Olympus ESG-400: troubleshooting and error codes
The ServiceForge knowledge base holds 40 documented symptoms for the Olympus ESG-400, with 104 ranked causes, 68 diagnostic checks with test points and expected values, and 137 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Generator does not respond to footswitch or handswitch activation high voltage
Footswitch or handswitch input does not activate generator output
Ranked causes
- 1. Improper footswitch or HF instrument connection (common)
- 2. Damaged footswitch, handswitch or connection cable (occasional)
- 3. Wrong switch pedal or button pressed (common)
- 4. Generator not switched on (common)
- 5. Another switch pedal or button is pressed (occasional)
- 6. Output activated by peripheral equipment (occasional)
- 7. Window displayed on touch-screen (common)
- 8. Incorrect screen displayed (occasional)
- 9. Output mode deactivated or power level set to off (common)
- 10. Generator malfunction (rare)
Verify connections
- Test point
- Footswitch sockets and output socket on rear panel
- Expected
- Footswitch and HF instrument properly seated
5 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrosurgical generator does not respond to footswitch or handswitch activation high voltage
Footswitch or handswitch does not activate output when pressed
Ranked causes
- 1. Improper connection of footswitch or HF instrument to output socket (common)
- 2. Footswitch, handswitch or connection cable is damaged (occasional)
- 3. Incorrect footswitch pedal or handswitch button is pressed (common)
- 4. Electrosurgical generator is not switched on (common)
- 5. Another footswitch pedal or handswitch button is already pressed (occasional)
- 6. Output is activated by peripheral equipment (occasional)
- 7. Window is displayed on touch-screen (occasional)
- 8. Output mode deactivated or power level set to off (occasional)
- 9. Malfunction of the electrosurgical generator (rare)
Verify electrosurgical generator is powered on
- Test point
- Power switch and display/indicators
- Expected
- Unit is on; display shows normal screen
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic mist and smoke evacuation does not work caution
Automatic mist and smoke evacuation system with compatible insufflation unit is not functioning
Ranked causes
- 1. Erroneous settings (common)
- 2. Communication cable not connected (common)
- 3. Erroneous communication cable connection (occasional)
- 4. Insufflation unit malfunction (occasional)
Verify insufflation unit settings
- Test point
- High flow insufflation unit configuration
- Expected
- Correct settings for automatic evacuation
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic mist and smoke evacuation system does not work caution
Automatic mist and smoke evacuation function not operating with compatible high flow insufflation unit
Ranked causes
- 1. Erroneous settings (common)
- 2. Communication cable not connected (common)
- 3. Communication cable connection erroneous (occasional)
- 4. Compatible high flow insufflation unit malfunction (rare)
Verify insufflation unit settings
- Test point
- High flow insufflation unit configuration
- Expected
- Settings match generator mist/smoke evacuation configuration
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrosurgical generator does not respond after pressing power switch high voltage
Unit does not respond when power switch is activated
Ranked causes
- 1. Improper connection of power cord to AC socket or grounded wall outlet (common)
- 2. Grounded wall outlet has wrong or no output voltage (common)
- 3. Power cord is damaged (occasional)
- 4. Malfunction of the electrosurgical generator (rare)
Verify power cord connection
- Test point
- AC power socket on rear panel and wall outlet
- Expected
- Power cord firmly connected at both ends
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator does not respond after pressing power switch high voltage
No response when power switch is activated on electrosurgical generator ESG-400
Ranked causes
- 1. Improper power cord connection (common)
- 2. Wrong or no output voltage at wall outlet (common)
- 3. Damaged power cord (occasional)
- 4. Malfunction of generator (rare)
Verify power cord connection
- Test point
- AC power socket on rear panel, grounded wall outlet
- Expected
- Power cord seated firmly in socket
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
HF instrument not recognized on UNIVERSAL socket
Olympus HF instrument connected to UNIVERSAL socket not detected or recognized by generator
Ranked causes
- 1. Improper HF instrument plug connection to UNIVERSAL socket (common)
- 2. HF instrument does not support recognition capability (occasional)
- 3. Damaged HF instrument or connection cable (occasional)
- 4. Generator malfunction (rare)
Verify HF instrument plug fully inserted
- Test point
- UNIVERSAL socket connection
- Expected
- HF instrument plug fully seated in UNIVERSAL socket
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
HF instrument not recognized when connected to UNIVERSAL socket high voltage
Generator does not recognize Olympus HF instrument connected to UNIVERSAL socket
Ranked causes
- 1. Improper HF instrument plug connection (common)
- 2. HF instrument lacks recognition capability (occasional)
- 3. HF instrument or cable damaged (occasional)
- 4. Generator malfunction (rare)
Verify plug connection
- Test point
- UNIVERSAL socket and HF instrument plug
- Expected
- Plug fully inserted and seated
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output power in RFCoag mode with end of activation signal high voltage
RFCoag mode produces end of activation signal but no output power is delivered
Ranked causes
- 1. Electrode has no tissue contact (common)
- 2. Improper HF instrument plug connection (occasional)
- 3. Damaged HF instrument connection cable (occasional)
- 4. Generator malfunction (rare)
Verify electrode contact
- Test point
- Electrode position on tissue
- Expected
- Electrode in contact with tissue
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output power is delivered with RFCoag mode selected and end-of-activation signal is audible patient safety
Electrosurgical output does not activate when RFCoag mode is selected; end-of-activation tone sounds but no RF power generated
Ranked causes
- 1. Electrode has no contact with tissue (common)
- 2. Improper connection of HF instrument plug with output socket (occasional)
- 3. Damaged HF instrument connection cable (occasional)
- 4. Malfunction of the electrosurgical generator (rare)
Verify electrode contact with tissue
- Test point
- Surgical site; electrode position
- Expected
- Electrode in firm contact with tissue
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not respond after power switch pressed high voltage
No response from electrosurgical generator after power switch activation; may include no display or audible feedback
Ranked causes
- 1. Improper power cord connection (common)
- 2. Grounded wall outlet wrong or no output voltage (common)
- 3. Power cord damaged (occasional)
- 4. Malfunction of electrosurgical generator (rare)
Verify power cord connection
- Test point
- AC power socket on rear panel, grounded wall outlet
- Expected
- Power cord fully seated and secure
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autostart function does not activate output when electrode touches tissue patient safety
Autostart feature fails to activate RF output when electrode makes contact with tissue; activation sound present but no power delivery
Ranked causes
- 1. Autostart function assigned to another output socket (common)
- 2. Long time delay selected for autostart (occasional)
- 3. Generator malfunction (rare)
Verify autostart function socket assignment
- Test point
- Autostart assignment in Mode/Set screen
- Expected
- Autostart assigned to currently used output socket
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autostart function does not activate when electrode touches tissue high voltage
Autostart feature fails to trigger generator output when electrode contacts tissue
Ranked causes
- 1. Autostart assigned to different output socket (occasional)
- 2. Long time delay selected for autostart (occasional)
- 3. Generator malfunction (rare)
Verify autostart socket assignment
- Test point
- Autostart configuration setting
- Expected
- Autostart assigned to correct output socket
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footswitch or handswitch pressed and activation sound audible but no output power delivered patient safety
Activation tone sounds when footswitch/handswitch pressed but no RF energy is generated
Ranked causes
- 1. Footswitch is assigned to another output socket (occasional)
- 2. Improper connection of HF instrument plug with output socket (occasional)
- 3. Malfunction of the electrosurgical generator (rare)
Verify footswitch assignment
- Test point
- Footswitch socket; Mode and Set screens
- Expected
- Footswitch assigned to currently active output socket
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Footswitch or handswitch pressed but no output power delivered high voltage
Activation sound present when footswitch or handswitch button pressed, but no RF power output to electrode or HF instrument
Ranked causes
- 1. Footswitch assigned to another output socket (common)
- 2. Improper HF instrument plug connection to output socket (common)
- 3. Generator malfunction (rare)
Verify footswitch socket assignment
- Test point
- Footswitch assignment setting in menu
- Expected
- Footswitch assigned to currently used output socket
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output power with audible activation sound and footswitch pressed high voltage
Activation sound plays but no power is delivered when footswitch or handswitch is pressed
Ranked causes
- 1. Footswitch assigned to different output socket (occasional)
- 2. Improper HF instrument plug connection (common)
- 3. Generator malfunction (rare)
Verify footswitch socket assignment
- Test point
- Footswitch configuration
- Expected
- Footswitch assigned to correct output socket
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output cannot be deactivated patient safety
Generator output remains active and cannot be turned off by releasing switch
Ranked causes
- 1. Autostart with both electrodes in tissue contact (common)
- 2. Malfunction of footswitch or handswitch (occasional)
- 3. Generator malfunction (rare)
Verify electrode contact
- Test point
- Electrode position on tissue
- Expected
- Electrodes removed from tissue
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output of generator cannot be deactivated patient safety
RF output continues despite releasing footswitch/handswitch; unable to stop power delivery during procedure
Ranked causes
- 1. Autostart function selected with both electrodes touching tissue (common)
- 2. Malfunction of footswitch or handswitch (occasional)
- 3. Generator malfunction (rare)
Check electrode tissue contact status
- Test point
- Both electrode positions at surgical site
- Expected
- At least one electrode not in contact with tissue
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator does not react to front panel push button during standby
Push buttons on front panel are not responsive during standby mode
Ranked causes
- 1. Push button already pressed (common)
- 2. Generator malfunction (rare)
Check for stuck button
- Test point
- Front panel push buttons
- Expected
- All buttons released
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator does not react when push button on front panel pressed during standby
Front panel push buttons not responding during standby mode
Ranked causes
- 1. Push button already pressed (common)
- 2. Malfunction of electrosurgical generator (rare)
Check all front panel buttons
- Test point
- Front panel push buttons
- Expected
- All buttons released and not stuck
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sound is audible during activation patient safety
Electrosurgical generator activates but produces no audible feedback
Ranked causes
- 1. Volume is set to inaudible level (common)
- 2. Malfunction of the electrosurgical generator (rare)
Check volume level
- Test point
- Volume control on rear panel or Menu screen
- Expected
- Volume set to audible level
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch-screen cannot be controlled
Touch-screen is unresponsive to user input
Ranked causes
- 2. Touch-screen not properly calibrated (occasional)
- 3. Touch-screen malfunction (occasional)
Inspect touch-screen surface
- Test point
- Touch-screen display
- Expected
- No objects obstructing
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Volume cannot be adjusted via menu or rear panel
Volume control settings not working when adjusted through software menu or physical rear panel control
Ranked causes
- 1. Error-related signal volume not adjustable (common)
- 2. Malfunction of electrosurgical generator (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Volume cannot be adjusted via menu or rear panel control
Volume adjustment controls are not responsive
Ranked causes
- 1. Error-related signal volume is not adjustable (common)
- 2. Generator malfunction (rare)
Determine signal type
- Test point
- Audio source during activation
- Expected
- Regular operation audio or error signal
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio volume adjustment not responding caution
Sound level does not change when adjusted via MENU menu or rear volume control knob, or no acoustic feedback heard
Ranked causes
- 1. Volume control malfunction (touch screen or rear knob) (unknown)
Volume adjustment via touch screen menu
- Test point
- MENU push button, rocker switch
- Expected
- Sound carried according to setting; acoustic feedback heard
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Communication link test
Verify LINK-IN and LINK-OUT communication socket connectivity via communication cable
Ranked causes
- 1. Communication circuit or cable fault (unknown)
Communication socket connectivity
- Test point
- LINK-IN and LINK-OUT sockets rear housing
- Expected
- Communication cable properly seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Display brightness and volume
Verify display brightness adjustment and audio output via touch screen menu controls
Ranked causes
- 1. Display or audio circuit malfunction (unknown)
Display brightness adjustment
- Test point
- Touch screen menu
- Expected
- Brightness changes according to rocker switch setting
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Push button functionality
Verify all push buttons respond correctly and perform functions per instruction manual
Ranked causes
- 1. Push button mechanical or electrical failure (unknown)
Push button response
- Test point
- All push buttons
- Expected
- Buttons perform per instruction manual description
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Self test
Generator boot sequence with alarm detection and power-on verification
Ranked causes
- 1. Power supply or initialization failure (unknown)
Power on and boot
- Test point
- Power Switch
- Expected
- Generator boots without alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display brightness adjustment not responding caution
Display brightness does not change when adjusted via MENU push button and rocker switch
Ranked causes
- 1. Display brightness control malfunction (unknown)
Display brightness adjustment
- Test point
- MENU push button, rocker switch
- Expected
- Brightness changes according to setting
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrosurgical generator cannot be switched off patient safety
Power switch does not turn off unit; generator continues operating after power switch activation
Ranked causes
- 1. Malfunction of electrosurgical generator (rare)
Attempt emergency power disconnect
- Test point
- Power cord plug, AC power socket, wall outlet
- Expected
- Unit powers off after cord disconnection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator alarm on power-up
Generator produces an alarm during boot sequence instead of powering on cleanly.
Ranked causes
- 1. Unknown fault condition triggering alarm during startup (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator cannot be switched off patient safety
Power switch does not turn off the electrosurgical generator
Ranked causes
- 1. Generator malfunction (common)
Emergency power removal
- Test point
- Power cord plug at AC power socket
- Expected
- Power cord disconnected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator does not react to front panel button during activation
Push buttons are unresponsive while generator is in activation mode
Ranked causes
- 1. Buttons not available during activation (common)
Verify activation status
- Test point
- Footswitch/hand switch status
- Expected
- Switch released and activation stopped
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Generator does not react when button on front panel pressed during activation patient safety
Front panel buttons not responding during activation/operation mode
Ranked causes
- 1. Buttons not available during activation (common)
Deactivate current operation
- Test point
- Footswitch or hand switch control
- Expected
- Footswitch/hand switch released, activation stopped
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Push button functionality check caution
One or more push buttons do not respond to pressing or do not perform their intended function as described in instruction for use
Ranked causes
- 1. Push button mechanical or electrical failure (unknown)
Push button response to press
- Test point
- All push buttons on unit
- Expected
- Each button performs according to instruction for use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch-screen remains dark after power on
Display dark after switching on unit; audible sound indicates partial power but no display output
Ranked causes
- 1. Malfunction of touch-screen (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch-screen remains dark after power-on with audible sound
Touch-screen display is off but system produces sound when powered on
Ranked causes
- 1. Touch-screen malfunction (common)
Verify system powers on
- Test point
- Audio output
- Expected
- Audible sound present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch-screen remains dark after switching on (sound is audible)
Display does not illuminate but unit powers up and produces sound
Ranked causes
- 1. Malfunction of the touch-screen (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not boot without alarm patient safety
Generator fails to start cleanly or triggers alarm during power-on self-test after connecting power cord and pressing Power Switch
Ranked causes
- 1. Unspecified fault detected during self-test (unknown)
Self-test on power-up
- Test point
- Power Switch
- Expected
- Unit boots without alarm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E001 | Open circuit detected | Electrodes of the HF instrument have no proper tissue contact | error |
E002 | Short circuit detected | Electrodes of the HF instrument touch each other | error |
E003 | Docking Connector Error | Docking connector malfunction | error |
E004 | Internal software error | Software malfunction detected during operation | critical |
E004 | Error | Internal software malfunction | critical |
E006 | Non-conductive fluid | Use of non-conductive fluid during bipolar resection or electrode in air environment | error |
E012 | Adjustment incomplete | Device adjustment is missing or incomplete | warning |
E016 | Burn-in incomplete | Device burn-in is missing or incomplete | warning |
E017 | Error | Wrong or broken footswitch or damaged hardware | error |
E017 | Footswitch error | Footswitch malfunction or connection problem detected | error |
E019 | Footswitch combination | Two single pedal or two double pedal footswitches connected | error |
E023 | Invalid serial number | Wrong or missing device serial number | error |
E031 | Error | Wrong docking connector coding signal or broken hardware | error |
E031 | Docking connector coding error | Wrong docking connector coding signal detected | error |
E032 | Docking connector coding error | Wrong docking connector coding signal detected | error |
E038 | Error | Internal hardware error | error |
E038 | Internal hardware error | Hardware malfunction detected | critical |
E045 | Communication error | Communication error with USG-400 or other connected devices | error |
E045 | Error | Communication error with USG-400 or other devices | error |
E048 | Communication error with restart | Communication error with USG-400 or other devices during operation | error |
E049 | Communication error | Communication error with USG-400 or other devices | error |
E052 | Communication error with restart | Communication error during operation | error |
E053 | Communication error | Communication with USG-400 or other devices lost | error |
E054 | Communication error with software update option | Communication error detected | error |
E055 | Communication error | Communication with USG-400 or other devices lost | error |
E056 | Communication error with restart | Communication error during operation | error |
E057 | Communication error with restart option | Communication failure detected | error |
E058 | Error | Internal software or hardware error | error |
E060 | Internal software or hardware error | Internal fault detected | critical |
E099 | Error | Internal hardware error | error |
E100 | Error | Spark Monitor communication error | error |
E103 | Push button pressed during power-up | Push button on front panel held during power-on | warning |
E104 | Footswitch Cut Pedal pressed during power-up | Cut Pedal on Footswitch 1 held during power-on | warning |
E105 | Footswitch Coag Pedal pressed during power-up | Coag Pedal on Footswitch 1 held during power-on | warning |
E106 | Footswitch Cut Pedal pressed during power-up | Cut Pedal on Footswitch 2 held during power-on | warning |
E107 | Footswitch Coag Pedal pressed during power-up | Coag Pedal on Footswitch 2 held during power-on | warning |
E108 | Handswitch pressed | Pressed Cut Handswitch at Monopolar 1 socket detected during power-up | warning |
E109 | Handswitch pressed | Pressed Coag Handswitch at Monopolar 1 socket detected during power-up | warning |
E110 | Handswitch pressed | Pressed Cut Handswitch at Monopolar 2 socket detected during power-up | warning |
E111 | Handswitch pressed | Pressed Coag Handswitch at Monopolar 2 socket detected during power-up | warning |
E112 | Cut Handswitch pressed during power-up | Cut Handswitch at Multifunction socket held during power-on | error |
E113 | Coag Handswitch pressed during power-up | Coag Handswitch at Multifunction socket held during power-on | error |
E114 | Touch-screen touched during power-up | Display screen touched during power-on | warning |
E115 | Application time limit exceeded | Maximum application time has been exceeded during operation | warning |
E131 | Unknown instrument | Olympus HF instrument not properly connected to UNIVERSAL socket or invalid default mode | error |
E132 | Internal hardware error | Hardware malfunction detected | critical |
E133 | Internal hardware error | Hardware malfunction detected | critical |
E134 | Internal hardware error | Hardware malfunction detected | critical |
E135 | Single pedal footswitch not assigned | Footswitch assignment missing during configuration | warning |
E136 | Double pedal footswitch not assigned | Footswitch assignment missing during configuration | warning |
E137 | Auto Start feature incorrectly assigned | Auto Start feature configuration error detected | warning |
E137 | Error | Auto Start feature is not assigned correctly | warning |
E138 | Invalid Thunder Beat mode setting | Thunder Beat mode configuration error detected | error |
E138 | Error | Invalid Thunder Beat mode setting detected | warning |
E139 | Invalid Thunder Beat seal mode setting | Thunder Beat seal mode configuration error detected | error |
E139 | Error | Invalid Thunder Beat seal mode setting detected | warning |
E140 | No mode selected | Mode not selected while attempting to activate | warning |
E141 | Power level set to zero | Activation attempted with zero power level | warning |
E147 | Unknown instrument | HF instrument not properly connected or cannot be identified | error |
E148 | Unknown instrument | HF instrument not properly connected or short circuit at socket | error |
E149 | Unknown instrument | HF instrument not properly connected or malfunction at socket | error |
E151 | Unknown instrument | HF instrument not properly connected or invalid power setting | error |
E152 | Unknown instrument | HF instrument not properly connected or invalid effect setting | error |
E153 | Unknown instrument | HF instrument not properly connected or invalid Cut Mode | error |
E154 | Unknown instrument | HF instrument not properly connected or invalid Coag Mode | error |
E159 | Unknown instrument | HF instrument not properly connected or invalid Mode detected | error |
E166 | Unknown instrument | HF instrument not properly connected or invalid version number | error |
E167 | Unknown instrument | HF instrument not properly connected or internal software error | critical |
E179 | Temperature below limit | Generator Board temperature is too low | unknown |
E180 | Temperature above limit | Generator Board temperature is too high | unknown |
E181 | Temperature below limit | HVPS Board temperature is too low | unknown |
E182 | Temperature above limit | HVPS Board temperature is too high | unknown |
E186 | Error | Internal hardware malfunction | critical |
E186 | Internal hardware error | Hardware malfunction detected | critical |
E187 | Increased HF leakage current | High frequency leakage current exceeded 150 mA (monopolar) or 100 mA (bipolar) | error |
E188 | Excessive HF leakage current | High frequency leakage current exceeded 300 mA (monopolar) or 200 mA (bipolar) | error |
E202 | Insufficient neutral electrode contact | Neutral electrode contact resistance too high or not connected | unknown |
E214 | Low battery | Motherboard battery voltage low causing data loss | error |
E220 | Error | Internal software or hardware malfunction | critical |
E295 | Error | Unable to read data from Multifunction socket | error |
E296 | Error | Unable to write data to Multifunction socket | error |
E297 | Error | Unable to read instrument name from Multifunction socket | error |
E390 | Communication error | Improper connection of communication cables to LINK-OUT/LINK-IN socket | error |
E394 | Communication error | Ring connection detected in communication cables to LINK-OUT/LINK-IN socket | error |
E396 | Temperature below limit | Generator Board temperature too low | unknown |
E397 | Temperature above limit | Generator Board temperature too high | unknown |
E398 | Temperature below limit | HVPS Board temperature too low | unknown |
E399 | Temperature above limit | HVPS Board temperature too high | unknown |
E409 | Communication error | Communication problem detected in cables to LINK-OUT/LINK-IN socket | error |
E410 | Error | Refer to E058 | error |
E411 | Communication error | Communication problem detected in cables to LINK-OUT/LINK-IN socket | error |
E412 | Error | Refer to E058 | error |
E413 | Error | Controlled output power at Burn-In is below limit | warning |
E414 | Error | Controlled output power at Burn-In is above limit | error |
E415 | Error | Refer to E004 | error |
E416 | Error | Refer to E004 | error |
E417 | Error | Refer to E004 | error |
E430 | Error | Power failure detected | critical |
E430 | Power fail | Power failure event | critical |
E431 | Error | Refer to E004 | error |
E432 | Error | Internal hardware error | error |
E433 | Error | Internal hardware error | error |
E438 | Procedure data error | One or more saved procedures have been deleted; inconsistent procedure data detected | error |
E439 | Error | Internal hardware error | error |
E440 | Error | Internal hardware error | error |
E441 | Device setting error | All settings reset to default due to SRAM and EEPROM initialization | warning |
E457 | Error | Internal software or hardware malfunction | critical |
E458 | Error | Refer to E058 | error |
E459 | Error | Refer to E133 | error |
E460 | Error | Refer to E133 | error |
E486 | No instrument connected | Olympus HF instrument not properly connected to UNIVERSAL socket | error |
E490 | Communication error | Communication problem detected in cables to LINK-OUT/LINK-IN socket | error |
E504 | Error | Refer to E058 | error |
E515 | Communication error | Communication problem detected in cables to LINK-OUT/LINK-IN socket | error |
E516 | Communication error | Communication problem detected in cables to LINK-OUT/LINK-IN socket | error |
E525 | Error | Refer to E020 | error |
E526 | Error | Refer to E020 | error |
E539 | Error | Refer to E058 | error |
E540 | Error | Refer to E058 | error |
E558 | Error | Refer to E058 | error |
E560 | Error | Refer to E058 | error |
E561 | Error | Refer to E058 | error |
E562 | Error | Refer to E058 | error |
E617 | Excess coagulum on tip | Tissue built up on instrument tip causing sparking | unknown |
E619 | Data transfer | Device activated during instrument recognition data transfer | unknown |
E646 | Mains voltage drop | Mains voltage too low | unknown |
E650 | Error | Refer to E058 or instrument malfunction | error |
E651 | Error | Refer to E058 or internal hardware error | error |
E652 | Error | Internal software error | error |
E653 | Error | Internal software error | error |
E657 | Short circuit | Electrodes of HF instrument may touch each other | unknown |
E658 | Short circuit | Loop electrode may contact other instruments or metal parts | unknown |
E659 | Error | Internal software error | error |
E661 | Short circuit | Electrodes may touch each other or instrument tip contacts grasper | unknown |
E665 | Application time limit exceeded | Cut and coagulation switches pressed simultaneously | unknown |
E672 | Error | Internal software or hardware error | error |
E673 | Error | Internal software or hardware error | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI