Mizuho OSI MTS Base 5890/5891: troubleshooting and error codes

The ServiceForge knowledge base holds 13 documented symptoms for the Mizuho OSI MTS Base 5890/5891, with 34 ranked causes, 29 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No column function either end

Head or foot end column does not respond to control input

Ranked causes

  1. 1. Power cord disconnected or unplugged (common)
  2. 2. Circuit breaker off or tripped (common)
  3. 3. Hand control module or coiled hose disconnected (occasional)
  4. 4. Electrical connector at column base disconnected (occasional)
First check

Verify power cord connection

Test point
power outlet
Expected
Power cord plugged in

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No column function foot end

Foot end column does not operate while head end may function

Ranked causes

  1. 1. Hand control or coiled hose disconnected (common)
  2. 2. Electrical connector at foot end column base disconnected (common)
  3. 3. Failed air/electrical switch or junction box defect (occasional)
First check

Verify hand control and coiled hose connection

Test point
hand control module
Expected
Coiled hose plugged in

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No column function head end

Head end column does not operate while foot end may function

Ranked causes

  1. 1. Hand control or coiled hose disconnected (common)
  2. 2. Electrical connector at head end column base disconnected (common)
  3. 3. Failed air/electrical switch or junction box defect (occasional)
First check

Verify hand control and coiled hose connection

Test point
hand control module
Expected
Coiled hose plugged in

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No table functions operational patient safety

Pressing buttons on Hand Control Module produces no movement or response from the surgical table

Ranked causes

  1. 1. Input power lost or disconnected (common)
  2. 2. Column connector disconnected or loose (common)
  3. 3. Pneumatic tube disconnection or blockage in Junction Box (occasional)
First check

Audible click when button pressed

Test point
Inside Junction Box
Expected
Audible click heard

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Retractable table base will not collapse caution

Retractable center beam cannot be retracted into base

Ranked causes

  1. 1. Center beam excessive soil or debris (common)
  2. 2. Roller excessive soil, debris or damaged (common)
  3. 3. Lock handle not released or over-tightened (occasional)
First check

Inspect center beam for soil and debris

Test point
center beam outer and inner surfaces
Expected
Clean surfaces without obstruction

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotation friction control will not stop rotation patient safety

Head or foot end rotation cannot be locked using friction control brake

Ranked causes

  1. 1. Brake tension insufficient or control not fully tightened (common)
  2. 2. Foot end control handle out of proper range (occasional)
  3. 3. Brake band worn or damaged (occasional)
First check

Tighten friction control lever to maximum

Test point
friction control lever
Expected
Lever at full tightened position

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Table top cannot be attached to H-frame at foot end caution

Table top attachment to foot end H-frame cannot be completed

Ranked causes

  1. 1. H-frame will not connect to crossbar (common)
  2. 2. T-pin bent or misaligned (occasional)
  3. 3. Head and foot end columns misaligned (occasional)
First check

Verify H-frame connection to crossbar

Test point
H-frame to crossbar connection
Expected
H-frame fully connected to crossbar

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transfer lock will not insert into crossbar patient safety

Patient transfer safety lock lever cannot be inserted into crossbar at head or foot end

Ranked causes

  1. 1. Crossbar pivot shaft bent or misaligned (occasional)
  2. 2. Transfer lock has burr or scratch preventing insertion (occasional)
  3. 3. Transfer lock spring weakened or broken (rare)
First check

Inspect crossbar pivot shaft for straightness

Test point
crossbar pivot shaft
Expected
Shaft straight and aligned

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot End Slider movement requires excessive or insufficient force

Foot End Slider requires more or less than 20 to 25 lbs. force to move the Crossbar along the column top plate

Ranked causes

  1. 1. Drag shoe friction setting requires adjustment (common)
  2. 2. Incorrect force measurement technique (occasional)
First check

Measure Foot End Slider movement force

Test point
Foot End Slider and Crossbar
Expected
20 to 25 lbs. force required, ±2.5 lbs.
Tools
Force gauge

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

H-frame will not connect to crossbar caution

H-frame attachment to head or foot end crossbar cannot be completed

Ranked causes

  1. 1. Mounting pin bent or misaligned (common)
  2. 2. H-frame damaged or deformed (occasional)
First check

Inspect mounting pin for straightness

Test point
H-frame mounting pin
Expected
Pin straight and properly aligned

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Rotational movement restricted or cannot lock at maximum friction

Head End or Foot End Crossbar cannot be rotated or does not lock completely when Rotational Friction Control Lever set to MAX position

Ranked causes

  1. 1. Head End Friction Control wear requiring adjustment (common)
  2. 2. Foot End Friction Control wear requiring adjustment (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Single table function inoperative patient safety

One function button on Hand Control Module does not actuate corresponding table movement while other functions work normally

Ranked causes

  1. 1. Air leak in Hand Control Module hose lumen (common)
  2. 2. Hand Control Module internal failure (occasional)
First check

Audible click when suspected button pressed

Test point
Inside Junction Box
Expected
No audible click

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Caster will not lock or unlock patient safety

Caster locking mechanism cannot engage or disengage

Ranked causes

  1. 1. Caster damaged or worn (common)
First check

Inspect caster and locking mechanism for damage

Test point
caster wheel assembly
Expected
Caster intact with functional locking mechanism

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Thermal switch activatedColumn motor automatically stopped due to continuous operation exceeding 60-80 secondsColumn operated continuously for longer than 60 to 80 secondsinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI