Mizuho OSI MTS Base 5890/5891: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Mizuho OSI MTS Base 5890/5891, with 34 ranked causes, 29 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No column function either end
Head or foot end column does not respond to control input
Ranked causes
- 1. Power cord disconnected or unplugged (common)
- 2. Circuit breaker off or tripped (common)
- 3. Hand control module or coiled hose disconnected (occasional)
- 4. Electrical connector at column base disconnected (occasional)
Verify power cord connection
- Test point
- power outlet
- Expected
- Power cord plugged in
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No column function foot end
Foot end column does not operate while head end may function
Ranked causes
- 1. Hand control or coiled hose disconnected (common)
- 2. Electrical connector at foot end column base disconnected (common)
- 3. Failed air/electrical switch or junction box defect (occasional)
Verify hand control and coiled hose connection
- Test point
- hand control module
- Expected
- Coiled hose plugged in
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No column function head end
Head end column does not operate while foot end may function
Ranked causes
- 1. Hand control or coiled hose disconnected (common)
- 2. Electrical connector at head end column base disconnected (common)
- 3. Failed air/electrical switch or junction box defect (occasional)
Verify hand control and coiled hose connection
- Test point
- hand control module
- Expected
- Coiled hose plugged in
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No table functions operational patient safety
Pressing buttons on Hand Control Module produces no movement or response from the surgical table
Ranked causes
- 1. Input power lost or disconnected (common)
- 2. Column connector disconnected or loose (common)
- 3. Pneumatic tube disconnection or blockage in Junction Box (occasional)
Audible click when button pressed
- Test point
- Inside Junction Box
- Expected
- Audible click heard
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Retractable table base will not collapse caution
Retractable center beam cannot be retracted into base
Ranked causes
- 1. Center beam excessive soil or debris (common)
- 2. Roller excessive soil, debris or damaged (common)
- 3. Lock handle not released or over-tightened (occasional)
Inspect center beam for soil and debris
- Test point
- center beam outer and inner surfaces
- Expected
- Clean surfaces without obstruction
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotation friction control will not stop rotation patient safety
Head or foot end rotation cannot be locked using friction control brake
Ranked causes
- 1. Brake tension insufficient or control not fully tightened (common)
- 2. Foot end control handle out of proper range (occasional)
- 3. Brake band worn or damaged (occasional)
Tighten friction control lever to maximum
- Test point
- friction control lever
- Expected
- Lever at full tightened position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table top cannot be attached to H-frame at foot end caution
Table top attachment to foot end H-frame cannot be completed
Ranked causes
- 1. H-frame will not connect to crossbar (common)
- 2. T-pin bent or misaligned (occasional)
- 3. Head and foot end columns misaligned (occasional)
Verify H-frame connection to crossbar
- Test point
- H-frame to crossbar connection
- Expected
- H-frame fully connected to crossbar
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transfer lock will not insert into crossbar patient safety
Patient transfer safety lock lever cannot be inserted into crossbar at head or foot end
Ranked causes
- 1. Crossbar pivot shaft bent or misaligned (occasional)
- 2. Transfer lock has burr or scratch preventing insertion (occasional)
- 3. Transfer lock spring weakened or broken (rare)
Inspect crossbar pivot shaft for straightness
- Test point
- crossbar pivot shaft
- Expected
- Shaft straight and aligned
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot End Slider movement requires excessive or insufficient force
Foot End Slider requires more or less than 20 to 25 lbs. force to move the Crossbar along the column top plate
Ranked causes
- 1. Drag shoe friction setting requires adjustment (common)
- 2. Incorrect force measurement technique (occasional)
Measure Foot End Slider movement force
- Test point
- Foot End Slider and Crossbar
- Expected
- 20 to 25 lbs. force required, ±2.5 lbs.
- Tools
- Force gauge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
H-frame will not connect to crossbar caution
H-frame attachment to head or foot end crossbar cannot be completed
Ranked causes
- 1. Mounting pin bent or misaligned (common)
- 2. H-frame damaged or deformed (occasional)
Inspect mounting pin for straightness
- Test point
- H-frame mounting pin
- Expected
- Pin straight and properly aligned
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotational movement restricted or cannot lock at maximum friction
Head End or Foot End Crossbar cannot be rotated or does not lock completely when Rotational Friction Control Lever set to MAX position
Ranked causes
- 1. Head End Friction Control wear requiring adjustment (common)
- 2. Foot End Friction Control wear requiring adjustment (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Single table function inoperative patient safety
One function button on Hand Control Module does not actuate corresponding table movement while other functions work normally
Ranked causes
- 1. Air leak in Hand Control Module hose lumen (common)
- 2. Hand Control Module internal failure (occasional)
Audible click when suspected button pressed
- Test point
- Inside Junction Box
- Expected
- No audible click
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Caster will not lock or unlock patient safety
Caster locking mechanism cannot engage or disengage
Ranked causes
- 1. Caster damaged or worn (common)
Inspect caster and locking mechanism for damage
- Test point
- caster wheel assembly
- Expected
- Caster intact with functional locking mechanism
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Thermal switch activated | Column motor automatically stopped due to continuous operation exceeding 60-80 seconds | Column operated continuously for longer than 60 to 80 seconds | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI