Mindray A5: troubleshooting and error codes

The ServiceForge knowledge base holds 301 documented symptoms for the Mindray A5, with 603 ranked causes, 399 diagnostic checks with test points and expected values, and 178 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Pressure calibration: System failure

Troubleshooting for pressure calibration process that fails with error code display.

Ranked causes

  1. 1. Drive gas pressure too low (error 00 00 00 02) (common)
  2. 2. Auto/Manual switch in Manual position (error 00 00 00 04) (common)
  3. 3. Airway pressure sensor zero point error (error 00 00 00 08) (occasional)
  4. 4. PEEP pressure sensor zero point error (error 00 00 00 10) (occasional)
  5. 5. Airway pressure sensor measurement range error (error 00 00 00 20) (occasional)
  6. 6. PEEP pressure sensor measurement range error (error 00 00 00 40) (occasional)
  7. 7. Airway pressure sensor calibration data not unidirectional (error 00 00 00 80) (rare)
  8. 8. PEEP pressure sensor calibration data not unidirectional (error 00 00 01 00) (rare)
  9. 9. Airway pressure sensor resolution error (error 00 00 02 00) (rare)
  10. 10. PEEP pressure sensor resolution error (error 00 00 04 00) (rare)
  11. 11. Valve output pressure too low (error 00 00 08 00) (occasional)
  12. 12. Flow change not unidirectional (error 00 00 10 00) (rare)
  13. 13. Communication with calibration device interrupted (error 00 00 20 00) (occasional)
  14. 14. System fails to write EEPROM (error 00 00 40 00) (rare)
  15. 15. ACGO switch in ON position (error 00 00 80 00) (common)
  16. 16. Valve resolution insufficient (error 00 01 00 00) (rare)
  17. 17. Communication error (error FF FF FF FF) (occasional)

17 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pneumatic circuit leaks patient safety

Loss of pressure or gas leakage from breathing system connections, seals, or components; may affect ventilation delivery.

Ranked causes

  1. 1. Damaged seals on lifting device connections (common)
  2. 2. Damaged manual bag (common)
  3. 3. Damaged breathing tube (common)
  4. 4. Damaged Paw pressure gauge seal (common)
  5. 5. Damaged water collection cup seal (common)
  6. 6. Damaged O2 sensor or plug seal (common)
  7. 7. Damaged check valve dome seal (common)
  8. 8. Damaged bag arm seal (common)
  9. 9. Damaged prepak assembly seal (common)
  10. 10. Defective APL valve seals (common)
  11. 11. Defective support tube seals on median plate (common)
  12. 12. Defective Auto/Manual ventilation switch seals (common)
  13. 13. Damaged CGO assembly or connectors (occasional)
  14. 14. Leaky flow sensor pressure sampling pipeline (occasional)
First check

Turn off system switch

Test point
System power
Expected
Off

14 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system leaks patient safety

Air leaks from breathing circuit components including CO2 absorber canister, seals, bellows, or valve assembly

Ranked causes

  1. 2. Sealing piece for absorbent canister assembly is damaged (occasional)
  2. 3. Seal for bag arm is damaged (occasional)
  3. 4. Water collection cup is loose (common)
  4. 5. Seal for water collection cup assembly is damaged (occasional)
  5. 6. Seal for circuit adapter assembly is damaged (occasional)
  6. 8. Bellows sealing cushion falls off or is damaged (occasional)
  7. 9. Valve cover of breathing valve assembly not installed properly (occasional)
  8. 10. Seal for valve cover of breathing valve assembly is damaged (occasional)
  9. 12. Seal for O2 sensor or sensor plug is damaged (occasional)
  10. 13. Breathing tube connecting patient is damaged (occasional)
  11. 14. Bellows is damaged (occasional)
  12. 15. Sealing connection of other breathing system parts is damaged (occasional)
  13. 16. Condensate valve not installed properly or internal seal is damaged (occasional)

16 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: System pressure and PEEP sensors

Automated pressure calibration procedure for airway and PEEP pressure sensors during anesthesia machine service.

Ranked causes

  1. 1. Drive gas pressure too low (common)
  2. 2. Airway or PEEP pressure sensor zero point error (common)
  3. 3. Pressure sensor sampling line disconnected or improperly connected (common)
  4. 4. PEEP valve maximum pressure insufficient (occasional)
  5. 5. Pneumatic circuit connection error between calibration device and ventilator control board (occasional)
  6. 6. Communication connection error between calibration device and anesthesia machine (occasional)
  7. 7. Calibration device settings incorrect (occasional)
  8. 8. Calibration data not correct (occasional)
First check

Verify drive gas pressure

Test point
O2 pressure gauge
Expected
At least 200 kPa

2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Seal damage in breathing system components patient safety

Seals damaged or deteriorated in O2 sensor, check valve dome, bag arm, prepak assembly, APL valve, support tube, or Auto/Manual switch.

Ranked causes

  1. 1. O2 sensor seal damaged (common)
  2. 2. Check valve dome seal damaged (common)
  3. 3. Bag arm seal damaged (common)
  4. 4. Prepak assembly seal damaged (common)
  5. 5. APL valve seal damaged (occasional)
  6. 6. Support tube on median plate seal damaged (occasional)
  7. 7. Auto/Manual ventilation switch seal damaged (occasional)
First check

Turn off system switch

Test point
Power switch
Expected
System powered off

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Anesthesia machine cannot be started caution

System does not power up; AC and battery indicator lamps not lit; screen not lit; machine cannot ventilate

Ranked causes

  1. 1. AC power supply not connected or battery capacity insufficient (common)
  2. 2. AC input breaker tripped with insufficient battery (common)
  3. 3. Power board fuse burned with insufficient battery (occasional)
  4. 4. System switch damaged or cable disconnected (occasional)
  5. 5. Power board hardware failure: no voltage output (15.2V, 3.3V, 5V, 12V) (rare)
  6. 6. Power board software code error: no voltage output (15.2V, 3.3V, 5V, 12V) (occasional)
First check

Verify AC power supply connection

Test point
AC power connector
Expected
AC cable firmly connected to wall outlet and machine connector

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow waveform displayed irregularly patient safety

Irregular or unstable flow measurement display on monitor screen

Ranked causes

  1. 1. Flow sensor pressure sampling pipeline leaks (common)
  2. 2. Flow sensor assembly not installed properly (common)
  3. 3. Water accumulated inside flow sensor assembly (occasional)
  4. 4. Flow sensor membrane distorted, dirty or resistance changed (occasional)
  5. 5. Flow sensor assembly damaged (rare)
  6. 6. Pressure sensor on fresh flow sensor board defective (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaccurate tidal volume patient safety

Tidal volume readings are incorrect or inconsistent; may show TVe and TVi discrepancies or inaccurate flow measurements.

Ranked causes

  1. 3. Significant breathing system leaks with low fresh gas flow (occasional)
  2. 4. Water buildup in flow sensor (occasional)
  3. 5. Flow sensor membrane distorted, dirty, or resistance changed; zero drift occurs (occasional)
  4. 8. Inlet gas flow regulator defective in expiratory valve assembly (occasional)
  5. 9. Plimit set too low, causing premature expiration (occasional)
  6. 10. TVe and TVi mismatch due to sensor measurement error (occasional)
First check

Turn off flow regulators

Test point
Flow regulators
Expected
Off

7 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure - Ventilation sound heard then failure patient safety

Ventilation sound occurs initially but calibration fails shortly after Start pressed

Ranked causes

  1. 1. Sensor sampling line not connected or connected in reverse order (common)
  2. 2. Inspiratory valve maximum flow less than 90 L/min (occasional)
  3. 3. Pneumatic circuit connection error between machine and calibration device (occasional)
  4. 4. Communication connection fault between machine and calibration device (occasional)
  5. 5. Calibration device settings error (occasional)
First check

Verify inspiratory and expiratory flow sensor sampling lines are connected correctly

Expected
All three sensor lines properly connected in correct order

4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration failure: Ventilation sound heard, then failure prompt

After Begin is selected, ventilation sound is heard but '[Calibration Failure! Please try again.]' prompt appears very soon

Ranked causes

  1. 1. Sampling line not connected or improperly connected on airway or PEEP pressure sensor (common)
  2. 2. PEEP valve maximum pressure less than 95 cmH2O (occasional)
  3. 3. Pneumatic circuit connection error between calibration device and ventilator control board (occasional)
  4. 4. Communication connection error between calibration device and anesthesia machine (occasional)
  5. 5. Calibration device settings incorrect (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow sensor patient safety

Flow sensor calibration fails or measurement deviations exceed 9% compared to setting value

Ranked causes

  1. 1. Flow sensor aging or environmental drift (common)
  2. 2. Flow sensor replaced (common)
  3. 3. O2 sensor replaced or recalibrated (common)
  4. 4. Breathing system leak (occasional)
  5. 5. Drive gas pressure out of specification (occasional)
First check

Perform breathing system leak test

Test point
Breathing circuit
Expected
Leak test passed in mechanical ventilation mode

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow sensors patient safety

Flow sensor calibration for breathing system to correct tidal volume measurement deviation or after component replacement

Ranked causes

  1. 1. Tidal volume measurement deviation exceeds 9% after prolonged use (common)
  2. 2. Flow sensor in patient circuit replaced (occasional)
  3. 3. Expiratory valve assembly replaced (occasional)
  4. 4. Ventilator control board replaced (occasional)
  5. 5. Flow measurement deviation exceeds 5% of reading or 1 L/min (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow sensor assembly problems patient safety

Flow waveform displays irregularly or flow measurement is incorrect due to flow sensor assembly issues

Ranked causes

  1. 1. Flow sensor assembly not installed properly (common)
  2. 2. Water built up inside flow sensor assembly (occasional)
  3. 3. Flow sensor membrane distorted, dirty, or internal resistance changes (occasional)
  4. 4. Flow sensor is damaged (occasional)
  5. 5. Pressure sensor on fresh flow sensor board is defective (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas supply tube leak caution

Gas escaping from supply tube or connection points

Ranked causes

  1. 1. Gas supply tube is damaged (common)
  2. 2. Connection seal is damaged (common)
  3. 3. Quick plug-in connector is leaking or damaged (occasional)
  4. 4. Pipeline gas supply inlet assembly check valve or seal is damaged (occasional)
  5. 5. Drive gas pipeline is leaking (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low-pressure pneumatic circuit system leak patient safety

Breathing system leaks detected during mechanical or manual ventilation modes; pressure loss during test

Ranked causes

  1. 1. Bellows housing or sealing cushion leaky (common)
  2. 2. Manual ventilation mode bag or APL valve defective (occasional)
  3. 3. Expiratory valve assembly or drive gas pipeline leak (occasional)
  4. 4. Fresh gas or ACGO pipeline leak (occasional)
  5. 5. Circuit adapter or patient circuit damaged (occasional)
First check

Leak test of breathing system in mechanical ventilation mode

Test point
Breathing system circuit
Expected
No detectable pressure loss during test
Tools
As specified in section 3.7.2.2

5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 supply inlet assembly leaks patient safety

Pressure falls dramatically and continuously after O2 pipeline gas supply is turned off, indicating gas leakage

Ranked causes

  1. 1. O2 flush button assembly seal damaged or leaking (common)
  2. 2. System switch assembly seal damaged (common)
  3. 3. Vaporizer improperly installed or seal damaged (occasional)
  4. 4. Flowmeter or ORC assembly leaks (occasional)
  5. 5. Flow regulator or restrictor leaks (occasional)
First check

Observe pressure gauge reading trend after gas supply turned off

Test point
Test pressure gauge (downstream of inlet assembly)
Expected
Pressure should hold steady or fall slowly; not fall dramatically and continuously

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power system communication stopped caution

Lost communication with CPU board for 10 seconds. Power board cannot communicate with main processor.

Ranked causes

  1. 1. Communication cable failure or disconnection (common)
  2. 2. Power board software incorrect or needs update (occasional)
  3. 3. Power board processor reset required (occasional)
  4. 4. Power board failure (occasional)
  5. 5. CPU board failure (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Anesthesia machine cannot start patient safety

System fails to start after power switch activation.

Ranked causes

  1. 1. System switch damaged (rare)
  2. 2. Cable disconnected from system switch (common)
  3. 3. Power board hardware circuit failure (occasional)
  4. 4. Power board software code error (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not detected caution

Battery is not detected by the system. Battery voltage may be out of specification or cable connection is loose.

Ranked causes

  1. 1. Battery voltage out of specification (common)
  2. 2. Cable not connected correctly (common)
  3. 3. Defective battery (occasional)
  4. 4. Power board failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system accessories damaged or leaky patient safety

Manual bag, breathing tubes, Paw pressure gauge, or water collection cup show signs of damage or leakage

Ranked causes

  1. 1. Manual bag damaged (occasional)
  2. 2. Breathing tube damaged (occasional)
  3. 3. Paw pressure gauge seal damaged (occasional)
  4. 4. Water collection cup seal damaged (occasional)
First check

Visually inspect manual bag for damage

Test point
Manual bag surface
Expected
No visible tears, cracks, or defects

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Breathing system leakage detected patient safety

Pressure reading falls below expected level during leak test; indicates faulty seals or connections in breathing system

Ranked causes

  1. 1. Breathing system leak in mechanical or manual ventilation mode (common)
  2. 2. Bellows housing or manual ventilation switch leak (occasional)
  3. 3. Circuit adapter pipeline leak (occasional)
  4. 4. Absorb canister assembly leakage (occasional)
First check

Perform leak test in manual ventilation mode

Test point
Paw pressure gauge
Expected
Pressure maintained at 30 cmH2O
Tools
O2 flow regulator, breathing tubes, Y piece

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 100% O2 patient safety

System calibration for oxygen sensor using 100% oxygen gas; failure indicated by fail code display or calibration not completing within ~3 minutes

Ranked causes

  1. 1. O2 sensor not connected (common)
  2. 2. O2 supply pressure insufficient (< 200 kPa) (common)
  3. 3. 21% O2 calibration not performed first (occasional)
  4. 4. O2 sensor count value out of normal range (450-2700) (occasional)
First check

Verify O2 sensor connection to cable

Test point
O2 sensor connector
Expected
Secure connection

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow sensor (Service) patient safety

Service-level flow sensor and inspiratory valve calibration using standard flow measurement device

Ranked causes

  1. 1. Ventilator control board replaced (common)
  2. 2. Drive gas assembly replaced (common)
  3. 3. Solenoid valve assembly replaced (common)
  4. 4. Flow sensor measurement deviation exceeds specification (occasional)
First check

Verify no sensor or valve technical alarms

Test point
Alarm display
Expected
No active sensor or valve alarms

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure and Flow Zeroing Failure patient safety

Anesthesia machine unable to zero pressure and flow sensors during calibration; zeroing process fails to complete successfully.

Ranked causes

  1. 1. Sensor zero point out of range (occasional)
  2. 2. Sampling line occlusion or three-way valve mechanical fault (occasional)
  3. 3. Ventilator control board faulty (occasional)
  4. 4. Solenoid valve assembly faulty (occasional)
First check

Verify sensor zero point is within acceptable range

Test point
Pressure and flow sensors
Expected
Zero point normal per section 3.7.3

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 absorber canister not locked caution

CO2 absorber canister is not properly mounted or not detected

Ranked causes

  1. 1. Canister not properly seated (common)
  2. 2. Canister detection cable fault (occasional)
  3. 3. Ventilator control board failure (occasional)
  4. 4. Canister switch failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code [00 00 00 08] displayed - Inspiratory sensor zero point error patient safety

Inspiratory flow sensor zero reference point has shifted or is invalid

Ranked causes

  1. 1. Fresh gas not turned off preventing zero reference measurement (common)
  2. 2. Inspiratory valve has not fully closed blocking zero reference (occasional)
  3. 3. Zero point calibration is out of specification (occasional)
  4. 4. Inspiratory flow sensor or ventilator control board failure (occasional)
First check

Ensure fresh gas flow is turned completely off

Expected
Fresh gas supply off

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Flow sensor zero drift or irregular waveform display patient safety

Flow measurement is inaccurate or waveform appears distorted on display; pressure sensor shows drift

Ranked causes

  1. 1. Flow sensor membrane distorted, dirty, or internal resistance changed (common)
  2. 2. Flow sensor assembly damaged (occasional)
  3. 3. Pressure sensor on fresh flow sensor board defective (occasional)
  4. 4. Flow sensor pressure sampling pipeline leak (common)
First check

Perform leak test of flow sensor pressure sampling pipeline (expiratory and inspiratory lines)

Test point
Flow sensor pressure sampling pipeline test fixture connectors
Expected
Pressure does not fall more than 5 cmH2O within 15 s at 70-90 cmH2O, within 15 s
Tools
Anesthesia machine calibration device, flow sensor pressure sampling pipeline test fixture, circuit adapter test fixture, injector, Φ6 silicone tube, Y piece

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas leak in pneumatic circuit patient safety

Escaping gas detected in supply lines, connectors, or gas inlet assembly; audible hiss or pressure loss

Ranked causes

  1. 1. Gas supply tube damaged or seal at connection damaged (common)
  2. 2. Quick plug-in connector leaks (common)
  3. 3. Pipeline gas supply inlet assembly leaks (occasional)
  4. 4. Drive gas pipeline leaks (occasional)
First check

Inspect negative pressure ball for leaks before use

Test point
Front sealing cover and gas inlet
Expected
Ball does not visibly inflate for at least 30 seconds when finger blocks inlet

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Heater ineffective or not heating patient safety

Unit does not reach or maintain set temperature; heater output insufficient; no heat felt at outlet

Ranked causes

  1. 1. Heater driver and control circuit on power board damaged (occasional)
  2. 2. Heater element damaged (occasional)
  3. 3. Heater internal sensor ineffective (occasional)
  4. 4. Heater cable disconnected or loose (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory valve failure caution

Inspiratory valve voltage error or inspiratory valve flow error. Inspiratory flow sensor out of specification or valve assembly connection issue.

Ranked causes

  1. 1. Inspiratory flow sensor out of specification (common)
  2. 2. Voltage issue at test point or valve assembly connection (occasional)
  3. 3. Valve assembly failure (occasional)
  4. 4. Power board failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak in gas supply system patient safety

Gas pressure not maintained or leakage detected in pipeline, connectors, or inlet assembly

Ranked causes

  1. 1. Gas supply tube is damaged or seal at connection is damaged (common)
  2. 2. Quick plug-in connector leaks (common)
  3. 3. Pipeline gas supply inlet assembly leaks (occasional)
  4. 4. Drive gas pipeline leaks (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Machine cannot start

Machine fails to start after power is applied

Ranked causes

  1. 1. System switch damaged (occasional)
  2. 2. Cable to system switch disconnected (common)
  3. 3. Power board hardware failure preventing voltage output (occasional)
  4. 4. Power board software error preventing voltage output (rare)
First check

Verify system switch cable connection

Test point
System switch connector
Expected
Cable properly seated

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PEEP valve failure caution

PEEP valve voltage error or PEEP valve pressure error. Pressure sensor on PEEP circuit out of specification or valve assembly connection issue.

Ranked causes

  1. 1. Pressure sensor out of specification (common)
  2. 2. Voltage issue at test point or valve assembly connection (occasional)
  3. 3. Expiratory valve assembly failure (occasional)
  4. 4. Power board failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen cannot be lit

Display screen remains off when machine powered on

Ranked causes

  1. 1. Cable to inverter disconnected (common)
  2. 2. Inverter damaged (occasional)
  3. 3. Power board hardware failure causing improper output (occasional)
  4. 4. Power board software failure causing improper output (rare)
First check

Verify inverter cable connection

Test point
Inverter connector
Expected
Cable properly seated

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen does not light caution

Display not powered; no illumination or visual output when machine switched on

Ranked causes

  1. 1. Inverter cable disconnected or inverter damaged (occasional)
  2. 2. Power board hardware failure: improper output (occasional)
  3. 3. Power board software failure: improper output (occasional)
  4. 4. Screen power supply fuse burned: no 3.3V output (occasional)
First check

Verify screen power supply 3.3V output

Test point
Screen power supply output
Expected
3.3V present at connector
Tools
Multimeter

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-way solenoid valve failure caution

Solenoid valve electrical signal control status error detected

Ranked causes

  1. 1. Solenoid valve connection fault (common)
  2. 2. Solenoid valve assembly failure (common)
  3. 3. Ventilator control board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Aux Control Module communication stopped caution

Lost communication with CPU board for 10 seconds

Ranked causes

  1. 1. Communication cable defect or disconnection (common)
  2. 2. Ventilator Control Module Board failure (occasional)
  3. 3. CPU Board failure (occasional)
First check

Restart the machine

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Aux control module selftest error on startup caution

Machine displays aux control module selftest error at power-on. CPU, Flash, watchdog timer error or communication failure between CPU and ventilator control board.

Ranked causes

  1. 1. Communication cable failure or disconnection (common)
  2. 2. Ventilator control board failure (occasional)
  3. 3. CPU board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery in use - AC power lost patient safety

Unit is operating on battery power. AC mains supply has been disconnected or failed.

Ranked causes

  1. 1. AC power supply disconnected (common)
  2. 2. AC supply voltage out of specification (occasional)
  3. 3. Power board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure - No ventilation sound after Begin selected patient safety

System does not produce ventilation sound and fails immediately after Begin pressed

Ranked causes

  1. 1. Auto/Manual ventilation switch in bag position instead of mechanical ventilation (common)
  2. 2. Drive gas (O2) pressure below 200 kPa specification (common)
  3. 3. Zero point error in inspiratory or expiratory flow sensor (occasional)
First check

Verify Auto/Manual ventilation switch is in mechanical ventilation position, not bag

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration failure with no ventilation sound patient safety

After Begin is selected, no ventilation sound is heard and Calibration Failure message is displayed

Ranked causes

  1. 1. Auto/Manual ventilation switch set to bag position (common)
  2. 2. Drive gas pressure below 200 kPa (common)
  3. 3. Zero point error in inspiratory/expiratory flow sensor (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure with ventilation sound heard patient safety

After Start is selected, ventilation sound heard then Calibration Failure message displayed

Ranked causes

  1. 1. Sensor sampling line not connected or reversed (common)
  2. 2. Inspiratory valve maximum flow below 90 L/min (occasional)
  3. 3. Pneumatic circuit, communication, or configuration error (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure: Code [00 01 00 00]

Insufficient valve resolution during calibration process

Ranked causes

  1. 1. Insufficient gas supply for calibration (common)
  2. 2. Calibration device malfunction (occasional)
  3. 3. PEEP valve defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure: communication error patient safety

Pressure calibration error [FF FF FF FF] displayed indicating communication error occurs.

Ranked causes

  1. 1. Transient communication fault (occasional)
  2. 2. Communication cable damage or poor connection (common)
  3. 3. Communication board failure (rare)
First check

Restart anesthesia machine

Test point
Machine power and startup
Expected
Machine restarts normally and enters standby

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure: No ventilation sound, immediate failure message patient safety

After Begin selected, no ventilation heard; failure message 'Calibration Failure! Please try again.' displays immediately

Ranked causes

  1. 1. Zero point error in airway pressure gauge or PEEP pressure sensor (common)
  2. 2. Sampling line not connected or improperly connected (common)
  3. 3. PEEP valve maximum pressure less than 95 cmH2O (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure: unidirectional flow not achieved patient safety

Pressure calibration error [00 02 00 00] displayed indicating flow change is not unidirectional.

Ranked causes

  1. 1. Tubes connected incorrectly (common)
  2. 2. Insufficient gas supply (occasional)
  3. 3. Calibration device malfunction (occasional)
First check

Verify all pneumatic connections are correct

Test point
Pressure sampling lines and four-way device
Expected
Tubes connected per Figures 4-52 and 4-61; P1 to monitor board, P2 to PEEP, tube #72 connected, tube #9 unconnected

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration failure: valve resolution insufficient patient safety

Pressure calibration error [00 01 00 00] displayed indicating valve resolution is not adequate.

Ranked causes

  1. 1. Insufficient gas supply (occasional)
  2. 2. Calibration device malfunction (occasional)
  3. 3. Inspiratory valve defective (common)
First check

Verify gas supply pressure and flow

Test point
Gas supply inlet
Expected
Adequate supply for calibration process

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration failure: Ventilation sound heard, delayed failure message patient safety

After Begin selected, ventilation sound is heard; failure message appears after several seconds: 'Calibration Failure! Please try again.'

Ranked causes

  1. 1. Pneumatic circuit connection error between calibration device and ventilator control board (common)
  2. 2. Communication connection error between calibration device and machine (common)
  3. 3. Calibration device settings error (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow sensor service calibration patient safety

Flow sensor and inspiratory valve calibration after equipment service or when measured flow deviation exceeds 5% of reading or 1 L/min

Ranked causes

  1. 1. Expiratory valve assembly replaced (occasional)
  2. 2. Ventilator control board replaced (occasional)
  3. 3. Flow measurement device reading deviation exceeds 5% or 1 L/min (occasional)
First check

Verify tubes are not leaking when connected

Test point
Calibration tube connections
Expected
No leaks detected

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow sensors (service level) patient safety

Service-level calibration of flow sensors and inspiratory valve after component replacement or when measured deviation exceeds threshold.

Ranked causes

  1. 1. Expiratory valve assembly replaced (occasional)
  2. 2. Ventilator control board replaced (occasional)
  3. 3. Ventilator flow sensor deviation exceeds 5% of reading or 1 L/min (occasional)
First check

Compare ventilator flow sensor reading with calibrated measurement device

Test point
Flow measurement comparison
Expected
Deviation within 5% of reading or 1 L/min (whichever greater), ±5% or ±1 L/min

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: O2 sensor patient safety

O2 sensor calibration fails or produces out-of-range readings during 21% or 100% calibration steps

Ranked causes

  1. 2. O2 supply pressure insufficient (common)
  2. 3. 21% calibration not completed before 100% calibration (common)
  3. 4. O2 sensor defective or aged (occasional)
First check

Verify O2 sensor cable connection

Test point
O2 sensor connector
Expected
Alarm [O2 Sensor Unconnected] not displayed

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: O2 sensor calibration patient safety

O2 sensor accuracy calibration at 21% (air) and 100% O2 when measurement deviation exceeds 3% or after sensor replacement

Ranked causes

  1. 1. O2 sensor measurement deviation exceeds 3% (occasional)
  2. 2. O2 sensor replaced (occasional)
  3. 3. Ventilator control board replaced (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensor service calibration patient safety

Pressure sensor and PEEP valve calibration after equipment service or when pressure measurement deviation exceeds 5% of reading or 2 cmH2O

Ranked causes

  1. 1. Ventilator control board replaced (occasional)
  2. 2. Expiratory valve assembly replaced (occasional)
  3. 3. Pressure sensor measurement deviation exceeds 5% or 2 cmH2O (occasional)

Calibration: Pressure sensors and PEEP valve patient safety

Airway pressure sensor, PEEP pressure sensor, and PEEP valve require calibration after replacing ventilator control board, drive gas assembly, or solenoid valve assembly, or when measured values deviate significantly from standard measurement device.

Ranked causes

  1. 2. Drive gas assembly replaced (common)
  2. 3. Solenoid valve assembly replaced (common)
  3. 4. Pressure sensor deviation from standard measurement device (occasional)
First check

Verify anesthesia machine is in standby mode

Test point
Machine mode
Expected
Standby mode

10 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensors and PEEP valve (service level) patient safety

Service-level calibration of pressure sensors and PEEP valve after component replacement or measurement deviation exceeds acceptable threshold.

Ranked causes

  1. 1. Ventilator control board replaced (occasional)
  2. 2. Expiratory valve assembly replaced (occasional)
  3. 3. Pressure sensor deviation exceeds 5% of reading or 2 cmH2O (occasional)
First check

Compare machine pressure sensor reading with standard pressure gauge

Test point
Pressure measurement comparison
Expected
Deviation within 5% of reading or 2 cmH2O (whichever greater), ±5% or ±2 cmH2O

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: System gas flow rates patient safety

O2, N2O, and AIR needle valve flow rates require calibration to meet specified ranges during anesthesia delivery

Ranked causes

  1. 1. O2 needle valve requires flow rate adjustment (common)
  2. 2. N2O resistor requires adjustment for correct flow rate (common)
  3. 3. AIR and N2O resistors on pneumatic block require adjustment (common)
First check

Measure O2 flow rate when N2O needle valve at maximum

Test point
O2 flow outlet
Expected
0.25 to 0.3 L/min, 0.25–0.3 L/min

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error code [00 00 00 10] displayed - Expiratory sensor zero point error patient safety

Expiratory flow sensor zero reference point has shifted or is invalid

Ranked causes

  1. 1. Fresh gas not turned off preventing zero reference measurement (common)
  2. 2. Inspiratory valve has not fully closed blocking zero reference (occasional)
  3. 3. Zero point calibration out of specification (occasional)
First check

Verify fresh gas supply is turned off

Expected
Fresh gas flow zero

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Flow sensor failure caution

Ventilator flow sensor output is out of normal operating range. Sensor zero point or measurement is out of specification.

Ranked causes

  1. 1. Flow sensor zero point out of specification (common)
  2. 2. Flow sensor measurement out of specification (common)
  3. 3. Ventilator control board failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow sensor sampling line connection error

Flow sensor has two sampling lines; errors include backward connection, one line not connected, or both lines not connected

Ranked causes

  1. 1. Flow sensor sampling lines connected backward (occasional)
  2. 2. Flow sensor sampling line broken or not connected (occasional)
  3. 3. Flow sensor low-pressure sampling line not connected (occasional)
First check

Connect anesthesia machine tubes using constant-flow connection method

Test point
Pneumatic circuit
Expected
Tubes properly connected per constant-flow method

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flowmeter selftest error on startup caution

Machine displays flowmeter selftest error at power-on. CPU, Flash, watchdog timer error or communication failure between CPU and fresh flow sensor board.

Ranked causes

  1. 1. Communication cable failure or disconnection (common)
  2. 2. Fresh flow sensor board failure (occasional)
  3. 3. CPU board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Heater ineffective
  • Low battery voltage
  • Network connection fails
  • O2 concentration measurement fails or shows deviations
  • PEEP safety valve failure
  • Power board voltage error
  • System shutting down - battery depleted
  • Touch panel not responding to input
  • Unit does not power on
  • Unit does not power on, indicators and screen dark
  • Vaporizer manifold assembly leaks
  • Ventilator communication stopped
  • Ventilator initialization error on startup
  • Ventilator selftest error on startup
  • Ventilator voltage error
  • Alarm messages without sound
  • Bellows housing or manual ventilation switch leak
  • Breathing system leak in manual ventilation mode
  • Calibrate O2 sensor
  • Calibration failure after 15 minutes
  • Calibration failure after 15 minutes of operation
  • Calibration failure: ACGO switch in ON position
  • Calibration failure: communication interrupted
  • Calibration failure: EEPROM write failed
  • Calibration failure: No ventilation sound, immediate failure prompt
  • Calibration failure: output flow too low
  • Calibration: Electronic flowmeter zeroing
  • Calibration: Flow sensor (user)
  • Calibration: Flow sensor user calibration
  • Calibration: Flow sensors (user level)
  • Calibration: Pressure and flow zeroing
  • CGO assembly leak or connection failure
  • CGO assembly seal or connector damage
  • Check flow sensors reverse flow detected
  • CPU to infrared board communication fails
  • Drive gas pressure low
  • Error code [00 00 00 02] displayed - Drive gas pressure too low
  • Error code [00 00 00 20] displayed - Internal sensor zero point error
  • Error code [00 00 00 40] displayed - Inspiratory sensor measurement range error
  • Error code [00 00 00 80] displayed - Expiratory sensor measurement range error
  • Error code [00 00 01 00] displayed - Internal sensor measurement range error
  • Error code [00 00 02 00] displayed - Inspiratory sensor non-unidirectional data
  • Error code [00 00 04 00] displayed - Expiratory sensor non-unidirectional data
  • Error code [00 00 08 00] displayed - Internal sensor non-unidirectional data
  • Error code [00 00 10 00] displayed - Inspiratory sensor resolution error
  • Error code [00 00 20 00] displayed - Expiratory sensor resolution error
  • Error code [00 00 40 00] displayed - Internal sensor resolution error
  • Fan failure
  • Flow sensor assembly not installed properly
  • Flow sensor measurement error
  • Flowmeter assembly leaks gas
  • Flowmeter voltage error on startup
  • Inspiratory valve does not open adequately
  • Inspiratory valve does not open correctly
  • N2O flow out of range
  • No Drive Gas Pressure Low alarm when pressure is low
  • No gas output from valve in mechanical ventilation mode
  • No O2 Supply Failure alarm when pressure is low
  • O2 sensor disconnected
  • O2 supply failure
  • Patient circuit leak detected
  • PEEP safety valve does not open or PEEP valve malfunction
  • PEEP safety valve does not operate correctly
  • PEEP valve does not open or close correctly
  • PEEP valve opening state abnormal during Diagnostic Tests
  • Pinsp not achieved
  • Pipeline pressure gauge readings fluctuate greatly
  • Power board overheating
  • Pressure sensor measurement error
  • Screen displays content but image flashes
  • Screen flashes
  • Screen illuminated but displays no content
  • Screen lights but displays no content
  • Screen lights but no content displayed
  • System switch assembly leaks
  • System switch assembly or connecting tube damaged
  • Tidal volume not achieved
  • Ventilator voltage error on startup
  • 3-way Valve Failure
  • Absorb canister assembly leak
  • AG airway occluded
  • AG calibration failure
  • AG communication failure
  • AG data limit error
  • AG hardware malfunction
  • AG initialization error
  • AG module hardware error
  • AG self-test failure
  • AG watertrap disconnected
  • AG watertrap requires replacement
  • AG watertrap type mismatch
  • AG zeroing failure
  • Air flow out of range
  • Air Flow Too High
  • AIR pipeline pressure gauge damaged
  • AIR pipeline pressure incorrect
  • AIR supply inlet assembly pressure drop excessive
  • AIR supply inlet assembly pressure out of specification
  • AIR supply inlet assembly regulator pressure out of specification
  • AIR supply pressure out of range
  • Airway pressure gauge defective
  • Airway pressure gauge inaccurate or pointer immobile
  • Airway pressure gauge shows inaccurate reading or pointer cannot move
  • APL valve pressure adjustment
  • Aux Control Module Comm Stop
  • Back pressure regulator leaks
  • Back pressure regulator leaks gas
  • Breathing Circuit Not Mounted
  • Calibration failure: Approximately 15 minutes after start
  • Calibration failure: Code [00 00 40 00]
  • Calibration failure: Code [00 00 80 00]
  • Calibration failure: Code [FF FF FF FF]
  • Calibration failure: Delay approximately 15 minutes after start
  • Calibration: 21% O2
  • Calibration: AG Module
  • Calibration: Cylinder yoke regulator pressure
  • Calibration: Electronic flowmeter zeroing (user level)
  • Calibration: O2 sensor at 100% O2
  • Calibration: O2 sensor at 21% O2
  • Calibration: ORC oxygen-nitrous oxide ratio control
  • Calibration: ORC Oxygen/Nitrous Oxide Ratio Control
  • Calibration: Pressure and flow zeroing (service level)
  • Calibration: Pressure sensors and PEEP proportional valve
  • Cannot exit standby mode
  • Check Flow Sensors
  • CO2 measurement accuracy unspecified
  • Desflurane measurement accuracy unspecified
  • Drive gas pipeline leak in circuit adapter
  • Drive gas pressure alarm incorrect or missing
  • Drive Gas Pressure Low alarm does not occur at 0.05 MPa
  • Drive Gas Pressure Low alarm incorrect trigger or no alarm when pressure low
  • Drive Gas Pressure Low alarm not triggered
  • Drive gas pressure low alarm not triggered at expected pressure
  • Drive Gas Pressure Low alarm occurs at wrong threshold
  • Drive Gas Pressure Low alarm occurs when pressure is normal
  • Drive gas pressure low alarm triggered at wrong pressure
  • End-tidal desflurane measurement overrange
  • End-tidal enflurane measurement overrange
  • End-tidal halothane measurement overrange
  • End-tidal isoflurane measurement overrange
  • End-tidal N2O measurement overrange
  • End-tidal O2 measurement overrange
  • End-tidal sevoflurane measurement overrange
  • Enflurane measurement accuracy unspecified
  • Error code [00 00 00 04] displayed - Manual ventilation switch issue
  • Exiting Standby fails
  • Expired CO2 measurement overrange
  • External AG module disconnected
  • Flow sensor calibration data incorrect
  • Flow sensor calibration failure
  • Flow sensor leaks
  • Flow sensor leaks gas
  • Flow sensor low pressure sampling line disconnected
  • Flow sensor measurement error or calibration data incorrect
  • Flow sensor sampling lines broken or disconnected
  • Flow sensor sampling lines connected backward
  • Flow sensor zero point outside specifications
  • Flowmeter assembly leaks
  • Flowmeter Comm Stop
  • Flowmeter Voltage Error
  • Fresh gas pipeline leak in circuit adapter
  • Gas delivery system leak detected
  • Gas mixer leaks
  • Gas mixer leaks gas
  • Halothane measurement accuracy unspecified
  • Heating Module Failure
  • Incompatible AG module software version
  • Insp Valve Failure
  • Inspiratory valve opening defective or inadequate flow
  • Inspired CO2 measurement overrange
  • Inspired desflurane measurement overrange
  • Inspired enflurane measurement overrange
  • Inspired halothane measurement overrange
  • Inspired isoflurane measurement overrange
  • Inspired N2O measurement overrange
  • Inspired O2 measurement overrange
  • Inspired sevoflurane measurement overrange
  • Internal Air Flow Failure
  • Internal N2O Flow Failure
  • Internal O2 Flow Failure
  • IP address conflict
  • Isoflurane measurement accuracy unspecified
  • Leak in gas delivery system
  • Mixed halogenated agents detected (low MAC)
  • Mixed halogenated agents with high MAC
  • Mixed halogenated agents with invalid reading
  • N2O Flow Too High
  • N2O measurement accuracy unspecified
  • Needle valve leaks
  • Needle valve leaks gas
  • Network connection failed
  • NO Fresh Gas
  • O2 concentration measurement fails or has great deviations
  • O2 flow out of range
  • O2 Flow Too High
  • O2 flush button assembly leaks
  • O2 measurement accuracy unspecified
  • O2 pressure switch does not trigger O2 Supply Failure alarm at 0.25 MPa
  • O2 sensor calibration failure
  • O2 sensor error
  • O2 supply alarm incorrect or missing
  • O2 Supply Failure alarm incorrect trigger or no alarm when pressure low
  • O2 Supply Failure alarm not triggered
  • O2 supply failure alarm not triggered at expected pressure
  • O2 Supply Failure alarm occurs at wrong threshold
  • O2 Supply Failure alarm occurs when pressure is normal
  • O2 supply failure alarm triggered at wrong pressure
  • O2 supply inlet assembly pressure out of specification
  • O2 supply inlet assembly regulator pressure out of specification
  • O2-N2O Ratio Error
  • Oxygen Ratio Controller (ORC) damaged
  • Oxygen Ratio Controller (ORC) leaks
  • Patient Circuit Leak
  • PEEP safety valve opening defective when drive gas pressure low
  • PEEP valve does not operate correctly
  • PEEP valve opening defective in mechanical ventilation
  • Pipeline pressure gauge reads inaccurately or shows no reading
  • Pipeline pressure gauge shows inaccurate readings or no readings
  • Power supply voltage error on startup
  • Pressure sensor calibration data incorrect
  • Pressure sensor calibration failure
  • Pressure sensor measurement error or calibration data incorrect
  • Pressure sensor sampling line connected incorrectly
  • Pressure sensor sampling line connection error
  • Pressure sensor sampling line not connected
  • Pump rate overrange
  • Real-time clock failure
  • Real-time clock needs battery
  • Replace O2 sensor
  • Screen lights with content but flashes repeatedly
  • Sensor zero point out of specification
  • Sevoflurane measurement accuracy unspecified
  • Standby mode exit fails
  • Total flowmeter assembly leaks
  • Total flowmeter leaks
  • Touch panel ineffective
  • Ventilation stops during operation
  • Ventilation stops suddenly but display and buttons work
  • Ventilation stops suddenly during operation
  • Ventilator Comm Stop
  • Vt Not Achieved

241 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
[00 00 00 02]Drive gas pressure too lowGas supply pressure insufficient during calibrationerror
00 00 00 02Drive gas pressure too lowPressure below threshold during calibration initializationerror
00 00 00 02O2 supply pressure is low during 100% calibration processO2 supply pressure was not sufficient during 100% calibration steperror
00 00 00 04Auto/Manual switch in Manual positionAuto/Manual switch detected in Manual mode during calibrationerror
00 00 00 04O2 sensor is disconnected; sampled data exceeds 2900 (AD value)Sampled AD value greater than 2900 indicates sensor open circuit or very high impedanceerror
[00 00 00 04]Auto/Manual switch in wrong positionAuto/Manual switch positioned to Manual during calibrationwarning
[00 00 00 08]Airway pressure sensor zero point errorZero point calibration failed for airway pressure sensorerror
00 00 00 08Airway pressure sensor zero point errorAirway pressure sensor zero point outside acceptable rangeerror
00 00 00 0821% calibration value is outside expected range 150~500 (AD value)21% O2 calibration step produced reading below 150 or above 500error
[00 00 00 08]Inspiratory flow sensor zero point errorSensor zero point AD greater than or equal to 2000 or less than 0error
00 00 00 10PEEP pressure sensor zero point errorPEEP pressure sensor zero point outside acceptable rangeerror
[00 00 00 10]Expiratory flow sensor zero point errorSensor zero point AD greater than or equal to 2000 or less than 0error
[00 00 00 10]PEEP pressure sensor zero point errorZero point calibration failed for PEEP pressure sensorerror
00 00 00 10100% calibration value is outside expected range 800~2028 (AD value)100% O2 calibration step produced reading below 800 or above 2028error
00 00 00 20Airway pressure sensor measurement range errorAirway sensor reading outside acceptable measurement rangeerror
00 00 00 20Error writing to EEPROM memorySystem unable to write calibration data to EEPROM storage during calibration completioncritical
[00 00 00 20]Measurement range error in airway pressure sensorAirway pressure sensor output exceeds or cannot reach expected measurement rangeerror
[00 00 00 20]Airway pressure sensor measurement range errorSensor unable to achieve full measurement range during calibrationerror
[00 00 00 40]Measurement range error in inspiratory flow sensorInspiratory flow sensor reading exceeds valid measurement range during calibrationerror
00 00 00 40PEEP pressure sensor measurement range errorPEEP sensor reading outside acceptable measurement rangeerror
[00 00 00 40]PEEP pressure sensor measurement range errorSensor unable to achieve full measurement range during calibrationerror
[00 00 00 80]Expiratory flow sensor measurement range errorSensor AD value greater than 3900 at 90 L/min flowerror
[00 00 00 80]Airway pressure sensor calibration data not unidirectionalCalibration data shows non-linear or backward response during pressure sweeperror
[00 00 00 80]Measurement range error in expiratory flow sensorExpiratory flow sensor reading exceeds valid measurement range during calibrationerror
00 00 00 80Airway pressure sensor calibration data not unidirectionalCalibration data shows non-linear or inconsistent responseerror
[00 00 01 00]Internal flow sensor measurement range errorSensor AD value greater than 3900 at 45 L/min flowerror
00 00 01 00PEEP pressure sensor calibration data not unidirectionalCalibration data shows non-linear or inconsistent responseerror
[00 00 01 00]PEEP pressure sensor calibration data not unidirectionalCalibration data shows non-linear or backward response during pressure sweeperror
00 00 02 00Airway pressure sensor resolution errorSensor resolution insufficient for accurate measurementerror
[00 00 02 00]Airway pressure sensor resolution errorInsufficient sensor resolution; difference between max and min AD <= 500 or single-point resolution < 1 cmH2Oerror
[00 00 02 00]Inspiratory flow sensor calibration data not unidirectionalCalibration data does not show expected unidirectional behaviorerror
[00 00 04 00]Expiratory flow sensor calibration data not unidirectionalCalibration data does not show expected unidirectional behaviorerror
[00 00 04 00]PEEP pressure sensor resolution errorInsufficient sensor resolution; difference between max and min AD <= 500 or single-point resolution < 1 cmH2Oerror
00 00 04 00PEEP pressure sensor resolution errorSensor resolution insufficient for accurate measurementerror
00 00 08 00Valve output pressure too lowPEEP valve output pressure below acceptable thresholderror
[00 00 08 00]Valve output pressure lowPEEP valve unable to generate sufficient output pressureerror
[00 00 08 00]Calibration data of internal flow sensor is not unidirectionalCalibration system detects bidirectional flow or inconsistent data trend in internal sensorerror
00 00 10 00Flow change not unidirectionalFlow measurement shows non-linear or inconsistent response during calibrationerror
[00 00 10 00]Resolution error in inspiratory flow sensorSensor resolution does not meet calibration requirementserror
[00 00 10 00]Change of flow is not unidirectionalFlow measurement shows non-monotonic or bidirectional behavior during calibrationerror
[00 00 20 00]Resolution error in expiratory flow sensorSensor resolution does not meet calibration requirementserror
00 00 20 00Communication with calibration device interruptedLoss of communication with external calibration deviceerror
[00 00 20 00]Communication with calibration device is interruptedData exchange between calibration device and machine stops or becomes unreliableerror
00 00 40 00System fails to write EEPROMData write to non-volatile memory unsuccessfulerror
[00 00 40 00]EEPROM write failureDuring calibration processerror
[00 00 40 00]Resolution error in internal flow sensorInternal flow sensor resolution insufficient or unstable during calibrationerror
[00 00 40 00]System fails to write EEPROMEEPROM write operation failed during calibration data saveerror
[00 00 80 00]Output flow of valve is lowValve output flow during calibration is less than specifiederror
00 00 80 00ACGO switch in ON positionACGO (automatic calibration gas outlet) switch detected in ON modeerror
[00 00 80 00]ACGO switch in ON positionAuto-calibration gas outlet switch activated during service calibrationwarning
[00 01 00 00]PEEP valve resolution insufficientValve control resolution inadequate for fine pressure controlerror
00 01 00 00Valve resolution insufficientPEEP valve resolution inadequate for precision calibrationerror
[00 02 00 00]Change of flow is not unidirectionalFlow changes exhibit bidirectional behavior during calibrationerror
[00 02 00 00]Maximum valve pressure not foundValve unable to reach specified maximum pressure during diagnostic sweeperror
[00 04 00 00]Minimum valve pressure not foundValve unable to reach specified minimum pressure during diagnostic sweeperror
[00 04 00 00]Communication with calibration device interruptedCommunication link between machine and calibration device failserror
[00 08 00 00]System fails to write EEPROMEEPROM write operation fails during calibrationerror
[00 10 00 00]ACGO switch positioned to ONACGO switch detected in ON position during calibrationerror
[00 20 00 00]Maximum value cannot be foundCalibration algorithm cannot determine maximum flow pointwarning
[00 40 00 00]Minimum value cannot be foundCalibration algorithm cannot determine minimum flow pointwarning
3-way Valve FailureThree-way solenoid valve malfunctionSolenoid valve electrical signal control status errorinfo
AG Airway OccludedAG sampling airway occludedPump rate drops below 20 ml/min for >1 secondcritical
AG Cal. FailedAG module calibration failedAG calibration procedure unsuccessfulcritical
AG Change WatertrapAG watertrap service interval expiredWatertrap has reached service life limiterror
AG Comm StopAG module communication failureLoss of communication with AG modulecritical
AG Data Limit ErrorAG module data transmission errorAG module data integrity check failserror
AG Hardware ErrorAG hardware error detectedAG module hardware malfunctionerror
AG Hardware Malfunction moduleAG module hardware malfunctionAG module improper installation or hardware faultcritical
AG Init ErrorAG module initialization errorAG module fails to initialize on startupcritical
AG No WatertrapAG module watertrap disconnectedWatertrap not detected when system initializeserror
AG Watertrap Type WrongAG watertrap type does not match patient typeAdult watertrap installed when patient type set to infanterror
AG Zero FailedAG module zeroing procedure failedAG module cannot zero sensor outputwarning
Air Flow Too HighAir flow exceeds maximum thresholdAir flow greater than 20 L/min for 1 secondinfo
Auto Ventilation DisabledAutomatic ventilation disabled due to system self-test failureSystem self-test failure during initializationinfo
Auto Ventilation Disabled - Leak Test FailedAutomatic ventilation disabled due to circuit leak test failureAutomatic circuit leak test failure during initializationinfo
Auto Ventilation is Non-FunctionalAutomatic ventilation non-functional while switch in Auto positionAuto/Manual switch in Auto position but automatic ventilation not functionalerror
Aux Control Module Comm StopCommunication loss between auxiliary control module and CPU boardLost communication with CPU board for 10 secondserror
Aux Control Module Selftest ErrorAuxiliary ventilator control module failed self-test during startupCPU communication failure or board internal error at power-onerror
Aux Control Module Selftest ErrorAuxiliary control module self-test failed. CPU, Flash, or watchdog timer error, or communication loss with ventilator control boardPower-oncritical
Battery in UseOperating on battery power due to AC failureAC power failure detectedinfo
Battery UndetectedBattery is not detected by the systemPower-on or runtimewarning
Breathing Circuit Not MountedBreathing circuit not properly installedBreathing circuit detection sensor indicates circuit not mountederror
Bundle Version ErrorSoftware versions are not compatibleIncompatible software versions detected at startuperror
Calibrate Flow Sensor and Insp ValveFlow sensor calibration table missing or corruptedCalibration table not found in EEPROM or checksum mismatchinfo
Calibrate Flow Sensor and Insp ValveCalibration table not found in EEPROM or checksum mismatch for flow sensor and inspiratory valvePower-on or runtimeinfo
Calibrate O2 SensorO2 sensor calibration table missing or corrupted in EEPROMEEPROM calibration data lost or checksum mismatch at startupwarning
Calibrate O2 SensorCalibration table not found in EEPROM or checksum mismatch for oxygen sensorPower-on or runtimeinfo
Calibrate Pressure Sensor and PEEP ValvePressure sensor calibration table missing or corrupted in EEPROMEEPROM calibration data lost or checksum mismatch at startupwarning
Calibrate Pressure Sensor and PEEP ValveCalibration table not found in EEPROM or checksum mismatch for pressure sensor and PEEP valvePower-on or runtimeinfo
Check Flow SensorsFlow sensor measurement error detectedInspiratory reverse flow OR expiratory reverse flow detectederror
CO2 Absorber Canister Not LockedCO2 absorber canister not properly mountedCO2 canister mounting detection sensor indicates canister not mountederror
CO2 Accuracy UnspecifiedCO2 measurement outside module accuracy rangeMeasured CO2 concentration exceeds specificationwarning
Des Accuracy UnspecifiedDesflurane measurement outside module accuracy rangeMeasured desflurane concentration exceeds specificationwarning
[Drive Gas Pressure Low]Drive gas (O2) pressure is lower than 200 kPaDrive gas supply pressure falls below threshold during calibration initiationerror
Drive Gas Pressure LowDrive gas (O2) pressure to expiratory valve assembly is below safe operating limitO2 supply pressure falls below pressure switch threshold (~0.2 MPa / 2.0 bar)warning
Drive Gas Pressure LowDrive gas (expiratory valve) pressure is below minimum threshold; alarm triggered by expiratory valve pressure switchExpiratory valve integrated pneumatic circuit pressure falls below 0.2 MPa (nominal switch point ~220 kPa)critical
Enf Accuracy UnspecifiedEnflurane measurement outside module accuracy rangeMeasured enflurane concentration exceeds specificationwarning
EtCO2 OverrangeEnd-tidal CO2 measurement exceeds module rangeEtCO2 concentration above maximum measurable valuewarning
EtDES OverrangeEnd-tidal desflurane measurement exceeds module rangeEtDES concentration above maximum measurable valuewarning
EtENF OverrangeEnd-tidal enflurane measurement exceeds module rangeEtENF concentration above maximum measurable valuewarning
EtHAL OverrangeEnd-tidal halothane measurement exceeds module rangeEtHAL concentration above maximum measurable valuewarning
EtISO OverrangeEnd-tidal isoflurane measurement exceeds module rangeEtISO concentration above maximum measurable valuewarning
EtN2O OverrangeEnd-tidal N2O measurement exceeds module rangeEtN2O concentration above maximum measurable valuewarning
EtO2 OverrangeEnd-tidal O2 measurement exceeds module rangeEtO2 concentration above maximum measurable valuewarning
EtSEV OverrangeEnd-tidal sevoflurane measurement exceeds module rangeEtSEV concentration above maximum measurable valuewarning
External AG Module DisconnectedExternal AG module not connected to systemFDA version expects external AG but module not foundcritical
External AG Self Test ErrorExternal anesthetic gas module self-test failureExternal AG module selftest malfunctioninfo
Fan FailureSystem fan speed below minimum thresholdFan speed 20% off nominal value (normal ~4000 RPM)warning
Fan FailureCooling fan speed 20% below nominal rotation speedFan speed monitoring indicates speed deviationwarning
Fan Failure 02Module rack fan speed below minimum thresholdModule rack fan speed less than 3640 RPM (normal ~4000 RPM)warning
[FF FF FF FF]Communication errorCommunication protocol or CRC error with control board or calibration devicecritical
FF FF FF FFCommunication errorGeneral communication failure between system componentscritical
FiCO2 OverrangeInspired CO2 measurement exceeds module rangeFiCO2 concentration above maximum measurable valuewarning
FiDES OverrangeInspired desflurane measurement exceeds module rangeFiDES concentration above maximum measurable valuewarning
FiENF OverrangeInspired enflurane measurement exceeds module rangeFiENF concentration above maximum measurable valuewarning
FiHAL OverrangeInspired halothane measurement exceeds module rangeFiHAL concentration above maximum measurable valuewarning
FiISO OverrangeInspired isoflurane measurement exceeds module rangeFiISO concentration above maximum measurable valuewarning
FiN2O OverrangeInspired N2O measurement exceeds module rangeFiN2O concentration above maximum measurable valuewarning
FiO2 OverrangeInspired O2 measurement exceeds module rangeFiO2 concentration above maximum measurable valuewarning
FiSEV OverrangeInspired sevoflurane measurement exceeds module rangeFiSEV concentration above maximum measurable valuewarning
Flowmeter Comm StopCommunication loss between flowmeter and CPU boardLost communication with CPU board for 10 secondserror
Flowmeter Selftest ErrorFresh flow sensor board failed internal self-test during startupCPU communication failure or board internal error at power-onerror
Flowmeter Selftest ErrorFlowmeter self-test failed during startup. CPU, Flash memory, or watchdog timer error, or communication failure with sensor boardPower-oncritical
Flowmeter Voltage ErrorElectronic flowmeter voltage (DVCC, AVDD or VC) is out of specificationPower-on or runtimecritical
Flowmeter Voltage ErrorVoltage error on digital voltage supply channels (DVCC, AVDD or VC)Power supply or communication failure to fresh flow sensor boarderror
Flow Sensor FailureVentilator flow sensor output is out of normal operating rangeRuntimeinfo
Hal Accuracy UnspecifiedHalothane measurement outside module accuracy rangeMeasured halothane concentration exceeds specificationwarning
Heating Module FailureHeating element temperature out of specificationHeating resistance temperature abnormal for extended perioderror
Incompatible AG Software VersionAG module software version too oldAG module software version < 1.7.3critical
Insp Valve FailureInspiratory valve malfunctionInspiratory valve voltage error or inspiratory valve flow errorwarning
Insp Valve FailureInspiratory valve has voltage error or flow sensor errorRuntimewarning
Internal Air Flow FailureCommunication failure between flowmeter CPU and Air flow sensorCommunication failure between flowmeter CPU and Air sensor for 3 seconds continuouslyinfo
Internal N2O Flow FailureCommunication failure between flowmeter CPU and N2O flow sensorCommunication failure between flowmeter CPU and N2O sensor for 3 seconds continuouslyinfo
Internal O2 Flow FailureCommunication failure between flowmeter CPU and O2 flow sensorCommunication failure between flowmeter CPU and O2 sensor for 3 seconds continuouslyinfo
IP Address ConflictDevice IP address duplicated on local networkIP address matches another device on same networkwarning
IP Address ConflictIP address is the same as another machine on the networkPower-on or runtimewarning
Iso Accuracy UnspecifiedIsoflurane measurement outside module accuracy rangeMeasured isoflurane concentration exceeds specificationwarning
Keyboard Self Test ErrorKeyboard malfunction detectedKeyboard self-test timeout during startuperror
LeakGas delivery system leak detected in one or more assembliesContinuous pressure drop detected during system test or operationerror
Low Battery Voltage!Battery voltage critically lowBattery voltage less than 10.6V for 5 secondserror
[Manual Vent.]Auto/Manual ventilation switch is in wrong positionCalibration system detects Auto/Manual switch set to bag position instead of mechanical ventilationerror
Mixed AgentTwo halogenated anesthetic agents detected with low combined MACMAC < 3 with two halogenated agents monitoredwarning
Mixed Agent and MAC≥3Two halogenated anesthetic agents with high combined MACMAC >= 3 with two halogenated agents monitorederror
N2O Accuracy UnspecifiedN2O measurement outside module accuracy rangeMeasured N2O concentration exceeds specificationwarning
N2O Flow Too HighNitrous oxide flow exceeds maximum thresholdN2O flow greater than 15 L/min for 1 secondinfo
NO Fresh GasFresh gas flow below minimum thresholdFresh gas flow less than 50 mL/minwarning
NO Fresh GasFresh gas flow insufficient for ventilationFresh gas flow less than 50 mL/minwarning
O2 Accuracy UnspecifiedO2 measurement outside module accuracy rangeMeasured O2 concentration exceeds specificationwarning
O2 Flow Too HighOxygen flow exceeds maximum thresholdO2 flow greater than 25 L/min for 1 secondinfo
O2-N2O Ratio ErrorOxygen-nitrous oxide ratio out of specificationN2O > 0.5 L/min AND N2O > 4x O2 flow for 1.6 secondserror
O2 Sensor DisconnectedOxygen sensor not connectedO2 sensor disconnected from cableinfo
O2 Sensor ErrorO2 sensor error detectedO2 sensor malfunctionerror
O2 Supply FailureO2 supply pressure to machine is below safe operating minimumO2 supply pressure falls below pressure switch threshold (~0.25 MPa / 2.5 bar)error
O2 Supply FailureOxygen supply failure or insufficient pressureO2 pressure switch detects inadequate oxygen supplyerror
Patient Circuit LeakBreathing circuit or patient connection leak detectedPressure < 2 cmH2O for 30 seconds during mechanical ventilation OR patient not connectedwarning
PEEP Safety Valve FailurePEEP safety valve electrical faultPEEP safety valve voltage errorwarning
PEEP Valve FailurePositive end-expiratory pressure valve malfunctionPEEP valve voltage error or PEEP valve pressure errorwarning
Pinsp Not AchievedInspiratory pressure not achievedIn pressure mode, Pinsp < 2/3 of setting (if >9 cmH2O) or < setting-3 cmH2O (if <9 cmH2O) for 6 cyclesinfo
Power Board High TempPower board temperature exceeds 95 degrees CelsiusRuntimecritical
Power Supply Voltage ErrorPower board output voltage (3.3V, 5V, or 12V) is out of specificationPower-on or runtimecritical
Power Supply Voltage Error (Runtime)3.3V, 5V, or 12V supply voltage out of specification during operationPower board detects output voltage out of rangeerror
Power System Comm StopCommunication loss between power board and CPU boardLost communication with CPU board for 10 secondserror
Rate OverrangePump rate measurement exceeds module rangeMeasured pump rate above maximum specificationwarning
Replace O2 sensorOxygen sensor reading critically lowO2 value less than 5%warning
RT Clock FailureReal-time clock chip malfunctionRT chip malfunction detectederror
RT Clock Needs BatteryReal-time clock backup battery is missing or depletedPower-on or runtimecritical
Sev Accuracy UnspecifiedSevoflurane measurement outside module accuracy rangeMeasured sevoflurane concentration exceeds specificationwarning
System going DOWN, Battery depleted!Battery voltage critically low; system shutting downBattery voltage less than 10.2V at power-oncritical
System going DOWN, Battery depleted!System shutdown imminent due to battery depletionBattery voltage less than 10.2Vcritical
Ventilator Comm StopCommunication loss between ventilator board and CPU boardLost communication with CPU board for 10 secondserror
Ventilator Initialization ErrorVentilator control board did not receive parameter settings at power-onCPU to ventilator communication failure during initializationerror
Ventilator Initialization ErrorCPU board cannot send parameter settings to ventilator control board at power-onPower-oncritical
Ventilator Initialization ErrorVentilator parameter initialization timeoutCPU board cannot send parameter settings to ventilator board after power onerror
Ventilator Selftest ErrorVentilator control board failed internal self-test during startupCPU communication failure or board internal error at power-onerror
Ventilator Selftest ErrorVentilator self-test failed. CPU, timer, RAM, watchdog, EEPROM or ADC error, or communication loss with ventilator control boardPower-oncritical
Ventilator Voltage ErrorVentilator control module power supply voltage (5V or 12V) is out of specificationPower-on or runtimecritical
Ventilator Voltage Error5V or 12V supply voltage out of specification to ventilator control modulePower board output voltage failureerror
Vt Not AchievedTidal volume not achievedTidal volume < TV setting for 6 cycles for >20% or 50 mL (whichever greater)info

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.