Nihon Kohden BSM-6000: troubleshooting and error codes

The ServiceForge knowledge base holds 305 documented symptoms for the Nihon Kohden BSM-6000, with 483 ranked causes, 245 diagnostic checks with test points and expected values, and 229 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

ECG noise in signal patient safety

High-amplitude noise corrupting ECG waveform; EMG muscle noise, motion artifact, or interference

Ranked causes

  1. 1. Baseline instability from respiration or body movement (common)
  2. 2. EMG muscle noise at electrode site (common)
  3. 3. Electrode pulled by lead tension (occasional)
  4. 4. Dry or degraded electrode (common)
  5. 5. Poor contact between lead and electrode (occasional)
  6. 6. High electrode impedance (occasional)
  7. 7. Electric blanket in use (rare)
  8. 8. Strong electromagnetic interference nearby (occasional)
  9. 9. Equipotential grounding not acquired (occasional)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electromagnetic interference affecting monitoring patient safety

Artifact appears on display, incorrect readings, or system malfunction due to external EM sources

Ranked causes

  1. 1. Strong EM interference from nearby emitter (radio station, cellular phone) (common)
  2. 2. RF interference through AC power supply (occasional)
  3. 3. Electrostatic discharge effects (occasional)
  4. 4. Equipment interferes with radio/TV receiver (occasional)
  5. 5. Lightning-induced voltage surge (rare)
  6. 6. Incompatible accessory, transducer, or cable connected (occasional)
  7. 7. Non-standard system configuration used (occasional)
  8. 8. Excessive measurement sensitivity setting (common)
  9. 9. Radiation therapy room environment interference (rare)
First check

Identify nearby EM emitters (radio, cellular phone, etc.)

Test point
Equipment location
Expected
No strong emitters within operating range

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG waveform does not appear on screen

No ECG waveform displayed; 'ECG CHECK ELECTRODES' message may appear

Ranked causes

  1. 1. Electrode peels or electrode gel is dry (common)
  2. 2. High contact impedance between electrode and skin (common)
  3. 3. Instrument not properly grounded (occasional)
  4. 4. ECG connection cord not firmly connected (common)
  5. 5. ECG connection cord failure (occasional)
  6. 6. Number of electrodes set incorrectly (occasional)
  7. 7. Input unit failure (rare)
First check

Inspect electrodes for peeling or dry gel

Test point
Patient electrode pads
Expected
Electrodes intact with wet gel

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Zeroing (zero adjustment) incomplete

IBP zero adjustment procedure does not complete successfully

Ranked causes

  1. 1. Noise (MPU MODULE ERROR displayed) (occasional)
  2. 2. IBP connection cord or blood pressure transducer failure (PRESS CHECK SENSOR displayed) (occasional)
  3. 3. Pressure less than -200 mmHg or more than +200 mmHg applied to transducer (occasional)
  4. 4. Transducer failure (occasional)
  5. 5. Some pressure remains (PRESS ZERO UNSTABLE displayed) (occasional)
  6. 6. External pressure or vibration applied to blood circuit (occasional)
  7. 7. MPU board or EXTMPU board failure (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power lamp does not light high voltage

Instrument does not turn on in AC power operation and AC power lamp is off

Ranked causes

  1. 1. Power cord failure (common)
  2. 2. Poor contact at cable between AC inlet and power supply unit (common)
  3. 3. Poor contact at cable between power supply unit and MAIN DIGITAL board (common)
  4. 4. Poor contact at cable between MAIN DIGITAL and FRONT IF board (common)
  5. 5. Poor contact at cable between FRONT IF and Power SW board (common)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 measurement out of range patient safety

CO2 calculated value outside measuring range; catheter retaining condition or coefficient value incorrect

Ranked causes

  1. 1. Injectate temperature not specified by catheter manufacturer (common)
  2. 2. Blood temperature out of range (occasional)
  3. 3. Catheter faulty (occasional)
  4. 4. Returned to thermodilution baseline too late (rare)
  5. 5. Injectate not cooled; room temperature above 27°C (occasional)
First check

Measure blood temperature

Test point
Temperature probe
Expected
15°C to 45°C, +15 to +45°C

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG electrode detached or lead problem patient safety

ECG electrode lead disconnected or detached from electrode; poor contact between lead and electrode

Ranked causes

  1. 1. Operator disconnected ECG lead or cord (common)
  2. 2. Electrode lead detached from electrode (common)
  3. 3. Electrode cannot attach firmly to skin (occasional)
  4. 4. Electrode lead disconnected from connection cord (common)
  5. 5. Poor contact between lead and electrode (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heart rate not counted correctly and ECG noise displayed

Heart rate detection inaccurate; 'ECG NOISE' message appears on display

Ranked causes

  1. 1. Electrode peels or electrode gel is dry (common)
  2. 2. Baseline drift or EMG due to body movement (common)
  3. 3. High contact impedance between electrode and skin (common)
  4. 4. Instrument not properly grounded (occasional)
  5. 5. Input unit failure (rare)
First check

Inspect electrodes for peeling or dry gel

Test point
Patient electrode pads
Expected
Electrodes intact with wet gel

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP MEASURING message and automatic remeasurement

Patient movement or weak pulse wave detected during active NIBP measurement, automatic remeasurement initiated

Ranked causes

  1. 1. Patient moved during measurement (common)
  2. 2. Weak patient pulse wave (occasional)
  3. 3. Cuff not properly attached (common)
  4. 4. Incorrect cuff size for patient (common)
  5. 5. Patient unstable pulse and heart rate (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2-2 DETECTING PULSE - Pulse detection in progress (NK/Masimo) patient safety

SpO2-2 sensor actively searching for and detecting pulse waveform

Ranked causes

  1. 1. Normal pulse detection process in progress (common)
  2. 2. Poor blood circulation at site (common)
  3. 3. Probe not attached at appropriate site (occasional)
  4. 4. Probe secured too tightly (occasional)
  5. 5. Probe disconnected from adapter (occasional)
First check

Monitor detection progress

Test point
SpO2 waveform display
Expected
Pulse waveform eventually detected

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

TEMP CHECK patient safety

Temperature probe or connection cord disconnected or damaged

Ranked causes

  1. 1. Probe disconnected from TEMP socket or cord disconnected from MULTI socket (common)
  2. 2. Probe or temperature connection cord damaged (occasional)
  3. 3. Measured value outside measuring range (occasional)
  4. 4. Temperature connection cord disconnected (occasional)
  5. 5. Temperature connection cord damaged (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCO CHECK SENSOR disconnected

FloTrac sensor disconnected from APCO/IBP processor or specified FloTrac sensor not being used.

Ranked causes

  1. 1. FloTrac sensor disconnected from APCO/IBP processor (common)
  2. 2. Wrong FloTrac sensor model used (occasional)
  3. 3. Cable or sensor damaged (occasional)
  4. 4. APCO/IBP processor internal failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCO CHECK WAVEFORM unstable

Arterial waveform quality insufficient for accurate CO measurement or transmission.

Ranked causes

  1. 1. Arterial waveform inadequate for CO measurement (common)
  2. 2. Waveform inadequate for extended measurement period (occasional)
  3. 3. IBP line improper setup (occasional)
  4. 4. Arterial pressure outside normal range (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Baseline drift or waveform synchronized with heartbeat in respiration

Baseline drift or heartbeat-synchronized waveform mixed in respiration measurement

Ranked causes

  1. 1. Body movement artifact in respiration signal (common)
  2. 2. Respiration noise not filtered (occasional)
  3. 3. Input unit failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

BIS SENSOR disconnected patient safety

BIS sensor disconnected from patient interface cable or sensor attachment failed.

Ranked causes

  1. 1. Patient interface cable disconnected from BIS/BISx processor (common)
  2. 2. BIS sensor disconnected from patient interface cable (common)
  3. 3. BIS sensor impedance too high (occasional)
First check

Verify patient interface cable connection

Test point
BIS/BISx processor interface
Expected
Cable firmly seated with click

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCO UNSTABLE measurement

Arterial waveform quality or IBP line conditions unstable for reliable CO measurement.

Ranked causes

  1. 1. Arterial waveform unstable (common)
  2. 2. IBP line not appropriate (occasional)
  3. 3. Patient pulse getting low (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication loss system network caution

System network cable disconnected or communication link failure between monitor and hub/router

Ranked causes

  1. 1. Network cable disconnected (common)
  2. 2. Cable or connector faulty (occasional)
  3. 3. Hub or router defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG electrode configuration incorrect patient safety

NUMBER OF ELECTRODES setting on ECG window not correct; differential offset voltage at electrodes

Ranked causes

  1. 1. Incorrect electrode count setting (common)
  2. 2. Electrode lead damaged (occasional)
  3. 3. Differential offset voltage at electrodes (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Frequent pacing marks appear on ECG

Pacing marks appear frequently on ECG waveform display

Ranked causes

  1. 1. Instrument not properly grounded (common)
  2. 2. Electric blanket not shielded (occasional)
  3. 3. Input unit failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

GAS CHECK WATERTRAP message displayed

Water trap not properly attached or clogged, or sampling line not properly connected

Ranked causes

  1. 1. Water trap not securely attached (common)
  2. 2. Sampling line not properly connected (common)
  3. 3. Water trap clogged (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

GAS LINEBLOCK re-alarm message

Sampling or exhaust gas line blocked, or water trap clogged or sample line blocked

Ranked causes

  1. 1. Water trap clogged (common)
  2. 2. Sampling line clogged or bent (common)
  3. 3. Exhaust gas tube clogged or bent (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

GAS MIXED GAS silence alarm patient safety

Two anesthetic agents detected simultaneously, typically during vaporizer change or dual vaporizer operation

Ranked causes

  1. 1. Vaporizer being changed with dual agents in circuit (common)
  2. 2. Two vaporizers operating simultaneously (occasional)
  3. 3. Wrong agent delivered to vaporizer (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

IBP DISCONNECT SENSOR - Blood pressure transducer disconnected patient safety

Invasive blood pressure monitoring disconnected or not receiving pulse and mean pressure

Ranked causes

  1. 1. Blood pressure transducer not properly connected (common)
  2. 2. Blood pressure transducer damaged (occasional)
  3. 3. IBP connection cord is damaged (occasional)
First check

Verify IBP transducer connection

Test point
IBP connector on monitor
Expected
Firmly connected and seated

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP measurement shows MEAS TIMEOUT patient safety

NIBP measurement fails after remeasurement, displays 'MEAS TIMEOUT' message

Ranked causes

  1. 1. Cuff not properly attached to patient (common)
  2. 2. Vibration or external pressure on cuff or air hose (common)
  3. 3. Failure in oscillation signal amplification circuit of MPU board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nothing displayed on external display

External display shows no output or is blank

Ranked causes

  1. 1. Poor contact or break in display cable when interface displayed in OPTIONS (common)
  2. 2. Display, interface, or MAIN DIGITAL board failure (occasional)
  3. 3. Interface or MAIN DIGITAL board failure when not in OPTIONS (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PRINT ERROR

Network printer communication or paper loading failure

Ranked causes

  1. 1. Too many print commands sent to printer (occasional)
  2. 2. Incorrect printer settings (common)
  3. 3. Recording paper not properly loaded (common)
First check

Wait for print queue to clear

Test point
Printer status
Expected
No pending print jobs

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 DETECTING PULSE - Cannot find pulse waveform yet patient safety

SpO2 sensor searching for pulse waveform but not yet detected

Ranked causes

  1. 1. System searching for pulse waveform detection (common)
  2. 2. Poor blood circulation (common)
  3. 3. Probe attached too tightly (occasional)
First check

Wait for initial pulse detection

Test point
SpO2 waveform display
Expected
Pulse waveform appears within normal time

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 LIGHT INTERFERENCE - Excessive light at probe site (NK/Masimo)

SpO2 sensor detecting too much ambient light affecting measurement

Ranked causes

  1. 1. Excessive ambient light at probe site (common)
  2. 2. Interference from surroundings (occasional)
  3. 3. Probe attachment site not appropriate (occasional)
First check

Assess ambient light level at probe site

Test point
Probe area illumination
Expected
Minimal direct or ambient light

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 LOW QUALITY SIGNAL - Poor signal quality or movement artifact (NK/Nellcor/Masimo) patient safety

SpO2 measuring unstable waveform due to motion or low perfusion

Ranked causes

  1. 1. Considerable body movement (common)
  2. 2. Probe not properly attached (common)
  3. 3. Low signal IQ (occasional)
First check

Assess patient motion

Test point
Patient extremity movement
Expected
Minimal involuntary movement

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 WEAK PULSE - Poor peripheral circulation or low perfusion (NK/Masimo) patient safety

SpO2 sensor detecting weak pulse signal due to poor circulation or tight probe

Ranked causes

  1. 1. Poor peripheral circulation (common)
  2. 2. Probe attached too tightly (occasional)
  3. 3. Low perfusion index (occasional)
First check

Assess patient peripheral perfusion

Test point
Patient extremity color, warmth, capillary refill
Expected
Good perfusion with normal parameters

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2-2 ALARM - Pulse oximeter alarm (Nellcor/Masimo) patient safety

Alarm condition on connected pulse oximeter (SpO2-2 channel)

Ranked causes

  1. 1. Alarm on pulse oximeter device (common)
  2. 2. Poor blood circulation at measurement site (common)
  3. 3. Probe attached too tightly (occasional)
First check

Review pulse oximeter alarm indication

Test point
SpO2 device display
Expected
Identify specific alarm condition

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2-2 CHANGE PROBE - Probe deteriorated or damaged (NK/Nellcor/Masimo) patient safety

SpO2-2 probe has deteriorated, damaged, or short-circuited and requires replacement

Ranked causes

  1. 1. Probe deteriorated from age or use (common)
  2. 2. Probe damaged or short-circuited (occasional)
  3. 3. SpO2 adapter is damaged (occasional)
First check

Inspect probe for visible deterioration

Test point
Probe body and connector
Expected
No discoloration, cracks, or wear

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2-2 DETECT PULSE - Cannot detect pulse (Masimo) patient safety

Masimo SpO2-2 module cannot detect pulse signal

Ranked causes

  1. 1. Pulse not detected by SpO2-2 sensor (common)
  2. 2. Poor blood circulation (common)
  3. 3. Probe attached too tightly (occasional)
First check

Assess patient perfusion status

Test point
Patient extremity
Expected
Good color and warmth indicating adequate circulation

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2-2 DETECT PULSE - Cannot detect pulse (NK) patient safety

NK (Nihon Kohden) SpO2-2 module cannot detect pulse signal

Ranked causes

  1. 1. Pulse not detected by SpO2-2 sensor (common)
  2. 2. Poor blood circulation (common)
  3. 3. Probe attached too tightly (occasional)
First check

Assess patient perfusion status

Test point
Patient extremity
Expected
Good color and warmth indicating adequate circulation

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2-2 LIGHT INTERFERENCE - Excessive light interference (NK/Masimo)

SpO2-2 module detecting excessive light interference at probe site

Ranked causes

  1. 1. Excessive ambient light at probe site (common)
  2. 2. Interference from surroundings (occasional)
  3. 3. Probe attachment site not appropriate (occasional)
First check

Assess light level at probe

Test point
Probe area
Expected
Minimal ambient light exposure

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2-2 WEAK PULSE - Weak pulse signal (NK/Masimo) patient safety

SpO2-2 weak pulse detection from poor circulation or probe tension issue

Ranked causes

  1. 1. Poor peripheral circulation (common)
  2. 2. Probe attached too tightly (occasional)
  3. 3. Low perfusion status (occasional)
First check

Evaluate patient hemodynamic status

Test point
Patient vital signs and appearance
Expected
Adequate blood pressure and perfusion

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

TOF CHECK SENSOR patient safety

Main cable, stimulation cable, electrodes disconnected or abnormal stimulation current

Ranked causes

  1. 1. Cables or electrodes not connected (common)
  2. 2. Twitch in patient too strong (rare)
  3. 3. Abnormal electric stimulation current detected (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch key response does not match display position

Pressing a key position on screen does not activate the intended function; mark appears elsewhere

Ranked causes

  1. 1. Edge protection rubber not properly attached (common)
  2. 2. Touch screen failure (occasional)
  3. 3. MAIN DIGITAL board failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not turn on in battery operation caution

Instrument fails to power on when operating from battery

Ranked causes

  1. 1. Poor contact at connectors between BATT MOTHER0 and MAIN DIGITAL boards (common)
  2. 2. Poor contact at connectors between BATT MOTHER1 and MAIN DIGITAL boards (common)
  3. 3. MAIN DIGITAL, BATT MOTHER0 or BATT MOTHER1 board failure (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm information not output from bedside monitor to nurse call system patient safety

Alarm signals from bedside monitor are not transmitted to nurse call system

Ranked causes

  1. 1. Short circuit in YJ-672P nurse call cable when interface displayed and signals not output (occasional)
  2. 2. Interface or MAIN DIGITAL board failure when not in OPTIONS (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot calibrate FiO2, O2CAL?? displayed

FiO2 calibration fails, message 'O2CAL??' shown on screen

Ranked causes

  1. 1. O2 sensor reached end of life (common)
  2. 2. MPU board or EXTMPU board failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot communicate with external equipment through RS-232C port

Communication failure through RS-232C port on QI-631P or QI-671P interface

Ranked causes

  1. 1. Wrong baud rate setting or connection cable failure when interface displayed (common)
  2. 2. Interface or MAIN DIGITAL board failure when not displayed in OPTIONS (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot communicate with network

Network communication fails; no response when connected to PC with cross cable and ping test

Ranked causes

  1. 1. MAIN DIGITAL board failure when ping response not returned (occasional)
  2. 2. Hub and cables have problems when ping response returned (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot communicate with optional unit through multilink port

Optional unit (QI-631P, QI-632P, QI-634P, QI-671P, QI-672P) not recognized; no error displayed

Ranked causes

  1. 1. Interface or MAIN DIGITAL board failure when not displayed in OPTIONS (occasional)
  2. 2. Optional unit or interface failure when port does not display model number (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure cardiac output (CO), temperatures, MPU MODULE ERROR displayed

CO, Tb and Ti temperatures cannot be measured, error message 'MPU MODULE ERROR?' shown

Ranked causes

  1. 1. Noise such as static electricity in CO circuit (common)
  2. 2. Input unit failure (occasional)
First check

Measure CO after disconnecting connection cord and reattaching input unit

Test point
CO connection cord
Expected
CO can be measured

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure CO and temperatures, no error message displayed

CO, Tb and Ti cannot be measured, no error message shown on screen

Ranked causes

  1. 1. MPU board or EXTMPU board failure (occasional)
  2. 2. CO connection cord or balloon catheter failure (occasional)
First check

Connect CO connection cord and balloon catheter to another MP socket

Test point
Alternative MP sockets
Expected
CO can be measured

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure CO, BASELINE DRIFT displayed

CO measurement fails, error message 'CO BASELINE DRIFT' shown on screen

Ranked causes

  1. 1. MPU board or EXTMPU board failure (occasional)
  2. 2. Physiological fluctuation or external noise interferes with baseline (occasional)
First check

Connect AX-410G to CO connector at MP socket

Test point
CO connector of AX-410G at MP socket
Expected
Problem disappears or remains

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure ECG and module error displayed

ECG measurement fails with 'ECG MODULE ERROR?' message on screen

Ranked causes

  1. 1. Noise such as static electricity (common)
  2. 2. Input unit failure (occasional)
First check

Disconnect ECG connection cord and remove then reattach input unit

Test point
ECG input
Expected
ECG can be measured

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure FiO2, MPU MODULE ERROR displayed

FiO2 concentration cannot be measured, error message 'MPU MODULE ERROR?' appears

Ranked causes

  1. 1. Noise such as static electricity in FiO2 circuit (common)
  2. 2. FiO2 input unit failure (occasional)
First check

Measure FiO2 after disconnecting connection cord and reattaching input unit

Test point
FiO2 connection cord
Expected
FiO2 can be measured

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure FiO2, no error message, no FiO2 value displayed

No FiO2 measurement, no error message shown on screen

Ranked causes

  1. 1. MPU board or EXTMPU board failure (occasional)
  2. 2. FiO2 connection cord or O2 sensor failure (occasional)
First check

Connect FiO2 connection cord and O2 sensor to another MP socket

Test point
Alternative MP sockets
Expected
FiO2 can be measured

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure IBP and module error displayed

Invasive blood pressure measurement fails with 'MPU MODULE ERROR?' message on screen

Ranked causes

  1. 1. Noise such as static electricity (common)
  2. 2. Input unit failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure IBP; no error message and no pressure value displayed

IBP measurement fails with no error message; blood pressure value not shown

Ranked causes

  1. 1. MPU board or EXTMPU board failure (common)
  2. 2. IBP connection cord or blood pressure transducer failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure respiration (thermistor), MPU MODULE ERROR displayed

Temperature cannot be measured, error message 'MPU MODULE ERROR?' appears on screen

Ranked causes

  1. 1. Noise such as static electricity in respiration pickup circuit (common)
  2. 2. Input unit failure (occasional)
First check

Measure respiration after disconnecting MP connector and reattaching input unit

Test point
MP connector of respiration pickup
Expected
Respiration can be measured

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure respiration, no error message, no respiration rate displayed

No temperature measurement, no error message on screen, respiration rate not shown

Ranked causes

  1. 1. MPU board or EXTMPU board failure (occasional)
  2. 2. Respiration pickup failure (occasional)
First check

Connect respiration pickup to another MP socket

Test point
Alternative MP sockets
Expected
Respiration can be measured

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure SpO2, WEAK PULSE or CANNOT DETECT PULSE message displayed

SpO2 measurement fails, message 'SpO2 WEAK PULSE' or 'SpO2 CANNOT DETECT PULSE?' shown

Ranked causes

  1. 1. SpO2 board or MPUPWR board failure (occasional)
  2. 2. Instrument and accessories functioning normally (unknown)
First check

Replace SpO2 connection cord and probe, then connect AX-410G to SpO2 socket

Test point
SpO2 socket
Expected
Messages disappear when AX-410G is connected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot measure temperature

Temperature measurement fails; temperature value not displayed

Ranked causes

  1. 1. Temperature probe failure (common)
  2. 2. Connector board or Analog board failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CATHETER not connected properly patient safety

Catheter connection error for hemodynamic monitoring (PCCO/ProAQT/CCO)

Ranked causes

  1. 1. JT-170P CO connection cord improperly connected (common)
  2. 2. ProAQT sensor expired (common)
First check

Verify ProAQT sensor connection

Test point
Connection cord
Expected
Properly connected to monitor

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCO AP SENSOR ERROR

ProAQT sensor not connected properly or sensor is faulty.

Ranked causes

  1. 1. ProAQT sensor not connected to sensor cable (common)
  2. 2. ProAQT sensor faulty (occasional)
First check

Verify ProAQT sensor connection

Test point
ProAQT sensor cable connector
Expected
Secure connection with click

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCO APCHECK SENSOR disconnected

ProAQT sensor or PiCCO catheter and injectate sensor disconnected from monitor.

Ranked causes

  1. 1. ProAQT sensor disconnected from sensor cable (common)
  2. 2. PiCCO catheter and/or injectate sensor disconnected (common)
First check

Verify ProAQT sensor connection

Test point
ProAQT sensor cable
Expected
Sensor firmly connected to cable

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CCO CHECK PATIENT INFORMATION missing patient safety

Patient demographic data (gender, age, BSA) not entered or out of valid range for CCO calculation.

Ranked causes

  1. 1. Patient gender, age, or BSA not entered (common)
  2. 2. Patient information out of range (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • CCO CONNECTOR OFF disconnected
  • CCO ZERO UNSTABLE
  • CO CHECK SENSOR disconnected
  • CO2 external device disconnected or interface error
  • EEG external device failure
  • FiO2 value is randomly unstable
  • GAS CAL ERROR message displayed
  • GAS CHECK EXTERNAL DEVICE Termination message
  • GAS OUT OF RANGE silence alarm
  • GAS UNSPECIFIED ACCURACY on AG-920R
  • IBP PRESS OUT OF RANGE
  • IBP PRESS ZERO OUT OF RANGE
  • IRREGULAR RR MONITOR OFF message
  • NIBP AIR HOSE LEAK Silence alarm
  • NIBP CANNOT DETECT PULSE Silence alarm
  • NIBP CONNECTOR OFF Termination message
  • NIBP display appears, cuff inflates, pressure does not reach target
  • NIBP display area does not appear after air hose connected
  • NIBP MEAS TIMEOUT Silence alarm
  • NIBP SAFETY CIRCUIT RUNNING Re-alarm message
  • NIBP SYS OUT OF RANGE Silence alarm
  • Noise on blood pressure waveform
  • O2 CAL message displayed
  • O2 CONNECTOR OFF Termination message
  • O2 FiO2 connection error
  • Recorder shows INSERT REC PAPER message with paper properly set
  • RESP CHECK SENSOR - Respiration pickup damaged or disconnected
  • RESP CONNECTOR OFF - Respiration pickup disconnected (thermistor)
  • Respiration rate discrepancy from manual count
  • Respiration waveform does not appear; dashes displayed
  • SpO2 CHECK PROBE - Probe attachment and damage check (Nellcor/Masimo)
  • SpO2 CHECK PROBE - Probe not properly attached or damaged (NK)
  • SpO2 CONNECTOR OFF - Probe cable or adapter disconnected
  • SpO2-2 CONNECTOR OFF - SpO2-2 probe cable disconnected
  • SpO2-2 LOW QUALITY SIGNAL - Poor signal quality (NK/Masimo)
  • StO2 CHECK EXTERNAL DEVICE - Cerebral oximeter disconnected or communication error
  • StO2 CHECK EXTERNAL DEVICE - Cerebral/somatic oximeter disconnected
  • TB SIGNAL UNSTABLE
  • tcPO2/PCO2 CHECK
  • TEC DATA RECEIVE ERROR
  • Touch screen mark stays at particular pressed position
  • Touch screen mark stays out of acceptable range
  • Vital signs cannot be transmitted to ZS-900P receiver
  • A-FIB detected
  • Adjacent equipment causes malfunction
  • AG-400R CO2 unit appearance and physical check
  • Alarm indicator does not light
  • All ports display optional unit model numbers on MULTILINK CHECK
  • Appearance check failed
  • ASYSTOLE detected
  • Battery operation time shorter than expected
  • Battery operation with SB-671P pack
  • Battery weak
  • Bigeminy detected
  • BIS CHECK EXTERNAL DEVICE disconnected
  • BIS CHECKING IMPEDANCE
  • BIS MODULE FAILURE
  • BIS NOISE detected
  • BIS REPLACE SENSOR
  • BIS SENSOR ERROR
  • BIS SENSOR EXPIRED
  • Both display lamp and power lamp light
  • BYPASS activated
  • Calibration: AE-918P Neuro Unit ABR generator check
  • Calibration: BIS/BISx processor impedance check
  • Calibration: Cardiac output measurement
  • Calibration: CO2 gas accuracy check and sensitivity calibration
  • Calibration: CO2 partial pressure accuracy check
  • Calibration: Held pressure (cuff pressure maintenance)
  • Calibration: NIBP air circuit operation
  • Calibration: NIBP cuff pressure measurement
  • Calibration: NMT module electrical stimulation check
  • Calibration: Pressurization speed (adult/child mode)
  • Calibration: Pressurization speed (neonatal mode)
  • Calibration: Safety check measurements
  • Calibration: Safety circuit pressure release (adult/child mode)
  • Calibration: Safety circuit pressure release (neonatal mode)
  • Calibration: Solenoid valve operation (Step Deflation)
  • Calibration: Touch panel position incorrect
  • Cannot analyze ECG
  • Cannot measure FiO2, displays "---", O2 CHECK SENSOR message shown
  • Cannot measure IBP; dashes displayed and sensor check message
  • Cannot measure respiration, RESP CHECK SENSOR displayed
  • Cannot measure SpO2, CONNECTOR OFF or CHECK PROBE message displayed
  • Cannot measure Tb and Ti, displays "---", CO CHECK SENSOR message shown
  • Cannot recognize AY-600P series optional unit
  • CCO CHECK EXTERNAL DEVICE disconnected
  • CCO MODULE FAILURE
  • CCO ZERO IMBALANCE
  • CCO ZEROING COMPLETE
  • CHECK INPUT UNIT BATTERY
  • CO CHECK TEMPERATURE baseline
  • CO INJECT failure
  • CO INJECTION TIME OUT
  • CO2 absorbent needs replacement
  • CO2 adapter damaged
  • CO2 airway adapter disconnected
  • CO2 bath probe or inline sensor faulty
  • CO2 calibration failed
  • CO2 external device communication error
  • CO2 filterline not connected properly
  • CO2 sampling line blocked
  • CO2 sensor damaged
  • CO2 sensor insufficient light
  • CO2 sensor kit connector disconnected
  • CO2 thermodilution baseline error
  • CO2 unspecified operating environment
  • CO2 zero calibration not performed
  • Connector-off alarm verification
  • CPU RAM check failed
  • CPU ROM check failed
  • Display lamp does not light
  • Display lamp lights but power lamp does not light
  • ECG module malfunction
  • ECG pacing spike detected
  • ECG QRS amplitude too low
  • EEG connection cord disconnected
  • EEG connection cord re-seated during operation
  • EEG cord damaged
  • EEG cord disconnected from neuro unit
  • EEG differential offset voltage high
  • EEG electrode dry or not attached
  • EEG electrode lead poor contact
  • EEG module error
  • EEG neuro unit disconnected
  • Extended arrhythmia network communication lost
  • Flow tube disconnected from unit
  • Flow/Paw calibration error
  • Flow/Paw communication error with monitor
  • Flow/Paw measurement disabled in settings
  • Flow/Paw measurement out of range
  • FRAM check failed
  • GAS CAL COMPLETE message displayed
  • GAS CHECK SAMPLE LINE clogged
  • GAS DEVICE error - exhaust gas tube clogged
  • GAS MEASUREMENT setting not enabled
  • GAS OVERHEAT silence alarm on GF-210R/220R/310R/320R
  • GAS power indicator shows unit not powered
  • GAS PURGING in progress on AG-920R
  • GAS WARMING UP message displayed
  • Hard key function check failed
  • Heart rate not accurate
  • High frequency noise on ECG waveform with electrosurgical unit
  • High frequency noise on respiration waveform with electrosurgical unit
  • HIS SYNC ERROR Termination message
  • IBP CALIBRATING - Zero balance in progress
  • IBP LABEL ALREADY REGISTERED
  • IBP PRESS ZERO UNSTABLE
  • IMPEDANCE MEASUREMENT RESP OFF - Impedance respiration monitoring disabled
  • Impedance respiration waveform does not appear; RESP OFF message
  • INPUT UNIT DISCONNECT Termination message
  • Input unit not connected to main unit
  • MPU MODULE ERROR TEMP Silence alarm
  • MULTILINK CONFIG ERROR message displayed
  • MULTILINK CONFIG ERROR Re-alarm message
  • MULTILINK POWER ERROR message displayed
  • MULTILINK POWER ERROR Re-alarm message
  • NIBP CHECK INTERVAL SETTING Termination message
  • NIBP CUFF OCCLUSION Re-alarm message
  • NIBP DIA OUT OF RANGE Silence alarm
  • NIBP display appears, cuff does not inflate
  • NIBP HIGH CUFF PRESS Silence alarm
  • NIBP INFLATION during measurement
  • NIBP INFLATION PRESS LOW
  • NIBP MODULE ERROR Silence alarm
  • NIBP OUT OF RANGE Silence alarm
  • NIBP PLEASE WAIT message displayed
  • NIBP PROLONGED RR or Arrhythmia detected, uncalibrated
  • NO PACER PULSE ECG alarm
  • NO PCCO-AP Hemodynamic measurement error
  • No ports display optional unit model numbers on MULTILINK CHECK
  • No sound or small volume
  • Pacemaker interaction with cardiac monitoring
  • Page-dependent recording does not start at page top
  • PCCO CHECK EXTERNAL DEVICE Termination message
  • PCCO CHECK MODE Silence alarm
  • PM FAILURE Re-alarm message
  • Power cord safety check failed
  • QF/IF series interface connection status check
  • QI-670P interface connection and version check
  • RECEIVING DATA - Receiving data from defibrillator
  • Recorder door shows CLOSE PAPER MAGAZINE message when firmly closed
  • Remote controller key check failed
  • SAMPLE LINE pressurized on AG-920R
  • ScvO2 AMBIENT TEMP HIGH - CeVOX detects high ambient temperature
  • ScvO2 CAL ERROR - CeVOX calibration error
  • ScvO2 CONNECTOR OFF - CeVOX module not connected
  • ScvO2 LIGHT INTERFERENCE - CeVOX detects excessive ambient light
  • ScvO2 MODULE ERROR - CeVOX module internal error
  • ScvO2 OVERHEAT - CeVOX module overheated
  • ScvO2 SIGNAL TOO HIGH - CeVOX probe signal excessive
  • ScvO2 SIGNAL TOO LOW - CeVOX probe signal weak
  • ScvO2 WARMING UP - CeVOX module warming up
  • SENDING DATA - Input unit data transmission to central monitor
  • SENSOR ERROR COMPLETE - ProAQT sensor issue
  • SIMULATED DATA - Display showing simulated/demo data
  • Sound is distorted
  • Sound output check failed
  • SpO2 MODULE ERROR - SpO2 hardware malfunction (NK/Nellcor/Masimo)
  • SpO2 NO PROBE - SpO2 probe not connected to connection cord (NK)
  • SpO2 PROBE - Probe identification or integrity error (Masimo)
  • SpO2-2 CHECK PROBE - Probe attachment issue (NK)
  • SpO2-2 MODULE ERROR - SpO2-2 hardware malfunction (NK/Masimo)
  • SpO2-2 NO PROBE - SpO2-2 probe not connected to cord (NK)
  • SUSPEND MONITORING
  • Tb temperature outside measurable range
  • tcPO2/PCO2 ALARM
  • TD MEASUREMENT LIMIT
  • TD TEMP SENSOR ERROR
  • TD TIME OUT
  • TEC INTERFACE ALREADY CONNECTED
  • TEC INTERFACE DISCONNECTED
  • TEC INTERFACE ERROR
  • TEMP THIS LABEL IS ALREADY REGISTERED
  • THIS PARAMETER IS ALREADY REGISTERED
  • TI INJECTATE TEMP HIGH
  • Ti temperature outside measurable range
  • TOF ALARM
  • TOF CAL??
  • TOF CALERROR
  • TOF CHECK EXTERNAL DEVICE
  • TOF CHECK TEMP SENSOR
  • TOF COMMUNICATION ERROR
  • TOF LOW TEMPERATURE
  • TOF PLEASE WAIT
  • TOF PRE PTC RESPONSE DETECTED
  • TOF SIGNAL OUT OF RANGE
  • TOF SIGNAL TOO WEAK
  • TOF SUPRAMAXIMAL NOT FOUND
  • TRANSMITTER CONNECTOR OFF
  • Unit does not turn on in AC power operation
  • UPDATING DATA
  • USB mouse does not work with interface not in OPTIONS
  • USB mouse does not work with optical mouse LED off
  • USB mouse does not work with optical mouse LED on
  • VARIABLE PULSE RATE
  • VENT ALARM
  • VENT CHECK EXTERNAL DEVICE
  • Water in CO2 respiration circuit
  • Waveforms and grid have faint parts
  • Waveforms and numeric data record when paper pulled out manually
  • YJ-672P nurse call cable connection check
  • ZEROING COMPLETE
  • ZS-900P transmitter operational status check
  • ZS-900P transmitter signal reception verification

245 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
A-FIBArrhythmia detected on ECGAbnormal heart rhythm detected during ECG analysiscritical
ALARM RESETALARM RESET function key was activatedALARM RESET function key pressed during alarm reset conditioninfo
ALARMS SILENCEDAlarm sounds have been silenced[Silence Alarms] key pressed or ALARM RESET function key touchedwarning
ALARMS SUSPENDEDAll alarms have been suspended[Silence Alarms] key pressed twice or SUSPEND ALARMS key pressedwarning
ALL ALARMS OFFAll alarm functions have been disabledALARMS OFF key pressed or BYPASS key activatedwarning
ARRHYTHMIA ANALYSIS OFFArrhythmia analysis function is disabledSYSTEM SETUP window set to OFF for arrhythmia analysiswarning
BATTERY WEAKBattery pack charge level critically low at approximately 25%Battery pack reaches 25% charge levelwarning
BIS CHECK EXTERNAL DEVICEBIS monitor is disconnected from the bedside monitorBIS interface cable disconnectederror
BIS CHECKING IMPEDANCEBIS sensor impedance check in progressImpedance verification routine running on sensorinfo
BIS CHECK SENSORBIS sensor connection error or impedance issuePatient interface cable disconnected or BIS sensor not attached properlyerror
BIS MODULE FAILUREBIS processor has encountered internal failureBIS processor self-test failed or internal communication errorcritical
BIS NOISENoise interference detected on BIS signalElectrical or motion noise on BIS sensor signalerror
BIS REPLACE SENSORBIS sensor has exceeded maximum number of usesSensor usage counter exceeds manufacturer limitwarning
BIS SENSOR ERRORBIS sensor failure detectedSensor communication failure or self-test failureerror
BIS SENSOR EXPIREDBIS sensor has passed its expiration dateSensor expiration date exceeded during startup or during usewarning
BISx CONNECTION CABLEBISx connection cable disconnected from monitor or damagedBISx cable removed from monitor or connection failure detectederror
BYPASSAlarm bypass mode activatedBYPASS key pressed during alarm conditionwarning
CANNOT ANALYZEECG arrhythmia analysis cannot be performedExcessive noise on ECG signal for more than 30 secondserror
CCO APCHECK SENSORProAQT sensor or PiCCO catheter cable disconnected from CO monitorSensor or cable removal from connection point detectederror
CCO AP SENSOR ERRORProAQT sensor connection error or sensor failureSensor not detected or communication failure with sensorerror
CCO CHECK EXTERNAL DEVICECCO monitor disconnected from bedside monitorCCO interface cable disconnectederror
CCO CHECK PATIENT INFORMATIONRequired patient demographic data missing or invalid for CCO calculationPatient gender, age, or BSA not entered or values out of acceptable rangeerror
CCO CHECK SENSORFloTrac sensor disconnected, wrong type, or cable/sensor damagedFloTrac sensor not detected or wrong sensor model selectederror
CCO CHECK WAVEFORMArterial waveform quality insufficient for reliable CO measurementArterial waveform signal degradation or IBP line issueerror
CCO CONNECTOR OFFAPCO/IBP processor is disconnected from monitorAPCO/IBP processor interface cable disconnectederror
CCO MODULE FAILUREAPCO/IBP processor is faulty and non-functionalProcessor self-test failed or internal error detectedcritical
CCO UNSTABLEArterial waveform too unstable for reliable CO measurementWaveform instability detected over timeerror
CCO ZERO IMBALANCEZero balance calibration not performed for APCO/IBP systemZero balance adjustment required but not completederror
CCO ZEROING COMPLETEZero balance adjustment successfully completedZero balance calibration routine finishedinfo
CCO ZERO UNSTABLEZero balance adjustment cannot complete due to circuit or pressure instabilityZero balance procedure initiated but cannot stabilizeerror
CHECK INPUT UNIT BATTERYBattery pack not inserted correctly in BSM-1700 series monitorBattery pack not properly seated in battery compartmenterror
CO CAL COMPLETE (Sidestream)CO sidestream calibration completed successfullyCalibration procedure finished without errorsinfo
CO CAL ERROR (Sidestream)CO sidestream calibration failedCalibration procedure did not complete successfullyerror
CO CALIBRATING (Mainstream TG-950P/970P/980P)CO mainstream module in calibration; measuring reference light amountCalibration procedure in progressinfo
CO CALIBRATING (Sidestream)CO sidestream module in calibration procedureCalibration procedure in progressinfo
CO CAL? (Mainstream TG-950P/970P/980P)CO zero calibration not performed on mainstream moduleSystem initialized without zero calibration on startuperror
CO CELL OFF (TG-970P/980P)CO sensor airway adapter disconnected or CO2 cell offlineAirway adapter unplugged from sensorwarning
CO CHANGE ABSORBENT (Sidestream)CO2 sidestream absorbent cartridge depleted and requires replacementAbsorbent material exhausted after extended usewarning
CO CHANGE ADAPTER (Mainstream)CO mainstream adapter damaged or needs replacementAdapter mechanical damage detected or functionality degradedwarning
CO CHECK CO2 CELL (TG-950P/970P/980P)Water collected in respiration circuit; unable to measure CO2Water accumulation detected in breathing circuiterror
CO CHECK EXTERNAL DEVICE (Microcap/MicroPod/Termination)CO external instrument interface disconnected from monitorCable disconnected between CO2 unit and bedside monitorerror
CO CHECK EXTERNAL DEVICE (Sidestream)AG-400R CO2 unit disconnected from monitorAG-400R unit interface cable unpluggederror
CO CHECK SENSORCO connection cord disconnected or CO connection issueCO connection cord removed from monitor interfaceerror
CO CHECK SENSOR (Mainstream)CO mainstream sensor light level insufficient for measurementInsufficient light reaching photodetector in sensorerror
CO CHECK TEMPERATUREInjectate temperature baseline not detected during CO measurementTemperature sensor not reading injectate temperatureerror
CO CONNECTOR OFF (Mainstream)CO sensor kit disconnected from monitor or connector damagedSensor kit cable unplugged or damaged connectorerror
CO DEVICE ERROR (Microcap/MicroPod/NanoPod)Communication error between external CO2 instrument and monitorCommunication link failure or timeouterror
CO INJECTInjectate not injected or monitor does not acknowledge injectionInjectate not detected at sensor or monitor communication failureerror
CO INJECTION TIME OUTInjectate not detected within 30 seconds after MANUAL key pressed30-second timeout elapsed without injectate detectionerror
CO LINE BLOCK (Sidestream/Microcap/MicroPod/NanoPod)CO2 sampling line clogged; cannot measure CO2Airway adapter, filterline, or exhaust tube blockederror
COMMUNICATION LOSS (System)Network communication failure between monitor and hub/central systemNetwork cable disconnected or communication link failureerror
CONNECT INPUT UNIT (System)Input unit not connected to main monitor; transport function enabled but patient not readyInput unit disconnected when monitor requires it for operationerror
CO NO FILTERLINE (Microcap/MicroPod/NanoPod)CO2 filterline not connected to external instrumentSpecified filterline not connected to external instrument porterror
CO NO FILTERLINE (Sidestream)CO2 filterline not connected or disconnected from unitSpecified filterline not seated in CO2 unit connectorerror
CO OUT OF RANGECO calculated value outside measuring rangeCalculated CO value exceeds measurement limitswarning
CO OUT OF RANGE (Mainstream TG-950P/970P/980P)CO measurement value exceeds valid measurement rangeCalculated CO exceeds sensor measurement limitswarning
CO PLEASE WAIT (Sidestream)CO2 calibration complete; calculation phase in progressZero balance calibration finished; calculation startedinfo
CO PURGING LINE BLOCK (Sidestream/Microcap/MicroPod/NanoPod)CO2 sampling line purge in progress to clear blockageSystem performing automatic purge cycle after clog detectioninfo
CO SENSOR ERROR (Mainstream)CO mainstream sensor damaged or defectiveSensor hardware failure or signal losserror
CO Tb TEMP ERRORBlood temperature outside valid range for CO measurementBlood temperature below 15°C or above 45°Cerror
CO THERMODILUTION ERRORCO thermodilution baseline error; returned too late to baseline curveReturn to baseline takes too long during thermodilution measurementwarning
CO Ti TEMP ERRORInjectate temperature outside valid range for CO measurementInjectate temperature not within manufacturer spec or room temp above 27°Cerror
CO UNSPECIFIED OPERATING ENVIRONMENT (TG-950P/970P/980P)CO module temperature or pressure exceeds operating limitsOperating environment outside specified temperature/pressure rangeerror
CO WARMING UP (Microcap/MicroPod/NanoPod)CO2 external instrument warming up; measurement not possibleDevice power-on initialization in progressinfo
CO WARMING UP (Sidestream)CO2 sidestream unit in warm-up phase; measurement not possiblePower-on warm-up cycle in progressinfo
CO ZERO CALIBRATING (Microcap/MicroPod/NanoPod)CO2 auto zero balance adjustment in progressDevice performing automatic zero calibration on startup or commandedinfo
CO ZERO CALIBRATING (Sidestream)CO2 sidestream zero calibration in progressZero calibration procedure initiatedinfo
DIAG NG CHRFlashRom character area checksum errorPower-on self-testcritical
DIAG NG DATSetting value management data check errorPower-on self-testerror
DIAG NG FRAMFRAM data area check errorPower-on self-testcritical
DIAG NG INFOFlashRom information block checksum errorPower-on self-testcritical
DIAG NG MACMAC address not correctPower-on self-testerror
DIAG NG RAMSDRAM data area errorPower-on self-testcritical
DIAG NG ROMFlashRom program area checksum errorPower-on self-testcritical
DIAG NG RTCDate and real time clock IC errorPower-on self-testerror
DIAG NG SDMemory card error or card not inserted at power-onPower-on self-testwarning
DIAG NG TCHTouch panel error detectedPower-on self-testerror
ECG AUTO LEAD CHANGE (ECG)ECG auto lead change activated; lead electrode detached for >5 secondsElectrode lead detached from patient when AUTO LEAD CHANGE is ONwarning
ECG CHECK ELECTRODES (ECG)ECG electrode connection problem or configuration errorOne or more electrode leads disconnected or contact failureerror
ECG LEARNING (ECG)ECG module learning QRS waveform for arrhythmia analysisQRS learning procedure initiated on startup or operator requestinfo
ECG Low mV (ECG)ECG QRS signal amplitude too low for proper analysisQRS peak voltage below minimum threshold for detectionwarning
ECG MODULE ERROR (ECG)ECG signal processing module malfunctionModule hardware or firmware error detectedcritical
ECG NOISE (ECG)High-amplitude noise corrupting ECG waveformExcessive noise superimposed on ECG signalwarning
ECG PACING (ECG)Paced QRS complex detected in ECG signalPacing spike and QRS detected when AUTO LEAD CHANGE ONinfo
EEG CALIBRATING (EEG)EEG window calibration in progress; calibration waveforms being generatedCalibration initiated via screen controlinfo
EEG CHECK ELECTRODES (EEG)EEG electrode attachment problem; dry or improperly positionedElectrode dry, degraded, or not firmly attached to patienterror
EEG CHECK EXTERNAL DEVICE (EEG)EEG neuro unit disconnected from bedside monitorNeuro unit cable unplugged or not fully seatederror
EEG CONNECTOR OFF (EEG)EEG connection cord disconnected from neuro unit; connector damage possibleEEG connection cord unplugged or damaged connectorerror
EEG EXTERNAL DEVICE FAILURE (EEG)Neuro unit or multi-link instrument failure; faulty external deviceNeuro unit or multi-link instrument malfunction detectederror
EEG IMPEDANCE (EEG)EEG electrode impedance too high; signal quality degradedElectrode impedance exceeds acceptable thresholdwarning
EEG MODULE ERROR (EEG)EEG signal processing module malfunction or firmware errorModule hardware or software error detected during operationcritical
EEG RESETTING BASELINE (EEG)EEG waveform baseline being reset to center positionRESET key on EEG window touchedinfo
ERRORGeneric monitor error conditionMultiple possible internal faultserror
FLOW CAL ERROR (FLOW/Paw)Flow and pressure calibration and zeroing failedCalibration procedure unable to complete successfullyerror
FLOW CALIBRATING (FLOW/Paw)Flow and pressure module calibration or zeroing in progressCalibration procedure initiated during startup or maintenanceinfo
FLOW CHECK EXTERNAL DEVICE (FLOW/Paw)Communication error between Flow/Paw unit and bedside monitorCommunication link failure or timeout between unit and monitorerror
FLOW CONNECTOR OFF (FLOW/Paw)Flow/Paw MEASUREMENT setting disabled in window configurationMEASUREMENT parameter set to OFF in Flow/Paw window settingsinfo
FLOW DEVICE ERROR (FLOW/Paw)Flow/Paw unit malfunction or defect detectedUnit self-test or operation reveals malfunctionerror
FLOW OUT OF RANGE (FLOW/Paw)Flow or pressure measured value exceeds valid measurement rangeSensor reading outside calibrated measurement limitswarning
FRAM Check: Manage Table ErrorSet value check error detected in the management areaPower-on self-test or during operationerror
FRAM Check: Setting Area Sum ErrorSet value check error detected in SUM checkPower-on self-test or during operationerror
FRAM Check: Setting Table Size ErrorSet value check error detected in memory size checkPower-on self-test or during operationerror
FREEZE (All parameters)All waveforms on display frozen; not updating in real-timeFreeze function activated or waveform update suspendedwarning
GAS CAL COMPLETEAir or gas calibration process successfully completedGas calibration routine finished on supported gas modulesinfo
GAS CAL ERRORGas calibration failed due to environmental contamination or equipment issueContaminated ambient air or gas leakage detected during calibrationerror
GAS CHECK EXTERNAL DEVICECommunication failure between external gas analysis module and monitorLoss of communication signal between monitor and connected gas moduleerror
GAS LINEBLOCKLine obstruction preventing proper gas sampling (water trap, sampling line, or exhaust tube blocked)Flow or pressure sensor detects blockage anywhere in gas circuiterror
GAS MEASUREMENT OFFGAS MEASUREMENT parameter disabled on gas measurement windowGAS MEASUREMENT setting switched to OFFwarning
GAS MIXED GASTwo anesthetic agents detected simultaneously (incompatible with monitoring settings)Sensor detects two distinct anesthetic agent concentrations above thresholdwarning
GAS OUT OF RANGEMeasured gas concentration outside calibrated measurement rangeGas concentration exceeds upper or lower measurement limitwarning
GAS OVERHEATGas module overheating condition detected (GF-210R/220R/310R/320R models)Internal temperature exceeds safe operating limiterror
GAS POWER OFFGas module not receiving power from monitorPower cord disconnected or loose connectionerror
GAS PURGINGAutomatic purging cycle in progress to clear sampling line obstruction (AG-920R)Blockage detected; automatic purge cycle initiatedinfo
GAS SAMPLE LINE BLOCKSampling line clogged or kinked, restricting gas flow to sensorPressure or flow sensor detects obstruction in sampling pathwayerror
GAS UNSPECIFIED ACCURACYMeasured gas value outside acceptable accuracy range (AG-920R)Measurement accuracy degrades or operating conditions changewarning
GAS WARMING UPGas module undergoing warm-up cycle after power-up; not ready for measurementPower applied to gas module; self-test and warm-up initiatedinfo
GAS WATERTRAP ALERTWater trap missing, not properly attached, or clogged with water or sampling line issuesWater trap detection mechanism or flow sensor activatederror
HIS SYNC ERRORHL7 interface configuration error preventing HIS synchronizationHL7 settings mismatch or incorrect network configurationerror
IBP DISCONNECT SENSORIBP transducer disconnected or not receiving signalNo IBP pulse and mean pressure detectedwarning
IBP LABEL ALREADY REGISTEREDDuplicate IBP connection cord labels detectedMore than one IBP cord with same label connectedwarning
IBP PRESS OUT OF RANGEIBP measurement exceeds measurable rangeBlood pressure reading outside monitor rangewarning
IBP PRESS ZERO OUT OF RANGEZero balance calibration out of acceptable rangeZero balance pressure reading not accuratewarning
IBP PRESS ZERO UNSTABLEZero balance adjustment unstable or driftingZero balance pressure value unstable or changingwarning
IMPEDANCE MEASUREMENT RESP OFFImpedance respiration measurement disabled<IMPEDANCE MEASUREMENT> parameter set to OFF when impedance method neededinfo
INPUT UNIT DISCONNECTData acquisition or input unit not communicating with bedside monitorLoss of communication link between input unit and main monitorerror
IRREGULAR RR MONITOR OFFArrhythmia detected on registered interbed monitor or interbed unit offline/disconnectedArrhythmia signal received or network communication loss with interbed unitwarning
MPU MODULE ERROR TEMPTemperature signal processor self-diagnostic failure; cannot cover full 0-45°C rangeTemperature measurement self-test failed during periodic validationerror
MULTILINK CONFIG ERRORCommunication failure between QF/IF interface and monitor on multilink systemLoss of communication signal on multilink interfaceerror
MULTILINK POWER ERRORMulti-link power supply failure preventing interface communicationMulti-link power supply voltage out of specificationerror
NIBP AIR HOSE LEAKAir hose disconnected or damaged, preventing cuff inflation or causing air leakLoss of cuff pressure or no inflation pressure developingerror
NIBP CANNOT DETECT PULSEUnable to detect patient pulse wave during NIBP measurementPulse wave amplitude below detection threshold during measurement cyclewarning
NIBP CHECK INTERVAL SETTINGNIBP measurement cannot restart after power interruption during auto/STAT/SIM measurementPower loss and restoration during active measurement sequenceerror
NIBP CONNECTOR OFFAir hose disconnected from monitor, disabling NIBP functionAir pressure connection missing or disconnectederror
NIBP CUFF OCCLUSIONCuff pressure not decreasing properly after NIBP measurement cycle completionResidual pressure remains in cuff after deflation cyclewarning
NIBP DIA OUT OF RANGEMeasured diastolic blood pressure outside acceptable measurement rangeMeasured diastolic pressure exceeds lower or upper range limitwarning
NIBP HIGH CUFF PRESSExcessive cuff pressure detected during measurement (patient movement or cuff malfunction)Cuff pressure exceeds safe upper limit during inflationwarning
NIBP INFLATION PRESS LOWInsufficient cuff inflation pressure at measurement startCuff inflation pressure below minimum required at cycle initiationwarning
NIBP MEAS TIMEOUTNIBP measurement duration exceeded specification due to arrhythmia or noise interferenceMeasurement cycle time exceeded maximum allowed durationwarning
NIBP MEASURINGNIBP measurement in progress with automatic remeasurement due to artifact or cuff issuePatient movement, weak pulse, cuff issue detected during measurement; retry initiatedinfo
NIBP MODULE ERRORNIBP module hardware malfunction preventing blood pressure measurementInternal module self-test failure or communication losserror
NIBP OUT OF RANGESystolic or diastolic blood pressure measured outside monitor's acceptable rangeEither measured pressure exceeds device measurement specificationwarning
NIBP PLEASE WAITCuff pressure not fully deflated before new measurement attemptCuff deflation incomplete; pressure still above minimum thresholdinfo
NIBP SAFETY CIRCUIT RUNNINGSafety protection activated due to improper measurement sequence or air circuit obstructionSafety circuit detects abnormal condition during measurementwarning
NIBP SYS OUT OF RANGEMeasured systolic blood pressure outside monitor's acceptable rangeSystolic pressure exceeds upper or lower measurement limitwarning
NO PACER PULSENo pacemaker artifact detected in ECG or arrhythmia condition presentECG signal analysis shows no pacemaker spikes or arrhythmia detectedwarning
NO PCCO-APPressure sensor not connected for PiCCO hemodynamic monitoringPressure sensor input not detected at module initializationerror
O2 CALOxygen sensor calibration cycle in progress or calibration error detectedCalibration mode activated; sensor signal verification in progress or failedwarning
O2 CONNECTOR OFFOxygen sensor cable disconnected from monitor or cable damagedNo pressure signal from oxygen sensor connectionerror
PCCO CHECK EXTERNAL DEVICEPiCCO hemodynamic monitor disconnected or interface communication lostLoss of communication with external PiCCO monitor uniterror
PCCO CHECK MODEPressure sensor type and PiCCO module technology mismatch preventing measurementSensor specification does not match selected measurement module/technologywarning
PM FAILUREData transmission failure between input unit and bedside/central monitor due to time zone mismatchTime zone settings differ between transmitting and receiving monitorwarning
PRINT ERRORPrint operation failedPrint command to network printer failedwarning
RESP CHECK SENSORRespiration sensor check requiredRespiration pickup damaged or disconnectedwarning
RESP CONNECTOR OFFRespiration connector disconnectedRespiration pickup disconnected from monitorwarning
ScvO2 AMBIENT TEMP HIGHAmbient temperature exceeds safe operating range for CeVOXEnvironmental temperature too high at module locationunknown
ScvO2 CAL ERRORCeVOX ScvO2 calibration failedScvO2 calibration procedure failed or probe error detectedwarning
ScvO2 CONNECTOR OFFCeVOX module not properly connectedCeVOX module connector not seated or disconnectedwarning
ScvO2 LIGHT INTERFERENCEExcessive ambient light detected by CeVOX moduleAmbient light level exceeds acceptable threshold at probe siteunknown
ScvO2 MODULE ERRORCeVOX module internal hardware errorCeVOX module internal fault detectederror
ScvO2 OVERHEATCeVOX module internal temperature exceededCeVOX module internal temperature exceeds safe operating limitwarning
ScvO2 SIGNAL TOO HIGHCeVOX signal strength excessiveProbe attached at inappropriate site for measurementwarning
ScvO2 SIGNAL TOO LOWCeVOX signal strength too weakProbe signal below minimum detection thresholdwarning
ScvO2 WARMING UPCeVOX module startup warm-up phaseCeVOX module is warming up after connection or power-oninfo
SENDING DATABedside monitor transmitting data to central monitorInput unit transferred from one bedside monitor to another in networkinfo
SENSOR ERROR COMPLETEProAQT sensor expired or malfunctionProAQT sensor reaches end of life or failswarning
SETLIB: Keycode ErrDuplicate key code detected during copying set valueDuring set value copy operationwarning
SIMULATED DATADisplay showing simulated/demo patient data<SIMULATION MODE> set to ON instead of OFFwarning
SpO2-2 ALARMPulse oximeter alarm condition on SpO2-2 channelAlarm triggered on Nellcor or Masimo pulse oximeter moduleunknown
SpO2-2 CHANGE PROBESpO2-2 probe deteriorated, damaged or short-circuitedProbe worn out, broken, or electrical malfunctionwarning
SpO2-2 CONNECTOR OFFSpO2-2 probe cable disconnected from cordProbe cable or adapter disconnected from monitor socketwarning
SpO2-2 DETECTING PULSESpO2-2 searching for pulse waveform (NK/Masimo)SpO2-2 module actively searching for pulse signalinfo
SpO2-2 DETECT PULSECannot detect pulse on SpO2-2 (Nellcor/Masimo)SpO2-2 module unable to find pulse signalwarning
SpO2-2 LIGHT INTERFERENCEExcessive light interference on SpO2-2 probeAmbient light level exceeds SpO2-2 sensor thresholdunknown
SpO2-2 LOW QUALITY SIGNALSpO2-2 signal quality poor (NK/Masimo)SpO2-2 measurement signal unstable or weakwarning
SpO2-2 MODULE ERRORSpO2-2 hardware malfunction (NK/Masimo)SpO2-2 module internal hardware failure detectederror
SpO2-2 NO PROBESpO2-2 probe not connected to cord (NK)Probe not attached to SpO2 connection cordcritical
SpO2-2 WEAK PULSESpO2-2 weak pulse signal detection (NK/Masimo)SpO2-2 pulse signal weak or marginalwarning
SpO2 CHECK PROBESpO2 probe attachment or damage check needed (NK)Probe not properly attached or damaged detectionwarning
SpO2 CONNECTOR OFFSpO2 probe cable or adapter disconnectedProbe cable disconnected from adapter or adapter from socketwarning
SpO2 DETECTING PULSESpO2 system searching for pulse waveformSensor actively detecting pulse after probe attachmentinfo
SpO2 LIGHT INTERFERENCEExcessive ambient light at SpO2 probe siteLight level exceeds acceptable threshold for SpO2 measurementunknown
SpO2 LOW QUALITY SIGNALSpO2 signal quality poor due to motion or low perfusionSpO2 waveform unstable or signal strength lowwarning
SpO2 MODULE ERRORSpO2 hardware malfunctionSpO2 module internal hardware failureerror
SpO2 NO PROBESpO2 probe not connected to connection cord (NK)Probe connector not attached to SpO2 cord on monitorcritical
SpO2 PROBESpO2 probe identification or integrity error (Masimo)Monitor cannot identify probe or probe damagederror
SpO2 WEAK PULSESpO2 weak pulse signal from poor circulation or tight probePulse signal strength below acceptable thresholdwarning
StO2 CHECK EXTERNAL DEVICECerebral/somatic oximeter disconnected or communication errorCerebral/somatic oximeter not communicating with bedside monitorwarning
SUSPEND MONITORINGPatient monitoring has been suspendedSUSPEND MONITORING key touchedwarning
SYSTEM DOWN DATA ABORT/WP ERRORUnauthorized address was accessedDuring system operationcritical
SYSTEM DOWN PREFETCH ABORTUnauthorized instruction was fetchedDuring system operationcritical
SYSTEM DOWN UNDEFINED INSTRUCTIONUndefined instruction was givenDuring system operationcritical
SYSTEM DOWN ZERO DIVIDEDivision by zero errorDuring system operationcritical
SYSTEM ERRORGeneral system error with unspecified large classificationDuring system operationerror
SYSTEM ERROR FILE CREATIONFile cannot be properly createdDuring data logging or file operationerror
SYSTEM ERROR LIBRARY RUNTIMELibrary runtime error occurredDuring system operationerror
SYSTEM ERROR SD CARD TRANSFERSD card transfer task error occurredDuring memory card data transfererror
TB SIGNAL UNSTABLEUnstable temperature signal in hemodynamic monitoring unitTD curve error or baseline error due to unstable Tb signalwarning
tcPO2/PCO2 ALARMTranscutaneous oxygen and carbon dioxide monitor alarmAlarm condition detected on transcutaneous monitorwarning
tcPO2/PCO2 CHECKCommunication check failure for transcutaneous gas monitorDisconnection or communication error between transcutaneous monitor and bedside monitorerror
TD MEASUREMENT LIMITMaximum number of thermodilution measurements reachedStorage limit for TD measurements exceededwarning
TD TEMP SENSOR ERRORInjectate temperature sensor malfunction on hemodynamic unitSensor error detectederror
TD TIME OUTThermodilution measurement timeoutNo injection detected within expected time windowwarning
TEC DATA RECEIVE ERRORCommunication error receiving data from defibrillatorDefibrillator data transmission failure to bedside monitorerror
TEC INTERFACE ALREADY CONNECTEDTwo defibrillator interfaces connected simultaneouslyMore than one QI-670P interface connected to monitorerror
TEC INTERFACE DISCONNECTEDQI-670P defibrillator interface USB cable disconnectedUSB cable not connected to monitorerror
TEC INTERFACE ERRORCommunication error with QI-670P defibrillator interfaceCommunication failure between QI-670P interface and monitorerror
TEMP CHECKTemperature probe or connection errorTemperature sensor disconnection or malfunctionwarning
TEMP THIS LABEL IS ALREADY REGISTEREDDuplicate temperature probe labelMore than one temperature probe with the same label detectedwarning
THIS PARAMETER IS ALREADY REGISTEREDExceeds specified number of channels for a parameterMore than specified number of parameter channels are activatedwarning
TI INJECTATE TEMP HIGHInjectate temperature too warm for thermodilution measurementInjectate temperature warmer than blood temperaturewarning
TOF ALARMNeuromuscular transmission monitoring device errorError detected on NMT monitoring devicewarning
TOF CAL??NMT module not calibratedCalibration not yet performed on NMT modulewarning
TOF CALERRORNMT module calibration failureCalibration procedure failed on NMT moduleerror
TOF CHECK EXTERNAL DEVICENMT monitoring device disconnected from bedside monitorNMT device communication loss with main monitorerror
TOF CHECK SENSORNMT sensor cables or electrodes not connectedMain cable, stimulation cable, electrodes or acceleration transducer disconnectionerror
TOF CHECK TEMP SENSORNMT temperature sensor not connectedTemperature sensor disconnection detectederror
TOF COMMUNICATION ERRORCommunication error between NMT module and bedside monitorCommunication failure with NMT moduleerror
TOF LOW TEMPERATUREBody temperature too low for NMT monitoringBody temperature below 32°C (89.6°F)warning
TOF PLEASE WAITInsufficient interval between NMT stimulation cyclesStimulation applied within 2 minutes of previous measurementinfo
TOF PRE PTC RESPONSE DETECTEDVital reaction detected before post-tetanic potentiationPatient response detected before PTC phasewarning
TOF SIGNAL OUT OF RANGETemperature sensor reading outside measurable rangeTemperature measurement outside 20.0 to 42.5°C rangeerror
TOF SIGNAL TOO WEAKAcceleration signal during NMT measurement too weakSignal amplitude below threshold during measurementwarning
TOF SUPRAMAXIMAL NOT FOUNDSupramaximal stimulation current cannot be determinedFailed to establish supramaximal current during calibrationerror
TRANSMITTER CONNECTOR OFFWireless transmitter cable disconnected from monitorTransmitter connection cord not connected to monitorerror
UPDATING DATAParameter update in progress on trend screenParameter being changed on trend screeninfo
VARIABLE PULSE RATESVV (Stroke Volume Variation) measurement may not be reliablePatient pulse rate is variable or unstable during measurementwarning
VENT ALARMAlarm occurring on connected ventilatorAlarm condition detected on ventilator devicewarning
VENT CHECK EXTERNAL DEVICEVentilator disconnected or communication errorVentilator communication loss with bedside monitorerror
ZEROING COMPLETEHemodynamic pressure zero adjustment completed successfullyZero adjustment procedure completedinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI