Mindray Accutorr Plus: troubleshooting and error codes
The ServiceForge knowledge base holds 52 documented symptoms for the Mindray Accutorr Plus, with 69 ranked causes, 115 diagnostic checks with test points and expected values, and 30 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No measurement or inaccurate temperature readings patient safety
AccuTemp fails to display temperature reading or reading is inconsistent with expected value for patient condition.
Ranked causes
- 1. Technique error - button not held throughout measurement (common)
- 2. Technique error - not scanning side-to-side (common)
- 3. Technique error - twisting or rotating probe (common)
- 4. Technique error - optical film cover not changed (common)
- 5. Debris or contamination on lens (occasional)
- 6. Ear canal opening not fully covered by probe (occasional)
- 7. Visible wax plug or dense hair blocking view (occasional)
- 8. Low battery voltage (occasional)
Verify button is pressed throughout measurement
- Test point
- Start button on AccuTemp
- Expected
- Button held continuously from initiation until temperature displays
6 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AccuTemp no measurement or inaccurate reading
AccuTemp fails to display measurement or readings are inaccurate; may be caused by technique error or equipment malfunction.
Ranked causes
- 1. Technique error - improper button or probe handling (common)
- 2. Debris or obstruction on lens or in ear canal (common)
- 3. Low battery voltage (occasional)
Verify proper measurement technique
- Test point
- Operator procedure
- Expected
- Button held throughout measurement; probe scanned side-to-side; new optical film cover used
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery operation abnormal
Battery operation LED is not steady on or is flashing, indicating near depletion or no battery operation.
Ranked causes
- 1. Battery depleted or disconnected (common)
- 2. Power supply/charger malfunction (occasional)
Check battery operation LED status
- Test point
- Battery operation LED
- Expected
- Steady ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Low battery indicator voltage high voltage
Determine threshold voltage where battery LED begins to flash during voltage decrease
Ranked causes
- 1. Test low battery indicator threshold (unknown)
- 2. Decrease voltage and observe indicator (unknown)
Monitor ON and initialize
- Test point
- Monitor display
- Expected
- Self-check completes successfully
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff over-inflation patient safety
NIBP cuff inflates beyond safe pressure limit for patient size
Ranked causes
- 1. Inadequate over-pressure transducer control (occasional)
- 2. Pump speed not reduced for neonatal cuff (occasional)
Monitor cuff pressure during inflation
- Test point
- DC pressure channel (A/D channel 1)
- Expected
- Reaches target pressure without exceeding over-pressure limit for selected patient size
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multiple temperature readings inconsistent
Taking repetitive measurements in the same location results in lower readings; temperature varies between measurements.
Ranked causes
- 1. Probe cooling effect from repeated measurements (common)
- 2. Patient lying on measured ear (occasional)
Verify wait time between successive measurements
- Test point
- Time elapsed between measurements
- Expected
- At least a few minutes between measurements in same location
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulsatile signal not detected patient safety
NIBP system fails to acquire pulse channel signal during cuff inflation
Ranked causes
- 1. Pulse channel filter malfunction or misconfiguration (rare)
- 2. Weak pulse amplitude for patient size (occasional)
Measure AC pressure (pulse) channel
- Test point
- A/D channel 2 (AC pressure/pulse)
- Expected
- Detectable oscillations during cuff deflation above systolic pressure Hz
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature readings not displayed patient safety
Optional temperature module readings not appearing in temperature display window.
Ranked causes
- 1. Temperature probe failure (common)
- 2. Temperature module failure (occasional)
Obtain one or more temperature readings
- Test point
- Temperature display window
- Expected
- Readings displayed
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Various temperature readings between opposite ears
Temperature measurements differ significantly between left and right ears when taken sequentially.
Ranked causes
- 1. Patient lying on one ear (common)
- 2. Ear covered (hat, blanket, bandage, telephone) (occasional)
Verify patient position and ear coverage
- Test point
- Both ears visually
- Expected
- Both ears uncovered, patient not lying on either ear
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Various temperature readings from same location
Quick repetitive measurements in same ear or axilla yield inconsistent results, temperature decreases with repeated measurements.
Ranked causes
- 1. Probe thermally cools measured area (common)
- 2. Dirty or wet lens (occasional)
Verify measurement interval in same location
- Test point
- Same ear or axilla
- Expected
- Temperature stable after several minutes rest
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AccuTemp critical low battery error
AccuTemp double beeps and displays '——' while holding Start button, then powers down; battery voltage below 5.7 VDC.
Ranked causes
- 1. Battery voltage critically low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AccuTemp displays E-00 or E-01 error code
AccuTemp shows error code E-00 or E-01 on display while holding Start button.
Ranked causes
- 1. Internal AccuTemp fault (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AccuTemp double beeps and displays "--" then powers down
Device sounds double beep, displays two dashes on screen, and automatically shuts off during measurement attempt.
Ranked causes
- 1. Battery voltage critically low (common)
Measure battery voltage
- Test point
- 9V battery terminals
- Expected
- ≥5.7 VDC, ≥5.7
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AccuTemp double beeps while holding Start button
Device sounds double beep during measurement attempt, but continues operation and displays reading.
Ranked causes
- 1. Low battery condition (common)
Measure battery voltage
- Test point
- 9V battery terminals
- Expected
- 6.2 to 5.7 VDC, 6.2-5.7
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AccuTemp error E-O0 or E-o1
AccuTemp displays 'E-O0' or 'E-o1' while holding Start button; internal malfunction.
Ranked causes
- 1. Internal AccuTemp malfunction (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AccuTemp low battery condition
AccuTemp double beeps while holding Start button; battery voltage between 6.2 and 5.7 VDC.
Ranked causes
- 1. Battery voltage low but not critically depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery operation LED not on or flashing
Battery operation LED is not steadily illuminated or is flashing, indicating near depleted or failed battery
Ranked causes
- 1. Battery depleted or power supply/charger failure (common)
Check battery operation LED status
- Test point
- Battery operation LED on front panel
- Expected
- Steady ON indicates battery operation; flashing indicates near depleted battery
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery test setup high voltage
Verify normal power-up sequence and battery indicator function using external power supply
Ranked causes
- 1. Initial setup for battery testing (unknown)
Power up UUT with external supply
- Test point
- Keypad battery indicator
- Expected
- Normal power-up sequence occurs and battery symbol indicator is lit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Infrared thermometer temperature verification
Verify infrared thermometer temperature accuracy using preheated water bath, Black Body device, and laboratory-grade reference thermometer
Ranked causes
- 1. Periodic verification of infrared thermometer accuracy required (unknown)
Pre-heat water bath to target temperature
- Test point
- Water bath
- Expected
- 100°F
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Low battery cut-off voltage high voltage
Determine voltage threshold at which monitor powers off due to low battery
Ranked causes
- 1. Test low battery cut-off threshold (unknown)
Monitor power-off point for SLA battery
- Test point
- Monitor power state
- Expected
- 9.8 to 10.7 VDC, to 10.7 VDC
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Main pressure transducer
Main pressure transducer reading disagrees with external reference gauge; MAP display does not match simulator manometer reading at 158 mmHg
Ranked causes
- 1. Main pressure transducer out of calibration (unknown)
Connect Cufflink simulator and 700 cc test chamber
- Test point
- Accutorr Plus cuff connector, simulator open port
- Expected
- Equipment connected and manometer function selected
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Main Pressure Transducer Verification Test
Verify main pressure transducer calibration in LOW, MED, and HIGH modes by comparing readings to external gauge
Ranked causes
- 1. Main pressure transducer out of calibration (unknown)
Select LOW mode calibration test
- Test point
- Room 11, Bed a
- Expected
- Pressure settles below 100 mmHg after 10 seconds
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Over-pressure transducer
Over-pressure transducer reading does not display 158 mmHg during Low pressure mode verification test
Ranked causes
- 1. Over-pressure transducer out of calibration (unknown)
Adjust potentiometer to set MAP readout
- Test point
- R4 on 0670-00-0584-01 PCB or R504 on 0670-00-1143 PCB
- Expected
- 158 mmHg
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Predictive Thermometer
Verify and adjust predictive thermometer (PTM) to factory specifications using water bath method or system calibration procedure. For infrared thermometer, perform temperature verification test with black body device.
Ranked causes
- 1. Thermometer readings out of specification (unknown)
Prepare water bath at specified temperature
- Test point
- Water bath thermometer
- Expected
- 89.6 to 110.0°F
- Tools
- Laboratory grade stirred water bath, precision thermometer
16 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Predictive thermometer system
Adjust output voltage and temperature readings to factory specifications using resistor decade box, DVM, and adjustable potentiometers R1 and R2
Ranked causes
- 1. Initial calibration or recalibration required (unknown)
Connect power source
- Test point
- Thermometer input
- Expected
- 9.1 V DC, ±0.2V nominal
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Predictive thermometer system calibration
Two-point calibration procedure for temperature measurement module using resistor decade box and voltage adjustment
Ranked causes
- 1. Temperature module requires periodic two-point calibration at reference resistance points (common)
Connect power supply to predictive thermometer
- Test point
- Thermometer power input
- Expected
- 9V battery (if module has one) or external 9.1V supply connected
- Tools
- 9V battery or adjustable external supply
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Predictive thermometer verification
Infrared thermometer probe verification against water bath black body device with ambient and target temperature recording
Ranked causes
- 1. Thermometer requires periodic verification against known temperature reference (common)
Pre-heat water bath to target temperature
- Test point
- Water bath
- Expected
- 100°F
- Tools
- Heat source, thermometer
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Set current time
Adjust real-time clock to correct date and time using TIMER/TEMP and arrow keys
Ranked causes
- 1. RTC reset during battery test (common)
Hour display active
- Test point
- Monitor display
- Expected
- Hour digit only displayed after TIMER/TEMP held 6 seconds
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Water Bath Method correction factors
Apply correction factors to thermometer readings based on temperature range to obtain calibrated measurements. Use Table 5-10 values.
Ranked causes
- 1. Reading correction required per specification (common)
Identify Celsius readout range and apply correction
- Test point
- Thermometer display (Celsius)
- Expected
- 32.0–41.0°C: subtract 0.6; 41.0–43.0°C: subtract 0.5
- Tools
- Calculator, Table 5-10
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disconnected cuff not detected patient safety
NIBP system does not detect disconnected or kinked cuff hoses during measurement cycle
Ranked causes
- 1. Cuff disconnection or kink in hose (common)
Verify cuff connection
- Test point
- Cuff connector
- Expected
- Firmly seated, no kinks in hose
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green LED not illuminated when AC power available
AC power is connected but green LED adjacent to front panel ON/OFF keys does not illuminate
Ranked causes
- 1. Power supply malfunction (common)
Verify AC power is available and green LED status
- Test point
- Green LED adjacent to front panel ON/OFF keys
- Expected
- Green LED is illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate or inconsistent infrared temperature readings patient safety
Infrared AccuTemp thermometer showing unreliable or variable measurements in same location.
Ranked causes
- 1. Probe cooler than measurement site, heat transfer occurs (common)
Wait 3-5 minutes between consecutive measurements in same ear
- Test point
- Measurement site
- Expected
- Consistent readings after recovery period
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect battery charge profile selected
Installed battery charge profile does not match battery type loaded in unit
Ranked causes
- 1. Operator configuration error (common)
Verify battery charge profile matches installed battery
- Test point
- Battery type and charge profile setting
- Expected
- Proper battery charge profile selected for installed battery
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect patient size or cuff mismatch patient safety
Wrong patient size selected (Adult, Neonate, Pediatric) or NIBP cuff does not match selected size
Ranked causes
- 1. Operator configuration error (common)
Verify patient size selection and cuff match
- Test point
- Front panel patient size setting and installed NIBP cuff
- Expected
- Correct patient size selected (Adult, Neonate, or Pediatric) with matching NIBP cuff
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP and/or SpO2 detection and display not working patient safety
NIBP measurements or SpO2 readings not displaying or functioning correctly.
Ranked causes
- 1. Failed Diagnostics Test 01-12 (unknown)
Obtain minimum five NIBP measurements
- Test point
- NIBP monitor
- Expected
- Valid measurements displayed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP hardware overpressure error 987 caution
Error code 987 displayed; excessive cuff pressure detected by hardware over-pressure sensor; triple beep tone generated.
Ranked causes
- 1. Hardware overpressure sensor activation (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP measurement cuff overpressure error 8812 caution
Error code 8812 displayed on front panel; excessive cuff pressure detected; unit generates triple beep tone.
Ranked causes
- 1. Overpressure circuit not programmed for current patient size (common)
Verify patient size selection matches overpressure circuit setting
- Test point
- System configuration
- Expected
- Overpressure circuit programmed for selected patient size
NIBP measurement retry error 8810
Error code 8810 displayed on front panel LED; unit generates triple beep tone and cannot complete NIBP measurement.
Ranked causes
- 1. Motion artifact or weak pulsations (common)
Verify patient size selection and cuff size match
- Test point
- System settings and attached cuff
- Expected
- Correct patient size selected; appropriate cuff attached
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Service Diagnostics: Software version verification
Technician enters Service Diagnostics mode to verify installed software versions; initial power-up sequence displays Room LED '0', Bed LED 'a', Systolic LED 'ddd'
Ranked causes
- 1. Need to verify software configuration (common)
Verify initial power-up display values
- Test point
- LED readouts
- Expected
- Room LED='0', Bed LED='a', Systolic LED='ddd'
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 failure error 996
Error code 996 displayed on front panel; SpO2 module failed self-test.
Ranked causes
- 1. SpO2 module internal failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 no sensor error 8850
Error code 8850 displayed on front panel; no SpO2 sensor is connected to the unit.
Ranked causes
- 1. SpO2 sensor not connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor not connected or incorrect sensor type/size patient safety
SpO2 section equipped but no sensor connected or sensor type/size does not match unit requirements
Ranked causes
- 1. Missing or mismatched SpO2 sensor (common)
Verify SpO2 sensor connection and type/size if equipped
- Test point
- SpO2 sensor connector and sensor specifications
- Expected
- SpO2 sensor connected with proper type/size for unit
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor off patient error 8851 patient safety
Error code 8851 displayed on front panel; sensor is not on patient (Masimo® SpO2 only).
Ranked causes
- 1. Sensor disconnected or removed from patient (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 uncalibrated error 995
Error code 995 displayed on front panel; SpO2 module fails calibration check.
Ranked causes
- 1. SpO2 calibration check failed (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature module battery low error 8832
Error code 8832 displayed on front panel; 9V battery in temperature module requires replacement.
Ranked causes
- 1. 9V battery depleted in temperature module (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature module calibration required error 988
Error code 988 displayed on front panel; thermometer needs calibration.
Ranked causes
- 1. Temperature module out of calibration (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature module internal failure error 991
Error code 991 displayed on front panel; thermometer module internal failure detected.
Ranked causes
- 1. Temperature module internal component failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature module switch error 990
Error code 990 displayed on front panel; thermometer switch is set incorrectly.
Ranked causes
- 1. Thermometer mode switch set to incorrect position (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature probe defective error 8831
Error code 8831 displayed on front panel; probe is defective or connection is faulty.
Ranked causes
- 1. Defective temperature probe or faulty connection (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature probe error 8830
Error code 8830 displayed on front panel; tissue contact may have been lost during temperature measurement.
Ranked causes
- 1. Probe lost contact with patient tissue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Monitor will not turn on or respond to power button.
Ranked causes
- 1. Power supply output voltages out of specification (common)
Measure 5V power supply output
- Test point
- Main power supply output
- Expected
- 4.9 - 5.1 V, 4.9 - 5.1
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not power on
Green LED adjacent to front panel ON/OFF keys is not illuminated when AC power is available.
Ranked causes
- 1. Power supply malfunction (common)
Check green LED adjacent to front panel ON/OFF keys
- Test point
- Front panel LED
- Expected
- Illuminated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
8810 | NIBP Retry - Unable to Measure | Motion artifact, cycle time-out, weak pulsations, or no pulsations detected | warning |
8811 | NIBP Retry - Pump Higher | Insufficient cuff pressure during measurement attempt | warning |
8812 | NIBP Stop - Cuff Overpressure | Software detects excessive cuff pressure during measurement | error |
8813 | NIBP Stop - Unable to Measure | Four successive NIBP measurement attempts fail | error |
8830 | Temperature Probe Check Required | Temperature probe detected loss of tissue contact during measurement | warning |
8830 | Check Probe (TEMP) | Temperature probe connection or contact lost | warning |
8831 | Temperature Probe Replace | Defective probe detected or connection failure | error |
8832 | Temperature Module Battery Low | Battery voltage in temperature module falls below operating threshold | warning |
8832 | Battery Low (TEMP) | 9V temperature probe battery depleted | warning |
8850 | SpO2 No Sensor | SpO2 module unable to detect connected sensor at startup | warning |
8851 | SpO2 Sensor Off Patient (Masimo only) | SpO2 sensor loses contact with patient or is removed during measurement | warning |
8851 | Sensor Off | SpO2 sensor not on patient | warning |
8852 | SpO2 Interference | Signal interference detected on SpO2 sensor line (Masimo only) | warning |
8853 | SpO2 Pulse Search | SpO2 module cannot locate acceptable signal | warning |
8854 | SpO2 Weak Pulse | SpO2 sensor detects weak arterial pulsations (Masimo only) | warning |
8856 | SpO2 Check Sensor | SpO2 sensor malfunction detected during operation (Masimo only) | error |
8857 | SpO2 Pulse Rate Below Threshold | Measured pulse rate falls below minimum threshold | info |
8857 | PR<21 or PR<26 | Pulse rate below threshold | warning |
8858 | SpO2 Pulse Rate Above Threshold | Measured pulse rate exceeds maximum threshold | info |
8858 | PR>249 or PR>239 | Pulse rate above threshold | warning |
984 | NIBP Hardware Failure | NIBP A/D conversion circuit malfunction detected | critical |
985 | NIBP Overpressure Circuit Not Programmed | Overpressure circuit not configured for current patient size | critical |
986 | NIBP Overpressure Circuit Not Tracking | Two pressure transducers not tracking each other during measurement | critical |
987 | Stop - Hardware Overpressure | Excessive cuff pressure detected by hardware over-pressure sensor | critical |
988 | Temperature Module Bad Calibration | Temperature module calibration check fails at startup or during operation | error |
990 | Temperature Module Illegal Mode | Temperature module mode switch set to incorrect position | error |
991 | Temperature Module Failed | Internal temperature module malfunction detected | critical |
995 | SpO2 Uncalibrated | SpO2 module fails calibration check at startup | error |
996 | SpO2 Failure | SpO2 module fails self-test at startup or during operation | critical |
E-00 or E-01 | Internal AccuTemp error code | Displayed when Start button is held during measurement | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.