Midmark 413: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Midmark 413, with 34 ranked causes, 33 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
SEAT UP and SEAT DOWN functions do not work caution
When SEAT DOWN or SEAT UP footswitch is depressed, nothing happens.
Ranked causes
- 1. Wire connections loose (common)
- 2. Treatment pan assembly not fully stowed (occasional)
- 3. Pan safety limit switch malfunctioning (occasional)
- 4. Seat capacitor weak or blown (occasional)
- 5. Thermal overload switch activated (occasional)
- 6. Seat actuator assembly malfunctioning (occasional)
Check wiring connections and voltage
- Test point
- SEAT DOWN footswitch to terminal block connections
- Expected
- proper voltage at connections VAC, +/- 5 VAC
- Tools
- multimeter
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BASE DOWN function does not work caution
When BASE DOWN footswitch is depressed, nothing happens.
Ranked causes
- 1. BASE DOWN footswitch malfunctioning (common)
- 2. BASE UP footswitch sticking open (occasional)
- 3. Base down limit switch tripped (common)
- 4. Base down limit switch malfunctioning (occasional)
- 5. Seat up limit switch malfunctioning (occasional)
Perform continuity check on BASE DOWN footswitch
- Test point
- BASE DOWN footswitch COM. and N.O. terminals
- Expected
- continuity when activated
- Tools
- continuity tester or multimeter
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BASE UP and BASE DOWN functions do not work caution
When BASE UP or BASE DOWN footswitch is depressed, nothing happens.
Ranked causes
- 1. Wire connections loose (common)
- 2. Base capacitor weak or blown (occasional)
- 3. Thermal overload switch in base actuator motor activated (occasional)
- 4. Base actuator assembly malfunctioning (occasional)
Check wiring connections and voltage
- Test point
- BASE UP footswitch to terminal block connections
- Expected
- proper voltage at connections VAC, +/- 5 VAC
- Tools
- multimeter
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chair will not operate caution
Foot control footswitch is depressed, actuator motor does not run or hum.
Ranked causes
- 1. Power cord not plugged in (common)
- 2. Facility circuit breaker tripped (common)
- 3. Fuses blown in AC connector (export models) (occasional)
- 4. Wire connections loose (occasional)
Verify power cord connection
- Test point
- facility wall outlet
- Expected
- power cord plugged in
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chair moves slowly or not at all for heavy patient patient safety
Chair moves fine for light patient, but will not move or moves slowly for very heavy patient.
Ranked causes
- 1. Low voltage supplied to chair (common)
- 2. Table overloaded with too heavy patient (common)
- 3. Capacitor weak (occasional)
Measure voltage at wall receptacle
- Test point
- wall receptacle where chair is plugged
- Expected
- 115 +/- 5 VAC (domestic) or 230 +/- 10 VAC (export), +/- 5 VAC
- Tools
- multimeter or voltmeter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SEAT DOWN function does not work caution
When SEAT DOWN footswitch is depressed, nothing happens.
Ranked causes
- 1. SEAT DOWN footswitch is malfunctioning (common)
- 2. Seat down limit switch is tripped (common)
- 3. Seat down limit switch is malfunctioning (occasional)
Perform continuity check on SEAT DOWN footswitch
- Test point
- SEAT DOWN footswitch COM. and N.O. terminals
- Expected
- continuity when footswitch activated
- Tools
- continuity tester or multimeter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BASE UP function does not work caution
When BASE UP footswitch is depressed, nothing happens.
Ranked causes
- 1. BASE UP footswitch malfunctioning (common)
- 2. Gas spring malfunctioning (occasional)
Perform continuity check on BASE UP footswitch
- Test point
- BASE UP footswitch COM. and N.O. terminals
- Expected
- continuity when activated
- Tools
- continuity tester or multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SEAT UP function does not work caution
When SEAT UP footswitch is depressed, nothing happens.
Ranked causes
- 1. SEAT UP footswitch is malfunctioning (common)
- 2. SEAT DOWN footswitch stuck open (occasional)
Perform continuity check on SEAT UP footswitch
- Test point
- SEAT UP footswitch COM. and N.O. terminals
- Expected
- continuity when footswitch activated
- Tools
- continuity tester or multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot extension locking mechanism malfunctioning caution
Foot extension does not lock into stowed position or extended position automatically.
Ranked causes
- 1. Foot extension lock/unlock mechanism malfunctioning (common)
Inspect return spring and linking rod
- Test point
- foot extension lock/unlock mechanism
- Expected
- return spring and linking rod present and intact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pan safety limit switch misalignment patient safety
Pan safety limit switch trip arm does not trip at correct position, causing pan holder to lock before or after switch activation.
Ranked causes
- 1. Incorrect pan safety limit switch position (common)
Observe switch trip timing during pan insertion
- Test point
- Pan holder assembly fully stowed position
- Expected
- Trip arm of switch trips just before pan holder locks into stowed position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Seat actuator installation alignment patient safety
Seat actuator swivel joint not properly aligned, causing seat section to not be level with floor.
Ranked causes
- 1. Swivel joint screw adjustment needed (common)
Test seat level after installation
- Test point
- Seat section relative to floor
- Expected
- Seat section level with floor and parallel spacing
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Seat up limit switch misalignment patient safety
Seat up limit switch trip arm not properly positioned in actuator shaft cutout, causing switch to trip at wrong time.
Ranked causes
- 1. Incorrect switch installation or adjustment (common)
Verify trip arm position in cutout
- Test point
- Seat actuator shaft cutout
- Expected
- Trip arm of switch in cutout of seat actuator shaft
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Whirling or squeaking noise from actuator caution
Whirling or squeaking noise is heard when an actuator assembly is being run.
Ranked causes
- 1. Foreign matter on ball screw or lack of lubricant (common)
Inspect ball screw threads
- Test point
- actuator assembly ball screw threads
- Expected
- clean, no foreign matter, adequate lubricant
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
SEAT_NOT_LEVEL | Seat section is not level with the floor after installation | Seat actuator assembly installed with swivel joint at incorrect angle | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI