Nellcor Oxi-Max N-595: troubleshooting and error codes
The ServiceForge knowledge base holds 22 documented symptoms for the Nellcor Oxi-Max N-595, with 44 ranked causes, 52 diagnostic checks with test points and expected values, and 81 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Alarm does not sound patient safety
Alarm is silent when expected to sound
Ranked causes
- 1. Alarm silenced or disabled (common)
- 2. Speaker connection loose (common)
- 3. Speaker defective (occasional)
- 4. User Interface PCB defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Low indicator lights steadily on AC with battery not discharged
Battery Low LED illuminates despite AC connection and adequate battery charge
Ranked causes
- 1. AC outlet not operational or unplugged (common)
- 2. Blown fuses (common)
- 3. Power supply defective (occasional)
- 4. Harness connection fault (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge
Battery fails to charge when connected to AC power
Ranked causes
- 1. Battery aged beyond 2 years (common)
- 2. Battery unable to hold charge (common)
- 3. Power supply output voltage or current incorrect (occasional)
Determine battery age
- Test point
- Battery
- Expected
- Battery less than 2 years old
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display pixels do not light
LCD display is dark or pixels are not illuminating
Ranked causes
- 1. Display PCB connection loose (common)
- 2. Display PCB defective (occasional)
- 3. User Interface PCB defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display values missing or erratic
Displayed readings are incomplete, intermittent or unstable
Ranked causes
- 1. Pulse oximetry cable faulty (common)
- 2. OXIMAX sensor damaged (common)
- 3. User Interface PCB defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power-up on AC and/or DC
Monitor does not turn on when connected to AC or DC power
Ranked causes
- 1. Blown fuses in Power Entry Module (common)
- 2. Power supply defective (occasional)
- 3. Harness connection fault (occasional)
Inspect fuses in Power Entry Module
- Test point
- Power Entry Module
- Expected
- Fuses intact and properly seated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 or Pulse values change rapidly with erratic Pulse Amplitude
Readings fluctuate excessively and waveform indicator is unstable
Ranked causes
- 1. OXIMAX sensor damp or reused excessively (common)
- 2. Electrosurgical unit interference (occasional)
- 3. User Interface PCB defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sounds for no apparent reason
Audible alarm triggers without valid cause or trigger condition
Ranked causes
- 1. Moisture or spilled liquids in monitor (common)
- 2. User Interface PCB defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Low error code with AC power disconnection
Monitor generates error when disconnected from AC power supply
Ranked causes
- 1. Battery discharged (common)
- 2. Battery defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Alarm operation and silence function patient safety
Verify audible and visual alarm activation, silence button functionality, and timeout behavior
Ranked causes
- 1. Alarm system requires functional verification (common)
- 2. Alarm volume requires adjustment capability (occasional)
Plethysmograph waveform tracking
- Test point
- Pleth display
- Expected
- Waveform tracks user pulse rate
12 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power-on alarm limits patient safety
Verify and adjust default alarm limits for SpO2, pulse rate, and SatSeconds parameters
Ranked causes
- 1. Alarm limits require verification after repair or maintenance (common)
- 2. Factory defaults require confirmation after power cycle (common)
SpO2 upper alarm limit minimum
- Test point
- LIMITS display
- Expected
- Minimum value 86 %
9 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to button presses
Buttons pressed do not trigger any response or function
Ranked causes
- 1. User Interface PCB defective (common)
- 2. Top case assembly defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse Amplitude indicator shows pulse but digital display shows zeroes
Pulse waveform indicator active but no numerical readings displayed
Ranked causes
- 1. OXIMAX sensor damaged (common)
- 2. User Interface PCB defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge verification
Charging battery to full capacity with monitor off or on
Ranked causes
- 1. Battery requires charging before use (common)
Connect to AC power
- Test point
- AC power inlet
- Expected
- Monitor connected to AC power source
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Low alarm during DC operation with 15 minutes charge remaining caution
Low battery alarm sounds indicating 15 or fewer minutes of usable charge left
Ranked causes
- 1. Normal battery depletion at end of charge cycle (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Analog output 1 volt BPM and Pleth
Verify analog BPM and Pleth output voltage at data port pins 13 and 14
Ranked causes
- 1. Analog BPM and Pleth outputs require calibration verification (common)
Analog BPM and Pleth output voltage at pins 13 and 14
- Test point
- Data port pins 10 (negative reference) and 13, 14 (positive)
- Expected
- +1.0 VDC, ±0.025 VDC
- Tools
- Digital voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Analog output 1 volt SpO2
Verify analog SpO2 output voltage at data port pin 6 relative to pin 10
Ranked causes
- 1. Analog SpO2 output requires calibration verification (common)
Connect voltmeter to analog output
- Test point
- Data port connector pins 10 (negative) and 6 (positive)
- Expected
- Voltmeter connected with correct polarity
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse call function RS-232
Verify RS-232 nurse call voltage output at data port pins 5 and 11
Ranked causes
- 1. RS-232 nurse call output requires verification (common)
Connect voltmeter to data port
- Test point
- Data port connector pins 5 (negative) and 11 (positive)
- Expected
- Voltmeter connected with correct polarity
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Nurse call function solid state relay
Verify solid state relay nurse call output continuity between data port pins
Ranked causes
- 1. Solid state nurse call relay requires functional verification (common)
Solid state relay pins 8-15 in alarm state
- Test point
- Data port pins 8 and 15
- Expected
- No continuity (≥1 megohm), ≥1 megohm
- Tools
- Digital voltmeter (DVM)
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pulse tone volume control
Verify pulse tone volume adjustment and audibility range
Ranked causes
- 1. Pulse tone volume requires adjustment capability (common)
Pulse tone volume increase
- Test point
- Monitor speaker
- Expected
- Pulse tone volume increases with ADJUST UP
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test verification
Monitor LEDs, LCD pixels, backlight, speaker, and indicators during startup sequence
Ranked causes
- 1. Normal operation expected (common)
All LEDs illuminate
- Test point
- Monitor front panel
- Expected
- All LEDs illuminated within 2 seconds of power-on
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
Monitor will not turn on or off; possible blown fuse in power module
Ranked causes
- 1. Blown fuse in power module (common)
Inspect fuses in power module
- Test point
- Fuse drawer in power entry module
- Expected
- Fuses intact and rated 5x20-mm slow blow 0.5-amp 250-volt
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | SpO2 front end RAM error | Non-correctable failure detected | critical |
10 | Over-current limit in SpO2 front end has tripped | Non-correctable failure detected | error |
11 | SpO2 front end reports incorrect system voltage | Non-correctable failure detected | error |
12 | SpO2 front end reports other hardware problem | Non-correctable failure detected | error |
14 | SpO2 front end reports communication channel overflow | Non-correctable failure detected | error |
16 | SpO2 front end reports watch dog time out | Non-correctable failure detected | critical |
17 | SpO2 front end reports that sensor appears defective | Non-correctable failure detected | error |
18 | SpO2 front end reports internal register appears modified from expected value | Non-correctable failure detected | error |
19 | SpO2 front end reports signal out-of-range | Non-correctable failure detected | warning |
2 | SpO2 front end ROM/code integrity error | Non-correctable failure detected | critical |
256 | SpO2 back end reports beginning of packet missing | Non-correctable failure detected | error |
257 | SpO2 back end reports packet start ID (SID) missing | Non-correctable failure detected | error |
258 | SpO2 back end reports packet length error | Non-correctable failure detected | error |
259 | SpO2 back end reports message length error | Non-correctable failure detected | error |
260 | SpO2 back end reports packet contains unsupported Key | Non-correctable failure detected | warning |
261 | SpO2 back end reports packet CRC error | Non-correctable failure detected | error |
262 | SpO2 back end reports end of packet missing | Non-correctable failure detected | error |
263 | SpO2 back end reports packet contains undefined key | Non-correctable failure detected | error |
264 | SpO2 back end reports corrupted variable | Non-correctable failure detected | error |
265 | SpO2 back end reports memory overflow | Non-correctable failure detected | error |
266 | SpO2 back end reports bad pointer | Non-correctable failure detected | error |
267 | SpO2 back end reports parameter value out-of-range | Non-correctable failure detected | error |
268 | SpO2 back end reports reset detected | Non-correctable failure detected | warning |
269 | SpO2 back end reports unexpected value | Non-correctable failure detected | warning |
270 | SpO2 back end reports time-out | Non-correctable failure detected | error |
271 | SpO2 back end reports not ready/not initialized | Non-correctable failure detected | error |
272 | SpO2 back end reports double fault | Non-correctable failure detected | critical |
273 | SpO2 back end reports date out-of-range error | Non-correctable failure detected | warning |
274 | SpO2 back end reports incompatible software version | Non-correctable failure detected | error |
275 | SpO2 back end reports incorrect registration number | Non-correctable failure detected | warning |
276 | SpO2 back end reports sensor read failure | Non-correctable failure detected | error |
277 | SpO2 back end reports sensor signature verification fails | Non-correctable failure detected | error |
278 | SpO2 back end reports warmed sensor temperature set point failure | Non-correctable failure detected | error |
279 | SpO2 back end reports warmed sensor/SpO2 front end incompatible | Non-correctable failure detected | error |
280 | SpO2 back end reports does not support feature required by sensor | Non-correctable failure detected | warning |
281 | SpO2 back end reports overflow/underflow | Non-correctable failure detected | error |
282 | SpO2 back end reports sensor activation failure | Non-correctable failure detected | error |
3 | SpO2 front end reported a bad CRC | Non-correctable failure detected | error |
4 | SpO2 front end reported FSP message not allowed | Non-correctable failure detected | error |
48 | SpO2 front end reports spurious interrupt | Non-correctable failure detected | warning |
49 | SpO2 front end reports internal buffer overflow | Non-correctable failure detected | error |
5 | SpO2 front end reported illegal value sent in FSP message | Non-correctable failure detected | error |
50 | SpO2 front end reports intermittent error | Non-correctable failure detected | warning |
51 | SpO2 front end reports digital communications error | Non-correctable failure detected | error |
512 | General failure of UIF Module generic post / Dead battery/Missing battery | Non-correctable failure detected | critical |
514 | Real time clock is non-operational | Non-correctable failure detected | error |
515 | Application code is not present in the flash | Non-correctable failure detected | critical |
516 | Invalid flash type | Non-correctable failure detected | critical |
517 | Serial clock line is not toggling or is toggling at an incorrect rate | Non-correctable failure detected | error |
518 | Application program is corrupt | Non-correctable failure detected | critical |
519 | Invalid FE102 version | Non-correctable failure detected | error |
52 | SpO2 front end reports warmer error | Non-correctable failure detected | error |
520 | Error in the start up sequence | Non-correctable failure detected | error |
521 | OS multitasking service failure | Non-correctable failure detected | error |
522 | A state machine has received an unknown state transition | Non-correctable failure detected | error |
523 | Operation not completed successfully | User attempted operation failed | warning |
524 | Unexpected value received | Out-of-range parameter passed to function | warning |
525 | EEPROM CRC failure | Non-correctable failure detected | error |
526 | SpO2 module not responded | Non-correctable failure detected | error |
527 | Institutional parameters lost | EEPROM section CRC corrupt | warning |
528 | Current settings lost | EEPROM section CRC corrupt | warning |
529 | Critical low battery | Battery voltage critically low | critical |
53 | Front end data not received | Non-correctable failure detected | error |
530 | Low battery error | Battery voltage below acceptable threshold | warning |
531 | External watchdog failure | Non-correctable failure detected | error |
532 | Power PC watchdog failure | Non-correctable failure detected | critical |
533 | Boot NVROM uninitialized error | Non-correctable failure detected | error |
534 | Failed CRC check of application code in flash | Non-correctable failure detected | critical |
535 | Failed periodic ram CRC check on application code running in RAM | Non-correctable failure detected | error |
562 | SpO2 front end reset | Front end module reset detected | warning |
563 | SpO2 reported error | SpO2 module reported error condition | error |
564 | Clinical mode was exited after input was received | User exited clinical mode during operation | info |
565 | Communication failures between software modules | Module communication link failed | error |
566 | Excessive resets before UIF runs | Multiple resets occur during startup | error |
567 | Unexpected interrupt has been asserted | Interrupt from unidentified source | warning |
568 | General failure in UIF module generic post | Non-correctable failure detected | error |
569 | BOOT application program is corrupt - CRC does not match | Non-correctable failure detected | critical |
570 | RTC was restarted | Real-time clock reinitialized | info |
574 | Excessive restarts within 1 minute | Device restarted multiple times rapidly | error |
6 | SpO2 front end reports calibration (offset) failure | Non-correctable failure detected | error |
9 | SpO2 front end reported syntax error in FSP message | Non-correctable failure detected | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI