Nihon Kohden BSM-6000: troubleshooting and error codes
The ServiceForge knowledge base holds 305 documented symptoms for the Nihon Kohden BSM-6000, with 483 ranked causes, 245 diagnostic checks with test points and expected values, and 229 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
ECG noise in signal patient safety
High-amplitude noise corrupting ECG waveform; EMG muscle noise, motion artifact, or interference
Ranked causes
- 1. Baseline instability from respiration or body movement (common)
- 2. EMG muscle noise at electrode site (common)
- 3. Electrode pulled by lead tension (occasional)
- 4. Dry or degraded electrode (common)
- 5. Poor contact between lead and electrode (occasional)
- 6. High electrode impedance (occasional)
- 7. Electric blanket in use (rare)
- 8. Strong electromagnetic interference nearby (occasional)
- 9. Equipotential grounding not acquired (occasional)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electromagnetic interference affecting monitoring patient safety
Artifact appears on display, incorrect readings, or system malfunction due to external EM sources
Ranked causes
- 1. Strong EM interference from nearby emitter (radio station, cellular phone) (common)
- 2. RF interference through AC power supply (occasional)
- 3. Electrostatic discharge effects (occasional)
- 4. Equipment interferes with radio/TV receiver (occasional)
- 5. Lightning-induced voltage surge (rare)
- 6. Incompatible accessory, transducer, or cable connected (occasional)
- 7. Non-standard system configuration used (occasional)
- 8. Excessive measurement sensitivity setting (common)
- 9. Radiation therapy room environment interference (rare)
Identify nearby EM emitters (radio, cellular phone, etc.)
- Test point
- Equipment location
- Expected
- No strong emitters within operating range
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG waveform does not appear on screen
No ECG waveform displayed; 'ECG CHECK ELECTRODES' message may appear
Ranked causes
- 1. Electrode peels or electrode gel is dry (common)
- 2. High contact impedance between electrode and skin (common)
- 3. Instrument not properly grounded (occasional)
- 4. ECG connection cord not firmly connected (common)
- 5. ECG connection cord failure (occasional)
- 6. Number of electrodes set incorrectly (occasional)
- 7. Input unit failure (rare)
Inspect electrodes for peeling or dry gel
- Test point
- Patient electrode pads
- Expected
- Electrodes intact with wet gel
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zeroing (zero adjustment) incomplete
IBP zero adjustment procedure does not complete successfully
Ranked causes
- 1. Noise (MPU MODULE ERROR displayed) (occasional)
- 2. IBP connection cord or blood pressure transducer failure (PRESS CHECK SENSOR displayed) (occasional)
- 3. Pressure less than -200 mmHg or more than +200 mmHg applied to transducer (occasional)
- 4. Transducer failure (occasional)
- 5. Some pressure remains (PRESS ZERO UNSTABLE displayed) (occasional)
- 6. External pressure or vibration applied to blood circuit (occasional)
- 7. MPU board or EXTMPU board failure (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power lamp does not light high voltage
Instrument does not turn on in AC power operation and AC power lamp is off
Ranked causes
- 1. Power cord failure (common)
- 2. Poor contact at cable between AC inlet and power supply unit (common)
- 3. Poor contact at cable between power supply unit and MAIN DIGITAL board (common)
- 4. Poor contact at cable between MAIN DIGITAL and FRONT IF board (common)
- 5. Poor contact at cable between FRONT IF and Power SW board (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 measurement out of range patient safety
CO2 calculated value outside measuring range; catheter retaining condition or coefficient value incorrect
Ranked causes
- 1. Injectate temperature not specified by catheter manufacturer (common)
- 2. Blood temperature out of range (occasional)
- 3. Catheter faulty (occasional)
- 4. Returned to thermodilution baseline too late (rare)
- 5. Injectate not cooled; room temperature above 27°C (occasional)
Measure blood temperature
- Test point
- Temperature probe
- Expected
- 15°C to 45°C, +15 to +45°C
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG electrode detached or lead problem patient safety
ECG electrode lead disconnected or detached from electrode; poor contact between lead and electrode
Ranked causes
- 1. Operator disconnected ECG lead or cord (common)
- 2. Electrode lead detached from electrode (common)
- 3. Electrode cannot attach firmly to skin (occasional)
- 4. Electrode lead disconnected from connection cord (common)
- 5. Poor contact between lead and electrode (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heart rate not counted correctly and ECG noise displayed
Heart rate detection inaccurate; 'ECG NOISE' message appears on display
Ranked causes
- 1. Electrode peels or electrode gel is dry (common)
- 2. Baseline drift or EMG due to body movement (common)
- 3. High contact impedance between electrode and skin (common)
- 4. Instrument not properly grounded (occasional)
- 5. Input unit failure (rare)
Inspect electrodes for peeling or dry gel
- Test point
- Patient electrode pads
- Expected
- Electrodes intact with wet gel
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP MEASURING message and automatic remeasurement
Patient movement or weak pulse wave detected during active NIBP measurement, automatic remeasurement initiated
Ranked causes
- 1. Patient moved during measurement (common)
- 2. Weak patient pulse wave (occasional)
- 3. Cuff not properly attached (common)
- 4. Incorrect cuff size for patient (common)
- 5. Patient unstable pulse and heart rate (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2-2 DETECTING PULSE - Pulse detection in progress (NK/Masimo) patient safety
SpO2-2 sensor actively searching for and detecting pulse waveform
Ranked causes
- 1. Normal pulse detection process in progress (common)
- 2. Poor blood circulation at site (common)
- 3. Probe not attached at appropriate site (occasional)
- 4. Probe secured too tightly (occasional)
- 5. Probe disconnected from adapter (occasional)
Monitor detection progress
- Test point
- SpO2 waveform display
- Expected
- Pulse waveform eventually detected
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TEMP CHECK patient safety
Temperature probe or connection cord disconnected or damaged
Ranked causes
- 1. Probe disconnected from TEMP socket or cord disconnected from MULTI socket (common)
- 2. Probe or temperature connection cord damaged (occasional)
- 3. Measured value outside measuring range (occasional)
- 4. Temperature connection cord disconnected (occasional)
- 5. Temperature connection cord damaged (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO CHECK SENSOR disconnected
FloTrac sensor disconnected from APCO/IBP processor or specified FloTrac sensor not being used.
Ranked causes
- 1. FloTrac sensor disconnected from APCO/IBP processor (common)
- 2. Wrong FloTrac sensor model used (occasional)
- 3. Cable or sensor damaged (occasional)
- 4. APCO/IBP processor internal failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO CHECK WAVEFORM unstable
Arterial waveform quality insufficient for accurate CO measurement or transmission.
Ranked causes
- 1. Arterial waveform inadequate for CO measurement (common)
- 2. Waveform inadequate for extended measurement period (occasional)
- 3. IBP line improper setup (occasional)
- 4. Arterial pressure outside normal range (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baseline drift or waveform synchronized with heartbeat in respiration
Baseline drift or heartbeat-synchronized waveform mixed in respiration measurement
Ranked causes
- 1. Body movement artifact in respiration signal (common)
- 2. Respiration noise not filtered (occasional)
- 3. Input unit failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
BIS SENSOR disconnected patient safety
BIS sensor disconnected from patient interface cable or sensor attachment failed.
Ranked causes
- 1. Patient interface cable disconnected from BIS/BISx processor (common)
- 2. BIS sensor disconnected from patient interface cable (common)
- 3. BIS sensor impedance too high (occasional)
Verify patient interface cable connection
- Test point
- BIS/BISx processor interface
- Expected
- Cable firmly seated with click
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO UNSTABLE measurement
Arterial waveform quality or IBP line conditions unstable for reliable CO measurement.
Ranked causes
- 1. Arterial waveform unstable (common)
- 2. IBP line not appropriate (occasional)
- 3. Patient pulse getting low (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication loss system network caution
System network cable disconnected or communication link failure between monitor and hub/router
Ranked causes
- 1. Network cable disconnected (common)
- 2. Cable or connector faulty (occasional)
- 3. Hub or router defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG electrode configuration incorrect patient safety
NUMBER OF ELECTRODES setting on ECG window not correct; differential offset voltage at electrodes
Ranked causes
- 1. Incorrect electrode count setting (common)
- 2. Electrode lead damaged (occasional)
- 3. Differential offset voltage at electrodes (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequent pacing marks appear on ECG
Pacing marks appear frequently on ECG waveform display
Ranked causes
- 1. Instrument not properly grounded (common)
- 2. Electric blanket not shielded (occasional)
- 3. Input unit failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
GAS CHECK WATERTRAP message displayed
Water trap not properly attached or clogged, or sampling line not properly connected
Ranked causes
- 1. Water trap not securely attached (common)
- 2. Sampling line not properly connected (common)
- 3. Water trap clogged (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
GAS LINEBLOCK re-alarm message
Sampling or exhaust gas line blocked, or water trap clogged or sample line blocked
Ranked causes
- 1. Water trap clogged (common)
- 2. Sampling line clogged or bent (common)
- 3. Exhaust gas tube clogged or bent (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
GAS MIXED GAS silence alarm patient safety
Two anesthetic agents detected simultaneously, typically during vaporizer change or dual vaporizer operation
Ranked causes
- 1. Vaporizer being changed with dual agents in circuit (common)
- 2. Two vaporizers operating simultaneously (occasional)
- 3. Wrong agent delivered to vaporizer (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP DISCONNECT SENSOR - Blood pressure transducer disconnected patient safety
Invasive blood pressure monitoring disconnected or not receiving pulse and mean pressure
Ranked causes
- 1. Blood pressure transducer not properly connected (common)
- 2. Blood pressure transducer damaged (occasional)
- 3. IBP connection cord is damaged (occasional)
Verify IBP transducer connection
- Test point
- IBP connector on monitor
- Expected
- Firmly connected and seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement shows MEAS TIMEOUT patient safety
NIBP measurement fails after remeasurement, displays 'MEAS TIMEOUT' message
Ranked causes
- 1. Cuff not properly attached to patient (common)
- 2. Vibration or external pressure on cuff or air hose (common)
- 3. Failure in oscillation signal amplification circuit of MPU board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nothing displayed on external display
External display shows no output or is blank
Ranked causes
- 1. Poor contact or break in display cable when interface displayed in OPTIONS (common)
- 2. Display, interface, or MAIN DIGITAL board failure (occasional)
- 3. Interface or MAIN DIGITAL board failure when not in OPTIONS (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PRINT ERROR
Network printer communication or paper loading failure
Ranked causes
- 1. Too many print commands sent to printer (occasional)
- 2. Incorrect printer settings (common)
- 3. Recording paper not properly loaded (common)
Wait for print queue to clear
- Test point
- Printer status
- Expected
- No pending print jobs
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 DETECTING PULSE - Cannot find pulse waveform yet patient safety
SpO2 sensor searching for pulse waveform but not yet detected
Ranked causes
- 1. System searching for pulse waveform detection (common)
- 2. Poor blood circulation (common)
- 3. Probe attached too tightly (occasional)
Wait for initial pulse detection
- Test point
- SpO2 waveform display
- Expected
- Pulse waveform appears within normal time
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 LIGHT INTERFERENCE - Excessive light at probe site (NK/Masimo)
SpO2 sensor detecting too much ambient light affecting measurement
Ranked causes
- 1. Excessive ambient light at probe site (common)
- 2. Interference from surroundings (occasional)
- 3. Probe attachment site not appropriate (occasional)
Assess ambient light level at probe site
- Test point
- Probe area illumination
- Expected
- Minimal direct or ambient light
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 LOW QUALITY SIGNAL - Poor signal quality or movement artifact (NK/Nellcor/Masimo) patient safety
SpO2 measuring unstable waveform due to motion or low perfusion
Ranked causes
- 1. Considerable body movement (common)
- 2. Probe not properly attached (common)
- 3. Low signal IQ (occasional)
Assess patient motion
- Test point
- Patient extremity movement
- Expected
- Minimal involuntary movement
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 WEAK PULSE - Poor peripheral circulation or low perfusion (NK/Masimo) patient safety
SpO2 sensor detecting weak pulse signal due to poor circulation or tight probe
Ranked causes
- 1. Poor peripheral circulation (common)
- 2. Probe attached too tightly (occasional)
- 3. Low perfusion index (occasional)
Assess patient peripheral perfusion
- Test point
- Patient extremity color, warmth, capillary refill
- Expected
- Good perfusion with normal parameters
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2-2 ALARM - Pulse oximeter alarm (Nellcor/Masimo) patient safety
Alarm condition on connected pulse oximeter (SpO2-2 channel)
Ranked causes
- 1. Alarm on pulse oximeter device (common)
- 2. Poor blood circulation at measurement site (common)
- 3. Probe attached too tightly (occasional)
Review pulse oximeter alarm indication
- Test point
- SpO2 device display
- Expected
- Identify specific alarm condition
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2-2 CHANGE PROBE - Probe deteriorated or damaged (NK/Nellcor/Masimo) patient safety
SpO2-2 probe has deteriorated, damaged, or short-circuited and requires replacement
Ranked causes
- 1. Probe deteriorated from age or use (common)
- 2. Probe damaged or short-circuited (occasional)
- 3. SpO2 adapter is damaged (occasional)
Inspect probe for visible deterioration
- Test point
- Probe body and connector
- Expected
- No discoloration, cracks, or wear
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2-2 DETECT PULSE - Cannot detect pulse (Masimo) patient safety
Masimo SpO2-2 module cannot detect pulse signal
Ranked causes
- 1. Pulse not detected by SpO2-2 sensor (common)
- 2. Poor blood circulation (common)
- 3. Probe attached too tightly (occasional)
Assess patient perfusion status
- Test point
- Patient extremity
- Expected
- Good color and warmth indicating adequate circulation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2-2 DETECT PULSE - Cannot detect pulse (NK) patient safety
NK (Nihon Kohden) SpO2-2 module cannot detect pulse signal
Ranked causes
- 1. Pulse not detected by SpO2-2 sensor (common)
- 2. Poor blood circulation (common)
- 3. Probe attached too tightly (occasional)
Assess patient perfusion status
- Test point
- Patient extremity
- Expected
- Good color and warmth indicating adequate circulation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2-2 LIGHT INTERFERENCE - Excessive light interference (NK/Masimo)
SpO2-2 module detecting excessive light interference at probe site
Ranked causes
- 1. Excessive ambient light at probe site (common)
- 2. Interference from surroundings (occasional)
- 3. Probe attachment site not appropriate (occasional)
Assess light level at probe
- Test point
- Probe area
- Expected
- Minimal ambient light exposure
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2-2 WEAK PULSE - Weak pulse signal (NK/Masimo) patient safety
SpO2-2 weak pulse detection from poor circulation or probe tension issue
Ranked causes
- 1. Poor peripheral circulation (common)
- 2. Probe attached too tightly (occasional)
- 3. Low perfusion status (occasional)
Evaluate patient hemodynamic status
- Test point
- Patient vital signs and appearance
- Expected
- Adequate blood pressure and perfusion
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
TOF CHECK SENSOR patient safety
Main cable, stimulation cable, electrodes disconnected or abnormal stimulation current
Ranked causes
- 1. Cables or electrodes not connected (common)
- 2. Twitch in patient too strong (rare)
- 3. Abnormal electric stimulation current detected (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch key response does not match display position
Pressing a key position on screen does not activate the intended function; mark appears elsewhere
Ranked causes
- 1. Edge protection rubber not properly attached (common)
- 2. Touch screen failure (occasional)
- 3. MAIN DIGITAL board failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not turn on in battery operation caution
Instrument fails to power on when operating from battery
Ranked causes
- 1. Poor contact at connectors between BATT MOTHER0 and MAIN DIGITAL boards (common)
- 2. Poor contact at connectors between BATT MOTHER1 and MAIN DIGITAL boards (common)
- 3. MAIN DIGITAL, BATT MOTHER0 or BATT MOTHER1 board failure (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm information not output from bedside monitor to nurse call system patient safety
Alarm signals from bedside monitor are not transmitted to nurse call system
Ranked causes
- 1. Short circuit in YJ-672P nurse call cable when interface displayed and signals not output (occasional)
- 2. Interface or MAIN DIGITAL board failure when not in OPTIONS (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot calibrate FiO2, O2CAL?? displayed
FiO2 calibration fails, message 'O2CAL??' shown on screen
Ranked causes
- 1. O2 sensor reached end of life (common)
- 2. MPU board or EXTMPU board failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot communicate with external equipment through RS-232C port
Communication failure through RS-232C port on QI-631P or QI-671P interface
Ranked causes
- 1. Wrong baud rate setting or connection cable failure when interface displayed (common)
- 2. Interface or MAIN DIGITAL board failure when not displayed in OPTIONS (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot communicate with network
Network communication fails; no response when connected to PC with cross cable and ping test
Ranked causes
- 1. MAIN DIGITAL board failure when ping response not returned (occasional)
- 2. Hub and cables have problems when ping response returned (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot communicate with optional unit through multilink port
Optional unit (QI-631P, QI-632P, QI-634P, QI-671P, QI-672P) not recognized; no error displayed
Ranked causes
- 1. Interface or MAIN DIGITAL board failure when not displayed in OPTIONS (occasional)
- 2. Optional unit or interface failure when port does not display model number (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure cardiac output (CO), temperatures, MPU MODULE ERROR displayed
CO, Tb and Ti temperatures cannot be measured, error message 'MPU MODULE ERROR?' shown
Ranked causes
- 1. Noise such as static electricity in CO circuit (common)
- 2. Input unit failure (occasional)
Measure CO after disconnecting connection cord and reattaching input unit
- Test point
- CO connection cord
- Expected
- CO can be measured
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure CO and temperatures, no error message displayed
CO, Tb and Ti cannot be measured, no error message shown on screen
Ranked causes
- 1. MPU board or EXTMPU board failure (occasional)
- 2. CO connection cord or balloon catheter failure (occasional)
Connect CO connection cord and balloon catheter to another MP socket
- Test point
- Alternative MP sockets
- Expected
- CO can be measured
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure CO, BASELINE DRIFT displayed
CO measurement fails, error message 'CO BASELINE DRIFT' shown on screen
Ranked causes
- 1. MPU board or EXTMPU board failure (occasional)
- 2. Physiological fluctuation or external noise interferes with baseline (occasional)
Connect AX-410G to CO connector at MP socket
- Test point
- CO connector of AX-410G at MP socket
- Expected
- Problem disappears or remains
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure ECG and module error displayed
ECG measurement fails with 'ECG MODULE ERROR?' message on screen
Ranked causes
- 1. Noise such as static electricity (common)
- 2. Input unit failure (occasional)
Disconnect ECG connection cord and remove then reattach input unit
- Test point
- ECG input
- Expected
- ECG can be measured
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure FiO2, MPU MODULE ERROR displayed
FiO2 concentration cannot be measured, error message 'MPU MODULE ERROR?' appears
Ranked causes
- 1. Noise such as static electricity in FiO2 circuit (common)
- 2. FiO2 input unit failure (occasional)
Measure FiO2 after disconnecting connection cord and reattaching input unit
- Test point
- FiO2 connection cord
- Expected
- FiO2 can be measured
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure FiO2, no error message, no FiO2 value displayed
No FiO2 measurement, no error message shown on screen
Ranked causes
- 1. MPU board or EXTMPU board failure (occasional)
- 2. FiO2 connection cord or O2 sensor failure (occasional)
Connect FiO2 connection cord and O2 sensor to another MP socket
- Test point
- Alternative MP sockets
- Expected
- FiO2 can be measured
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure IBP and module error displayed
Invasive blood pressure measurement fails with 'MPU MODULE ERROR?' message on screen
Ranked causes
- 1. Noise such as static electricity (common)
- 2. Input unit failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure IBP; no error message and no pressure value displayed
IBP measurement fails with no error message; blood pressure value not shown
Ranked causes
- 1. MPU board or EXTMPU board failure (common)
- 2. IBP connection cord or blood pressure transducer failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure respiration (thermistor), MPU MODULE ERROR displayed
Temperature cannot be measured, error message 'MPU MODULE ERROR?' appears on screen
Ranked causes
- 1. Noise such as static electricity in respiration pickup circuit (common)
- 2. Input unit failure (occasional)
Measure respiration after disconnecting MP connector and reattaching input unit
- Test point
- MP connector of respiration pickup
- Expected
- Respiration can be measured
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure respiration, no error message, no respiration rate displayed
No temperature measurement, no error message on screen, respiration rate not shown
Ranked causes
- 1. MPU board or EXTMPU board failure (occasional)
- 2. Respiration pickup failure (occasional)
Connect respiration pickup to another MP socket
- Test point
- Alternative MP sockets
- Expected
- Respiration can be measured
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure SpO2, WEAK PULSE or CANNOT DETECT PULSE message displayed
SpO2 measurement fails, message 'SpO2 WEAK PULSE' or 'SpO2 CANNOT DETECT PULSE?' shown
Ranked causes
- 1. SpO2 board or MPUPWR board failure (occasional)
- 2. Instrument and accessories functioning normally (unknown)
Replace SpO2 connection cord and probe, then connect AX-410G to SpO2 socket
- Test point
- SpO2 socket
- Expected
- Messages disappear when AX-410G is connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot measure temperature
Temperature measurement fails; temperature value not displayed
Ranked causes
- 1. Temperature probe failure (common)
- 2. Connector board or Analog board failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CATHETER not connected properly patient safety
Catheter connection error for hemodynamic monitoring (PCCO/ProAQT/CCO)
Ranked causes
- 1. JT-170P CO connection cord improperly connected (common)
- 2. ProAQT sensor expired (common)
Verify ProAQT sensor connection
- Test point
- Connection cord
- Expected
- Properly connected to monitor
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO AP SENSOR ERROR
ProAQT sensor not connected properly or sensor is faulty.
Ranked causes
- 1. ProAQT sensor not connected to sensor cable (common)
- 2. ProAQT sensor faulty (occasional)
Verify ProAQT sensor connection
- Test point
- ProAQT sensor cable connector
- Expected
- Secure connection with click
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO APCHECK SENSOR disconnected
ProAQT sensor or PiCCO catheter and injectate sensor disconnected from monitor.
Ranked causes
- 1. ProAQT sensor disconnected from sensor cable (common)
- 2. PiCCO catheter and/or injectate sensor disconnected (common)
Verify ProAQT sensor connection
- Test point
- ProAQT sensor cable
- Expected
- Sensor firmly connected to cable
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CCO CHECK PATIENT INFORMATION missing patient safety
Patient demographic data (gender, age, BSA) not entered or out of valid range for CCO calculation.
Ranked causes
- 1. Patient gender, age, or BSA not entered (common)
- 2. Patient information out of range (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- CCO CONNECTOR OFF disconnected
- CCO ZERO UNSTABLE
- CO CHECK SENSOR disconnected
- CO2 external device disconnected or interface error
- EEG external device failure
- FiO2 value is randomly unstable
- GAS CAL ERROR message displayed
- GAS CHECK EXTERNAL DEVICE Termination message
- GAS OUT OF RANGE silence alarm
- GAS UNSPECIFIED ACCURACY on AG-920R
- IBP PRESS OUT OF RANGE
- IBP PRESS ZERO OUT OF RANGE
- IRREGULAR RR MONITOR OFF message
- NIBP AIR HOSE LEAK Silence alarm
- NIBP CANNOT DETECT PULSE Silence alarm
- NIBP CONNECTOR OFF Termination message
- NIBP display appears, cuff inflates, pressure does not reach target
- NIBP display area does not appear after air hose connected
- NIBP MEAS TIMEOUT Silence alarm
- NIBP SAFETY CIRCUIT RUNNING Re-alarm message
- NIBP SYS OUT OF RANGE Silence alarm
- Noise on blood pressure waveform
- O2 CAL message displayed
- O2 CONNECTOR OFF Termination message
- O2 FiO2 connection error
- Recorder shows INSERT REC PAPER message with paper properly set
- RESP CHECK SENSOR - Respiration pickup damaged or disconnected
- RESP CONNECTOR OFF - Respiration pickup disconnected (thermistor)
- Respiration rate discrepancy from manual count
- Respiration waveform does not appear; dashes displayed
- SpO2 CHECK PROBE - Probe attachment and damage check (Nellcor/Masimo)
- SpO2 CHECK PROBE - Probe not properly attached or damaged (NK)
- SpO2 CONNECTOR OFF - Probe cable or adapter disconnected
- SpO2-2 CONNECTOR OFF - SpO2-2 probe cable disconnected
- SpO2-2 LOW QUALITY SIGNAL - Poor signal quality (NK/Masimo)
- StO2 CHECK EXTERNAL DEVICE - Cerebral oximeter disconnected or communication error
- StO2 CHECK EXTERNAL DEVICE - Cerebral/somatic oximeter disconnected
- TB SIGNAL UNSTABLE
- tcPO2/PCO2 CHECK
- TEC DATA RECEIVE ERROR
- Touch screen mark stays at particular pressed position
- Touch screen mark stays out of acceptable range
- Vital signs cannot be transmitted to ZS-900P receiver
- A-FIB detected
- Adjacent equipment causes malfunction
- AG-400R CO2 unit appearance and physical check
- Alarm indicator does not light
- All ports display optional unit model numbers on MULTILINK CHECK
- Appearance check failed
- ASYSTOLE detected
- Battery operation time shorter than expected
- Battery operation with SB-671P pack
- Battery weak
- Bigeminy detected
- BIS CHECK EXTERNAL DEVICE disconnected
- BIS CHECKING IMPEDANCE
- BIS MODULE FAILURE
- BIS NOISE detected
- BIS REPLACE SENSOR
- BIS SENSOR ERROR
- BIS SENSOR EXPIRED
- Both display lamp and power lamp light
- BYPASS activated
- Calibration: AE-918P Neuro Unit ABR generator check
- Calibration: BIS/BISx processor impedance check
- Calibration: Cardiac output measurement
- Calibration: CO2 gas accuracy check and sensitivity calibration
- Calibration: CO2 partial pressure accuracy check
- Calibration: Held pressure (cuff pressure maintenance)
- Calibration: NIBP air circuit operation
- Calibration: NIBP cuff pressure measurement
- Calibration: NMT module electrical stimulation check
- Calibration: Pressurization speed (adult/child mode)
- Calibration: Pressurization speed (neonatal mode)
- Calibration: Safety check measurements
- Calibration: Safety circuit pressure release (adult/child mode)
- Calibration: Safety circuit pressure release (neonatal mode)
- Calibration: Solenoid valve operation (Step Deflation)
- Calibration: Touch panel position incorrect
- Cannot analyze ECG
- Cannot measure FiO2, displays "---", O2 CHECK SENSOR message shown
- Cannot measure IBP; dashes displayed and sensor check message
- Cannot measure respiration, RESP CHECK SENSOR displayed
- Cannot measure SpO2, CONNECTOR OFF or CHECK PROBE message displayed
- Cannot measure Tb and Ti, displays "---", CO CHECK SENSOR message shown
- Cannot recognize AY-600P series optional unit
- CCO CHECK EXTERNAL DEVICE disconnected
- CCO MODULE FAILURE
- CCO ZERO IMBALANCE
- CCO ZEROING COMPLETE
- CHECK INPUT UNIT BATTERY
- CO CHECK TEMPERATURE baseline
- CO INJECT failure
- CO INJECTION TIME OUT
- CO2 absorbent needs replacement
- CO2 adapter damaged
- CO2 airway adapter disconnected
- CO2 bath probe or inline sensor faulty
- CO2 calibration failed
- CO2 external device communication error
- CO2 filterline not connected properly
- CO2 sampling line blocked
- CO2 sensor damaged
- CO2 sensor insufficient light
- CO2 sensor kit connector disconnected
- CO2 thermodilution baseline error
- CO2 unspecified operating environment
- CO2 zero calibration not performed
- Connector-off alarm verification
- CPU RAM check failed
- CPU ROM check failed
- Display lamp does not light
- Display lamp lights but power lamp does not light
- ECG module malfunction
- ECG pacing spike detected
- ECG QRS amplitude too low
- EEG connection cord disconnected
- EEG connection cord re-seated during operation
- EEG cord damaged
- EEG cord disconnected from neuro unit
- EEG differential offset voltage high
- EEG electrode dry or not attached
- EEG electrode lead poor contact
- EEG module error
- EEG neuro unit disconnected
- Extended arrhythmia network communication lost
- Flow tube disconnected from unit
- Flow/Paw calibration error
- Flow/Paw communication error with monitor
- Flow/Paw measurement disabled in settings
- Flow/Paw measurement out of range
- FRAM check failed
- GAS CAL COMPLETE message displayed
- GAS CHECK SAMPLE LINE clogged
- GAS DEVICE error - exhaust gas tube clogged
- GAS MEASUREMENT setting not enabled
- GAS OVERHEAT silence alarm on GF-210R/220R/310R/320R
- GAS power indicator shows unit not powered
- GAS PURGING in progress on AG-920R
- GAS WARMING UP message displayed
- Hard key function check failed
- Heart rate not accurate
- High frequency noise on ECG waveform with electrosurgical unit
- High frequency noise on respiration waveform with electrosurgical unit
- HIS SYNC ERROR Termination message
- IBP CALIBRATING - Zero balance in progress
- IBP LABEL ALREADY REGISTERED
- IBP PRESS ZERO UNSTABLE
- IMPEDANCE MEASUREMENT RESP OFF - Impedance respiration monitoring disabled
- Impedance respiration waveform does not appear; RESP OFF message
- INPUT UNIT DISCONNECT Termination message
- Input unit not connected to main unit
- MPU MODULE ERROR TEMP Silence alarm
- MULTILINK CONFIG ERROR message displayed
- MULTILINK CONFIG ERROR Re-alarm message
- MULTILINK POWER ERROR message displayed
- MULTILINK POWER ERROR Re-alarm message
- NIBP CHECK INTERVAL SETTING Termination message
- NIBP CUFF OCCLUSION Re-alarm message
- NIBP DIA OUT OF RANGE Silence alarm
- NIBP display appears, cuff does not inflate
- NIBP HIGH CUFF PRESS Silence alarm
- NIBP INFLATION during measurement
- NIBP INFLATION PRESS LOW
- NIBP MODULE ERROR Silence alarm
- NIBP OUT OF RANGE Silence alarm
- NIBP PLEASE WAIT message displayed
- NIBP PROLONGED RR or Arrhythmia detected, uncalibrated
- NO PACER PULSE ECG alarm
- NO PCCO-AP Hemodynamic measurement error
- No ports display optional unit model numbers on MULTILINK CHECK
- No sound or small volume
- Pacemaker interaction with cardiac monitoring
- Page-dependent recording does not start at page top
- PCCO CHECK EXTERNAL DEVICE Termination message
- PCCO CHECK MODE Silence alarm
- PM FAILURE Re-alarm message
- Power cord safety check failed
- QF/IF series interface connection status check
- QI-670P interface connection and version check
- RECEIVING DATA - Receiving data from defibrillator
- Recorder door shows CLOSE PAPER MAGAZINE message when firmly closed
- Remote controller key check failed
- SAMPLE LINE pressurized on AG-920R
- ScvO2 AMBIENT TEMP HIGH - CeVOX detects high ambient temperature
- ScvO2 CAL ERROR - CeVOX calibration error
- ScvO2 CONNECTOR OFF - CeVOX module not connected
- ScvO2 LIGHT INTERFERENCE - CeVOX detects excessive ambient light
- ScvO2 MODULE ERROR - CeVOX module internal error
- ScvO2 OVERHEAT - CeVOX module overheated
- ScvO2 SIGNAL TOO HIGH - CeVOX probe signal excessive
- ScvO2 SIGNAL TOO LOW - CeVOX probe signal weak
- ScvO2 WARMING UP - CeVOX module warming up
- SENDING DATA - Input unit data transmission to central monitor
- SENSOR ERROR COMPLETE - ProAQT sensor issue
- SIMULATED DATA - Display showing simulated/demo data
- Sound is distorted
- Sound output check failed
- SpO2 MODULE ERROR - SpO2 hardware malfunction (NK/Nellcor/Masimo)
- SpO2 NO PROBE - SpO2 probe not connected to connection cord (NK)
- SpO2 PROBE - Probe identification or integrity error (Masimo)
- SpO2-2 CHECK PROBE - Probe attachment issue (NK)
- SpO2-2 MODULE ERROR - SpO2-2 hardware malfunction (NK/Masimo)
- SpO2-2 NO PROBE - SpO2-2 probe not connected to cord (NK)
- SUSPEND MONITORING
- Tb temperature outside measurable range
- tcPO2/PCO2 ALARM
- TD MEASUREMENT LIMIT
- TD TEMP SENSOR ERROR
- TD TIME OUT
- TEC INTERFACE ALREADY CONNECTED
- TEC INTERFACE DISCONNECTED
- TEC INTERFACE ERROR
- TEMP THIS LABEL IS ALREADY REGISTERED
- THIS PARAMETER IS ALREADY REGISTERED
- TI INJECTATE TEMP HIGH
- Ti temperature outside measurable range
- TOF ALARM
- TOF CAL??
- TOF CALERROR
- TOF CHECK EXTERNAL DEVICE
- TOF CHECK TEMP SENSOR
- TOF COMMUNICATION ERROR
- TOF LOW TEMPERATURE
- TOF PLEASE WAIT
- TOF PRE PTC RESPONSE DETECTED
- TOF SIGNAL OUT OF RANGE
- TOF SIGNAL TOO WEAK
- TOF SUPRAMAXIMAL NOT FOUND
- TRANSMITTER CONNECTOR OFF
- Unit does not turn on in AC power operation
- UPDATING DATA
- USB mouse does not work with interface not in OPTIONS
- USB mouse does not work with optical mouse LED off
- USB mouse does not work with optical mouse LED on
- VARIABLE PULSE RATE
- VENT ALARM
- VENT CHECK EXTERNAL DEVICE
- Water in CO2 respiration circuit
- Waveforms and grid have faint parts
- Waveforms and numeric data record when paper pulled out manually
- YJ-672P nurse call cable connection check
- ZEROING COMPLETE
- ZS-900P transmitter operational status check
- ZS-900P transmitter signal reception verification
245 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
A-FIB | Arrhythmia detected on ECG | Abnormal heart rhythm detected during ECG analysis | critical |
ALARM RESET | ALARM RESET function key was activated | ALARM RESET function key pressed during alarm reset condition | info |
ALARMS SILENCED | Alarm sounds have been silenced | [Silence Alarms] key pressed or ALARM RESET function key touched | warning |
ALARMS SUSPENDED | All alarms have been suspended | [Silence Alarms] key pressed twice or SUSPEND ALARMS key pressed | warning |
ALL ALARMS OFF | All alarm functions have been disabled | ALARMS OFF key pressed or BYPASS key activated | warning |
ARRHYTHMIA ANALYSIS OFF | Arrhythmia analysis function is disabled | SYSTEM SETUP window set to OFF for arrhythmia analysis | warning |
BATTERY WEAK | Battery pack charge level critically low at approximately 25% | Battery pack reaches 25% charge level | warning |
BIS CHECK EXTERNAL DEVICE | BIS monitor is disconnected from the bedside monitor | BIS interface cable disconnected | error |
BIS CHECKING IMPEDANCE | BIS sensor impedance check in progress | Impedance verification routine running on sensor | info |
BIS CHECK SENSOR | BIS sensor connection error or impedance issue | Patient interface cable disconnected or BIS sensor not attached properly | error |
BIS MODULE FAILURE | BIS processor has encountered internal failure | BIS processor self-test failed or internal communication error | critical |
BIS NOISE | Noise interference detected on BIS signal | Electrical or motion noise on BIS sensor signal | error |
BIS REPLACE SENSOR | BIS sensor has exceeded maximum number of uses | Sensor usage counter exceeds manufacturer limit | warning |
BIS SENSOR ERROR | BIS sensor failure detected | Sensor communication failure or self-test failure | error |
BIS SENSOR EXPIRED | BIS sensor has passed its expiration date | Sensor expiration date exceeded during startup or during use | warning |
BISx CONNECTION CABLE | BISx connection cable disconnected from monitor or damaged | BISx cable removed from monitor or connection failure detected | error |
BYPASS | Alarm bypass mode activated | BYPASS key pressed during alarm condition | warning |
CANNOT ANALYZE | ECG arrhythmia analysis cannot be performed | Excessive noise on ECG signal for more than 30 seconds | error |
CCO APCHECK SENSOR | ProAQT sensor or PiCCO catheter cable disconnected from CO monitor | Sensor or cable removal from connection point detected | error |
CCO AP SENSOR ERROR | ProAQT sensor connection error or sensor failure | Sensor not detected or communication failure with sensor | error |
CCO CHECK EXTERNAL DEVICE | CCO monitor disconnected from bedside monitor | CCO interface cable disconnected | error |
CCO CHECK PATIENT INFORMATION | Required patient demographic data missing or invalid for CCO calculation | Patient gender, age, or BSA not entered or values out of acceptable range | error |
CCO CHECK SENSOR | FloTrac sensor disconnected, wrong type, or cable/sensor damaged | FloTrac sensor not detected or wrong sensor model selected | error |
CCO CHECK WAVEFORM | Arterial waveform quality insufficient for reliable CO measurement | Arterial waveform signal degradation or IBP line issue | error |
CCO CONNECTOR OFF | APCO/IBP processor is disconnected from monitor | APCO/IBP processor interface cable disconnected | error |
CCO MODULE FAILURE | APCO/IBP processor is faulty and non-functional | Processor self-test failed or internal error detected | critical |
CCO UNSTABLE | Arterial waveform too unstable for reliable CO measurement | Waveform instability detected over time | error |
CCO ZERO IMBALANCE | Zero balance calibration not performed for APCO/IBP system | Zero balance adjustment required but not completed | error |
CCO ZEROING COMPLETE | Zero balance adjustment successfully completed | Zero balance calibration routine finished | info |
CCO ZERO UNSTABLE | Zero balance adjustment cannot complete due to circuit or pressure instability | Zero balance procedure initiated but cannot stabilize | error |
CHECK INPUT UNIT BATTERY | Battery pack not inserted correctly in BSM-1700 series monitor | Battery pack not properly seated in battery compartment | error |
CO CAL COMPLETE (Sidestream) | CO sidestream calibration completed successfully | Calibration procedure finished without errors | info |
CO CAL ERROR (Sidestream) | CO sidestream calibration failed | Calibration procedure did not complete successfully | error |
CO CALIBRATING (Mainstream TG-950P/970P/980P) | CO mainstream module in calibration; measuring reference light amount | Calibration procedure in progress | info |
CO CALIBRATING (Sidestream) | CO sidestream module in calibration procedure | Calibration procedure in progress | info |
CO CAL? (Mainstream TG-950P/970P/980P) | CO zero calibration not performed on mainstream module | System initialized without zero calibration on startup | error |
CO CELL OFF (TG-970P/980P) | CO sensor airway adapter disconnected or CO2 cell offline | Airway adapter unplugged from sensor | warning |
CO CHANGE ABSORBENT (Sidestream) | CO2 sidestream absorbent cartridge depleted and requires replacement | Absorbent material exhausted after extended use | warning |
CO CHANGE ADAPTER (Mainstream) | CO mainstream adapter damaged or needs replacement | Adapter mechanical damage detected or functionality degraded | warning |
CO CHECK CO2 CELL (TG-950P/970P/980P) | Water collected in respiration circuit; unable to measure CO2 | Water accumulation detected in breathing circuit | error |
CO CHECK EXTERNAL DEVICE (Microcap/MicroPod/Termination) | CO external instrument interface disconnected from monitor | Cable disconnected between CO2 unit and bedside monitor | error |
CO CHECK EXTERNAL DEVICE (Sidestream) | AG-400R CO2 unit disconnected from monitor | AG-400R unit interface cable unplugged | error |
CO CHECK SENSOR | CO connection cord disconnected or CO connection issue | CO connection cord removed from monitor interface | error |
CO CHECK SENSOR (Mainstream) | CO mainstream sensor light level insufficient for measurement | Insufficient light reaching photodetector in sensor | error |
CO CHECK TEMPERATURE | Injectate temperature baseline not detected during CO measurement | Temperature sensor not reading injectate temperature | error |
CO CONNECTOR OFF (Mainstream) | CO sensor kit disconnected from monitor or connector damaged | Sensor kit cable unplugged or damaged connector | error |
CO DEVICE ERROR (Microcap/MicroPod/NanoPod) | Communication error between external CO2 instrument and monitor | Communication link failure or timeout | error |
CO INJECT | Injectate not injected or monitor does not acknowledge injection | Injectate not detected at sensor or monitor communication failure | error |
CO INJECTION TIME OUT | Injectate not detected within 30 seconds after MANUAL key pressed | 30-second timeout elapsed without injectate detection | error |
CO LINE BLOCK (Sidestream/Microcap/MicroPod/NanoPod) | CO2 sampling line clogged; cannot measure CO2 | Airway adapter, filterline, or exhaust tube blocked | error |
COMMUNICATION LOSS (System) | Network communication failure between monitor and hub/central system | Network cable disconnected or communication link failure | error |
CONNECT INPUT UNIT (System) | Input unit not connected to main monitor; transport function enabled but patient not ready | Input unit disconnected when monitor requires it for operation | error |
CO NO FILTERLINE (Microcap/MicroPod/NanoPod) | CO2 filterline not connected to external instrument | Specified filterline not connected to external instrument port | error |
CO NO FILTERLINE (Sidestream) | CO2 filterline not connected or disconnected from unit | Specified filterline not seated in CO2 unit connector | error |
CO OUT OF RANGE | CO calculated value outside measuring range | Calculated CO value exceeds measurement limits | warning |
CO OUT OF RANGE (Mainstream TG-950P/970P/980P) | CO measurement value exceeds valid measurement range | Calculated CO exceeds sensor measurement limits | warning |
CO PLEASE WAIT (Sidestream) | CO2 calibration complete; calculation phase in progress | Zero balance calibration finished; calculation started | info |
CO PURGING LINE BLOCK (Sidestream/Microcap/MicroPod/NanoPod) | CO2 sampling line purge in progress to clear blockage | System performing automatic purge cycle after clog detection | info |
CO SENSOR ERROR (Mainstream) | CO mainstream sensor damaged or defective | Sensor hardware failure or signal loss | error |
CO Tb TEMP ERROR | Blood temperature outside valid range for CO measurement | Blood temperature below 15°C or above 45°C | error |
CO THERMODILUTION ERROR | CO thermodilution baseline error; returned too late to baseline curve | Return to baseline takes too long during thermodilution measurement | warning |
CO Ti TEMP ERROR | Injectate temperature outside valid range for CO measurement | Injectate temperature not within manufacturer spec or room temp above 27°C | error |
CO UNSPECIFIED OPERATING ENVIRONMENT (TG-950P/970P/980P) | CO module temperature or pressure exceeds operating limits | Operating environment outside specified temperature/pressure range | error |
CO WARMING UP (Microcap/MicroPod/NanoPod) | CO2 external instrument warming up; measurement not possible | Device power-on initialization in progress | info |
CO WARMING UP (Sidestream) | CO2 sidestream unit in warm-up phase; measurement not possible | Power-on warm-up cycle in progress | info |
CO ZERO CALIBRATING (Microcap/MicroPod/NanoPod) | CO2 auto zero balance adjustment in progress | Device performing automatic zero calibration on startup or commanded | info |
CO ZERO CALIBRATING (Sidestream) | CO2 sidestream zero calibration in progress | Zero calibration procedure initiated | info |
DIAG NG CHR | FlashRom character area checksum error | Power-on self-test | critical |
DIAG NG DAT | Setting value management data check error | Power-on self-test | error |
DIAG NG FRAM | FRAM data area check error | Power-on self-test | critical |
DIAG NG INFO | FlashRom information block checksum error | Power-on self-test | critical |
DIAG NG MAC | MAC address not correct | Power-on self-test | error |
DIAG NG RAM | SDRAM data area error | Power-on self-test | critical |
DIAG NG ROM | FlashRom program area checksum error | Power-on self-test | critical |
DIAG NG RTC | Date and real time clock IC error | Power-on self-test | error |
DIAG NG SD | Memory card error or card not inserted at power-on | Power-on self-test | warning |
DIAG NG TCH | Touch panel error detected | Power-on self-test | error |
ECG AUTO LEAD CHANGE (ECG) | ECG auto lead change activated; lead electrode detached for >5 seconds | Electrode lead detached from patient when AUTO LEAD CHANGE is ON | warning |
ECG CHECK ELECTRODES (ECG) | ECG electrode connection problem or configuration error | One or more electrode leads disconnected or contact failure | error |
ECG LEARNING (ECG) | ECG module learning QRS waveform for arrhythmia analysis | QRS learning procedure initiated on startup or operator request | info |
ECG Low mV (ECG) | ECG QRS signal amplitude too low for proper analysis | QRS peak voltage below minimum threshold for detection | warning |
ECG MODULE ERROR (ECG) | ECG signal processing module malfunction | Module hardware or firmware error detected | critical |
ECG NOISE (ECG) | High-amplitude noise corrupting ECG waveform | Excessive noise superimposed on ECG signal | warning |
ECG PACING (ECG) | Paced QRS complex detected in ECG signal | Pacing spike and QRS detected when AUTO LEAD CHANGE ON | info |
EEG CALIBRATING (EEG) | EEG window calibration in progress; calibration waveforms being generated | Calibration initiated via screen control | info |
EEG CHECK ELECTRODES (EEG) | EEG electrode attachment problem; dry or improperly positioned | Electrode dry, degraded, or not firmly attached to patient | error |
EEG CHECK EXTERNAL DEVICE (EEG) | EEG neuro unit disconnected from bedside monitor | Neuro unit cable unplugged or not fully seated | error |
EEG CONNECTOR OFF (EEG) | EEG connection cord disconnected from neuro unit; connector damage possible | EEG connection cord unplugged or damaged connector | error |
EEG EXTERNAL DEVICE FAILURE (EEG) | Neuro unit or multi-link instrument failure; faulty external device | Neuro unit or multi-link instrument malfunction detected | error |
EEG IMPEDANCE (EEG) | EEG electrode impedance too high; signal quality degraded | Electrode impedance exceeds acceptable threshold | warning |
EEG MODULE ERROR (EEG) | EEG signal processing module malfunction or firmware error | Module hardware or software error detected during operation | critical |
EEG RESETTING BASELINE (EEG) | EEG waveform baseline being reset to center position | RESET key on EEG window touched | info |
ERROR | Generic monitor error condition | Multiple possible internal faults | error |
FLOW CAL ERROR (FLOW/Paw) | Flow and pressure calibration and zeroing failed | Calibration procedure unable to complete successfully | error |
FLOW CALIBRATING (FLOW/Paw) | Flow and pressure module calibration or zeroing in progress | Calibration procedure initiated during startup or maintenance | info |
FLOW CHECK EXTERNAL DEVICE (FLOW/Paw) | Communication error between Flow/Paw unit and bedside monitor | Communication link failure or timeout between unit and monitor | error |
FLOW CONNECTOR OFF (FLOW/Paw) | Flow/Paw MEASUREMENT setting disabled in window configuration | MEASUREMENT parameter set to OFF in Flow/Paw window settings | info |
FLOW DEVICE ERROR (FLOW/Paw) | Flow/Paw unit malfunction or defect detected | Unit self-test or operation reveals malfunction | error |
FLOW OUT OF RANGE (FLOW/Paw) | Flow or pressure measured value exceeds valid measurement range | Sensor reading outside calibrated measurement limits | warning |
FRAM Check: Manage Table Error | Set value check error detected in the management area | Power-on self-test or during operation | error |
FRAM Check: Setting Area Sum Error | Set value check error detected in SUM check | Power-on self-test or during operation | error |
FRAM Check: Setting Table Size Error | Set value check error detected in memory size check | Power-on self-test or during operation | error |
FREEZE (All parameters) | All waveforms on display frozen; not updating in real-time | Freeze function activated or waveform update suspended | warning |
GAS CAL COMPLETE | Air or gas calibration process successfully completed | Gas calibration routine finished on supported gas modules | info |
GAS CAL ERROR | Gas calibration failed due to environmental contamination or equipment issue | Contaminated ambient air or gas leakage detected during calibration | error |
GAS CHECK EXTERNAL DEVICE | Communication failure between external gas analysis module and monitor | Loss of communication signal between monitor and connected gas module | error |
GAS LINEBLOCK | Line obstruction preventing proper gas sampling (water trap, sampling line, or exhaust tube blocked) | Flow or pressure sensor detects blockage anywhere in gas circuit | error |
GAS MEASUREMENT OFF | GAS MEASUREMENT parameter disabled on gas measurement window | GAS MEASUREMENT setting switched to OFF | warning |
GAS MIXED GAS | Two anesthetic agents detected simultaneously (incompatible with monitoring settings) | Sensor detects two distinct anesthetic agent concentrations above threshold | warning |
GAS OUT OF RANGE | Measured gas concentration outside calibrated measurement range | Gas concentration exceeds upper or lower measurement limit | warning |
GAS OVERHEAT | Gas module overheating condition detected (GF-210R/220R/310R/320R models) | Internal temperature exceeds safe operating limit | error |
GAS POWER OFF | Gas module not receiving power from monitor | Power cord disconnected or loose connection | error |
GAS PURGING | Automatic purging cycle in progress to clear sampling line obstruction (AG-920R) | Blockage detected; automatic purge cycle initiated | info |
GAS SAMPLE LINE BLOCK | Sampling line clogged or kinked, restricting gas flow to sensor | Pressure or flow sensor detects obstruction in sampling pathway | error |
GAS UNSPECIFIED ACCURACY | Measured gas value outside acceptable accuracy range (AG-920R) | Measurement accuracy degrades or operating conditions change | warning |
GAS WARMING UP | Gas module undergoing warm-up cycle after power-up; not ready for measurement | Power applied to gas module; self-test and warm-up initiated | info |
GAS WATERTRAP ALERT | Water trap missing, not properly attached, or clogged with water or sampling line issues | Water trap detection mechanism or flow sensor activated | error |
HIS SYNC ERROR | HL7 interface configuration error preventing HIS synchronization | HL7 settings mismatch or incorrect network configuration | error |
IBP DISCONNECT SENSOR | IBP transducer disconnected or not receiving signal | No IBP pulse and mean pressure detected | warning |
IBP LABEL ALREADY REGISTERED | Duplicate IBP connection cord labels detected | More than one IBP cord with same label connected | warning |
IBP PRESS OUT OF RANGE | IBP measurement exceeds measurable range | Blood pressure reading outside monitor range | warning |
IBP PRESS ZERO OUT OF RANGE | Zero balance calibration out of acceptable range | Zero balance pressure reading not accurate | warning |
IBP PRESS ZERO UNSTABLE | Zero balance adjustment unstable or drifting | Zero balance pressure value unstable or changing | warning |
IMPEDANCE MEASUREMENT RESP OFF | Impedance respiration measurement disabled | <IMPEDANCE MEASUREMENT> parameter set to OFF when impedance method needed | info |
INPUT UNIT DISCONNECT | Data acquisition or input unit not communicating with bedside monitor | Loss of communication link between input unit and main monitor | error |
IRREGULAR RR MONITOR OFF | Arrhythmia detected on registered interbed monitor or interbed unit offline/disconnected | Arrhythmia signal received or network communication loss with interbed unit | warning |
MPU MODULE ERROR TEMP | Temperature signal processor self-diagnostic failure; cannot cover full 0-45°C range | Temperature measurement self-test failed during periodic validation | error |
MULTILINK CONFIG ERROR | Communication failure between QF/IF interface and monitor on multilink system | Loss of communication signal on multilink interface | error |
MULTILINK POWER ERROR | Multi-link power supply failure preventing interface communication | Multi-link power supply voltage out of specification | error |
NIBP AIR HOSE LEAK | Air hose disconnected or damaged, preventing cuff inflation or causing air leak | Loss of cuff pressure or no inflation pressure developing | error |
NIBP CANNOT DETECT PULSE | Unable to detect patient pulse wave during NIBP measurement | Pulse wave amplitude below detection threshold during measurement cycle | warning |
NIBP CHECK INTERVAL SETTING | NIBP measurement cannot restart after power interruption during auto/STAT/SIM measurement | Power loss and restoration during active measurement sequence | error |
NIBP CONNECTOR OFF | Air hose disconnected from monitor, disabling NIBP function | Air pressure connection missing or disconnected | error |
NIBP CUFF OCCLUSION | Cuff pressure not decreasing properly after NIBP measurement cycle completion | Residual pressure remains in cuff after deflation cycle | warning |
NIBP DIA OUT OF RANGE | Measured diastolic blood pressure outside acceptable measurement range | Measured diastolic pressure exceeds lower or upper range limit | warning |
NIBP HIGH CUFF PRESS | Excessive cuff pressure detected during measurement (patient movement or cuff malfunction) | Cuff pressure exceeds safe upper limit during inflation | warning |
NIBP INFLATION PRESS LOW | Insufficient cuff inflation pressure at measurement start | Cuff inflation pressure below minimum required at cycle initiation | warning |
NIBP MEAS TIMEOUT | NIBP measurement duration exceeded specification due to arrhythmia or noise interference | Measurement cycle time exceeded maximum allowed duration | warning |
NIBP MEASURING | NIBP measurement in progress with automatic remeasurement due to artifact or cuff issue | Patient movement, weak pulse, cuff issue detected during measurement; retry initiated | info |
NIBP MODULE ERROR | NIBP module hardware malfunction preventing blood pressure measurement | Internal module self-test failure or communication loss | error |
NIBP OUT OF RANGE | Systolic or diastolic blood pressure measured outside monitor's acceptable range | Either measured pressure exceeds device measurement specification | warning |
NIBP PLEASE WAIT | Cuff pressure not fully deflated before new measurement attempt | Cuff deflation incomplete; pressure still above minimum threshold | info |
NIBP SAFETY CIRCUIT RUNNING | Safety protection activated due to improper measurement sequence or air circuit obstruction | Safety circuit detects abnormal condition during measurement | warning |
NIBP SYS OUT OF RANGE | Measured systolic blood pressure outside monitor's acceptable range | Systolic pressure exceeds upper or lower measurement limit | warning |
NO PACER PULSE | No pacemaker artifact detected in ECG or arrhythmia condition present | ECG signal analysis shows no pacemaker spikes or arrhythmia detected | warning |
NO PCCO-AP | Pressure sensor not connected for PiCCO hemodynamic monitoring | Pressure sensor input not detected at module initialization | error |
O2 CAL | Oxygen sensor calibration cycle in progress or calibration error detected | Calibration mode activated; sensor signal verification in progress or failed | warning |
O2 CONNECTOR OFF | Oxygen sensor cable disconnected from monitor or cable damaged | No pressure signal from oxygen sensor connection | error |
PCCO CHECK EXTERNAL DEVICE | PiCCO hemodynamic monitor disconnected or interface communication lost | Loss of communication with external PiCCO monitor unit | error |
PCCO CHECK MODE | Pressure sensor type and PiCCO module technology mismatch preventing measurement | Sensor specification does not match selected measurement module/technology | warning |
PM FAILURE | Data transmission failure between input unit and bedside/central monitor due to time zone mismatch | Time zone settings differ between transmitting and receiving monitor | warning |
PRINT ERROR | Print operation failed | Print command to network printer failed | warning |
RESP CHECK SENSOR | Respiration sensor check required | Respiration pickup damaged or disconnected | warning |
RESP CONNECTOR OFF | Respiration connector disconnected | Respiration pickup disconnected from monitor | warning |
ScvO2 AMBIENT TEMP HIGH | Ambient temperature exceeds safe operating range for CeVOX | Environmental temperature too high at module location | unknown |
ScvO2 CAL ERROR | CeVOX ScvO2 calibration failed | ScvO2 calibration procedure failed or probe error detected | warning |
ScvO2 CONNECTOR OFF | CeVOX module not properly connected | CeVOX module connector not seated or disconnected | warning |
ScvO2 LIGHT INTERFERENCE | Excessive ambient light detected by CeVOX module | Ambient light level exceeds acceptable threshold at probe site | unknown |
ScvO2 MODULE ERROR | CeVOX module internal hardware error | CeVOX module internal fault detected | error |
ScvO2 OVERHEAT | CeVOX module internal temperature exceeded | CeVOX module internal temperature exceeds safe operating limit | warning |
ScvO2 SIGNAL TOO HIGH | CeVOX signal strength excessive | Probe attached at inappropriate site for measurement | warning |
ScvO2 SIGNAL TOO LOW | CeVOX signal strength too weak | Probe signal below minimum detection threshold | warning |
ScvO2 WARMING UP | CeVOX module startup warm-up phase | CeVOX module is warming up after connection or power-on | info |
SENDING DATA | Bedside monitor transmitting data to central monitor | Input unit transferred from one bedside monitor to another in network | info |
SENSOR ERROR COMPLETE | ProAQT sensor expired or malfunction | ProAQT sensor reaches end of life or fails | warning |
SETLIB: Keycode Err | Duplicate key code detected during copying set value | During set value copy operation | warning |
SIMULATED DATA | Display showing simulated/demo patient data | <SIMULATION MODE> set to ON instead of OFF | warning |
SpO2-2 ALARM | Pulse oximeter alarm condition on SpO2-2 channel | Alarm triggered on Nellcor or Masimo pulse oximeter module | unknown |
SpO2-2 CHANGE PROBE | SpO2-2 probe deteriorated, damaged or short-circuited | Probe worn out, broken, or electrical malfunction | warning |
SpO2-2 CONNECTOR OFF | SpO2-2 probe cable disconnected from cord | Probe cable or adapter disconnected from monitor socket | warning |
SpO2-2 DETECTING PULSE | SpO2-2 searching for pulse waveform (NK/Masimo) | SpO2-2 module actively searching for pulse signal | info |
SpO2-2 DETECT PULSE | Cannot detect pulse on SpO2-2 (Nellcor/Masimo) | SpO2-2 module unable to find pulse signal | warning |
SpO2-2 LIGHT INTERFERENCE | Excessive light interference on SpO2-2 probe | Ambient light level exceeds SpO2-2 sensor threshold | unknown |
SpO2-2 LOW QUALITY SIGNAL | SpO2-2 signal quality poor (NK/Masimo) | SpO2-2 measurement signal unstable or weak | warning |
SpO2-2 MODULE ERROR | SpO2-2 hardware malfunction (NK/Masimo) | SpO2-2 module internal hardware failure detected | error |
SpO2-2 NO PROBE | SpO2-2 probe not connected to cord (NK) | Probe not attached to SpO2 connection cord | critical |
SpO2-2 WEAK PULSE | SpO2-2 weak pulse signal detection (NK/Masimo) | SpO2-2 pulse signal weak or marginal | warning |
SpO2 CHECK PROBE | SpO2 probe attachment or damage check needed (NK) | Probe not properly attached or damaged detection | warning |
SpO2 CONNECTOR OFF | SpO2 probe cable or adapter disconnected | Probe cable disconnected from adapter or adapter from socket | warning |
SpO2 DETECTING PULSE | SpO2 system searching for pulse waveform | Sensor actively detecting pulse after probe attachment | info |
SpO2 LIGHT INTERFERENCE | Excessive ambient light at SpO2 probe site | Light level exceeds acceptable threshold for SpO2 measurement | unknown |
SpO2 LOW QUALITY SIGNAL | SpO2 signal quality poor due to motion or low perfusion | SpO2 waveform unstable or signal strength low | warning |
SpO2 MODULE ERROR | SpO2 hardware malfunction | SpO2 module internal hardware failure | error |
SpO2 NO PROBE | SpO2 probe not connected to connection cord (NK) | Probe connector not attached to SpO2 cord on monitor | critical |
SpO2 PROBE | SpO2 probe identification or integrity error (Masimo) | Monitor cannot identify probe or probe damaged | error |
SpO2 WEAK PULSE | SpO2 weak pulse signal from poor circulation or tight probe | Pulse signal strength below acceptable threshold | warning |
StO2 CHECK EXTERNAL DEVICE | Cerebral/somatic oximeter disconnected or communication error | Cerebral/somatic oximeter not communicating with bedside monitor | warning |
SUSPEND MONITORING | Patient monitoring has been suspended | SUSPEND MONITORING key touched | warning |
SYSTEM DOWN DATA ABORT/WP ERROR | Unauthorized address was accessed | During system operation | critical |
SYSTEM DOWN PREFETCH ABORT | Unauthorized instruction was fetched | During system operation | critical |
SYSTEM DOWN UNDEFINED INSTRUCTION | Undefined instruction was given | During system operation | critical |
SYSTEM DOWN ZERO DIVIDE | Division by zero error | During system operation | critical |
SYSTEM ERROR | General system error with unspecified large classification | During system operation | error |
SYSTEM ERROR FILE CREATION | File cannot be properly created | During data logging or file operation | error |
SYSTEM ERROR LIBRARY RUNTIME | Library runtime error occurred | During system operation | error |
SYSTEM ERROR SD CARD TRANSFER | SD card transfer task error occurred | During memory card data transfer | error |
TB SIGNAL UNSTABLE | Unstable temperature signal in hemodynamic monitoring unit | TD curve error or baseline error due to unstable Tb signal | warning |
tcPO2/PCO2 ALARM | Transcutaneous oxygen and carbon dioxide monitor alarm | Alarm condition detected on transcutaneous monitor | warning |
tcPO2/PCO2 CHECK | Communication check failure for transcutaneous gas monitor | Disconnection or communication error between transcutaneous monitor and bedside monitor | error |
TD MEASUREMENT LIMIT | Maximum number of thermodilution measurements reached | Storage limit for TD measurements exceeded | warning |
TD TEMP SENSOR ERROR | Injectate temperature sensor malfunction on hemodynamic unit | Sensor error detected | error |
TD TIME OUT | Thermodilution measurement timeout | No injection detected within expected time window | warning |
TEC DATA RECEIVE ERROR | Communication error receiving data from defibrillator | Defibrillator data transmission failure to bedside monitor | error |
TEC INTERFACE ALREADY CONNECTED | Two defibrillator interfaces connected simultaneously | More than one QI-670P interface connected to monitor | error |
TEC INTERFACE DISCONNECTED | QI-670P defibrillator interface USB cable disconnected | USB cable not connected to monitor | error |
TEC INTERFACE ERROR | Communication error with QI-670P defibrillator interface | Communication failure between QI-670P interface and monitor | error |
TEMP CHECK | Temperature probe or connection error | Temperature sensor disconnection or malfunction | warning |
TEMP THIS LABEL IS ALREADY REGISTERED | Duplicate temperature probe label | More than one temperature probe with the same label detected | warning |
THIS PARAMETER IS ALREADY REGISTERED | Exceeds specified number of channels for a parameter | More than specified number of parameter channels are activated | warning |
TI INJECTATE TEMP HIGH | Injectate temperature too warm for thermodilution measurement | Injectate temperature warmer than blood temperature | warning |
TOF ALARM | Neuromuscular transmission monitoring device error | Error detected on NMT monitoring device | warning |
TOF CAL?? | NMT module not calibrated | Calibration not yet performed on NMT module | warning |
TOF CALERROR | NMT module calibration failure | Calibration procedure failed on NMT module | error |
TOF CHECK EXTERNAL DEVICE | NMT monitoring device disconnected from bedside monitor | NMT device communication loss with main monitor | error |
TOF CHECK SENSOR | NMT sensor cables or electrodes not connected | Main cable, stimulation cable, electrodes or acceleration transducer disconnection | error |
TOF CHECK TEMP SENSOR | NMT temperature sensor not connected | Temperature sensor disconnection detected | error |
TOF COMMUNICATION ERROR | Communication error between NMT module and bedside monitor | Communication failure with NMT module | error |
TOF LOW TEMPERATURE | Body temperature too low for NMT monitoring | Body temperature below 32°C (89.6°F) | warning |
TOF PLEASE WAIT | Insufficient interval between NMT stimulation cycles | Stimulation applied within 2 minutes of previous measurement | info |
TOF PRE PTC RESPONSE DETECTED | Vital reaction detected before post-tetanic potentiation | Patient response detected before PTC phase | warning |
TOF SIGNAL OUT OF RANGE | Temperature sensor reading outside measurable range | Temperature measurement outside 20.0 to 42.5°C range | error |
TOF SIGNAL TOO WEAK | Acceleration signal during NMT measurement too weak | Signal amplitude below threshold during measurement | warning |
TOF SUPRAMAXIMAL NOT FOUND | Supramaximal stimulation current cannot be determined | Failed to establish supramaximal current during calibration | error |
TRANSMITTER CONNECTOR OFF | Wireless transmitter cable disconnected from monitor | Transmitter connection cord not connected to monitor | error |
UPDATING DATA | Parameter update in progress on trend screen | Parameter being changed on trend screen | info |
VARIABLE PULSE RATE | SVV (Stroke Volume Variation) measurement may not be reliable | Patient pulse rate is variable or unstable during measurement | warning |
VENT ALARM | Alarm occurring on connected ventilator | Alarm condition detected on ventilator device | warning |
VENT CHECK EXTERNAL DEVICE | Ventilator disconnected or communication error | Ventilator communication loss with bedside monitor | error |
ZEROING COMPLETE | Hemodynamic pressure zero adjustment completed successfully | Zero adjustment procedure completed | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.