Nova Biomedical Stat Profile Prime Plus: troubleshooting and error codes

The ServiceForge knowledge base holds 30 documented symptoms for the Nova Biomedical Stat Profile Prime Plus, with 36 ranked causes, 35 diagnostic checks with test points and expected values, and 177 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Dye calibration absorbance value outside valid range

CO-Ox calibration fails because measured absorbance of calibration dye (Standard B) at 563 nm does not fall within 0.5 to 1.5 mAbs

Ranked causes

  1. 1. Air bubbles in the cuvette (common)
  2. 2. Air leaks in the flowpath (common)
  3. 3. Contamination on inside surface of cuvette gradually dissolving (occasional)
  4. 4. Change in sensor card and/or calibration pack (occasional)
First check

Verify dye absorbance value is within acceptable range

Test point
Calibration dye absorbance at 563 nm
Expected
0.5 to 1.5 mAbs, 0.5 to 1.5

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer door open

Analyzer's thermal printer door sensor indicates printer door is open

Ranked causes

  1. 1. Printer door not closed (common)
  2. 2. Door sensor malfunction (occasional)
First check

Inspect printer door and close securely

Test point
Printer door
Expected
Door fully closed and latched

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer out of paper

Analyzer's thermal printer paper sensor does not detect available roll of paper

Ranked causes

  1. 1. Printer paper depleted (common)
  2. 2. Paper sensor malfunction (occasional)
First check

Verify printer paper is installed and available

Test point
Printer paper slot
Expected
Paper roll loaded and sensor detects it

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Probe failed to home

Sampler unable to move sample probe to required position

Ranked causes

  1. 1. Sampler obstruction or mechanical jam (occasional)
  2. 2. Sampler control electronics failure (occasional)
First check

Visually inspect sampler path for obstructions

Test point
Sampler probe and housing
Expected
No obstructions visible

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air detector voltage reading abnormal

Real-time air detector voltage readings do not fall within expected range or status indicates malfunction

Ranked causes

  1. 1. Air detector sensor malfunction or calibration drift (unknown)
First check

Monitor air detector voltage in real time

Test point
AD-50, AD-130, AD-145, AD-REF, AD-Hemo
Expected
Voltage within normal operating range and status indicates functional

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Analyte A-to-A drift

During last analysis sequence, indicated analyte's performance changed significantly from previous analysis

Ranked causes

  1. 1. Sensor cartridge degradation (common)
First check

Compare current analysis result to previous analysis for same analyte

Test point
Analyte result from consecutive samples
Expected
Result variance within normal limits

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Analyte slope drift

During last calibration, indicated analyte's performance changed significantly from previous calibration

Ranked causes

  1. 1. Sensor cartridge degradation (common)
First check

Monitor sensor slope response during calibration

Test point
Sensor slope during 2-point calibration
Expected
Slope within acceptable range, consistent with previous calibrations

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Analyte zero drift after calibration

During last analysis sequence, indicated analyte's performance changed significantly from successful 2-point calibration

Ranked causes

  1. 1. Sensor cartridge degradation or contamination (common)
First check

Run calibration sequence and monitor analyte drift

Test point
Analyte response during 2-point calibration
Expected
Analyte performance within acceptable range after successful calibration

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration lamp not producing signal to spectrometer

Calibration lamp test shows no peaks or incorrect peak position on spectrometer graph

Ranked causes

  1. 1. Calibration lamp light source malfunction (unknown)
First check

Read calibration lamp pixel output to spectrometer

Test point
Spectrometer graph display
Expected
2 tall peaks on right side of graph, rightmost peak within ±10 pixels of 800th pixel, ±10 pixels

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibrator days expiration

Indicated cartridge has exceeded its use life expiration date

Ranked causes

  1. 1. Calibrator use life exceeded (common)
First check

Verify calibrator days-in-use expiration

Test point
Calibrator installation date and use days
Expected
Days in use has not exceeded maximum allowed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibrator invalid part number

Indicated cartridge programmed with part number not intended for use with analyzer

Ranked causes

  1. 1. Wrong calibrator installed (occasional)
First check

Verify calibrator cartridge part number

Test point
Calibrator cartridge part number
Expected
Part number matches analyzer specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibrator lot expiration

Indicated cartridge has exceeded its shelf life expiration date

Ranked causes

  1. 1. Expired calibrator (common)
First check

Check calibrator lot expiration date

Test point
Calibrator lot code and expiration date
Expected
Expiration date has not been reached

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibrator sample use expiration

Indicated cartridge has exceeded maximum number of cycles (samples)

Ranked causes

  1. 1. Sample cycle limit reached (common)
First check

Check calibrator sample count versus maximum

Test point
Calibrator sample counter
Expected
Sample count below maximum allowable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Clamp positioning failure

Peristaltic pump clamp assembly failed to open or close correctly

Ranked causes

  1. 1. Clamp actuator control failure (occasional)
First check

Power cycle analyzer to reset clamp control

Test point
Analyzer power and clamp response
Expected
Clamp opens and closes properly after power on

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Component temperature out of specification during diagnostics

Real-time temperature readings of ABG plate, nest, coil, preheater, spectrometer, BUN/Creat plate, or LED heater do not match expected 37°C

Ranked causes

  1. 1. Heater malfunction or thermal sensor drift (unknown)
First check

Measure real-time component temperature

Test point
ABG Plate, Nest, Coil, Preheater, Spectrometer, BUN/Creat Plate
Expected
37°C for all components (Ambient temperature per specifications)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Conditional calibration flagged

Dye absorbance value is within 0.5-1.5 mAbs range but differs from 5-value average by more than tolerance, marked with C flag

Ranked causes

  1. 1. Absorbance variation exceeds ±1.5% from 5-value average (common)
First check

Review CO-Ox Sequence Execution menu for conditional calibrations

Test point
CO-Ox Sequence Execution menu
Expected
List showing valid (V) and conditional (C) calibrations

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Contamination detected in sensor card cuvette caution

During last light scan, debris was detected in sensor card cuvette

Ranked causes

  1. 1. Cuvette contamination or debris (occasional)
First check

Perform Flush procedure

Test point
Flowpath during flush
Expected
Debris cleared from cuvette

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

COOX Calibration drift with existing card/pack

During last calibration, CO-Ox dye absorbance value fell outside expected range or did not agree within certain percentage of average from earlier calibrations

Ranked causes

  1. 1. Sensor card or calibrator pack degradation (occasional)
First check

Review dye calibration validity conditions per section 5.4.2

Test point
Dye absorbance value during calibration
Expected
Absorbance within expected range and consistent with previous calibrations

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failed to X-Calibrate Spectrometer

CO-Oximeter module unable to initiate or complete X-Axis Optical Calibration

Ranked causes

  1. 1. Spectrometer optical calibration failure (occasional)
First check

Run manual X Calibration from CO-Ox menu

Test point
X Calibration sequence in CO-Ox Sequence Execution Menu
Expected
X Calibration completes successfully

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

I2C communication failure caution

During last sequence, inter-integrated circuit (I2C) channel unable to communicate with module or component

Ranked causes

  1. 1. Fluid contamination on EEPROM or sensor card microchip pins (occasional)
First check

Visually inspect SC Nest, Preheater, and sensor card connectors for fluid

Test point
EEPROM pins and sensor card microchip contacts
Expected
No fluid or corrosion on pins

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LED light not responding to brightness adjustment

LED in sensor module does not change brightness when DAC value is adjusted during diagnostics

Ranked causes

  1. 1. LED illumination optics malfunction (unknown)
First check

Observe LED brightness response to DAC adjustment

Test point
LED in Sensor Module Door
Expected
Light brightness changes proportionally to DAC value adjustment

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

pCO2 slope drop or time check failure

During last 2-point calibration, PCO2 sensor did not meet minimum specifications for properly performing sensor

Ranked causes

  1. 1. Sensor cartridge failure (common)
First check

Run 2-point calibration and monitor PCO2 slope and time response

Test point
pCO2 sensor during calibration
Expected
pCO2 slope and response time meet minimum specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotary valve positioning failure

Rotary valve unable to move to required position

Ranked causes

  1. 1. Rotary valve control failure (occasional)
First check

Power cycle analyzer to reset valve control

Test point
Analyzer power and rotary valve response
Expected
Valve moves to required position after power on

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sample contamination in cuvette

Blood debris detected in Sensor Card cuvette; CO-Ox parameters are unusable until cuvette is cleared

Ranked causes

  1. 1. Blood debris accumulated in Sensor Card cuvette (common)
First check

Perform Flush Flowpath procedure

Test point
Cuvette
Expected
Clear with no visible blood debris

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scanner communications failure

System bar code scanner did not respond or did not respond correctly when expected

Ranked causes

  1. 1. Scanner control electronics failure (occasional)
First check

Power cycle analyzer to reset scanner control

Test point
Analyzer power and scanner response
Expected
Scanner responds to bar codes after power on

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor card invalid part number

Consumable is programmed with part number not intended for use with analyzer

Ranked causes

  1. 1. Wrong sensor cartridge installed (occasional)
First check

Verify sensor card part number matches analyzer specifications

Test point
Sensor card part number
Expected
Part number matches analyzer documentation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor cartridge days expiration

Consumable has exceeded its use life expiration date

Ranked causes

  1. 1. Sensor cartridge use life exceeded (common)
First check

Verify cartridge days-in-use expiration

Test point
Cartridge installation date and use days
Expected
Days in use has not exceeded maximum allowed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor cartridge lot expiration

Consumable has exceeded its shelf life expiration date

Ranked causes

  1. 1. Expired sensor cartridge (common)
First check

Check sensor cartridge lot expiration date

Test point
Cartridge lot code and expiration date
Expected
Expiration date has not been reached

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor cartridge sample use expiration

Consumable has reached maximum number of allowable cycles (samples)

Ranked causes

  1. 1. Sample cycle limit reached (common)
First check

Check cartridge sample count versus maximum allowed

Test point
Cartridge sample counter
Expected
Sample count below maximum allowable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor voltage reading abnormal

Real-time sensor voltage readings appear out of specification during diagnostics

Ranked causes

  1. 1. Sensor cartridge malfunction or connector issue (unknown)
First check

Monitor sensor voltage output in real time

Test point
Sensor module connectors
Expected
Voltage within normal operating range for installed sensor cards

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
ABG Plate Heater Out Of ToleranceABG plate temperature fell outside tolerances during last sequenceABG plate heater temperature out of range during operationwarning
ABG Sensor Card Void WarrantyABG sensor cartridge warranty has been voided; warranty error codes no longer generatedCard removed for over 2 hours, reference cartridge over 150 days old, warranty sample/day limits reachedwarning
Air Threshold Out Of RangeAir detector mV reading did not fall within established air or fluid mV thresholdDuring last sequence, air detector reading falls outside threshold limitswarning
Auto QC Days ExpirationAuto QC use life expiredAuto QC cartridge use life expiration date exceededwarning
Auto QC Invalid Part NumberAuto QC cartridge has invalid part numberAuto QC cartridge detected with invalid part numbererror
Auto QC Lot ExpirationAuto QC lot shelf life expiredAuto QC cartridge lot expiration date exceededwarning
Auto QC Sample Use ExpirationAuto QC reached maximum sample cyclesAuto QC sample count reached maximum allowedwarning
BUN AcquisitionBlood urea nitrogen sensor reading could not be taken during calibration or analysisBUN sensor fails to produce readable output during calibration or sample analysiserror
BUN A to A DriftBUN sensor drift from previous analysisDuring analysis sequence, BUN sensor performance changed significantly from previous analysiserror
BUN CalculationBlood urea nitrogen sensor reading could not be calculated due to sensor or dependent-sensor errorBUN calculation fails during calibration or analysis sequenceerror
BUN/Creat Heater Out Of ToleranceBUN/Creatinine plate temperature fell outside tolerances during last sequenceBUN/Creat heater temperature out of range during operationwarning
BUN/Creat Sensor Card Days ExpirationBUN/Creatinine sensor use life expiredBUN/Creat sensor cartridge use life expiration exceededwarning
BUN/Creat Sensor Card Invalid Part No.BUN/Creatinine sensor card has invalid part numberBUN/Creat sensor card detected with invalid part numbererror
BUN/Creat Sensor Card Lot ExpirationBUN/Creatinine sensor lot shelf life expiredBUN/Creat sensor cartridge lot expiration date exceededwarning
BUN/Creat Sensor Card Sample Use ExpirationBUN/Creatinine sensor reached maximum sample cyclesBUN/Creat sensor sample count reached maximum allowedwarning
BUN E-zero DriftBUN sensor zero point drift from calibrationDuring analysis sequence, BUN sensor performance changed significantly from successful 2-point calibrationerror
BUN OverloadBlood urea nitrogen sensor reading exceeded software limits during calibration or analysisBUN sensor output exceeds maximum threshold during calibration or sample analysiserror
BUN Sensor Card Void WarrantyBUN sensor cartridge warranty has been voided; warranty error codes no longer generatedCard removed for over 2 hours, reference cartridge over 150 days old, warranty sample/day limits reachedwarning
BUN SlopeBlood urea nitrogen calibration standard difference did not meet minimum specificationsDuring 2-point calibration, BUN sensor response between standards below thresholderror
BUN Slope DriftBUN sensor slope drift from previous calibrationDuring last calibration, BUN sensor performance changed significantly from previous calibrationerror
BUN StabilityBlood urea nitrogen sensor reading did not reach stable endpoint during calibration or analysisBUN sensor reading continues to drift or fluctuate beyond stability thresholderror
Ca AcquisitionCalcium sensor reading could not be taken during calibration or analysisCa sensor fails to produce readable output during calibration or sample analysiserror
Ca A to A DriftCalcium sensor drift from previous analysisDuring analysis sequence, Ca sensor performance changed significantly from previous analysiserror
Ca CalculationCalcium sensor reading could not be calculated due to sensor or dependent-sensor errorCa calculation fails during calibration or analysis sequenceerror
Ca E-zero DriftCalcium sensor zero point drift from calibrationDuring analysis sequence, Ca sensor performance changed significantly from successful 2-point calibrationerror
Calibrator Days ExpirationCalibrator use life expiredCalibrator cartridge use life expiration date exceededwarning
Calibrator Invalid Part NumberCalibrator programmed with wrong part numberCalibrator cartridge detected with invalid part numbererror
Calibrator Lot ExpirationCalibrator shelf life expiredCalibrator cartridge lot expiration date exceededwarning
Calibrator Sample Use ExpirationCalibrator reached maximum sample cyclesCalibrator sample count reached maximum allowedwarning
Ca OverloadCalcium sensor reading exceeded software limits during calibration or analysisCa sensor output exceeds maximum threshold during calibration or sample analysiserror
Cartridge Read FailedAnalyzer unable to read data from consumable cartridgeRead operation from cartridge failserror
Cartridge Write FailedAnalyzer unable to write data to consumable cartridgeWrite operation to cartridge failserror
Ca SlopeCalcium calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Ca sensor response between standards below thresholderror
Ca Slope DriftCalcium sensor slope drift from previous calibrationDuring last calibration, Ca sensor performance changed significantly from previous calibrationerror
Ca StabilityCalcium sensor reading did not reach stable endpoint during calibration or analysisCa sensor reading continues to drift or fluctuate beyond stability thresholderror
Cl AcquisitionChloride sensor reading could not be taken during calibration or analysisCl sensor fails to produce readable output during calibration or sample analysiserror
Clamp Positioning FailedPeristaltic pump clamp assembly positioning failureClamp assembly failed to open or close correctlyerror
Cl A to A DriftChloride sensor drift from previous analysisDuring analysis sequence, Cl sensor performance changed significantly from previous analysiserror
Cl CalculationChloride sensor reading could not be calculated due to sensor or dependent-sensor errorCl calculation fails during calibration or analysis sequenceerror
Cl E-zero DriftChloride sensor zero point drift from calibrationDuring analysis sequence, Cl sensor performance changed significantly from successful 2-point calibrationerror
Cl OverloadChloride sensor reading exceeded software limits during calibration or analysisCl sensor output exceeds maximum threshold during calibration or sample analysiserror
Cl SlopeChloride calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Cl sensor response between standards below thresholderror
Cl Slope DriftChloride sensor slope drift from previous calibrationDuring last calibration, Cl sensor performance changed significantly from previous calibrationerror
Cl StabilityChloride sensor reading did not reach stable endpoint during calibration or analysisCl sensor reading continues to drift or fluctuate beyond stability thresholderror
Coil Heater Out Of ToleranceCoil temperature fell outside tolerances during last sequenceCoil heater temperature out of range during operationwarning
Contamination DetectedDebris detected in sensor card cuvetteDuring last light scan, debris was detected in sensor card cuvettewarning
COOX Calibration Drift With Existing Card/PackCO-Ox dye absorbance value out of rangeDuring last calibration, CO-Ox dye absorbance fell outside expected range or disagreed with previous averageerror
CO-Ox drift with existing card/packCO-Ox parameters indicate drift or drift in absorbance values with current sensor card or calibration pack installedOne or more conditional calibrations remain in CO-Ox Sequence Execution menuerror
Creat AcquisitionCreatinine sensor reading could not be taken during calibration or analysisCreat sensor fails to produce readable output during calibration or sample analysiserror
Creat A to A DriftCreatinine sensor drift from previous analysisDuring analysis sequence, Creat sensor performance changed significantly from previous analysiserror
Creat CalculationCreatinine sensor reading could not be calculated due to sensor or dependent-sensor errorCreat calculation fails during calibration or analysis sequenceerror
Creat E-zero DriftCreatinine sensor zero point drift from calibrationDuring analysis sequence, Creat sensor performance changed significantly from successful 2-point calibrationerror
Creat OverloadCreatinine sensor reading exceeded software limits during calibration or analysisCreat sensor output exceeds maximum threshold during calibration or sample analysiserror
Creat SlopeCreatinine calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Creat sensor response between standards below thresholderror
Creat Slope DriftCreatinine sensor slope drift from previous calibrationDuring last calibration, Creat sensor performance changed significantly from previous calibrationerror
Creat StabilityCreatinine sensor reading did not reach stable endpoint during calibration or analysisCreat sensor reading continues to drift or fluctuate beyond stability thresholderror
Delta mV CheckMinimum mV delta between flush solution and calibration standards not achievedDuring last sequence, calibration standard voltage delta below minimum thresholderror
Dye calibrationCO-Ox calibration failure due to absorbance value outside acceptable range or varianceAbsorbance value outside 0.5-1.5 mAbs or conditional calibration exceeds ±1.5% varianceerror
EEPROM Read FailedEEPROM on consumable cartridge could not be read by analyzerAnalyzer unable to read EEPROM data during cartridge validationerror
Failed to XCalibrate SpectrometerCO-Oximeter X-Axis Optical Calibration failedCO-Oximeter module unable to successfully initiate or complete X-Axis Optical Calibrationerror
Fluid Threshold Out Of RangeAir detector mV reading did not fall within established air or fluid mV thresholdDuring last sequence, air detector reading falls outside threshold limitswarning
Glu AcquisitionGlucose sensor reading could not be taken during calibration or analysisGlu sensor fails to produce readable output during calibration or sample analysiserror
Glu A to A DriftGlucose sensor drift from previous analysisDuring analysis sequence, Glu sensor performance changed significantly from previous analysiserror
Glu CalculationGlucose sensor reading could not be calculated due to sensor or dependent-sensor errorGlu calculation fails during calibration or analysis sequenceerror
Glu E-zero DriftGlucose sensor zero point drift from calibrationDuring analysis sequence, Glu sensor performance changed significantly from successful 2-point calibrationerror
Glu OverloadGlucose sensor reading exceeded software limits during calibration or analysisGlu sensor output exceeds maximum threshold during calibration or sample analysiserror
Glu SlopeGlucose calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Glu sensor response between standards below thresholderror
Glu Slope DriftGlucose sensor slope drift from previous calibrationDuring last calibration, Glu sensor performance changed significantly from previous calibrationerror
Glu StabilityGlucose sensor reading did not reach stable endpoint during calibration or analysisGlu sensor reading continues to drift or fluctuate beyond stability thresholderror
Hct AcquisitionHematocrit sensor reading could not be taken during calibration or analysisHct sensor fails to produce readable output during calibration or sample analysiserror
Hct A to A DriftHematocrit sensor drift from previous analysisDuring analysis sequence, Hct sensor performance changed significantly from previous analysiserror
Hct CalculationHematocrit sensor reading could not be calculated due to sensor or dependent-sensor errorHct calculation fails during calibration or analysis sequenceerror
Hct E-zero DriftHematocrit sensor zero point drift from calibrationDuring analysis sequence, Hct sensor performance changed significantly from successful 2-point calibrationerror
Hct OverloadHematocrit sensor reading exceeded software limits during calibration or analysisHct sensor output exceeds maximum threshold during calibration or sample analysiserror
Hct SlopeHematocrit calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Hct sensor response between standards below thresholderror
Hct Slope DriftHematocrit sensor slope drift from previous calibrationDuring last calibration, Hct sensor performance changed significantly from previous calibrationerror
Hct StabilityHematocrit sensor reading did not reach stable endpoint during calibration or analysisHct sensor reading continues to drift or fluctuate beyond stability thresholderror
I2C in ProgressI2C communication channel failureDuring last sequence, I2C channel unable to communicate with module or componenterror
Insufficient SampleLeading edge of aspirated sample was not detected by AD-Ref air detectorDuring last sequence, sample leading edge not detected by reference air detectorerror
Invalid Auto QC PackAuto QC cartridge was deemed invalid by softwareQC cartridge fails validation checkerror
Invalid BUN CardBUN sensor cartridge was deemed invalid by softwareBUN card fails validation checkerror
Invalid CalibratorCalibrator cartridge was deemed invalid by softwareCalibrator cartridge fails validation checkerror
Invalid ReferenceReference cartridge was deemed invalid by softwareReference cartridge fails validation checkerror
Invalid Rotary Valve PositionRotary valve in invalid positionRotary valve unable to move to required positionerror
Invalid Sensor CardABG sensor cartridge was deemed invalid by softwareSensor card fails validation checkerror
K AcquisitionPotassium sensor reading could not be taken during calibration or analysisK sensor fails to produce readable output during calibration or sample analysiserror
K A to A DriftPotassium sensor drift from previous analysisDuring analysis sequence, K sensor performance changed significantly from previous analysiserror
K CalculationPotassium sensor reading could not be calculated due to sensor or dependent-sensor errorK calculation fails during calibration or analysis sequenceerror
K E-zero DriftPotassium sensor zero point drift from calibrationDuring analysis sequence, K sensor performance changed significantly from successful 2-point calibrationerror
K OverloadPotassium sensor reading exceeded software limits during calibration or analysisK sensor output exceeds maximum threshold during calibration or sample analysiserror
K SlopePotassium calibration standard difference did not meet minimum specificationsDuring 2-point calibration, K sensor response between standards below thresholderror
K Slope DriftPotassium sensor slope drift from previous calibrationDuring last calibration, K sensor performance changed significantly from previous calibrationerror
K StabilityPotassium sensor reading did not reach stable endpoint during calibration or analysisK sensor reading continues to drift or fluctuate beyond stability thresholderror
Lac AcquisitionLactate sensor reading could not be taken during calibration or analysisLac sensor fails to produce readable output during calibration or sample analysiserror
Lac A to A DriftLactate sensor drift from previous analysisDuring analysis sequence, Lac sensor performance changed significantly from previous analysiserror
Lac CalculationLactate sensor reading could not be calculated due to sensor or dependent-sensor errorLac calculation fails during calibration or analysis sequenceerror
Lac E-zero DriftLactate sensor zero point drift from calibrationDuring analysis sequence, Lac sensor performance changed significantly from successful 2-point calibrationerror
Lac OverloadLactate sensor reading exceeded software limits during calibration or analysisLac sensor output exceeds maximum threshold during calibration or sample analysiserror
Lac SlopeLactate calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Lac sensor response between standards below thresholderror
Lac Slope DriftLactate sensor slope drift from previous calibrationDuring last calibration, Lac sensor performance changed significantly from previous calibrationerror
Lac StabilityLactate sensor reading did not reach stable endpoint during calibration or analysisLac sensor reading continues to drift or fluctuate beyond stability thresholderror
LED Heater Out Of ToleranceLED module temperature fell outside tolerances during last sequenceLED heater temperature out of range during operationwarning
Mg AcquisitionMagnesium sensor reading could not be taken during calibration or analysisMg sensor fails to produce readable output during calibration or sample analysiserror
Mg A to A DriftMagnesium sensor drift from previous analysisDuring analysis sequence, Mg sensor performance changed significantly from previous analysiserror
Mg CalculationMagnesium sensor reading could not be calculated due to sensor or dependent-sensor errorMg calculation fails during calibration or analysis sequenceerror
Mg E-zero DriftMagnesium sensor zero point drift from calibrationDuring analysis sequence, Mg sensor performance changed significantly from successful 2-point calibrationerror
Mg OverloadMagnesium sensor reading exceeded software limits during calibration or analysisMg sensor output exceeds maximum threshold during calibration or sample analysiserror
Mg SlopeMagnesium calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Mg sensor response between standards below thresholderror
Mg Slope DriftMagnesium sensor slope drift from previous calibrationDuring last calibration, Mg sensor performance changed significantly from previous calibrationerror
Mg StabilityMagnesium sensor reading did not reach stable endpoint during calibration or analysisMg sensor reading continues to drift or fluctuate beyond stability thresholderror
Na AcquisitionSodium sensor reading could not be taken during calibration or analysisNa sensor fails to produce readable output during calibration or sample analysiserror
Na A to A DriftSodium sensor drift from previous analysisDuring analysis sequence, Na sensor performance changed significantly from previous analysiserror
Na CalculationSodium sensor reading could not be calculated due to sensor or dependent-sensor errorNa calculation fails during calibration or analysis sequenceerror
Na E-zero DriftSodium sensor zero point drift from calibrationDuring analysis sequence, Na sensor performance changed significantly from successful 2-point calibrationerror
Na OverloadSodium sensor reading exceeded software limits during calibration or analysisNa sensor output exceeds maximum threshold during calibration or sample analysiserror
Na SlopeSodium calibration standard difference did not meet minimum specificationsDuring 2-point calibration, Na sensor response between standards below thresholderror
Na Slope DriftSodium sensor slope drift from previous calibrationDuring last calibration, Na sensor performance changed significantly from previous calibrationerror
Na StabilitySodium sensor reading did not reach stable endpoint during calibration or analysisNa sensor reading continues to drift or fluctuate beyond stability thresholderror
Nest Heater Out Of ToleranceNest temperature fell outside tolerances during last sequenceNest heater temperature out of range during operationwarning
No Air When RequiredAir detector did not detect air when expected during last sequenceAir not detected by air detector at expected timeerror
No Flush When RequiredAir detector did not detect flush solution when expected during last sequenceFlush solution not detected by air detector at expected timeerror
No Internal QC When RequiredAir detector did not detect QC solution when expected during internal QC analysisDuring last internal QC analysis, QC solution not detected by air detectorerror
No Sample For HctNo sample detected at air detector when hematocrit measurement expectedDuring last sequence, sample not detected for hematocrit analysiserror
No Sample When RequiredAir detector did not detect the sample when expected during analysis sequenceDuring last sequence, air detector failed to detect sample at expected timeerror
No Standard A When RequiredAir detector did not detect Standard A calibration fluid when expectedStandard A not detected by air detector during calibration sequenceerror
No Standard B When RequiredAir detector did not detect Standard B calibration fluid when expectedStandard B not detected by air detector during calibration sequenceerror
No Standard C When RequiredAir detector did not detect Standard C calibration fluid when expectedStandard C not detected by air detector during calibration sequenceerror
No Standard D When RequiredAir detector did not detect Standard D calibration fluid when expectedStandard D not detected by air detector during calibration sequenceerror
pCO2 AcquisitionpCO2 sensor reading could not be taken during calibration or analysispCO2 sensor fails to produce readable output during calibration or sample analysiserror
pCO2 A to A DriftpCO2 sensor drift from previous analysisDuring analysis sequence, pCO2 sensor performance changed significantly from previous analysiserror
pCO2 CalculationpCO2 sensor reading could not be calculated due to sensor or dependent-sensor errorpCO2 calculation fails during calibration or analysis sequenceerror
pCO2 E-zero DriftpCO2 sensor zero point drift from calibrationDuring analysis sequence, pCO2 sensor performance changed significantly from successful 2-point calibrationerror
pCO2 OverloadpCO2 sensor reading exceeded software limits during calibration or analysispCO2 sensor output exceeds maximum threshold during calibration or sample analysiserror
pCO2 SlopepCO2 calibration standard difference did not meet minimum specificationsDuring 2-point calibration, pCO2 sensor response between standards below thresholderror
pCO2 Slope DriftpCO2 sensor slope drift from previous calibrationDuring last calibration, pCO2 sensor performance changed significantly from previous calibrationerror
pCO2 Slope DropPCO2 sensor did not meet minimum specificationsDuring last 2-point calibration, pCO2 sensor did not meet minimum specificationserror
pCO2 Slope Time CheckPCO2 sensor response time failureDuring last 2-point calibration, pCO2 sensor did not meet minimum response specificationserror
pCO2 StabilitypCO2 sensor reading did not reach stable endpoint during calibration or analysispCO2 sensor reading continues to drift or fluctuate beyond stability thresholderror
pH AcquisitionpH sensor reading could not be taken during calibration or analysispH sensor fails to produce readable output during calibration or sample analysiserror
pH A to A DriftpH sensor drift from previous analysisDuring analysis sequence, pH sensor performance changed significantly from previous analysiserror
pH CalculationpH sensor reading could not be calculated due to sensor or dependent-sensor errorpH calculation fails during calibration or analysis sequenceerror
pH E-zero DriftpH sensor zero point drift from calibrationDuring analysis sequence, pH sensor performance changed significantly from successful 2-point calibrationerror
pH OverloadpH sensor reading exceeded software limits during calibration or analysispH sensor output exceeds maximum threshold during calibration or sample analysiserror
pH SlopepH calibration standard difference did not meet minimum specificationsDuring 2-point calibration, pH sensor response between standards below thresholderror
pH Slope DriftpH sensor slope drift from previous calibrationDuring last calibration, pH sensor performance changed significantly from previous calibrationerror
pH StabilitypH sensor reading did not reach stable endpoint during calibration or analysispH sensor reading continues to drift or fluctuate beyond stability thresholderror
pO2 AcquisitionpO2 sensor reading could not be taken during calibration or analysispO2 sensor fails to produce readable output during calibration or sample analysiserror
pO2 A to A DriftpO2 sensor drift from previous analysisDuring analysis sequence, pO2 sensor performance changed significantly from previous analysiserror
pO2 CalculationpO2 sensor reading could not be calculated due to sensor or dependent-sensor errorpO2 calculation fails during calibration or analysis sequenceerror
pO2 E-zero DriftpO2 sensor zero point drift from calibrationDuring analysis sequence, pO2 sensor performance changed significantly from successful 2-point calibrationerror
pO2 OverloadpO2 sensor reading exceeded software limits during calibration or analysispO2 sensor output exceeds maximum threshold during calibration or sample analysiserror
pO2 SlopepO2 calibration standard difference did not meet minimum specificationsDuring 2-point calibration, pO2 sensor response between standards below thresholderror
pO2 Slope DriftpO2 sensor slope drift from previous calibrationDuring last calibration, pO2 sensor performance changed significantly from previous calibrationerror
pO2 StabilitypO2 sensor reading did not reach stable endpoint during calibration or analysispO2 sensor reading continues to drift or fluctuate beyond stability thresholderror
Preheater Out Of TolerancePreheater temperature fell outside tolerances during last sequencePreheater temperature out of range during operationwarning
Printer Door OpenThermal printer door not fully closedPrinter door sensor indicates printer door is openwarning
Printer Out Of PaperThermal printer paper depleted or not detectedPrinter paper sensor does not detect available paper rollwarning
Probe Failed To HomeSample probe unable to reach home positionSampler unable to move sample probe to required positionerror
Reference Days ExpirationReference cartridge use life expiredReference cartridge use life expiration exceededwarning
Reference Invalid Part NumberReference cartridge has invalid part numberReference cartridge detected with invalid part numbererror
Reference Lot ExpirationReference cartridge shelf life expiredReference cartridge lot expiration date exceededwarning
Reference Sample Use ExpirationReference cartridge reached maximum sample cyclesReference cartridge sample count reached maximum allowedwarning
RFID Tag Not FoundRFID tag for one or more consumables was not found by readerConsumable cartridge installed without readable RFID tagerror
Rotary Valve Failed To HomeRotary valve unable to reach home positionRotary valve unable to move to required positionerror
Sample ContaminationBlood debris detected in Sensor Card cuvette; CO-Ox parameters are unreliableAnalyzer detects blood debris inside cuvette during sample analysiserror
Sample PositionNo sample detected at one or more air detectors when expected during sequenceDuring last sequence, sample position not confirmed by air detectorserror
Sampler Failed Reaching Top SensorSampler unable to reach top sensor positionSampler unable to move to top sensor positionerror
Sampler Homing FailedSampler homing sequence failedSampler unable to complete homing sequenceerror
Sampler Positioning FailedSampler unable to reach required positionSampler unable to move to required positionerror
Scanner CommunicationsBarcode scanner communication failureSystem bar code scanner did not respond or respond correctlyerror
Scanner TimeoutBarcode scanner timeoutSystem bar code scanner did not respond when expectederror
Sensor Card Days ExpirationSensor card use life expiredSensor cartridge use life expiration date exceededwarning
Sensor Card Invalid Part NumberSensor card programmed with wrong part numberSensor cartridge detected with invalid part numbererror
Sensor Card Lot ExpirationSensor card shelf life expiredSensor cartridge lot expiration date exceededwarning
Sensor Card Sample Use ExpirationSensor card reached maximum sample cyclesSensor cartridge sample count reached maximum allowedwarning
Short SampleNo sample detected at one or more air detectors when expectedDuring last sequence, sample not detected at one or more air detector positionserror
Spectro Heater Out Of ToleranceSpectrometer temperature fell outside tolerances during last sequenceSpectrometer heater temperature out of range during operationwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI