Mindray A7: troubleshooting and error codes
The ServiceForge knowledge base holds 257 documented symptoms for the Mindray A7, with 365 ranked causes, 467 diagnostic checks with test points and expected values, and 211 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Breathing system leak patient safety
Audible hiss or visible gas escape from breathing system connections, or inability to maintain pressure during ventilation.
Ranked causes
- 1. CO2 absorber canister not installed properly (common)
- 2. CO2 absorber canister sealing component damaged (occasional)
- 3. Bag arm seal damaged (occasional)
- 4. Water collection cup loose or seal damaged (common)
- 5. Circuit adapter assembly seal damaged (occasional)
- 6. Bellows housing or bellows not installed properly (common)
- 7. Bellows sealing cushion damaged or fallen off (occasional)
- 8. Breathing valve cover not installed properly or seal damaged (common)
- 9. Patient breathing tube damaged (occasional)
- 10. Bellows damaged (rare)
- 11. Other breathing system sealing connection damaged (rare)
- 12. Condensate valve not installed properly or internal seal damaged (occasional)
Visually inspect all breathing system connections
- Test point
- all breathing circuit connections
- Expected
- All connections seated securely, no visible gaps or hissing
1 more check with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate tidal volume patient safety
Measured tidal volume does not match set value; may include inconsistency between TVe and TVi readings.
Ranked causes
- 1. Flow sensor not installed properly (common)
- 2. Inappropriate fresh gas flow setting (common)
- 3. Breathing system leaks with insufficient fresh gas flow (occasional)
- 4. Water buildup inside flow sensor (occasional)
- 5. Flow sensor membrane distorted, dirty or zero drift in pressure sensor (occasional)
- 6. Flow sensor pressure sampling pipeline leak (occasional)
- 7. Flow sensor damaged (rare)
- 8. Pressure sensor on ventilator control board defective (rare)
- 9. Inlet gas flow regulator of expiratory valve assembly defective (rare)
- 10. Plimit set too low causing early expiration (occasional)
- 11. TVe and TVi measurements differ; sensor calibration error (occasional)
Verify flow sensor installation
- Test point
- Flow sensor seating and alignment
- Expected
- Flow sensor properly seated and aligned
4 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accessories and breathing circuit component leaks caution
Leakage detected at Paw gauge, water collection cup, O2 sensor, check valve dome, bag arm, or pre-pak assembly connection points.
Ranked causes
- 1. Damaged manual bag (occasional)
- 2. Damaged breathing tube (occasional)
- 3. Damaged Paw pressure gauge seal (occasional)
- 4. Damaged water collection cup seal (occasional)
- 5. Damaged O2 sensor seal or plug seal (occasional)
- 6. Damaged check valve dome seal (occasional)
- 7. Damaged bag arm seal (occasional)
- 8. Damaged pre-pak assembly seal (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic circuit leak test fails patient safety
Safety valve control test fails or automatic circuit leak test reports excessive leak (≥200 mL/min and ≤1000 mL or >1000 mL)
Ranked causes
- 1. Bellows not fully inflated before test (common)
- 2. Y piece not connected to test plug (common)
- 3. Bellows housing not properly installed (common)
- 4. Leak in bellows, breathing tubes, or CO2 absorber canister (common)
- 5. Drive gas pressure low (occasional)
- 6. Safety valve malfunction (occasional)
Verify bellows fully inflated
- Test point
- Bellows housing
- Expected
- Bellows fully expanded
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow waveform displayed irregularly patient safety
Flow waveform on display shows abnormal shape, dropouts, or erratic behavior instead of smooth sinusoidal or expected pattern.
Ranked causes
- 1. Flow sensor assembly not installed properly (common)
- 2. Water accumulated inside flow sensor assembly (occasional)
- 3. Flow sensor membrane distorted, dirty, or pressure sensor zero drift (occasional)
- 4. Flow sensor damaged (rare)
- 5. Pressure sensor on fresh flow sensor board defective (rare)
- 6. Flow sensor pressure sampling pipeline leaks (occasional)
Verify flow sensor mechanical installation
- Test point
- flow sensor assembly mounting
- Expected
- Flow sensor seated firmly with no movement or gaps
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic system leak in O2 supply or related assemblies caution
Test pressure reading falls dramatically and continuously after gas supply is turned off, indicating leak in O2 inlet, expiratory valve, flush button, system switch, or O2 flow regulator.
Ranked causes
- 1. Leak in O2 supply inlet assembly (occasional)
- 2. Leak in expiratory valve assembly (occasional)
- 3. Leak in O2 flush button assembly (occasional)
- 4. Leak in system switch assembly (occasional)
- 5. Leak in O2 flow regulator (occasional)
Monitor test pressure gauge after turning off pipeline gas supply
- Test point
- System pressure via O2 inlet test setup
- Expected
- Pressure does not fall dramatically and continuously; stable reading indicates no leak
- Tools
- 1 MPa test pressure gauge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Machine cannot be started after power is applied
System fails to boot or initialize despite power being present
Ranked causes
- 1. System switch is damaged (rare)
- 2. Cable connected to system switch is disconnected (occasional)
- 3. Power board hardware circuit failure with no voltage output (occasional)
- 4. Power board software code error (occasional)
Verify system switch connection
- Test point
- System switch cable
- Expected
- Connected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen is not lit
Display remains dark, no backlight visible
Ranked causes
- 1. Inverter cable disconnected (occasional)
- 2. Screen backlight board damaged (occasional)
- 3. Power board hardware failure causing improper output (rare)
- 4. Power board software failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilation sound heard then Calibration Failure displayed patient safety
Machine ventilates initially but calibration fails immediately after Start selected
Ranked causes
- 1. Sampling line disconnected or reversed on flow sensor (common)
- 2. Maximum inspiratory valve flow less than 90 L/min (occasional)
- 3. Pneumatic circuit connection error between machine and calibration device (occasional)
- 4. Communication connection error or settings error on calibration device (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AG module insertion and alarm test patient safety
Verify AG module detection, watertrap detection, and airway occlusion alarm function after module insertion.
Ranked causes
- 1. AG module not properly detected (occasional)
- 2. Watertrap not detected (occasional)
- 3. Airway occlusion alarm malfunction (rare)
Remove and reinstall AG module
- Test point
- Module rack
- Expected
- Module inserted fully and locked
6 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
APL valve, median plate support tube or Auto/Manual switch seal damage caution
Leakage from APL valve, support tube on median plate, or Auto/Manual ventilation switch sealing components.
Ranked causes
- 1. Damaged APL valve seals (occasional)
- 2. Damaged median plate support tube seals (occasional)
- 3. Damaged Auto/Manual ventilation switch seals (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AG Module (Anesthetic Gas) patient safety
Calibration of internal or external AG module using standard calibration gas; requires specific gas concentrations and setup.
Ranked causes
- 1. Machine in BFCS state instead of EFCS (occasional)
- 2. N2O supply pressure low (occasional)
- 3. Calibration points not conforming to monotonicity (occasional)
Prepare calibration gas system
- Test point
- Gas cylinder, T-connector, tubing
- Expected
- Gas concentration: AA≥1.5%, CO2≥1.5%, N2O≥40%, O2≥40%, a/c≤0.01
9 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electronic Flow Control System (EFCS) zeroing failure
Troubleshooting steps when EFCS zeroing fails; addresses mechanical and electronic faults.
Ranked causes
- 1. Trapped gas in machine or flowmeter imbalance (common)
- 2. Mechanical failure in three-way valve or proportional valve fault (occasional)
- 3. Sensirion flow sensor or EFCS board fault (occasional)
Disconnect gas supplies and bleed machine
- Test point
- Gas inlet, flowmeter
- Expected
- Pressure gauge pointer returns to zero
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater is ineffective caution
Heater does not warm as expected or produces insufficient heat
Ranked causes
- 1. Heater driver and control circuit in power board damaged (occasional)
- 2. Heater is damaged (occasional)
- 3. Heater internal sensor is ineffective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network connection fails
Machine cannot establish or maintain network connectivity
Ranked causes
- 1. Network connection board cables are loose (common)
- 2. Network cable is too long (occasional)
- 3. Network cable is used incorrectly (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ventilation sound after Begin; Calibration Failure displayed patient safety
Machine fails to ventilate and error message appears immediately after Begin is selected
Ranked causes
- 1. Auto/Manual ventilation switch in bag position (common)
- 2. Drive gas (O2) pressure below 200 kPa (common)
- 3. Zero point error on inspiratory/expiratory flow sensor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pinsp Not Achieved patient safety
In pressure mode, inspiratory pressure less than target for 6 consecutive cycles
Ranked causes
- 1. Breathing circuit leakage (common)
- 2. Pressure sensor measurement accuracy error (occasional)
- 3. Ventilator control board failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure sensor sampling line connected incorrectly or not connected patient safety
During normal ventilation, Paw waveform decreases with increasing actual pressure. Alarms 'Paw Too Low' or 'Patient Circuit Leak' occur. A/D value of pressure sensor decreases or remains unchanged with increasing PEEP D/A value.
Ranked causes
- 1. Airway pressure sensor sampling line connected incorrectly (common)
- 2. Airway pressure sensor sampling line not connected (occasional)
- 3. Airway pressure gauge defective (rare)
Observe Paw waveform during normal ventilation with increasing actual pressure
- Test point
- Patient airway pressure waveform display
- Expected
- Paw waveform should increase with increasing actual pressure
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch panel is unresponsive
Touch screen does not respond to user input or touch commands
Ranked causes
- 1. Touch panel is damaged (occasional)
- 2. Touch panel controller is damaged (rare)
- 3. Touch panel cable is disconnected (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Troubleshooting: Total Flow Sensor calibration failure
Common failures during total flow sensor calibration and recommended corrective actions.
Ranked causes
- 1. Machine in BFCS state instead of EFCS (common)
- 2. N2O supply pressure low (common)
- 3. Calibration points not conforming to monotonicity (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on and screen is not lit high voltage
AC and battery indicator lamps are not lit, machine cannot ventilate, screen is dark
Ranked causes
- 1. AC power supply not connected and battery capacity insufficient (common)
- 2. AC input breaker tripped and battery capacity insufficient (common)
- 3. Power board fuse burned and battery capacity insufficient (occasional)
Verify AC power supply connection
- Test point
- AC inlet
- Expected
- Properly connected
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vaporizer manifold assembly leaks
Negative pressure ball fully expands within 30 seconds during vaporizer manifold leak test, indicating damaged seals or washers.
Ranked causes
- 1. Rubber plain washer upper surface or contacted mechanical surface damaged (unknown)
- 2. Rubber plain washer lower surface or contacted mechanical surface damaged (unknown)
- 3. Seal rings damaged (unknown)
Test upper rubber washer seal by negative pressure ball
- Test point
- tube 53 outlet at vaporizer manifold
- Expected
- Negative pressure ball fully expands within 30 seconds
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACGO mode pressure insufficient patient safety
ACGO mode pressure fails to reach 30 cmH2O within 1 minute at 0.2 L/min O2 flow
Ranked causes
- 1. Leak in ACGO circuit or port seal defective (common)
- 2. O2 flow regulator malfunction or blockage (occasional)
Verify ACGO port is properly sealed
- Test point
- ACGO port on system
- Expected
- Port is plugged securely
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AIR cylinder pressure drops excessively
AIR cylinder pressure gauge drops more than 10% of initial pressure within 1 minute after closing cylinder
Ranked causes
- 1. Leak at AIR yoke connection or cylinder (common)
- 2. Defective AIR cylinder valve (occasional)
Inspect AIR yoke connection and washer
- Test point
- AIR cylinder yoke on rear panel
- Expected
- Cylinder seated flush, clean washer in place
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm messages display without sound patient safety
Alarm text appears on screen but no audible alarm sound is produced
Ranked causes
- 1. Speaker is damaged (occasional)
- 2. Speaker cable is disconnected (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic backup flow control system (BFCS) fails to deploy patient safety
BFCS does not extend during automatic backup flow control test
Ranked causes
- 1. BFCS mechanically stuck (occasional)
- 2. Solenoid of BFCS is faulty (occasional)
Visually inspect BFCS for mechanical obstruction
- Test point
- BFCS assembly
- Expected
- No visible obstruction or damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BFCS mode O2/AIR line pressure drops excessively
Line pressure gauge shows drop exceeding 10 psi per 100 seconds in BFCS mode when hoses disconnected
Ranked causes
- 1. Leak in O2 or AIR line pressure circuit (common)
- 2. Defective back pressure regulator or related valve (occasional)
Prepare system for O2/AIR line pressure test in BFCS mode
- Test point
- System settings
- Expected
- BFCS deployed, O2 and AIR needle valves fully open
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BFCS three-way valve failure
Three-way valve in BFCS circuit not functioning correctly, measured flow less than expected
Ranked causes
- 1. Three-way valve malfunction or flow sensor error (common)
- 2. O2 flow sensor error in total limb (occasional)
Set O2 flows around 1LPM under EFCS and check total limb flow vs O2 limb flow
- Test point
- [Setup]?[Service]?[Data Monitors]?[Component]?[FCS]
- Expected
- Total limb flow not obviously smaller than O2 limb flow by more than 0.2LPM, ≤0.2LPM difference
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system has significant leak in mechanical ventilation patient safety
Folding bag collapses or excessive fresh gas compensation required during mechanical ventilation
Ranked causes
- 1. Leak in bellows, sodalime canister, patient tubes, flow sensors or connectors (common)
- 2. PEEP safety valve not functioning correctly (occasional)
Observe fresh gas compensation rate and bellows bag behavior
- Test point
- Bellows housing and fresh gas inlet
- Expected
- Minimal fresh gas compensation; bellows bag does not collapse
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow sensor (user level)
Tidal volume measurement shows deviation greater than 9% from set value, or flow sensors have been replaced in breathing circuit.
Ranked causes
- 1. Flow sensor ageing or environmental drift (common)
- 2. Flow sensor replacement or reinstallation (occasional)
Perform breathing system leak test in mechanical ventilation mode
- Test point
- Mechanical ventilation mode test result
- Expected
- Leak test passed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PEEP valve pressure patient safety
Verify pressure sensor accuracy across PEEP valve settings (5, 20, 50, 70, 90 cmH2O) using calibration device.
Ranked causes
- 1. Pressure sensor calibration required (common)
- 2. Deviation exceeds tolerance after calibration (occasional)
Set safety valve to ON in diagnostic menu
- Test point
- Calibration Front Control Panel
- Expected
- Safety valve ON
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Flow Sensors
Inspiratory or expiratory reverse flow detected, or sampling lines incorrectly connected
Ranked causes
- 1. Check valve failure allowing reverse flow (common)
- 2. Sampling lines connected in incorrect order (common)
Check valve inspection
- Test point
- Check valve
- Expected
- Functioning correctly, no reverse flow
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 Absorber Canister Not Locked
CO2 canister not mounted or not locked properly
Ranked causes
- 1. CO2 absorber canister not mounted or improperly seated (common)
- 2. Cable between canister and ventilator control board disconnected or defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication between CPU and infrared communication board fails
Data transmission fails between main CPU board and infrared communication module
Ranked causes
- 1. Cables between CPU board and infrared board are loose (common)
- 2. Communication board or CPU board is damaged (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive Gas Pressure Low caution
Drive gas pressure below normal operating range
Ranked causes
- 1. Actual gas supply pressure low or defective pressure switch (common)
- 2. Pressure switch or ventilator control board connection issue (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EFCS mode O2/N2O/AIR line pressure drops excessively
Line pressure gauge shows drop exceeding 10 psi per 100 seconds in EFCS mode when hoses disconnected
Ranked causes
- 1. Leak in O2, N2O, or AIR line pressure circuit (common)
- 2. Defective proportional valve or 3-way valve assembly (occasional)
Prepare EFCS mode with proportional valves set to maximum
- Test point
- SERVICE menu FCS TESTS
- Expected
- 3-Way Valve (CPU/FPGA) set to H; O2/N2O/AIR Prop Valve DA all set to 4000
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow Sensor Failure
Inspiratory or expiratory flow sensor out of range or disconnected
Ranked causes
- 1. Flow sensor zero point or measurement out of specifications (common)
- 2. Flow sensor cable disconnected or loose connection (common)
Check flow sensor zero point within specifications
- Test point
- Flow sensor zero point
- Expected
- Within specifications
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor sampling lines connected incorrectly
A/D value decreases with increasing actual flow, or A/D value stays unchanged, or A/D value reaches saturation quickly. Diagnostic Tests shows anomalous flow sensor readings.
Ranked causes
- 1. Two sampling lines connected backward (occasional)
- 2. One or both sampling lines not connected or broken (occasional)
Observe A/D value of flow sensor with gradual increase of D/A value of inspiratory valve
- Test point
- Flow sensor A/D value in Diagnostic Tests
- Expected
- A/D value should increase gradually with increased gas supply
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insp Valve Failure
Inspiratory valve voltage error or flow error detected
Ranked causes
- 1. Inspiratory flow sensor error (common)
- 2. Inspiratory valve voltage error or power supply connection issue (common)
Check inspiratory flow sensor specifications
- Test point
- Inspiratory flow sensor
- Expected
- Within specifications
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory valve does not open correctly patient safety
With increased D/A value in Diagnostic Tests, flow sensor values change very little and low gas flow felt at water collection cup outlet. Standard flow measurement device fails to reach 90 L/min target.
Ranked causes
- 1. Inspiratory valve assembly defective (common)
- 2. D/A converter on Ventilator Control Board defective (occasional)
Gradually increase D/A value of inspiratory valve; observe flow sensor readings and gas flow at water collection cup outlet
- Test point
- Flow sensor A/D values and outlet gas flow
- Expected
- Flow sensor values should change substantially and adequate gas flow at outlet
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
N2O cylinder pressure drops excessively
N2O cylinder pressure gauge drops more than 10% of initial pressure within 1 minute after closing cylinder
Ranked causes
- 1. Leak at N2O yoke connection or cylinder (common)
- 2. Defective N2O cylinder valve (occasional)
Inspect N2O yoke connection and washer
- Test point
- N2O cylinder yoke on rear panel
- Expected
- Cylinder seated flush, clean washer in place
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No gas output through valve in mechanical ventilation mode patient safety
Gas does not flow from valve during mechanical ventilation even though mode is selected
Ranked causes
- 1. Auto/Manual ventilation switch is defective (occasional)
- 2. Valve cannot be opened (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 concentration measurement fails or has great deviations patient safety
AG Module reports oxygen concentration readings that are inaccurate or vary significantly from expected values during operation.
Ranked causes
- 1. AG Module not calibrated (common)
- 2. AG Module damaged (rare)
Verify AG Module calibration status
- Test point
- service mode AG Module status
- Expected
- AG Module calibrated per last service date
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 cylinder pressure drops excessively
O2 cylinder pressure gauge drops more than 10% of initial pressure within 1 minute after closing cylinder
Ranked causes
- 1. Leak at O2 yoke connection or cylinder (common)
- 2. Defective O2 cylinder valve (occasional)
Inspect O2 yoke connection and washer
- Test point
- O2 cylinder yoke on rear panel
- Expected
- Cylinder seated flush, clean washer in place
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Circuit Leak patient safety
Pressure less than 2cmH2O for 30 seconds during mechanical ventilation or patient not connected
Ranked causes
- 1. Breathing circuit connection loose or flow sensor misconfigured (common)
- 2. Breathing system leakage (common)
Check tidal volume measurement accuracy
- Test point
- Flow sensor, breathing circuit
- Expected
- Accurate measurement within specifications
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PEEP safety valve does not switch on or off correctly patient safety
When PEEP safety valve is switched on, no click heard, or pressure does not respond to valve switching. 'Drive Gas Pressure Low' alarm when PEEP safety valve is permanently OFF and gas supplies within specifications.
Ranked causes
- 1. PEEP safety valve assembly defective (common)
- 2. Safety valve drive voltage circuit on Ventilator Control Board defective (occasional)
In Diagnostic Tests, switch PEEP safety valve on; listen for click sound
- Test point
- PEEP safety valve click audible
- Expected
- Subtle click should be heard when switching on
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PEEP Valve Failure
PEEP valve voltage error or pressure error detected
Ranked causes
- 1. PEEP valve pressure sensor error (common)
- 2. PEEP valve voltage error or power supply connection issue (common)
Check PEEP circuit pressure sensor
- Test point
- PEEP circuit pressure sensor
- Expected
- Within specifications
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen flashes or flickers
Display is lit and shows content but screen flickers or flashes intermittently
Ranked causes
- 1. Power board failure causing power fluctuations (occasional)
- 2. Main control board LVDS time sequence abnormal (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen is lit but displays no content
Display backlight is on but screen appears blank without any visible content or menu
Ranked causes
- 1. Screen power supply fuse burned, no 3.3V output (occasional)
- 2. Main control board failure with no display output (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor zero point check
Verify pressure and flow sensor zero points are within normal range; indicates sensor aging and calibration status.
Ranked causes
- 1. Sensor zero point out of range (aging or drift) (occasional)
- 2. Sensor requires replacement (zero point critically out of range) (rare)
Turn off all fresh gases
- Test point
- Fresh gas flow controls
- Expected
- All gas flows set to zero
7 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System switch assembly leaks
Negative pressure ball fully expands within 30 seconds during system switch assembly leak test, indicating gas leakage or damaged tubing.
Ranked causes
- 1. Tubing connected to system switch assembly damaged (unknown)
- 2. System switch assembly damaged (unknown)
Test tubes connected to system switch assembly for leaks
- Test point
- tube 45 at flow regulator and tube 43 at Y piece
- Expected
- Negative pressure ball fully expands within 30 seconds
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Three-way valve malfunction patient safety
Three-way valve fails to switch properly or measured flow is insufficient when switched to BFCS limb during test
Ranked causes
- 1. Gas supply connection abnormal or supply gas pressure low (occasional)
- 2. Three-way valve is defective (occasional)
Verify gas supply connection and pressure
- Test point
- Gas supply line to three-way valve, drive gas pressure gauge
- Expected
- Gas supply secure, pressure 350-450 kPa, 350-450 kPa
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vt Not Achieved patient safety
Tidal volume less than setting value for 6 consecutive cycles, difference >20% or 50ml
Ranked causes
- 1. Breathing circuit leak (common)
- 2. Pressure sensor measurement accuracy error (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-way Valve Failure
Error in solenoid valve electrical signal control status
Ranked causes
- 1. Solenoid valve connection line issue (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Absorb canister assembly leak patient safety
Pressure drops more than 10 cmH2O in 30 seconds during absorb canister pressure hold test, indicating internal leakage.
Ranked causes
- 1. Damaged seals or connectors on absorb canister assembly (unknown)
Absorb canister assembly pressure retention
- Test point
- Anesthesia machine calibration device pressure reading
- Expected
- Pressure drop of 10cmH2O or less within 30s after stopping injector push, or less
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACGO assembly leak caution
Negative pressure ball expands within 30 seconds when ACGO assembly is isolated, indicating internal leakage.
Ranked causes
- 1. Damaged seals in ACGO assembly or circuit adapter connectors (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACGO Failure
Status of ACGO switch is incorrect
Ranked causes
- 1. Drive gas pressure out of normal range or ACGO switch malfunction (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AG airway occluded patient safety
AG module pump rate dropped below minimum threshold indicating airway obstruction or sampling line blockage
Ranked causes
- 1. AG module sampling line or airway occluded, pump rate below 20 ml/min (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AG airway occluded alarm not activating patient safety
Alarm 'AG Airway Occluded' does not appear when inlet hose is occluded
Ranked causes
- 1. Airway occlusion detection sensor failure (unknown)
Verify 'AG Airway Occluded' alarm when inlet hose occluded
- Test point
- AG module inlet hose and monitor screen
- Expected
- Alarm 'AG Airway Occluded' appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AG calibration failed patient safety
AG module calibration procedure did not complete or calibration values out of acceptable range
Ranked causes
- 1. AG module calibration failure or sensor drift (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- AG communication stopped
- AG data limit exceeded
- AG hardware error
- AG initialization error
- AG module 'Airway Occluded' alarm does not activate
- AG module 'No Watertrap' alarm does not appear or disappear appropriately
- AG Module airway occlusion detection failure
- AG module calibration verification
- AG module does not load or display correctly
- AG module does not unload cleanly
- AG module does not unload or unload message does not appear
- AG module hardware malfunction
- AG module measurement abnormal
- AG Module missing watertrap alarm
- AG module not detected or not loading correctly
- AG Module not loading or error on insertion
- AG module software version incompatible
- AG no watertrap alarm not appearing or not clearing
- AG watertrap disconnected
- AG watertrap requires replacement
- AG watertrap type mismatch
- AG zeroing failed
- AIR pipeline pressure gauge reading differs from test gauge by >0.1 MPa
- AIR supply failure alarm does not activate
- AIR supply failure not triggering alarm
- AIR supply inlet pressure outside specification range
- Airway pressure gauge shows inaccurate reading or pointer cannot move
- Auto Ventilation Disabled
- Auto Ventilation Disabled - Leak Test Failed
- Auto Ventilation is Non-Functional
- Automatic backup flow control system (BFCS) fails to retract
- Aux Control Module Comm Stop
- Auxiliary light does not illuminate
- Battery replacement
- Bellows housing or manual ventilation valve leak
- Breathing system leak in manual ventilation mode
- Breathing system leak in mechanical ventilation mode
- Breathing system leaks in manual ventilation mode
- Breathing system leaks in mechanical ventilation mode
- Calibration Failure after 15 minutes
- Calibration failure: Flow sensor (automatic)
- Calibration: ACGO mode pressure verification
- Calibration: AG Module concentration
- Calibration: BFCS line pressure leak test
- Calibration: EFCS flowmeter zero offset
- Calibration: EFCS line pressure leak test
- Calibration: Flow accuracy of EFCS
- Calibration: Flow sensor accuracy
- Calibration: Flow sensor measurement deviations
- Calibration: Flow sensor system
- Calibration: Flow sensors
- Calibration: O2 sensor at 100% oxygen
- Calibration: O2 sensor at 21% oxygen
- Calibration: Pressure and flow sensor zero point
- Calibration: Pressure sensor system
- CGO assembly leaks
- Check pressure sensor accuracy
- Check valve pressure integrity test
- Check valve pressure retention test
- Clear package software installation
- CO2 measurement accuracy out of specification
- Desflurane measurement accuracy out of specification
- DGSS circuit pressure below minimum or exceeds maximum
- DGSS malfunction or improper pressure
- Drive gas pipeline leak in circuit adapter
- EFCS gas concentration or flow verification failure
- EFCS gas flow accuracy out of specification
- Electronic ACGO Undetected
- Electronic flowmeter measurement abnormal
- End-tidal CO2 measurement out of range
- End-tidal desflurane measurement out of range
- End-tidal enflurane measurement out of range
- End-tidal halogenated agent measurement out of range
- End-tidal isoflurane measurement out of range
- End-tidal N2O measurement out of range
- End-tidal O2 measurement out of range
- End-tidal sevoflurane measurement out of range
- Enflurane measurement accuracy out of specification
- Error code [00 00 00 02] displayed
- Error code [00 00 00 04] displayed
- Error code [00 00 00 08] displayed
- Error code [00 00 00 10] displayed
- Error code [00 00 00 20] displayed
- Error code [00 00 00 40] displayed
- Error code [00 00 00 80] displayed
- Error code [00 00 01 00] displayed
- Expiratory valve drive gas pressure switch defective
- External AG module not connected
- External AG self-test failure
- Flow regulator knob gets loose
- Flow regulator leaks
- Flow sensor accuracy check
- Flow sensor measurements inconsistent or inaccurate
- Flow sensor readings inconsistent across three sensors
- Flowmeter float indicates inaccurate value or remains unmoved
- Fresh gas circuit connections leak
- Fresh gas pipeline leak in circuit adapter
- Gas mixer leaks
- Gas pump rate out of range
- Gas supplies cannot be turned off after machine is turned off
- Halothane measurement accuracy out of specification
- High inspiratory O2 alarm activation test
- High minute volume alarm activation test
- High peak airway pressure alarm activation test
- Inspired CO2 measurement out of range
- Inspired desflurane measurement out of range
- Inspired enflurane measurement out of range
- Inspired halogenated agent measurement out of range
- Inspired isoflurane measurement out of range
- Inspired N2O measurement out of range
- Inspired O2 measurement out of range
- Inspired sevoflurane measurement out of range
- Internal AG calibration failure error 08
- Internal AG communication failure or malfunction error 05
- Internal AG hardware error 01
- Internal AG initialization error 04
- Internal AG installation or malfunction error 03
- Internal AG parameter values out of range error 12
- Internal AG pump rate low error 10
- Internal AG self-test error 02
- Internal AG watertrap change required error 11
- Internal AG watertrap device loose error 09
- Internal AG zeroing failure error 07
- Isoflurane measurement accuracy out of specification
- License activation
- Line pressure gauge reading inaccurate
- Line pressure gauge readings inaccurate
- Line voltage alarm does not activate when AC power lost
- Line voltage alarm not activating on AC loss
- Low inspiratory O2 alarm activation test
- Low minute volume alarm activation test
- Low peak airway pressure alarm activation test
- Machine cannot be powered on after turning on
- Manual circuit leak test fails
- Mechanical ventilation not provided normally
- Mixed halogenated agents with high MAC
- Mixed halogenated agents with invalid measurement
- Mixed halogenated anesthetic agents detected
- N2O measurement accuracy out of specification
- N2O pipeline pressure gauge reading differs from test gauge by >0.1 MPa
- N2O supply failure alarm does not activate
- N2O supply failure not triggering alarm or cutoff
- Needle valve leaks
- O2 and AIR line isolation failure in BFCS mode
- O2 flush button assembly leaks
- O2 flush flow verification
- O2 measurement accuracy out of specification
- O2 needle valve basal flow out of adjustment
- O2 pipeline pressure gauge reading differs from test gauge by >0.1 MPa
- O2 sensor error
- O2 Supply Failure
- O2 supply failure alarm does not activate
- O2 supply failure not triggering alarm or cutoff
- O2 supply inlet pressure outside specification
- O2 supply pressure switch defective or out of specification
- O2:N2O ratio test (with AG Module, EFCS mode)
- O2:N2O ratio test (without AG Module, EFCS mode)
- O2:N2O ratio test with EFCS (with AG Module)
- O2:N2O ratio test with EFCS (without AG Module)
- PEEP Safety Valve Failure
- PEEP valve does not open correctly
- Position switch where BFCS ejects is faulty
- Position switch where BFCS retracts is faulty
- Preparation for pneumatic tests
- Preset pressure and tidal volume do not match measured values
- Pressure and flow zeroing (service)
- Pressure relief valve at breathing connection leaks
- Pressure sensor accuracy check
- Pressure sensor measurements inconsistent or inaccurate
- Remove calibration gas regulator
- Sensor zero points out of normal range
- Sevoflurane measurement accuracy out of specification
- Software package selection for board replacement
- Software update installation
- Software upgrade sequence execution
- Suction regulator not functioning correctly
- Suction regulator pressure not responding correctly
- System does not return to Standby mode after AG module removal
- Top light does not illuminate
- Top light not illuminating
- Total Flow Sensor Self Test failure
- Total flowmeter leaks
- Touchpad is not responsive
- Touchpad not functional
- Unable to exit standby mode
- Vaporizer installed improperly causing leak
- Vaporizer interlock malfunction
- Vaporizer interlock not functioning
- Vaporizer leak test
- Vaporizer manifold assembly damaged
- Vaporizer manifold assembly seal damaged
- Vaporizer manifold internal seal damaged
- Ventilation stops during operation
- Ventilator Comm Stop
- Ventilator Voltage Error
- Waste gas scavenger does not adjust between MIN and MAX
- Waste gas scavenger malfunction
197 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
[00 00 00 02] | Drive gas pressure too low condition | Drive gas (O2) pressure falls below specified minimum during calibration | error |
00 00 00 02 | O2 supply pressure is low during 100% calibration process | O2 supply pressure insufficient during 100% calibration | error |
00 00 00 04 | O2 sensor is disconnected with sampled data greater than 2900 AD value | O2 sensor disconnection detected during calibration | error |
[00 00 00 04] | Auto/Manual ventilation switch in incorrect position | Auto/Manual switch detected on Manual (bag) position during calibration startup | error |
[00 00 00 08] | Inspiratory flow sensor zero point error | Zero point AD value out of range: ≥2000 or <0 (EPSON); ≥26457 or <554 (DSP) | error |
00 00 00 08 | 21% calibration value outside expected range 150~500 AD value | O2 sensor reads outside normal 21% oxygen range during calibration | error |
[00 00 00 10] | Expiratory flow sensor zero point error | Zero point AD value out of range: ≥2000 or <0 (EPSON); ≥26457 or <554 (DSP) | error |
00 00 00 10 | 100% calibration value outside expected range 800~2028 AD value | O2 sensor reads outside normal 100% oxygen range during calibration | error |
[00 00 00 20] | Internal flow sensor zero point error | Zero point AD value out of range: ≥2000 or <0 (EPSON); ≥26457 or <554 (DSP) | error |
00 00 00 20 | Error writing to EEPROM memory during calibration | System fails to save calibration data to EEPROM | error |
[00 00 00 40] | Inspiratory flow sensor measurement range error | Sensor sampling AD value exceeds upper limit: >3900 (EPSON) or >60000 (DSP) at ~90 L/min flow | error |
[00 00 00 80] | Airway pressure sensor calibration data not unidirectional | Calibration detects non-linear or scattered sensor response | error |
[00 00 00 80] | Expiratory flow sensor measurement range error | Sensor sampling AD value exceeds upper limit: >3900 (EPSON) or >60000 (DSP) at ~90 L/min flow | error |
[00 00 01 00] | Internal (ventilator) flow sensor measurement range error | Sensor sampling AD value exceeds upper limit: >3900 (EPSON) or >60000 (DSP) at ~45 L/min flow | error |
[00 00 01 00] | PEEP pressure sensor calibration data not unidirectional | Calibration detects non-linear or scattered sensor response | error |
[00 00 02 00] | Inspiratory flow sensor bidirectional flow detected | Calibration data indicates flow in both directions rather than unidirectional only | error |
[00 00 02 00] | Airway pressure sensor resolution error | Calibration detects insufficient sensor resolution or limited AD range | error |
[00 00 04 00] | PEEP pressure sensor resolution error | Calibration detects insufficient sensor resolution or limited AD range | error |
[00 00 04 00] | Expiratory flow sensor bidirectional flow detected | Calibration data indicates flow in both directions rather than unidirectional only | error |
[00 00 08 00] | Valve output pressure too low | Calibration detects PEEP valve output below acceptable minimum | error |
[00 00 08 00] | Internal flow sensor bidirectional flow detected | Calibration data indicates flow in both directions rather than unidirectional only | error |
[00 00 10 00] | Flow change not unidirectional | Calibration detects non-monotonic flow response | error |
[00 00 10 00] | Inspiratory flow sensor resolution error | Sensor output resolution insufficient; signal noise or quantization error detected | error |
[00 00 20 00] | Communication with calibration device interrupted | Calibration device connection lost or communication timeout | error |
[00 00 20 00] | Resolution error in expiratory flow sensor | System calibration process detects expiratory flow sensor resolution fault | error |
[00 00 40 00] | Resolution error in internal flow sensor | System calibration detects internal flow sensor resolution fault | error |
[00 00 40 00] | System fails to write EEPROM | Data storage error during calibration | error |
[00 00 80 00] | Valve output flow is low | Calibration detects valve output flow below minimum threshold | error |
[00 00 80 00] | ACGO switch in ON position | Calibration attempted with ACGO switch ON | error |
[00 01 00 00] | Valve resolution is not enough | Calibration detects insufficient valve resolution | error |
[00 02 00 00] | Maximum pressure value cannot be found | Calibration unable to reach maximum pressure setpoint | error |
[00 02 00 00] | Flow change is not unidirectional | Calibration detects bidirectional or non-monotonic flow changes | error |
[00 04 00 00] | Minimum pressure value cannot be found | Calibration unable to reach minimum pressure setpoint | error |
[00 04 00 00] | Communication with calibration device is interrupted | Calibration process loses connection to external calibration device | error |
[00 08 00 00] | System fails to write EEPROM | Calibration cannot write calibration data to EEPROM memory | error |
[00 08 00 00] | PEEP valve closed pressure exceeds threshold (DSP only) | When PEEP valve is closed, calibration device reads over 3 cmH2O | error |
[00 10 00 00] | Sampling line leakage or disconnection (DSP only) | Calibration detects pressure loss in sampling circuit | error |
[00 10 00 00] | ACGO switch is in ON position | Calibration detects ACGO switch not in OFF position | error |
[00 20 00 00] | Maximum valve opening point cannot be found | Pressure calibration cannot establish maximum flow reference point | error |
[00 40 00 00] | Minimum valve opening point cannot be found | Pressure calibration cannot establish minimum flow reference point | error |
[00 80 00 00] | DSP platform: Excessive time finding mid-range DA value | Search for DA causing flow between 9-25L/min takes longer than 50 seconds | warning |
[01 00 00 00] | DSP platform: Calibration device zero point out of range | Calibration device zero point measurement exceeds acceptable range | error |
[02 00 00 00] | DSP platform: Tube is disconnected | Calibration detects disconnected or loose tube during pressure calibration | error |
[04 00 00 00] | DSP platform: Inspiratory valve leakage detected | When inspiratory valve is closed, flow measured by sensor exceeds 1L/min | error |
0x00000001 | CPU AVDD voltage too low | AVDD supply voltage drops below threshold | error |
0x00000001 | CPU self-test error | EFCS startup self-test | critical |
0x00000002 | CPU AVDD voltage too high | AVDD supply voltage exceeds threshold | error |
0x00000002 | RAM self-test error | EFCS startup self-test | critical |
0x00000004 | Address line self-test error | EFCS startup self-test | critical |
0x00000004 | Position switch where BFCS ejects is faulty | During system operation when BFCS ejection position detected as faulty | error |
0x00000004 | CPU DVDD voltage too low | DVDD supply voltage drops below threshold | error |
0x00000008 | Watchdog self-test error | EFCS startup self-test | critical |
0x00000008 | CPU DVDD voltage too high | DVDD supply voltage exceeds threshold | error |
0x00000008 | Position switch where BFCS retracts is faulty | During system operation when BFCS retraction position detected as faulty | error |
0x00000010 | CPU DVCC voltage too low | DVCC supply voltage drops below threshold | error |
0x00000010 | FLASH self-test error | EFCS startup self-test | critical |
0x00000010 | BFCS three-way valve failure | Run BFCS auto selftest; measured flow by total flow sensor less than 0.2/min | error |
0x00000020 | CPU DVCC voltage too high | DVCC supply voltage exceeds threshold | error |
0x00000020 | O2 proportional valve self-test error | EFCS startup self-test | critical |
0x00000040 | AIR proportional valve self-test error | EFCS startup self-test | critical |
0x00000040 | FPGA VPP voltage too low | VPP supply voltage drops below threshold | error |
0x00000080 | FPGA VPP voltage too high | VPP supply voltage exceeds threshold | error |
0x00000080 | N2O proportional valve self-test error | EFCS startup self-test | critical |
0x00000100 | FPGA DVCC voltage too low | FPGA DVCC supply voltage drops below threshold | error |
0x00000100 | O2 flow leakage detected | EFCS startup or runtime self-test | critical |
0x00000200 | AIR flow leakage detected | EFCS startup or runtime self-test | critical |
0x00000200 | FPGA DVCC voltage too high | FPGA DVCC supply voltage exceeds threshold | error |
0x00000400 | N2O flow leakage detected | EFCS startup or runtime self-test | critical |
0x00000400 | FPGA 3.3V voltage too low | 3.3V supply voltage drops below threshold | error |
0x00000800 | Startup zero reading error | EFCS startup initialization | critical |
0x00000800 | FPGA 3.3V voltage too high | 3.3V supply voltage exceeds threshold | error |
0x00001000 | FPGA 1.2V voltage too low | 1.2V supply voltage drops below threshold | error |
0x00001000 | FPGA configuration error | Startup initialization | critical |
0x00002000 | FPGA 1.2V voltage too high | 1.2V supply voltage exceeds threshold | error |
0x00004000 | AVCC voltage too low | AVCC supply voltage drops below threshold | error |
0x00008000 | AVCC voltage too high | AVCC supply voltage exceeds threshold | error |
0x00010000 | 3-way valve power-off voltage too low | Three-way valve control voltage below threshold when powered off | error |
0x00020000 | 3-way valve power-off voltage too high | Three-way valve control voltage exceeds threshold when powered off | error |
0x00040000 | 3-way valve power-on voltage too high | Three-way valve control voltage exceeds threshold when powered on | error |
0x00080000 | O2 limb flow sensor communication failure | I2C communication between FPGA and Sensirion O2 flow sensor fails for 1 second continuous | error |
0x00100000 | AIR limb flow sensor communication failure | I2C communication between FPGA and Sensirion AIR flow sensor fails for 1 second continuous | error |
0x00200000 | N2O limb flow sensor communication failure | I2C communication between FPGA and Sensirion N2O flow sensor fails for 1 second continuous | error |
0x00400000 | O2 branch flow not achieved (exceeds 0.4 LPM) | O2 flow deviation exceeds 0.4 L/min from setpoint | error |
0x00800000 | Balance gas AIR branch flow not achieved (exceeds 0.4 LPM) | AIR flow deviation exceeds 0.4 L/min from setpoint | error |
0x01000000 | Balance gas N2O branch flow not achieved (exceeds 0.4 LPM) | N2O flow deviation exceeds 0.4 L/min from setpoint | error |
0x02000000 | O2 temperature limit exceeded | O2 supply gas temperature exceeds 50°C | warning |
0x0400000 | AIR temperature limit exceeded | AIR supply gas temperature exceeds 50°C | warning |
0x0800000 | N2O temperature limit exceeded | N2O supply gas temperature exceeds 50°C | warning |
0x1000000 | FPGA failure | FPGA internal self-test or operation failure | critical |
0x2000000 | Flow sensor single selftest error | Flow sensor output inconsistency between limb and total flow sensors | error |
3-way Valve Failure | Solenoid valve electrical signal control error | Solenoid valve electrical control signal status error detected | info |
ACGO Failure | ACGO switch status is incorrect | ACGO switch status does not match expected state | warning |
AG Airway Occluded | AG module sampling line or airway occluded | Pump rate dropped below 20 ml/min for more than one second | error |
AG Cal. Failed | AG module calibration procedure failed | Calibration routine did not complete or calibration values invalid | error |
AG Change Watertrap | AG module watertrap requires replacement | Watertrap maintenance interval reached or performance degraded | error |
AG Comm Stop | Communication with AG module lost or stopped | AG module communication timeout or communication error detected | error |
AG Data Limit Error | AG module measurement data out of valid range | Measured values exceeded data validity limits during operation | error |
AG Hardware Error | AG hardware error detected | AG module hardware malfunction detected during operation or self-test | error |
AG Hardware Malfunction module | AG module hardware malfunction or improper installation | AG module hardware failure detected or module not seated properly | error |
AG Init Error | AG module initialization failed | AG module failed to initialize during power-up or reset | error |
AG No Watertrap | AG module watertrap is not connected | Watertrap sensor detects missing or disconnected watertrap during operation | warning |
AG Watertrap Type Wrong | Wrong watertrap type installed for patient type | Adult watertrap detected when patient type is set to infant | warning |
AG Zero Failed | AG module zeroing calibration failed | Zero calibration routine did not complete or produced invalid results | warning |
Air Supply Failure | Compressed air supply pressure loss | Air supply pressure drops below threshold | warning |
Auto Ventilation Disabled | System self test failure, automatic ventilation disabled, manual only | System self test failure in one or more modules | warning |
Auto Ventilation Disabled - Leak Test Failed | Automatic circuit leak test failure | Automatic leak test procedure detects circuit failure | warning |
Auto Ventilation is Non-Functional | Auto ventilation non-functional but Auto/Manual switch in Auto position | Machine in manual-only mode while Auto/Manual switch is in Auto position | critical |
Aux Control Module Comm Stop | Lost communication with CPU board for 10 seconds | Communication link between CPU Board and ventilator control board interrupted for 10 seconds | critical |
Aux Control Module Selftest Error | Auxiliary control module self-test failed | Startup initialization | error |
Backup Flow Control Deployment Failure | Backup flow control system solenoid actuator malfunction | Solenoid actuator fails to deploy backup flow control | critical |
Backup Flow Control Enabled | Backup flow control system active | Manual activation or automatic activation due to flow sensor malfunction | info |
Backup Flow Control Error | Backup flow control position sensor or LED circuit malfunction | BFCS ejection/retraction position sensor malfunction or LED voltage input error | warning |
Backup Flow Control Retraction Failure | Backup flow control system fails to retract | Stepper motor fails to retract backup flow control | warning |
Backup Flow Control Valves Open | Backup flow control needle valve not fully closed during EFCS operation | Position switch detects needle valve in open position during EFCS use | warning |
Balance Gas Branch Flow Not Achieved | Balance gas flow below setpoint | Balance gas flow exceeds preset by >10% or >0.2 L/min (greater prevails) | info |
Battery in Use | Backup power active due to AC mains loss | AC power supply disconnected or failed | warning |
Battery Undetected | Battery not detected by system | System battery monitoring | warning |
Breathing Circuit Not Mounted | Breathing circuit not properly installed or connected | Breathing circuit connection not detected | critical |
Bundle Version Error | Software versions are not compatible | Startup initialization | error |
Calibrate Flow Sensor and Insp Valve | Flow sensor calibration table invalid | Startup or calibration check | warning |
Calibrate Pressure Sensor and PEEP Valve | Pressure sensor calibration table invalid | Startup or calibration check | warning |
Calibration Failure! Please try again. | Calibration process failed to complete successfully | Detected after Begin button selected with ventilation sound or error conditions | error |
Check Flow Sensors | Inspiratory or expiratory reverse flow detected | Check valve detects reverse flow in insp or exp line | warning |
CO2 Absorber Canister Not Locked | CO2 canister not mounted or not locked properly | CO2 absorber canister is not detected as properly mounted or locked | critical |
CO2 Accuracy Unspecified | CO2 measurement accuracy out of specification | Measured CO2 value exceeded module accuracy specification range | warning |
Des Accuracy Unspecified | Desflurane measurement accuracy out of specification | Measured desflurane value exceeded module accuracy specification range | warning |
Drive Gas Pressure Low | Drive gas pressure below normal operating range | Drive gas pressure detected below normal specifications | critical |
Electronic ACGO Undetected | No electronic ACGO module detected by system | System initialization unable to detect electronic ACGO module | info |
Electronic Flow Control Error | Gas flow control system malfunction | Voltage error in CPU/FPGA supplies; valve malfunction; flow sensor fault; temperature or communication error | warning |
Electronic Flow Control Error | Total flow sensor self test failed after sensor replacement | Self-test initiated after total flow sensor replacement; sensor failed to validate during initialization | error |
Enf Accuracy Unspecified | Enflurane measurement accuracy out of specification | Measured enflurane value exceeded module accuracy specification range | warning |
EtCO2 Overrange | End-tidal CO2 measurement exceeds module measurement range | Measured EtCO2 value exceeded maximum specification during operation | warning |
EtDES Overrange | End-tidal desflurane measurement exceeds module measurement range | Measured EtDES value exceeded maximum specification during operation | warning |
EtENF Overrange | End-tidal enflurane measurement exceeds module measurement range | Measured EtENF value exceeded maximum specification during operation | warning |
EtHAL Overrange | End-tidal halogenated agent measurement exceeds module measurement range | Measured EtHAL value exceeded maximum specification during operation | warning |
EtISO Overrange | End-tidal isoflurane measurement exceeds module measurement range | Measured EtISO value exceeded maximum specification during operation | warning |
EtN2O Overrange | End-tidal N2O measurement exceeds module measurement range | Measured EtN2O value exceeded maximum specification during operation | warning |
EtO2 Overrange | End-tidal O2 measurement exceeds module measurement range | Measured EtO2 value exceeded maximum specification during operation | warning |
EtSEV Overrange | End-tidal sevoflurane measurement exceeds module measurement range | Measured EtSEV value exceeded maximum specification during operation | warning |
External AG Module Disconnected | External AG module not connected in FDA standard configuration | FDA configuration detected but external AG module not connected to system | error |
External AG Self Test Error | External anesthetic gas module self-test failed | Startup initialization | warning |
Fan Failure | CPU board fan malfunction | Fan speed monitoring | warning |
Fan Failure 02 | Module rack fan malfunction | Module rack fan speed monitoring | warning |
[FF FF FF FF] | Communication error | System unable to communicate with calibration device or control board | critical |
FiCO2 Overrange | Inspired CO2 measurement exceeds module measurement range | Measured FiCO2 value exceeded maximum specification during operation | warning |
FiDES Overrange | Inspired desflurane measurement exceeds module measurement range | Measured FiDES value exceeded maximum specification during operation | warning |
FiENF Overrange | Inspired enflurane measurement exceeds module measurement range | Measured FiENF value exceeded maximum specification during operation | warning |
FiHAL Overrange | Inspired halogenated agent measurement exceeds module measurement range | Measured FiHAL value exceeded maximum specification during operation | warning |
FiISO Overrange | Inspired isoflurane measurement exceeds module measurement range | Measured FiISO value exceeded maximum specification during operation | warning |
FiN2O Overrange | Inspired N2O measurement exceeds module measurement range | Measured FiN2O value exceeded maximum specification during operation | warning |
FiO2 Overrange | Inspired O2 measurement exceeds module measurement range | Measured FiO2 value exceeded maximum specification during operation | warning |
FiO2 Too High | Inspired oxygen concentration above high alarm limit | Inspired O2 exceeds setpoint during ventilation | warning |
FiO2 Too Low | Inspired oxygen concentration below low alarm limit | Inspired O2 falls below setpoint during ventilation | warning |
FiSEV Overrange | Inspired sevoflurane measurement exceeds module measurement range | Measured FiSEV value exceeded maximum specification during operation | warning |
Flowmeter Comm Stop | Communication lost between CPU and flow sensor board | CPU fails to communicate with flow sensor board for 10 seconds | warning |
Flowmeter Selftest Error | Flow sensor board self-test failed | Startup initialization | error |
Flow Sensor Failure | Inspiratory or expiratory flow sensor out of range or disconnected | Insp or Exp flow out of range, or internal flow sensor connection lost | info |
Hal Accuracy Unspecified | Halothane measurement accuracy out of specification | Measured halothane value exceeded module accuracy specification range | warning |
Heating Module Failure | Anesthesia heating module temperature out of control | Both resistance temps >106°C for 20s OR one >110°C for 15s | error |
Incompatible AG Software Version | AG module software version below minimum required version 1.7.3 | AG module firmware version detected at startup or during self-test | error |
Insp Valve Failure | Inspiratory valve voltage error or flow error | Inspiratory valve voltage out of spec or flow sensor error detected | warning |
Internal AG Error 01 | Internal AG module hardware error | Internal AG module detected hardware malfunction during operation or self-test | warning |
Internal AG Error 02 | Internal AG module self-test error | Internal AG module self-test failed during startup or periodic diagnostic | warning |
Internal AG Error 03 | Internal AG module not properly installed or malfunction | Internal AG module not seated properly or hardware failure detected | warning |
Internal AG Error 04 | Internal AG module initialization error | Internal AG module failed to initialize during startup or reset | warning |
Internal AG Error 05 | Internal AG module malfunction or communication failure | Internal AG module detected malfunction or communication error with main unit | warning |
Internal AG Error 07 | Internal AG module zeroing failure | Internal AG module zeroing calibration routine failed to complete | warning |
Internal AG Error 08 | Internal AG module calibration failure | Internal AG module calibration procedure failed to complete successfully | warning |
Internal AG Error 09 | Internal AG module watertrap substitute device loose or disconnected | Internal AG module detected watertrap substitute device not securely attached | warning |
Internal AG Error 10 | Internal AG module pump rate below 20 ml/min for more than 1 second | Internal AG pump rate dropped below minimum threshold during operation | warning |
Internal AG Error 11 | Internal AG module watertrap substitute device requires replacement | Internal AG module detected watertrap substitute device maintenance interval reached | warning |
Internal AG Error 12 | Internal AG module parameter values exceed normal range | Internal AG module measured parameters outside normal operating range | warning |
Internal AG Selftest Error 02 | Internal anesthetic gas module self-test failed | Startup initialization | warning |
IP Address Conflict | Network IP address already in use | Network initialization | warning |
Iso Accuracy Unspecified | Isoflurane measurement accuracy out of specification | Measured isoflurane value exceeded module accuracy specification range | warning |
Keyboard Comm Stop | Communication lost between CPU and keyboard/display | CPU fails to communicate with keyboard for 10 seconds | warning |
Keyboard Self Test Error | Keyboard self-test failed | Startup initialization | error |
Low Battery Voltage! | Battery voltage critically low | Battery voltage below 10.6V for 5 seconds | error |
Mixed Agent | Two halogenated agents with low MAC and one agent reading invalid | AG module detects multiple agents with low MAC and measurement error on one agent | error |
Mixed Agent and MAC≥3 | Two halogenated agents detected with MAC ≥3 | AG module detects multiple halogenated agents with high combined MAC value | error |
MV Too High | Minute volume exceeds high alarm limit | Total minute ventilation exceeds setpoint | warning |
MV Too Low | Minute volume below low alarm limit | Total minute ventilation falls below setpoint | warning |
N2O Accuracy Unspecified | N2O measurement accuracy out of specification | Measured N2O value exceeded module accuracy specification range | warning |
N2O Supply Failure | Nitrous oxide supply pressure loss | N2O supply pressure drops below threshold | warning |
NO Fresh Gas | Fresh gas delivery disabled or not flowing | Fresh gas turned off in manual or mechanical ventilation mode | warning |
O2 Accuracy Unspecified | O2 measurement accuracy out of specification | Measured O2 value exceeded module accuracy specification range | warning |
O2 Branch Flow Not Achieved | Oxygen flow below setpoint | O2 flow exceeds preset value by >10% or >0.2 L/min (greater prevails) | info |
O2 Sensor Error | O2 sensor malfunction in AG module | O2 sensor failure detected during operation or self-test | error |
O2 Supply Failure | O2 supply pressure failure detected | O2 pressure switch detects supply pressure failure | critical |
Patient Circuit Leak | Pressure <2cmH2O for 30s during mechanical ventilation or patient not connected | Breathing system pressure drops below 2cmH2O for 30 seconds during ventilation | warning |
Paw Too High | Peak airway pressure exceeds high alarm limit | Peak pressure reaches or exceeds setpoint during inspiration | warning |
Paw Too Low | Peak airway pressure below low alarm limit | Peak pressure falls below setpoint during ventilation | warning |
PEEP Safety Valve Failure | PEEP safety valve malfunction | Safety valve operation | warning |
PEEP Valve Failure | PEEP expiratory valve malfunction | Valve operation or pressure monitoring | warning |
PEEP Valve Failure | PEEP valve voltage error or pressure error | PEEP valve voltage out of spec or pressure sensor error detected | warning |
Pinsp Not Achieved | Inspiratory pressure less than target value for 6 consecutive cycles | In pressure mode, Pinsp is <2/3 of setting (if >9cmH2O) or <setting-3cmH2O (if <9cmH2O) for 6 cycles | info |
Power Board High Temp | Power board temperature exceeds safe limits | Power board temperature greater than 95°C for 10 seconds continuously | warning |
Power Supply Voltage Error | Power supply voltage out of specification | Voltage monitoring at startup or runtime | error |
Power Supply Voltage Error (Power Board) | Power supply voltage out of specification | Voltage monitoring on power board | error |
Power System Comm Stop | Communication lost with CPU board | Power board communication timeout (10 seconds) | error |
Rate Overrange | Gas pump rate measurement exceeds module measurement range | Measured pump rate value exceeded maximum specification during operation | warning |
RT Clock Failure | Real-time clock chip malfunction | Startup or time-dependent operation | error |
RT Clock Needs Battery | Real-time clock battery depleted or missing | Startup initialization | error |
Sev Accuracy Unspecified | Sevoflurane measurement accuracy out of specification | Measured sevoflurane value exceeded module accuracy specification range | warning |
System going DOWN, Battery depleted! | Battery voltage completely depleted | Battery voltage below 10.2V | error |
Ventilator Comm Stop | Lost communication with CPU board for 10 seconds | Communication link to CPU board interrupted for 10 seconds | critical |
Ventilator Initialization Error | Ventilator control board initialization failed | Startup initialization | error |
Ventilator Selftest Error | Ventilator control board self-test failed | Startup initialization | error |
Ventilator Voltage Error | 5V or 12V voltage error in ventilator control module | Input voltage on power cable detected out of specification | error |
Ventilator Voltage Error | Ventilator control module voltage out of specification | Startup or runtime voltage measurement | error |
Vt Not Achieved | Tidal volume less than setting value for 6 consecutive cycles, >20% or 50ml difference | TVi <TV setting value for 6 cycles for >20% or 50ml difference | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.