Nellcor N-395: troubleshooting and error codes
The ServiceForge knowledge base holds 23 documented symptoms for the Nellcor N-395, with 46 ranked causes, 48 diagnostic checks with test points and expected values, and 20 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not power up on AC or DC caution
No power from either AC or DC source; Battery Low indicator may light or fail to light depending on condition
Ranked causes
- 1. Defective AC outlet or no AC power (common)
- 2. Open fuses in Power Entry Module (common)
- 3. Defective power supply (occasional)
- 4. Defective User Interface PCB or harness connection (occasional)
Verify AC power supply and indicator
- Test point
- AC outlet and AC indicator light
- Expected
- AC indicator on
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not sound patient safety
Audible alarm function non-operational; alarms do not sound when triggered or condition occurs
Ranked causes
- 1. Alarm silenced or muted by user setting (common)
- 2. Speaker defective or disconnected (common)
- 3. User Interface PCB defective (occasional)
Verify alarm sound setting is enabled
- Test point
- Monitor alarm silence/mute setting
- Expected
- Alarm sound setting enabled (not silenced)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge
Battery fails to accept or hold charge when connected to AC power; Battery Low indicator remains on or battery depletes during charging
Ranked causes
- 1. Battery age exceeds 2 years (common)
- 2. Battery unable to hold charge (internal degradation) (common)
- 3. Defective power supply (occasional)
Check battery age
- Test point
- Battery manufacture date label
- Expected
- Battery less than 2 years old
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cooling fan does not run caution
Fan not operating; unit may overheat.
Ranked causes
- 1. Temperature below fan activation threshold (common)
- 2. Temperature sensor U3 failure (occasional)
- 3. Unit running on battery power (common)
Verify AC power connection
- Test point
- Power cord and outlet
- Expected
- AC power connected and active
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display pixels do not light
LCD display or sections of display do not illuminate; pixels remain dark or inactive
Ranked causes
- 1. Poor connection between User Interface PCB and Display PCB (common)
- 2. Display PCB defective (occasional)
- 3. User Interface PCB defective (occasional)
Check connection between User Interface PCB and Display PCB
- Test point
- PCB interconnect connectors
- Expected
- Connection secure and properly seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display values are missing or erratic patient safety
Displayed values on screen are incomplete, flickering, or unstable; readings jump or disappear intermittently
Ranked causes
- 1. Sensor extension cable defective or loose (common)
- 2. Pulse oximetry sensor defective (common)
- 3. Interface PCB defective (occasional)
Inspect sensor connection and cable
- Test point
- Sensor extension cable connectors
- Expected
- Cable firmly connected, no damage or moisture
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS-232 nurse call not working
Nurse call signal on data port pin 11 not functioning when alarm occurs
Ranked causes
- 1. Improper connections on data port (common)
- 2. Incorrect output voltage on nurse call pin (occasional)
- 3. User interface PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 or Pulse values change rapidly; Pulse Amplitude indicator erratic patient safety
Readings fluctuate significantly cycle to cycle; waveform bar jumps erratically; unstable measurements
Ranked causes
- 1. Sensor damp or reused excessively (common)
- 2. Electrosurgical unit (ESU) electromagnetic interference (occasional)
- 3. User Interface PCB defective (occasional)
Inspect sensor for moisture and reuse condition
- Test point
- Pulse oximetry sensor
- Expected
- Sensor dry and within acceptable reuse limit
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sounds for no apparent reason
Audible alarm triggered spontaneously without an alarm condition; nuisance alarms occurring repeatedly
Ranked causes
- 1. Moisture or spilled liquid on or in monitor (common)
- 2. User Interface PCB defective (occasional)
Inspect monitor for moisture or liquid contamination
- Test point
- Monitor interior and PCBs
- Expected
- No visible moisture or liquid present
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery alarm triggered
Unit displays low battery alarm despite charging.
Ranked causes
- 1. Battery voltage below threshold (common)
- 2. Defective battery (occasional)
Measure battery charging voltage
- Test point
- Battery charging circuit output
- Expected
- 6.8 VDC ±0.15 V, ±0.15 V
- Tools
- Multimeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No data port printout received
Monitor not sending data to external printer via RS-232 data port
Ranked causes
- 1. Baud rate mismatch between monitor and printer (common)
- 2. User interface PCB failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse Amplitude indicator shows pulse but digital displays show zeroes patient safety
Waveform indicator bar bounces or shows activity, but numeric SpO2 and Pulse Rate displays remain at zero
Ranked causes
- 1. Sensor damaged or failed (common)
- 2. User Interface PCB defective (occasional)
Inspect sensor condition and reconnect
- Test point
- Pulse oximetry sensor
- Expected
- Sensor intact and properly seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
No output from power supply PCB; unit fails to run.
Ranked causes
- 1. Blown fuse in PEM (Hot or Neutral line) (common)
- 2. Short circuit in DC circuitry (occasional)
Verify AC input voltage at PEM
- Test point
- PEM inlet
- Expected
- 120 VAC or 240 VAC depending on SW1 setting
- Tools
- Multimeter
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not respond to buttons
Monitor powers on but does not respond to some or all button inputs; buttons non-functional
Ranked causes
- 1. Top Case assembly defective or misaligned (common)
- 2. Interface PCB defective (occasional)
Test all buttons for response
- Test point
- All user interface buttons
- Expected
- All buttons respond to presses
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog output signal levels
Three analog output channels (saturation, pulse rate, plethysmographic waveform) provide 0.0–1.0 VDC. 1.0 VDC = 100% SpO2, 250 bpm, or 254 pulse amplitude units. Output drops as parameter values decrease. If no data available, output is 1.0 VDC.
Ranked causes
- 1. Normal analog output operation during measurement (common)
Measure saturation output during normal operation
- Test point
- Saturation analog output
- Expected
- 0.0–1.0 VDC linear with SpO2 (1.0 VDC = 100%)
- Tools
- DMM
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery connected backwards caution
Reverse battery connection is protected by circuit components on User Interface PCB and Power Supply
Ranked causes
- 1. Battery polarity reversed during installation (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Low indicator on during DC operation with alarm sounding
Battery Low indicator illuminated during battery-powered operation accompanied by audible alarm
Ranked causes
- 1. Battery charge critically low (15 minutes or less remaining) (common)
Monitor remaining battery time
- Test point
- Battery Low indicator and alarm status
- Expected
- Battery Low indicator on, alarm sounding, 15 min or less charge remaining
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery short circuit detected caution
Fuse F3 opens to protect battery when output current exceeds 5.0 amps
Ranked causes
- 1. Battery output current exceeds 5.0 amps (short to ground) (rare)
Measure battery output current
- Test point
- Battery output circuit or power supply output
- Expected
- Below 5.0 amps, < 5.0 amps
- Tools
- Ammeter or clamp meter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery temperature exceeds 50°C caution
Battery protection circuit opens charging circuit when battery temperature rises above 50°C
Ranked causes
- 1. Battery temperature exceeds 50°C (occasional)
Verify battery temperature
- Test point
- Battery temperature sensor or battery casing
- Expected
- Below 50°C
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: analog output signals
Automatic three-step calibration signal at power-on after POST completes. Signal sequence: 0.0 VDC for 60 sec, then 1.0 VDC for 60 sec, then stair-step 0.0–1.0 VDC in 0.1 VDC increments (1 sec each).
Ranked causes
- 1. Normal power-on calibration procedure (common)
Verify first segment of calibration signal
- Test point
- Analog output (saturation, pulse rate, or plethysmographic)
- Expected
- 0.0 VDC held for 60 seconds
- Tools
- DMM or oscilloscope
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Critical battery level error code displayed patient safety
Unit displays error code and will not monitor patient when powered on at critical battery voltage
Ranked causes
- 1. Battery voltage at or below 5.67 VDC critical level (common)
Measure battery voltage
- Test point
- Battery terminals or internal battery connection
- Expected
- 5.67 ± 0.1 VDC (critical threshold), ±0.1 VDC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery alarm declared
Unit displays low battery alarm and will power down after 15 minutes of continued operation
Ranked causes
- 1. Battery voltage dropped below 5.85 VDC (common)
Measure battery voltage
- Test point
- Battery terminals or internal battery connection
- Expected
- 5.85 ± 0.1 VDC or higher, ±0.1 VDC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not operate when disconnected from AC power
Device non-functional on DC battery power; may be intermittent or complete loss of operation
Ranked causes
- 1. Battery is discharged (common)
Check battery charge state and recharge
- Test point
- Battery and AC power
- Expected
- Battery fully charged after connection to AC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Power-on self-test (POST) failure | POST failure detected during startup | critical |
11 | Flash ROM corruption | Flash memory checksum failed | error |
12 | Excessive microprocessor resets | Too many CPU reset events | error |
4 | Battery dead or non-functional | Battery cannot supply power | error |
5 | Too many microprocessor resets within period of time | Excessive CPU reset events detected | error |
52 | Loss of monitor settings | Settings lost or corrupted | warning |
6 | Boot CRC error | Checksum error in boot code | error |
76 | Error accessing EPROM | EPROM read/write failure | error |
8 | Boot CRC error | Checksum error in boot code | error |
80 | Institutional default values lost, reset to factory defaults | Institutional settings loss | warning |
81 | Settings lost (non-default values lost) | Configured settings lost from memory | warning |
82 | Real-time clock lost | RTC data loss or failure | warning |
84 | Internal communications error | Communication between internal components failed | error |
DISALLOWED ON BATTERY | Print or download operation attempted while on battery power | User attempted print/download during battery operation | info |
DISALLOWED ON LOW BATTERY | Backlight activation attempted during low battery condition | User turned on backlight during low battery | info |
INVALID SILENCE DURATION | Alarm silence duration power-on default set to OFF | User attempted to set silence duration to OFF | info |
INVALID SpO2 LIMIT | SpO2 alarm limit set below 80 | User attempted to set SpO2 limit power-on default below 80 | info |
READING TRENDS | Monitor gathering trend information for display | System collecting historical trend data | info |
SENSOR DISCONNECTED | Pulse oximetry sensor or cable disconnected | Sensor unplugged from patient or cable unplugged from monitor | warning |
SENSOR OFF | Pulse oximetry sensor removed from patient | Sensor no longer on patient | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.