Nellcor A7390: troubleshooting and error codes
The ServiceForge knowledge base holds 27 documented symptoms for the Nellcor A7390, with 65 ranked causes, 24 diagnostic checks with test points and expected values, and 70 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
USB data port does not function properly
USB port fails to transfer data or communicate with PC
Ranked causes
- 1. USB cable not firmly connected (common)
- 2. Covidien bridge driver corrupted (occasional)
- 3. Mismatched baud rates between monitoring system and PC (occasional)
- 4. Mismatched COM port between PC and HyperTerminal (occasional)
- 5. USB port damaged (rare)
- 6. Main board failure (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Questionable patient physiological measurements or missing/mistagged data patient safety
Unreliable SpO2 readings, patient data incorrectly tagged, or patient data missing
Ranked causes
- 1. Sensor not attached correctly (common)
- 2. Sensor is malfunctioning (occasional)
- 3. Electromagnetic interference (occasional)
- 4. Ambient light interfering with sensing (occasional)
- 5. Coin cell battery removed from main board (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Date and time incorrect on LCD and in monitoring history download patient safety
System displays wrong date/time or monitoring history shows incorrect timestamps
Ranked causes
- 1. Date and time not set correctly during installation (common)
- 2. Coin cell battery depleted or removed and replaced (common)
- 3. Coin cell battery tab does not provide firm contact (occasional)
- 4. Main board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not function properly or looks distorted
LCD shows distorted image, improper display, or visual artifacts
Ranked causes
- 1. Electromagnetic interference (occasional)
- 2. LCD cable disconnected or improperly connected (common)
- 3. Main board is malfunctioning (occasional)
- 4. LCD visibly cracked or broken (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No response when control buttons are pressed
Buttons do not respond to user input or commands
Ranked causes
- 1. Keypad ribbon cable not making contact (common)
- 2. System is frozen (occasional)
- 3. Buttons are defective (occasional)
- 4. Button activators on main board malfunctioning (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor messages: SpO2 Pulse Search, Sensor Off, Cable/Sensor Disconnect, Signal Artifact patient safety
SpO2 sensor connectivity or signal quality issues indicated by monitor messages
Ranked causes
- 1. Sensor disconnected from cable or monitoring system (common)
- 2. Sensor detached from patient (common)
- 3. Interference from substances or motion (common)
- 4. Defective cable or sensor (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sound during POST
Buzzer or speaker does not produce sound during power-on self-test
Ranked causes
- 1. Speaker cable is loose or disconnected (common)
- 2. Speaker is malfunctioning (occasional)
- 3. Main board is malfunctioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not turn on when Power button is pressed
Monitoring system fails to power on despite pressing Power On/Off button
Ranked causes
- 1. Power button not held long enough (common)
- 2. Battery not properly seated (common)
- 3. Depleted or unsuitable battery (common)
Verify power button activation
- Test point
- Power On/Off button
- Expected
- Button held for 1 second
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio alarm cannot be silenced patient safety
Alarm sound continues after pressing Silence button
Ranked causes
- 1. Main board is malfunctioning (occasional)
- 2. Keypad is damaged (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio alarm does not silence for specified time patient safety
Silence button press does not mute alarm for duration set in institutional defaults
Ranked causes
- 1. Institutional defaults not set correctly (common)
- 2. Silence Alarm button activator malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank display after normal POST
LCD screen shows no image or text following successful power-on self-test
Ranked causes
- 1. LCD cable disconnected (common)
- 2. LCD is damaged (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low or critically low battery condition
Monitor displays low battery or critically low battery message
Ranked causes
- 1. Battery depleted after extended use (common)
- 2. Unsuitable battery type installed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitoring system frozen
System is frozen or unresponsive
Ranked causes
- 1. NELL-1 malfunction requiring reset (common)
- 2. Main board malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Permission to Mute Alarm not functioning
User cannot mute alarm through Permission to Mute Alarm setting; alarm audio continues despite setting enabled
Ranked causes
- 1. Permission to Mute Alarm setting not enabled in Service Menu (common)
- 2. Alarm audio not disabled in Device Settings Sound Settings (common)
Verify Permission to Mute Alarm is enabled
- Test point
- Service Menu Permission to Mute Alarm
- Expected
- Yes
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse Volume not adjustable or not audible
Pulse tone does not sound with heartbeats or cannot be adjusted in volume
Ranked causes
- 1. Pulse Volume setting not configured correctly (common)
- 2. Sensor not connected properly or patient data not being detected (occasional)
Access Pulse Volume setting
- Test point
- Device Settings Sound Settings Pulse Volume
- Expected
- menu accessible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen Saver not activating or not darkening display
Screen does not darken during inactivity or Screen Saver setting not functioning
Ranked causes
- 1. Screen Saver Time setting not configured correctly (common)
- 2. Screen Saver only activates when no alarm condition and no buttons pressed (occasional)
Access Screen Saver Time setting
- Test point
- Service Menu Power Saving Settings Screen Saver Time
- Expected
- setting accessible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor Disconnect alarm not triggering or not at correct priority patient safety
Sensor disconnect is not detected or alarm priority does not match setting
Ranked causes
- 1. Sensor Disconnect alarm priority not set correctly (common)
- 2. Sensor not properly disconnected or connection is intermittent (occasional)
Disconnect sensor from monitoring system
- Test point
- Sensor port connection
- Expected
- sensor fully disconnected, not making intermittent contact
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor Off alarm not triggering or not at correct priority patient safety
Sensor off condition is not detected or alarm priority does not match setting
Ranked causes
- 1. Sensor Off alarm priority not set correctly (common)
- 2. Sensor still in contact with patient or connection not fully removed (occasional)
Disconnect sensor from patient finger
- Test point
- Sensor contact with finger
- Expected
- sensor fully removed from finger
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure (910) message displayed with buzzer sound
Monitor displays error code 910 and buzzer sounds during POST
Ranked causes
- 1. Speaker cable is disconnected (common)
- 2. Speaker or main board is malfunctioning (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System Failure (920) message displayed
Monitoring system displays error message 920 after POST or abnormal shutdown
Ranked causes
- 1. Abnormally shut down or harnesses disconnected (common)
- 2. Batteries depleted (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System failure message and error code appear on display
Monitor displays system failure message with associated error code
Ranked causes
- 1. Firmware is not loading correctly (occasional)
- 2. Monitoring system is malfunctioning (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Silenced function not working
Alarm silence button does not silence alarms; alarm tone continues or does not silence for expected duration
Ranked causes
- 1. Alarm Silence Duration not set to 30 seconds (common)
Verify alarm silence duration setting
- Test point
- Service Menu Alarm Silence Duration
- Expected
- 30 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Volume Control not adjusting
Alarm volume cannot be adjusted or does not respond to volume control changes
Ranked causes
- 1. Alarm volume setting not accessible or responsive (occasional)
Access Alarm Volume control
- Test point
- Device Settings Sound Settings Alarm Volume
- Expected
- menu accessible and responding to Up/Down buttons
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display brightness cannot be adjusted
Screen brightness is too dim or too bright and does not respond to adjustment controls
Ranked causes
- 1. Brightness setting not accessible or not responding to controls (common)
Access Brightness Setting
- Test point
- Device Settings Brightness Setting menu
- Expected
- menu accessible and responsive
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Key Beep Volume not responding
Key beep sounds on every button press when disabled, or does not sound when enabled
Ranked causes
- 1. Key Beep Volume setting not properly configured (common)
Access Key Beep Volume setting
- Test point
- Device Settings Sound Settings Key Beep Volume
- Expected
- menu accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Save Spot Reading function not working
Spot readings cannot be saved or saved data is not displayed in monitoring history
Ranked causes
- 1. Spot Reading not being saved or saved data cannot be retrieved (occasional)
Attempt to save spot reading
- Test point
- Menu Save Spot Reading function
- Expected
- Spot Reading Saved message appears
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor Failure alarm not functioning patient safety
Sensor Failure condition is not detected or alarm does not sound
Ranked causes
- 1. Sensor Failure alarm priority setting not configured or sensor not actually failing (unknown)
Verify Sensor Failure priority setting available
- Test point
- Service Menu Sensor Alarm Priority Settings
- Expected
- Sensor Failure option available with Low, Medium, or High selection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
001 | SpO2 front end RAM error | RAM memory failure detected in SpO2 front end | critical |
002 | SpO2 front end ROM/code integrity error | ROM or firmware code integrity check failed in SpO2 front end | critical |
004 | SpO2 front end reports command not allowed | Invalid or out-of-sequence command sent to SpO2 front end | error |
005 | SpO2 front end reports illegal value sent by SpO2 back end | Invalid parameter value received from SpO2 back end module | error |
006 | SpO2 front end reports calibration (offset) failure | Calibration offset calculation failed during initialization | error |
010 | Over-current limit in SpO2 front end has tripped | Overcurrent condition detected in SpO2 front end power supply | error |
012 | SpO2 front end reports other hardware problem | Unspecified hardware fault detected in SpO2 front end | error |
017 | SpO2 front end reports other hardware problem | Unspecified hardware fault detected in SpO2 front end | error |
018 | SpO2 front end reports internal register modified from expected value | Internal register value changed unexpectedly during operation | error |
048 | SpO2 front end reports spurious interrupt detected | Unexpected interrupt signal detected in SpO2 front end | warning |
050 | SpO2 front end reports internal software consistency check failed | Internal software checksum or integrity check failed in front end | error |
051 | SpO2 front end reports DigiCAL communications error | Communication error on DigiCAL interface of SpO2 front end | error |
256 | SpO2 back end reports beginning of packet missing | Packet received without proper start delimiter in back end | error |
257 | SpO2 back end reports packet state ID (SID) missing | Received packet missing State ID field in back end | error |
258 | SpO2 back end reports packet length error | Received packet length does not match expected value in back end | error |
259 | SpO2 back end reports message length error | Message length exceeds expected size or is invalid in back end | error |
260 | SpO2 back end reports packet contains unsupported Key | Packet received containing unrecognized Key field in back end | warning |
261 | SpO2 back end reports packet CRC error | Cyclic Redundancy Check failed on received packet in back end | error |
262 | SpO2 back end reports end of packet missing | Packet received without proper end delimiter in back end | error |
263 | SpO2 back end reports packet contains undefined Key | Packet received containing undefined Key field in back end | warning |
264 | SpO2 back end reports corrupted variable | Variable data integrity check failed in back end software | error |
265 | SpO2 back end reports memory overflow | Memory allocation exceeded available space in back end | error |
266 | SpO2 back end reports bad pointer | Invalid memory pointer detected in back end software | error |
267 | SpO2 back end reports parameter value out-of-range | Parameter value received outside valid range in back end | warning |
268 | SpO2 back end reports reset detected | Unexpected reset or restart detected in back end module | warning |
269 | SpO2 back end reports unexpected value | Unexpected data value received in back end processing | warning |
270 | SpO2 back end reports timeout | Operation exceeded timeout threshold in back end | warning |
271 | SpO2 back end reports not ready/not initialized | Back end module not initialized or ready for operation | error |
272 | SpO2 back end reports double fault | Multiple fault conditions detected simultaneously in back end | critical |
273 | SpO2 back end reports data out of range error | Measurement or data value outside acceptable range in back end | warning |
274 | SpO2 back end reports incompatible software version | Software version mismatch between modules in back end | error |
275 | SpO2 back end reports incorrect registration number | Registration or license number mismatch detected in back end | error |
276 | SpO2 back end reports sensor read failure | Unable to read sensor data in back end module | error |
277 | SpO2 back end reports sensor signature verification fails | Sensor authentication or signature check failed in back end | error |
280 | SpO2 back end reports does not support feature required by sensor | Sensor requires feature not supported by back end module | warning |
281 | SpO2 back end reports overflow/underflow | Data value overflow or underflow condition in back end processing | warning |
282 | SpO2 back end reports sensor activation failure | Sensor could not be activated during initialization in back end | error |
283 | SpO2 back end reports sensor write failure | Unable to write data to sensor in back end module | error |
284 | Host is providing both HW and SW ECG triggers | Both hardware and software ECG trigger signals sent to monitor | error |
285 | Host attempted Read or Close of Sensor Event before successful Open | Sensor Event accessed without proper initialization in host | error |
286 | Host attempted redundant Open of Sensor Event | Sensor Event opened multiple times without closing | error |
287 | An error occurred with the Sensor Event data and is unavailable for reading | Sensor Event data corrupted or unable to be accessed by host | error |
288 | No more Sensor Event data available for reading by host | All available Sensor Event data has been read by host | info |
289 | Sensor Private Label/Host Sensor Key incompatible; will not operate sensor | Sensor key does not match host requirements in back end | error |
512 | SpO2 back end reports internal software consistency check failed | Back end software checksum or integrity verification failed | error |
513 | SpO2 back end reports software functions executed before initialization completed | Software function called before initialization sequence complete | error |
514 | SpO2 back end reports internal memory buffer overflow | Internal memory buffer exceeded capacity in back end | error |
516 | SpO2 back end ROM/code integrity error | ROM or firmware code integrity check failed in back end | critical |
517 | SpO2 back end RAM error | RAM memory failure detected in back end module | critical |
518 | SpO2 back end reports internal software consistency check failed | Back end software checksum or integrity verification failed | error |
519 | SpO2 back end reports a state machine is in an unexpected state | Internal state machine detected in invalid state in back end | error |
520 | SpO2 back end reports memory corruption detected | Memory corruption detected by diagnostic check in back end | critical |
521 | SpO2 back end reports spurious interrupt detected | Unexpected interrupt signal detected in back end module | warning |
522 | SpO2 back end reports unable to issue commands to SpO2 front end | Communication failure when sending commands to front end | error |
523 | SpO2 back end reports unable to write to external Flash | External Flash memory write operation failed in back end | critical |
524 | SpO2 back end reports that communication with SpO2 front end lost | Communication link lost between back end and front end modules | critical |
525 | SpO2 back end reports internal register modified from expected value | Internal register value changed unexpectedly in back end | error |
701 | Flash writing transaction error | Error occurred during Flash memory write operation | error |
901 | SpO2 module communication error/setting value mismatch between host and SpO2 module | Communication error or configuration mismatch with SpO2 module | error |
902 | Program memory error | Error detected in program memory during operation | critical |
903 | RAM error | Error detected in RAM memory during operation | critical |
904 | PSRAM error | Error detected in PSRAM (pseudo-static RAM) during operation | critical |
905 | Sub MCU communication error | Communication failure with sub-microcontroller unit | error |
907 | RTC error | Real-time clock (RTC) error detected during operation | warning |
908 | Serial flash memory error | Error detected in serial Flash memory during operation | error |
909 | Internal voltage error | Internal power supply voltage out of acceptable range | warning |
910 | Sound/speaker error | Speaker or audio system malfunction detected during POST | warning |
912 | LCD error | LCD display malfunction detected during POST | warning |
913 | Button error | Button or keypad malfunction detected during POST | warning |
920 | Abnormally shut down last time | System was not shut down properly on previous startup | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI