Mindray Trio: troubleshooting and error codes
The ServiceForge knowledge base holds 80 documented symptoms for the Mindray Trio, with 93 ranked causes, 30 diagnostic checks with test points and expected values, and 85 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Black screen during operation with power and fan running
Display shows black or no content while device is running; power indicator and fan both operating normally.
Ranked causes
- 1. Main control board failure (occasional)
- 2. Display failure (occasional)
- 3. HV inverter PCB damage (occasional)
Connect external VGA display to monitor output
- Test point
- VGA connector
- Expected
- External display shows image if CPU and inverter are functional
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: IBP Pressure patient safety
Verify accurate blood pressure measurements via invasive pressure monitoring. System displays calibration completion messages (SUCCESSFUL CALIBRATE, SENSOR OFF FAIL, IN DEMO FAIL, PRESSURE OVER RANGE FAIL).
Ranked causes
- 1. IBP transducer not connected or sensor offline (occasional)
- 2. Monitor in DEMO mode (occasional)
- 3. Incorrect transducer value selected (occasional)
Verify pressure transducer connection
- Test point
- IBP socket on right side panel
- Expected
- Transducer Interface Cable securely connected
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device responds slowly
System response time is sluggish; commands take longer to execute than normal.
Ranked causes
- 1. Poor performance of power board (common)
- 2. Poor performance of main control board (occasional)
- 3. Poor connection of power supply or CPU PCB (common)
Verify power supply output voltages
- Test point
- Power board connector pins
- Expected
- 5V, 12V, 18V within tolerance
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display after power-on
Display shows nothing when device is powered on; power indicator may or may not be lit.
Ranked causes
- 1. Blown fuse (common)
- 2. Power Supply defective (common)
- 3. Main control board failure or display failure (occasional)
Inspect fuse for visible damage or open circuit
- Test point
- Fuse location on power supply board
- Expected
- Fuse intact and continuous
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder will not print
Recording/printing device does not produce output on paper.
Ranked causes
- 1. Recorder has no paper or roller lever is not pressed down (common)
- 2. Recorder failure (occasional)
- 3. Recorder cable is damaged (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keys or Navigator knob disabled
Keyboard input or navigation knob controls do not respond to user input.
Ranked causes
- 1. Keypad or Navigator knob is damaged (common)
- 2. Keypad cable is damaged (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operation or measurement function disabled
Specific monitoring or measurement feature is not responding or operational.
Ranked causes
- 1. CPU PCB damage (occasional)
- 2. Corresponding parameter PCB damage (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sound is distorted or mute
Audio output is either not present or quality is poor with audible distortion.
Ranked causes
- 1. Keypad failure (occasional)
- 2. Speaker or cable failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm limit error
Alarm limit of parameter has deviated from standard range
Ranked causes
- 1. Alarm settings deviation from standard (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage too low
Monitor automatically shuts down within 5 minutes when battery voltage is insufficient
Ranked causes
- 1. Battery depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG Amplitude patient safety
Verify ECG signal amplitude accuracy at different gain settings (x1, x2, x0.25, x0.5). Measure displayed calibration pulse with metric ruler; pulse should be 1 cm ± 5% at x1 gain.
Ranked causes
- 1. ECG amplifier gain or signal path out of calibration (rare)
Set ECG lead to I and ECG size to x1
- Test point
- ECG parameter settings
- Expected
- ECG lead I selected, gain set to x1
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP Pneumatic patient safety
Test for air leaks in NIBP pneumatic airway per EN 1060-1 standard. System inflates to ~180 mmHg, deflates after 20 seconds, and reports any pneumatic leaks.
Ranked causes
- 1. Air leaks in NIBP pneumatic system (occasional)
Connect test chamber to NIBP quick connect fitting
- Test point
- NIBP quick connect fitting on Trio Monitor
- Expected
- Test chamber (P/N 0138-00-0001-03) securely connected via hose
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP Pressure patient safety
Verify non-invasive blood pressure measurement accuracy. Measure difference between digital mercury manometer readings and monitor NIBP parameter display at 0, 50, and 200 mmHg.
Ranked causes
- 1. NIBP module pressure measurement out of calibration (unknown)
Connect test chamber and manometer to NIBP quick connect fitting
- Test point
- NIBP quick connect fitting on Trio Monitor
- Expected
- Test chamber (P/N 0138-00-0001-03) and calibrated digital mercury manometer securely connected via T fitting
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cell battery voltage too high
Cell battery is defective or incorrect type installed
Ranked causes
- 1. Defective or incorrect cell battery (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cell battery voltage too low
Cell battery is depleted, not installed, or connection is loose
Ranked causes
- 1. Depleted, missing or loose cell battery (common)
Defective display cable
Visible damage or poor connection in display ribbon cable causes no display or black screen.
Ranked causes
- 1. Display cable damaged or disconnected (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG C lead off
C lead wire of ECG is not connected correctly
Ranked causes
- 1. C lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG F lead off
F lead wire of ECG is not connected correctly
Ranked causes
- 1. F lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG L lead off
L lead wire of ECG is not connected correctly
Ranked causes
- 1. L lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG LA lead off
LA lead wire of ECG is not connected correctly
Ranked causes
- 1. LA lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG lead off
ECG lead is not connected correctly
Ranked causes
- 1. Lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG LL lead off
LL lead wire of ECG is not connected correctly
Ranked causes
- 1. LL lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG noise excessive
Excessive interference appears on ECG signals
Ranked causes
- 1. Poor electrode or lead wire connection or patient movement (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG R lead off
R lead wire of ECG is not connected correctly
Ranked causes
- 1. R lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG RA lead off
RA lead wire of ECG is not connected correctly
Ranked causes
- 1. RA lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG V lead off
V lead wire of ECG is not connected correctly
Ranked causes
- 1. V lead wire disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG weak signal patient safety
Patient's ECG signal is too weak to detect properly
Ranked causes
- 1. Poor electrode or lead wire connection (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan does not run caution
Cooling fan motor does not spin when device is powered on.
Ranked causes
- 1. Defective component or PCB (common)
Verify fan power supply
- Test point
- Fan connector on power board
- Expected
- 18V DC present at fan connector
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse blows after connecting specific PCB high voltage
Fuse opens only when a particular module or PCB is connected to the system.
Ranked causes
- 1. PCB may be defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse blows upon power-on high voltage
Fuse opens immediately or very shortly after device is powered on.
Ranked causes
- 1. Short-circuit in power supply or other module (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse blows with all loads disconnected high voltage
Fuse opens even when all peripheral modules are disconnected from power board.
Ranked causes
- 1. Power Supply failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP lead off
IBP cable is not connected correctly
Ranked causes
- 1. Cable disconnection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP sensor removal alarm patient safety
Unit displays sensor off message in Message Area when IBP sensor is disconnected during operation.
Ranked causes
- 1. IBP sensor disconnected or loose (common)
Verify 'IBP SENSOR OFF' message displays in Message Area when sensor unplugged
- Test point
- Message area top right corner of display
- Expected
- 'IBP SENSOR OFF' message displayed in Message Area
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard communication error caution
The keypad has a failure
Ranked causes
- 1. Keyboard communication failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard not available caution
The keypad cannot be used
Ranked causes
- 1. Keyboard hardware failure or stuck keys (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP air leak
NIBP cuff not connected correctly or airway leaks present
Ranked causes
- 1. Poor cuff connection or airway leaks (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP communication error
No NIBP communication
Ranked causes
- 1. NIBP module communication failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff loose
NIBP cuff is not connected correctly
Ranked causes
- 1. Loose cuff connection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP cuff type error
NIBP cuff may be wrong size
Ranked causes
- 1. Incorrect cuff size for patient (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP illegal reset caution
Illegal reset occurs during NIBP measurement
Ranked causes
- 1. Airway occlusion or hardware issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP initialization error
NIBP initialization error occurred
Ranked causes
- 1. NIBP module initialization failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement failure
Problem occurs when measuring NIBP; system cannot perform measurement
Ranked causes
- 1. Poor connection or patient condition issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement timeout
Problem occurs when measuring NIBP; system cannot complete measurement
Ranked causes
- 1. Poor connection or patient condition issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP overpressure caution
NIBP is in an overpressure condition
Ranked causes
- 1. Occlusion in cuff or hose (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP pneumatic leak
NIBP airway has leaks
Ranked causes
- 1. Airway leaks in cuff or hose (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP signal saturated
Problem occurs when measuring NIBP; system cannot perform measurement or analysis
Ranked causes
- 1. Poor cuff/hose connection or patient condition issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP system failure
Problem occurs when measuring NIBP; system cannot perform measurement
Ranked causes
- 1. NIBP system hardware or communication failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pulse detected patient safety
Patient's pulse signal is too weak or non-existent
Ranked causes
- 1. Sensor connection failure (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker not paced (PNP) patient safety
Pacemaker is not delivering pacing output
Ranked causes
- 1. Pacemaker or electrode connection failure (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker signal not captured (PNC) patient safety
No pacemaker signal is captured by the monitor
Ranked causes
- 1. Pacemaker or electrode connection failure (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter range exceeded
Measured value of parameter has exceeded system measurement range
Ranked causes
- 1. Measurement out of range (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power and control board indicators dark but fan and keypad indicators lit
Power and main control board LEDs do not illuminate but fan runs and keypad LED is on.
Ranked causes
- 1. +5V DC power is damaged (common)
Measure +5V DC output
- Test point
- Power board output connector pins for 5V rail
- Expected
- 5V DC ±5%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power and control board indicators lit but fan and keypad indicators dark
Power and main control board LEDs illuminate but fan does not run and keypad LED is off.
Ranked causes
- 1. +12V DC power is damaged (common)
Measure +12V DC output
- Test point
- Power board output connector pins for 12V rail
- Expected
- 12V DC ±10%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power indicator does not light
Power LED on front panel does not illuminate when device is powered on.
Ranked causes
- 1. Power Supply defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltage too high or too low high voltage
Power supply has a failure affecting 5V, 12V or 3.3V rails
Ranked causes
- 1. Power supply voltage out of specification (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real clock needs set
System time is inaccurate, displayed as 2000-1-1
Ranked causes
- 1. System time not set or lost (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Real clock not exist
System has no cell battery or battery has run down
Ranked causes
- 1. Missing or depleted cell battery (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Record paper skews
Paper output from recorder is misaligned or skewed during printing.
Ranked causes
- 1. Bad recorder paper installation or positioning (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder communication error
Communication of the recorder is abnormal
Ranked causes
- 1. Recorder communication failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder head in wrong position
Recorder roller lever is not pressed down
Ranked causes
- 1. Roller lever not engaged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Recorder head too hot
- Recorder not available
- Recorder out of paper
- Recorder paper jammed
- Recorder paper not in correct position
- Recorder self test error
- Recorder voltage failure
- SpO2 board fault
- SpO2 communication error
- SpO2 interference detected
- SpO2 low perfusion
- SpO2 pulse search in progress
- SpO2 sensor fault
- SpO2 sensor off
- SpO2 sensor removal alarm
- SpO2 too much ambient light
- SpO2 unrecognized sensor
- System watchdog failure
- Temperature sensor off
- Waveforms displayed intermittently with normal characters
20 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
12V TOO HIGH | 12V power supply voltage too high | 12V supply exceeds specification | error |
12V TOO LOW | 12V power supply voltage too low | 12V supply below specification | error |
3.3V TOO HIGH | 3.3V power supply voltage too high | 3.3V supply exceeds specification | error |
3.3V TOO LOW | 3.3V power supply voltage too low | 3.3V supply below specification | error |
5V TOO HIGH | 5V power supply voltage too high | 5V supply exceeds specification | error |
5V TOO LOW | 5V power supply voltage too low | 5V supply below specification | error |
AIR LEAK | NIBP airway leak detected | Pressure drops indicating air leak | warning |
BATTERY VOLTAGE TOO LOW | Battery voltage is insufficient for operation | Battery charge falls below operational threshold | warning |
CELL BAT TOO HIGH | Cell battery voltage too high | Cell battery voltage exceeds specification | warning |
CELL BAT TOO LOW | Cell battery voltage too low | Cell battery charge depleted | warning |
CUFF TYPE ERR | NIBP cuff type mismatch | Cuff size incorrect for patient | error |
ECG C LEAD OFF | ECG C lead not connected | C lead signal disconnected | warning |
ECG F LEAD OFF | ECG F lead not connected | F lead signal disconnected | warning |
ECG LA LEAD OFF | ECG LA lead not connected | LA lead signal disconnected | warning |
ECG LEAD OFF | ECG lead not connected | Lead signal disconnected | warning |
ECG L LEAD OFF | ECG L lead not connected | L lead signal disconnected | warning |
ECG LL LEAD OFF | ECG LL lead not connected | LL lead signal disconnected | warning |
ECG NOISE | Excessive ECG signal interference | Interference above threshold detected | warning |
ECG RA LEAD OFF | ECG RA lead not connected | RA lead signal disconnected | warning |
ECG R LEAD OFF | ECG R lead not connected | R lead signal disconnected | warning |
ECG V LEAD OFF | ECG V lead not connected | V lead signal disconnected | warning |
ECG WEAK SIGNAL | Patient ECG signal is too weak | ECG amplitude below detection threshold | warning |
IBP LEAD OFF | IBP cable not connected | IBP cable signal lost | warning |
KEYBOARD COMM ERR | Keyboard communication error | Keyboard communication failure | error |
KEYBOARD ERR1 | Keyboard error 1 | Keyboard error condition 1 | error |
KEYBOARD ERR2 | Keyboard error 2 | Keyboard error condition 2 | error |
KEYBOARD ERROR | Keyboard generic error | Keyboard operation failure | error |
KEYBOARD FAILURE | Keyboard failure | Keyboard hardware failure | error |
KEYBOARD NOT AVAILABLE | Keyboard not usable | Keyboard interface failure | warning |
LOOSE CUFF | NIBP cuff not connected properly | Cuff connection lost during operation | warning |
MEASURE FAIL | NIBP measurement failed | Measurement could not complete | error |
NIBP COMM ERR | NIBP communication error | Host-NIBP communication failure | error |
NIBP ILLEGALLY RESET | NIBP module received illegal reset | Reset signal received during measurement | error |
NIBP INIT ERR | NIBP initialization error | NIBP module boot failure | error |
NIBP SELFTEST ERR | NIBP self-test error | NIBP module self-test fails | error |
NIBP SYSTEM FAILURE | NIBP system failure | NIBP subsystem failure | critical |
NO PULSE | Patient pulse signal not detected | Pulse signal too weak or absent | warning |
OVER PRESSURE | NIBP over-pressure condition | Pressure exceeds safe limits | error |
PNC | Pacemaker signal not captured | Pacemaker output not detected | error |
PNEUMATIC LEAK | NIBP pneumatic leak detected | Sustained pressure drop indicating leak | error |
PNP | Pacemaker not paced | Pacemaker not delivering output | error |
POWER ERR3 | Power supply error 3 | Power supply error condition 3 | error |
POWER ERR4 | Power supply error 4 | Power supply error condition 4 | error |
POWER ERR7 | Power supply error 7 | Power supply error condition 7 | error |
POWER ERR8 | Power supply error 8 | Power supply error condition 8 | error |
REAL CLOCK NEED SET | System time is inaccurate | System displays 2000-1-1 or time verification fails | warning |
REAL CLOCK NOT EXIST | Cell battery missing or depleted | Real-time clock battery failure | warning |
REC HEAD IN WRONG POSITION | Recorder head in wrong position | Recorder roller lever not engaged | warning |
REC NOT AVAILABLE | Recorder not available for communication | Recorder communication unavailable | error |
RECORDER COMM ERR | Recorder communication error | Host-recorder communication failure | error |
RECORDER HEAD HOT | Recorder head temperature too high | Recorder head thermal limit exceeded | warning |
RECORDER OUT OF PAPER | Recorder paper depleted | Paper end detected | info |
RECORDER PAPER JAM | Recorder paper jammed | Paper jam detected during operation | warning |
RECORDER PAPER W.P. | Recorder paper in wrong position | Paper position verification fails | warning |
RECORDER SELF TEST ERR | Recorder module self-test failure | Recorder self-test fails during boot | error |
RECORDER VLT HIGH | Recorder module voltage too high | Recorder power supply exceeds spec | error |
RECORDER VLT LOW | Recorder module voltage too low | Recorder power supply below spec | error |
SIGNAL SATURATED | NIBP signal saturation during measurement | Signal exceeds dynamic range | warning |
SpO2: BOARD FAULT | MasimaSET board hardware failure | Board self-test failure | critical |
SpO2: COMMUNICATION ERROR | SpO2 module communication failure | Communication link between monitor and SpO2 module lost | error |
SpO2: INTERFERENCE | Noise detected on SpO2 pulse signal | Pulse signal corrupted by interference | warning |
SpO2: LOW PERFUSION | Patient perfusion is low | Inadequate blood perfusion at sensor site | warning |
SpO2: PULSE SEARCH | SpO2 module searching for valid pulse | Module adjusting settings to detect pulse | info |
SpO2: SENSOR FAULT | SpO2 sensor is defective | Sensor self-test failure | error |
SpO2 SENSOR OFF | SpO2 sensor not connected | SpO2 sensor signal lost | warning |
SpO2: TOO MUCH LIGHT | Excessive ambient light detected | Room light interferes with sensor | warning |
SpO2: UNRECOGNIZED SENSOR | Sensor not recognized by monitor | Sensor identification fails | error |
SYSTEM CMOS ERR | CMOS error detected | CMOS data corruption or failure | error |
SYSTEM CMOS FULL | CMOS memory full | CMOS storage capacity exceeded | error |
SYSTEM EPGA FAILURE | FPGA initialization failure | FPGA boot or operation failure | error |
SYSTEM FAILURE10 | System generic failure 10 | Generic system error condition 10 | error |
SYSTEM FAILURE11 | System generic failure 11 | Generic system error condition 11 | error |
SYSTEM FAILURE12 | System generic failure 12 | Generic system error condition 12 | error |
SYSTEM FAILURE2 | System generic failure 2 | Generic system error condition 2 | error |
SYSTEM FAILURE3 | System generic failure 3 | Generic system error condition 3 | error |
SYSTEM FAILURE4 | System generic failure 4 | Generic system error condition 4 | error |
SYSTEM FAILURE5 | System generic failure 5 | Generic system error condition 5 | error |
SYSTEM FAILURE6 | System generic failure 6 | Generic system error condition 6 | error |
SYSTEM FAILURE7 | System generic failure 7 | Generic system error condition 7 | error |
SYSTEM FAILURE8 | System generic failure 8 | Generic system error condition 8 | error |
SYSTEM FAILURE9 | System generic failure 9 | Generic system error condition 9 | error |
SYSTEM SOFTWARE ERR | System software error | Software execution failure | error |
SYSTEM WD FAILURE | System watchdog detected fatal error | Watchdog timeout or system crash | critical |
TEMP SENSOR OFF | Temperature sensor not connected | Temperature sensor signal lost | warning |
TIME OUT | NIBP measurement timeout | Measurement exceeds time limit | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI