Nellcor Oxi-Max N-595: troubleshooting and error codes

The ServiceForge knowledge base holds 22 documented symptoms for the Nellcor Oxi-Max N-595, with 44 ranked causes, 52 diagnostic checks with test points and expected values, and 81 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Alarm does not sound patient safety

Alarm is silent when expected to sound

Ranked causes

  1. 1. Alarm silenced or disabled (common)
  2. 2. Speaker connection loose (common)
  3. 3. Speaker defective (occasional)
  4. 4. User Interface PCB defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Low indicator lights steadily on AC with battery not discharged

Battery Low LED illuminates despite AC connection and adequate battery charge

Ranked causes

  1. 1. AC outlet not operational or unplugged (common)
  2. 2. Blown fuses (common)
  3. 3. Power supply defective (occasional)
  4. 4. Harness connection fault (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery does not charge

Battery fails to charge when connected to AC power

Ranked causes

  1. 1. Battery aged beyond 2 years (common)
  2. 2. Battery unable to hold charge (common)
  3. 3. Power supply output voltage or current incorrect (occasional)
First check

Determine battery age

Test point
Battery
Expected
Battery less than 2 years old

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display pixels do not light

LCD display is dark or pixels are not illuminating

Ranked causes

  1. 1. Display PCB connection loose (common)
  2. 2. Display PCB defective (occasional)
  3. 3. User Interface PCB defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display values missing or erratic

Displayed readings are incomplete, intermittent or unstable

Ranked causes

  1. 1. Pulse oximetry cable faulty (common)
  2. 2. OXIMAX sensor damaged (common)
  3. 3. User Interface PCB defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No power-up on AC and/or DC

Monitor does not turn on when connected to AC or DC power

Ranked causes

  1. 1. Blown fuses in Power Entry Module (common)
  2. 2. Power supply defective (occasional)
  3. 3. Harness connection fault (occasional)
First check

Inspect fuses in Power Entry Module

Test point
Power Entry Module
Expected
Fuses intact and properly seated

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 or Pulse values change rapidly with erratic Pulse Amplitude

Readings fluctuate excessively and waveform indicator is unstable

Ranked causes

  1. 1. OXIMAX sensor damp or reused excessively (common)
  2. 2. Electrosurgical unit interference (occasional)
  3. 3. User Interface PCB defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm sounds for no apparent reason

Audible alarm triggers without valid cause or trigger condition

Ranked causes

  1. 1. Moisture or spilled liquids in monitor (common)
  2. 2. User Interface PCB defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Low error code with AC power disconnection

Monitor generates error when disconnected from AC power supply

Ranked causes

  1. 1. Battery discharged (common)
  2. 2. Battery defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Alarm operation and silence function patient safety

Verify audible and visual alarm activation, silence button functionality, and timeout behavior

Ranked causes

  1. 1. Alarm system requires functional verification (common)
  2. 2. Alarm volume requires adjustment capability (occasional)
First check

Plethysmograph waveform tracking

Test point
Pleth display
Expected
Waveform tracks user pulse rate

12 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Power-on alarm limits patient safety

Verify and adjust default alarm limits for SpO2, pulse rate, and SatSeconds parameters

Ranked causes

  1. 1. Alarm limits require verification after repair or maintenance (common)
  2. 2. Factory defaults require confirmation after power cycle (common)
First check

SpO2 upper alarm limit minimum

Test point
LIMITS display
Expected
Minimum value 86 %

9 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not respond to button presses

Buttons pressed do not trigger any response or function

Ranked causes

  1. 1. User Interface PCB defective (common)
  2. 2. Top case assembly defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pulse Amplitude indicator shows pulse but digital display shows zeroes

Pulse waveform indicator active but no numerical readings displayed

Ranked causes

  1. 1. OXIMAX sensor damaged (common)
  2. 2. User Interface PCB defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charge verification

Charging battery to full capacity with monitor off or on

Ranked causes

  1. 1. Battery requires charging before use (common)
First check

Connect to AC power

Test point
AC power inlet
Expected
Monitor connected to AC power source

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Low alarm during DC operation with 15 minutes charge remaining caution

Low battery alarm sounds indicating 15 or fewer minutes of usable charge left

Ranked causes

  1. 1. Normal battery depletion at end of charge cycle (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Analog output 1 volt BPM and Pleth

Verify analog BPM and Pleth output voltage at data port pins 13 and 14

Ranked causes

  1. 1. Analog BPM and Pleth outputs require calibration verification (common)
First check

Analog BPM and Pleth output voltage at pins 13 and 14

Test point
Data port pins 10 (negative reference) and 13, 14 (positive)
Expected
+1.0 VDC, ±0.025 VDC
Tools
Digital voltmeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Analog output 1 volt SpO2

Verify analog SpO2 output voltage at data port pin 6 relative to pin 10

Ranked causes

  1. 1. Analog SpO2 output requires calibration verification (common)
First check

Connect voltmeter to analog output

Test point
Data port connector pins 10 (negative) and 6 (positive)
Expected
Voltmeter connected with correct polarity

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Nurse call function RS-232

Verify RS-232 nurse call voltage output at data port pins 5 and 11

Ranked causes

  1. 1. RS-232 nurse call output requires verification (common)
First check

Connect voltmeter to data port

Test point
Data port connector pins 5 (negative) and 11 (positive)
Expected
Voltmeter connected with correct polarity

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Nurse call function solid state relay

Verify solid state relay nurse call output continuity between data port pins

Ranked causes

  1. 1. Solid state nurse call relay requires functional verification (common)
First check

Solid state relay pins 8-15 in alarm state

Test point
Data port pins 8 and 15
Expected
No continuity (≥1 megohm), ≥1 megohm
Tools
Digital voltmeter (DVM)

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pulse tone volume control

Verify pulse tone volume adjustment and audibility range

Ranked causes

  1. 1. Pulse tone volume requires adjustment capability (common)
First check

Pulse tone volume increase

Test point
Monitor speaker
Expected
Pulse tone volume increases with ADJUST UP

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power-on self-test verification

Monitor LEDs, LCD pixels, backlight, speaker, and indicators during startup sequence

Ranked causes

  1. 1. Normal operation expected (common)
First check

All LEDs illuminate

Test point
Monitor front panel
Expected
All LEDs illuminated within 2 seconds of power-on

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on high voltage

Monitor will not turn on or off; possible blown fuse in power module

Ranked causes

  1. 1. Blown fuse in power module (common)
First check

Inspect fuses in power module

Test point
Fuse drawer in power entry module
Expected
Fuses intact and rated 5x20-mm slow blow 0.5-amp 250-volt

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1SpO2 front end RAM errorNon-correctable failure detectedcritical
10Over-current limit in SpO2 front end has trippedNon-correctable failure detectederror
11SpO2 front end reports incorrect system voltageNon-correctable failure detectederror
12SpO2 front end reports other hardware problemNon-correctable failure detectederror
14SpO2 front end reports communication channel overflowNon-correctable failure detectederror
16SpO2 front end reports watch dog time outNon-correctable failure detectedcritical
17SpO2 front end reports that sensor appears defectiveNon-correctable failure detectederror
18SpO2 front end reports internal register appears modified from expected valueNon-correctable failure detectederror
19SpO2 front end reports signal out-of-rangeNon-correctable failure detectedwarning
2SpO2 front end ROM/code integrity errorNon-correctable failure detectedcritical
256SpO2 back end reports beginning of packet missingNon-correctable failure detectederror
257SpO2 back end reports packet start ID (SID) missingNon-correctable failure detectederror
258SpO2 back end reports packet length errorNon-correctable failure detectederror
259SpO2 back end reports message length errorNon-correctable failure detectederror
260SpO2 back end reports packet contains unsupported KeyNon-correctable failure detectedwarning
261SpO2 back end reports packet CRC errorNon-correctable failure detectederror
262SpO2 back end reports end of packet missingNon-correctable failure detectederror
263SpO2 back end reports packet contains undefined keyNon-correctable failure detectederror
264SpO2 back end reports corrupted variableNon-correctable failure detectederror
265SpO2 back end reports memory overflowNon-correctable failure detectederror
266SpO2 back end reports bad pointerNon-correctable failure detectederror
267SpO2 back end reports parameter value out-of-rangeNon-correctable failure detectederror
268SpO2 back end reports reset detectedNon-correctable failure detectedwarning
269SpO2 back end reports unexpected valueNon-correctable failure detectedwarning
270SpO2 back end reports time-outNon-correctable failure detectederror
271SpO2 back end reports not ready/not initializedNon-correctable failure detectederror
272SpO2 back end reports double faultNon-correctable failure detectedcritical
273SpO2 back end reports date out-of-range errorNon-correctable failure detectedwarning
274SpO2 back end reports incompatible software versionNon-correctable failure detectederror
275SpO2 back end reports incorrect registration numberNon-correctable failure detectedwarning
276SpO2 back end reports sensor read failureNon-correctable failure detectederror
277SpO2 back end reports sensor signature verification failsNon-correctable failure detectederror
278SpO2 back end reports warmed sensor temperature set point failureNon-correctable failure detectederror
279SpO2 back end reports warmed sensor/SpO2 front end incompatibleNon-correctable failure detectederror
280SpO2 back end reports does not support feature required by sensorNon-correctable failure detectedwarning
281SpO2 back end reports overflow/underflowNon-correctable failure detectederror
282SpO2 back end reports sensor activation failureNon-correctable failure detectederror
3SpO2 front end reported a bad CRCNon-correctable failure detectederror
4SpO2 front end reported FSP message not allowedNon-correctable failure detectederror
48SpO2 front end reports spurious interruptNon-correctable failure detectedwarning
49SpO2 front end reports internal buffer overflowNon-correctable failure detectederror
5SpO2 front end reported illegal value sent in FSP messageNon-correctable failure detectederror
50SpO2 front end reports intermittent errorNon-correctable failure detectedwarning
51SpO2 front end reports digital communications errorNon-correctable failure detectederror
512General failure of UIF Module generic post / Dead battery/Missing batteryNon-correctable failure detectedcritical
514Real time clock is non-operationalNon-correctable failure detectederror
515Application code is not present in the flashNon-correctable failure detectedcritical
516Invalid flash typeNon-correctable failure detectedcritical
517Serial clock line is not toggling or is toggling at an incorrect rateNon-correctable failure detectederror
518Application program is corruptNon-correctable failure detectedcritical
519Invalid FE102 versionNon-correctable failure detectederror
52SpO2 front end reports warmer errorNon-correctable failure detectederror
520Error in the start up sequenceNon-correctable failure detectederror
521OS multitasking service failureNon-correctable failure detectederror
522A state machine has received an unknown state transitionNon-correctable failure detectederror
523Operation not completed successfullyUser attempted operation failedwarning
524Unexpected value receivedOut-of-range parameter passed to functionwarning
525EEPROM CRC failureNon-correctable failure detectederror
526SpO2 module not respondedNon-correctable failure detectederror
527Institutional parameters lostEEPROM section CRC corruptwarning
528Current settings lostEEPROM section CRC corruptwarning
529Critical low batteryBattery voltage critically lowcritical
53Front end data not receivedNon-correctable failure detectederror
530Low battery errorBattery voltage below acceptable thresholdwarning
531External watchdog failureNon-correctable failure detectederror
532Power PC watchdog failureNon-correctable failure detectedcritical
533Boot NVROM uninitialized errorNon-correctable failure detectederror
534Failed CRC check of application code in flashNon-correctable failure detectedcritical
535Failed periodic ram CRC check on application code running in RAMNon-correctable failure detectederror
562SpO2 front end resetFront end module reset detectedwarning
563SpO2 reported errorSpO2 module reported error conditionerror
564Clinical mode was exited after input was receivedUser exited clinical mode during operationinfo
565Communication failures between software modulesModule communication link failederror
566Excessive resets before UIF runsMultiple resets occur during startuperror
567Unexpected interrupt has been assertedInterrupt from unidentified sourcewarning
568General failure in UIF module generic postNon-correctable failure detectederror
569BOOT application program is corrupt - CRC does not matchNon-correctable failure detectedcritical
570RTC was restartedReal-time clock reinitializedinfo
574Excessive restarts within 1 minuteDevice restarted multiple times rapidlyerror
6SpO2 front end reports calibration (offset) failureNon-correctable failure detectederror
9SpO2 front end reported syntax error in FSP messageNon-correctable failure detectederror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI