Mindray Passport 5-Lead: troubleshooting and error codes
The ServiceForge knowledge base holds 155 documented symptoms for the Mindray Passport 5-Lead, with 227 ranked causes, 298 diagnostic checks with test points and expected values, and 15 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
CHK Lead message displayed
Monitor displays 'CHK Lead' error message
Ranked causes
- 1. Increased impedance from poor skin preparation (common)
- 2. Chest hair under electrodes (common)
- 3. Dried electrode gel (common)
- 4. Electrode off (common)
- 5. Lead off (common)
- 6. Cracked lead wires (occasional)
Inspect skin preparation
- Test point
- Skin under electrode pads
- Expected
- Skin clean, abraded, and dry
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveform acquisition failure
Monitor unable to acquire and process waveform data
Ranked causes
- 1. Front end failure (rare)
- 2. External supply failure (rare)
- 3. +12/-23 volt board failure (rare)
- 4. CPU board failure (rare)
- 5. CO2 interface board failure (rare)
Check front end board
- Test point
- Monitor circuit board assembly
- Expected
- Front end board functioning without error
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Baseline wander
Baseline shifts up and down, obscuring waveform details
Ranked causes
- 1. Patient moving excessively (common)
- 2. Patient respiration (common)
- 3. Electrodes dry (common)
- 4. Static buildup around patient (rare)
Secure lead wires and cable
- Test point
- Lead routing to patient
- Expected
- Leads secured and not moving with patient
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 adapter installation or condition problem
CO2: CHECK ADAPTER message displayed, sensor not correctly positioned on adapter or adapter issue detected
Ranked causes
- 1. Sensor not correctly positioned on adapter (common)
- 2. Incorrect or different airway adapter type installed (occasional)
- 3. Dirty or damaged adapter (occasional)
- 4. Persistent installation or compatibility issue (occasional)
Verify sensor position on airway adapter
- Test point
- Sensor and adapter interface
- Expected
- Sensor properly seated on adapter connector
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 waveform not displaying
Capnostat CO2 sensor connected but waveform does not appear on monitor display
Ranked causes
- 1. Sensor not warmed up (common)
- 2. Incorrect Resp Source setting (common)
- 3. Airway adapter windows blocked or dirty (occasional)
- 4. Adapter calibration required (occasional)
Verify sensor warm-up status
- Test point
- Monitor display
- Expected
- 'Sensor Warming Up' message should disappear before numeric data available
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Damped invasive waveform caution
Invasive pressure waveform appears flattened or attenuated
Ranked causes
- 1. Air bubbles in tubing (common)
- 2. Kinked catheter (common)
- 3. Catheter against wall of blood vessel (occasional)
- 4. Blood in tubing (occasional)
Inspect tubing for air bubbles
- Test point
- Pressure tubing between catheter and transducer
- Expected
- No air bubbles visible
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive heart rate and lead fault alarms
Frequent false alarms for heart rate or lead faults
Ranked causes
- 1. Electrodes dry (common)
- 2. Alarm limits set too close to patient normal heart rate (common)
- 3. R-wave wrong size (occasional)
- 4. Excessive patient movement or muscle tremor (occasional)
Inspect electrode condition
- Test point
- Electrode pads on patient
- Expected
- Electrodes moist and properly adhered
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interference on ECG
Poor electrode site preparation, improper lead selection, or excessive 50/60 Hz electrical noise on ECG waveform
Ranked causes
- 1. Poor electrode site preparation (common)
- 2. Improper lead selection (common)
- 3. Excessive 50 or 60 Hz electrical noise (common)
- 4. Wrong ECG cable in use (occasional)
Inspect electrode sites
- Test point
- Electrode placement areas
- Expected
- Clean skin with good contact
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low amplitude ECG signal
ECG waveform displayed with very small amplitude
Ranked causes
- 1. Gain set too low (common)
- 2. Skin improperly prepared (common)
- 3. Not patient's normal ECG complex (rare)
- 4. Electrode positioned over bone or muscle mass (occasional)
Verify gain setting
- Test point
- SIZE key display
- Expected
- Gain at appropriate level for visible ECG waveform
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No respiration waveform or rate displayed patient safety
Respiration monitoring not functioning; no respiratory data shown
Ranked causes
- 1. Patient not connected to patient safety cable (common)
- 2. Respiration parameter turned off (common)
- 3. CO2 off and no CO2 module (occasional)
- 4. Patient connected using ESIS cable (occasional)
Verify patient cable connection
- Test point
- Patient cable to monitor
- Expected
- Patient safety cable properly connected
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC noise on ECG signal
Excessive AC noise visible on ECG waveform display
Ranked causes
- 1. Gain set too high (common)
- 2. Electrodes dry (common)
- 3. Patient cable entwined with other electrical devices (occasional)
Verify gain setting
- Test point
- SIZE key display
- Expected
- Gain at appropriate level for patient signal
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Artifact on ECG waveform
50 or 60 Hz noise detected on ECG; displayed heart rate may be inaccurate
Ranked causes
- 1. Poor electrode site preparation (common)
- 2. Improper lead selection (common)
- 3. Excessive electrical noise in environment (common)
Assess electrode placement
- Test point
- Electrode sites and contact
- Expected
- Good skin contact on all electrodes
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 occlusion detected in sampling system caution
Sampling pump line or system blockage detected while sidestream pump is on
Ranked causes
- 1. Sampling line blocked (common)
- 2. Filter blocked (common)
- 3. Sampling cell replaced (occasional)
Visually inspect sampling line
- Test point
- Full length of sampling line
- Expected
- No visible blockage or kinks
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display appears off
Monitor screen is blank or not illuminated
Ranked causes
- 1. Mains power switch not on (common)
- 2. Unit not plugged into AC outlet (common)
- 3. Battery pack drained (occasional)
Check power switch
- Test point
- Mains power switch on side panel
- Expected
- Power switch in ON position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG baseline with no waveform
Baseline displayed but no ECG signal visible
Ranked causes
- 1. Gain control not set high enough (common)
- 2. Lead wires and patient cable not fully inserted (common)
- 3. Cable or lead wires damaged (occasional)
Verify gain setting
- Test point
- SIZE key display
- Expected
- Gain set to appropriate level
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ESU interference detected
Poor electrode site preparation or excessive electrosurgical unit (ESU) cautery noise affecting ECG signal
Ranked causes
- 1. Poor electrode site preparation (common)
- 2. Faulty or dirty electrodes (common)
- 3. Excessive high frequency noise (occasional)
Inspect electrode application sites
- Test point
- Electrode placement areas on patient skin
- Expected
- Clean skin with good electrode-to-skin contact
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
False apnea alarm patient safety
Apnea alarm triggers when patient is actually breathing
Ranked causes
- 1. Apnea delay improperly set (common)
- 2. Patient having frequent CVA episodes (occasional)
- 3. Scale size too low (occasional)
Verify apnea delay setting
- Test point
- Monitor alarm settings menu
- Expected
- Apnea delay set appropriately for patient
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas Module air leak detected caution
Pneumatic leak detected in system; may also indicate missing sample line, disconnected module, or prolonged offline operation
Ranked causes
- 1. Pneumatic leak in system (common)
- 2. Gas Module on without sample line attached (occasional)
- 3. Gas Module on extended time without Passport (occasional)
Turn off Gas Module and Passport
- Test point
- Power switches
- Expected
- Both units powered down
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas Module occlusion in sampling line caution
Obstruction detected in sampling line or water trap bottle full
Ranked causes
- 1. Water trap bottle full (common)
- 2. Sampling line blocked (common)
- 3. Filter blocked (common)
Check water trap level
- Test point
- Water trap reservoir
- Expected
- Fluid level below fill mark
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent ECG signal
Signal drops out or disconnects intermittently during monitoring
Ranked causes
- 1. Connections not tight and properly secured (common)
- 2. Electrodes dry (common)
- 3. Cable or lead wires damaged (occasional)
Verify electrode to lead connection
- Test point
- Electrode snap connectors
- Expected
- Connections tight and fully seated
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noisy ECG trace
Excessive noise or interference on ECG waveform
Ranked causes
- 1. Loose or dry electrodes (common)
- 2. Defective electrode wires (occasional)
- 3. Patient cable too close to electrical devices (common)
Inspect electrode condition and contact
- Test point
- All electrode pads
- Expected
- Electrodes fresh, adhesive intact, good skin contact
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm tone or QRS tone disabled
Audio alert not sounding when expected
Ranked causes
- 1. Mute key pressed (common)
- 2. Beep volume low (common)
Check mute symbol
- Test point
- Monitor display screen
- Expected
- No alarm mute symbol visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery Voltage
Battery voltage verification and low battery warning threshold confirmation
Ranked causes
- 1. Battery voltage below nominal or triggering low battery warning (unknown)
- 2. System shutdown voltage threshold reached (unknown)
Measure battery voltage
- Test point
- Battery supply
- Expected
- 12 V DC nominal
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 Sensor
Perform Capnostat CO2 sensor calibration procedure using zero cell and airway adapter when sensor is first connected or monitor requests it
Ranked causes
- 1. First sensor connection to monitor (common)
- 2. Sensor replacement or monitor request (occasional)
Plug Capnostat CO2 sensor into monitor CO2 connector
- Test point
- Passport XG CO2 connector (34)
- Expected
- CO2: Sensor Warming Up message displayed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sidestream pump calibration and flow verification caution
Verify proper flow rate of Sidestream CO2 pump. Tubing configuration or sampling cell adapter changes require recalibration. Pump calibration also calibrates adapter.
Ranked causes
- 1. Tubing configuration or adapter changed (occasional)
- 2. Occlusion detection verification (occasional)
Power on unit and assemble calibration parts
- Test point
- Calibration kit assembly
- Expected
- Parts assembled per Figure 7-13, connected to pump input
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Video Controller Test
Video controller diagnostics - verifies CPU Board LCD output signals driving video controller
Ranked causes
- 1. CPU Board defective - incorrect LCD signals output (occasional)
- 2. Display Assembly faulty - loading LCD signals incorrectly (occasional)
Run Video Controller Test to generate vertical bar pattern
- Test point
- Display output and CPU Board LCD output signals
- Expected
- Vertical bars fill entire screen or half screen depending on software version
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 adapter calibration failure
Adapter calibration did not complete successfully
Ranked causes
- 1. Sensor not fully warmed up (common)
- 2. Sensor not correctly installed on airway adapter (common)
Verify sensor warmup status
- Test point
- Warmup indicator
- Expected
- Warmup cycle complete
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 sensor out of calibration or damaged
CO2: CHECK SENSOR message displayed, sensor out of calibration or defective
Ranked causes
- 1. CO2 sensor requires zero calibration (common)
- 2. CO2 sensor damaged or defective (occasional)
Perform CO2 zero calibration
- Test point
- CO2 menu calibration function
- Expected
- Calibration completes with message clears
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas Module disconnected
Passport cannot detect signals from Gas Module
Ranked causes
- 1. Gas Module not powered on (common)
- 2. Interface cable not properly connected (common)
Verify Gas Module power status
- Test point
- Gas Module power indicator
- Expected
- Power indicator lit or display active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement unable to complete patient safety
NIBP: UNABLE TO MEASURE message displayed after unsuccessful measurement cycle
Ranked causes
- 1. Cuff leaks or poor peripheral pulses (common)
- 2. Excessive patient movement (common)
Inspect cuff and tubing for leaks
- Test point
- Cuff connection to monitor
- Expected
- No air leaks
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No trace for desired parameter
Waveform not displaying for requested parameter
Ranked causes
- 1. Improper transducer attachment to monitor (common)
- 2. Faulty transducer (occasional)
Inspect transducer connection
- Test point
- Transducer cable and monitor input
- Expected
- Cable fully seated and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration waveform too small
Respiration waveform displayed with insufficient amplitude
Ranked causes
- 1. Patient breathing shallow or turned on side (common)
- 2. Scale set inappropriately (common)
Verify patient position and breathing
- Test point
- Patient observation
- Expected
- Patient breathing deeply and upright
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sidestream CO2 setup for intubated patients patient safety
Installation and initialization of Capnostat CO2 sensor, airway adapter, disc filter, capillary tubing, and nafion tubing for monitoring intubated patients with sampling from breathing circuit
Ranked causes
- 1. Initial setup procedure for intubated CO2 monitoring (common)
- 2. Tubing configuration changes requiring pump recalibration (occasional)
Verify sensor warming message clears
- Test point
- Monitor display
- Expected
- 'CO2: Sensor Warming Up' message disappears
2 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 below calibrated range
SpO2 UNCALIBRATED message displayed, detected SpO2 falls below calibrated range
Ranked causes
- 1. Patient desaturation or critical SpO2 level (occasional)
- 2. Sensor malfunction or contamination (occasional)
Assess patient oxygenation status
- Test point
- Patient respiratory status
- Expected
- Adequate oxygenation given clinical context
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 pulse rate above 250 bpm patient safety
SpO2 PR OVER 250 message displayed, detected pulse rate extremely high
Ranked causes
- 1. Excessive patient movement or artifact (common)
- 2. Severe tachycardia (occasional)
Reduce patient movement
- Test point
- Patient extremity with sensor
- Expected
- No visible motion during measurement
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 pulse rate below 30 bpm patient safety
SpO2 PR UNDER 30 message displayed, detected pulse rate extremely low
Ranked causes
- 1. Severe bradycardia or patient condition change (occasional)
- 2. Sensor misalignment or loose placement (occasional)
Assess patient's clinical status
- Test point
- Patient vital signs and condition
- Expected
- HR appropriate for patient condition
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor malfunction or misalignment
SpO2 CHECK SENSOR message displayed, insufficient light reaching detector
Ranked causes
- 1. Emitter and detector not properly aligned (common)
- 2. Sensor malfunction or damage (occasional)
Inspect sensor alignment on patient extremity
- Test point
- Sensor emitter and detector position
- Expected
- Proper alignment across extremity
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 signal interference detected
SpO2 INTERFERENCE message displayed, noise prevents pulse discrimination
Ranked causes
- 1. Patient movement during measurement (common)
- 2. Loose or improperly positioned sensor (common)
Instruct patient to remain still
- Test point
- Patient extremity
- Expected
- No visible movement during measurement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adapter calibration patient safety
Procedure to calibrate airway adapter when replaced or 'Check Adapter' message displays
Ranked causes
- 1. Airway adapter replaced or Check Adapter message (occasional)
Wait for sensor warming
- Test point
- Monitor display
- Expected
- 'CO2: Sensor Warming Up' message disappears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery operation verification caution
Verify battery-backed operation when main power removed. Two batteries required; unit must operate on either battery alone.
Ranked causes
- 1. Power supply or battery verification required (common)
Verify power supply connected and batteries removed
- Test point
- Power Supply Module, battery bay
- Expected
- Power supply connected to appropriate source, D.C. plug connected, no batteries installed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 adapter
CO2 sensor must be calibrated on airway adapter after zero calibration. Sensor should be removed from zero or reference cell and placed on the airway adapter to complete adapter calibration.
Ranked causes
- 1. Adapter calibration required after zero calibration (common)
Ensure sensor is on airway adapter
- Test point
- airway adapter
- Expected
- sensor seated on adapter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 adapter calibration
Compensate for optical differences when switching between adult and neonatal airway adapters for mainstream CO2 operation
Ranked causes
- 1. Adapter type switched (common)
Ensure sensor warm-up complete
- Test point
- Monitor display
- Expected
- 'Sensor Warming Up' message must not be displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 sensor calibration
Calibrate Capnostat CO2 sensor when first connected to monitor or when monitor requests it
Ranked causes
- 1. New sensor connected or monitor requests calibration (common)
Verify monitor power and sensor connection
- Test point
- Passport XG display
- Expected
- Monitor on, Capnostat plugged in
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 sensor calibration verification
Verify CO2 sensor calibration accuracy using reference cell; should be performed at least once weekly
Ranked causes
- 1. Routine maintenance (common)
Verify monitor power status
- Test point
- Passport display
- Expected
- Monitor turned on
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 sensor zero
CO2 sensor must be calibrated on zero cell to establish baseline. ETCO2 reading should be 0 when sensor is positioned on the zero cell located on the sensor cable adjacent to connector.
Ranked causes
- 1. Zero calibration required for sensor baseline (common)
Verify ETCO2 reading on zero cell
- Test point
- zero cell
- Expected
- 0
- Tools
- monitor display
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Date and time verification
Verify date and time displayed in upper left of LCD. If incorrect, use freeze key during self-test to enter configuration menu.
Ranked causes
- 1. Incorrect date or time in unit (unknown)
Observe date and time in upper left LCD corner during startup
- Test point
- LCD display upper left area
- Expected
- Correct current date and time
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG amplitude and gain verification
Verify ECG display scaling accuracy by testing multiple amplitude and size settings with expected pixel measurements
Ranked causes
- 1. ECG gain adjustment or display scaling error (unknown)
Set simulator to 5 Hz 1 mV sine wave, Passport lead 1, size 1 cm/mV
- Test point
- Waveform 1 screen with FREEZE active
- Expected
- 10 mm peak-to-peak +/- 1 mm; overall trace 10 cm +/- 2 mm
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG heart rate detection and alarm patient safety
Verify heart rate calculation accuracy across range and alarm function with HI/LO thresholds and muting capability
Ranked causes
- 1. ECG heart rate detection algorithm error or processing lag (unknown)
Set simulator to ECG QRS, rate 252 bpm
- Test point
- HR display on LCD
- Expected
- 252 +/- 8 bpm, +/- 8 bpm
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG lead detection and messaging
Verify ECG responds to disconnected leads by displaying flat line and appropriate LEAD OFF message for each lead
Ranked causes
- 1. ECG lead detection circuit malfunction (unknown)
Disconnect RA lead from simulator
- Test point
- RA input connector
- Expected
- Flat line display and 'LEAD OFF: RA' message
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG noise and pacer detection
Verify noise suppression and pacer pulse detection with simulator set to SHORT LEADS and Ventricular Pacer mode
Ranked causes
- 1. ECG input filtering or pacer detection circuit failure (unknown)
Set simulator to SHORT LEADS
- Test point
- ECG display
- Expected
- Noise does not exceed one pixel resolution
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG signal display and sweep timing
Verify ECG trace sweeps across waveform screens in correct timing with complete cycles displayed at specified speeds
Ranked causes
- 1. ECG processing or display timing error (unknown)
Observe ECG trace sweep timing on Waveform 1
- Test point
- Waveform 1 screen
- Expected
- Trace completes sweep in 2 seconds with 2 complete ECG cycles
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG trend data recording
Verify high heart rate value recorded in trend data and highlighted in reverse video after alarm violation
Ranked causes
- 1. Trend data storage or display error (unknown)
Press TREND/RETURN key after HR alarm violation
- Test point
- Trend data display
- Expected
- High HR value (125 bpm) shown in reverse video
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: External Communication
External communication channel loop test - data transmission and echo verification
Ranked causes
- 1. Character echo error on external communication channel (unknown)
Verify character echo on external communication
- Test point
- External communication channel
- Expected
- Every character echoed back correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: External Voltage high voltage
External DC power supply voltage verification during operation
Ranked causes
- 1. External voltage out of specification (unknown)
Measure external power supply voltage
- Test point
- External DC power input
- Expected
- 15.6 - 17.6 Volts, 15.6 to 17.6 V
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas Module patient safety
Perform gas module zero and span calibration to verify and adjust accuracy of anesthetic gas, O2, N2O and CO2 measurements
Ranked causes
- 1. Initial setup or periodic accuracy verification (common)
Warm up gas module to full temperature
- Test point
- Gas module
- Expected
- Module fully warmed (30 minutes recommended)
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: IBP1 and IBP2 zeroing and accuracy
Verify IBP pressure channel zeroing function and measurement accuracy across pressure range with dual channels at different values
Ranked causes
- 1. IBP transducer not zeroed or calibration error (common)
Set simulator to 0 mmHg for both IBP1 and IBP2
- Test point
- Pressure source
- Expected
- Both channels display dashed lines before zeroing
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LCD display contrast
Verify front panel contrast control adjusts display brightness from completely white to completely blue
Ranked causes
- 1. Contrast control out of range or LCD failure (rare)
Set contrast control fully up
- Test point
- Front panel contrast control
- Expected
- LCD display completely white
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Mainstream CO2 adapter calibration caution
Perform adapter calibration when switching between adult and neonatal airway adapters or after Check Adapter message. Must occur when sensor not warming.
Ranked causes
- 1. CO2 adapter optical mismatch between adult and neonatal (occasional)
Verify Sensor Warming Up message NOT displayed
- Test point
- CO2 waveform window
- Expected
- No 'Sensor Warming Up' message visible
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Mainstream CO2 sensor verification
Reference cell calibration verification to ensure sensor accuracy. Passport turned on, Capnostat connected and warmed-up.
Ranked causes
- 1. Sensor requires periodic verification (common)
Verify Passport powered on and Capnostat connected
- Test point
- Capnostat connection and warm-up indicator
- Expected
- Capnostat warmed-up and connected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Mainstream CO2 sensor zero calibration
Full sensor calibration required first time sensor connects to monitor or when monitor requests it. Sensor must be warmed up before starting.
Ranked causes
- 1. New sensor or sensor requires recalibration (common)
Verify Passport powered on and sensor warmed
- Test point
- Capnostat warm-up indicator
- Expected
- Sensor Warming Up message not displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: NIBP Communication
- Calibration: NIBP measurement cycle and cuff inflation/deflation
- Calibration: Sidestream adapter calibration
- Calibration: Sidestream CO2 sensor verification
- Calibration: Sidestream CO2 sensor zero calibration
- Calibration: SpO2 Accuracy Test
- Calibration: SpO2 sensor detection and pleth waveform
- Calibration: SpO2 shared RAM memory
- Calibration: Temperature sensor accuracy with 700 and 400 series probes
- CO2 breaths detected during pump calibration
- CO2 calibration failed - adapter not inserted
- CO2 calibration failed - zero cell not inserted
- CO2 sensor not connected
- CO2 sensor positioned on reference cell
- CO2 sensor positioned on zero cell
- CO2 sensor reference cell check
- CO2 sensor warming up
- CO2 sidestream on neonate not allowed
- CO2 sidestream pump calibration failure
- CO2 sidestream pump failure
- CO2 subsystem failure
- CO2 subsystem initialization in progress
- CO2 zero calibration failure
- CO2 zero calibration in progress
- Cosmetic checks and hardware inspection
- CVA message displayed
- DAC waveform irregular
- ECG trace not moving
- Excessive electrosurgical interference
- Gas Module agent mismatch - Desflurane
- Gas Module agent mismatch - Enflurane
- Gas Module agent mismatch - Halothane
- Gas Module agent mismatch - Isoflurane
- Gas Module agent mismatch - Sevoflurane
- Gas Module anesthetic agent sensor uncalibrated
- Gas Module anesthetic agent zero error
- Gas Module calibration error - sampling error all channels
- Gas Module calibration error - zeroing error
- Gas Module calibration not started
- Gas Module cannot zero - retrying
- Gas Module CO2 sensor uncalibrated
- Gas Module CO2 zero error
- Gas Module end tidal anesthetic agent high
- Gas Module end tidal anesthetic agent low
- Gas Module end tidal CO2 high
- Gas Module end tidal CO2 low
- Gas Module end tidal O2 high
- Gas Module end tidal O2 low
- Gas Module exhaust blocked
- Gas Module failed
- Gas Module inspired anesthetic agent high
- Gas Module inspired anesthetic agent low
- Gas Module inspired CO2 high
- Gas Module inspired N2O high
- Gas Module inspired O2 high
- Gas Module inspired O2 low
- Gas Module mixed anesthetic agents detected
- Gas Module N2O sensor uncalibrated
- Gas Module N2O zero error
- Gas Module O2 sensor uncalibrated
- Gas Module O2 zero error
- Gas Module pump off due to pneumatic error
- Gas Module respiration rate high
- Gas Module respiration rate low
- Gas Module sampling error during calibration
- Gas Module unknown anesthetic agent detected
- Gas Module warm-up for measurement
- Gas Module water trap needs replacement
- Lead off of patient
- Leakage current test: Lead to ground sink current with line voltage
- Leakage current test: Source current chassis to ground
- Low amplitude ECG
- NIBP cuff overpressure detected
- NIBP measurement fails with motion detected
- NIBP system failure detected
- NIBP system initialization delayed
- NIBP unable to detect pulse at highest cuff pressure
- Pump calibration
- Recorder door open
- Recorder paper depleted
- Recorder report appears blank
- Respiration waveform too large
- Sidestream CO2 setup for non-intubated patients
- SpO2 excessive ambient light interference
- SpO2 low perfusion condition
- SpO2 no pulse detected
- SpO2 pulse search in progress
- SpO2 sensor not connected to patient
- SpO2 sensor not detected
- SpO2 system failure with ROM, RAM or filter error
- SpO2 system self-test in progress
- SpO2 unknown or unrecognized sensor detected
- SpO2 weak pulse detected
- Temperature probes not working
- Wandering ECG baseline
95 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ARTIFACT | Line frequency noise detected on ECG waveform | 50 or 60 Hz noise detected on ECG | warning |
CO2: ADAPTER CAL COMPLETE | Adapter calibration of CO2 sensor completed successfully | Adapter calibration finished | info |
CO2: ADAPTER CAL FAILURE | Adapter calibration of CO2 sensor failed | Adapter calibration did not complete successfully | warning |
CO2: ADAPTER CAL IN PROGRESS | Adapter calibration of CO2 sensor is in progress | CO2 adapter calibration requested | info |
CO2: CAL FAILED - INSERT ADAPTER | Adapter calibration attempted while sensor on zero or reference cell | Adapter calibration attempted while sensor attached to zero or reference cell | error |
CO2: CAL FAILED - INSERT ZERO CELL | Zero calibration attempted without sensor attached to zero cell | Zero calibration attempted while sensor not attached to zero cell | error |
CO2: FAILURE | CO2 subsystem has failed | CO2 subsystem malfunction detected | critical |
CO2: ZERO CAL COMPLETE | Zero calibration of CO2 sensor completed successfully | Zero calibration finished | info |
CO2: ZERO CAL FAILURE | Zero calibration of CO2 sensor failed | Zero calibration did not complete successfully | warning |
CO2: ZERO CAL IN PROGRESS | Zero calibration of CO2 sensor is in progress | Sensor placed on zero cell and zero calibration requested | info |
DIAGNOSTICS IN PROGRESS | Self test is being performed | Monitor self-test initiated | info |
ESU INTERFERENCE | Excessive electrosurgical unit noise detected on ECG | Poor electrode preparation or excessive cautery noise | warning |
INTERFERENCE | Excessive electrical noise detected on ECG | Poor electrode site preparation, improper lead selection, or excessive line frequency noise | warning |
LEAD FAULT | Lead off of patient detected | ECG lead disconnected from patient | warning |
SELF TEST COMPLETE | Diagnostics OK | Self-test completed successfully | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.