Midmark Preva: troubleshooting and error codes
The ServiceForge knowledge base holds 23 documented symptoms for the Midmark Preva, with 53 ranked causes, 80 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power is ON but no operator display high voltage
System powered on but operator panel displays no information, may have buzzing sound
Ranked causes
- 1. No AC mains power to USB hub (common)
- 2. Service disconnect switch failure (occasional)
- 3. Communication cable between Logic Board and Operator Panel damaged (occasional)
- 4. Ribbon cable between Power Supply and Logic Board loose or damaged (occasional)
- 5. Logic Board voltage supply failure (occasional)
- 6. Power Supply Board failure (occasional)
- 7. Logic Board failure (occasional)
- 8. Operator Panel Board failure (occasional)
Check for 100-230V AC at Power Supply Board J4 USB Hub Power connector
- Expected
- 100-230 volts AC present
- Tools
- Multimeter
10 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No kV produced high voltage
Operator panel displays message that kV is too low; no high voltage produced during exposure
Ranked causes
- 1. D11 light on Power Board not illuminated (common)
- 2. Power Supply Board output voltage out of range (common)
- 3. Logic Board missing supply voltages (common)
- 4. Power Supply Board missing supply voltages (common)
- 5. Logic Board kV Setpoint voltage incorrect (occasional)
- 6. Ribbon cable between boards damaged or unseated (occasional)
- 7. Cable between Control Unit and Tube Head assembly damaged (occasional)
Verify D11 indicator lamp illumination
- Test point
- D11 on Power Board
- Expected
- D11 light illuminated
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High-voltage breakdown during exposure high voltage
Loud snap sound occurs during exposure attempt. High voltage present but excessive and uncontrolled.
Ranked causes
- 1. Circuit breaker tripped during exposure (common)
- 2. Power Board failure (occasional)
- 3. Power Supply Board voltage out of range (occasional)
- 4. Faulty cabling between boards or to tube head (occasional)
- 5. Tube Head Assembly malfunction or feedback circuit failure (occasional)
- 6. Logic Board failure (occasional)
Verify D11 indicator light status on Power Board
- Test point
- D11 light on Power Board
- Expected
- illuminated
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device ready but no radiation produced radiation
System powers on, Ready indicator illuminated green, exposure switch activated, but no X-ray output; relay and LED may or may not engage; audible indicators may not sound
Ranked causes
- 1. Operator Panel indicator does not sound upon exposure (common)
- 2. Relay or D2 LED do not activate on Power Board (occasional)
- 3. D401 on Logic Board does not illuminate during exposure (occasional)
- 4. Exposure signal conductor in Cat 5 cable is open (occasional)
- 5. Logic Board indicator sounds but Power Board does not activate (rare)
Verify system power and Ready indicator
- Test point
- Operator Panel Ready Indicator LED
- Expected
- Green light illuminated when Preva is powered on and idle
5 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Electrical integrity inspection high voltage
Verify integrity of power cord, circuit breaker, power switch, X-ray irradiation switches, and operator panel functionality.
Ranked causes
- 1. Power cord damage or loose ground connection (unknown)
- 2. Circuit breaker or power switch malfunction (unknown)
- 3. Exposure switch malfunction (unknown)
- 4. Operator panel malfunction or display issue (unknown)
Visually inspect power cord for damage
- Test point
- Power cord
- Expected
- No cuts, breaks, or damage visible
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous beep at power-up
Continuous beeping sound during power-up; display may be backlit or normal
Ranked causes
- 1. Operator Display Board fault with missing or incorrect technique factors display (common)
- 2. Logic Board fault with missing or incorrect technique factors display (common)
- 3. Logic Board fault with normal display (occasional)
Inspect display condition at power-up
- Test point
- Display screen
- Expected
- Display backlit with technique factors visible, or display with normal operation
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error – Serial Communication displays
Error message displayed indicating no communication between Logic Board and Operator Panel
Ranked causes
- 1. Cat 5 communication cable open between Logic Board and Operator Panel (common)
- 2. Logic Board or Operator Panel Board fault (occasional)
Inspect Cat 5 communication cable continuity and condition
- Test point
- Cat 5 cable between Logic Board and Operator Panel
- Expected
- Cable intact, properly connected, no breaks or open circuits
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display or abnormal display at power-up
Operator Display Board is backlit but does not show technique factors, or no display appears at all
Ranked causes
- 1. Operator Display Board failure (common)
- 2. Logic Board failure (common)
Verify display backlight status
- Test point
- Operator Display
- Expected
- Display backlit
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No X-ray produced but exposure indicators detected radiation
Unit operates exposure indicators but produces no X-ray output. Diagnostic mode shows either no kV or no mA produced.
Ranked causes
- 1. No kV produced (unknown)
- 2. No mA produced (unknown)
Activate Diagnostic Mode
- Test point
- Operator Panel
- Expected
- Diagnostic Mode active per page 192 procedure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzing Noise at Power-Up
System produces buzzing sound during power-up sequence
Ranked causes
- 1. Ribbon cable between Power Supply Board and Logic Board not properly seated or damaged (common)
Inspect and reseat ribbon cable between Power Supply Board and Logic Board
- Test point
- Ribbon cable connection
- Expected
- Cable fully seated with no gaps
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Articulated Arm balance and friction caution
Articulated Arm balance or friction out of specification; arm not steady in all positions or does not move smoothly
Ranked causes
- 1. Balance or friction adjustment required (occasional)
Verify Articulated Arm is steady in all positions when Tube head is moved
- Test point
- Articulated Arm assembly
- Expected
- Arm remains steady in all positions and moves before Horizontal Arm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Caster tightening (Mobile units only) patient safety
Inspect and tighten loose caster bolts on mobile base to prevent system wobbling and unsafe operation.
Ranked causes
- 1. Loose caster bolts from repeated rolling over rough surfaces (common)
Visually inspect casters for visible looseness
- Test point
- Caster attachment to mobile base
- Expected
- Caster snug against base, no visible threads between caster and base
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Device support inspection patient safety
Verify structural integrity of wall attachment and ensure Preva system is securely fastened to the wall.
Ranked causes
- 1. Inadequate wall support or loose attachment (unknown)
Visually inspect wall support adequacy
- Test point
- Wall support structure
- Expected
- Wall support adequate and secure
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Horizontal Arm brake adjustment caution
Horizontal Arm does not move with proper resistance; should begin to move after Articulated Arm
Ranked causes
- 1. Brake assembly screws not properly tightened (common)
Verify Horizontal Arm moves after Articulated Arm when Tube head is repositioned
- Test point
- Horizontal Arm pivot post
- Expected
- Horizontal Arm begins to move after Articulated Arm; Articulated Arm steady in all positions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Horizontal Arm pivot braking
Adjust brake assembly screws to ensure Horizontal Arm begins to move only after Articulated Arm
Ranked causes
- 1. Factory pre-set braking requires verification during installation (unknown)
Verify Articulated Arm movement before Horizontal Arm
- Test point
- Brake assembly clamping bar at Control Unit top
- Expected
- Articulated Arm moves before Horizontal Arm begins to move; Tube head remains steady in all positions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Imaging performance verification radiation
Verify that Preva produces acceptable radiographic images of a test object as intended.
Ranked causes
- 1. Imaging system malfunction or degradation (unknown)
Acquire radiograph of test object
- Test point
- Test object radiograph
- Expected
- Image quality acceptable for clinical use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Irradiation interlocks operation verification radiation
Verify that irradiation interlock circuits safely inhibit X-ray operation when safety switches open (e.g., door open), and that X-ray resumes when interlock is closed.
Ranked causes
- 1. Interlock circuit malfunction or damage (unknown)
Visually inspect irradiation interlock circuit and exposed wiring
- Test point
- Interlock circuit and related wiring
- Expected
- No damage visible
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: kilovoltage (kV) radiation
kV output does not match desired technique factor setting; requires X-ray kV meter (Radcal Accugold or similar) to verify and adjust output
Ranked causes
- 1. kV output requires adjustment to match set technique factor (unknown)
Enable Diagnostic Mode
- Test point
- System menu
- Expected
- Diagnostic Mode active; operational display returned
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Mechanical integrity inspection patient safety
Verify mechanical integrity of Control Unit, Horizontal Arm, Articulated Arm, Yoke, Tube Head, and attachment points are structurally sound without excessive wear or wobbling.
Ranked causes
- 1. Mechanical wear, damage, or loose connections (unknown)
Visually inspect Control Unit
- Test point
- Control Unit
- Expected
- No visible damage, excessive wear, or looseness
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: On-machine labels inspection patient safety
Verify legibility of all device on-machine labels that communicate important safety messages.
Ranked causes
- 1. Labels illegible, damaged, or missing (unknown)
Visually inspect all on-machine labels for legibility
- Test point
- All on-machine labels
- Expected
- All labels clearly legible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: X-ray performance QA by Qualified Physicist radiation
Arrange for X-ray performance inspection by a Qualified Physicist to verify essential X-ray functionality.
Ranked causes
- 1. X-ray performance degradation or non-compliance (unknown)
Arrange and complete X-ray performance QA by Qualified Physicist
- Test point
- X-ray system performance
- Expected
- Performance meets specification per Qualification Procedure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pre-Termination Error displays without cause high voltage
Error screen appears when exposure button is released early or from high-voltage breakdown in tube head; system normally returns to normal operation after five seconds
Ranked causes
- 1. Exposure button released early during exposure cycle (common)
Verify operator exposure button release timing
- Test point
- Operator technique
- Expected
- Button held for full exposure duration per technique selection
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ribbon cable defect between Power Supply and Logic Boards caution
Faulty ribbon cable connection between Power Supply and Logic Boards causing system malfunction.
Ranked causes
- 1. Ribbon cable damaged or disconnected (unknown)
Inspect ribbon cable between Power Supply and Logic Boards
- Test point
- Ribbon cable connector
- Expected
- Cable intact, fully seated, no visible damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Error – Serial Communication | No communication between Logic Board and Operator Panel | Cat 5 cable open or Logic Board/Operator Panel Board failure | error |
Pre-Termination Error | Exposure terminated prematurely resulting in underexposed image | Early release of exposure button or high-voltage breakdown in Tube Head | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.