Nellcor N-20: troubleshooting and error codes

The ServiceForge knowledge base holds 12 documented symptoms for the Nellcor N-20, with 25 ranked causes, 26 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Pulse Search indicator appears for more than 5-10 seconds patient safety

Pulse Search indicator remains on longer than normal, indicating difficulty finding a pulse.

Ranked causes

  1. 1. Sensor improperly positioned (common)
  2. 2. Incorrect sensor in use (common)
  3. 3. Perfusion too low (occasional)
  4. 4. Foreign material on sensor LEDs or photodetector (occasional)
  5. 5. Patient or environmental motion interfering (occasional)
  6. 6. Sensor too tight or excessive illumination (occasional)
  7. 7. DB-9 sensor connector broken (rare)
First check

Verify sensor placement

Test point
Sensor application site
Expected
Sensor correctly positioned per directions for use

5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No response to Measure button

Pressing the Measure button produces no response from the oximeter.

Ranked causes

  1. 1. Battery access door not properly latched (common)
  2. 2. Batteries discharged (common)
  3. 3. Batteries installed incorrectly (occasional)
  4. 4. Battery contacts not making electrical connection (occasional)
  5. 5. Fuse F1 open on auxiliary PCB (occasional)
  6. 6. Dust under Measure button causing contact loss (rare)
First check

Verify battery access door is properly latched

Test point
Access door
Expected
Door fully latched and secure

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pulse Search indicator appears after successful measurements patient safety

After making successful measurements, Pulse Search indicator suddenly appears, indicating loss of signal.

Ranked causes

  1. 1. Patient perfusion too low (common)
  2. 2. Patient motion interfering with pulse tracking (common)
  3. 3. Sensor too tight or excessive illumination (occasional)
First check

Assess patient perfusion status

Test point
Patient vital signs and limb perfusion
Expected
Adequate and stable perfusion

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Performance verification after repair

Required materials and test setup for N-20/N-20P pulse oximeter post-repair validation including display backlight, battery, and self-test verification

Ranked causes

  1. 1. Post-repair verification required (common)
First check

Main backlight illumination in low light

Test point
Oxygen Saturation and Pulse Rate display, 14-segment pulse rate amplitude indicator
Expected
Backlight illuminates in darkened room

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Dashes appear in the display

Display shows dashes (– – –) instead of pulse oximetry readings.

Ranked causes

  1. 1. Sensor not connected to instrument (common)
First check

Verify sensor connection

Test point
Sensor DB-9 connector and instrument input
Expected
Sensor firmly connected to instrument; no extension cable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display backlight not illuminating

One or more sections of electroluminescent backlight fail to illuminate under low light conditions or during power-on

Ranked causes

  1. 1. Backlight component failure or connection issue (unknown)
First check

Ambient light detector function

Test point
Circular window in top right corner of display
Expected
Main section backlight activates in low light conditions

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error message with number displayed

Err followed by a number appears on display; specific error code not identified at this stage

Ranked causes

  1. 1. Unknown error code (unknown)
First check

Note error number on display

Test point
LCD display field
Expected
Err followed by numeric code

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper mechanism jams during loading or printing caution

Paper stuck between print head and printer mechanism or in paper path; difficulty advancing or removing paper

Ranked causes

  1. 1. Paper caught in printer mechanism (occasional)
First check

Assess print head position

Test point
Print head carriage location
Expected
Print head at home position (right side)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer fails to operate

Printer is non-functional; N-20P model only

Ranked causes

  1. 1. Fuse F2 on auxiliary PCB open (occasional)
First check

Inspect fuse F2 on auxiliary PCB

Test point
Auxiliary PCB fuse F2
Expected
Fuse intact and conducting

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer not operational while monitor continues measuring

Paper jam or print head not at home position; patient measurements continue unaffected

Ranked causes

  1. 1. Paper jammed in printer mechanism (common)
First check

Inspect paper compartment for jammed paper

Test point
Paper path in printer
Expected
No paper obstruction; print head at home position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer paper advances but no printing occurs

Paper moves through mechanism but thermal print head does not mark; N-20P only

Ranked causes

  1. 1. Thermal paper improperly loaded (common)
First check

Verify thermal paper orientation

Test point
Paper loaded in printer compartment
Expected
Thermal side facing print head; paper seated in slot correctly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Time or date incorrect on display

Clock shows wrong time or date; N-20P model only

Ranked causes

  1. 1. Real-time clock (RTC) battery exhausted (occasional)
First check

Verify displayed time and date

Test point
LCD display after startup
Expected
Correct time and date matching current date/time

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
101Error in internal RAM registers testPower-on self-test or performance test detects internal RAM register failurecritical
102Error in zero register testPower-on self-test or performance test detects zero register failurecritical
103Error in register contents clearing testPower-on self-test or performance test detects register clearing failurecritical
104Error in register contents increment testPower-on self-test or performance test detects register increment failurecritical
105Error in register contents decrement testPower-on self-test or performance test detects register decrement failurecritical
106Errors in logical operations testPower-on self-test or performance test detects logical operation failurecritical
110Error in exchange testPower-on self-test or performance test detects exchange operation failurecritical
111Error in timer testsPower-on self-test or performance test detects timer failurecritical
112Error in window select register testPower-on self-test or performance test detects window select register failurecritical
113Errors in stack manipulation testPower-on self-test or performance test detects stack manipulation failurecritical
115Errors in CPU flags testPower-on self-test or performance test detects CPU flags failurecritical
118Error in interrupt pending register testPower-on self-test or performance test detects interrupt pending register failurecritical
119Error in program counter testPower-on self-test or performance test detects program counter failurecritical
120Error in CPU serial port testPower-on self-test or performance test detects serial port failurecritical
121Error in pulse width modulation register testPower-on self-test or performance test detects PWM register failurecritical
122Error in A/D register testPower-on self-test or performance test detects A/D register failurecritical
123Error in addressing modes testPower-on self-test or performance test detects addressing mode failurecritical
124Error in high speed input register testPower-on self-test or performance test detects high speed input register failurecritical
125Error in content addressable memory testPower-on self-test or performance test detects CAM failurecritical
201Errors in external RAM testPower-on self-test or performance test detects external RAM failurecritical
301Error in PROM testPower-on self-test or performance test detects PROM failurecritical
401Errors in I/O port testPower-on self-test or performance test detects I/O port failurecritical
601Failure of real-time clockPower-on self-test or performance test detects real-time clock failurecritical
602Errors in real-time clockPower-on self-test or performance test detects RTC errorcritical
701Errors in watchdog-timerPower-on self-test or performance test detects watchdog-timer failurecritical
Pr ErrPrinter error conditionPrinter detects error during operationerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI