Midmark M4.9 Easyclave: troubleshooting and error codes
The ServiceForge knowledge base holds 54 documented symptoms for the Midmark M4.9 Easyclave, with 218 ranked causes, 104 diagnostic checks with test points and expected values, and 73 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Chamber pressure does not build during heat-up patient safety
Pressure fails to reach 1 psig in 1 minute or 4.93 psig within 16 minutes of heat-up mode start
Ranked causes
- 1. Door gasket leaking (common)
- 2. Pressure relief valve leaking (common)
- 3. Vent valve leaking (common)
- 4. Air valve leaking (common)
- 5. Leakage around fittings (occasional)
- 6. Steam heater malfunction (occasional)
- 7. Main relay contacts not closing (occasional)
- 8. Pressure transducer out of calibration (occasional)
- 9. Pressure transducer malfunction (rare)
- 10. Main PC board malfunction (rare)
Inspect all seals and gaskets for leakage
- Expected
- No visible leakage
3 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sterilization temperature regulation questionable patient safety
For second time during Sterilization Mode (first time unit automatically restarts), questionable regulation of sterilization conditions (pressure/temperature) inside chamber
Ranked causes
- 1. Inadequate water in chamber trough (common)
- 2. Chamber overloaded with articles (common)
- 3. Steam leakage from chamber (occasional)
- 4. Temperature RTD out of calibration (occasional)
- 5. Pressure Transducer out of calibration (occasional)
- 6. Air Valve malfunctioning (occasional)
- 7. Pressure Transducer malfunctioning (occasional)
- 8. Steam Temperature Sensor (RTD) malfunctioning (occasional)
- 9. P.C. Board malfunctioning (rare)
Visual inspection of water level in chamber trough
- Test point
- Chamber trough after FILL cycle
- Expected
- Adequate water level per equipment manual
- Tools
- Visual inspection
1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sterilizer not drying instruments properly
Instruments are wet after drying cycle is complete
Ranked causes
- 1. Sterilizer overloaded (common)
- 2. Operator fully opened or closed door before Dry Cycle complete (occasional)
- 3. Sterilizer not level (occasional)
- 4. Chamber filter clogged (common)
- 5. Vent valve malfunctioning or lines restricted (occasional)
- 6. Heater not functioning (occasional)
- 7. High limit switch stuck open (occasional)
- 8. Solid State Relay malfunctioning (rare)
- 9. Main relay malfunctioning (rare)
Verify sterilizer is level
- Test point
- Base of unit
- Expected
- Level surface
3 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchpad and display do not work with power present
Sterilizer has power but touchpad and display show no response or undefined characters with intermittent beep
Ranked causes
- 1. J12 connector loose on Main P.C. board (common)
- 2. Wire harness from transformer to J12 has broken wire (occasional)
- 3. Transformer has open or shorted winding (occasional)
- 4. Fuses F1 and F2 on Main P.C. board blown (occasional)
- 5. Ribbon cable from J1 to J3 disconnected (occasional)
- 6. J1/J3 ribbon cable has open lead (occasional)
- 7. Main P.C. Board malfunction (rare)
- 8. Display P.C. Board malfunction (rare)
Inspect J12 connector condition
- Test point
- J12 on Main P.C. board
- Expected
- Connector fully seated and undamaged
7 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature high during sterilization high voltage
Auxiliary temperature exceeds 568.4°F (298°C) during STERILIZE mode; items may not be sterile
Ranked causes
- 1. Sterilizer not level or water fill inadequate (common)
- 2. Level Sensor Probe dirty or corroded (common)
- 3. Vent Valve malfunctioning, staying open (occasional)
- 4. Air Valve doesn't close (occasional)
- 5. Auxiliary Temperature Sensor (RTD) malfunctioning (occasional)
- 6. SSR shorted causing constant heater power (occasional)
- 7. Main P.C. Board malfunctioning (rare)
Check SSR resistance when heaters energized
- Test point
- SSR terminals 1 & 2
- Expected
- approximately 700 ohms when heaters on, +/- 50 ohms
- Tools
- Ohmmeter
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C232 displayed
Display shows FILL MODE, WATER LOW, ITEMS NOT STERILE, PUSH STOP TO RESTART
Ranked causes
- 1. Reservoir water level low (common)
- 2. Mesh filter restricted (common)
- 3. Water Fill Level Sensor dirty (occasional)
- 4. Fill Valve or related plumbing clogged (occasional)
- 5. Fill Valve solenoid coil malfunctioning (occasional)
- 6. J6 Plug connection loose or broken wires (occasional)
- 7. PC Board malfunctioning (rare)
Check water level in Reservoir
- Test point
- Reservoir
- Expected
- water in correct range on the water level label
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature exceeds 276.1°F during operation patient safety
Unit displays error C570, C573, or C574. Steam temperature exceeds 276.1°F (135.6°C) during power-up, heat-up, or sterilize mode.
Ranked causes
- 1. Steam Temperature Sensor (RTD) malfunction (common)
- 2. Solid State Relay (SSR) shorted (occasional)
- 3. Main Relay malfunctioning (occasional)
- 4. Main P.C. Board malfunction (occasional)
- 5. Pressure lines restricted or clogged (occasional)
- 6. Pressure Transducer malfunction (occasional)
- 7. Cassette Tray Rack installed backward (rare)
Measure RTD resistance at J5 connector
- Test point
- J5 on P.C. Board
- Expected
- 1090 ohms at 75°F or 2066 ohms at 545°F, ±10 ohms
- Tools
- Ohm meter
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature high during heatup high voltage
Auxiliary temperature exceeds 568.4°F (298°C) during HEATUP mode; items may not be sterile
Ranked causes
- 1. Sterilizer not level or water fill inadequate (common)
- 2. Level Sensor Probe dirty or corroded (common)
- 3. Vent Valve malfunctioning, staying open (occasional)
- 4. Air Valve doesn't close (occasional)
- 5. Auxiliary Temperature Sensor (RTD) malfunctioning (occasional)
- 6. Main P.C. Board malfunctioning (rare)
Check sterilizer level and water fill
- Test point
- Chamber water level and unit level
- Expected
- Unit level per manual; adequate water in trough
- Tools
- Level tool, visual inspection
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Biological indicator strip designates unit not sterilizing patient safety
Positive biological indicator or no color change on chemical indicator strip
Ranked causes
- 1. Sterilizer overloaded (common)
- 2. Improper operator procedure (common)
- 3. Wrong instrument trays used (occasional)
- 4. Wrong type of biological indicator (occasional)
- 5. Chemical indicator strip in contact with water (occasional)
- 6. Indicator strips stored in damp or hot environment (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dry mode pressure high caution
Unit displays Code C647; gauge pressure inside chamber equals or exceeds 2.18 PSIG (15 kPa) during DRY mode
Ranked causes
- 1. Door did not fully open; door gasket stuck (common)
- 2. Filter screen in trough restricted (common)
- 3. Vent Valve malfunctioning or lines restricted (common)
- 4. Pressure Transducer out of calibration (occasional)
- 5. Pressure Transducer malfunctioning (occasional)
- 6. P.C. Board malfunctioning (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High-limit switch open during power up caution
Unit displays Code C980; Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during POWER UP mode; requires unplug/re-plug
Ranked causes
- 1. Water level covering condensing coil outlet (common)
- 2. High Limit Switch weak, opens prematurely (common)
- 3. High Limit Switch opened normally due to overheat (occasional)
- 4. Wire or connector loose, disconnected or broken (occasional)
- 5. Insulation deteriorated between High Limit Switch and chamber (occasional)
- 6. P.C. Board malfunctioning (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure over limit in sterilize mode caution
Chamber pressure exceeds 34.8 PSIG (240 kPa) during STERILIZE mode. Error C674 displayed. Items marked not sterile.
Ranked causes
- 1. Air valve malfunction (common)
- 2. Pressure transducer malfunction (common)
- 3. Temperature RTD out of calibration (occasional)
- 4. Steam temperature sensor (RTD) malfunction (occasional)
- 5. Solid State Relay (SSR) shorted (occasional)
- 6. P.C. board malfunction (rare)
Test solid state relay resistance
- Test point
- SSR terminals 1 and 2
- Expected
- approximately 700 ohms when heaters on
- Tools
- Multimeter (ohms setting), run Service Diagnostics Steam/Auxiliary Heater Test
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer does not print
Printer does not generate data
Ranked causes
- 1. Printer out of paper (common)
- 2. Cartridge ribbon is dry (common)
- 3. Printer wire harness disconnected (occasional)
- 4. Printer wire harness broken or open lead (occasional)
- 5. Printer malfunctioning (occasional)
- 6. Main P.C. board malfunctioning (rare)
Verify printer has paper
- Test point
- Paper roll compartment
- Expected
- Paper installed
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature exceeds limit during vent mode high voltage
Temperature exceeds 276.1°F (135.6°C) when system should be venting, items marked not sterile
Ranked causes
- 1. Filter screen in trough restricted (common)
- 2. Vent valve malfunctioning or lines restricted (occasional)
- 3. Steam temperature sensor malfunction (occasional)
- 4. Solid state relay shorted, constant heater power (occasional)
- 5. Main relay malfunctioning (rare)
- 6. Main PC board malfunction (rare)
Inspect filter screen for obstruction
- Expected
- Screen clean and unobstructed
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature low during heatup high voltage
During Heat Up Mode steam temperature does not rise at least 5°F within 10 minutes; sterilization mode not reached by 20 minutes; items not sterile
Ranked causes
- 1. Steam heater malfunctioning (common)
- 2. Steam leakage from chamber (common)
- 3. Solid State Relay (SSR) malfunctioning (occasional)
- 4. Main Relay malfunctioning (occasional)
- 5. Steam Temperature sensor (RTD) malfunctioning (occasional)
- 6. Main P.C. Board malfunctioning (rare)
Measure continuity of steam heater
- Test point
- Steam (Lower) heater terminals
- Expected
- 115 VAC heaters 10 ohms cold; 230 VAC heaters 39 ohms cold, varies by voltage
- Tools
- Ohmmeter
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature fails to reach target during drying high voltage
Display shows C437: DRY MODE AUX TEMP LOW. Chamber temperature below 158°F (70°C) after 5 minutes into DRY mode.
Ranked causes
- 1. Door left wide open and tray rack not in position (common)
- 2. Solid State Relay (SSR) malfunctioning (occasional)
- 3. Auxiliary (upper) or Steam (lower) heater has open winding (occasional)
- 4. Auxiliary Temperature Sensor (RTD) malfunctioning (occasional)
- 5. Main P.C. Board malfunctioning (rare)
Verify door position and tray rack placement
- Expected
- Door in partially open or closed position; tray rack positioned in chamber
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dry mode pressure overlim caution
Unit displays Code C677; gauge pressure inside chamber exceeds 34.8 PSIG (240 kPa) during DRY mode; requires unplug/re-plug
Ranked causes
- 1. Door did not fully open; door gasket stuck (common)
- 2. Vent Valve malfunctioning or lines restricted (common)
- 3. Pressure Transducer malfunctioning (occasional)
- 4. Pressure Transducer out of calibration (occasional)
- 5. P.C. Board malfunctioning (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High-limit switch open during dry mode caution
Unit displays Code C987; Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during DRY mode
Ranked causes
- 1. Water level covering condensing coil outlet (common)
- 2. High Limit Switch weak, opens prematurely (common)
- 3. High Limit Switch opened normally from overheat (occasional)
- 4. Wire or connector loose or broken (occasional)
- 5. Insulation deteriorated (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
High-limit switch open during select mode caution
Unit displays Code C981; Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during SELECT mode
Ranked causes
- 1. High Limit Switch weak, opens prematurely (common)
- 2. High Limit Switch opened normally from overheat (occasional)
- 3. Wire or connector loose or broken (occasional)
- 4. Insulation deteriorated (rare)
- 5. P.C. Board malfunctioning (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure over limit in heatup mode caution
Chamber pressure exceeds 34.8 PSIG (240 kPa) during HEATUP mode. Error C673 displayed. Items marked not sterile.
Ranked causes
- 1. Air valve malfunction (common)
- 2. Pressure transducer malfunction (common)
- 3. Temperature RTD out of calibration (occasional)
- 4. Steam temperature sensor (RTD) malfunction (occasional)
- 5. P.C. board malfunction (rare)
Measure RTD resistance at J5 connector on P.C. Board
- Test point
- J5 connector
- Expected
- 1090 ohms at 75°F (23°C) or 2066 ohms at 545°F (285°C), +/- 10 ohms
- Tools
- Multimeter (ohms setting)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature exceeds limit during dry mode high voltage
Temperature exceeds 276.1°F (135.6°C) during dry cycle when door should open
Ranked causes
- 1. Door stuck closed due to gasket swelling (common)
- 2. Steam temperature sensor malfunction (occasional)
- 3. Solid state relay shorted (occasional)
- 4. Main relay malfunction (rare)
- 5. Main PC board malfunction (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent mode pressure overlim caution
Unit displays Code C675; gauge pressure inside chamber exceeds 34.8 PSIG (240 kPa) during VENT mode; items not sterile; requires unplug/re-plug
Ranked causes
- 1. Filter screen in trough restricted (common)
- 2. Vent Valve malfunctioning or lines restricted (common)
- 3. Pressure Transducer malfunctioning (occasional)
- 4. Pressure Transducer out of calibration (occasional)
- 5. P.C. Board malfunctioning (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature exceeds safe limit during power-up high voltage
Display shows C470: POWER UP MODE AUX TEMP OVERLIM with UNPLUG / RE-PLUG UNIT prompt. Auxiliary temperature exceeds 568.4°F (298°C) during POWER UP mode.
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) malfunctioning (resistance above 2113 ohms) (common)
- 2. Previous malfunction causing false limit trigger (occasional)
- 3. Auxiliary high limit switch weak or opened prematurely (occasional)
- 4. P.C. Board malfunctioning (rare)
Measure RTD resistance at P.C. Board J3 connector
- Test point
- J3 connector on P.C. Board
- Expected
- 75°F (23°C) = 1090 ohms; 545°F (285°C) = 2066 ohms; +/- 10 ohms, +/- 10 ohms
- Tools
- Multimeter
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C326 displayed
Display shows DOOR MODE, DOOR CLOSED, OPEN DOOR; door remains closed after solenoid energized or is open when code displays
Ranked causes
- 1. Door hanging up on door pulse solenoid latch (occasional)
- 2. Instrument inside chamber depressing door interlock plunger (occasional)
- 3. Door Switch malfunctioning (occasional)
- 4. PC Board malfunctioning (rare)
Check position of door (open or closed); run Service Diagnostics and check door pulse solenoid operation
- Test point
- Door, door pulse solenoid
- Expected
- door opens after solenoid pulse
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C431 displayed
Display shows SELECT MODE, AUX TEMP LOW, PUSH STOP TO RESTART; auxiliary temperature has not risen to at least 212°F (100°C) in 10 minutes during SELECT/PREHEAT mode
Ranked causes
- 1. Door left wide open and tray rack not in position (common)
- 2. Water level of reservoir covering outlet of condensing coil (occasional)
- 3. Solid State Relay (SSR) malfunctioning (occasional)
- 4. Auxiliary (upper) or Steam (lower) heater has open winding (occasional)
Check position of door and placement of tray rack after cycle
- Test point
- Door position, tray rack position
- Expected
- door partially open or closed; tray rack in place
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fill mode pressure high caution
Unit displays Code C642 with 'Items Not Sterile' after filling begins; gauge pressure inside chamber equals or exceeds 2.18 PSIG (15 kPa)
Ranked causes
- 1. Air Valve does not close (common)
- 2. Vent Valve malfunctioning, staying open (common)
- 3. Atmospheric Pressure out of calibration (occasional)
- 4. Pressure Transducer out of calibration (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure over limit in fill mode caution
Chamber pressure exceeds 34.8 PSIG (240 kPa) during FILL mode. Error C672 displayed. Items marked not sterile.
Ranked causes
- 1. Air valve malfunction (common)
- 2. Pressure transducer malfunction (common)
- 3. Pressure transducer out of calibration (occasional)
- 4. P.C. board malfunction (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure over limit in select mode caution
Chamber pressure exceeds 34.8 PSIG (240 kPa) during SELECT/FILL mode. Error C671 displayed. Items marked not sterile.
Ranked causes
- 1. Pressure transducer malfunction (common)
- 2. Pressure transducer out of calibration (occasional)
- 3. Vent valve malfunction or blocked (occasional)
- 4. P.C. board malfunction (rare)
Check pressure transducer connector J14 ribbon cable
- Test point
- J14 connector on P.C. Board
- Expected
- connector secure
- Tools
- Visual inspection
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature exceeds limit during door open mode high voltage
Temperature exceeds 276.1°F (135.6°C) when door is open
Ranked causes
- 1. Steam temperature sensor malfunction (common)
- 2. Solid state relay shorted (occasional)
- 3. Main relay malfunction (occasional)
- 4. Main PC board malfunction (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sterilizer has no power
Touchpad and display do not respond when pressed; no indication of operation
Ranked causes
- 1. Power cord not plugged in (common)
- 2. No power at outlet or circuit breaker tripped (common)
- 3. Main L1 fuse blown (occasional)
- 4. Transformer has open windings (occasional)
Verify power cord connection
- Test point
- Power cord connector at sterilizer and outlet
- Expected
- Cord fully inserted
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature exceeds safe limit during select mode high voltage
Display shows C471: SELECT MODE AUX TEMP OVERLIM with UNPLUG / RE-PLUG UNIT prompt. Auxiliary temperature exceeds 568.4°F (298°C) during SELECT mode.
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) malfunctioning (common)
- 2. Solid State Relay (SSR) shorted causing constant heater power (occasional)
- 3. P.C. Board malfunctioning (rare)
Measure RTD resistance at P.C. Board J3 connector
- Test point
- J3 connector on P.C. Board
- Expected
- 75°F (23°C) = 1090 ohms; 545°F (285°C) = 2066 ohms; +/- 10 ohms, +/- 10 ohms
- Tools
- Multimeter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature high during dry high voltage
Auxiliary temperature exceeds 568.4°F (298°C) during DRY mode; items may not be sterile
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) malfunctioning (common)
- 2. Main Relay malfunctioning (occasional)
- 3. SSR shorted causing constant heater power (occasional)
Measure Auxiliary RTD resistance
- Test point
- J3 connector
- Expected
- 75°F = 1090 ohms +/- 10; 545°F = 2066 ohms +/- 10, +/- 10%
- Tools
- Ohmmeter
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature high during vent or heatup (C47X codes) high voltage
Auxiliary temperature exceeds 568.4°F (298°C) during VENT mode or HEATUP mode; items may not be sterile
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) malfunctioning (common)
- 2. SSR shorted causing constant heater power (occasional)
- 3. Main P.C. Board malfunctioning (rare)
Measure Auxiliary RTD resistance
- Test point
- J3 connector
- Expected
- 75°F = 1090 ohms +/- 10; 545°F = 2066 ohms +/- 10, +/- 10 ohms
- Tools
- Ohmmeter
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature sensor hardware error during drying caution
Display shows C467: DRY MODE AUX TEMP HARDWARE. A/D converter reports value outside normal limits during DRY mode.
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) unplugged from P.C. Board (common)
- 2. Auxiliary Temperature Sensor (RTD) malfunctioning (occasional)
- 3. Main P.C. Board malfunctioning (rare)
Inspect Auxiliary RTD plug condition and security at P.C. Board
- Test point
- J3 connector on P.C. Board
- Expected
- Connector seated securely; no corrosion or damage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature sensor hardware error during select mode caution
Display shows C461: SELECT MODE AUX TEMP HARDWARE. A/D converter reports value outside normal limits during SELECT mode (occurs during warm start/back-to-back cycles).
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) unplugged from P.C. Board (common)
- 2. Auxiliary Temperature Sensor (RTD) malfunctioning (occasional)
- 3. Main P.C. Board malfunctioning (rare)
Inspect Auxiliary RTD plug condition and security at P.C. Board
- Test point
- J3 connector on P.C. Board
- Expected
- Connector seated securely; no corrosion or damage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature hardware error high voltage
Steam A/D converter reports value outside normal limits during startup or mode operation
Ranked causes
- 1. Steam RTD unplugged from P.C. Board (common)
- 2. Steam RTD malfunctioning (occasional)
- 3. Main P.C. Board malfunctioning (rare)
Verify Steam RTD connector is seated at J5
- Test point
- J5 pin connector on P.C. Board
- Expected
- Connector fully seated and secure
- Tools
- Visual inspection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature high during sterilization high voltage
During Sterilization Mode chamber temperature increases 6°F or more above cycle's designated sterilization temperature; items may not be sterile
Ranked causes
- 1. Water dissipated or leaked through Vent or Fill Valve (common)
- 2. Steam Temperature sensor (RTD) malfunctioning (occasional)
- 3. Main P.C. Board malfunctioning (rare)
Compare STEAM temp with probe reading during Service Diagnostics
- Test point
- Chamber interior with calibrated probe
- Expected
- STEAM sensor reading matches calibrated probe reading
- Tools
- Calibrated temperature probe, Service Diagnostics
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent mode pressure high caution
Unit displays Code C645 with 'Items Not Sterile'; absolute pressure inside chamber greater than 0.73 PSIG (5 kPa) for longer than 5 minutes
Ranked causes
- 1. Vent Valve malfunctioning, staying open (common)
- 2. Pressure Transducer malfunctioning (common)
- 3. Main P.C. Board malfunctioning (occasional)
Check Vent Valve operation and leakage
- Test point
- Vent Valve
- Expected
- No leakage when closed
- Tools
- Service Diagnostics tool
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature fails to reach target during preheat caution
Display shows C431: SELECT MODE AUX TEMP LOW. Trough temperature has not risen to 212°F (100°C) in 10 minutes while unit is in SELECT/PREHEAT mode.
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) malfunctioning (common)
- 2. Main P.C. Board malfunctioning (occasional)
Run Service Diagnostics and compare temperature readings
- Expected
- STEAM and AUX temperature readings within normal range and in line with each other
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary temperature high during fill high voltage
Auxiliary temperature exceeds 568.4°F (298°C) during FILL mode; items may not be sterile
Ranked causes
- 1. Auxiliary Temperature Sensor (RTD) malfunctioning (common)
- 2. Main P.C. Board malfunctioning (rare)
Measure Auxiliary RTD resistance
- Test point
- J3 connector on P.C. Board
- Expected
- 75°F = 1090 ohms +/- 10; 545°F = 2066 ohms +/- 10, +/- 10 ohms
- Tools
- Ohmmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C382 displayed
Display shows FILL MODE, DOOR OPEN, ITEMS NOT STERILE, CLOSE DOOR; door opened during Fill Mode
Ranked causes
- 1. Operator opened door during Fill Mode (common)
- 2. Door Switch malfunctioning (occasional)
Check with person operating unit to see if door was opened during Fill Mode
- Test point
- Operator action
- Expected
- door remained closed during Fill Mode
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C383 displayed
Display shows HEATUP MODE, DOOR OPEN, ITEMS NOT STERILE, PUSH STOP TO RESTART; door opened during Heatup Mode
Ranked causes
- 1. Operator opened door during Heatup Mode (common)
- 2. Door Switch malfunctioning (occasional)
Check with person operating unit to see if door was opened during Heatup
- Test point
- Operator action
- Expected
- door remained closed during Heatup
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C385 displayed
Display shows VENT MODE, DOOR OPEN, ITEMS NOT STERILE, PUSH STOP TO RESTART; door opened during Vent Mode
Ranked causes
- 1. Operator opened door during Vent Mode (common)
- 2. Door Switch malfunctioning (occasional)
Check with person operating unit to see if door was opened during Vent
- Test point
- Operator action
- Expected
- door remained closed during Vent
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Display with touchpad working
Display is blank or shows undefined characters with intermittent beep; touchpad responds but display does not function
Ranked causes
- 1. Ribbon cable from J2 to touchpad disconnected (common)
- 2. J2 ribbon cable has open lead (occasional)
Inspect J2 connector condition
- Test point
- J2 Plug connection
- Expected
- Connector fully seated with pins aligned
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C010 displayed
Display shows POWER UP MODE, SYSTEM PWR LOSS, ITEMS NOT STERILE, PUSH STOP TO RESTART
Ranked causes
- 1. Unit lost power during Fill, Heat Up, or Sterilization Mode (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C060 displayed
Display shows POWER UP MODE, SYSTEM HARDWARE, ITEMS NOT STERILE, UNPLUG/RE-PLUG UNIT
Ranked causes
- 1. Internal glitch from brief power interruption affecting PC Board components (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C102 displayed
Display shows FILL MODE, STOP PRESSED, ITEMS NOT STERILE, PUSH STOP TO RESTART
Ranked causes
- 1. Operator pressed STOP button during Fill Mode (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Code C384 displayed patient safety
Display shows STERILIZE MODE, DOOR OPEN, ITEMS NOT STERILE, UNPLUG/RE-PLUG UNIT; door opened during Sterilize Mode
Ranked causes
- 1. Door Switch malfunctioning (common)
Check Door Switch contact is closed when door is in closed position
- Test point
- Door Switch contact
- Expected
- closed when door is closed
- Tools
- continuity meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Door interlock plunger sticks or does not operate freely patient safety
Door interlock plunger does not spring back readily when depressed, may be stuck
Ranked causes
- 1. Door interlock plunger contamination or lack of lubrication (common)
Depress door interlock plunger and observe if it operates freely
- Test point
- Door interlock plunger
- Expected
- Plunger depresses and springs back freely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure relief valve does not close properly patient safety
Valve remains open or will not seal completely when pressure relief tab is released during sterilizing cycle
Ranked causes
- 1. Pressure relief valve seat leakage or internal damage (common)
Run UNWRAPPED cycle and lift relief tab at STERILIZING
- Test point
- Pressure relief tab (2)
- Expected
- Valve opens easily and steam vents
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure relief valve requires excessive force to open caution
High force required when lifting pressure relief tab during sterilizing cycle; indicates valve stiction or blockage
Ranked causes
- 1. Pressure relief valve internal blockage or wear (common)
Select UNWRAPPED cycle and press START
- Test point
- Control display
- Expected
- Cycle initiates
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Steam temperature exceeds upper limit high voltage
Steam temperature exceeds 276.1°F (135.6°C) during Power Up mode
Ranked causes
- 1. Previous malfunction or power interruption (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tray rack installed backwards or incorrectly patient safety
Tray rack positioned backwards in chamber, may cause water level sensor grounding issue
Ranked causes
- 1. Tray rack installed backwards in chamber (occasional)
Verify tray rack front is facing toward front of chamber
- Test point
- Tray rack orientation in chamber
- Expected
- Tray rack front positioned toward sterilizer front
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit will not fill
Display shows CHAMBER VENTING BEFORE FILL; unit fails to fill chamber with water
Ranked causes
- 1. Atmospheric pressure setting out of calibration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
C010 | POWER UP MODE, SYSTEM PWR LOSS | Unit lost power for a few seconds during Fill, Heat Up, or Sterilization Mode | warning |
C060 | POWER UP MODE, SYSTEM HARDWARE | An error that occurred in Fill, Heatup, Sterilization or Vent modes was not properly cleared before unit was unplugged and replugged | error |
C102 | FILL MODE, STOP PRESSED | Operator pressed STOP button during Fill Mode | info |
C103 | HEATUP MODE, STOP PRESSED | Operator pressed STOP button during Heatup Mode | info |
C104 | STERILIZE MODE, STOP PRESSED | Operator pressed STOP button during Sterilize Mode | info |
C105 | VENT MODE, STOP PRESSED | Operator pressed STOP button during Vent Mode | info |
C106 | DOOR MODE, STOP PRESSED | Operator pressed STOP button during Door Mode | info |
C232 | FILL MODE, WATER LOW | Water fill sensor did not detect water level within 2 minutes after door closed or chamber vented | warning |
C326 | DOOR MODE, DOOR CLOSED | Door remains closed one second after door pulse solenoid energized, or door is open but code displays | error |
C382 | FILL MODE, DOOR OPEN | Door Switch contacts opened during Fill Mode | warning |
C383 | HEATUP MODE, DOOR OPEN | Door Switch contacts opened during Heatup Mode | warning |
C384 | STERILIZE MODE, DOOR OPEN | Door Switch contacts opened during Sterilize Mode | critical |
C385 | VENT MODE, DOOR OPEN | Door Switch contacts opened during Vent Mode | warning |
C431 | SELECT MODE, AUX TEMP LOW | Auxiliary temperature (trough) has not risen to at least 212°F (100°C) in 10 minutes during SELECT/PREHEAT mode | warning |
C431 | Auxiliary temperature does not reach 212°F (100°C) in 10 minutes during SELECT/PREHEAT mode | Unit in SELECT/PREHEAT mode; auxiliary temp not reaching 212°F within time limit | warning |
C437 | Auxiliary temperature below 158°F (70°C) after 5 minutes in DRY mode | Unit in DRY mode; auxiliary temp not reaching 158°F within 5 minutes | warning |
C461 | Auxiliary A/D converter reports value outside normal limits during SELECT mode (warm start) | Back-to-back cycles in SELECT mode; A/D converter reading exceeds acceptable range | warning |
C467 | Auxiliary A/D converter reports value outside normal limits during DRY mode | DRY mode operation; A/D converter reading exceeds acceptable range | warning |
C470 | Auxiliary temperature exceeds 568.4°F (298°C) during POWER UP mode | POWER UP mode; auxiliary temperature exceeds safe maximum threshold | critical |
C471 | Auxiliary temperature exceeds 568.4°F (298°C) during SELECT mode | SELECT mode operation; auxiliary temperature exceeds safe maximum threshold | critical |
C472 | Fill Mode - Auxiliary temperature exceeds upper limit | Auxiliary temperature exceeds 568.4°F (298°C) during FILL mode | critical |
C473 | Heatup Mode - Auxiliary temperature exceeds upper limit | Auxiliary temperature exceeds 568.4°F (298°C) during HEATUP mode | critical |
C474 | Sterilize Mode - Auxiliary temperature exceeds upper limit | Auxiliary temperature exceeds 568.4°F (298°C) during STERILIZE mode | critical |
C475 | Vent Mode - Auxiliary temperature exceeds upper limit | Auxiliary temperature exceeds 568.4°F (298°C) during VENT mode | critical |
C476 | Heatup Mode - Auxiliary temperature exceeds upper limit (C47X series) | Auxiliary temperature exceeds 568.4°F (298°C) during HEATUP mode | critical |
C477 | Dry Mode - Auxiliary temperature exceeds upper limit | Auxiliary temperature exceeds 568.4°F (298°C) during DRY mode | critical |
C533 | Heatup Mode - Steam temperature low | During Heat Up Mode steam temperature does not rise at least 5°F within 10 minutes | critical |
C544 | Sterilize Mode - Steam temperature high | Chamber temperature increases 6°F or more above designated sterilization temperature | critical |
C554 | Sterilize Mode - Steam temperature regulation issue | For second time during sterilization (first time unit restarts), poor temperature/pressure regulation | critical |
C560 | Power Up Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Power Up | critical |
C561 | Select Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Select mode | critical |
C562 | Fill Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Fill mode | critical |
C563 | Heatup Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Heatup mode | critical |
C564 | Sterilize Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Sterilize mode | critical |
C565 | Vent Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Vent mode | critical |
C566 | Door Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Door mode | critical |
C567 | Dry Mode - Steam temperature hardware error | Steam A/D converter reports value outside normal limits during Dry mode | critical |
C570 | Power Up Mode - Steam temperature exceeds upper limit | Steam temperature exceeds 276.1°F (135.6°C) during Power Up mode | critical |
C573 | Heat-up mode - steam temperature over limit | Steam temperature exceeds 276.1°F (135.6°C) during HEAT UP mode | critical |
C574 | Sterilize mode - steam temperature over limit | Steam temperature exceeds 276.1°F (135.6°C) during STERILIZE mode | critical |
C575 | Vent Mode Steam Temperature Over Limit | Steam temperature exceeds 276.1°F (135.6°C) during VENT mode | critical |
C576 | Door Mode Steam Temperature Over Limit | Steam temperature exceeds 276.1°F (135.6°C) during DOOR OPEN mode | critical |
C577 | Dry Mode Steam Temperature Over Limit | Steam temperature exceeds 276.1°F (135.6°C) during DRY OPEN mode | critical |
C633 | Heat-Up Mode Pressure Low | Chamber pressure fails to reach 1 psig in 1 minute or 4.93 psig in 16 minutes after air valve closes | critical |
C642 | Fill Mode Pressure High; Items Not Sterile | After filling begins, gauge pressure inside chamber equals or exceeds 2.18 PSIG (15 kPa) | warning |
C645 | Vent Mode Pressure High; Items Not Sterile | Absolute pressure inside chamber exceeds 0.73 PSIG (5 kPa) for longer than 5 minutes | warning |
C647 | Dry Mode Pressure High | Gauge pressure inside chamber equals or exceeds 2.18 PSIG (15 kPa) during DRY mode | warning |
C660 | Power Up Mode Pressure Hardware | Pressure A/D converter reports average value outside limits for normal operation during POWER UP mode | critical |
C661 | Select Mode Pressure Hardware | Pressure A/D converter reports average value outside limits during SELECT mode | critical |
C662 | Fill Mode Pressure Hardware | Pressure A/D converter reports average value outside limits during FILL mode | critical |
C663 | Heatup Mode Pressure Hardware | Pressure A/D converter reports average value outside limits during HEATUP mode | critical |
C664 | Sterilize Mode Pressure Hardware | Pressure A/D converter reports average value outside limits during STERILIZE mode | critical |
C665 | Vent Mode Pressure Hardware | Pressure A/D converter reports average value outside limits during VENT mode | critical |
C666 | Door Mode Pressure Hardware | Pressure A/D converter reports average value outside limits during DOOR mode | critical |
C667 | Dry Mode Pressure Hardware | Pressure A/D converter reports average value outside limits during DRY mode | critical |
C670 | Power Up Mode Pressure Overlim | Gauge pressure inside chamber exceeds 34.8 PSIG (240 kPa) during POWER UP mode | critical |
C671 | SELECT MODE pressure over limit | Chamber pressure exceeds 34.8 PSIG (240 kPa) during SELECT/FILL mode | error |
C672 | FILL MODE pressure over limit | Chamber pressure exceeds 34.8 PSIG (240 kPa) during FILL mode | error |
C673 | HEATUP MODE pressure over limit | Chamber pressure exceeds 34.8 PSIG (240 kPa) during HEATUP mode | error |
C674 | STERILIZE MODE pressure over limit | Chamber pressure exceeds 34.8 PSIG (240 kPa) during STERILIZE mode | error |
C675 | Vent Mode Pressure Overlim; Items Not Sterile | Gauge pressure inside chamber exceeds 34.8 PSIG (240 kPa) during VENT mode | critical |
C676 | Door Mode Pressure Overlim; Items Not Sterile | Gauge pressure inside chamber exceeds 34.8 PSIG (240 kPa) during DOOR mode | critical |
C677 | Dry Mode Pressure Overlim | Gauge pressure inside chamber exceeds 34.8 PSIG (240 kPa) during DRY mode | critical |
C980 | Power Up Mode High Limit Open | Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during POWER UP mode | critical |
C981 | Select Mode High Limit Open | Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during SELECT mode | critical |
C982 | Fill Mode High Limit Open | Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during FILL mode | critical |
C983 | Heatup Mode High Limit Open | Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during HEATUP mode | critical |
C984 | Sterilize Mode High Limit Open | Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during STERILIZE mode | critical |
C985 | Vent Mode High Limit Open | Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during VENT mode | critical |
C986 | Door Mode High Limit Open | Steam and/or Auxiliary High Limit switch opened for at least 1/4 second during DOOR mode | critical |
C987 | Dry Mode High-Limit Open | Steam and/or Aux. High Limit switch opened for at least 1/4 second during DRY mode | critical |
PERFORM MONTHLY MAINTENANCE | Monthly maintenance reminder | Cycle count reaches a multiple of 160 | info |
PERFORM WEEKLY MAINTENANCE | Weekly maintenance reminder | Cycle count reaches a multiple of 40 | info |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI