Nova Biomedical Stat Profile Prime Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 30 documented symptoms for the Nova Biomedical Stat Profile Prime Plus, with 36 ranked causes, 35 diagnostic checks with test points and expected values, and 177 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Dye calibration absorbance value outside valid range
CO-Ox calibration fails because measured absorbance of calibration dye (Standard B) at 563 nm does not fall within 0.5 to 1.5 mAbs
Ranked causes
- 1. Air bubbles in the cuvette (common)
- 2. Air leaks in the flowpath (common)
- 3. Contamination on inside surface of cuvette gradually dissolving (occasional)
- 4. Change in sensor card and/or calibration pack (occasional)
Verify dye absorbance value is within acceptable range
- Test point
- Calibration dye absorbance at 563 nm
- Expected
- 0.5 to 1.5 mAbs, 0.5 to 1.5
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer door open
Analyzer's thermal printer door sensor indicates printer door is open
Ranked causes
- 1. Printer door not closed (common)
- 2. Door sensor malfunction (occasional)
Inspect printer door and close securely
- Test point
- Printer door
- Expected
- Door fully closed and latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer out of paper
Analyzer's thermal printer paper sensor does not detect available roll of paper
Ranked causes
- 1. Printer paper depleted (common)
- 2. Paper sensor malfunction (occasional)
Verify printer paper is installed and available
- Test point
- Printer paper slot
- Expected
- Paper roll loaded and sensor detects it
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Probe failed to home
Sampler unable to move sample probe to required position
Ranked causes
- 1. Sampler obstruction or mechanical jam (occasional)
- 2. Sampler control electronics failure (occasional)
Visually inspect sampler path for obstructions
- Test point
- Sampler probe and housing
- Expected
- No obstructions visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detector voltage reading abnormal
Real-time air detector voltage readings do not fall within expected range or status indicates malfunction
Ranked causes
- 1. Air detector sensor malfunction or calibration drift (unknown)
Monitor air detector voltage in real time
- Test point
- AD-50, AD-130, AD-145, AD-REF, AD-Hemo
- Expected
- Voltage within normal operating range and status indicates functional
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analyte A-to-A drift
During last analysis sequence, indicated analyte's performance changed significantly from previous analysis
Ranked causes
- 1. Sensor cartridge degradation (common)
Compare current analysis result to previous analysis for same analyte
- Test point
- Analyte result from consecutive samples
- Expected
- Result variance within normal limits
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analyte slope drift
During last calibration, indicated analyte's performance changed significantly from previous calibration
Ranked causes
- 1. Sensor cartridge degradation (common)
Monitor sensor slope response during calibration
- Test point
- Sensor slope during 2-point calibration
- Expected
- Slope within acceptable range, consistent with previous calibrations
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analyte zero drift after calibration
During last analysis sequence, indicated analyte's performance changed significantly from successful 2-point calibration
Ranked causes
- 1. Sensor cartridge degradation or contamination (common)
Run calibration sequence and monitor analyte drift
- Test point
- Analyte response during 2-point calibration
- Expected
- Analyte performance within acceptable range after successful calibration
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration lamp not producing signal to spectrometer
Calibration lamp test shows no peaks or incorrect peak position on spectrometer graph
Ranked causes
- 1. Calibration lamp light source malfunction (unknown)
Read calibration lamp pixel output to spectrometer
- Test point
- Spectrometer graph display
- Expected
- 2 tall peaks on right side of graph, rightmost peak within ±10 pixels of 800th pixel, ±10 pixels
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibrator days expiration
Indicated cartridge has exceeded its use life expiration date
Ranked causes
- 1. Calibrator use life exceeded (common)
Verify calibrator days-in-use expiration
- Test point
- Calibrator installation date and use days
- Expected
- Days in use has not exceeded maximum allowed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibrator invalid part number
Indicated cartridge programmed with part number not intended for use with analyzer
Ranked causes
- 1. Wrong calibrator installed (occasional)
Verify calibrator cartridge part number
- Test point
- Calibrator cartridge part number
- Expected
- Part number matches analyzer specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibrator lot expiration
Indicated cartridge has exceeded its shelf life expiration date
Ranked causes
- 1. Expired calibrator (common)
Check calibrator lot expiration date
- Test point
- Calibrator lot code and expiration date
- Expected
- Expiration date has not been reached
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibrator sample use expiration
Indicated cartridge has exceeded maximum number of cycles (samples)
Ranked causes
- 1. Sample cycle limit reached (common)
Check calibrator sample count versus maximum
- Test point
- Calibrator sample counter
- Expected
- Sample count below maximum allowable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clamp positioning failure
Peristaltic pump clamp assembly failed to open or close correctly
Ranked causes
- 1. Clamp actuator control failure (occasional)
Power cycle analyzer to reset clamp control
- Test point
- Analyzer power and clamp response
- Expected
- Clamp opens and closes properly after power on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Component temperature out of specification during diagnostics
Real-time temperature readings of ABG plate, nest, coil, preheater, spectrometer, BUN/Creat plate, or LED heater do not match expected 37°C
Ranked causes
- 1. Heater malfunction or thermal sensor drift (unknown)
Measure real-time component temperature
- Test point
- ABG Plate, Nest, Coil, Preheater, Spectrometer, BUN/Creat Plate
- Expected
- 37°C for all components (Ambient temperature per specifications)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Conditional calibration flagged
Dye absorbance value is within 0.5-1.5 mAbs range but differs from 5-value average by more than tolerance, marked with C flag
Ranked causes
- 1. Absorbance variation exceeds ±1.5% from 5-value average (common)
Review CO-Ox Sequence Execution menu for conditional calibrations
- Test point
- CO-Ox Sequence Execution menu
- Expected
- List showing valid (V) and conditional (C) calibrations
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Contamination detected in sensor card cuvette caution
During last light scan, debris was detected in sensor card cuvette
Ranked causes
- 1. Cuvette contamination or debris (occasional)
Perform Flush procedure
- Test point
- Flowpath during flush
- Expected
- Debris cleared from cuvette
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
COOX Calibration drift with existing card/pack
During last calibration, CO-Ox dye absorbance value fell outside expected range or did not agree within certain percentage of average from earlier calibrations
Ranked causes
- 1. Sensor card or calibrator pack degradation (occasional)
Review dye calibration validity conditions per section 5.4.2
- Test point
- Dye absorbance value during calibration
- Expected
- Absorbance within expected range and consistent with previous calibrations
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failed to X-Calibrate Spectrometer
CO-Oximeter module unable to initiate or complete X-Axis Optical Calibration
Ranked causes
- 1. Spectrometer optical calibration failure (occasional)
Run manual X Calibration from CO-Ox menu
- Test point
- X Calibration sequence in CO-Ox Sequence Execution Menu
- Expected
- X Calibration completes successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
I2C communication failure caution
During last sequence, inter-integrated circuit (I2C) channel unable to communicate with module or component
Ranked causes
- 1. Fluid contamination on EEPROM or sensor card microchip pins (occasional)
Visually inspect SC Nest, Preheater, and sensor card connectors for fluid
- Test point
- EEPROM pins and sensor card microchip contacts
- Expected
- No fluid or corrosion on pins
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED light not responding to brightness adjustment
LED in sensor module does not change brightness when DAC value is adjusted during diagnostics
Ranked causes
- 1. LED illumination optics malfunction (unknown)
Observe LED brightness response to DAC adjustment
- Test point
- LED in Sensor Module Door
- Expected
- Light brightness changes proportionally to DAC value adjustment
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
pCO2 slope drop or time check failure
During last 2-point calibration, PCO2 sensor did not meet minimum specifications for properly performing sensor
Ranked causes
- 1. Sensor cartridge failure (common)
Run 2-point calibration and monitor PCO2 slope and time response
- Test point
- pCO2 sensor during calibration
- Expected
- pCO2 slope and response time meet minimum specifications
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotary valve positioning failure
Rotary valve unable to move to required position
Ranked causes
- 1. Rotary valve control failure (occasional)
Power cycle analyzer to reset valve control
- Test point
- Analyzer power and rotary valve response
- Expected
- Valve moves to required position after power on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sample contamination in cuvette
Blood debris detected in Sensor Card cuvette; CO-Ox parameters are unusable until cuvette is cleared
Ranked causes
- 1. Blood debris accumulated in Sensor Card cuvette (common)
Perform Flush Flowpath procedure
- Test point
- Cuvette
- Expected
- Clear with no visible blood debris
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scanner communications failure
System bar code scanner did not respond or did not respond correctly when expected
Ranked causes
- 1. Scanner control electronics failure (occasional)
Power cycle analyzer to reset scanner control
- Test point
- Analyzer power and scanner response
- Expected
- Scanner responds to bar codes after power on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor card invalid part number
Consumable is programmed with part number not intended for use with analyzer
Ranked causes
- 1. Wrong sensor cartridge installed (occasional)
Verify sensor card part number matches analyzer specifications
- Test point
- Sensor card part number
- Expected
- Part number matches analyzer documentation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cartridge days expiration
Consumable has exceeded its use life expiration date
Ranked causes
- 1. Sensor cartridge use life exceeded (common)
Verify cartridge days-in-use expiration
- Test point
- Cartridge installation date and use days
- Expected
- Days in use has not exceeded maximum allowed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cartridge lot expiration
Consumable has exceeded its shelf life expiration date
Ranked causes
- 1. Expired sensor cartridge (common)
Check sensor cartridge lot expiration date
- Test point
- Cartridge lot code and expiration date
- Expected
- Expiration date has not been reached
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor cartridge sample use expiration
Consumable has reached maximum number of allowable cycles (samples)
Ranked causes
- 1. Sample cycle limit reached (common)
Check cartridge sample count versus maximum allowed
- Test point
- Cartridge sample counter
- Expected
- Sample count below maximum allowable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor voltage reading abnormal
Real-time sensor voltage readings appear out of specification during diagnostics
Ranked causes
- 1. Sensor cartridge malfunction or connector issue (unknown)
Monitor sensor voltage output in real time
- Test point
- Sensor module connectors
- Expected
- Voltage within normal operating range for installed sensor cards
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ABG Plate Heater Out Of Tolerance | ABG plate temperature fell outside tolerances during last sequence | ABG plate heater temperature out of range during operation | warning |
ABG Sensor Card Void Warranty | ABG sensor cartridge warranty has been voided; warranty error codes no longer generated | Card removed for over 2 hours, reference cartridge over 150 days old, warranty sample/day limits reached | warning |
Air Threshold Out Of Range | Air detector mV reading did not fall within established air or fluid mV threshold | During last sequence, air detector reading falls outside threshold limits | warning |
Auto QC Days Expiration | Auto QC use life expired | Auto QC cartridge use life expiration date exceeded | warning |
Auto QC Invalid Part Number | Auto QC cartridge has invalid part number | Auto QC cartridge detected with invalid part number | error |
Auto QC Lot Expiration | Auto QC lot shelf life expired | Auto QC cartridge lot expiration date exceeded | warning |
Auto QC Sample Use Expiration | Auto QC reached maximum sample cycles | Auto QC sample count reached maximum allowed | warning |
BUN Acquisition | Blood urea nitrogen sensor reading could not be taken during calibration or analysis | BUN sensor fails to produce readable output during calibration or sample analysis | error |
BUN A to A Drift | BUN sensor drift from previous analysis | During analysis sequence, BUN sensor performance changed significantly from previous analysis | error |
BUN Calculation | Blood urea nitrogen sensor reading could not be calculated due to sensor or dependent-sensor error | BUN calculation fails during calibration or analysis sequence | error |
BUN/Creat Heater Out Of Tolerance | BUN/Creatinine plate temperature fell outside tolerances during last sequence | BUN/Creat heater temperature out of range during operation | warning |
BUN/Creat Sensor Card Days Expiration | BUN/Creatinine sensor use life expired | BUN/Creat sensor cartridge use life expiration exceeded | warning |
BUN/Creat Sensor Card Invalid Part No. | BUN/Creatinine sensor card has invalid part number | BUN/Creat sensor card detected with invalid part number | error |
BUN/Creat Sensor Card Lot Expiration | BUN/Creatinine sensor lot shelf life expired | BUN/Creat sensor cartridge lot expiration date exceeded | warning |
BUN/Creat Sensor Card Sample Use Expiration | BUN/Creatinine sensor reached maximum sample cycles | BUN/Creat sensor sample count reached maximum allowed | warning |
BUN E-zero Drift | BUN sensor zero point drift from calibration | During analysis sequence, BUN sensor performance changed significantly from successful 2-point calibration | error |
BUN Overload | Blood urea nitrogen sensor reading exceeded software limits during calibration or analysis | BUN sensor output exceeds maximum threshold during calibration or sample analysis | error |
BUN Sensor Card Void Warranty | BUN sensor cartridge warranty has been voided; warranty error codes no longer generated | Card removed for over 2 hours, reference cartridge over 150 days old, warranty sample/day limits reached | warning |
BUN Slope | Blood urea nitrogen calibration standard difference did not meet minimum specifications | During 2-point calibration, BUN sensor response between standards below threshold | error |
BUN Slope Drift | BUN sensor slope drift from previous calibration | During last calibration, BUN sensor performance changed significantly from previous calibration | error |
BUN Stability | Blood urea nitrogen sensor reading did not reach stable endpoint during calibration or analysis | BUN sensor reading continues to drift or fluctuate beyond stability threshold | error |
Ca Acquisition | Calcium sensor reading could not be taken during calibration or analysis | Ca sensor fails to produce readable output during calibration or sample analysis | error |
Ca A to A Drift | Calcium sensor drift from previous analysis | During analysis sequence, Ca sensor performance changed significantly from previous analysis | error |
Ca Calculation | Calcium sensor reading could not be calculated due to sensor or dependent-sensor error | Ca calculation fails during calibration or analysis sequence | error |
Ca E-zero Drift | Calcium sensor zero point drift from calibration | During analysis sequence, Ca sensor performance changed significantly from successful 2-point calibration | error |
Calibrator Days Expiration | Calibrator use life expired | Calibrator cartridge use life expiration date exceeded | warning |
Calibrator Invalid Part Number | Calibrator programmed with wrong part number | Calibrator cartridge detected with invalid part number | error |
Calibrator Lot Expiration | Calibrator shelf life expired | Calibrator cartridge lot expiration date exceeded | warning |
Calibrator Sample Use Expiration | Calibrator reached maximum sample cycles | Calibrator sample count reached maximum allowed | warning |
Ca Overload | Calcium sensor reading exceeded software limits during calibration or analysis | Ca sensor output exceeds maximum threshold during calibration or sample analysis | error |
Cartridge Read Failed | Analyzer unable to read data from consumable cartridge | Read operation from cartridge fails | error |
Cartridge Write Failed | Analyzer unable to write data to consumable cartridge | Write operation to cartridge fails | error |
Ca Slope | Calcium calibration standard difference did not meet minimum specifications | During 2-point calibration, Ca sensor response between standards below threshold | error |
Ca Slope Drift | Calcium sensor slope drift from previous calibration | During last calibration, Ca sensor performance changed significantly from previous calibration | error |
Ca Stability | Calcium sensor reading did not reach stable endpoint during calibration or analysis | Ca sensor reading continues to drift or fluctuate beyond stability threshold | error |
Cl Acquisition | Chloride sensor reading could not be taken during calibration or analysis | Cl sensor fails to produce readable output during calibration or sample analysis | error |
Clamp Positioning Failed | Peristaltic pump clamp assembly positioning failure | Clamp assembly failed to open or close correctly | error |
Cl A to A Drift | Chloride sensor drift from previous analysis | During analysis sequence, Cl sensor performance changed significantly from previous analysis | error |
Cl Calculation | Chloride sensor reading could not be calculated due to sensor or dependent-sensor error | Cl calculation fails during calibration or analysis sequence | error |
Cl E-zero Drift | Chloride sensor zero point drift from calibration | During analysis sequence, Cl sensor performance changed significantly from successful 2-point calibration | error |
Cl Overload | Chloride sensor reading exceeded software limits during calibration or analysis | Cl sensor output exceeds maximum threshold during calibration or sample analysis | error |
Cl Slope | Chloride calibration standard difference did not meet minimum specifications | During 2-point calibration, Cl sensor response between standards below threshold | error |
Cl Slope Drift | Chloride sensor slope drift from previous calibration | During last calibration, Cl sensor performance changed significantly from previous calibration | error |
Cl Stability | Chloride sensor reading did not reach stable endpoint during calibration or analysis | Cl sensor reading continues to drift or fluctuate beyond stability threshold | error |
Coil Heater Out Of Tolerance | Coil temperature fell outside tolerances during last sequence | Coil heater temperature out of range during operation | warning |
Contamination Detected | Debris detected in sensor card cuvette | During last light scan, debris was detected in sensor card cuvette | warning |
COOX Calibration Drift With Existing Card/Pack | CO-Ox dye absorbance value out of range | During last calibration, CO-Ox dye absorbance fell outside expected range or disagreed with previous average | error |
CO-Ox drift with existing card/pack | CO-Ox parameters indicate drift or drift in absorbance values with current sensor card or calibration pack installed | One or more conditional calibrations remain in CO-Ox Sequence Execution menu | error |
Creat Acquisition | Creatinine sensor reading could not be taken during calibration or analysis | Creat sensor fails to produce readable output during calibration or sample analysis | error |
Creat A to A Drift | Creatinine sensor drift from previous analysis | During analysis sequence, Creat sensor performance changed significantly from previous analysis | error |
Creat Calculation | Creatinine sensor reading could not be calculated due to sensor or dependent-sensor error | Creat calculation fails during calibration or analysis sequence | error |
Creat E-zero Drift | Creatinine sensor zero point drift from calibration | During analysis sequence, Creat sensor performance changed significantly from successful 2-point calibration | error |
Creat Overload | Creatinine sensor reading exceeded software limits during calibration or analysis | Creat sensor output exceeds maximum threshold during calibration or sample analysis | error |
Creat Slope | Creatinine calibration standard difference did not meet minimum specifications | During 2-point calibration, Creat sensor response between standards below threshold | error |
Creat Slope Drift | Creatinine sensor slope drift from previous calibration | During last calibration, Creat sensor performance changed significantly from previous calibration | error |
Creat Stability | Creatinine sensor reading did not reach stable endpoint during calibration or analysis | Creat sensor reading continues to drift or fluctuate beyond stability threshold | error |
Delta mV Check | Minimum mV delta between flush solution and calibration standards not achieved | During last sequence, calibration standard voltage delta below minimum threshold | error |
Dye calibration | CO-Ox calibration failure due to absorbance value outside acceptable range or variance | Absorbance value outside 0.5-1.5 mAbs or conditional calibration exceeds ±1.5% variance | error |
EEPROM Read Failed | EEPROM on consumable cartridge could not be read by analyzer | Analyzer unable to read EEPROM data during cartridge validation | error |
Failed to XCalibrate Spectrometer | CO-Oximeter X-Axis Optical Calibration failed | CO-Oximeter module unable to successfully initiate or complete X-Axis Optical Calibration | error |
Fluid Threshold Out Of Range | Air detector mV reading did not fall within established air or fluid mV threshold | During last sequence, air detector reading falls outside threshold limits | warning |
Glu Acquisition | Glucose sensor reading could not be taken during calibration or analysis | Glu sensor fails to produce readable output during calibration or sample analysis | error |
Glu A to A Drift | Glucose sensor drift from previous analysis | During analysis sequence, Glu sensor performance changed significantly from previous analysis | error |
Glu Calculation | Glucose sensor reading could not be calculated due to sensor or dependent-sensor error | Glu calculation fails during calibration or analysis sequence | error |
Glu E-zero Drift | Glucose sensor zero point drift from calibration | During analysis sequence, Glu sensor performance changed significantly from successful 2-point calibration | error |
Glu Overload | Glucose sensor reading exceeded software limits during calibration or analysis | Glu sensor output exceeds maximum threshold during calibration or sample analysis | error |
Glu Slope | Glucose calibration standard difference did not meet minimum specifications | During 2-point calibration, Glu sensor response between standards below threshold | error |
Glu Slope Drift | Glucose sensor slope drift from previous calibration | During last calibration, Glu sensor performance changed significantly from previous calibration | error |
Glu Stability | Glucose sensor reading did not reach stable endpoint during calibration or analysis | Glu sensor reading continues to drift or fluctuate beyond stability threshold | error |
Hct Acquisition | Hematocrit sensor reading could not be taken during calibration or analysis | Hct sensor fails to produce readable output during calibration or sample analysis | error |
Hct A to A Drift | Hematocrit sensor drift from previous analysis | During analysis sequence, Hct sensor performance changed significantly from previous analysis | error |
Hct Calculation | Hematocrit sensor reading could not be calculated due to sensor or dependent-sensor error | Hct calculation fails during calibration or analysis sequence | error |
Hct E-zero Drift | Hematocrit sensor zero point drift from calibration | During analysis sequence, Hct sensor performance changed significantly from successful 2-point calibration | error |
Hct Overload | Hematocrit sensor reading exceeded software limits during calibration or analysis | Hct sensor output exceeds maximum threshold during calibration or sample analysis | error |
Hct Slope | Hematocrit calibration standard difference did not meet minimum specifications | During 2-point calibration, Hct sensor response between standards below threshold | error |
Hct Slope Drift | Hematocrit sensor slope drift from previous calibration | During last calibration, Hct sensor performance changed significantly from previous calibration | error |
Hct Stability | Hematocrit sensor reading did not reach stable endpoint during calibration or analysis | Hct sensor reading continues to drift or fluctuate beyond stability threshold | error |
I2C in Progress | I2C communication channel failure | During last sequence, I2C channel unable to communicate with module or component | error |
Insufficient Sample | Leading edge of aspirated sample was not detected by AD-Ref air detector | During last sequence, sample leading edge not detected by reference air detector | error |
Invalid Auto QC Pack | Auto QC cartridge was deemed invalid by software | QC cartridge fails validation check | error |
Invalid BUN Card | BUN sensor cartridge was deemed invalid by software | BUN card fails validation check | error |
Invalid Calibrator | Calibrator cartridge was deemed invalid by software | Calibrator cartridge fails validation check | error |
Invalid Reference | Reference cartridge was deemed invalid by software | Reference cartridge fails validation check | error |
Invalid Rotary Valve Position | Rotary valve in invalid position | Rotary valve unable to move to required position | error |
Invalid Sensor Card | ABG sensor cartridge was deemed invalid by software | Sensor card fails validation check | error |
K Acquisition | Potassium sensor reading could not be taken during calibration or analysis | K sensor fails to produce readable output during calibration or sample analysis | error |
K A to A Drift | Potassium sensor drift from previous analysis | During analysis sequence, K sensor performance changed significantly from previous analysis | error |
K Calculation | Potassium sensor reading could not be calculated due to sensor or dependent-sensor error | K calculation fails during calibration or analysis sequence | error |
K E-zero Drift | Potassium sensor zero point drift from calibration | During analysis sequence, K sensor performance changed significantly from successful 2-point calibration | error |
K Overload | Potassium sensor reading exceeded software limits during calibration or analysis | K sensor output exceeds maximum threshold during calibration or sample analysis | error |
K Slope | Potassium calibration standard difference did not meet minimum specifications | During 2-point calibration, K sensor response between standards below threshold | error |
K Slope Drift | Potassium sensor slope drift from previous calibration | During last calibration, K sensor performance changed significantly from previous calibration | error |
K Stability | Potassium sensor reading did not reach stable endpoint during calibration or analysis | K sensor reading continues to drift or fluctuate beyond stability threshold | error |
Lac Acquisition | Lactate sensor reading could not be taken during calibration or analysis | Lac sensor fails to produce readable output during calibration or sample analysis | error |
Lac A to A Drift | Lactate sensor drift from previous analysis | During analysis sequence, Lac sensor performance changed significantly from previous analysis | error |
Lac Calculation | Lactate sensor reading could not be calculated due to sensor or dependent-sensor error | Lac calculation fails during calibration or analysis sequence | error |
Lac E-zero Drift | Lactate sensor zero point drift from calibration | During analysis sequence, Lac sensor performance changed significantly from successful 2-point calibration | error |
Lac Overload | Lactate sensor reading exceeded software limits during calibration or analysis | Lac sensor output exceeds maximum threshold during calibration or sample analysis | error |
Lac Slope | Lactate calibration standard difference did not meet minimum specifications | During 2-point calibration, Lac sensor response between standards below threshold | error |
Lac Slope Drift | Lactate sensor slope drift from previous calibration | During last calibration, Lac sensor performance changed significantly from previous calibration | error |
Lac Stability | Lactate sensor reading did not reach stable endpoint during calibration or analysis | Lac sensor reading continues to drift or fluctuate beyond stability threshold | error |
LED Heater Out Of Tolerance | LED module temperature fell outside tolerances during last sequence | LED heater temperature out of range during operation | warning |
Mg Acquisition | Magnesium sensor reading could not be taken during calibration or analysis | Mg sensor fails to produce readable output during calibration or sample analysis | error |
Mg A to A Drift | Magnesium sensor drift from previous analysis | During analysis sequence, Mg sensor performance changed significantly from previous analysis | error |
Mg Calculation | Magnesium sensor reading could not be calculated due to sensor or dependent-sensor error | Mg calculation fails during calibration or analysis sequence | error |
Mg E-zero Drift | Magnesium sensor zero point drift from calibration | During analysis sequence, Mg sensor performance changed significantly from successful 2-point calibration | error |
Mg Overload | Magnesium sensor reading exceeded software limits during calibration or analysis | Mg sensor output exceeds maximum threshold during calibration or sample analysis | error |
Mg Slope | Magnesium calibration standard difference did not meet minimum specifications | During 2-point calibration, Mg sensor response between standards below threshold | error |
Mg Slope Drift | Magnesium sensor slope drift from previous calibration | During last calibration, Mg sensor performance changed significantly from previous calibration | error |
Mg Stability | Magnesium sensor reading did not reach stable endpoint during calibration or analysis | Mg sensor reading continues to drift or fluctuate beyond stability threshold | error |
Na Acquisition | Sodium sensor reading could not be taken during calibration or analysis | Na sensor fails to produce readable output during calibration or sample analysis | error |
Na A to A Drift | Sodium sensor drift from previous analysis | During analysis sequence, Na sensor performance changed significantly from previous analysis | error |
Na Calculation | Sodium sensor reading could not be calculated due to sensor or dependent-sensor error | Na calculation fails during calibration or analysis sequence | error |
Na E-zero Drift | Sodium sensor zero point drift from calibration | During analysis sequence, Na sensor performance changed significantly from successful 2-point calibration | error |
Na Overload | Sodium sensor reading exceeded software limits during calibration or analysis | Na sensor output exceeds maximum threshold during calibration or sample analysis | error |
Na Slope | Sodium calibration standard difference did not meet minimum specifications | During 2-point calibration, Na sensor response between standards below threshold | error |
Na Slope Drift | Sodium sensor slope drift from previous calibration | During last calibration, Na sensor performance changed significantly from previous calibration | error |
Na Stability | Sodium sensor reading did not reach stable endpoint during calibration or analysis | Na sensor reading continues to drift or fluctuate beyond stability threshold | error |
Nest Heater Out Of Tolerance | Nest temperature fell outside tolerances during last sequence | Nest heater temperature out of range during operation | warning |
No Air When Required | Air detector did not detect air when expected during last sequence | Air not detected by air detector at expected time | error |
No Flush When Required | Air detector did not detect flush solution when expected during last sequence | Flush solution not detected by air detector at expected time | error |
No Internal QC When Required | Air detector did not detect QC solution when expected during internal QC analysis | During last internal QC analysis, QC solution not detected by air detector | error |
No Sample For Hct | No sample detected at air detector when hematocrit measurement expected | During last sequence, sample not detected for hematocrit analysis | error |
No Sample When Required | Air detector did not detect the sample when expected during analysis sequence | During last sequence, air detector failed to detect sample at expected time | error |
No Standard A When Required | Air detector did not detect Standard A calibration fluid when expected | Standard A not detected by air detector during calibration sequence | error |
No Standard B When Required | Air detector did not detect Standard B calibration fluid when expected | Standard B not detected by air detector during calibration sequence | error |
No Standard C When Required | Air detector did not detect Standard C calibration fluid when expected | Standard C not detected by air detector during calibration sequence | error |
No Standard D When Required | Air detector did not detect Standard D calibration fluid when expected | Standard D not detected by air detector during calibration sequence | error |
pCO2 Acquisition | pCO2 sensor reading could not be taken during calibration or analysis | pCO2 sensor fails to produce readable output during calibration or sample analysis | error |
pCO2 A to A Drift | pCO2 sensor drift from previous analysis | During analysis sequence, pCO2 sensor performance changed significantly from previous analysis | error |
pCO2 Calculation | pCO2 sensor reading could not be calculated due to sensor or dependent-sensor error | pCO2 calculation fails during calibration or analysis sequence | error |
pCO2 E-zero Drift | pCO2 sensor zero point drift from calibration | During analysis sequence, pCO2 sensor performance changed significantly from successful 2-point calibration | error |
pCO2 Overload | pCO2 sensor reading exceeded software limits during calibration or analysis | pCO2 sensor output exceeds maximum threshold during calibration or sample analysis | error |
pCO2 Slope | pCO2 calibration standard difference did not meet minimum specifications | During 2-point calibration, pCO2 sensor response between standards below threshold | error |
pCO2 Slope Drift | pCO2 sensor slope drift from previous calibration | During last calibration, pCO2 sensor performance changed significantly from previous calibration | error |
pCO2 Slope Drop | PCO2 sensor did not meet minimum specifications | During last 2-point calibration, pCO2 sensor did not meet minimum specifications | error |
pCO2 Slope Time Check | PCO2 sensor response time failure | During last 2-point calibration, pCO2 sensor did not meet minimum response specifications | error |
pCO2 Stability | pCO2 sensor reading did not reach stable endpoint during calibration or analysis | pCO2 sensor reading continues to drift or fluctuate beyond stability threshold | error |
pH Acquisition | pH sensor reading could not be taken during calibration or analysis | pH sensor fails to produce readable output during calibration or sample analysis | error |
pH A to A Drift | pH sensor drift from previous analysis | During analysis sequence, pH sensor performance changed significantly from previous analysis | error |
pH Calculation | pH sensor reading could not be calculated due to sensor or dependent-sensor error | pH calculation fails during calibration or analysis sequence | error |
pH E-zero Drift | pH sensor zero point drift from calibration | During analysis sequence, pH sensor performance changed significantly from successful 2-point calibration | error |
pH Overload | pH sensor reading exceeded software limits during calibration or analysis | pH sensor output exceeds maximum threshold during calibration or sample analysis | error |
pH Slope | pH calibration standard difference did not meet minimum specifications | During 2-point calibration, pH sensor response between standards below threshold | error |
pH Slope Drift | pH sensor slope drift from previous calibration | During last calibration, pH sensor performance changed significantly from previous calibration | error |
pH Stability | pH sensor reading did not reach stable endpoint during calibration or analysis | pH sensor reading continues to drift or fluctuate beyond stability threshold | error |
pO2 Acquisition | pO2 sensor reading could not be taken during calibration or analysis | pO2 sensor fails to produce readable output during calibration or sample analysis | error |
pO2 A to A Drift | pO2 sensor drift from previous analysis | During analysis sequence, pO2 sensor performance changed significantly from previous analysis | error |
pO2 Calculation | pO2 sensor reading could not be calculated due to sensor or dependent-sensor error | pO2 calculation fails during calibration or analysis sequence | error |
pO2 E-zero Drift | pO2 sensor zero point drift from calibration | During analysis sequence, pO2 sensor performance changed significantly from successful 2-point calibration | error |
pO2 Overload | pO2 sensor reading exceeded software limits during calibration or analysis | pO2 sensor output exceeds maximum threshold during calibration or sample analysis | error |
pO2 Slope | pO2 calibration standard difference did not meet minimum specifications | During 2-point calibration, pO2 sensor response between standards below threshold | error |
pO2 Slope Drift | pO2 sensor slope drift from previous calibration | During last calibration, pO2 sensor performance changed significantly from previous calibration | error |
pO2 Stability | pO2 sensor reading did not reach stable endpoint during calibration or analysis | pO2 sensor reading continues to drift or fluctuate beyond stability threshold | error |
Preheater Out Of Tolerance | Preheater temperature fell outside tolerances during last sequence | Preheater temperature out of range during operation | warning |
Printer Door Open | Thermal printer door not fully closed | Printer door sensor indicates printer door is open | warning |
Printer Out Of Paper | Thermal printer paper depleted or not detected | Printer paper sensor does not detect available paper roll | warning |
Probe Failed To Home | Sample probe unable to reach home position | Sampler unable to move sample probe to required position | error |
Reference Days Expiration | Reference cartridge use life expired | Reference cartridge use life expiration exceeded | warning |
Reference Invalid Part Number | Reference cartridge has invalid part number | Reference cartridge detected with invalid part number | error |
Reference Lot Expiration | Reference cartridge shelf life expired | Reference cartridge lot expiration date exceeded | warning |
Reference Sample Use Expiration | Reference cartridge reached maximum sample cycles | Reference cartridge sample count reached maximum allowed | warning |
RFID Tag Not Found | RFID tag for one or more consumables was not found by reader | Consumable cartridge installed without readable RFID tag | error |
Rotary Valve Failed To Home | Rotary valve unable to reach home position | Rotary valve unable to move to required position | error |
Sample Contamination | Blood debris detected in Sensor Card cuvette; CO-Ox parameters are unreliable | Analyzer detects blood debris inside cuvette during sample analysis | error |
Sample Position | No sample detected at one or more air detectors when expected during sequence | During last sequence, sample position not confirmed by air detectors | error |
Sampler Failed Reaching Top Sensor | Sampler unable to reach top sensor position | Sampler unable to move to top sensor position | error |
Sampler Homing Failed | Sampler homing sequence failed | Sampler unable to complete homing sequence | error |
Sampler Positioning Failed | Sampler unable to reach required position | Sampler unable to move to required position | error |
Scanner Communications | Barcode scanner communication failure | System bar code scanner did not respond or respond correctly | error |
Scanner Timeout | Barcode scanner timeout | System bar code scanner did not respond when expected | error |
Sensor Card Days Expiration | Sensor card use life expired | Sensor cartridge use life expiration date exceeded | warning |
Sensor Card Invalid Part Number | Sensor card programmed with wrong part number | Sensor cartridge detected with invalid part number | error |
Sensor Card Lot Expiration | Sensor card shelf life expired | Sensor cartridge lot expiration date exceeded | warning |
Sensor Card Sample Use Expiration | Sensor card reached maximum sample cycles | Sensor cartridge sample count reached maximum allowed | warning |
Short Sample | No sample detected at one or more air detectors when expected | During last sequence, sample not detected at one or more air detector positions | error |
Spectro Heater Out Of Tolerance | Spectrometer temperature fell outside tolerances during last sequence | Spectrometer heater temperature out of range during operation | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.