- Monitor display visible when Generator Control Panel opened
- Expected
- Display shows current CO content of feed air, proper flow through sensor, dew point of medical air
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO alarm activated (≥10 PPM detected) patient safety
ENMET Med-Air 2000 sounds loud audible alarm when CO ≥10 PPM detected in feed air; visual alarm on control panel; danger of carbon monoxide poisoning
Ranked causes
- 1. CO source near Supply Air Filter Assembly on Feed Air Compressor (common)
- 2. Interfering substances (methane, propane, alcohols, aerosols, solvents, etc.) (occasional)
Recognize audible and visual CO alarm
- Test point
- Monitor display and Audible Alarm Fig 2.7
- Expected
- Loud audible alarm sounding; visual indicator on control panel
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO Dew Point alarm
Dew point measurement from CO analyzer is inaccurate or sensor is faulty
Ranked causes
- 1. Faulty drier (occasional)
- 2. Dew Point sensor fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Crossover Valve does not operate properly
Crossover Valve fails to open in Operate Mode or fails to close in Standby Mode or when power is off. Valve normally warm during operation as it remains energized in Operate Mode.
Ranked causes
- 1. Wiring problem in Crossover Valve circuit (occasional)
- 2. Defective Crossover Valve (occasional)
Verify Crossover Valve opens in Operate Mode
- Test point
- White button at top of Crossover Valve
- Expected
- White button pulled in (valve open)
- Tools
- Pen or small screwdriver
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper sample flow through CO and Dewpoint monitor
Med-Air 2000 front panel does not show expected 1.5 to 2.4 SCFH of sample flow after system pressurization
Ranked causes
- 1. Medical Air Outlet Pressure exceeds 48 PSIG (common)
- 2. Particulate filter in Med-Air 2000 is clogged (occasional)
Verify Medical Air Outlet Pressure
- Test point
- Medical Air Outlet Pressure Gauge
- Expected
- <48 PSIG
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Pressure - Med Air System alarm patient safety
Medical air output pressure or flow is below acceptable levels
Ranked causes
- 1. Leaks in medical air system (common)
- 2. Medical Air flow setting too high (occasional)
Verify Medical Air output flow rate
- Expected
- less than 30 LPM
- Tools
- flow meter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen purity under 90%
Oxygen analyzer displays purity reading below 90%
Ranked causes
- 1. Generator not allowed to build purity (common)
- 2. Oxygen analyzer reading incorrect (occasional)
Generator purge cycle completion
- Test point
- system operation time and oxygen flow rate
- Expected
- 15-20 LPM flowing for 45-60 minutes
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System requires depressurization before maintenance caution
Air and oxygen systems must be depressurized by slowly disconnecting air hose or opening outlet valves before maintenance work
Ranked causes
- 1. Pressurized system can cause sudden air/oxygen release (common)
- 2. Pressurized oxygen supply requires venting (common)
Disconnect air hose slowly
- Test point
- Air hose connection at POGS unit
- Expected
- Hose disconnects with controlled air release
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Timer alarm
Compressor timing sequence cycle is incorrect or not meeting specifications
Ranked causes
- 1. Timing out of specification (occasional)
- 2. Process valves require inspection (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Too Much O2 Flow alarm patient safety
Oxygen output flow rate exceeds safe maximum of 33 LPM
Ranked causes
- 1. Leaks in system (common)
- 2. Too Much O2 Flow - flow meter setting (occasional)
Verify O2 output flow rate
- Expected
- less than 33 LPM
- Tools
- flow meter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inlet filters clogged
Reduced air flow or pressure output from compressor due to dirty inlet filters
Ranked causes
- 1. Air inlet filters dirty or clogged (common)
Inspect air inlet filters
- Test point
- Two black filter housings
- Expected
- Filters clean, no visible dirt or debris
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inlet filters contaminated
Black filter housings require inspection; filter elements may be clogged with environmental debris
Ranked causes
- 1. Environmental dust and debris accumulation (common)
Inspect air inlet filters for dirt and debris
- Test point
- Two black filter housings
- Expected
- Clean filter elements
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inlet filters need inspection
Black filter housings at air inlet require regular inspection for contamination or clogging based on environmental conditions
Ranked causes
- 1. Filter contamination from environmental conditions (common)
Inspect air inlet filters for contamination
- Test point
- Black filter housings
- Expected
- Clean filter elements without visible contamination
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air inlet filters require inspection
Two air inlet filters in compressor need regular inspection; filter housings are black and twist-open to access filter elements
Ranked causes
- 1. Regular maintenance requirement based on environmental conditions (common)
Inspect two air inlet filters
- Test point
- Black filter housings on compressor
- Expected
- Clean filter elements
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air line filter due for replacement caution
Air line filter has reached three-month service interval and requires replacement; located behind lower panel
Ranked causes
- 1. Scheduled maintenance interval reached (common)
Depressurize air line before filter change
- Test point
- Black valve knob bottom left of cabinet
- Expected
- System depressurized
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air line filter needs replacement
Air line filter requires replacement at three month intervals and maintenance cleaning
Ranked causes
- 1. Air line filter aged and requires replacement (common)
Depressurize air line
- Test point
- Black valve knob at bottom left of cabinet
- Expected
- Air pressure released
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air line filter requires replacement
Air line filter element at three-month service interval or when flow restriction detected
Ranked causes
- 1. Three-month service interval reached (common)
Depressurize air line
- Test point
- Black valve knob, bottom left of cabinet
- Expected
- Air line fully depressurized
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air Tank Pressure alarm caution
Compressor air tank pressure is outside operating range (58-93 psig) or not shutting down at 110 psi
Ranked causes
- 1. Tank pressure out of range (occasional)
Measure air tank pressure
- Test point
- air tank pressure gauge
- Expected
- 58 psig to 93 psig, shutdown at 110 psig, 58-93 psig operating, 110 psig shutdown
- Tools
- pressure gauge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air tank water bottle full
External water collection bottle fills with condensed moisture from compressed air
Ranked causes
- 1. Routine moisture condensation (common)
Inspect external water bottle level
- Test point
- Air tank water bottle outside cabinet
- Expected
- Bottle empty or at acceptable level
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air tank water bottle requires emptying
External water bottle on POGS oxygen generator accumulates condensate requiring daily replacement
Ranked causes
- 1. Daily drainage requirement for collected water (common)
Inspect and empty air tank water bottle
- Test point
- External water bottle outside POGS cabinet
- Expected
- Bottle contains water requiring removal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms sound on power-up
Alarms may sound briefly when Generator Power Switch is turned to ON position. Silenced by pressing ALARM ACKNOWLEDGE/AUDIO DEFEAT button on monitor.
Ranked causes
- 1. Normal initialization sequence (common)
Press ALARM ACKNOWLEDGE/AUDIO DEFEAT button
- Test point
- Monitor control panel
- Expected
- Alarms silent; monitor display active
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto Drain Restrictor .031" obstruction clearance caution
The .031 inch restrictor orifice in the auto drain valve is clogged and requires inspection and clearing.
Ranked causes
- 1. Restrictor orifice blocked with foreign object or debris (occasional)
Shut down Generator and Compressor
- Test point
- Equipment controls
- Expected
- Generator and compressor offline
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto start/pressure switch not functioning
Booster does not auto-start at correct pressure
Ranked causes
- 1. Input pressure below switch threshold (common)
Verify input pressure above threshold
- Test point
- Input pressure gauge
- Expected
- > 28 psig
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet filter clogged
Reduced compressor performance or airflow from dirty cabinet filter
Ranked causes
- 1. Cabinet filter dirty or clogged (common)
Inspect cabinet filter
- Test point
- Center of compressor cabinet
- Expected
- Filter clean, no visible dirt
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet filter contaminated
Center cabinet filter requires inspection and replacement; dust and environmental debris may clog the filter
Ranked causes
- 1. Environmental dust accumulation in cabinet (common)
Inspect cabinet filter for dirt accumulation
- Test point
- Center of compressor cabinet
- Expected
- Clean filter element
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet filter needs inspection
Center-mounted cabinet filter requires regular inspection for clogging or contamination based on environmental conditions
Ranked causes
- 1. Filter contamination from environmental conditions (common)
Inspect cabinet filter for contamination
- Test point
- Center of compressor cabinet
- Expected
- Clean filter without visible contamination or clogging
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cabinet filter requires inspection
Cabinet filter located in center of compressor cabinet needs regular inspection and replacement
Ranked causes
- 1. Regular maintenance requirement based on environmental conditions (common)
Inspect cabinet filter
- Test point
- Center of compressor cabinet
- Expected
- Clean filter element
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration of Oxygen Analyzer with Calibration Gas
Two-point calibration procedure using ambient air and calibration gas (80-99.9% purity) to achieve accurate Oxygen Analyzer readings. Regulator must maintain 50 PSIG +/- 5 PSIG calibration gas pressure.
Ranked causes
- 1. Calibration procedure required for accurate oxygen purity measurement (common)
Verify generator warmup time
- Test point
- Oxygen Analyzer
- Expected
- On for at least 30 minutes prior to calibration
- Tools
- Timer
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Air compressor pressure relief system caution
Systematic test and maintenance of the auto drain restrictor to ensure proper pressure relief operation.
Ranked causes
- 1. Routine maintenance and inspection (common)
Verify restrictor fitting is accessible
- Test point
- Lower panel and valve housing
- Expected
- Lower panel removed, valve visible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO monitor zero and span
ENMET Med-Air 2000 CO sensor pre-calibrated from factory; field calibration uses 0 PPM and 20 PPM calibration gases every 3 months
Ranked causes
- 1. Routine calibration maintenance (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Compressor motor replacement high voltage
Procedure to remove and install a replacement compressor motor on the POGS 33C feed air compressor
Ranked causes
- 1. Compressor motor failure or end-of-life replacement (unknown)
Disconnect hoses from Supply Air Inlet and Feed Air Outlet
- Test point
- Underside of Compressor Interior Cover
- Expected
- Hoses disconnected by loosening hose clamps and pulling off hose barb
27 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dew point sensor maintenance
Dew point sensor cannot be field calibrated; requires replacement every 6 months to maintain accurate readings
Ranked causes
- 1. Sensor aging and drift (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: O2 Sensor
O2 sensor requires calibration using certified calibration gas at 90-95% O2 concentration and ambient air reference.
Ranked causes
- 1. O2 sensor drift requiring calibration (common)
Perform 30 minute warm-up of generator
- Test point
- Generator system
- Expected
- System reaches operating temperature
10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen analyzer single-point (ambient air)
Stabilize analyzer for 30 minutes, set valve to CALIBRATE, adjust reading to 20.9% using ambient air reference
Ranked causes
- 1. Routine oxygen purity monitoring calibration (common)
Allow oxygen analyzer temperature stabilization
- Test point
- Generator power
- Expected
- On for at least 30 minutes
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen analyzer two-point with calibration gas
Two-point calibration using ambient air and regulated calibration gas (80-99.9% purity, 92-95% recommended); requires regulator set to 50 PSIG +/- 5 PSIG and 5/32 SMC tubing adapter
Ranked causes
- 1. Need for improved accuracy with known reference gas (occasional)
Warm up oxygen analyzer for temperature stabilization
- Test point
- Generator power
- Expected
- On for at least 30 minutes
13 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure Switch adjustment
Adjust pressure switch to set compressor standby mode pressure at 110 psig; plastic barrel clockwise increases reading, counter-clockwise decreases
Ranked causes
- 1. Pressure switch requires adjustment to 110 psig (common)
Verify pressure switch adjustment
- Test point
- Pressure Switch
- Expected
- 110 psig
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scroll compressor replacement caution
Procedure to remove and install a replacement scroll compressor on the POGS 33C feed air compressor
Ranked causes
- 1. Scroll compressor failure or end-of-life replacement (unknown)
Shutdown feed air compressor and generator
- Test point
- Power switches
- Expected
- Equipment powered down per section 10.6 procedures
34 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Timing Assembly finger adjustment
Align timing assembly fingers on three gears to control compressor valve actuation; fingers must be positioned in specific directions and angular spacing based on 60-second rotation cycle
Ranked causes
- 1. Timing assembly requires finger alignment (common)
Set bottom gear fingers
- Test point
- Bottom gear
- Expected
- Two fingers 180 degrees apart, one up and one down
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO Calibration OFF alarm
Carbon monoxide analyzer calibration is disabled or not set
Ranked causes
- 1. Calibration not performed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO Contamination alarm high voltage
Carbon monoxide detected in system output, exceeding safe limits
Ranked causes
- 1. CO source from fuel powered motors or forklifts (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- CO Faulty Analyzer alarm
- CO monitor calibration fault
- Compressor Pressure Regulator Open too Much
- Cooling Air Filter Element maintenance
- Dew point alarm activated (>39°F detected)
- Equipment fails to supply clean medical air
- External air tank water bottle full
- HEPA Filter replacement - Medical Air
- HEPA Filter replacement - Oxygen
- HEPA filters due for replacement
- HEPA filters need replacement
- HEPA filters require replacement
- Humidity alarm
- Lack of flow to Medical Air Outlet ports
- Low Flow Pressure alarm
- Low Pressure - Check Valves alarm
- Low Pressure - Process Valves alarm
- Low purity output
- Medical Air Filter replacement required
- Monitor resumes operation after power loss
- No air output from oxygen generator
- No pressure reading in Medical Air Gage
- No water drains from manual condensate drain valve
- Oxygen analyzer alarm: Excessive flow condition
- Oxygen analyzer ERx error (serious internal fault)
- Oxygen analyzer FAULT light illuminated
- Oxygen Analyzer low purity reading
- Oxygen check valve leakage
- Oxygen purity drops below alarm threshold
- Particulate Filter Assembly replacement
- Pre-maintenance power disconnection required
- Purity Air temperature high
- Purity Analyzer Calibration alarm
- Purity Excessive Flow alarm
- Purity Faulty Setting alarm
- Purity Leaks alarm
- Purity Sensor Age alarm
- Purity Sensor Orifice Closed alarm
- Purity Sensor Out of Cal alarm
- Purity Warm Up Time alarm
- Replacement: Pressure Switch
- Sensor orifice blocked or restricted
- Side Cooling Air Filter Element maintenance
- Sieve bed check valve failure
- Sieve bed check valve vibration
- Sieve bed protection during removal
- Sieve beds require protection when removed
- System depressurization required
- Test oxygen analyzer alarm function
- Unit requires power disconnection before maintenance
- Water accumulation in air tank
- Water accumulation in POGS air tank
- Water bottle full in POGS oxygen generator
- Water in air tank
54 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
01, 02, 03 | Calibration fault codes for ENMET Med-Air 2000 CO/dew point monitor | Internal sensor calibration drift or fault detected during operation | warning |
≥10 PPM CO | Carbon monoxide alarm indicating dangerous CO level in feed air | CO concentration in feed air reaches or exceeds 10 PPM | critical |
>39°F Dew Point | Dew point alarm indicating excessive moisture in medical air output | Dew point of medical air outlet exceeds 39 degrees Fahrenheit | error |
ALM1 | Oxygen purity alarm indicating purity below preset threshold (89% factory default) | Oxygen purity drops below 89% at generator outlet | warning |
ALM2 | Secondary oxygen purity alarm indicating backup threshold (85% set in analyzer) | Oxygen purity drops below 85% at generator outlet | critical |
ERx | Serious internal fault in Oxygen Analyzer requiring manufacturer repair | Critical internal fault in oxygen analyzer electronics detected | critical |
FAULT | Internal fault detected by Oxygen Analyzer; multiple fault codes possible | Internal electronic or sensor fault detected by oxygen analyzer | warning |
Low Flow | Insufficient flow detected through CO monitor sensor | Feed air flow to CO monitor sensor drops below adequate level | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI