Mindray A5: troubleshooting and error codes
The ServiceForge knowledge base holds 301 documented symptoms for the Mindray A5, with 603 ranked causes, 399 diagnostic checks with test points and expected values, and 178 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Pressure calibration: System failure
Troubleshooting for pressure calibration process that fails with error code display.
Ranked causes
- 1. Drive gas pressure too low (error 00 00 00 02) (common)
- 2. Auto/Manual switch in Manual position (error 00 00 00 04) (common)
- 3. Airway pressure sensor zero point error (error 00 00 00 08) (occasional)
- 4. PEEP pressure sensor zero point error (error 00 00 00 10) (occasional)
- 5. Airway pressure sensor measurement range error (error 00 00 00 20) (occasional)
- 6. PEEP pressure sensor measurement range error (error 00 00 00 40) (occasional)
- 7. Airway pressure sensor calibration data not unidirectional (error 00 00 00 80) (rare)
- 8. PEEP pressure sensor calibration data not unidirectional (error 00 00 01 00) (rare)
- 9. Airway pressure sensor resolution error (error 00 00 02 00) (rare)
- 10. PEEP pressure sensor resolution error (error 00 00 04 00) (rare)
- 11. Valve output pressure too low (error 00 00 08 00) (occasional)
- 12. Flow change not unidirectional (error 00 00 10 00) (rare)
- 13. Communication with calibration device interrupted (error 00 00 20 00) (occasional)
- 14. System fails to write EEPROM (error 00 00 40 00) (rare)
- 15. ACGO switch in ON position (error 00 00 80 00) (common)
- 16. Valve resolution insufficient (error 00 01 00 00) (rare)
- 17. Communication error (error FF FF FF FF) (occasional)
17 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic circuit leaks patient safety
Loss of pressure or gas leakage from breathing system connections, seals, or components; may affect ventilation delivery.
Ranked causes
- 1. Damaged seals on lifting device connections (common)
- 2. Damaged manual bag (common)
- 3. Damaged breathing tube (common)
- 4. Damaged Paw pressure gauge seal (common)
- 5. Damaged water collection cup seal (common)
- 6. Damaged O2 sensor or plug seal (common)
- 7. Damaged check valve dome seal (common)
- 8. Damaged bag arm seal (common)
- 9. Damaged prepak assembly seal (common)
- 10. Defective APL valve seals (common)
- 11. Defective support tube seals on median plate (common)
- 12. Defective Auto/Manual ventilation switch seals (common)
- 13. Damaged CGO assembly or connectors (occasional)
- 14. Leaky flow sensor pressure sampling pipeline (occasional)
Turn off system switch
- Test point
- System power
- Expected
- Off
14 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system leaks patient safety
Air leaks from breathing circuit components including CO2 absorber canister, seals, bellows, or valve assembly
Ranked causes
- 2. Sealing piece for absorbent canister assembly is damaged (occasional)
- 3. Seal for bag arm is damaged (occasional)
- 4. Water collection cup is loose (common)
- 5. Seal for water collection cup assembly is damaged (occasional)
- 6. Seal for circuit adapter assembly is damaged (occasional)
- 8. Bellows sealing cushion falls off or is damaged (occasional)
- 9. Valve cover of breathing valve assembly not installed properly (occasional)
- 10. Seal for valve cover of breathing valve assembly is damaged (occasional)
- 12. Seal for O2 sensor or sensor plug is damaged (occasional)
- 13. Breathing tube connecting patient is damaged (occasional)
- 14. Bellows is damaged (occasional)
- 15. Sealing connection of other breathing system parts is damaged (occasional)
- 16. Condensate valve not installed properly or internal seal is damaged (occasional)
16 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System pressure and PEEP sensors
Automated pressure calibration procedure for airway and PEEP pressure sensors during anesthesia machine service.
Ranked causes
- 1. Drive gas pressure too low (common)
- 2. Airway or PEEP pressure sensor zero point error (common)
- 3. Pressure sensor sampling line disconnected or improperly connected (common)
- 4. PEEP valve maximum pressure insufficient (occasional)
- 5. Pneumatic circuit connection error between calibration device and ventilator control board (occasional)
- 6. Communication connection error between calibration device and anesthesia machine (occasional)
- 7. Calibration device settings incorrect (occasional)
- 8. Calibration data not correct (occasional)
Verify drive gas pressure
- Test point
- O2 pressure gauge
- Expected
- At least 200 kPa
2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Seal damage in breathing system components patient safety
Seals damaged or deteriorated in O2 sensor, check valve dome, bag arm, prepak assembly, APL valve, support tube, or Auto/Manual switch.
Ranked causes
- 1. O2 sensor seal damaged (common)
- 2. Check valve dome seal damaged (common)
- 3. Bag arm seal damaged (common)
- 4. Prepak assembly seal damaged (common)
- 5. APL valve seal damaged (occasional)
- 6. Support tube on median plate seal damaged (occasional)
- 7. Auto/Manual ventilation switch seal damaged (occasional)
Turn off system switch
- Test point
- Power switch
- Expected
- System powered off
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthesia machine cannot be started caution
System does not power up; AC and battery indicator lamps not lit; screen not lit; machine cannot ventilate
Ranked causes
- 1. AC power supply not connected or battery capacity insufficient (common)
- 2. AC input breaker tripped with insufficient battery (common)
- 3. Power board fuse burned with insufficient battery (occasional)
- 4. System switch damaged or cable disconnected (occasional)
- 5. Power board hardware failure: no voltage output (15.2V, 3.3V, 5V, 12V) (rare)
- 6. Power board software code error: no voltage output (15.2V, 3.3V, 5V, 12V) (occasional)
Verify AC power supply connection
- Test point
- AC power connector
- Expected
- AC cable firmly connected to wall outlet and machine connector
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow waveform displayed irregularly patient safety
Irregular or unstable flow measurement display on monitor screen
Ranked causes
- 1. Flow sensor pressure sampling pipeline leaks (common)
- 2. Flow sensor assembly not installed properly (common)
- 3. Water accumulated inside flow sensor assembly (occasional)
- 4. Flow sensor membrane distorted, dirty or resistance changed (occasional)
- 5. Flow sensor assembly damaged (rare)
- 6. Pressure sensor on fresh flow sensor board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate tidal volume patient safety
Tidal volume readings are incorrect or inconsistent; may show TVe and TVi discrepancies or inaccurate flow measurements.
Ranked causes
- 3. Significant breathing system leaks with low fresh gas flow (occasional)
- 4. Water buildup in flow sensor (occasional)
- 5. Flow sensor membrane distorted, dirty, or resistance changed; zero drift occurs (occasional)
- 8. Inlet gas flow regulator defective in expiratory valve assembly (occasional)
- 9. Plimit set too low, causing premature expiration (occasional)
- 10. TVe and TVi mismatch due to sensor measurement error (occasional)
Turn off flow regulators
- Test point
- Flow regulators
- Expected
- Off
7 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure - Ventilation sound heard then failure patient safety
Ventilation sound occurs initially but calibration fails shortly after Start pressed
Ranked causes
- 1. Sensor sampling line not connected or connected in reverse order (common)
- 2. Inspiratory valve maximum flow less than 90 L/min (occasional)
- 3. Pneumatic circuit connection error between machine and calibration device (occasional)
- 4. Communication connection fault between machine and calibration device (occasional)
- 5. Calibration device settings error (occasional)
Verify inspiratory and expiratory flow sensor sampling lines are connected correctly
- Expected
- All three sensor lines properly connected in correct order
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration failure: Ventilation sound heard, then failure prompt
After Begin is selected, ventilation sound is heard but '[Calibration Failure! Please try again.]' prompt appears very soon
Ranked causes
- 1. Sampling line not connected or improperly connected on airway or PEEP pressure sensor (common)
- 2. PEEP valve maximum pressure less than 95 cmH2O (occasional)
- 3. Pneumatic circuit connection error between calibration device and ventilator control board (occasional)
- 4. Communication connection error between calibration device and anesthesia machine (occasional)
- 5. Calibration device settings incorrect (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow sensor patient safety
Flow sensor calibration fails or measurement deviations exceed 9% compared to setting value
Ranked causes
- 1. Flow sensor aging or environmental drift (common)
- 2. Flow sensor replaced (common)
- 3. O2 sensor replaced or recalibrated (common)
- 4. Breathing system leak (occasional)
- 5. Drive gas pressure out of specification (occasional)
Perform breathing system leak test
- Test point
- Breathing circuit
- Expected
- Leak test passed in mechanical ventilation mode
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow sensors patient safety
Flow sensor calibration for breathing system to correct tidal volume measurement deviation or after component replacement
Ranked causes
- 1. Tidal volume measurement deviation exceeds 9% after prolonged use (common)
- 2. Flow sensor in patient circuit replaced (occasional)
- 3. Expiratory valve assembly replaced (occasional)
- 4. Ventilator control board replaced (occasional)
- 5. Flow measurement deviation exceeds 5% of reading or 1 L/min (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor assembly problems patient safety
Flow waveform displays irregularly or flow measurement is incorrect due to flow sensor assembly issues
Ranked causes
- 1. Flow sensor assembly not installed properly (common)
- 2. Water built up inside flow sensor assembly (occasional)
- 3. Flow sensor membrane distorted, dirty, or internal resistance changes (occasional)
- 4. Flow sensor is damaged (occasional)
- 5. Pressure sensor on fresh flow sensor board is defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas supply tube leak caution
Gas escaping from supply tube or connection points
Ranked causes
- 1. Gas supply tube is damaged (common)
- 2. Connection seal is damaged (common)
- 3. Quick plug-in connector is leaking or damaged (occasional)
- 4. Pipeline gas supply inlet assembly check valve or seal is damaged (occasional)
- 5. Drive gas pipeline is leaking (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low-pressure pneumatic circuit system leak patient safety
Breathing system leaks detected during mechanical or manual ventilation modes; pressure loss during test
Ranked causes
- 1. Bellows housing or sealing cushion leaky (common)
- 2. Manual ventilation mode bag or APL valve defective (occasional)
- 3. Expiratory valve assembly or drive gas pipeline leak (occasional)
- 4. Fresh gas or ACGO pipeline leak (occasional)
- 5. Circuit adapter or patient circuit damaged (occasional)
Leak test of breathing system in mechanical ventilation mode
- Test point
- Breathing system circuit
- Expected
- No detectable pressure loss during test
- Tools
- As specified in section 3.7.2.2
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 supply inlet assembly leaks patient safety
Pressure falls dramatically and continuously after O2 pipeline gas supply is turned off, indicating gas leakage
Ranked causes
- 1. O2 flush button assembly seal damaged or leaking (common)
- 2. System switch assembly seal damaged (common)
- 3. Vaporizer improperly installed or seal damaged (occasional)
- 4. Flowmeter or ORC assembly leaks (occasional)
- 5. Flow regulator or restrictor leaks (occasional)
Observe pressure gauge reading trend after gas supply turned off
- Test point
- Test pressure gauge (downstream of inlet assembly)
- Expected
- Pressure should hold steady or fall slowly; not fall dramatically and continuously
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power system communication stopped caution
Lost communication with CPU board for 10 seconds. Power board cannot communicate with main processor.
Ranked causes
- 1. Communication cable failure or disconnection (common)
- 2. Power board software incorrect or needs update (occasional)
- 3. Power board processor reset required (occasional)
- 4. Power board failure (occasional)
- 5. CPU board failure (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthesia machine cannot start patient safety
System fails to start after power switch activation.
Ranked causes
- 1. System switch damaged (rare)
- 2. Cable disconnected from system switch (common)
- 3. Power board hardware circuit failure (occasional)
- 4. Power board software code error (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not detected caution
Battery is not detected by the system. Battery voltage may be out of specification or cable connection is loose.
Ranked causes
- 1. Battery voltage out of specification (common)
- 2. Cable not connected correctly (common)
- 3. Defective battery (occasional)
- 4. Power board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system accessories damaged or leaky patient safety
Manual bag, breathing tubes, Paw pressure gauge, or water collection cup show signs of damage or leakage
Ranked causes
- 1. Manual bag damaged (occasional)
- 2. Breathing tube damaged (occasional)
- 3. Paw pressure gauge seal damaged (occasional)
- 4. Water collection cup seal damaged (occasional)
Visually inspect manual bag for damage
- Test point
- Manual bag surface
- Expected
- No visible tears, cracks, or defects
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Breathing system leakage detected patient safety
Pressure reading falls below expected level during leak test; indicates faulty seals or connections in breathing system
Ranked causes
- 1. Breathing system leak in mechanical or manual ventilation mode (common)
- 2. Bellows housing or manual ventilation switch leak (occasional)
- 3. Circuit adapter pipeline leak (occasional)
- 4. Absorb canister assembly leakage (occasional)
Perform leak test in manual ventilation mode
- Test point
- Paw pressure gauge
- Expected
- Pressure maintained at 30 cmH2O
- Tools
- O2 flow regulator, breathing tubes, Y piece
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 100% O2 patient safety
System calibration for oxygen sensor using 100% oxygen gas; failure indicated by fail code display or calibration not completing within ~3 minutes
Ranked causes
- 1. O2 sensor not connected (common)
- 2. O2 supply pressure insufficient (< 200 kPa) (common)
- 3. 21% O2 calibration not performed first (occasional)
- 4. O2 sensor count value out of normal range (450-2700) (occasional)
Verify O2 sensor connection to cable
- Test point
- O2 sensor connector
- Expected
- Secure connection
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow sensor (Service) patient safety
Service-level flow sensor and inspiratory valve calibration using standard flow measurement device
Ranked causes
- 1. Ventilator control board replaced (common)
- 2. Drive gas assembly replaced (common)
- 3. Solenoid valve assembly replaced (common)
- 4. Flow sensor measurement deviation exceeds specification (occasional)
Verify no sensor or valve technical alarms
- Test point
- Alarm display
- Expected
- No active sensor or valve alarms
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure and Flow Zeroing Failure patient safety
Anesthesia machine unable to zero pressure and flow sensors during calibration; zeroing process fails to complete successfully.
Ranked causes
- 1. Sensor zero point out of range (occasional)
- 2. Sampling line occlusion or three-way valve mechanical fault (occasional)
- 3. Ventilator control board faulty (occasional)
- 4. Solenoid valve assembly faulty (occasional)
Verify sensor zero point is within acceptable range
- Test point
- Pressure and flow sensors
- Expected
- Zero point normal per section 3.7.3
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 absorber canister not locked caution
CO2 absorber canister is not properly mounted or not detected
Ranked causes
- 1. Canister not properly seated (common)
- 2. Canister detection cable fault (occasional)
- 3. Ventilator control board failure (occasional)
- 4. Canister switch failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code [00 00 00 08] displayed - Inspiratory sensor zero point error patient safety
Inspiratory flow sensor zero reference point has shifted or is invalid
Ranked causes
- 1. Fresh gas not turned off preventing zero reference measurement (common)
- 2. Inspiratory valve has not fully closed blocking zero reference (occasional)
- 3. Zero point calibration is out of specification (occasional)
- 4. Inspiratory flow sensor or ventilator control board failure (occasional)
Ensure fresh gas flow is turned completely off
- Expected
- Fresh gas supply off
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Flow sensor zero drift or irregular waveform display patient safety
Flow measurement is inaccurate or waveform appears distorted on display; pressure sensor shows drift
Ranked causes
- 1. Flow sensor membrane distorted, dirty, or internal resistance changed (common)
- 2. Flow sensor assembly damaged (occasional)
- 3. Pressure sensor on fresh flow sensor board defective (occasional)
- 4. Flow sensor pressure sampling pipeline leak (common)
Perform leak test of flow sensor pressure sampling pipeline (expiratory and inspiratory lines)
- Test point
- Flow sensor pressure sampling pipeline test fixture connectors
- Expected
- Pressure does not fall more than 5 cmH2O within 15 s at 70-90 cmH2O, within 15 s
- Tools
- Anesthesia machine calibration device, flow sensor pressure sampling pipeline test fixture, circuit adapter test fixture, injector, Φ6 silicone tube, Y piece
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas leak in pneumatic circuit patient safety
Escaping gas detected in supply lines, connectors, or gas inlet assembly; audible hiss or pressure loss
Ranked causes
- 1. Gas supply tube damaged or seal at connection damaged (common)
- 2. Quick plug-in connector leaks (common)
- 3. Pipeline gas supply inlet assembly leaks (occasional)
- 4. Drive gas pipeline leaks (occasional)
Inspect negative pressure ball for leaks before use
- Test point
- Front sealing cover and gas inlet
- Expected
- Ball does not visibly inflate for at least 30 seconds when finger blocks inlet
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heater ineffective or not heating patient safety
Unit does not reach or maintain set temperature; heater output insufficient; no heat felt at outlet
Ranked causes
- 1. Heater driver and control circuit on power board damaged (occasional)
- 2. Heater element damaged (occasional)
- 3. Heater internal sensor ineffective (occasional)
- 4. Heater cable disconnected or loose (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory valve failure caution
Inspiratory valve voltage error or inspiratory valve flow error. Inspiratory flow sensor out of specification or valve assembly connection issue.
Ranked causes
- 1. Inspiratory flow sensor out of specification (common)
- 2. Voltage issue at test point or valve assembly connection (occasional)
- 3. Valve assembly failure (occasional)
- 4. Power board failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak in gas supply system patient safety
Gas pressure not maintained or leakage detected in pipeline, connectors, or inlet assembly
Ranked causes
- 1. Gas supply tube is damaged or seal at connection is damaged (common)
- 2. Quick plug-in connector leaks (common)
- 3. Pipeline gas supply inlet assembly leaks (occasional)
- 4. Drive gas pipeline leaks (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Machine cannot start
Machine fails to start after power is applied
Ranked causes
- 1. System switch damaged (occasional)
- 2. Cable to system switch disconnected (common)
- 3. Power board hardware failure preventing voltage output (occasional)
- 4. Power board software error preventing voltage output (rare)
Verify system switch cable connection
- Test point
- System switch connector
- Expected
- Cable properly seated
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PEEP valve failure caution
PEEP valve voltage error or PEEP valve pressure error. Pressure sensor on PEEP circuit out of specification or valve assembly connection issue.
Ranked causes
- 1. Pressure sensor out of specification (common)
- 2. Voltage issue at test point or valve assembly connection (occasional)
- 3. Expiratory valve assembly failure (occasional)
- 4. Power board failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen cannot be lit
Display screen remains off when machine powered on
Ranked causes
- 1. Cable to inverter disconnected (common)
- 2. Inverter damaged (occasional)
- 3. Power board hardware failure causing improper output (occasional)
- 4. Power board software failure causing improper output (rare)
Verify inverter cable connection
- Test point
- Inverter connector
- Expected
- Cable properly seated
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen does not light caution
Display not powered; no illumination or visual output when machine switched on
Ranked causes
- 1. Inverter cable disconnected or inverter damaged (occasional)
- 2. Power board hardware failure: improper output (occasional)
- 3. Power board software failure: improper output (occasional)
- 4. Screen power supply fuse burned: no 3.3V output (occasional)
Verify screen power supply 3.3V output
- Test point
- Screen power supply output
- Expected
- 3.3V present at connector
- Tools
- Multimeter
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-way solenoid valve failure caution
Solenoid valve electrical signal control status error detected
Ranked causes
- 1. Solenoid valve connection fault (common)
- 2. Solenoid valve assembly failure (common)
- 3. Ventilator control board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aux Control Module communication stopped caution
Lost communication with CPU board for 10 seconds
Ranked causes
- 1. Communication cable defect or disconnection (common)
- 2. Ventilator Control Module Board failure (occasional)
- 3. CPU Board failure (occasional)
Restart the machine
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Aux control module selftest error on startup caution
Machine displays aux control module selftest error at power-on. CPU, Flash, watchdog timer error or communication failure between CPU and ventilator control board.
Ranked causes
- 1. Communication cable failure or disconnection (common)
- 2. Ventilator control board failure (occasional)
- 3. CPU board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery in use - AC power lost patient safety
Unit is operating on battery power. AC mains supply has been disconnected or failed.
Ranked causes
- 1. AC power supply disconnected (common)
- 2. AC supply voltage out of specification (occasional)
- 3. Power board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure - No ventilation sound after Begin selected patient safety
System does not produce ventilation sound and fails immediately after Begin pressed
Ranked causes
- 1. Auto/Manual ventilation switch in bag position instead of mechanical ventilation (common)
- 2. Drive gas (O2) pressure below 200 kPa specification (common)
- 3. Zero point error in inspiratory or expiratory flow sensor (occasional)
Verify Auto/Manual ventilation switch is in mechanical ventilation position, not bag
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration failure with no ventilation sound patient safety
After Begin is selected, no ventilation sound is heard and Calibration Failure message is displayed
Ranked causes
- 1. Auto/Manual ventilation switch set to bag position (common)
- 2. Drive gas pressure below 200 kPa (common)
- 3. Zero point error in inspiratory/expiratory flow sensor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure with ventilation sound heard patient safety
After Start is selected, ventilation sound heard then Calibration Failure message displayed
Ranked causes
- 1. Sensor sampling line not connected or reversed (common)
- 2. Inspiratory valve maximum flow below 90 L/min (occasional)
- 3. Pneumatic circuit, communication, or configuration error (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure: Code [00 01 00 00]
Insufficient valve resolution during calibration process
Ranked causes
- 1. Insufficient gas supply for calibration (common)
- 2. Calibration device malfunction (occasional)
- 3. PEEP valve defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure: communication error patient safety
Pressure calibration error [FF FF FF FF] displayed indicating communication error occurs.
Ranked causes
- 1. Transient communication fault (occasional)
- 2. Communication cable damage or poor connection (common)
- 3. Communication board failure (rare)
Restart anesthesia machine
- Test point
- Machine power and startup
- Expected
- Machine restarts normally and enters standby
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure: No ventilation sound, immediate failure message patient safety
After Begin selected, no ventilation heard; failure message 'Calibration Failure! Please try again.' displays immediately
Ranked causes
- 1. Zero point error in airway pressure gauge or PEEP pressure sensor (common)
- 2. Sampling line not connected or improperly connected (common)
- 3. PEEP valve maximum pressure less than 95 cmH2O (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure: unidirectional flow not achieved patient safety
Pressure calibration error [00 02 00 00] displayed indicating flow change is not unidirectional.
Ranked causes
- 1. Tubes connected incorrectly (common)
- 2. Insufficient gas supply (occasional)
- 3. Calibration device malfunction (occasional)
Verify all pneumatic connections are correct
- Test point
- Pressure sampling lines and four-way device
- Expected
- Tubes connected per Figures 4-52 and 4-61; P1 to monitor board, P2 to PEEP, tube #72 connected, tube #9 unconnected
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration failure: valve resolution insufficient patient safety
Pressure calibration error [00 01 00 00] displayed indicating valve resolution is not adequate.
Ranked causes
- 1. Insufficient gas supply (occasional)
- 2. Calibration device malfunction (occasional)
- 3. Inspiratory valve defective (common)
Verify gas supply pressure and flow
- Test point
- Gas supply inlet
- Expected
- Adequate supply for calibration process
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration failure: Ventilation sound heard, delayed failure message patient safety
After Begin selected, ventilation sound is heard; failure message appears after several seconds: 'Calibration Failure! Please try again.'
Ranked causes
- 1. Pneumatic circuit connection error between calibration device and ventilator control board (common)
- 2. Communication connection error between calibration device and machine (common)
- 3. Calibration device settings error (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow sensor service calibration patient safety
Flow sensor and inspiratory valve calibration after equipment service or when measured flow deviation exceeds 5% of reading or 1 L/min
Ranked causes
- 1. Expiratory valve assembly replaced (occasional)
- 2. Ventilator control board replaced (occasional)
- 3. Flow measurement device reading deviation exceeds 5% or 1 L/min (occasional)
Verify tubes are not leaking when connected
- Test point
- Calibration tube connections
- Expected
- No leaks detected
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow sensors (service level) patient safety
Service-level calibration of flow sensors and inspiratory valve after component replacement or when measured deviation exceeds threshold.
Ranked causes
- 1. Expiratory valve assembly replaced (occasional)
- 2. Ventilator control board replaced (occasional)
- 3. Ventilator flow sensor deviation exceeds 5% of reading or 1 L/min (occasional)
Compare ventilator flow sensor reading with calibrated measurement device
- Test point
- Flow measurement comparison
- Expected
- Deviation within 5% of reading or 1 L/min (whichever greater), ±5% or ±1 L/min
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: O2 sensor patient safety
O2 sensor calibration fails or produces out-of-range readings during 21% or 100% calibration steps
Ranked causes
- 2. O2 supply pressure insufficient (common)
- 3. 21% calibration not completed before 100% calibration (common)
- 4. O2 sensor defective or aged (occasional)
Verify O2 sensor cable connection
- Test point
- O2 sensor connector
- Expected
- Alarm [O2 Sensor Unconnected] not displayed
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: O2 sensor calibration patient safety
O2 sensor accuracy calibration at 21% (air) and 100% O2 when measurement deviation exceeds 3% or after sensor replacement
Ranked causes
- 1. O2 sensor measurement deviation exceeds 3% (occasional)
- 2. O2 sensor replaced (occasional)
- 3. Ventilator control board replaced (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure sensor service calibration patient safety
Pressure sensor and PEEP valve calibration after equipment service or when pressure measurement deviation exceeds 5% of reading or 2 cmH2O
Ranked causes
- 1. Ventilator control board replaced (occasional)
- 2. Expiratory valve assembly replaced (occasional)
- 3. Pressure sensor measurement deviation exceeds 5% or 2 cmH2O (occasional)
Calibration: Pressure sensors and PEEP valve patient safety
Airway pressure sensor, PEEP pressure sensor, and PEEP valve require calibration after replacing ventilator control board, drive gas assembly, or solenoid valve assembly, or when measured values deviate significantly from standard measurement device.
Ranked causes
- 2. Drive gas assembly replaced (common)
- 3. Solenoid valve assembly replaced (common)
- 4. Pressure sensor deviation from standard measurement device (occasional)
Verify anesthesia machine is in standby mode
- Test point
- Machine mode
- Expected
- Standby mode
10 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure sensors and PEEP valve (service level) patient safety
Service-level calibration of pressure sensors and PEEP valve after component replacement or measurement deviation exceeds acceptable threshold.
Ranked causes
- 1. Ventilator control board replaced (occasional)
- 2. Expiratory valve assembly replaced (occasional)
- 3. Pressure sensor deviation exceeds 5% of reading or 2 cmH2O (occasional)
Compare machine pressure sensor reading with standard pressure gauge
- Test point
- Pressure measurement comparison
- Expected
- Deviation within 5% of reading or 2 cmH2O (whichever greater), ±5% or ±2 cmH2O
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: System gas flow rates patient safety
O2, N2O, and AIR needle valve flow rates require calibration to meet specified ranges during anesthesia delivery
Ranked causes
- 1. O2 needle valve requires flow rate adjustment (common)
- 2. N2O resistor requires adjustment for correct flow rate (common)
- 3. AIR and N2O resistors on pneumatic block require adjustment (common)
Measure O2 flow rate when N2O needle valve at maximum
- Test point
- O2 flow outlet
- Expected
- 0.25 to 0.3 L/min, 0.25–0.3 L/min
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error code [00 00 00 10] displayed - Expiratory sensor zero point error patient safety
Expiratory flow sensor zero reference point has shifted or is invalid
Ranked causes
- 1. Fresh gas not turned off preventing zero reference measurement (common)
- 2. Inspiratory valve has not fully closed blocking zero reference (occasional)
- 3. Zero point calibration out of specification (occasional)
Verify fresh gas supply is turned off
- Expected
- Fresh gas flow zero
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Flow sensor failure caution
Ventilator flow sensor output is out of normal operating range. Sensor zero point or measurement is out of specification.
Ranked causes
- 1. Flow sensor zero point out of specification (common)
- 2. Flow sensor measurement out of specification (common)
- 3. Ventilator control board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor sampling line connection error
Flow sensor has two sampling lines; errors include backward connection, one line not connected, or both lines not connected
Ranked causes
- 1. Flow sensor sampling lines connected backward (occasional)
- 2. Flow sensor sampling line broken or not connected (occasional)
- 3. Flow sensor low-pressure sampling line not connected (occasional)
Connect anesthesia machine tubes using constant-flow connection method
- Test point
- Pneumatic circuit
- Expected
- Tubes properly connected per constant-flow method
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flowmeter selftest error on startup caution
Machine displays flowmeter selftest error at power-on. CPU, Flash, watchdog timer error or communication failure between CPU and fresh flow sensor board.
Ranked causes
- 1. Communication cable failure or disconnection (common)
- 2. Fresh flow sensor board failure (occasional)
- 3. CPU board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Heater ineffective
- Low battery voltage
- Network connection fails
- O2 concentration measurement fails or shows deviations
- PEEP safety valve failure
- Power board voltage error
- System shutting down - battery depleted
- Touch panel not responding to input
- Unit does not power on
- Unit does not power on, indicators and screen dark
- Vaporizer manifold assembly leaks
- Ventilator communication stopped
- Ventilator initialization error on startup
- Ventilator selftest error on startup
- Ventilator voltage error
- Alarm messages without sound
- Bellows housing or manual ventilation switch leak
- Breathing system leak in manual ventilation mode
- Calibrate O2 sensor
- Calibration failure after 15 minutes
- Calibration failure after 15 minutes of operation
- Calibration failure: ACGO switch in ON position
- Calibration failure: communication interrupted
- Calibration failure: EEPROM write failed
- Calibration failure: No ventilation sound, immediate failure prompt
- Calibration failure: output flow too low
- Calibration: Electronic flowmeter zeroing
- Calibration: Flow sensor (user)
- Calibration: Flow sensor user calibration
- Calibration: Flow sensors (user level)
- Calibration: Pressure and flow zeroing
- CGO assembly leak or connection failure
- CGO assembly seal or connector damage
- Check flow sensors reverse flow detected
- CPU to infrared board communication fails
- Drive gas pressure low
- Error code [00 00 00 02] displayed - Drive gas pressure too low
- Error code [00 00 00 20] displayed - Internal sensor zero point error
- Error code [00 00 00 40] displayed - Inspiratory sensor measurement range error
- Error code [00 00 00 80] displayed - Expiratory sensor measurement range error
- Error code [00 00 01 00] displayed - Internal sensor measurement range error
- Error code [00 00 02 00] displayed - Inspiratory sensor non-unidirectional data
- Error code [00 00 04 00] displayed - Expiratory sensor non-unidirectional data
- Error code [00 00 08 00] displayed - Internal sensor non-unidirectional data
- Error code [00 00 10 00] displayed - Inspiratory sensor resolution error
- Error code [00 00 20 00] displayed - Expiratory sensor resolution error
- Error code [00 00 40 00] displayed - Internal sensor resolution error
- Fan failure
- Flow sensor assembly not installed properly
- Flow sensor measurement error
- Flowmeter assembly leaks gas
- Flowmeter voltage error on startup
- Inspiratory valve does not open adequately
- Inspiratory valve does not open correctly
- N2O flow out of range
- No Drive Gas Pressure Low alarm when pressure is low
- No gas output from valve in mechanical ventilation mode
- No O2 Supply Failure alarm when pressure is low
- O2 sensor disconnected
- O2 supply failure
- Patient circuit leak detected
- PEEP safety valve does not open or PEEP valve malfunction
- PEEP safety valve does not operate correctly
- PEEP valve does not open or close correctly
- PEEP valve opening state abnormal during Diagnostic Tests
- Pinsp not achieved
- Pipeline pressure gauge readings fluctuate greatly
- Power board overheating
- Pressure sensor measurement error
- Screen displays content but image flashes
- Screen flashes
- Screen illuminated but displays no content
- Screen lights but displays no content
- Screen lights but no content displayed
- System switch assembly leaks
- System switch assembly or connecting tube damaged
- Tidal volume not achieved
- Ventilator voltage error on startup
- 3-way Valve Failure
- Absorb canister assembly leak
- AG airway occluded
- AG calibration failure
- AG communication failure
- AG data limit error
- AG hardware malfunction
- AG initialization error
- AG module hardware error
- AG self-test failure
- AG watertrap disconnected
- AG watertrap requires replacement
- AG watertrap type mismatch
- AG zeroing failure
- Air flow out of range
- Air Flow Too High
- AIR pipeline pressure gauge damaged
- AIR pipeline pressure incorrect
- AIR supply inlet assembly pressure drop excessive
- AIR supply inlet assembly pressure out of specification
- AIR supply inlet assembly regulator pressure out of specification
- AIR supply pressure out of range
- Airway pressure gauge defective
- Airway pressure gauge inaccurate or pointer immobile
- Airway pressure gauge shows inaccurate reading or pointer cannot move
- APL valve pressure adjustment
- Aux Control Module Comm Stop
- Back pressure regulator leaks
- Back pressure regulator leaks gas
- Breathing Circuit Not Mounted
- Calibration failure: Approximately 15 minutes after start
- Calibration failure: Code [00 00 40 00]
- Calibration failure: Code [00 00 80 00]
- Calibration failure: Code [FF FF FF FF]
- Calibration failure: Delay approximately 15 minutes after start
- Calibration: 21% O2
- Calibration: AG Module
- Calibration: Cylinder yoke regulator pressure
- Calibration: Electronic flowmeter zeroing (user level)
- Calibration: O2 sensor at 100% O2
- Calibration: O2 sensor at 21% O2
- Calibration: ORC oxygen-nitrous oxide ratio control
- Calibration: ORC Oxygen/Nitrous Oxide Ratio Control
- Calibration: Pressure and flow zeroing (service level)
- Calibration: Pressure sensors and PEEP proportional valve
- Cannot exit standby mode
- Check Flow Sensors
- CO2 measurement accuracy unspecified
- Desflurane measurement accuracy unspecified
- Drive gas pipeline leak in circuit adapter
- Drive gas pressure alarm incorrect or missing
- Drive Gas Pressure Low alarm does not occur at 0.05 MPa
- Drive Gas Pressure Low alarm incorrect trigger or no alarm when pressure low
- Drive Gas Pressure Low alarm not triggered
- Drive gas pressure low alarm not triggered at expected pressure
- Drive Gas Pressure Low alarm occurs at wrong threshold
- Drive Gas Pressure Low alarm occurs when pressure is normal
- Drive gas pressure low alarm triggered at wrong pressure
- End-tidal desflurane measurement overrange
- End-tidal enflurane measurement overrange
- End-tidal halothane measurement overrange
- End-tidal isoflurane measurement overrange
- End-tidal N2O measurement overrange
- End-tidal O2 measurement overrange
- End-tidal sevoflurane measurement overrange
- Enflurane measurement accuracy unspecified
- Error code [00 00 00 04] displayed - Manual ventilation switch issue
- Exiting Standby fails
- Expired CO2 measurement overrange
- External AG module disconnected
- Flow sensor calibration data incorrect
- Flow sensor calibration failure
- Flow sensor leaks
- Flow sensor leaks gas
- Flow sensor low pressure sampling line disconnected
- Flow sensor measurement error or calibration data incorrect
- Flow sensor sampling lines broken or disconnected
- Flow sensor sampling lines connected backward
- Flow sensor zero point outside specifications
- Flowmeter assembly leaks
- Flowmeter Comm Stop
- Flowmeter Voltage Error
- Fresh gas pipeline leak in circuit adapter
- Gas delivery system leak detected
- Gas mixer leaks
- Gas mixer leaks gas
- Halothane measurement accuracy unspecified
- Heating Module Failure
- Incompatible AG module software version
- Insp Valve Failure
- Inspiratory valve opening defective or inadequate flow
- Inspired CO2 measurement overrange
- Inspired desflurane measurement overrange
- Inspired enflurane measurement overrange
- Inspired halothane measurement overrange
- Inspired isoflurane measurement overrange
- Inspired N2O measurement overrange
- Inspired O2 measurement overrange
- Inspired sevoflurane measurement overrange
- Internal Air Flow Failure
- Internal N2O Flow Failure
- Internal O2 Flow Failure
- IP address conflict
- Isoflurane measurement accuracy unspecified
- Leak in gas delivery system
- Mixed halogenated agents detected (low MAC)
- Mixed halogenated agents with high MAC
- Mixed halogenated agents with invalid reading
- N2O Flow Too High
- N2O measurement accuracy unspecified
- Needle valve leaks
- Needle valve leaks gas
- Network connection failed
- NO Fresh Gas
- O2 concentration measurement fails or has great deviations
- O2 flow out of range
- O2 Flow Too High
- O2 flush button assembly leaks
- O2 measurement accuracy unspecified
- O2 pressure switch does not trigger O2 Supply Failure alarm at 0.25 MPa
- O2 sensor calibration failure
- O2 sensor error
- O2 supply alarm incorrect or missing
- O2 Supply Failure alarm incorrect trigger or no alarm when pressure low
- O2 Supply Failure alarm not triggered
- O2 supply failure alarm not triggered at expected pressure
- O2 Supply Failure alarm occurs at wrong threshold
- O2 Supply Failure alarm occurs when pressure is normal
- O2 supply failure alarm triggered at wrong pressure
- O2 supply inlet assembly pressure out of specification
- O2 supply inlet assembly regulator pressure out of specification
- O2-N2O Ratio Error
- Oxygen Ratio Controller (ORC) damaged
- Oxygen Ratio Controller (ORC) leaks
- Patient Circuit Leak
- PEEP safety valve opening defective when drive gas pressure low
- PEEP valve does not operate correctly
- PEEP valve opening defective in mechanical ventilation
- Pipeline pressure gauge reads inaccurately or shows no reading
- Pipeline pressure gauge shows inaccurate readings or no readings
- Power supply voltage error on startup
- Pressure sensor calibration data incorrect
- Pressure sensor calibration failure
- Pressure sensor measurement error or calibration data incorrect
- Pressure sensor sampling line connected incorrectly
- Pressure sensor sampling line connection error
- Pressure sensor sampling line not connected
- Pump rate overrange
- Real-time clock failure
- Real-time clock needs battery
- Replace O2 sensor
- Screen lights with content but flashes repeatedly
- Sensor zero point out of specification
- Sevoflurane measurement accuracy unspecified
- Standby mode exit fails
- Total flowmeter assembly leaks
- Total flowmeter leaks
- Touch panel ineffective
- Ventilation stops during operation
- Ventilation stops suddenly but display and buttons work
- Ventilation stops suddenly during operation
- Ventilator Comm Stop
- Vt Not Achieved
241 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
[00 00 00 02] | Drive gas pressure too low | Gas supply pressure insufficient during calibration | error |
00 00 00 02 | Drive gas pressure too low | Pressure below threshold during calibration initialization | error |
00 00 00 02 | O2 supply pressure is low during 100% calibration process | O2 supply pressure was not sufficient during 100% calibration step | error |
00 00 00 04 | Auto/Manual switch in Manual position | Auto/Manual switch detected in Manual mode during calibration | error |
00 00 00 04 | O2 sensor is disconnected; sampled data exceeds 2900 (AD value) | Sampled AD value greater than 2900 indicates sensor open circuit or very high impedance | error |
[00 00 00 04] | Auto/Manual switch in wrong position | Auto/Manual switch positioned to Manual during calibration | warning |
[00 00 00 08] | Airway pressure sensor zero point error | Zero point calibration failed for airway pressure sensor | error |
00 00 00 08 | Airway pressure sensor zero point error | Airway pressure sensor zero point outside acceptable range | error |
00 00 00 08 | 21% calibration value is outside expected range 150~500 (AD value) | 21% O2 calibration step produced reading below 150 or above 500 | error |
[00 00 00 08] | Inspiratory flow sensor zero point error | Sensor zero point AD greater than or equal to 2000 or less than 0 | error |
00 00 00 10 | PEEP pressure sensor zero point error | PEEP pressure sensor zero point outside acceptable range | error |
[00 00 00 10] | Expiratory flow sensor zero point error | Sensor zero point AD greater than or equal to 2000 or less than 0 | error |
[00 00 00 10] | PEEP pressure sensor zero point error | Zero point calibration failed for PEEP pressure sensor | error |
00 00 00 10 | 100% calibration value is outside expected range 800~2028 (AD value) | 100% O2 calibration step produced reading below 800 or above 2028 | error |
00 00 00 20 | Airway pressure sensor measurement range error | Airway sensor reading outside acceptable measurement range | error |
00 00 00 20 | Error writing to EEPROM memory | System unable to write calibration data to EEPROM storage during calibration completion | critical |
[00 00 00 20] | Measurement range error in airway pressure sensor | Airway pressure sensor output exceeds or cannot reach expected measurement range | error |
[00 00 00 20] | Airway pressure sensor measurement range error | Sensor unable to achieve full measurement range during calibration | error |
[00 00 00 40] | Measurement range error in inspiratory flow sensor | Inspiratory flow sensor reading exceeds valid measurement range during calibration | error |
00 00 00 40 | PEEP pressure sensor measurement range error | PEEP sensor reading outside acceptable measurement range | error |
[00 00 00 40] | PEEP pressure sensor measurement range error | Sensor unable to achieve full measurement range during calibration | error |
[00 00 00 80] | Expiratory flow sensor measurement range error | Sensor AD value greater than 3900 at 90 L/min flow | error |
[00 00 00 80] | Airway pressure sensor calibration data not unidirectional | Calibration data shows non-linear or backward response during pressure sweep | error |
[00 00 00 80] | Measurement range error in expiratory flow sensor | Expiratory flow sensor reading exceeds valid measurement range during calibration | error |
00 00 00 80 | Airway pressure sensor calibration data not unidirectional | Calibration data shows non-linear or inconsistent response | error |
[00 00 01 00] | Internal flow sensor measurement range error | Sensor AD value greater than 3900 at 45 L/min flow | error |
00 00 01 00 | PEEP pressure sensor calibration data not unidirectional | Calibration data shows non-linear or inconsistent response | error |
[00 00 01 00] | PEEP pressure sensor calibration data not unidirectional | Calibration data shows non-linear or backward response during pressure sweep | error |
00 00 02 00 | Airway pressure sensor resolution error | Sensor resolution insufficient for accurate measurement | error |
[00 00 02 00] | Airway pressure sensor resolution error | Insufficient sensor resolution; difference between max and min AD <= 500 or single-point resolution < 1 cmH2O | error |
[00 00 02 00] | Inspiratory flow sensor calibration data not unidirectional | Calibration data does not show expected unidirectional behavior | error |
[00 00 04 00] | Expiratory flow sensor calibration data not unidirectional | Calibration data does not show expected unidirectional behavior | error |
[00 00 04 00] | PEEP pressure sensor resolution error | Insufficient sensor resolution; difference between max and min AD <= 500 or single-point resolution < 1 cmH2O | error |
00 00 04 00 | PEEP pressure sensor resolution error | Sensor resolution insufficient for accurate measurement | error |
00 00 08 00 | Valve output pressure too low | PEEP valve output pressure below acceptable threshold | error |
[00 00 08 00] | Valve output pressure low | PEEP valve unable to generate sufficient output pressure | error |
[00 00 08 00] | Calibration data of internal flow sensor is not unidirectional | Calibration system detects bidirectional flow or inconsistent data trend in internal sensor | error |
00 00 10 00 | Flow change not unidirectional | Flow measurement shows non-linear or inconsistent response during calibration | error |
[00 00 10 00] | Resolution error in inspiratory flow sensor | Sensor resolution does not meet calibration requirements | error |
[00 00 10 00] | Change of flow is not unidirectional | Flow measurement shows non-monotonic or bidirectional behavior during calibration | error |
[00 00 20 00] | Resolution error in expiratory flow sensor | Sensor resolution does not meet calibration requirements | error |
00 00 20 00 | Communication with calibration device interrupted | Loss of communication with external calibration device | error |
[00 00 20 00] | Communication with calibration device is interrupted | Data exchange between calibration device and machine stops or becomes unreliable | error |
00 00 40 00 | System fails to write EEPROM | Data write to non-volatile memory unsuccessful | error |
[00 00 40 00] | EEPROM write failure | During calibration process | error |
[00 00 40 00] | Resolution error in internal flow sensor | Internal flow sensor resolution insufficient or unstable during calibration | error |
[00 00 40 00] | System fails to write EEPROM | EEPROM write operation failed during calibration data save | error |
[00 00 80 00] | Output flow of valve is low | Valve output flow during calibration is less than specified | error |
00 00 80 00 | ACGO switch in ON position | ACGO (automatic calibration gas outlet) switch detected in ON mode | error |
[00 00 80 00] | ACGO switch in ON position | Auto-calibration gas outlet switch activated during service calibration | warning |
[00 01 00 00] | PEEP valve resolution insufficient | Valve control resolution inadequate for fine pressure control | error |
00 01 00 00 | Valve resolution insufficient | PEEP valve resolution inadequate for precision calibration | error |
[00 02 00 00] | Change of flow is not unidirectional | Flow changes exhibit bidirectional behavior during calibration | error |
[00 02 00 00] | Maximum valve pressure not found | Valve unable to reach specified maximum pressure during diagnostic sweep | error |
[00 04 00 00] | Minimum valve pressure not found | Valve unable to reach specified minimum pressure during diagnostic sweep | error |
[00 04 00 00] | Communication with calibration device interrupted | Communication link between machine and calibration device fails | error |
[00 08 00 00] | System fails to write EEPROM | EEPROM write operation fails during calibration | error |
[00 10 00 00] | ACGO switch positioned to ON | ACGO switch detected in ON position during calibration | error |
[00 20 00 00] | Maximum value cannot be found | Calibration algorithm cannot determine maximum flow point | warning |
[00 40 00 00] | Minimum value cannot be found | Calibration algorithm cannot determine minimum flow point | warning |
3-way Valve Failure | Three-way solenoid valve malfunction | Solenoid valve electrical signal control status error | info |
AG Airway Occluded | AG sampling airway occluded | Pump rate drops below 20 ml/min for >1 second | critical |
AG Cal. Failed | AG module calibration failed | AG calibration procedure unsuccessful | critical |
AG Change Watertrap | AG watertrap service interval expired | Watertrap has reached service life limit | error |
AG Comm Stop | AG module communication failure | Loss of communication with AG module | critical |
AG Data Limit Error | AG module data transmission error | AG module data integrity check fails | error |
AG Hardware Error | AG hardware error detected | AG module hardware malfunction | error |
AG Hardware Malfunction module | AG module hardware malfunction | AG module improper installation or hardware fault | critical |
AG Init Error | AG module initialization error | AG module fails to initialize on startup | critical |
AG No Watertrap | AG module watertrap disconnected | Watertrap not detected when system initializes | error |
AG Watertrap Type Wrong | AG watertrap type does not match patient type | Adult watertrap installed when patient type set to infant | error |
AG Zero Failed | AG module zeroing procedure failed | AG module cannot zero sensor output | warning |
Air Flow Too High | Air flow exceeds maximum threshold | Air flow greater than 20 L/min for 1 second | info |
Auto Ventilation Disabled | Automatic ventilation disabled due to system self-test failure | System self-test failure during initialization | info |
Auto Ventilation Disabled - Leak Test Failed | Automatic ventilation disabled due to circuit leak test failure | Automatic circuit leak test failure during initialization | info |
Auto Ventilation is Non-Functional | Automatic ventilation non-functional while switch in Auto position | Auto/Manual switch in Auto position but automatic ventilation not functional | error |
Aux Control Module Comm Stop | Communication loss between auxiliary control module and CPU board | Lost communication with CPU board for 10 seconds | error |
Aux Control Module Selftest Error | Auxiliary ventilator control module failed self-test during startup | CPU communication failure or board internal error at power-on | error |
Aux Control Module Selftest Error | Auxiliary control module self-test failed. CPU, Flash, or watchdog timer error, or communication loss with ventilator control board | Power-on | critical |
Battery in Use | Operating on battery power due to AC failure | AC power failure detected | info |
Battery Undetected | Battery is not detected by the system | Power-on or runtime | warning |
Breathing Circuit Not Mounted | Breathing circuit not properly installed | Breathing circuit detection sensor indicates circuit not mounted | error |
Bundle Version Error | Software versions are not compatible | Incompatible software versions detected at startup | error |
Calibrate Flow Sensor and Insp Valve | Flow sensor calibration table missing or corrupted | Calibration table not found in EEPROM or checksum mismatch | info |
Calibrate Flow Sensor and Insp Valve | Calibration table not found in EEPROM or checksum mismatch for flow sensor and inspiratory valve | Power-on or runtime | info |
Calibrate O2 Sensor | O2 sensor calibration table missing or corrupted in EEPROM | EEPROM calibration data lost or checksum mismatch at startup | warning |
Calibrate O2 Sensor | Calibration table not found in EEPROM or checksum mismatch for oxygen sensor | Power-on or runtime | info |
Calibrate Pressure Sensor and PEEP Valve | Pressure sensor calibration table missing or corrupted in EEPROM | EEPROM calibration data lost or checksum mismatch at startup | warning |
Calibrate Pressure Sensor and PEEP Valve | Calibration table not found in EEPROM or checksum mismatch for pressure sensor and PEEP valve | Power-on or runtime | info |
Check Flow Sensors | Flow sensor measurement error detected | Inspiratory reverse flow OR expiratory reverse flow detected | error |
CO2 Absorber Canister Not Locked | CO2 absorber canister not properly mounted | CO2 canister mounting detection sensor indicates canister not mounted | error |
CO2 Accuracy Unspecified | CO2 measurement outside module accuracy range | Measured CO2 concentration exceeds specification | warning |
Des Accuracy Unspecified | Desflurane measurement outside module accuracy range | Measured desflurane concentration exceeds specification | warning |
[Drive Gas Pressure Low] | Drive gas (O2) pressure is lower than 200 kPa | Drive gas supply pressure falls below threshold during calibration initiation | error |
Drive Gas Pressure Low | Drive gas (O2) pressure to expiratory valve assembly is below safe operating limit | O2 supply pressure falls below pressure switch threshold (~0.2 MPa / 2.0 bar) | warning |
Drive Gas Pressure Low | Drive gas (expiratory valve) pressure is below minimum threshold; alarm triggered by expiratory valve pressure switch | Expiratory valve integrated pneumatic circuit pressure falls below 0.2 MPa (nominal switch point ~220 kPa) | critical |
Enf Accuracy Unspecified | Enflurane measurement outside module accuracy range | Measured enflurane concentration exceeds specification | warning |
EtCO2 Overrange | End-tidal CO2 measurement exceeds module range | EtCO2 concentration above maximum measurable value | warning |
EtDES Overrange | End-tidal desflurane measurement exceeds module range | EtDES concentration above maximum measurable value | warning |
EtENF Overrange | End-tidal enflurane measurement exceeds module range | EtENF concentration above maximum measurable value | warning |
EtHAL Overrange | End-tidal halothane measurement exceeds module range | EtHAL concentration above maximum measurable value | warning |
EtISO Overrange | End-tidal isoflurane measurement exceeds module range | EtISO concentration above maximum measurable value | warning |
EtN2O Overrange | End-tidal N2O measurement exceeds module range | EtN2O concentration above maximum measurable value | warning |
EtO2 Overrange | End-tidal O2 measurement exceeds module range | EtO2 concentration above maximum measurable value | warning |
EtSEV Overrange | End-tidal sevoflurane measurement exceeds module range | EtSEV concentration above maximum measurable value | warning |
External AG Module Disconnected | External AG module not connected to system | FDA version expects external AG but module not found | critical |
External AG Self Test Error | External anesthetic gas module self-test failure | External AG module selftest malfunction | info |
Fan Failure | System fan speed below minimum threshold | Fan speed 20% off nominal value (normal ~4000 RPM) | warning |
Fan Failure | Cooling fan speed 20% below nominal rotation speed | Fan speed monitoring indicates speed deviation | warning |
Fan Failure 02 | Module rack fan speed below minimum threshold | Module rack fan speed less than 3640 RPM (normal ~4000 RPM) | warning |
[FF FF FF FF] | Communication error | Communication protocol or CRC error with control board or calibration device | critical |
FF FF FF FF | Communication error | General communication failure between system components | critical |
FiCO2 Overrange | Inspired CO2 measurement exceeds module range | FiCO2 concentration above maximum measurable value | warning |
FiDES Overrange | Inspired desflurane measurement exceeds module range | FiDES concentration above maximum measurable value | warning |
FiENF Overrange | Inspired enflurane measurement exceeds module range | FiENF concentration above maximum measurable value | warning |
FiHAL Overrange | Inspired halothane measurement exceeds module range | FiHAL concentration above maximum measurable value | warning |
FiISO Overrange | Inspired isoflurane measurement exceeds module range | FiISO concentration above maximum measurable value | warning |
FiN2O Overrange | Inspired N2O measurement exceeds module range | FiN2O concentration above maximum measurable value | warning |
FiO2 Overrange | Inspired O2 measurement exceeds module range | FiO2 concentration above maximum measurable value | warning |
FiSEV Overrange | Inspired sevoflurane measurement exceeds module range | FiSEV concentration above maximum measurable value | warning |
Flowmeter Comm Stop | Communication loss between flowmeter and CPU board | Lost communication with CPU board for 10 seconds | error |
Flowmeter Selftest Error | Fresh flow sensor board failed internal self-test during startup | CPU communication failure or board internal error at power-on | error |
Flowmeter Selftest Error | Flowmeter self-test failed during startup. CPU, Flash memory, or watchdog timer error, or communication failure with sensor board | Power-on | critical |
Flowmeter Voltage Error | Electronic flowmeter voltage (DVCC, AVDD or VC) is out of specification | Power-on or runtime | critical |
Flowmeter Voltage Error | Voltage error on digital voltage supply channels (DVCC, AVDD or VC) | Power supply or communication failure to fresh flow sensor board | error |
Flow Sensor Failure | Ventilator flow sensor output is out of normal operating range | Runtime | info |
Hal Accuracy Unspecified | Halothane measurement outside module accuracy range | Measured halothane concentration exceeds specification | warning |
Heating Module Failure | Heating element temperature out of specification | Heating resistance temperature abnormal for extended period | error |
Incompatible AG Software Version | AG module software version too old | AG module software version < 1.7.3 | critical |
Insp Valve Failure | Inspiratory valve malfunction | Inspiratory valve voltage error or inspiratory valve flow error | warning |
Insp Valve Failure | Inspiratory valve has voltage error or flow sensor error | Runtime | warning |
Internal Air Flow Failure | Communication failure between flowmeter CPU and Air flow sensor | Communication failure between flowmeter CPU and Air sensor for 3 seconds continuously | info |
Internal N2O Flow Failure | Communication failure between flowmeter CPU and N2O flow sensor | Communication failure between flowmeter CPU and N2O sensor for 3 seconds continuously | info |
Internal O2 Flow Failure | Communication failure between flowmeter CPU and O2 flow sensor | Communication failure between flowmeter CPU and O2 sensor for 3 seconds continuously | info |
IP Address Conflict | Device IP address duplicated on local network | IP address matches another device on same network | warning |
IP Address Conflict | IP address is the same as another machine on the network | Power-on or runtime | warning |
Iso Accuracy Unspecified | Isoflurane measurement outside module accuracy range | Measured isoflurane concentration exceeds specification | warning |
Keyboard Self Test Error | Keyboard malfunction detected | Keyboard self-test timeout during startup | error |
Leak | Gas delivery system leak detected in one or more assemblies | Continuous pressure drop detected during system test or operation | error |
Low Battery Voltage! | Battery voltage critically low | Battery voltage less than 10.6V for 5 seconds | error |
[Manual Vent.] | Auto/Manual ventilation switch is in wrong position | Calibration system detects Auto/Manual switch set to bag position instead of mechanical ventilation | error |
Mixed Agent | Two halogenated anesthetic agents detected with low combined MAC | MAC < 3 with two halogenated agents monitored | warning |
Mixed Agent and MAC≥3 | Two halogenated anesthetic agents with high combined MAC | MAC >= 3 with two halogenated agents monitored | error |
N2O Accuracy Unspecified | N2O measurement outside module accuracy range | Measured N2O concentration exceeds specification | warning |
N2O Flow Too High | Nitrous oxide flow exceeds maximum threshold | N2O flow greater than 15 L/min for 1 second | info |
NO Fresh Gas | Fresh gas flow below minimum threshold | Fresh gas flow less than 50 mL/min | warning |
NO Fresh Gas | Fresh gas flow insufficient for ventilation | Fresh gas flow less than 50 mL/min | warning |
O2 Accuracy Unspecified | O2 measurement outside module accuracy range | Measured O2 concentration exceeds specification | warning |
O2 Flow Too High | Oxygen flow exceeds maximum threshold | O2 flow greater than 25 L/min for 1 second | info |
O2-N2O Ratio Error | Oxygen-nitrous oxide ratio out of specification | N2O > 0.5 L/min AND N2O > 4x O2 flow for 1.6 seconds | error |
O2 Sensor Disconnected | Oxygen sensor not connected | O2 sensor disconnected from cable | info |
O2 Sensor Error | O2 sensor error detected | O2 sensor malfunction | error |
O2 Supply Failure | O2 supply pressure to machine is below safe operating minimum | O2 supply pressure falls below pressure switch threshold (~0.25 MPa / 2.5 bar) | error |
O2 Supply Failure | Oxygen supply failure or insufficient pressure | O2 pressure switch detects inadequate oxygen supply | error |
Patient Circuit Leak | Breathing circuit or patient connection leak detected | Pressure < 2 cmH2O for 30 seconds during mechanical ventilation OR patient not connected | warning |
PEEP Safety Valve Failure | PEEP safety valve electrical fault | PEEP safety valve voltage error | warning |
PEEP Valve Failure | Positive end-expiratory pressure valve malfunction | PEEP valve voltage error or PEEP valve pressure error | warning |
Pinsp Not Achieved | Inspiratory pressure not achieved | In pressure mode, Pinsp < 2/3 of setting (if >9 cmH2O) or < setting-3 cmH2O (if <9 cmH2O) for 6 cycles | info |
Power Board High Temp | Power board temperature exceeds 95 degrees Celsius | Runtime | critical |
Power Supply Voltage Error | Power board output voltage (3.3V, 5V, or 12V) is out of specification | Power-on or runtime | critical |
Power Supply Voltage Error (Runtime) | 3.3V, 5V, or 12V supply voltage out of specification during operation | Power board detects output voltage out of range | error |
Power System Comm Stop | Communication loss between power board and CPU board | Lost communication with CPU board for 10 seconds | error |
Rate Overrange | Pump rate measurement exceeds module range | Measured pump rate above maximum specification | warning |
Replace O2 sensor | Oxygen sensor reading critically low | O2 value less than 5% | warning |
RT Clock Failure | Real-time clock chip malfunction | RT chip malfunction detected | error |
RT Clock Needs Battery | Real-time clock backup battery is missing or depleted | Power-on or runtime | critical |
Sev Accuracy Unspecified | Sevoflurane measurement outside module accuracy range | Measured sevoflurane concentration exceeds specification | warning |
System going DOWN, Battery depleted! | Battery voltage critically low; system shutting down | Battery voltage less than 10.2V at power-on | critical |
System going DOWN, Battery depleted! | System shutdown imminent due to battery depletion | Battery voltage less than 10.2V | critical |
Ventilator Comm Stop | Communication loss between ventilator board and CPU board | Lost communication with CPU board for 10 seconds | error |
Ventilator Initialization Error | Ventilator control board did not receive parameter settings at power-on | CPU to ventilator communication failure during initialization | error |
Ventilator Initialization Error | CPU board cannot send parameter settings to ventilator control board at power-on | Power-on | critical |
Ventilator Initialization Error | Ventilator parameter initialization timeout | CPU board cannot send parameter settings to ventilator board after power on | error |
Ventilator Selftest Error | Ventilator control board failed internal self-test during startup | CPU communication failure or board internal error at power-on | error |
Ventilator Selftest Error | Ventilator self-test failed. CPU, timer, RAM, watchdog, EEPROM or ADC error, or communication loss with ventilator control board | Power-on | critical |
Ventilator Voltage Error | Ventilator control module power supply voltage (5V or 12V) is out of specification | Power-on or runtime | critical |
Ventilator Voltage Error | 5V or 12V supply voltage out of specification to ventilator control module | Power board output voltage failure | error |
Vt Not Achieved | Tidal volume not achieved | Tidal volume < TV setting for 6 cycles for >20% or 50 mL (whichever greater) | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI