Nellcor N-20: troubleshooting and error codes
The ServiceForge knowledge base holds 12 documented symptoms for the Nellcor N-20, with 25 ranked causes, 26 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Pulse Search indicator appears for more than 5-10 seconds patient safety
Pulse Search indicator remains on longer than normal, indicating difficulty finding a pulse.
Ranked causes
- 1. Sensor improperly positioned (common)
- 2. Incorrect sensor in use (common)
- 3. Perfusion too low (occasional)
- 4. Foreign material on sensor LEDs or photodetector (occasional)
- 5. Patient or environmental motion interfering (occasional)
- 6. Sensor too tight or excessive illumination (occasional)
- 7. DB-9 sensor connector broken (rare)
Verify sensor placement
- Test point
- Sensor application site
- Expected
- Sensor correctly positioned per directions for use
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No response to Measure button
Pressing the Measure button produces no response from the oximeter.
Ranked causes
- 1. Battery access door not properly latched (common)
- 2. Batteries discharged (common)
- 3. Batteries installed incorrectly (occasional)
- 4. Battery contacts not making electrical connection (occasional)
- 5. Fuse F1 open on auxiliary PCB (occasional)
- 6. Dust under Measure button causing contact loss (rare)
Verify battery access door is properly latched
- Test point
- Access door
- Expected
- Door fully latched and secure
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse Search indicator appears after successful measurements patient safety
After making successful measurements, Pulse Search indicator suddenly appears, indicating loss of signal.
Ranked causes
- 1. Patient perfusion too low (common)
- 2. Patient motion interfering with pulse tracking (common)
- 3. Sensor too tight or excessive illumination (occasional)
Assess patient perfusion status
- Test point
- Patient vital signs and limb perfusion
- Expected
- Adequate and stable perfusion
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance verification after repair
Required materials and test setup for N-20/N-20P pulse oximeter post-repair validation including display backlight, battery, and self-test verification
Ranked causes
- 1. Post-repair verification required (common)
Main backlight illumination in low light
- Test point
- Oxygen Saturation and Pulse Rate display, 14-segment pulse rate amplitude indicator
- Expected
- Backlight illuminates in darkened room
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dashes appear in the display
Display shows dashes (– – –) instead of pulse oximetry readings.
Ranked causes
- 1. Sensor not connected to instrument (common)
Verify sensor connection
- Test point
- Sensor DB-9 connector and instrument input
- Expected
- Sensor firmly connected to instrument; no extension cable
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display backlight not illuminating
One or more sections of electroluminescent backlight fail to illuminate under low light conditions or during power-on
Ranked causes
- 1. Backlight component failure or connection issue (unknown)
Ambient light detector function
- Test point
- Circular window in top right corner of display
- Expected
- Main section backlight activates in low light conditions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message with number displayed
Err followed by a number appears on display; specific error code not identified at this stage
Ranked causes
- 1. Unknown error code (unknown)
Note error number on display
- Test point
- LCD display field
- Expected
- Err followed by numeric code
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper mechanism jams during loading or printing caution
Paper stuck between print head and printer mechanism or in paper path; difficulty advancing or removing paper
Ranked causes
- 1. Paper caught in printer mechanism (occasional)
Assess print head position
- Test point
- Print head carriage location
- Expected
- Print head at home position (right side)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer fails to operate
Printer is non-functional; N-20P model only
Ranked causes
- 1. Fuse F2 on auxiliary PCB open (occasional)
Inspect fuse F2 on auxiliary PCB
- Test point
- Auxiliary PCB fuse F2
- Expected
- Fuse intact and conducting
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not operational while monitor continues measuring
Paper jam or print head not at home position; patient measurements continue unaffected
Ranked causes
- 1. Paper jammed in printer mechanism (common)
Inspect paper compartment for jammed paper
- Test point
- Paper path in printer
- Expected
- No paper obstruction; print head at home position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer paper advances but no printing occurs
Paper moves through mechanism but thermal print head does not mark; N-20P only
Ranked causes
- 1. Thermal paper improperly loaded (common)
Verify thermal paper orientation
- Test point
- Paper loaded in printer compartment
- Expected
- Thermal side facing print head; paper seated in slot correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Time or date incorrect on display
Clock shows wrong time or date; N-20P model only
Ranked causes
- 1. Real-time clock (RTC) battery exhausted (occasional)
Verify displayed time and date
- Test point
- LCD display after startup
- Expected
- Correct time and date matching current date/time
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
101 | Error in internal RAM registers test | Power-on self-test or performance test detects internal RAM register failure | critical |
102 | Error in zero register test | Power-on self-test or performance test detects zero register failure | critical |
103 | Error in register contents clearing test | Power-on self-test or performance test detects register clearing failure | critical |
104 | Error in register contents increment test | Power-on self-test or performance test detects register increment failure | critical |
105 | Error in register contents decrement test | Power-on self-test or performance test detects register decrement failure | critical |
106 | Errors in logical operations test | Power-on self-test or performance test detects logical operation failure | critical |
110 | Error in exchange test | Power-on self-test or performance test detects exchange operation failure | critical |
111 | Error in timer tests | Power-on self-test or performance test detects timer failure | critical |
112 | Error in window select register test | Power-on self-test or performance test detects window select register failure | critical |
113 | Errors in stack manipulation test | Power-on self-test or performance test detects stack manipulation failure | critical |
115 | Errors in CPU flags test | Power-on self-test or performance test detects CPU flags failure | critical |
118 | Error in interrupt pending register test | Power-on self-test or performance test detects interrupt pending register failure | critical |
119 | Error in program counter test | Power-on self-test or performance test detects program counter failure | critical |
120 | Error in CPU serial port test | Power-on self-test or performance test detects serial port failure | critical |
121 | Error in pulse width modulation register test | Power-on self-test or performance test detects PWM register failure | critical |
122 | Error in A/D register test | Power-on self-test or performance test detects A/D register failure | critical |
123 | Error in addressing modes test | Power-on self-test or performance test detects addressing mode failure | critical |
124 | Error in high speed input register test | Power-on self-test or performance test detects high speed input register failure | critical |
125 | Error in content addressable memory test | Power-on self-test or performance test detects CAM failure | critical |
201 | Errors in external RAM test | Power-on self-test or performance test detects external RAM failure | critical |
301 | Error in PROM test | Power-on self-test or performance test detects PROM failure | critical |
401 | Errors in I/O port test | Power-on self-test or performance test detects I/O port failure | critical |
601 | Failure of real-time clock | Power-on self-test or performance test detects real-time clock failure | critical |
602 | Errors in real-time clock | Power-on self-test or performance test detects RTC error | critical |
701 | Errors in watchdog-timer | Power-on self-test or performance test detects watchdog-timer failure | critical |
Pr Err | Printer error condition | Printer detects error during operation | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI