Nihon Kohden CSM-1901: troubleshooting and error codes
The ServiceForge knowledge base holds 439 documented symptoms for the Nihon Kohden CSM-1901, with 811 ranked causes, 373 diagnostic checks with test points and expected values, and 292 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Hum noise interference in EEG waveform
Visible AC hum or electrical noise in the EEG waveform
Ranked causes
- 1. The electrode lead is damaged (occasional)
- 2. The Z electrode is not attached to the patient (AE-918P) or Cz and Z electrodes not attached (AE-920P) (common)
- 3. Electrodes 1 and 2 or the EEG1 (+) and (-) electrodes are not attached (common)
- 4. If the bed is metal, the bed is not grounded (occasional)
- 5. The provided power cord is not used (occasional)
- 6. Several medical devices are used on the patient (occasional)
- 7. There is an AC outlet or table tap near the patient or bed (occasional)
- 8. A desk lamp or fluorescent light is turned on (occasional)
- 9. The patient is touching some metal part (occasional)
- 10. An electric blanket is used (occasional)
- 11. There is a cell phone or radio near the patient (occasional)
- 12. The electrodes are not properly attached to the patient's head (AE-120A) (common)
- 13. The electrodes are not properly connected to the EEG head set (AE-120A) (common)
13 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG noise
ECG baseline unstable with artifact, may be from respiration, body movement, EMG, electrode issues, or electromagnetic interference
Ranked causes
- 1. Baseline instability from respiration or body movement (common)
- 2. EMG noise superimposed (common)
- 3. Electrode lead pulled (occasional)
- 4. Electrode dry or poor contact (common)
- 5. Electrode lead clip poor contact (occasional)
- 6. High contact resistance between skin and electrode (occasional)
- 7. Electromagnetic interference from electric blanket (occasional)
- 8. Noise from ESU (Electrosurgical Unit) (occasional)
- 9. Inadequate equipotential grounding (occasional)
Assess electrode site for body movement
- Test point
- Electrode position on patient
- Expected
- Site selected with minimal body movement
8 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check electrodes
ECG electrode connection problem: lead disconnected, electrode detached, cord disconnected, or poor contact between clip and electrode
Ranked causes
- 1. Operator disconnected electrode lead or ECG cord (common)
- 2. Electrode lead detached from electrode (common)
- 3. Electrode cannot attach firmly to skin (common)
- 4. Electrode lead disconnected from ECG connection cord (occasional)
- 5. Poor contact between electrode clip and electrode (occasional)
- 6. Electrode lead damaged (occasional)
- 7. Different electrode types used together (rare)
- 8. Polarization voltage too high (rare)
Verify ECG lead connections to monitor
- Test point
- ECG connection port at monitor
- Expected
- All ECG leads firmly connected
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 low quality signal
SpO2 probe signal quality degraded due to patient movement, low pulse, ESU, or light
Ranked causes
- 1. Patient body movement (common)
- 2. Low pulse rate (occasional)
- 3. Respiratory fluctuation (occasional)
- 4. Probe not attached properly (common)
- 5. During IABP procedure (occasional)
- 6. Electrosurgical unit (ESU) in use (occasional)
- 7. Probe exposed to strong ambient light (occasional)
- 8. Measurement during CPR (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EEG electrode check needed
Screen message 'Silence EEG Check Electrodes1' appears; poor signal quality
Ranked causes
- 1. Electrode attachment failure (common)
- 2. Electrode lead disconnected from EEG cord (common)
- 3. Poor electrode lead clip contact (occasional)
- 4. Electrode lead damaged (occasional)
- 5. Abnormal polarization voltage (occasional)
- 6. Electrode impedance too high (common)
- 7. EEG cord temporarily disconnected and reconnected (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 value is unstable
Oxygen saturation reading fluctuates or varies excessively on display
Ranked causes
- 1. Probe size inappropriate for patient (common)
- 2. Probe on limb with NIBP cuff or measurement catheter (common)
- 3. Measuring at site with large blood flow fluctuation (occasional)
- 4. Electrosurgical unit (ESU) interference (occasional)
- 5. Probe not attached at appropriate site (common)
- 6. Probe deteriorated or damaged (occasional)
- 7. Measuring on venous pulse (rare)
Verify probe size matches patient limb
- Test point
- Probe and patient limb circumference
- Expected
- Probe appropriate size for patient
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arterial blood pressure cannot be measured patient safety
Arterial pressure monitoring system does not produce valid blood pressure measurements via hemodynamic unit.
Ranked causes
- 1. Measurement data is out of range (common)
- 2. Blood pressure transducer malfunction (occasional)
- 3. Connection cable is damaged (occasional)
- 4. Transducer not appropriate for technology used (occasional)
- 5. Technology settings are incorrect on bedside monitor (common)
- 6. Unspecified transducer is used (common)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas module malfunction or power issue
Gas module displays 'Gas Device Failure' termination message or 'Check GAS Device' silence message due to hardware or power failure
Ranked causes
- 1. Gas module hardware failure (rare)
- 2. Power cord disconnected (AG-920RK, GF-200/GF-300 series) (occasional)
- 3. Measurement switch turned off (common)
- 4. Gas Measurement set to Off in menu (GF-100/GF-200/GF-300) (occasional)
- 5. Multi-link or connection cable disconnected (common)
- 6. Communication error between unit and monitor (occasional)
Verify measurement switch status on unit
- Test point
- Measurement switch on gas module
- Expected
- Switch in ON position
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measured value of gas is low
Gas concentration measurement (e.g., CO2, N2O, O2) displays lower value than expected clinically.
Ranked causes
- 1. Sampling gas leaking from sampling line or T-piece (common)
- 2. Unspecified sampling line is used (common)
- 3. Patient ventilation is poor (occasional)
- 4. Measurement sensitivity is inaccurate (occasional)
- 5. Abnormality in anesthetic machine (occasional)
- 6. Multigas unit is faulty (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 value on bedside monitor and CO oximeter do not match patient safety
Discrepancy between SpO2 reading on patient monitor and external CO oximeter measurement
Ranked causes
- 1. Probe not attached properly - emitter and detector misaligned (common)
- 2. Attachment site inappropriate for probe type (common)
- 3. Measuring site not clean (common)
- 4. Abnormal hemoglobin present (rare)
- 5. Dye injected in blood (rare)
- 6. SpO2 measured during CPR (rare)
Verify probe emitter and detector alignment
- Test point
- SpO2 probe attachment on patient
- Expected
- Emitter and detector directly facing each other
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
The waveform is not stable
EEG waveform appears unstable, fluctuates, or cannot maintain consistent baseline
Ranked causes
- 1. Electrode Z not attached (AE-918P) or Cz and Z not attached (AE-920P) (common)
- 2. Electrodes 1 and 2 or the EEG1 (+) and (-) electrodes are not attached (common)
- 3. New and old electrodes or different types of electrodes are used together (occasional)
- 4. Insufficient paste remaining on the disposable EEG electrodes (AE-120A) (common)
- 5. The belts are not fastened (AE-120A) (occasional)
- 6. A belt is loose (AE-120A) (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN transport does not start
Wireless LAN transport fails to initialize between monitors
Ranked causes
- 1. Central and bedside monitors have different software versions (common)
- 2. Incompatible host monitor connected (occasional)
- 3. Host monitor data saved in incompatible central monitor (occasional)
- 4. Incorrect transport settings on host monitor (common)
- 5. Incorrect transport settings on BSM-1700 bedside monitor (common)
- 6. QI-170P wireless LAN station not mounted on BSM-1700 (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accuracy of airway volume measurements is low
Airway volume data shows poor accuracy or appears unreliable.
Ranked causes
- 1. Flow gas is leaking (common)
- 2. Water droplets on flow tube (occasional)
- 3. Temperature and humidity settings not set on bedside monitor (occasional)
- 4. Gas composition not set on bedside monitor (occasional)
- 5. Multigas/flow unit is faulty (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cursor position misaligned or shaking on LCD
Cursor position misalignment or shaking observed on VL-190P LCD unit screen
Ranked causes
- 1. Metal objects or cables near screen (common)
- 2. Two displays located too close together (occasional)
- 3. Core unit and LCD unit not docked properly or cable disconnected (common)
- 4. Screen dirty (common)
- 5. Touch panel cleaned with unspecified cleaners containing antistatic agents (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG learning in progress
ECG learning process is running, system awaits completion before analysis can proceed
Ranked causes
- 1. Learning process initiated or in progress (common)
- 2. Immediately after starting measurement (common)
- 3. Lead change in ECG settings (occasional)
- 4. After patient admission (common)
- 5. Check Electrodes message appeared and cleared (occasional)
Monitor system status for learning completion
- Test point
- ECG window display
- Expected
- Learning process complete, analysis enabled
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hemodynamic unit measurement data not displayed on bedside monitor
Hemodynamic data from AP-170P hemodynamic unit does not appear on bedside monitor display.
Ranked causes
- 1. Necessary settings are not set on bedside monitor (common)
- 2. Multi-link cable is not connected (common)
- 3. Hemodynamic unit detected error and operation recovered (occasional)
- 4. Bedside monitor software version incompatible with hemodynamic unit (rare)
- 5. Hemodynamic unit is malfunctioning (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP measured value differs from estimated value
Invasive blood pressure reading does not match expected or calculated pressure value
Ranked causes
- 1. Air bubbles in measurement circuit (common)
- 2. Extra tube connected in circuit (occasional)
- 3. Blood pressure transducer position inappropriate (occasional)
- 4. Blood pressure transducer with different sensitivity used (rare)
- 5. Zero balance adjustment needed (common)
Inspect measurement circuit for air bubbles
- Test point
- Transducer line and circuit tubing
- Expected
- No visible air bubbles in tubing
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 check probe
SpO2 probe not attached properly or probe damaged/short-circuited
Ranked causes
- 1. Probe not attached to patient properly (common)
- 2. Poor connector contact (occasional)
- 3. Probe damaged or short-circuited; pulse waveform flat (occasional)
- 4. Probe not at appropriate site; waveform displayed (common)
- 5. Probe deteriorated; waveform displayed (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 connector off
SpO2 connection cord removed or disconnected from SpO2 socket of input unit
Ranked causes
- 1. SpO2 connection cord removed (common)
- 2. SpO2 connection cord disconnected (common)
- 3. SpO2 connection cord faulty (occasional)
- 4. Probe not properly connected to patient cable (occasional)
- 5. Monitor cannot identify probe (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Warming up message does not disappear
The Warming up message persists on display longer than normal or indefinitely after power-on.
Ranked causes
- 1. Unit is still warming up normally (common)
- 2. Sampling line or exhaust gas tube is bent (occasional)
- 3. Water trap or sampling line is kinked or blocked (occasional)
- 4. Exhaust gas tube directly connected to facility scavenging system (occasional)
- 5. Multigas unit is faulty (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC interference on ECG waveform
50/60 Hz noise visible as regular oscillations on ECG trace
Ranked causes
- 1. Electric blanket in use (occasional)
- 2. Electrode dry or poor attachment to skin (common)
- 3. Filter set to Diag instead of Mon or Max (common)
- 4. Bedside monitor and external instruments not grounded properly (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AE-120A EEG head set electrode impedance high or disconnected patient safety
Silence EEG Check Electrodes message; electrode not contacting scalp, lead disconnected, or impedance too high
Ranked causes
- 1. No gel in contact with electrode (common)
- 2. Electrode lead not connected to head set (common)
- 3. Electrode paste not adhering to scalp or electrodes not securely attached (common)
- 4. Electrode paste dried out or electrodes came off (occasional)
Visual inspection of electrode gel contact
- Test point
- Electrode Z and other electrodes on patient scalp
- Expected
- Gel visibly in contact with scalp at all electrode locations
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baseline wandering on ECG
ECG baseline drifts or moves up and down, obscuring the trace
Ranked causes
- 1. Baseline unstable due to respiration or body movement (common)
- 2. Electrode dry (common)
- 3. High contact resistance between skin and electrode (common)
- 4. Filter set to Diag instead of Mon or Max (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bedside monitor cannot connect to network
Network connectivity failure; monitor not appearing on central monitor or unable to establish link
Ranked causes
- 1. Network cable not connected properly (common)
- 2. Network settings incorrect (common)
- 3. Monitor not selected as monitored bed on central monitor (common)
- 4. Network cable or hub faulty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS Check Sensor message (BIS Processor/BISx) patient safety
Silence alarm, BIS sensor, interface cable, or digital converter disconnected or impedance too high.
Ranked causes
- 1. Digital signal converter disconnected from PIC Plus interface (common)
- 2. BIS sensor disconnected from interface cable (common)
- 3. BIS sensor physically disconnected from patient (common)
- 4. BIS sensor impedance too high (occasional)
Verify digital converter to PIC Plus cable connection
- Test point
- Light gray connectors
- Expected
- Converter light gray side connected to cable light gray side
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blood pressure data differs from expected value patient safety
Arterial pressure measurement on bedside monitor does not match expected clinical value or other reference measurements.
Ranked causes
- 1. Air bubble in blood pressure circuit (common)
- 2. Extra tube connected to blood pressure circuit (occasional)
- 3. Blood pressure transducer is not positioned correctly (common)
- 4. Other causes requiring zero adjustment (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO sensor connection error
Screen message 'Termination CCO Check Sensor' appears; no CCO output
Ranked causes
- 1. Connection cable disconnected from APCO/IBP processor (common)
- 2. Flo Trac cable or sensor faulty (occasional)
- 3. Non-Nihon Kohden sensor in use (occasional)
- 4. Internal system failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO Unstable Baseline message patient safety
Physiological change, respiration, injection interval, or external noise affecting baseline.
Ranked causes
- 1. Physiological change affecting baseline (common)
- 2. Patient on mechanical respiration (occasional)
- 3. Injection interval too short (occasional)
- 4. Noise interference from external instruments (occasional)
Assess injectate temperature and volume
- Test point
- Injectate bath and syringe
- Expected
- Cool temperature and adequate volume (10 mL for adults)
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 check sensor alarm
CO2 sensor measurement signal insufficient or sensor deterioration detected
Ranked causes
- 1. Insufficient sensor light; water in respiration circuit (common)
- 2. CO2 sensor requires cleaning (common)
- 3. Airway or nasal adapter faulty or contaminated (occasional)
- 4. Light emitter deteriorated (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 measured value displayed as dashes
CO2 value shows '---' on monitor display instead of numeric value
Ranked causes
- 1. Fewer than three respirations detected (common)
- 2. Respiration not detected for 40+ seconds (occasional)
- 3. Zero calibration in progress (occasional)
- 4. HFO ventilation in use (rare)
Count patient respirations
- Test point
- Patient breathing pattern
- Expected
- Three or more breaths detected
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 measured value is inaccurate
CO2 reading does not match expected values or shows instability
Ranked causes
- 1. Airway adapter is dirty or degraded (common)
- 2. Patient respiration rate very high or irregular (occasional)
- 3. Suction catheter in airway adapter during measurement (occasional)
- 4. Respirator oscillation (occasional)
Inspect airway adapter condition
- Test point
- Airway adapter
- Expected
- Clean, intact adapter with no debris
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 measured value is low
End-tidal CO2 reading lower than expected on mainstream CO2 sensor
Ranked causes
- 1. CO2 mixed in inspiration (common)
- 2. Low atmospheric pressure at high altitude (occasional)
- 3. Zero calibration not performed (common)
- 4. Incorrect respiration circuit type (occasional)
Verify CO2 sensor type and inspired gas composition
- Test point
- Monitor display
- Expected
- Correct sensor kit installed per manual
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data send error to central monitor
Network connection or configuration issue prevents sending data to central monitoring system
Ranked causes
- 1. Network cable not properly connected (common)
- 2. Incorrect network settings (occasional)
- 3. Bedside monitor not selected as monitored bed (occasional)
- 4. Network cable damaged or hub faulty (rare)
Verify network cable connection to bedside monitor
- Test point
- Monitor network port
- Expected
- Network cable firmly connected
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow/Paw device check needed
Screen message 'Termination Flow/Paw Check Device' appears
Ranked causes
- 1. Flow/Paw measurement switch off (common)
- 2. Connection cable disconnected (common)
- 3. Communication error between unit and core (occasional)
- 4. Power cord disconnected (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas calibration error
Gas module displays 'Silence Gas Cal Error' message; air or gas calibration failed
Ranked causes
- 1. Sampling line or exhaust gas tube bent or clogged (common)
- 2. Contaminated air around unit (occasional)
- 3. Gas value outside air calibration range (rare)
- 4. Calibration failed; low gas cylinder pressure (AG-920RK, GF-100 series) (occasional)
Inspect sampling line and exhaust tube for obstruction or bending
- Test point
- Entire length of sampling line and exhaust tube
- Expected
- Tubing straight and unobstructed
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate inaccurate patient safety
ECG monitor displays incorrect heart rate value or unstable rate readings
Ranked causes
- 1. QRS amplitude too small (common)
- 2. QRS not detected correctly (common)
- 3. Lead or electrode position suboptimal for QRS detection (common)
- 4. Pacing detection setting inappropriate (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injectate or blood temperature cannot be measured
Temperature measurement system in hemodynamic unit does not measure or display injectate or blood temperature values.
Ranked causes
- 1. Measurement data is out of range (common)
- 2. Catheter or cable is disconnected (common)
- 3. Catheter is not Nihon Kohden specified (occasional)
- 4. Technology settings are incorrect (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measured value of gas is high
Gas concentration measurement displays higher value than expected clinically.
Ranked causes
- 1. Unspecified sampling line is used (common)
- 2. Measurement sensitivity is inaccurate (occasional)
- 3. Abnormality in anesthetic machine (occasional)
- 4. Multigas unit is faulty (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP abnormal measurement results displayed patient safety
Blood pressure measurement values are inconsistent, unrealistic or irregular on display
Ranked causes
- 1. Cuff size incorrect for patient (common)
- 2. Cuff wrapped incorrectly around arm (common)
- 3. Body movement or severe arrhythmia during measurement (occasional)
- 4. Measuring on wrong arms (common)
Verify cuff size appropriate for patient limb
- Test point
- Cuff and patient arm circumference
- Expected
- Cuff size matches patient arm measurement
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor uncalibrated or past end of life
FiO2 module displays '— O2 Uncal' message; sensor not calibrated or reached end of service life
Ranked causes
- 1. Oxygen sensor reached end of life (occasional)
- 2. Oxygen sensor not connected properly; unstable signal during calibration (common)
- 3. Incorrect calibration method selected (occasional)
- 4. Oxygen sensor not yet calibrated (common)
Verify oxygen sensor connection to FiO2 cord
- Test point
- Oxygen sensor connector and cord
- Expected
- Sensor fully seated and locked in connector
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacing detected
Pacemaker-like complexes detected by the monitor
Ranked causes
- 1. Patient has pacemaker (expected) (common)
- 2. Patient does not have pacemaker (occasional)
- 3. Electric blanket causing false pacing detection (occasional)
- 4. Monitoring neonate ECG (occasional)
Confirm patient has pacemaker
- Test point
- Patient medical history and chart
- Expected
- Pacemaker implant confirmed if present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote control does not function
Remote control fails to operate or control the bedside monitor
Ranked causes
- 1. Batteries not inserted or polarity incorrect (common)
- 2. Remote control batteries weak (common)
- 3. Wrong bed selected (common)
- 4. Remote control display number incorrect (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform and rate unstable (impedance method)
Respiratory impedance signal and rate readings are inconsistent or fluctuate
Ranked causes
- 1. Electrode positions inappropriate for respiration measurement (common)
- 2. Electrode contact poor (common)
- 3. Resp-Imp Noise Reduction off and waveform amplitude too small (occasional)
- 4. Resp-Imp Noise Reduction on and respiration-heart rate timing coincide (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform not displayed (impedance method)
Respiratory impedance measurement does not produce waveform on display
Ranked causes
- 1. Resp-Imp Measurement set to Off (common)
- 2. Electrodes, leads, or ECG connection cord not connected properly (common)
- 3. Electrode dry or poor attachment to skin (common)
- 4. High contact resistance between skin and electrode (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 cannot detect pulse patient safety
SpO2 unable to detect pulse wave; peripheral circulation insufficient
Ranked causes
- 1. Insufficient peripheral circulation for SpO2 measurement (common)
- 2. Probe attached too tightly; obstructing circulation (common)
- 3. Probe not attached to patient properly (common)
- 4. Related alarms ('Light Interference', 'Check Probe Site', 'Detecting Pulse') (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 probe failure
SpO2 probe is damaged, short-circuited, or connection cord is faulty
Ranked causes
- 1. Probe damaged (occasional)
- 2. Probe short-circuited (occasional)
- 3. SpO2 connection cord faulty (occasional)
- 4. Probe faulty; replace cord if problem persists (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 signal level degradation or probe attachment issue patient safety
Probe attachment is improper or signal degradation is detected when checking SpO2 signal level with AY-661P or AY-663P probe
Ranked causes
- 1. Attachment site is thick (common)
- 2. Probe is deteriorated (occasional)
- 3. Probe is deteriorated or damaged (occasional)
- 4. Attachment site is cold or probe attached too tightly (occasional)
Connect SpO2 probe to unit
- Test point
- SpO2 input connector
- Expected
- Probe properly connected
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
The 12-lead analysis result is not correct patient safety
12-lead ECG analysis produces inaccurate or incorrect interpretation results
Ranked causes
- 1. There was body movement when the waveforms for analysis were taken in (common)
- 2. Electrodes were detached (common)
- 3. The patient's gender is not entered for the patient information (occasional)
- 4. The patient's date of birth is not entered for the patient information (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
The waveform or measured value is not displayed
EEG waveform or parameter values missing from screen despite unit connection
Ranked causes
- 1. The multi-link cable is not connected properly (common)
- 2. The software does not support this unit (rare)
- 3. The screen layout is Fixed and the parameter is not set to display (common)
- 4. The disposable EEG electrodes are not applied or are detached (AE-120A) (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator measured value not displayed
Ventilation data from connected ventilator does not appear on bedside monitor display.
Ranked causes
- 1. Ventilator not connected to bedside monitor properly (common)
- 2. Interface or cable connected to wrong socket on ventilator (common)
- 3. Communication settings do not match (common)
- 4. Measured value is out of display range (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accuracy of C and R measurements is low
Compliance and airway resistance values show poor accuracy or appear unreliable.
Ranked causes
- 1. Patient has spontaneous breathing (common)
- 2. Auto PEEP is applied (occasional)
- 3. Flow gas is leaking (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthetic gas measuring accuracy not within acceptable range patient safety
CO2, O2, N2O, or isoflurane concentration values displayed on bedside monitor fall outside specified tolerance margins during calibration gas flow.
Ranked causes
- 1. Calibration gas cylinder pressure too low (common)
- 2. Multigas unit requires gas calibration after accuracy check failure (common)
- 3. Water trap or sampling line connection improper or contaminated (occasional)
Check flow regulator pressure meter reading
- Test point
- Pressure meter on flow regulator
- Expected
- 0.1 MPa or higher, ≥0.1 MPa
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS Connector Off message (BIS Processor/BISx) patient safety
Termination alarm, BIS digital signal converter cable or connection cord disconnected.
Ranked causes
- 1. Digital signal converter cable disconnected from module (common)
- 2. BIS parameter connection cord disconnected (common)
- 3. Connection cord faulty (occasional)
Verify digital signal converter cable to module
- Test point
- Converter cable (dark gray) at module
- Expected
- Cable fully seated on module connector
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS monitor measured value not displayed
Bispectral index value when connecting standalone BIS monitor does not appear on bedside monitor
Ranked causes
- 1. BIS monitor not connected to bedside monitor properly (common)
- 2. Interface connected to wrong socket on BIS monitor (common)
- 3. Communication settings do not match (occasional)
Verify BIS monitor connected via QF interface to bedside monitor
- Test point
- QF series interface connectors at both units
- Expected
- QF interface properly connected between BIS monitor and bedside monitor per manual
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS Sensor Failure message (BIS Processor/BISx) patient safety
Silence alarm, water on sensor, sensor dirty, or BIS sensor faulty.
Ranked causes
- 1. Water on BIS sensor (occasional)
- 2. BIS sensor dirty (occasional)
- 3. BIS sensor faulty (occasional)
Inspect BIS sensor surface for moisture or contamination
- Test point
- Sensor electrodes and adhesive
- Expected
- Dry, clean sensor surface
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure CO2 with sidestream AG-400RK unit
CO2 value cannot be obtained when using AG-400RK sidestream unit
Ranked causes
- 1. FilterLine is clogged (common)
- 2. Exhaust gas adapter is clogged (common)
- 3. CO2 unit is faulty (rare)
Inspect FilterLine for blockage
- Test point
- FilterLine
- Expected
- Clear, unobstructed flow path
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure NIBP - general failure
Non-invasive blood pressure measurement fails to complete or obtain results
Ranked causes
- 1. Noise interference preventing pressure calculation (occasional)
- 2. Air hose bent or squeezed (common)
- 3. Cuff worn out or damaged (common)
Inspect air hose for obstructions or damage
- Test point
- Air hose along entire length
- Expected
- Hose straight, unfolded, free of bends or kinks
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiac output measurement differs from expected value patient safety
Calculated cardiac output value displayed on bedside monitor does not match expected clinical value or previous measurements.
Ranked causes
- 1. Settings are incorrect (common)
- 2. ScvO2 cannot be measured (affects calculation) (occasional)
- 3. Cardiac output was measured by dye dilution method (temporal issue) (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO measured value not displayed with CCO monitor
Continuous cardiac output data from connected CCO monitor does not appear on bedside monitor display.
Ranked causes
- 1. CCO monitor not connected to bedside monitor properly (common)
- 2. Interface or cable connected to wrong socket on CCO monitor (common)
- 3. Communication settings do not match (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO waveform inaccuracy
Screen message 'Silence CCO Check Waveform' appears; CO measurement unreliable
Ranked causes
- 1. Poor arterial waveform quality (common)
- 2. Inappropriate IBP line (occasional)
- 3. Prolonged waveform inaccuracy (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- CO Catheter Not Connected message
- CO2 check external device (Microcap/Micropod)
- CO2 connector off
- CO2 overheat (AG-400RK unit)
- CO2 value is low with sidestream AG-400RK unit
- Communication loss with network device
- Core unit overheating
- Dashes displayed on window
- Device communication interrupted
- External device communication interrupted
- FiO2 connection cord disconnected from input unit
- Gas calibration cannot be done
- IBP Check Sensor message
- Incompatible parameter connected
- Input unit communication error
- Measure lamp on AG-400RK unit does not light
- Microcap or Micropod external instrument value not displayed
- Mixed anesthetic agents detected simultaneously
- Multigas unit front panel lamp does not light
- Multigas/flow unit front panel lamp does not light
- Multilink configuration error
- NIBP Weak Pulse message
- NMT sensor connections incomplete
- No sync sound output from bedside monitor
- Oxygen sensor disconnected or damaged
- Respiration impedance signal noise
- Respiration thermistor connector off
- Respiration waveform amplitude small or becomes baseline (thermistor method)
- Sample gas outlet connection error
- ScvO2 measurement data differs from expected value
- SpO2 check external device
- SpO2 detecting pulse
- SpO2 weak pulse
- Temperature Check Sensor message
- The measured value for the transcutaneous monitor is not displayed
- Water trap missing or improperly attached
- WLAN transport failed to start
- AA-910P multi amp unit disconnected
- AC power disconnected or failure
- AE-120A EEG head set power or start switch held during measurement
- AE-120A EEG head set wireless communication interrupted
- AE-920P neuro unit multilink cable disconnected
- AE-920P neuro unit receiver disconnected
- AP sensor error
- AP zero calibration error
- Arrhythmia alarm occurs frequently with normal heart rate
- Backup battery charging
- Bedside monitor clock display incorrect
- Bedside monitor screen dark or dim
- BIS Check External Device message (BIS Monitor)
- BIS processor/BISx measured value not displayed
- Calculated parameter values differ from bedside monitor display
- Cannot turn on bedside monitor power
- CCO connector disconnected
- CCO zero balance unstable
- CCO/PCCO measured value not displayed with APCO/IBP processor
- Clock failure
- CO Blood Temperature Out Of Range message
- CO Connector Off message
- CO Injectate Temperature Out Of Range message
- CO Irregular Curve message
- CO2 calibration error
- CO2 check CO2 cell
- CO2 check device (AG-400RK unit)
- CO2 device failure (AG-400RK unit)
- CO2 device failure (Microcap/Micropod)
- CO2 waveform does not appear
- Communication Error message with central monitor or receiving instrument
- EEG connector disconnected
- Electrode impedance out of range
- Flow/Paw calibration error
- Gas warming up in progress
- IBP Check Label message
- IBP Connector Off message
- IBP Out Of Range message
- IBP Zero Out of Range message
- IBP Zero Unstable message
- Input signal amplitude incorrect (standard sensitivity check failed)
- Input unit disconnected or improperly connected
- Measured value and waveform not displayed on window
- Measurement value out of calibration range
- Multigas unit front panel lamp lights in orange
- Multigas/flow unit front panel lamp lights in orange
- Multigas/flow unit measured value and waveform not displayed
- Multiple anesthetic machines or multigas units connected
- Network printer connection failed
- NIBP Air Leak alarm
- NIBP Cannot Detect Pulse message
- NIBP Connector Off message
- NIBP cuff suddenly inflates unexpectedly
- NMT communication error with bedside monitor
- No flow sensor detected
- No invasive blood pressure value appears on window
- No SpO2 data displayed with Masimo probe
- No SpO2-2 data displayed with Nihon Kohden probe
- O2 measured value displays as dashes
- O2 measured value is abnormal
- Paper not feeding from recorder
- PiCCO calibration error
- PiCCO catheter unavailable
- ProAQT calibration error
- Recording is faint
- Red LED on CO2 adapter blinking
- Respiration thermistor check sensor alarm
- ScvO2 cannot be measured
- Sine wave noise on pulse wave
- Small injectate temperature change
- SpO2 check probe site
- SpO2 external device check (non-Nihon Kohden)
- System does not start normally
- Temperature Check Label message
- Temperature Connector Off message
- Temperature value not displayed on window
- The waveform trace is interrupted (AE-120A)
- TOF monitoring device disconnected
- Water trap, sample line or exhaust tubing clogged or bent
- Waveforms recorded but trend and list cannot
- A-Fib recall data cannot be displayed on destination monitor
- Abnormal noise or smell from EEG head set
- Acquiring PWTT Data message
- AE-120A EEG head set wireless signal quality poor
- Alarm occurring on connected ventilator
- Analog external instrument disconnected from bedside monitor
- Anesthetic machine disconnected from bedside monitor
- AP sensor not connected
- AP zero imbalance
- Arrhythmia analysis disabled
- Auto lead change occurred
- Balloon burst during CO measurement
- Bedside monitor overheating
- BIS Alarm message (BIS Monitor)
- BIS Error message (BIS Monitor)
- BIS High Impedance message
- BIS Impedance Checking message
- BIS Module Failure (BIS Processor/BISx)
- BIS Noise message
- BIS Parameter Duplicated message (BIS Monitor)
- BIS Parameter Duplicated message (BIS Processor/BISx)
- BIS processor operation check
- BIS Sensor Expired message (BIS Processor/BISx)
- BIS Simulated Data message
- Blood congestion from frequent NIBP measurement
- Blood pressure waveform not output from BP OUT socket
- BSM-1700 battery pack not installed
- Calibration: Gas measuring accuracy for AG-920RK and GF-100 series multigas units
- Calibration: Gas measuring accuracy for GF-200 and GF-300 series multigas units (accuracy check only)
- Cannot check BIS impedance
- Cannot connect cuff to air hose
- CCO measurement unstable
- CCO monitor alarm active
- CCO monitor error
- CCO parameter duplicated
- CCO patient information missing
- CCO zero imbalance
- CCO zeroing complete
- Cerebral/somatic oximeter disconnected from bedside monitor
- CeVOX module disconnected
- CeVOX module warming up
- CO Computing CO message
- CO Detecting Baseline message
- CO Inject message
- CO Injection Timeout message
- CO Out Of Range message
- CO Parameter Duplicated message
- CO2 adapter failure
- CO2 calibration in progress (AG-400RK unit)
- CO2 cell off (TG-970P, TG-980P)
- CO2 change scrubber (AG-400RK unit)
- CO2 external device failure (Microcap/Micropod)
- CO2 line block (AG-400RK unit)
- CO2 line block (Microcap/Micropod)
- CO2 measured value is high
- CO2 no filterline (AG-400RK unit)
- CO2 no filterline (Microcap/Micropod)
- CO2 not calibrated
- CO2 parameter duplicated
- CO2 parameter duplicated (Microcap/Micropod)
- CO2 please wait (AG-400RK unit)
- CO2 purging (AG-400RK unit)
- CO2 purging (Microcap/Micropod)
- CO2 sensor failure
- CO2 value is high with sidestream AG-400RK unit
- CO2 warming up (AG-400RK unit)
- CO2 warming up (Microcap/Micropod)
- Connected parameter device failure
- Core unit CPU reset
- Cuff pressure does not hold or leaks excessively
- Cursor moves or window opens without touch
- Dashes displayed on flow/paw window
- Dots missing from recording
- Duplicate analog communication cable labels
- Duplicate PiCCO parameter
- Duplicate TOF parameter
- ECG cannot analyze
- ECG low millivolt signal
- ECG module failure
- EEG baseline reset
- EEG calibrating
- EEG device communication error
- EEG device failure
- EEG electrode impedance too high
- EEG head set abnormally hot
- EEG head set LEDs do not light on power-on
- EEG impedance checking
- Electrode attached correctly but message displayed; waveform not shown
- Electrode detachment not detected
- Error condition on connected ventilator
- Expiration and inspiration phases reversed (thermistor method)
- External CCO device disconnected
- Flow/Paw calibration complete
- Flow/Paw device failure
- Flow/Paw measurement out of range
- Flow/Paw zeroing in progress
- Gas calibration complete (informational)
- Gas calibration in progress (informational)
- Gas measurement PCWP measuring (informational)
- Gas module overheat (GF-200/GF-300 series)
- Gas purging operation in progress
- High-cut filter response out of specification
- IBP Catheter Disconnect message
- IBP displays 0 mmHg when using JP-911P interface
- IBP Not Zero message
- IBP Parameter Duplicated message
- IBP Zeroing Complete message
- IBP Zeroing message
- Inflation speed outside acceptable range (adult/child mode)
- Inflation speed outside acceptable range (neonate mode)
- Inflation Time Passed message
- Injectate temperature too high
- Input unit battery pack problem
- Input unit disconnected during transport
- Input unit not connected or not ready
- Input unit not transmitting data to bedside monitor
- Input unit power error
- Internal gas module tube clogged
- LED on remote control does not blink or light
- Menu window does not close
- MOTION waveform not responding to head set movement
- Multilink power error
- Multiple AA-910P multi amp units connected
- Multiple FiO2 connection cords connected
- Multiple JA-920P interface units connected
- Multiple neuro units connected to bedside monitor
- Multiple oximeters connected to bedside monitor
- Multiple transcutaneous monitors connected
- Network cable not properly connected to printer
- Neuro unit (AE-918P) EEG accuracy check
- NIBP auto measurement does not start despite interval passing
- NIBP calibration sensor readings out of tolerance
- NIBP Check Interval Setting message
- NIBP cuff does not inflate
- NIBP cuff inflates rapidly with pressure rising quickly
- NIBP Cuff Occlusion message
- NIBP cuff suddenly deflates during inflation
- NIBP data appears dark or shows dashes on window
- NIBP Diastolic Out Of Range message
- NIBP High Cuff Pressure alarm
- NIBP Inflation Pressure low message
- NIBP Measurement Time-Out
- NIBP Module Failure
- NIBP Out Of Range message
- NIBP pressure does not rise
- NIBP Remeasuring message
- NIBP Safety Circuit Running message displayed
- NIBP Systolic Out Of Range message
- NIBP Zeroing message
- NMT module calibration failed
- NMT module malfunction
- NMT module not calibrated
- NMT temperature sensor not connected
- No BIS data displayed on window
- No CO data displayed on window
- No O2 data on window
- No printing from recorder (paper feeding only)
- No recording paper loaded
- No SpO2 data displayed with Nellcor probe
- Not enough free space in USB memory
- Nothing appears on remote control display window
- Nothing appears on the EEG window
- Oxygen sensor calibrating (informational)
- Parameter connector disconnected
- Patient body temperature too low for TOF measurement
- PCWP automatic measurement fails
- PCWP measurement not recognized
- PiCCO catheter disconnected
- PiCCO catheter not connected
- PiCCO check external device
- PiCCO external device disconnected
- PiCCO external device error
- PiCCO monitor alarm active
- PiCCO monitor error
- PiCCO parameter duplicated
- Please Wait message
- Power lamps blinking on VL-190P LCD and core unit
- Power unit failure in core unit
- Printed characters garbled; data cannot print
- Printer is busy
- Printer network settings incorrect
- Printer not powered on
- Printer out of order
- ProAQT sensor expired
- ProAQT sensor LED not pulsating or orange LED lights
- PWTT Uncal message
- Recorder door open
- Recorder operates only some of the time
- Recorder unit faulty
- Recorder unit not connected
- Recorder unit not installed or icon disabled
- Recording paper not loaded properly
- Recording start position inaccurate
- Recording starts suddenly without key operation
- Respiration impedance measurement off
- Respiration rate inaccurate (thermistor method)
- Respiration thermistor parameter duplicated
- Respiration waveform not displayed (thermistor method)
- Safety circuit over-pressure limit not functioning (adult/child mode)
- Safety circuit over-pressure limit not functioning (neonate mode)
- ScvO2 ambient temperature high
- ScvO2 calibration error
- ScvO2 light interference
- ScvO2 module internal error
- ScvO2 module overheat
- ScvO2 signal too high
- ScvO2 signal too low
- Solenoid valve operation (V1 quick exhaust, V2 slow exhaust) out of specification
- Sometimes the waveform becomes flat
- SpO2 cannot measure
- SpO2 external device alarm
- SpO2 external device error
- SpO2 external device parameter duplicated
- SpO2 light interference
- SpO2 module failure
- SpO2 no probe
- SpO2-2 parameter duplicated
- SQI bar graph not displayed on home screen
- Standby lamp on AG-400RK CO2 unit does not light
- SVV highly variable pulse rate
- Tb signal unstable
- TD measurement limit reached
- TD measurement timeout
- TD slow injection
- TD temperature sensor error
- Temperature Parameter Duplicated message
- The analog waveform is not displayed
- The impedance check was not performed correctly (AE-120A)
- The measured value for the cerebral/somatic oximeter is not displayed
- The same waveform appears in a different channel (AE-120A)
- The tcPO2/tcPCO2 measured value displayed on the bedside monitor is abnormal
- The waveform of a channel including optional electrodes is flat (AE-120A)
- Thrombosis risk from NIBP measurement
- Time constant out of specification
- Time zone mismatch between monitors
- TOF acceleration signal too high during calibration
- TOF acceleration signal too low during calibration
- TOF acceleration signal unstable during calibration
- TOF device error
- TOF measured value not displayed
- TOF measurement stimulus applied too soon after previous measurement
- TOF supramaximal stimulation current not found
- TOF temperature sensor out of measurement range
- Touch position misaligned on LCD screen
- Touch screen keys not responding to input
- Touch screen location misalignment
- Transcutaneous monitor disconnected from bedside monitor
- Unspecified accuracy (AG-920RK only)
- Upper or lower limit alarm occurring
- USB keyboard or mouse not functioning
- Ventilator connection conflict
- Ventilator disconnected from bedside monitor
- Visual inspection of BIS processor (QE-910P)
- Visual inspection of EEG head set (AE-120A) and accessories
- Visual inspection of multigas unit (AG-920RK, GF-100, GF-200, GF-300 series)
- Visual inspection of multilink cable and wireless receiver (AE-920P neuro unit)
- Visual inspection of neuro unit (AE-918P)
- Vital reaction detected during TOF pre-PTC phase
- Water in water trap container
- Waveform amplitude fails simple method check
- Wireless signal loss or interrupted waveforms at 5 m distance
- Zeroing complete
379 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AC POWER ERROR | AC power disconnected from monitor; AC power supply failure or power outlet failure | AC power cable disconnected or power failure detected | critical |
Acquiring PWTT Data | System acquiring pulse wave transit time calibration data | PWTT measurement initiation | info |
— AP Zero Cal Error | AP zero calibration failed | Circuit not open to air or pressure unstable | error |
BIS Alarm | Alarm condition on external BIS monitor | BIS monitor generates alarm | warning |
BIS Check External Device | BIS monitor disconnected or communication error | BIS monitor cable disconnect or comm failure | error |
BIS Check Sensor | BIS sensor, interface cable, or converter disconnected; impedance too high | Sensor/cable disconnect; high impedance | error |
BIS Connector Off | BIS digital signal converter or parameter cord disconnected | BIS cable or connection cord removed | error |
BIS Error | Error condition on external BIS monitor | BIS monitor detects error | error |
BIS High Impedance | High impedance between BIS sensor and skin | Sensor-skin interface impedance elevated | warning |
BIS Impedance Checking | System checking electrode impedance | Automatic impedance check during initialization | info |
BIS Module Failure | BIS module hardware is faulty | Module component failure detected | critical |
BIS Noise | Electromagnetic or physiological noise detected | Noise interference during BIS measurement | warning |
BIS Parameter Duplicated (Monitor) | Two or more BIS monitors connected | Multiple BIS monitor interfaces detected | error |
BIS Parameter Duplicated (Processor) | Two or more BIS sensors connected to BIS processor | Multiple BIS sensors detected | error |
BIS Sensor Expired | BIS sensor past expiration date or use limit | Sensor expires after set date or number of uses | warning |
BIS Sensor Failure | BIS sensor dirty, wet, or faulty | Sensor condition deterioration detected | error |
BIS Simulated Data | Demonstration BIS sensor in use; data for demo only | Demo sensor detected | warning |
— (blank screen during startup) | PC function is starting or in recovery process | Normal system startup or recovery in progress | info |
— (blank screen or error during operation) | PC function was down and restart is in progress | PC function shutdown detected and restart initiated | warning |
Change/Save PCWP | Operator instruction to register PCWP data in hemodynamic system | PCWP measurement completed or manual measurement frozen | info |
Charging Battery | Backup battery is weak from prolonged AC disconnection or undergoing periodic charging cycle | Backup battery voltage low or scheduled charging cycle active | info |
— Check SSD card | SSD card has exceeded its service life expiration date | SSD card age reaches manufacturer end-of-life date | warning |
Clock Failure | Real-time clock failure due to low lithium battery or faulty RTC circuit | RTC chip detects battery voltage low or circuit malfunction | warning |
CO Catheter Not Connected | Thermodilution catheter not connected to CO cord | Catheter disconnect or missing connection | error |
CO Computing CO | System measuring and calculating cardiac output | Thermodilution curve being processed | info |
CO Connector Off | CO connection cord removed or disconnected from MULTI socket | CO cord disconnect detected | error |
CO Detecting Baseline | System searching for baseline temperature curve | CO measurement initiated | info |
CO Inject | System ready for injectate bolus | CO measurement initialized and awaiting injection | info |
CO Injection Timeout | Injectate not detected within timeout period | No injectate detected after 4 min or 30 sec | warning |
CO Irregular Curve | Abnormal thermodilution curve suggesting late return or shunt | Curve analysis shows abnormality | warning |
Communication Loss | Network communication lost with bedside monitor or networked device; interbed monitor disconnected | Network cable disconnected; network settings incorrect; network communication timeout; hub failure | warning |
CO Out Of Range | Calculated CO value outside monitor range | CO result exceeds monitor limits | warning |
CO Parameter Duplicated | Two or more CO connection cords connected | Multiple CO cords detected | error |
CO Tb Out Of Range | Blood temperature outside 15-45°C range | Temperature measurement outside physiological range | warning |
CO Ti Out Of Range | Injectate temperature 27°C (81°F) or higher | Injectate too warm for measurement | warning |
CO Unstable Baseline | Baseline unstable due to physiology or noise | Baseline instability detected | warning |
Deflate Balloon | Operator instruction to deflate PCWP balloon | PCWP measurement completed | info |
— EEG CAL | EEG system performing calibration | EEG calibration process initiated | info |
— EEG Calibrating | EEG calibration in progress | Calibration routine running | info |
— EEG High Impedance | EEG electrode impedance exceeds acceptable level during impedance check | Electrode impedance measurement exceeds system threshold | info |
— EEG Impedance Checking | EEG system performing electrode impedance check automatically | Automatic impedance check initiated by system | info |
— EEG Low Battery | AE-120A EEG head set battery charge is low but not empty | Remaining battery charge approximately one hour or less | info |
— EEG Resetting Baseline | EEG baseline reset in progress | Baseline reset routine running | info |
— EEG Resetting Baseline | EEG system resetting baseline for signal | Baseline reset process initiated | info |
— EEG Start Switch Press and Held During Meas | START switch held down while measurement started on AE-120A head set | START switch pressed and held during measurement initiation | info |
— EEG Start Switch Pressed | START switch activated during ongoing EEG measurement | START switch pressed while measurement already in progress | info |
— esCCO Ready For Cal | esCCO system ready for calibration; required parameter data available | All required parameters obtained and ready for calibration | info |
— Flow/Paw Cal Complete | Flow/Paw calibration successfully completed | Zero calibration completes | info |
— Flow/Paw Out Of Range | Flow/Paw measurement exceeds valid range | Measurement value outside acceptable limits | info |
— Flow/Paw Zeroing | Flow/Paw zero calibration in progress | Calibration procedure running | info |
Freeze Waveform | PCWP automatic measurement failed; manual mode required | Automatic PCWP analysis unable to complete | warning |
— - Gas Cal Complete | Air or gas calibration completed successfully | Calibration procedure finished without errors | info |
— - Gas Calibrating | Gas or air calibration procedure in progress | Calibration started and running | info |
— - Gas Purging | Gas module performing automatic sampling line purge to clear clogging (AG-920RK only) | Sampling line clogging detected during operation (AG-920RK) | info |
— - Gas Warming Up | Gas module initializing after power-on or configuration change | Power applied to gas module or instrument configuration changed (GF-100) | info |
Inflate Balloon | PCWP measurement awaiting balloon inflation | PCWP measurement initiated without balloon inflation | info |
Inflation Time Passed | Cuff inflation time exceeded maximum for patient category | Venous puncture mode inflation exceeded time limit | warning |
Input Unit Disconnect | Input unit cable or connector not properly inserted; abnormality in input unit detected | Input unit physically disconnected or communication loss to input unit | error |
Input Unit Failure | Communication error between input unit and core unit; abnormality in memory unit; or data transferred with mismatched time zone settings | Input unit cannot communicate with core unit; memory error detected; or time zone mismatch detected during data transfer | error |
Input Unit Pwr Error | Abnormality in power voltage supplied to input unit | Power supply to input unit fails or produces abnormal voltage | error |
Interface Unit Duplicated | Two or more JA-920P interface units connected to system simultaneously | System detects multiple JA-920P interface units | error |
Main Unit Overheat | Core unit temperature exceeds safe operating threshold | Internal temperature sensor detects overheating condition | error |
Measuring | PCWP value measurement in progress | Automatic PCWP measurement started | info |
Monitor Off | Interbed monitor device is turned off or disconnected from network | Interbed device powered down or network disconnected | warning |
Multiamp Unit Disconnect | AA-910P multi amp unit accidentally or intentionally disconnected from system | Multi-link cable disconnected from AA-910P unit | warning |
MULTI AMP UNIT DUPLICATED | Two or more AA-910P multi amp units connected to system simultaneously | System detects multiple AA-910P units on multilink | error |
Multilink Config Error | Abnormality in QF series interface or IF series communication cable; multi-link cable damaged or short-circuited; faulty circuit operation | Cable communication fails or short circuit detected on multi-link cable; circuit board detects malfunction | error |
Multilink Power Error | Power supply to multilink system is faulty or bedside monitor has internal power fault | Power monitoring detects fault in multilink power supply | error |
NIBP Cannot Detect Pulse | System unable to detect patient's pulse during measurement | Pulse wave amplitude too small or cuff incorrectly positioned | warning |
NIBP Check Interval Setting | NIBP measuring at 1-minute intervals for extended duration | Continuous NIBP measurement at 1-minute intervals for more than 30 minutes | info |
NIBP Connector Off | Air hose disconnected from NIBP socket | Air hose removed or loose connection | error |
NIBP Cuff Occlusion | Air flow blocked in cuff tubing | Air hose kinked or compressed | error |
NIBP Dia Out Of Range | Diastolic pressure exceeds monitor upper limit | Patient diastolic blood pressure above monitor range | warning |
NIBP High Cuff Press | Excessive cuff pressure applied | Cuff pressure exceeds safety limit during measurement | warning |
NIBP Inflation Press low | Cuff inflation pressure insufficient | Patient systolic pressure higher than expected or automatic inflation limit reached | warning |
NIBP Meas Time-Out | NIBP measurement exceeded time limit | Measurement time exceeded due to artifact or physiological factors | warning |
NIBP Module Failure | NIBP measurement module is faulty | Module hardware failure detected | critical |
NIBP Out Of Range | Both systolic and diastolic pressures exceed monitor range | Patient blood pressure above measurement range | warning |
NIBP Remeasuring | NIBP measurement restarted due to artifact or error | Arrhythmia; body movement; shaking; air hose squeezing detected during measurement | warning |
NIBP Safety Circuit Running | NIBP measurement stopped by safety circuit | Air hose bent or squeezed during measurement | warning |
NIBP Sys Out Of Range | Systolic pressure exceeds monitor upper limit | Patient systolic blood pressure above monitor range (typically >260 mmHg) | warning |
NIBP Weak Pulse | Patient's pulse wave too small for NIBP detection | Inadequate pulse amplitude or poor cuff application | warning |
NIBP Zeroing | Zero balance adjustment in progress | System initializing NIBP channel | info |
Not Enough Free Space in USB Memory | Insufficient free space on USB flash drive to save captured screen data | USB drive free space falls below required threshold during screen capture | warning |
— - O2 Calibrating | Oxygen sensor calibration in progress | Calibration procedure started | info |
— - O2 Uncal | Oxygen sensor not calibrated or past end of service life | Sensor reaches end of life or lacks valid calibration | warning |
Parameter Duplicated | Same parameter type connected multiple times to system | System detects duplicate parameter or sensor connections | warning |
Parameter Not Available | Incompatible parameter connected; mixed Nihon Kohden and third-party SpO2 probes; or incompatible sensor on restricted MULTI socket | Incompatible parameter detected; multiple SpO2 measurement methods active; sensor connected to restricted socket | warning |
— PiCCO Cal Error | PiCCO calibration failed | Invalid TD measurements or calibration data | error |
Please Wait | Measurement restarted too soon after pause | New measurement started within 2 seconds of previous measurement stop | info |
— Power Switch Press and Held During Meas | POWER switch held down while measurement started on AE-120A head set | POWER switch pressed and held during measurement initiation | info |
Power Unit Failure | Abnormality in power unit of CU-191R or CU-192R core unit detected | Internal power supply monitoring detects fault condition in power unit | critical |
— ProAQT Cal Error | ProAQT calibration failed | Missing CO value or AP measurement | error |
PWTT Uncal | NIBP calibration not completed for PWTT | PWTT measurement attempted without prior NIBP calibration | warning |
Recovery Anes Alarm | Alarm condition detected on connected anesthetic machine | Alarm occurrence on anesthetic machine device | warning |
Recovery Anes Error | Error condition detected on connected anesthetic machine | Error on anesthetic machine device | warning |
Recovery Battery Calibration Error | Battery capacity calibration failed or is out of accuracy | Battery calibration process failure or checksum error | warning |
Recovery Battery Charge FET Error | Battery charge control FET (Field Effect Transistor) failure | FET control circuit malfunction in battery pack | error |
Recovery Battery Charging Abnormal | Battery charging process abnormal or blocked | Charging circuit malfunction or battery not accepting charge | error |
Recovery Battery Charging Cell Imbalance | Battery cells have unequal charge levels causing imbalance | Cell voltage imbalance detected during charging | warning |
Recovery Battery Discharge Abnormal | Battery discharge process abnormal or blocked | Battery discharge circuit malfunction detected | error |
Recovery Battery Discharge Cell Imbalance | Battery cells have unequal discharge levels causing imbalance | Cell voltage imbalance detected during discharge | warning |
Recovery Battery EEPROM Access Error | Battery pack EEPROM memory read/write error | Memory access failure during battery status check | error |
Recovery Battery Fuse Cut | SB-920P battery pack internal fuse has opened | Battery pack circuit protection fuse activated | error |
Recovery Battery Impedance Abnormal | SB-920P battery pack internal impedance out of specification | Battery impedance measurement exceeds threshold | warning |
Recovery Battery Remaining Capacity Abnormal | Battery remaining capacity calculation abnormal | Capacity estimation failure or fuel gauge error | warning |
Recovery Battery Replacement Life | SB-920P battery pack has reached end of service life | Battery age or cycle count reaches replacement threshold | warning |
Recovery Battery RSOC Unstable | Remaining State Of Charge calculation unstable or inaccurate | Battery charging capacity not accurately detected | warning |
Recovery Battery Temperature Abnormal | SB-920P battery pack temperature is abnormally high | Battery pack operating temperature exceeds safe limit | warning |
Recovery Battery Thermistor Error | Battery temperature sensor malfunction | Thermistor circuit failure or sensor reading out of range | error |
Recovery Battery Trickle Charge Abnormal | Trickle charge maintenance function abnormal | Trickle charge circuit malfunction detected | warning |
Recovery Battery Voltage High | SB-920P battery pack voltage exceeds safe operating level | Battery charging circuit abnormality detected | error |
Recovery Battery Voltage Low | SB-920P battery pack voltage below safe operating level | Battery pack voltage drops below minimum threshold | warning |
Recovery CCO Alarm | Alarm active on external CCO monitor | Alarm condition on CCO monitor | error |
Recovery CCO Error | Error detected on external CCO monitor | Error condition on CCO monitor | error |
Recovery Check Input Unit Battery (BSM-1700 series) | Battery pack status abnormal in BSM-1700 series input unit | Battery pack missing or defective in BSM-1700 unit detected on startup | warning |
Recovery ECG Module Failure | ECG measurement module hardware malfunction | ECG module circuit failure or internal communication error | error |
Recovery esCCO Error | esCCO measurement value out of range or PWTT invalid | Measurement outside acceptable range or invalid pulse transit time | warning |
Recovery - Gas Line Block | Gas measurement recovered after line blockage detected | Blockage detected and possibly cleared during operation | warning |
Recovery Input Unit Failure | Communication error or abnormality between input unit and core unit | Communication failure or memory abnormality in input unit detected | error |
Recovery or Silence A-Fib | Atrial fibrillation detected | Irregular atrial activity without P waves detected for sustained period | warning |
Recovery PC Board Error | Abnormality in CU-191R or CU-192R core unit PC board | PC board voltage or internal communication error detected | error |
Recovery PCCO Alarm | Alarm active on external PiCCO monitor | Alarm condition on PiCCO monitor | error |
Recovery PCCO Error | Error detected on external PiCCO monitor | Error condition on PiCCO monitor | error |
Recovery Speaker Failure | Speaker in CU-191R or CU-192R core unit is faulty | Speaker circuit failure or audio output error detected | warning |
Recovery StO2 Error | Error condition detected on connected cerebral oximeter | Error on cerebral oximeter device | warning |
Recovery System Board Failure XXX (Board name) | System board abnormality in specified board with voltage or communication error | Board voltage out of specification or internal communication bus error | error |
Recovery System Failure PC | PC function abnormality preventing normal operation | PC subsystem failure or internal communication error detected | error |
Recovery System Start Error XXX (Board name) | Abnormality in specified system board or internal memory during startup | Board failure or memory error detected during system initialization | error |
Recovery tcPO2/tcPCO2 Alarm | Alarm condition detected on connected transcutaneous monitor | Alarm occurrence on transcutaneous monitor device | warning |
Recovery TOF Error | Error on the TOF monitoring device | TOF device detects internal error | error |
Recovery Unit Not Available | Incompatible unit is connected to the system | Incompatible module or interface device detected during connection | error |
Recovery - Vent Alarm | Active alarm on connected ventilator detected by monitor | Ventilator alarm condition transmitted to bedside monitor | warning |
Recovery - Vent Error | Error condition detected on connected ventilator | Ventilator error state transmitted to bedside monitor | warning |
— ScvO2 Cal Error | ScvO2 calibration error | ScvO2 calibration procedure failed | error |
Silence Analog Parameter Duplicated | Duplicate communication cable labels on analog external instruments | Two communication cables with same label connected simultaneously | warning |
Silence Anes Parameter Duplicated | Multiple anesthetic machines or multigas units connected to bedside monitor | Two or more anesthetic machines or multigas units connected simultaneously | warning |
Silence AP Sensor Error | ProAQT sensor connection error | ProAQT sensor not connected or cable faulty | warning |
Silence AP Zero Imbalance | AP zero adjustment failed | AP zero adjustment process failed | warning |
Silence Asystole | Absence of cardiac electrical activity detected | No QRS complexes detected for sustained period | critical |
Silence Bigeminy | Bigeminy pattern detected (every other beat premature) | Alternating pattern of normal and premature beats | warning |
Silence Bradycardia | General bradycardia (heart rate <60 bpm) detected | Sustained heart rate below bradycardia threshold | warning |
Silence CCO Check Patient Information | Patient demographics not entered | Patient gender, age, or BSA missing from monitor | warning |
Silence CCO Check Waveform | Arterial waveform not accurate for CO measurement | Poor waveform quality detected during CO measurement | warning |
Silence CCO Parameter Duplicated | Multiple CCO devices connected to core unit | CCO monitor and APCO/IBP processor both connected to core unit | warning |
Silence CCO Unstable | CCO measurement unstable or unreliable | Unstable CO output detected during measurement | warning |
Silence CCO Zero Imbalance | Zero balance not adjusted or invalid | Zero balance adjustment not performed or incomplete | warning |
Silence CCO Zeroing Complete | Zero balance adjustment successfully completed | Zero balance adjustment process completes | info |
Silence CCO Zero Unstable | Zero balance adjustment unstable or unsuccessful | Zero balance adjustment failure detected | warning |
Silence Couplet | Couplet (two consecutive premature beats) detected | Two consecutive premature beats detected | warning |
Silence Display Duplicated | Multiple displays connected in invalid configuration | Display detected on both docking socket and Display 1 socket simultaneously | error |
Silence Early VPC | Early ventricular premature complex detected | Premature ventricular contraction on T wave (R-on-T) | warning |
Silence EEG Battery Empty | Batteries in AE-120A EEG head set completely discharged | Battery level in head set reached zero | warning |
Silence EEG Check Electrodes | EEG electrode impedance issue or disconnection detected | Electrode not in contact, disconnected, or impedance too high | warning |
Silence EEG Check Electrodes1 | EEG electrode attachment or impedance problem | Poor electrode contact or impedance too high | warning |
Silence EEG Device Failure | AE-918P neuro unit or multi-link unit faulty | Neuro unit or multi-link unit hardware failure | warning |
Silence EEG No Communication | Wireless communication lost between AE-120A EEG head set and AE-920P neuro unit | Wireless signal interruption between head set and receiver | warning |
Silence EEG Parameter Duplicated | Multiple EEG neuro units with same parameter connected to bedside monitor | Two or more cables with same parameter or module connected to bedside monitor simultaneously | warning |
Silence EEG Place the Head Facing Up | Patient's head is positioned to the side during EEG measurement | Patient head turned to side during active measurement | warning |
Silence EEG Unstable Communication | Poor quality wireless signal between AE-120A EEG head set and AE-920P neuro unit | Weak wireless signal quality between head set and receiver | warning |
Silence esCCO Cal Complete | esCCO calibration completed successfully | Calibration process finished without errors | info |
Silence esCCO Cal Error | esCCO calibration failed due to invalid CO value or parameter changes | Calibration CO value out of range or patient parameters changed during calibration | warning |
Silence esCCO Check Cal Parameter | esCCO calibration parameter error or missing data | Required calibration parameters unavailable or invalid | warning |
Silence Extreme Brady | Extreme bradycardia (very slow heart rate) detected | Heart rate falls below extreme bradycardia threshold (typically <30-40 bpm) | critical |
Silence Extreme Tachy | Extreme tachycardia (very rapid heart rate) detected | Heart rate exceeds extreme tachycardia threshold (typically >180-200 bpm) | critical |
Silence Flow/Paw Cal Error | Flow/Paw calibration failed | Vibration or electromagnetic noise during calibration | warning |
Silence Flow/Paw Device Failure | Flow/Paw unit hardware failure | Flow/Paw unit malfunction detected | warning |
Silence Freq VPC | Frequent ventricular premature complexes detected | VPC frequency exceeds threshold (typically >5 per minute) | warning |
Silence - Gas Cal Error | Gas or air calibration procedure failed | Automatic calibration unable to complete successfully | warning |
Silence - Gas Check GAS Device | Gas module communication or configuration error; multiple possible causes | Module communication loss; measurement switch off; menu setting off; cable disconnect; communication error; power disconnect | error |
Silence - Gas Check Sample Line | Gas module detects improper sample line connection or excessive pressure (AG-920RK only) | Sample outlet connection or pressure fault detected during measurement (AG-920RK) | warning |
Silence - Gas Check Watertrap or Sample Line | Gas module detects water or blockage in sampling path | System detects water in trap or clogging during operation | warning |
Silence - Gas Out Of Range | Gas measurement value or calibration outside acceptable range | Measurement exceeds defined range or calibration invalid | warning |
Silence - Gas Overheat | Gas module overheating detected; fan malfunction (GF-200/GF-300 series only) | Internal temperature threshold exceeded | warning |
Silence - Gas Unspecified Accuracy | Gas measurement outside guaranteed accuracy range (AG-920RK only) | Measurement value exceeds specified accuracy bounds (AG-920RK) | warning |
Silence HiQ-View Setting Inconsistent | Settings on bedside monitor and telemetry device are mismatched | Settings discrepancy detected between monitor and telemetry device | warning |
Silence Irregular RR | Irregular RR interval pattern detected | Beat-to-beat intervals vary beyond normal variation | warning |
Silence - Mixed GAS | Multiple anesthetic agents detected simultaneously in gas stream | System detects two different anesthetic agents present at same time | warning |
Silence MPU Module Failure | Main processor unit module malfunction | Circuit operation abnormality in MPU | critical |
Silence Multiform | Multiform premature complexes detected | Premature beats of different morphologies detected | warning |
Silence NIBP Air Leak | Air leak detected in NIBP system | Cuff or air hose not connected or damaged | error |
Silence No Pacer Pulse | Pacemaker pulse not detected when expected | Pacing spike not detected during demand pacing | warning |
Silence - O2 Parameter Duplicated | Multiple FiO2 connection cords connected simultaneously | System detects more than one FiO2 cord in MULTI socket(s) | warning |
Silence Pacer Non-Capt | Pacemaker pulse not capturing ventricle (loss of capture) | Pacing spike present without corresponding QRS complex | warning |
Silence Pause | Pause in cardiac rhythm detected | Beat-to-beat interval exceeds pause threshold duration | warning |
Silence PCCO Catheter Disconnect | PiCCO catheter was removed | PiCCO catheter disconnected from cord | warning |
Silence PCCO Catheter Unavailable | PiCCO catheter faulty or incompatible | Catheter malfunction or incompatibility detected | warning |
Silence PCCO Check Ext Device | AP-170P hemodynamic unit internal error | Internal error detected on AP-170P unit | warning |
Silence PCCO Ext Device Error | AP-170P hemodynamic unit system initialization failed | System initialization error on AP-170P unit | warning |
Silence PCCO No Catheter | PiCCO catheter not connected | PiCCO catheter not connected to JT-170P cord | warning |
Silence PCCO Parameter Duplicated | Multiple CCO devices connected | Two or more CCO devices detected | warning |
Silence PCCO ProAQT Sensor Error | ProAQT sensor expired | Expired ProAQT sensor in use | warning |
Silence Prolonged RR | Prolonged interval between consecutive beats detected | Beat-to-beat interval exceeds normal range | warning |
Silence rSO2 Parameter Duplicated | Multiple cerebral/somatic oximeters connected to bedside monitor | Two or more oximeters connected simultaneously | warning |
Silence ScvO2 Ambient Temp High | Ambient temperature too high for CeVOX operation | Ambient temperature exceeds operational limits | warning |
Silence ScvO2 Light Interference | Excessive ambient light detected | Ambient light exceeds safe level for measurement | warning |
Silence ScvO2 Module Error | CeVOX module internal error | CeVOX module hardware/software failure | warning |
Silence ScvO2 Overheat | CeVOX module overheated | Module internal temperature too high | warning |
Silence ScvO2 Signal Too High | ScvO2 probe signal too strong | Excessive ScvO2 signal detected | warning |
Silence ScvO2 Signal Too Low | ScvO2 probe signal too weak | Weak or missing ScvO2 signal | warning |
Silence ScvO2 Warming Up | CeVOX module warming up after power-on | CeVOX module initializing | info |
Silence StO2 Parameter Duplicated | Multiple cerebral oximeters connected to bedside monitor | Two or more cerebral oximeters connected simultaneously | warning |
Silence SV Tachy | Supraventricular tachycardia detected | Rapid supraventricular rhythm >100-150 bpm | warning |
Silence SVV:Highly Variable Pulse Rate | SVV measurement affected by high pulse fluctuations | Excessive pulse rate variation detected | warning |
Silence System Failure Settings | Abnormality detected in system settings configuration | System settings checksum or validation error during startup | error |
Silence Tachycardia | General tachycardia (heart rate >100 bpm) detected | Sustained heart rate above tachycardia threshold | warning |
Silence tcPO2/tcPCO2 Parameter Duplicated | Multiple transcutaneous monitors connected to bedside monitor | Two or more transcutaneous monitors connected simultaneously | warning |
Silence Telemetry not Connected | Connection between telemetry device and bedside monitor is interrupted | Signal loss or communication dropout during HiQ-View operation | warning |
Silence - TOF Communication Error | Communication error between NMT module and bedside monitor | Communication link between NMT and monitor fails or becomes unstable | warning |
Silence - TOF Low Temperature | Patient body temperature below 32°C (89.6°F); TOF measurement unreliable | Temperature sensor reading drops below 32°C (89.6°F) | warning |
Silence - TOF Module Failure | NMT module internal hardware error detected | NMT module internal fault condition | critical |
Silence TOF Parameter Duplicated | Two or more TOF monitoring devices are connected | Multiple TOF devices detected on single interface | warning |
Silence Trigeminy | Trigeminy pattern detected (every third beat premature) | Pattern of two normal beats followed by premature beat repeating | warning |
Silence V Brady | Ventricular bradycardia detected | Slow ventricular rhythm with wide QRS complexes at slow rate | warning |
Silence - Vent Parameter Duplicated | Ventilator communication connection already established; duplicate connection attempt | Multiple ventilator connections or cable already connected when another connection initiated | warning |
Silence VF | Ventricular fibrillation detected | Chaotic rapid ventricular electrical activity without organized complexes | critical |
Silence VPC | Ventricular premature complex detected | Single or occasional ventricular premature beats detected | warning |
Silence VPC Run | Run of ventricular premature complexes detected | Multiple consecutive premature ventricular contractions without intervening sinus beat | warning |
Silence V Rhythm | Abnormal ventricular rhythm detected | Sustained ventricular rhythm pattern detected | warning |
Silence VT | Ventricular tachycardia detected | Rapid sustained ventricular rhythm >100 bpm without P waves | critical |
Silence XXX (Parameter name) External Device Failure | External monitoring device connected to parameter is faulty | External device hardware malfunction detected | error |
Silence XXX (Parameter name) Module Failure | Parameter measurement module hardware is faulty | Module malfunction detected during measurement | error |
Silence Zeroing Complete | Zero adjustment successfully completed | Zero adjustment process completes | info |
— Tb Signal Unstable | Blood temperature signal unstable | Unstable Tb signal detected | info |
— Tb Small Temp Change | Injectate temperature change too small | Temperature change less than 0.1°C between injectate and blood | info |
— TD Measurement Limit | Maximum TD measurements reached | TD measurement storage full | info |
— TD Slow Injection | Injection speed too slow | Slow injection detected during TD measurement | info |
— TD Temp Sensor Error | Injectate temperature sensor error | Temperature sensor malfunction | error |
— TD Time Out | No injection detected within timeout period | Injection not detected in expected time window | info |
— Telemetry Alarm Inactivated | Alarms are silenced or suspended on vital sign telemeter during HiQ-View use | GZ-120P, GZ-130P, or GZ-140P telemeter has alarms silenced or suspended | info |
Termination Analog Check External Device | Analog external instrument disconnected from bedside monitor | Communication cable disconnection between analog instrument and bedside monitor | error |
Termination Analog Ext X Check Sensor | Analog external instrument sensor or connection issue | Sensor disconnection or error on analog external instrument | error |
Termination Anes Check External Device | Anesthetic machine disconnected from bedside monitor | Communication cable disconnection between anesthetic machine and bedside monitor | error |
Termination AP Check Sensor | AP sensor or catheter disconnected | ProAQT sensor or PiCCO sensor cable disconnection | error |
Termination Cannot Print | Network printer unable to accept print job | Printer busy, offline, network issue, or configuration error | error |
Termination CCO Check External Device | CCO monitor disconnected or communication error | CCO monitor not detected or communication failure | error |
Termination CCO Check Sensor | CCO sensor connection error or fault | CCO sensor disconnection, cable fault, or wrong sensor type | error |
Termination CCO Connector Off | APCO/IBP processor not properly connected to MULTI socket | Processor cord removed or disconnected from input unit MULTI socket | error |
Termination Data Send Error | Network data transmission from bedside monitor to central monitor failed | Network communication error during data transmission | error |
Termination EEG Check Device | EEG communication error or multi-link cable disconnected | Communication failure or cable disconnection | error |
Termination EEG Check External Device | EEG multilink cable communication error between AE-920P neuro unit and bedside monitor | Multilink cable disconnected or communication error occurred | error |
Termination EEG Check Scalp | EEG electrode has been in prolonged contact with patient skin; skin condition check required | Prolonged EEG measurement over long period | warning |
Termination EEG Connector Off | EEG connection cord removed or disconnected | EEG cord not connected to socket | error |
Termination EEG Connector Off | AE-920P neuro unit receiver disconnection or communication error | Receiver disconnected from AE-920P neuro unit or communication error occurred | error |
Termination EEG Head Set Failure | AE-120A EEG head set internal hardware failure detected | Abnormal current or voltage levels detected in head set electronics | error |
Termination End A-Fib | Atrial fibrillation has ended and delay duration exceeded | A-Fib pattern ceases and configured delay time period elapses | info |
Termination Flow/Paw Check Device | Flow/Paw measurement switch off or connection error | Measurement switch off, cable disconnected, or communication error | error |
Termination - Gas Device Failure | Gas module has encountered an internal hardware fault | Internal gas module error detected | critical |
Termination - Gas No Watertrap or Sample Line | Gas module unable to detect water trap or sample line | System detects missing or improperly connected water trap/sample line at startup or during operation | error |
Termination HiQ-View Connection Failure | HiQ-View could not restart after bedside monitor restart | HiQ-View restart failed after monitor reboot | error |
Termination Input Unit Disconnect | Input unit was disconnected without proper removal preparation | Input unit physically removed while removal preparation not completed | error |
Termination No Flow Sensor | Flow tube disconnected or faulty | Flow tube not detected or malfunction | error |
Termination - O2 Check Sensor | Oxygen sensor disconnected or damaged | FiO2 sensor connection lost or hardware failure detected | error |
Termination - O2 Connector Off | FiO2 connection cord not connected to MULTI socket of input unit | FiO2 cord removed, disconnected, or loose in MULTI socket | error |
Termination PCCO Check External Device | PiCCO monitor disconnected or communication error | PiCCO monitor not detected or communication failure | error |
Termination rSO2 Check External Device | Cerebral/somatic oximeter disconnected from bedside monitor | Communication cable disconnection between oximeter and bedside monitor | error |
Termination ScvO2 Check Module | CeVOX module disconnected | CeVOX module not connected to AP-170P unit | error |
Termination StO2 Check External Device | Cerebral oximeter disconnected from bedside monitor | Communication cable disconnection between cerebral oximeter and bedside monitor | error |
Termination tcPO2/tcPCO2 Check External Device | Transcutaneous monitor disconnected from bedside monitor | Communication cable disconnection between transcutaneous monitor and bedside monitor | error |
Termination - TOF Cal Error1 | NMT module calibration procedure failed | Automatic NMT calibration unable to complete successfully | error |
Termination TOF Check External Device | TOF monitoring device is disconnected or communication error | TOF interface disconnection or communication error detected | error |
Termination - TOF Check Sensor | NMT sensor connections incomplete or loose | NMT module unable to detect proper sensor connections at startup or during operation | error |
Termination - TOF Check Temp Sensor | NMT temperature sensor not connected to module | Temperature sensor connection lost or not established | error |
Termination - TOF Not Calibrated | NMT module requires initial calibration before use | NMT module connected without prior calibration or reset | error |
Termination - TOF Pre PTC Response Detected | Vital patient response detected during TOF pre-PTC calibration phase | Patient muscle response detected during pre-PTC stimulus | warning |
Termination - TOF Signal Too High1 | Acceleration signal during TOF calibration exceeds upper limit | Measured acceleration signal during calibration exceeds maximum threshold | error |
Termination - TOF Signal Too Low1 | Acceleration signal during TOF calibration below lower limit | Measured acceleration signal during calibration falls below minimum threshold | error |
Termination - TOF Signal Unstable1 | Acceleration signal during TOF calibration fluctuating or unstable | Acceleration signal during calibration shows instability or excessive noise | error |
Termination - TOF Supramaximal Not Found1 | TOF supramaximal stimulation current could not be determined during calibration | NMT module unable to identify supramaximal stimulation level | error |
Termination - TOF Temp Out Of Range | Temperature sensor reading outside valid measurement range 20.0–42.5°C (68.0–108.5°F) | Temperature reading exceeds or falls below sensor measurement limits | warning |
Termination - Vent Check External Device | Ventilator communication interface lost or disconnected | Ventilator interface or cable disconnected from bedside monitor | error |
— Ti Injectate Temp High | Injectate temperature warmer than blood | Injectate temperature exceeds blood temperature | info |
— - TOF Please Wait | Insufficient interval since previous TOF stimulus operation | TET or PTC stimulus applied within 2 minutes of prior measurement | info |
XXX Check Device | Communication interrupted with connected parameter device; communication error occurred; connection between unit and monitor disconnected | Parameter device communication timeout; cable connection lost; communication protocol error | warning |
XXX Check External Device | Communication interrupted with external instrument; communication error; connection between external device and monitor disconnected | External device communication timeout; cable disconnected; communication protocol error | warning |
XXX Device Failure | Connected parameter unit or external device is faulty or not functioning | Parameter device fails self-test or communication lost after connection confirmed | error |
XXX(Press) Catheter Disconnect | Arterial catheter disconnected from IBP line | IBP pulse absent and mean pressure ≤10 mmHg | critical |
XXX(Press) Check Label | Duplicate or unrecognized IBP channel label | Two or more IBP cords with same label or unrecognized label | critical |
XXX(Press) Check Sensor | IBP transducer disconnected or faulty; IBP cord damaged | Transducer disconnect; cord damage; sensor failure | error |
XXX(Press) Connector Off | IBP connection cord removed from MULTI socket | Cord disconnected from monitor input | error |
XXX(Press) Not Zero | IBP zero balance not adjusted | Measurement attempted without zero balance calibration | warning |
XXX(Press) Out Of Range | IBP pressure measurement outside monitor range | Pressure reading exceeds monitor limits | warning |
XXX(Press) Parameter Duplicated | More than eight IBP cords connected | Attempt to connect >8 IBP channels | error |
XXX(Press) Zeroing | Zero balance adjustment in progress | Automatic zero balance procedure active | info |
XXX(Press) Zeroing Complete | Zero balance adjustment successfully completed | Zero balance procedure finished | info |
XXX(Press) Zero Out of Range | Zero balance value out of acceptable range | Zero point outside normal limits | error |
XXX(Press) Zero Unstable | Zero balance adjustment unstable or circuit not exposed to air | Zero balance procedure encountered instability | warning |
XXX(Temp) Check Label | Duplicate or unrecognized temperature probe label | Two or more probes with same label or unrecognized label | error |
XXX(Temp) Check Sensor | Temperature probe disconnected or cord damaged | Probe disconnect; cord damage detected | error |
XXX(Temp) Connector Off | Temperature cord disconnected from MULTI socket | Temperature cord removed from monitor | error |
XXX(Temp) Parameter Duplicated | More than eight temperature cords connected | Attempt to connect >8 temperature channels | error |
— XXX (Unit name) Power Failure | Power output abnormality in specified unit | Power supply output voltage out of specification | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI