Nellcor N-600: troubleshooting and error codes
The ServiceForge knowledge base holds 16 documented symptoms for the Nellcor N-600, with 36 ranked causes, 8 diagnostic checks with test points and expected values, and 155 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Alarm does not sound high voltage
No alarm tone when alert condition occurs
Ranked causes
- 1. Alarm silenced (common)
- 2. Speaker connection loose (common)
- 3. Speaker defective (occasional)
- 4. Main Board PCB defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Battery indicator lights on AC power
Low battery indicator illuminates steadily while connected to AC outlet with adequate charge
Ranked causes
- 1. Blown or open fuse in Power Entry Module (common)
- 2. Defective power supply (occasional)
- 3. Battery defective or aged (occasional)
- 4. Loose or corroded harness connection on Main Board (occasional)
Verify unit plugged into operational AC outlet
- Expected
- AC indicator lit
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge
Battery fails to accept charge when connected to AC power
Ranked causes
- 1. Battery aged over 2 years (common)
- 2. Battery defective (common)
- 3. Defective power supply (occasional)
Verify charging voltage on new battery
- Test point
- Battery terminals while on AC power
- Expected
- 7.05-7.35 volts initial climb, then decrease with time V, +0.30 -0.15 V
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display pixels do not light
Display screen shows no illuminated pixels
Ranked causes
- 1. Connection issue between Main and Display PCB (common)
- 2. Display PCB defective (occasional)
- 3. Main Board PCB defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display values missing or erratic
Display shows missing data or values fluctuate erratically
Ranked causes
- 1. OXIMAX sensor issue (common)
- 2. OXIMAX sensor defective (common)
- 3. Main Board PCB defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No printout received from data port
Monitor fails to send data to connected printer via data port
Ranked causes
- 1. Printer not operational (common)
- 2. Baud rate mismatch between monitor and printer (common)
- 3. Main Board PCB defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 or Pulse values change rapidly with erratic amplitude patient safety
Measured values fluctuate rapidly; Pulse Amplitude indicator shows erratic behavior
Ranked causes
- 1. OXIMAX sensor damp or reused excessively (common)
- 2. Electrosurgical unit interference (occasional)
- 3. Main Board PCB defective (occasional)
Verify performance per Performance Verification procedure
- Expected
- Normal SpO2 and Pulse values within expected range
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sounds without apparent reason
Alarm generates unexpectedly with no apparent trigger
Ranked causes
- 1. Moisture or liquid contamination (common)
- 2. Main Board PCB defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buttons do not respond
Monitor does not respond when buttons are pressed while powered on
Ranked causes
- 1. Main Board PCB defective (common)
- 2. Top enclosure assembly defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse Amplitude indicator shows pulse but displays zero patient safety
Pulse Amplitude indicator responds to patient signal but SpO2/Pulse values show zero
Ranked causes
- 1. OXIMAX sensor damaged (common)
- 2. Main Board PCB defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS-232 nurse call inoperable
Nurse call signal over RS-232 interface not functioning
Ranked causes
- 1. Connection issue on data port (common)
- 2. Main Board PCB defective (occasional)
Measure voltage between pins 5 and 11 of data port
- Test point
- Data port pins 5 (GND) and 11 (nurse call)
- Expected
- -5 to -12 VDC during normal operation; +5 to +12 VDC during alarm, +3.5 -3.5 V (no alarm)
- Tools
- DC Voltmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display PCB requires replacement caution
Display PCB removal and reinstallation procedure for monitor maintenance or repair
Ranked causes
- 1. Display PCB damaged or defective (unknown)
Verify CCFL harness disconnected from J4 on Main Board PCB
- Test point
- J4
- Expected
- CCFL harness removed
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code on disconnect from AC
Unit generates error code when powered from battery after AC disconnection
Ranked causes
- 1. Battery discharged or defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low Battery alarm during DC operation
Low Battery indicator illuminates and alarm sounds during battery-powered operation
Ranked causes
- 1. Battery nearing depletion (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary speaker fails with high-pitched tone
Primary speaker failure resulting in slow, high-pitched audible tone
Ranked causes
- 1. Primary speaker malfunction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary speaker failure detected
Monitor displays 'PRIMARY SPKR FAILURE: NOTIFY SVC PERSONNEL' message with high-pitched, slow-pulsing piezo tone. Once silenced, tone repeats every three minutes.
Ranked causes
- 1. Primary speaker hardware failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | %SpO2 front end RAM error | RAM error detected during operation | error |
10 | Over-current limit in %SpO2 front end has tripped | Current draw exceeds maximum threshold | error |
11 | %SpO2 front end reports incorrect system voltage | Supply voltage outside acceptable range | error |
12 | %SpO2 front end reports other hardware problem | Unspecified hardware failure detected | error |
14 | %SpO2 front end reports communication channel overflow | Receive buffer overflows during message reception | warning |
16 | %SpO2 front end reports watch dog time out | Watchdog timer expires; front end does not respond within timeout period | critical |
17 | %SpO2 front end reports that sensor appears defective | Sensor communication fails or sensor reports invalid data | error |
18 | %SpO2 front end reports internal register appears modified from expected value | Register verification detects unexpected value change | error |
19 | %SpO2 front end reports signal out-of-range | Sensor signal amplitude or frequency outside valid range | warning |
2 | %SpO2 front end ROM/code integrity error | ROM code checksum or integrity verification fails | error |
256 | %SpO2 back end reports beginning of packet missing | Packet start marker not found in expected location | error |
257 | %SpO2 back end reports packet start ID (SID) missing | Start ID field not found in packet header | error |
258 | %SpO2 back end reports packet length error | Declared packet length does not match actual received length | error |
259 | %SpO2 back end reports message length error | Message length field invalid or inconsistent | error |
260 | %SpO2 back end reports packet contains unsupported key | Packet contains a command or key not recognized by back end | warning |
261 | %SpO2 back end reports packet CRC error | Packet checksum does not match calculated CRC | error |
262 | %SpO2 back end reports end of packet missing | Packet terminator not found where expected | error |
263 | %SpO2 back end reports packet contains undefined key | Packet contains unknown command or identifier | warning |
264 | %SpO2 back end reports corrupted variable | Variable value fails integrity check or is out of range | error |
265 | %SpO2 back end reports memory overflow | Memory allocation request exceeds available space | error |
266 | %SpO2 back end reports bad pointer | Invalid pointer value detected during dereference | error |
267 | %SpO2 back end reports parameter value out-of-range | Parameter value exceeds valid bounds | warning |
268 | %SpO2 back end reports reset detected | Unexpected system reset or reboot detected during operation | warning |
269 | %SpO2 back end reports unexpected value | Received value does not match expected result or range | warning |
270 | %SpO2 back end reports time-out | Operation does not complete within timeout period | error |
271 | %SpO2 back end reports not ready/not initialized | Module or subsystem not yet initialized when accessed | error |
272 | %SpO2 back end reports double fault | Two faults detected simultaneously or fault during fault handling | critical |
273 | %SpO2 back end reports date out-of-range error | Date value is invalid or outside acceptable range | warning |
274 | %SpO2 back end reports incompatible software version | Software component version mismatch detected | error |
275 | %SpO2 back end reports incorrect registration number | Registration or license number verification fails | warning |
276 | %SpO2 back end reports sensor read failure | Unable to read sensor data or sensor communication fails | error |
277 | %SpO2 back end reports sensor signature verification fails | Sensor identification or authentication check fails | error |
278 | %SpO2 back end reports warmed sensor temperature set point failure | Temperature control for warmed sensor cannot reach or maintain setpoint | error |
279 | %SpO2 back end reports warmed sensor / %SpO2 front end incompatible | Warmed sensor and front end firmware/hardware versions do not match | error |
280 | %SpO2 back end reports does not support feature required by sensor | Sensor requires a feature not implemented in current firmware | warning |
281 | %SpO2 back end reports overflow/underflow | Numerical value exceeds maximum or falls below minimum | error |
282 | %SpO2 back end reports sensor activation failure | Sensor cannot be activated or powered on | error |
283 | %SpO2 back end reports sensor write failure | Unable to write configuration or data to sensor | error |
284 | ECG trigger error | ECG trigger signal absent, invalid, or out of range | error |
285 | Sensor trend not open | Request to access sensor trend data but trend not opened | warning |
286 | Sensor trend already open | Attempt to open trend data that is already open | warning |
287 | Sensor trend data unavailable | Trend data requested but not available or accessible | warning |
288 | All sensor trend data read | All available trend data has been read; end of trend reached | info |
289 | Incompatible private label | Private label configuration mismatch or unsupported variant | warning |
3 | %SpO2 front end reported bad Cyclic Redundancy Check (CRC) | CRC check fails during data transmission or storage verification | error |
4 | %SpO2 front end reported FSP message not allowed | Illegal FSP message received by front end | warning |
48 | %SpO2 front end reports spurious interrupt | Unexpected interrupt asserted to front end CPU | warning |
49 | %SpO2 front end reports internal buffer overflow | Internal message buffer exceeds capacity | error |
5 | %SpO2 front end reported illegal value sent in FSP message | Out-of-range or invalid parameter value in FSP message | warning |
50 | %SpO2 front end reports intermittent error | Intermittent failure detected during operation | warning |
51 | %SpO2 front end reports digital communications error | Communication protocol error or checksum failure | error |
512 | General failure of UIF Module generic post or Dead battery/Missing battery | UIF module POST fails or battery missing/disconnected | critical |
514 | Real time clock is non-operational | RTC fails to respond or is not functioning | error |
515 | Application code is not present in the Flash | Flash memory does not contain valid application code | critical |
516 | Invalid Flash type | Flash memory type does not match expected configuration | critical |
517 | Serial clock line is not toggling or is toggling at an incorrect rate | Serial clock signal missing or frequency incorrect | critical |
518 | Application program is corrupt | Application code CRC check fails | critical |
519 | Invalid Nell-1A version | Nell-1A firmware version not recognized or too old | error |
52 | %SpO2 front end reports warmer error | Warmer/heater control failure or temperature sensor error | error |
520 | Error in the start up sequence | Startup initialization fails at some point | critical |
521 | OS multi-tasking service failure | Real-time operating system fails to manage tasks | critical |
522 | A state machine has received an unknown state transition | State machine receives unexpected or invalid transition | error |
523 | The operation just attempted was not completed successfully - for example, Institutional Defaults could not be reset | Requested operation fails to complete or returns error | error |
524 | An unexpected value was received - for example, an out-of-range parameter was passed to a function | Received value outside valid range or not as expected | warning |
525 | EEPROM CRC failure | EEPROM checksum verification fails | error |
526 | %SpO2 module not responded | No response from %SpO2 module within timeout period | critical |
527 | Institutional parameters lost - e.g. for UIF: Institutional EEPROM section CRC corrupt | Institutional parameters cannot be read or CRC fails | error |
528 | Current settings lost - e.g. for UIF: Institutional EEPROM section CRC corrupt | Current settings data is corrupted or cannot be read | error |
529 | Critical low battery | Battery voltage drops below critical threshold | critical |
53 | Front end data not received | No data packet received from front end within timeout period | critical |
530 | Low battery error | Battery voltage drops below normal operating threshold | warning |
531 | External watchdog failure | External watchdog timer fails to function | critical |
532 | Power PC watchdog failure | PowerPC watchdog timer fails or does not reset | critical |
533 | Boot NVROM uninitialized error | Boot ROM/NVRAM is not properly initialized | critical |
534 | Failed CRC check of application code in flash | Application code in flash fails CRC verification | critical |
535 | Failed periodic ram CRC check on application code running in RAM | Running application code fails CRC check periodically | critical |
560 | Memory corruption detected | Memory integrity check detects corruption | error |
561 | RTOS Resource unavailable | Real-time OS resource (semaphore, queue, etc.) not available | error |
562 | %SpO2 front end reset | %SpO2 front end has reset or rebooted unexpectedly | warning |
563 | %SpO2 reported error | %SpO2 module reports an error condition | error |
564 | Clinical mode was exited after input was received | Clinical mode exits unexpectedly after receiving input | warning |
565 | Communication failures between software modules | Inter-module communication fails or times out | error |
566 | Excessive resets before UIF runs | System resets repeatedly before UIF module starts | critical |
567 | An unexpected interrupt has been asserted | Interrupt occurs when none was expected | warning |
568 | General failure in UIF module generic post | UIF module power-on self-test fails | critical |
569 | BOOT application program is corrupt - CRC does not match | Boot loader code fails CRC verification | critical |
570 | RTC was restarted | Real-time clock has been restarted or lost time | info |
574 | Excessive restarts within one minute | System restarts too many times in a short time period | critical |
575 | Trend data corrupted | Trend data fails integrity check | warning |
576 | Logic Processing Board (LPS) is going through excessive resets | LPS module resets repeatedly | error |
577 | LPS status messages not arriving in a timely manner | LPS communication lags or status messages delayed | warning |
578 | LPS driver has received too many corrupt (protocol error) messages | LPS driver detects excessive protocol errors from LPS | error |
579 | Status report AC OK bit does not equal the hardware AC OK signal | Software AC status does not match hardware signal | warning |
580 | LPS driver received only a partial LPS message | LPS message reception incomplete or truncated | error |
581 | Too many corrupt (invalid data) message | Excessive number of messages with invalid data received | error |
582 | Unable to enter shelf-mode, this could mean that the LPS is dead or there is a serial port transmit problem | Shelf mode transition fails | critical |
583 | Primary speaker failure detected | Speaker does not respond or test fails | warning |
6 | %SpO2 front end reports calibration (offset) failure | Calibration or offset adjustment fails during startup or operation | error |
650 | Oximetry module reports internal software consistency check failed | Oximetry module software integrity verification fails | error |
651 | Oximetry module reports software functions executed before initialization completed | Oximetry function called before module ready | error |
652 | Oximetry module reports internal memory buffer overflow | Oximetry module internal buffer exceeds capacity | error |
653 | Oximetry module reports host reset | Oximetry module detected unexpected host reset | warning |
654 | Oximetry module ROM/Code integrity error | Oximetry module firmware fails integrity check | error |
655 | Oximetry module RAM error | Oximetry module RAM test fails | error |
656 | Oximetry module reports background self-test failed to complete in the allocated time | Oximetry self-test does not finish within timeout | warning |
657 | Oximetry module reports a state machine is in an unexpected state | Oximetry state machine in invalid state | error |
658 | Oximetry module reports memory corruption detected | Oximetry module memory integrity check fails | error |
659 | Oximetry module reports spurious interrupt detected | Oximetry module receives unexpected interrupt | warning |
660 | Oximetry module reports unable to issue commands to %SpO2 front end | Oximetry cannot communicate with %SpO2 front end | error |
661 | Oximetry module reports unable to write to external Flash | Oximetry cannot write data to external flash memory | error |
662 | Oximetry module reports communication with %SpO2 front end lost | Oximetry loses communication with %SpO2 front end | critical |
663 | Oximetry module reports internal register modified from expected value | Register verification detects unexpected value change | error |
701 | POST external hardware failsafe circuit error | External hardware failsafe circuit test fails | critical |
702 | POST PWR_ENA fail | Power enable circuit test fails during POST | critical |
703 | Unknown state machine # or charge state | Invalid state machine number or charge state detected | error |
704 | Unexpected interrupt | Interrupt asserted when none expected | warning |
705 | Stack over/under error | Stack pointer exceeds bounds or underflows | error |
706 | ROM CRC error - POST or background | ROM code fails checksum verification | critical |
707 | POST RAM error | RAM test fails during power-on self-test | critical |
708 | Unexpected reset | System resets unexpectedly during operation | error |
709 | Unexpected internal WD instance #1 | Internal watchdog #1 times out unexpectedly | error |
710 | Unexpected internal WD instance #2 | Internal watchdog #2 times out unexpectedly | error |
711 | Unexpected internal WD instance #3 | Internal watchdog #3 times out unexpectedly | error |
712 | Unexpected internal WD instance #4 | Internal watchdog #4 times out unexpectedly | error |
713 | Unexpected internal WD instance #5 | Internal watchdog #5 times out unexpectedly | error |
714 | Unexpected internal WD instance #6 | Internal watchdog #6 times out unexpectedly | error |
715 | LINE_V & AC_OK disagree | AC line voltage and AC_OK signal mismatch | warning |
716 | POST Internal WD circuit error | Internal watchdog circuit test fails during POST | critical |
717 | Battery shorted with AC power | Battery and AC power connected simultaneously | critical |
718 | Battery open with AC power | Battery disconnected or missing when AC power present | critical |
719 | EEPROM W/R failure - POST or background | EEPROM write or read operation fails | error |
720 | Battery charger error - instance #1 | First battery charger circuit fails | error |
721 | Battery charger error - instance #2 | Second battery charger circuit fails | error |
722 | Battery charger error - instance #3 | Third battery charger circuit fails | error |
723 | Battery charger error - instance #4 | Fourth battery charger circuit fails | error |
724 | Battery charger error - instance #5 | Fifth battery charger circuit fails | error |
725 | Battery charger error - instance #6 | Sixth battery charger circuit fails | error |
726 | Battery charger error - instance #7 | Seventh battery charger circuit fails | error |
727 | Battery charger error - instance #8 | Eighth battery charger circuit fails | error |
728 | Battery charger error - instance #9 | Ninth battery charger circuit fails | error |
729 | Battery critical without AC power | Battery voltage critical and no AC power available | critical |
730 | Battery fail without AC power | Battery failure detected with no AC power backup | critical |
731 | EEPROM CRC Error - EEPROM cleared | EEPROM checksum fails and contents automatically cleared | error |
732 | Vref error on DC | DC voltage reference circuit fails or out of spec | error |
733 | Vref error on AC | AC voltage reference circuit fails or out of spec | error |
734 | Overtemperature (> 58 C) | Internal temperature exceeds 58°C threshold | critical |
735 | Battery I charge | Battery charging current measurement error or out of spec | error |
736 | Battery I discharge | Battery discharge current measurement error or out of spec | error |
737 | Temperature low | Internal temperature drops below minimum operating threshold | warning |
738 | Hardware failsafe occurred, maybe overtemperature | Hardware safety circuit activates, possibly due to overtemperature | critical |
739 | Hardware failsafe occurred, maybe overcurrent | Hardware safety circuit activates, possibly due to overcurrent | critical |
740 | Hardware failsafe occurred, cause unknown | Hardware safety circuit activates for unknown reason | critical |
741 | Message RX CRC error | Received message fails CRC checksum verification | warning |
743 | Error in the shut down sequence - occurs if the power is cycled off and on too quickly | Power cycled off and back on before shutdown completes | warning |
9 | %SpO2 front end reported syntax error in FSP message | Malformed FSP message received by front end | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.