Olympus ESG-400: troubleshooting and error codes

The ServiceForge knowledge base holds 40 documented symptoms for the Olympus ESG-400, with 104 ranked causes, 68 diagnostic checks with test points and expected values, and 137 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Generator does not respond to footswitch or handswitch activation high voltage

Footswitch or handswitch input does not activate generator output

Ranked causes

  1. 1. Improper footswitch or HF instrument connection (common)
  2. 2. Damaged footswitch, handswitch or connection cable (occasional)
  3. 3. Wrong switch pedal or button pressed (common)
  4. 4. Generator not switched on (common)
  5. 5. Another switch pedal or button is pressed (occasional)
  6. 6. Output activated by peripheral equipment (occasional)
  7. 7. Window displayed on touch-screen (common)
  8. 8. Incorrect screen displayed (occasional)
  9. 9. Output mode deactivated or power level set to off (common)
  10. 10. Generator malfunction (rare)
First check

Verify connections

Test point
Footswitch sockets and output socket on rear panel
Expected
Footswitch and HF instrument properly seated

5 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrosurgical generator does not respond to footswitch or handswitch activation high voltage

Footswitch or handswitch does not activate output when pressed

Ranked causes

  1. 1. Improper connection of footswitch or HF instrument to output socket (common)
  2. 2. Footswitch, handswitch or connection cable is damaged (occasional)
  3. 3. Incorrect footswitch pedal or handswitch button is pressed (common)
  4. 4. Electrosurgical generator is not switched on (common)
  5. 5. Another footswitch pedal or handswitch button is already pressed (occasional)
  6. 6. Output is activated by peripheral equipment (occasional)
  7. 7. Window is displayed on touch-screen (occasional)
  8. 8. Output mode deactivated or power level set to off (occasional)
  9. 9. Malfunction of the electrosurgical generator (rare)
First check

Verify electrosurgical generator is powered on

Test point
Power switch and display/indicators
Expected
Unit is on; display shows normal screen

2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Automatic mist and smoke evacuation does not work caution

Automatic mist and smoke evacuation system with compatible insufflation unit is not functioning

Ranked causes

  1. 1. Erroneous settings (common)
  2. 2. Communication cable not connected (common)
  3. 3. Erroneous communication cable connection (occasional)
  4. 4. Insufflation unit malfunction (occasional)
First check

Verify insufflation unit settings

Test point
High flow insufflation unit configuration
Expected
Correct settings for automatic evacuation

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Automatic mist and smoke evacuation system does not work caution

Automatic mist and smoke evacuation function not operating with compatible high flow insufflation unit

Ranked causes

  1. 1. Erroneous settings (common)
  2. 2. Communication cable not connected (common)
  3. 3. Communication cable connection erroneous (occasional)
  4. 4. Compatible high flow insufflation unit malfunction (rare)
First check

Verify insufflation unit settings

Test point
High flow insufflation unit configuration
Expected
Settings match generator mist/smoke evacuation configuration

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrosurgical generator does not respond after pressing power switch high voltage

Unit does not respond when power switch is activated

Ranked causes

  1. 1. Improper connection of power cord to AC socket or grounded wall outlet (common)
  2. 2. Grounded wall outlet has wrong or no output voltage (common)
  3. 3. Power cord is damaged (occasional)
  4. 4. Malfunction of the electrosurgical generator (rare)
First check

Verify power cord connection

Test point
AC power socket on rear panel and wall outlet
Expected
Power cord firmly connected at both ends

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator does not respond after pressing power switch high voltage

No response when power switch is activated on electrosurgical generator ESG-400

Ranked causes

  1. 1. Improper power cord connection (common)
  2. 2. Wrong or no output voltage at wall outlet (common)
  3. 3. Damaged power cord (occasional)
  4. 4. Malfunction of generator (rare)
First check

Verify power cord connection

Test point
AC power socket on rear panel, grounded wall outlet
Expected
Power cord seated firmly in socket

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

HF instrument not recognized on UNIVERSAL socket

Olympus HF instrument connected to UNIVERSAL socket not detected or recognized by generator

Ranked causes

  1. 1. Improper HF instrument plug connection to UNIVERSAL socket (common)
  2. 2. HF instrument does not support recognition capability (occasional)
  3. 3. Damaged HF instrument or connection cable (occasional)
  4. 4. Generator malfunction (rare)
First check

Verify HF instrument plug fully inserted

Test point
UNIVERSAL socket connection
Expected
HF instrument plug fully seated in UNIVERSAL socket

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

HF instrument not recognized when connected to UNIVERSAL socket high voltage

Generator does not recognize Olympus HF instrument connected to UNIVERSAL socket

Ranked causes

  1. 1. Improper HF instrument plug connection (common)
  2. 2. HF instrument lacks recognition capability (occasional)
  3. 3. HF instrument or cable damaged (occasional)
  4. 4. Generator malfunction (rare)
First check

Verify plug connection

Test point
UNIVERSAL socket and HF instrument plug
Expected
Plug fully inserted and seated

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No output power in RFCoag mode with end of activation signal high voltage

RFCoag mode produces end of activation signal but no output power is delivered

Ranked causes

  1. 1. Electrode has no tissue contact (common)
  2. 2. Improper HF instrument plug connection (occasional)
  3. 3. Damaged HF instrument connection cable (occasional)
  4. 4. Generator malfunction (rare)
First check

Verify electrode contact

Test point
Electrode position on tissue
Expected
Electrode in contact with tissue

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No output power is delivered with RFCoag mode selected and end-of-activation signal is audible patient safety

Electrosurgical output does not activate when RFCoag mode is selected; end-of-activation tone sounds but no RF power generated

Ranked causes

  1. 1. Electrode has no contact with tissue (common)
  2. 2. Improper connection of HF instrument plug with output socket (occasional)
  3. 3. Damaged HF instrument connection cable (occasional)
  4. 4. Malfunction of the electrosurgical generator (rare)
First check

Verify electrode contact with tissue

Test point
Surgical site; electrode position
Expected
Electrode in firm contact with tissue

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not respond after power switch pressed high voltage

No response from electrosurgical generator after power switch activation; may include no display or audible feedback

Ranked causes

  1. 1. Improper power cord connection (common)
  2. 2. Grounded wall outlet wrong or no output voltage (common)
  3. 3. Power cord damaged (occasional)
  4. 4. Malfunction of electrosurgical generator (rare)
First check

Verify power cord connection

Test point
AC power socket on rear panel, grounded wall outlet
Expected
Power cord fully seated and secure

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Autostart function does not activate output when electrode touches tissue patient safety

Autostart feature fails to activate RF output when electrode makes contact with tissue; activation sound present but no power delivery

Ranked causes

  1. 1. Autostart function assigned to another output socket (common)
  2. 2. Long time delay selected for autostart (occasional)
  3. 3. Generator malfunction (rare)
First check

Verify autostart function socket assignment

Test point
Autostart assignment in Mode/Set screen
Expected
Autostart assigned to currently used output socket

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Autostart function does not activate when electrode touches tissue high voltage

Autostart feature fails to trigger generator output when electrode contacts tissue

Ranked causes

  1. 1. Autostart assigned to different output socket (occasional)
  2. 2. Long time delay selected for autostart (occasional)
  3. 3. Generator malfunction (rare)
First check

Verify autostart socket assignment

Test point
Autostart configuration setting
Expected
Autostart assigned to correct output socket

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Footswitch or handswitch pressed and activation sound audible but no output power delivered patient safety

Activation tone sounds when footswitch/handswitch pressed but no RF energy is generated

Ranked causes

  1. 1. Footswitch is assigned to another output socket (occasional)
  2. 2. Improper connection of HF instrument plug with output socket (occasional)
  3. 3. Malfunction of the electrosurgical generator (rare)
First check

Verify footswitch assignment

Test point
Footswitch socket; Mode and Set screens
Expected
Footswitch assigned to currently active output socket

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Footswitch or handswitch pressed but no output power delivered high voltage

Activation sound present when footswitch or handswitch button pressed, but no RF power output to electrode or HF instrument

Ranked causes

  1. 1. Footswitch assigned to another output socket (common)
  2. 2. Improper HF instrument plug connection to output socket (common)
  3. 3. Generator malfunction (rare)
First check

Verify footswitch socket assignment

Test point
Footswitch assignment setting in menu
Expected
Footswitch assigned to currently used output socket

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No output power with audible activation sound and footswitch pressed high voltage

Activation sound plays but no power is delivered when footswitch or handswitch is pressed

Ranked causes

  1. 1. Footswitch assigned to different output socket (occasional)
  2. 2. Improper HF instrument plug connection (common)
  3. 3. Generator malfunction (rare)
First check

Verify footswitch socket assignment

Test point
Footswitch configuration
Expected
Footswitch assigned to correct output socket

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Output cannot be deactivated patient safety

Generator output remains active and cannot be turned off by releasing switch

Ranked causes

  1. 1. Autostart with both electrodes in tissue contact (common)
  2. 2. Malfunction of footswitch or handswitch (occasional)
  3. 3. Generator malfunction (rare)
First check

Verify electrode contact

Test point
Electrode position on tissue
Expected
Electrodes removed from tissue

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Output of generator cannot be deactivated patient safety

RF output continues despite releasing footswitch/handswitch; unable to stop power delivery during procedure

Ranked causes

  1. 1. Autostart function selected with both electrodes touching tissue (common)
  2. 2. Malfunction of footswitch or handswitch (occasional)
  3. 3. Generator malfunction (rare)
First check

Check electrode tissue contact status

Test point
Both electrode positions at surgical site
Expected
At least one electrode not in contact with tissue

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator does not react to front panel push button during standby

Push buttons on front panel are not responsive during standby mode

Ranked causes

  1. 1. Push button already pressed (common)
  2. 2. Generator malfunction (rare)
First check

Check for stuck button

Test point
Front panel push buttons
Expected
All buttons released

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator does not react when push button on front panel pressed during standby

Front panel push buttons not responding during standby mode

Ranked causes

  1. 1. Push button already pressed (common)
  2. 2. Malfunction of electrosurgical generator (rare)
First check

Check all front panel buttons

Test point
Front panel push buttons
Expected
All buttons released and not stuck

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No sound is audible during activation patient safety

Electrosurgical generator activates but produces no audible feedback

Ranked causes

  1. 1. Volume is set to inaudible level (common)
  2. 2. Malfunction of the electrosurgical generator (rare)
First check

Check volume level

Test point
Volume control on rear panel or Menu screen
Expected
Volume set to audible level

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch-screen cannot be controlled

Touch-screen is unresponsive to user input

Ranked causes

  1. 2. Touch-screen not properly calibrated (occasional)
  2. 3. Touch-screen malfunction (occasional)
First check

Inspect touch-screen surface

Test point
Touch-screen display
Expected
No objects obstructing

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Volume cannot be adjusted via menu or rear panel

Volume control settings not working when adjusted through software menu or physical rear panel control

Ranked causes

  1. 1. Error-related signal volume not adjustable (common)
  2. 2. Malfunction of electrosurgical generator (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Volume cannot be adjusted via menu or rear panel control

Volume adjustment controls are not responsive

Ranked causes

  1. 1. Error-related signal volume is not adjustable (common)
  2. 2. Generator malfunction (rare)
First check

Determine signal type

Test point
Audio source during activation
Expected
Regular operation audio or error signal

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio volume adjustment not responding caution

Sound level does not change when adjusted via MENU menu or rear volume control knob, or no acoustic feedback heard

Ranked causes

  1. 1. Volume control malfunction (touch screen or rear knob) (unknown)
First check

Volume adjustment via touch screen menu

Test point
MENU push button, rocker switch
Expected
Sound carried according to setting; acoustic feedback heard

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Communication link test

Verify LINK-IN and LINK-OUT communication socket connectivity via communication cable

Ranked causes

  1. 1. Communication circuit or cable fault (unknown)
First check

Communication socket connectivity

Test point
LINK-IN and LINK-OUT sockets rear housing
Expected
Communication cable properly seated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Display brightness and volume

Verify display brightness adjustment and audio output via touch screen menu controls

Ranked causes

  1. 1. Display or audio circuit malfunction (unknown)
First check

Display brightness adjustment

Test point
Touch screen menu
Expected
Brightness changes according to rocker switch setting

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Push button functionality

Verify all push buttons respond correctly and perform functions per instruction manual

Ranked causes

  1. 1. Push button mechanical or electrical failure (unknown)
First check

Push button response

Test point
All push buttons
Expected
Buttons perform per instruction manual description

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Self test

Generator boot sequence with alarm detection and power-on verification

Ranked causes

  1. 1. Power supply or initialization failure (unknown)
First check

Power on and boot

Test point
Power Switch
Expected
Generator boots without alarm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display brightness adjustment not responding caution

Display brightness does not change when adjusted via MENU push button and rocker switch

Ranked causes

  1. 1. Display brightness control malfunction (unknown)
First check

Display brightness adjustment

Test point
MENU push button, rocker switch
Expected
Brightness changes according to setting

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electrosurgical generator cannot be switched off patient safety

Power switch does not turn off unit; generator continues operating after power switch activation

Ranked causes

  1. 1. Malfunction of electrosurgical generator (rare)
First check

Attempt emergency power disconnect

Test point
Power cord plug, AC power socket, wall outlet
Expected
Unit powers off after cord disconnection

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator alarm on power-up

Generator produces an alarm during boot sequence instead of powering on cleanly.

Ranked causes

  1. 1. Unknown fault condition triggering alarm during startup (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator cannot be switched off patient safety

Power switch does not turn off the electrosurgical generator

Ranked causes

  1. 1. Generator malfunction (common)
First check

Emergency power removal

Test point
Power cord plug at AC power socket
Expected
Power cord disconnected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator does not react to front panel button during activation

Push buttons are unresponsive while generator is in activation mode

Ranked causes

  1. 1. Buttons not available during activation (common)
First check

Verify activation status

Test point
Footswitch/hand switch status
Expected
Switch released and activation stopped

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator does not react when button on front panel pressed during activation patient safety

Front panel buttons not responding during activation/operation mode

Ranked causes

  1. 1. Buttons not available during activation (common)
First check

Deactivate current operation

Test point
Footswitch or hand switch control
Expected
Footswitch/hand switch released, activation stopped

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Push button functionality check caution

One or more push buttons do not respond to pressing or do not perform their intended function as described in instruction for use

Ranked causes

  1. 1. Push button mechanical or electrical failure (unknown)
First check

Push button response to press

Test point
All push buttons on unit
Expected
Each button performs according to instruction for use

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch-screen remains dark after power on

Display dark after switching on unit; audible sound indicates partial power but no display output

Ranked causes

  1. 1. Malfunction of touch-screen (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch-screen remains dark after power-on with audible sound

Touch-screen display is off but system produces sound when powered on

Ranked causes

  1. 1. Touch-screen malfunction (common)
First check

Verify system powers on

Test point
Audio output
Expected
Audible sound present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch-screen remains dark after switching on (sound is audible)

Display does not illuminate but unit powers up and produces sound

Ranked causes

  1. 1. Malfunction of the touch-screen (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not boot without alarm patient safety

Generator fails to start cleanly or triggers alarm during power-on self-test after connecting power cord and pressing Power Switch

Ranked causes

  1. 1. Unspecified fault detected during self-test (unknown)
First check

Self-test on power-up

Test point
Power Switch
Expected
Unit boots without alarm

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
E001Open circuit detectedElectrodes of the HF instrument have no proper tissue contacterror
E002Short circuit detectedElectrodes of the HF instrument touch each othererror
E003Docking Connector ErrorDocking connector malfunctionerror
E004Internal software errorSoftware malfunction detected during operationcritical
E004ErrorInternal software malfunctioncritical
E006Non-conductive fluidUse of non-conductive fluid during bipolar resection or electrode in air environmenterror
E012Adjustment incompleteDevice adjustment is missing or incompletewarning
E016Burn-in incompleteDevice burn-in is missing or incompletewarning
E017ErrorWrong or broken footswitch or damaged hardwareerror
E017Footswitch errorFootswitch malfunction or connection problem detectederror
E019Footswitch combinationTwo single pedal or two double pedal footswitches connectederror
E023Invalid serial numberWrong or missing device serial numbererror
E031ErrorWrong docking connector coding signal or broken hardwareerror
E031Docking connector coding errorWrong docking connector coding signal detectederror
E032Docking connector coding errorWrong docking connector coding signal detectederror
E038ErrorInternal hardware errorerror
E038Internal hardware errorHardware malfunction detectedcritical
E045Communication errorCommunication error with USG-400 or other connected deviceserror
E045ErrorCommunication error with USG-400 or other deviceserror
E048Communication error with restartCommunication error with USG-400 or other devices during operationerror
E049Communication errorCommunication error with USG-400 or other deviceserror
E052Communication error with restartCommunication error during operationerror
E053Communication errorCommunication with USG-400 or other devices losterror
E054Communication error with software update optionCommunication error detectederror
E055Communication errorCommunication with USG-400 or other devices losterror
E056Communication error with restartCommunication error during operationerror
E057Communication error with restart optionCommunication failure detectederror
E058ErrorInternal software or hardware errorerror
E060Internal software or hardware errorInternal fault detectedcritical
E099ErrorInternal hardware errorerror
E100ErrorSpark Monitor communication errorerror
E103Push button pressed during power-upPush button on front panel held during power-onwarning
E104Footswitch Cut Pedal pressed during power-upCut Pedal on Footswitch 1 held during power-onwarning
E105Footswitch Coag Pedal pressed during power-upCoag Pedal on Footswitch 1 held during power-onwarning
E106Footswitch Cut Pedal pressed during power-upCut Pedal on Footswitch 2 held during power-onwarning
E107Footswitch Coag Pedal pressed during power-upCoag Pedal on Footswitch 2 held during power-onwarning
E108Handswitch pressedPressed Cut Handswitch at Monopolar 1 socket detected during power-upwarning
E109Handswitch pressedPressed Coag Handswitch at Monopolar 1 socket detected during power-upwarning
E110Handswitch pressedPressed Cut Handswitch at Monopolar 2 socket detected during power-upwarning
E111Handswitch pressedPressed Coag Handswitch at Monopolar 2 socket detected during power-upwarning
E112Cut Handswitch pressed during power-upCut Handswitch at Multifunction socket held during power-onerror
E113Coag Handswitch pressed during power-upCoag Handswitch at Multifunction socket held during power-onerror
E114Touch-screen touched during power-upDisplay screen touched during power-onwarning
E115Application time limit exceededMaximum application time has been exceeded during operationwarning
E131Unknown instrumentOlympus HF instrument not properly connected to UNIVERSAL socket or invalid default modeerror
E132Internal hardware errorHardware malfunction detectedcritical
E133Internal hardware errorHardware malfunction detectedcritical
E134Internal hardware errorHardware malfunction detectedcritical
E135Single pedal footswitch not assignedFootswitch assignment missing during configurationwarning
E136Double pedal footswitch not assignedFootswitch assignment missing during configurationwarning
E137Auto Start feature incorrectly assignedAuto Start feature configuration error detectedwarning
E137ErrorAuto Start feature is not assigned correctlywarning
E138Invalid Thunder Beat mode settingThunder Beat mode configuration error detectederror
E138ErrorInvalid Thunder Beat mode setting detectedwarning
E139Invalid Thunder Beat seal mode settingThunder Beat seal mode configuration error detectederror
E139ErrorInvalid Thunder Beat seal mode setting detectedwarning
E140No mode selectedMode not selected while attempting to activatewarning
E141Power level set to zeroActivation attempted with zero power levelwarning
E147Unknown instrumentHF instrument not properly connected or cannot be identifiederror
E148Unknown instrumentHF instrument not properly connected or short circuit at socketerror
E149Unknown instrumentHF instrument not properly connected or malfunction at socketerror
E151Unknown instrumentHF instrument not properly connected or invalid power settingerror
E152Unknown instrumentHF instrument not properly connected or invalid effect settingerror
E153Unknown instrumentHF instrument not properly connected or invalid Cut Modeerror
E154Unknown instrumentHF instrument not properly connected or invalid Coag Modeerror
E159Unknown instrumentHF instrument not properly connected or invalid Mode detectederror
E166Unknown instrumentHF instrument not properly connected or invalid version numbererror
E167Unknown instrumentHF instrument not properly connected or internal software errorcritical
E179Temperature below limitGenerator Board temperature is too lowunknown
E180Temperature above limitGenerator Board temperature is too highunknown
E181Temperature below limitHVPS Board temperature is too lowunknown
E182Temperature above limitHVPS Board temperature is too highunknown
E186ErrorInternal hardware malfunctioncritical
E186Internal hardware errorHardware malfunction detectedcritical
E187Increased HF leakage currentHigh frequency leakage current exceeded 150 mA (monopolar) or 100 mA (bipolar)error
E188Excessive HF leakage currentHigh frequency leakage current exceeded 300 mA (monopolar) or 200 mA (bipolar)error
E202Insufficient neutral electrode contactNeutral electrode contact resistance too high or not connectedunknown
E214Low batteryMotherboard battery voltage low causing data losserror
E220ErrorInternal software or hardware malfunctioncritical
E295ErrorUnable to read data from Multifunction socketerror
E296ErrorUnable to write data to Multifunction socketerror
E297ErrorUnable to read instrument name from Multifunction socketerror
E390Communication errorImproper connection of communication cables to LINK-OUT/LINK-IN socketerror
E394Communication errorRing connection detected in communication cables to LINK-OUT/LINK-IN socketerror
E396Temperature below limitGenerator Board temperature too lowunknown
E397Temperature above limitGenerator Board temperature too highunknown
E398Temperature below limitHVPS Board temperature too lowunknown
E399Temperature above limitHVPS Board temperature too highunknown
E409Communication errorCommunication problem detected in cables to LINK-OUT/LINK-IN socketerror
E410ErrorRefer to E058error
E411Communication errorCommunication problem detected in cables to LINK-OUT/LINK-IN socketerror
E412ErrorRefer to E058error
E413ErrorControlled output power at Burn-In is below limitwarning
E414ErrorControlled output power at Burn-In is above limiterror
E415ErrorRefer to E004error
E416ErrorRefer to E004error
E417ErrorRefer to E004error
E430ErrorPower failure detectedcritical
E430Power failPower failure eventcritical
E431ErrorRefer to E004error
E432ErrorInternal hardware errorerror
E433ErrorInternal hardware errorerror
E438Procedure data errorOne or more saved procedures have been deleted; inconsistent procedure data detectederror
E439ErrorInternal hardware errorerror
E440ErrorInternal hardware errorerror
E441Device setting errorAll settings reset to default due to SRAM and EEPROM initializationwarning
E457ErrorInternal software or hardware malfunctioncritical
E458ErrorRefer to E058error
E459ErrorRefer to E133error
E460ErrorRefer to E133error
E486No instrument connectedOlympus HF instrument not properly connected to UNIVERSAL socketerror
E490Communication errorCommunication problem detected in cables to LINK-OUT/LINK-IN socketerror
E504ErrorRefer to E058error
E515Communication errorCommunication problem detected in cables to LINK-OUT/LINK-IN socketerror
E516Communication errorCommunication problem detected in cables to LINK-OUT/LINK-IN socketerror
E525ErrorRefer to E020error
E526ErrorRefer to E020error
E539ErrorRefer to E058error
E540ErrorRefer to E058error
E558ErrorRefer to E058error
E560ErrorRefer to E058error
E561ErrorRefer to E058error
E562ErrorRefer to E058error
E617Excess coagulum on tipTissue built up on instrument tip causing sparkingunknown
E619Data transferDevice activated during instrument recognition data transferunknown
E646Mains voltage dropMains voltage too lowunknown
E650ErrorRefer to E058 or instrument malfunctionerror
E651ErrorRefer to E058 or internal hardware errorerror
E652ErrorInternal software errorerror
E653ErrorInternal software errorerror
E657Short circuitElectrodes of HF instrument may touch each otherunknown
E658Short circuitLoop electrode may contact other instruments or metal partsunknown
E659ErrorInternal software errorerror
E661Short circuitElectrodes may touch each other or instrument tip contacts grasperunknown
E665Application time limit exceededCut and coagulation switches pressed simultaneouslyunknown
E672ErrorInternal software or hardware errorerror
E673ErrorInternal software or hardware errorerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.