Mindray Passport V: troubleshooting and error codes
The ServiceForge knowledge base holds 115 documented symptoms for the Mindray Passport V, with 161 ranked causes, 33 diagnostic checks with test points and expected values, and 9 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
CHK Lead message displayed
Monitor displays 'CHK Lead' warning; indicates elevated electrode impedance
Ranked causes
- 1. Increased impedance caused by chest hair under electrodes (common)
- 2. Dried electrode gel (common)
- 3. Electrode off or lead off (occasional)
- 4. Cracked lead wires (rare)
- 5. Poor skin prep (common)
Inspect electrode sites for chest hair
- Test point
- Chest wall under electrodes
- Expected
- No visible chest hair
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Damped invasive waveform
IBP waveform appears flattened, rounded, or with reduced amplitude indicating poor signal quality
Ranked causes
- 1. Air bubbles in tubing (common)
- 2. Kinked catheter (common)
- 3. Catheter against wall of blood vessel (occasional)
- 4. Blood in tubing (occasional)
- 5. Catheter partially occluded with clot (occasional)
Inspect tubing for air bubbles
- Test point
- Entire pressure line from transducer to catheter
- Expected
- No visible air bubbles
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IBP not displayed or no IBP waveform
Invasive blood pressure signal absent from monitor display; no waveform or numerical values shown
Ranked causes
- 1. Improper setup (common)
- 2. Cable not plugged in (common)
- 3. Transducer not connected (common)
- 4. Stopcock turned improperly (occasional)
- 5. Transducer not zeroed (common)
Verify IBP display enabled
- Test point
- Monitor setup menu
- Expected
- IBP channel enabled
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baseline wander in ECG patient safety
ECG baseline drifts and fluctuates during waveform display
Ranked causes
- 1. Patient moving excessively (common)
- 2. Patient respiration or dry/loose electrodes (common)
- 3. Static buildup around patient (occasional)
- 4. ECG filter set to incorrect mode (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive alarms for heart rate or lead fault patient safety
Frequent false alarms triggered for heart rate or lead disconnection
Ranked causes
- 1. Electrodes dry (common)
- 2. Alarm limits set too close to patient's normal heart rate (common)
- 3. R-wave wrong size (occasional)
- 4. Excessive patient movement or muscle tremor (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display appears to be off
Monitor screen is dark or does not power on; no video output visible
Ranked causes
- 1. Mains power switch not activated (common)
- 2. Power cord not connected to AC outlet (common)
- 3. Battery pack drained (portable operation) (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal invalid patient safety
Monitor does not recognize valid ECG signal from electrode connections
Ranked causes
- 1. Connections not tight or properly secured (common)
- 2. Electrodes dry or loose (common)
- 3. Cable or lead wires damaged (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
False apnea alarm patient safety
System triggers apnea alarm when patient is actually breathing normally
Ranked causes
- 1. Apnea delay may be improperly set (common)
- 2. Patient may be having frequent episodes of CVA (occasional)
- 3. Scale size may be too low (common)
Review apnea delay setting
- Test point
- Monitor setup menu
- Expected
- Appropriate apnea delay for patient type
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas Module air leak detected caution
System displays 'GM: Air Leak' message; pneumatic leak detected or sample line not attached
Ranked causes
- 1. Pneumatic leak in system (common)
- 2. Sample line not attached or disconnected (occasional)
- 3. Gas Module on without Monitor for extended period (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG waveform patient safety
ECG waveform not displayed on monitor
Ranked causes
- 1. Gain set too low (occasional)
- 2. Lead wires and patient cable not fully or properly inserted (common)
- 3. Cable or lead wires damaged (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No respiration waveform or rate displayed patient safety
Respiration signal absent from monitor display; no numerical rate shown
Ranked causes
- 1. Respiration turned off (common)
- 2. Patient connected using ESIS choke cable (occasional)
- 3. Cable not connected (common)
Verify respiration is enabled
- Test point
- Monitor settings menu
- Expected
- Respiration enabled; 'Off' not displayed in Resp window
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ECG traces patient safety
ECG waveform contains noise and artifacts
Ranked causes
- 1. Loose or dry electrodes (common)
- 2. Defective electrode wires (occasional)
- 3. Patient cable routed too close to electrical devices (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote view LAN connectivity issue
Patient monitor connected to LAN but cannot view other patients in Remote View window
Ranked causes
- 1. Incorrect LAN cable connection or cable too long (common)
- 2. Excessive concurrent remote view requests (occasional)
- 3. Incorrect IP address configuration or conflict (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration high impedance
System detects abnormally high electrode impedance in respiration circuit
Ranked causes
- 1. Connections not tight and/or properly secured (common)
- 2. Electrodes dry or loose (common)
- 3. Cable or lead wires damaged (occasional)
Inspect all connections
- Test point
- Electrode to lead, lead to cable, cable to monitor
- Expected
- All connectors firmly seated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trends not printing
Trend data not output to printer when Print Trend requested
Ranked causes
- 1. Print Trend button not pressed (common)
- 2. No trends available to print (occasional)
- 3. Printer out of paper (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm tone disabled patient safety
Alarm audible alert not sounding; alarm silence feature active or volume reduced
Ranked causes
- 1. Alarm silence key pressed (common)
- 2. Beep volume set too low (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CVA message displayed
Monitor displays CVA (Cardiopulmonary variability/artifact) message indicating respiration-cardiac interference
Ranked causes
- 1. Can be caused by shallow breathing or an apnea event (occasional)
- 2. Patient HR and respiratory rate identical (rare)
Observe patient respiratory pattern
- Test point
- Patient bedside observation
- Expected
- Normal steady breathing pattern
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG artifact patient safety
ECG waveform contains distortion and artifact
Ranked causes
- 1. Electrical interference from auxiliary devices (common)
- 2. Patient movement (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG noise patient safety
ECG signal contains noise and interference
Ranked causes
- 1. External interference (occasional)
- 2. Patient movement (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas Module sampling line occlusion caution
System displays 'GM: Occlusion' message; obstruction detected in sampling line or water trap bottle full
Ranked causes
- 1. Water trap bottle full or sampling line blocked (common)
- 2. Exhaust line blocked (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No trace for desired parameter
Desired waveform or parameter trace not displayed on monitor screen
Ranked causes
- 1. Improper transducer or cable attachment (common)
- 2. Faulty transducer or cable (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient information not displayed patient safety
Patient demographics or data not visible on monitor screen after entry
Ranked causes
- 1. No patient data entered (common)
- 2. Done button not selected after data entry (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform too small
Respiratory signal displayed at insufficient amplitude; difficult to discern on monitor
Ranked causes
- 1. Patient breathing is shallow or patient is turned on side (common)
- 2. Scale set inappropriately (common)
Assess patient position and breathing depth
- Test point
- Patient observation
- Expected
- Patient upright with normal breathing pattern
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormally high or low IBP readings patient safety
Blood pressure values displayed are significantly elevated or reduced compared to expected patient values
Ranked causes
- 1. Transducer too high or too low (common)
Verify transducer level relative to patient heart
- Test point
- Transducer height
- Expected
- Transducer at same level as patient right atrium
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 module
Verify CO2 measurement accuracy; recommended annually or when readings appear suspicious
Ranked causes
- 1. CO2 module calibration drift (unknown)
Warm CO2 module
- Test point
- Module warm-up time
- Expected
- Minimum 20 minutes at power-on
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG channels
ECG signal inaccuracy; wave amplitude greater or smaller than expected
Ranked causes
- 1. ECG module hardware or software problem (unknown)
Compare ECG square wave amplitude to scale
- Test point
- Monitor screen
- Expected
- Difference within 5%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas module
Verify Gas Module 3 accuracy; recommended annually or when readings appear suspicious
Ranked causes
- 1. Gas Module 3 calibration drift (unknown)
Warm Gas Module 3
- Test point
- Module warm-up time
- Expected
- Minimum 10 minutes at power-on
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: IBP transducer
Verify invasive blood pressure transducer sensitivity for optimal accuracy
Ranked causes
- 1. IBP transducer calibration drift (unknown)
Zero calibration with simulator at 0 mmHg
- Test point
- Module pressure connector
- Expected
- Zero calibration accepted
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP accuracy
NIBP pressure readings may be inaccurate
Ranked causes
- 1. NIBP module accuracy drift (unknown)
Compare NIBP displayed pressure to manometer reference at 300 mmHg
- Test point
- Monitor screen vs. manometer
- Expected
- Difference not greater than +/-3 mmHg, +/-3 mmHg
- Tools
- Digital/Mercury manometer, test chamber
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP leakage
Verify pneumatic system does not leak
Ranked causes
- 1. Pneumatic leak in NIBP system (unknown)
Observe leakage test result message
- Test point
- Monitor screen prompt message area
- Expected
- No leakage detected; no NIBP Pneumatic Leak message
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NIBP transducer
Verify pressure transducer sensitivity for optimal NIBP accuracy
Ranked causes
- 1. NIBP transducer calibration drift (unknown)
Set calibration reference pressure
- Test point
- Calibration menu
- Expected
- Reference pressure 240-260 mmHg (default 250 mmHg), 240-260 mmHg, 1 mmHg steps
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check remote printer
Remote printer is busy, disconnected, out of paper, or has fault condition
Ranked causes
- 1. Remote printer operational issue (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 15V overrange high voltage
Microstream CO2 module 15V power supply voltage is out of acceptable range
Ranked causes
- 1. Power supply to module is out of range (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 calibration caused by measurement error
Microstream CO2 calibration failed; sensor unable to measure gas concentration accurately
Ranked causes
- 1. Measurement error (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 calibration caused by no gas or wrong gas concentration
Microstream CO2 calibration failed due to absent or incorrect calibration gas supply
Ranked causes
- 1. No gas or wrong gas concentration (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 calibration caused by no stable gas flow
Microstream CO2 calibration failed; gas flow rate unstable during procedure
Ranked causes
- 1. No stable gas flow (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 calibration completed successfully
CO2 module zero or span calibration procedure completed without error
Ranked causes
- 1. Calibration finished and successful (common)
CO2 calibration error
CO2 calibration failed; actual gas concentration differs significantly from entered value
Ranked causes
- 1. The actual concentration of calibration gas applied is in wide discrepancy with the concentration value entered by the user (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 check airway
Microstream CO2 module detects airway circuit error or disconnection
Ranked causes
- 1. Airway error (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 check flow rate
Microstream CO2 module detects abnormal (high or low) sample gas flow rate
Ranked causes
- 1. The system has detected a high or low flow rate (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 check sensor
Microstream CO2 sensor is faulty or malfunctioning; unreliable readings
Ranked causes
- 1. Possible faulty sensor (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 communication error
Monitor receives incorrect or corrupted response from CO2 module
Ranked causes
- 1. The monitor receives wrong response from CO2 module (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 communication stop
CO2 module communication has stopped; command transmission failing
Ranked causes
- 1. Sending out order is failed during work (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 filtering disconnected
Microstream CO2 filterline is disconnected from monitor module
Ranked causes
- 1. Filtering disconnected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 filterline error
DPM CO2 filterline is leaking or blocked; cannot deliver sample gas
Ranked causes
- 1. Filterline is leaked or blocked (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 hardware error
DPM CO2 module hardware malfunction; device cannot operate
Ranked causes
- 1. Module has problems (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 high airway pressure
DPM flux sensor measurement exceeds 790 mmHg; airway pressure abnormally elevated
Ranked causes
- 1. Measurement of Flux Sensor in module is higher than 790 mmHg (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 high barometric pressure
DPM detects atmospheric pressure higher than 790 mmHg; device location unsuitable
Ranked causes
- 1. Atmospheric pressure is too high (higher than 790 mmHg) (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 initialization error
CO2 module fails to respond during startup; initialization sequence incomplete
Ranked causes
- 1. No response after send order during initialization (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 low airway pressure
DPM flux sensor measurement is below 400 mmHg; airway pressure abnormally low
Ranked causes
- 1. Measurement of Flux Sensor in module is less than 400 mmHg (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 low barometric pressure
DPM detects atmospheric pressure lower than 428 mmHg; device location unsuitable
Ranked causes
- 1. Atmospheric pressure is too low (lower than 428 mmHg) (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 main board error
Microstream CO2 module main board malfunction detected
Ranked causes
- 1. Module has problems (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 no water trap
DPM CO2 module detects water trap is missing or not installed
Ranked causes
- 1. No watertrap on CO2 module (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 not ready for calibration
CO2 module unable to initialize calibration sequence; not in proper state
Ranked causes
- 1. Module is unable to initialize calibration (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 occlusion detected
Microstream CO2 module detects occlusion (blockage) in sample line preventing gas flow
Ranked causes
- 1. Sample line is blocked or kinked (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 purge required
Microstream CO2 module detects blockage or kink in sample line during purge cycle
Ranked causes
- 1. Sample line is blocked or kinked (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 replace scrubber and pump
Microstream CO2 module scrubber and pump have reached end of service life
Ranked causes
- 1. Module has problems (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 self-test error
CO2 module fails internal self-test; software or hardware malfunction detected
Ranked causes
- 1. Module cannot work because of software fault, circuit error, etc. (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor high temperature caution
DPM CO2 sensor temperature exceeds upper operating limit
Ranked causes
- 1. Temperature over ranges high limit checked by module (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor low temperature caution
DPM CO2 sensor temperature is below minimum operating limit
Ranked causes
- 1. Temperature over ranges low limit checked by module (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- CO2 startup
- CO2 temperature overrange
- CO2 warming up
- CO2 zero failed
- CO2 zeroing in progress
- Cooling fan failure
- Excessive electrosurgical interference
- Gas module agent zero error
- Gas module agents uncalibrated
- Gas Module cannot zero
- Gas module CO2 uncalibrated
- Gas module CO2 zero error
- Gas Module communication error
- Gas Module disconnected
- Gas Module exhaust blocked
- Gas module N2O uncalibrated
- Gas module N2O zero error
- Gas Module not ready for calibration
- Gas module O2 uncalibrated
- Gas module O2 zero error
- Gas Module pump turned off
- Gas Module sampling error during calibration
- Gas Module unrecoverable error
- Gas Module warming up
- Gas Module zeroing error during calibration
- IBP communication error
- IBP communication stop
- IBP initialization error
- Incorrect date or time displayed
- Local printer busy
- Local printer communication error
- Local printer door open
- Local printer out of paper
- Local printer report appears blank
- Local printer self-test error
- Local printer unable to print
- Monitor log review
- Muscle noise in ECG
- No print on alarm
- No printer selected or available
- No trends displayed
- Patient data missing after power cycle
- Pressure overrange cannot zero
- Printer buffer full
- Pulsatile pressure cannot zero
- Remote printer not available
- Respiration communication error
- Respiration communication stop
- Respiration initialization error
- Respiration waveform too large
- Sensor off
- Sensor off cannot zero
- Thermal printhead overheated
- Unknown anesthetic agent detected
- Water trap requires replacement
55 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Communication Error | ECG module unable to send commands correctly | The ECG module communication error. The command cannot be sent correctly. | error |
Communication Stop | ECG module communication is not operational | As the ECG module communication stops | error |
ECG Channels Checking | ECG module is performing calibration procedure | Appears when the ECG Module is calibrating. | info |
ESU-Resp Off | Respiration measurement disabled due to electrosurgery interference | The high frequency electrosurgery unit interference signal is detected. | warning |
Initialization Error | System cannot establish communication with ECG module during startup | During the ECG module power-on, as the ECG module communication stops | error |
Learning | ECG module performing arrhythmia or ST learning cycle | Displayed when a learning cycle has been requested for Arrhythmia or ST. | info |
No Arrhythmia Detection at Central | Central station lacks arrhythmia analysis capability | Central Station does not have arrhythmia Analysis capability. | info |
Pacer Rejection On | Pacemaker rejection filter is enabled | When Pacer Reject is set to On. | info |
Self Test Error | ECG module unable to complete self-test during power-on | During power-on of the unit, if ECG module cannot self-test successfully | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI