Midmark 411: troubleshooting and error codes
The ServiceForge knowledge base holds 24 documented symptoms for the Midmark 411, with 78 ranked causes, 20 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Foot up and down functions do not work
FOOT UP and FOOT DOWN buttons pressed but table does not move; all other functions operate normally
Ranked causes
- 1. Pan holder assembly not fully inserted; pan safety limit switch not tripped (common)
- 2. Foot capacitor weak or blown (common)
- 3. Thermal overload switch in foot actuator activated (common)
- 4. 5 amp FOOT fuse blown on interface board (common)
- 5. Foot actuator assembly malfunctioning (occasional)
- 6. Wire connections loose (common)
- 7. Pan safety switch out of adjustment (occasional)
- 8. Pan safety limit switch malfunctioning (occasional)
- 9. Interface board malfunctioning (occasional)
Check that pan holder assembly is fully inserted
- Test point
- Pan holder assembly position
- Expected
- Pan holder fully pushed in
3 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table does not operate caution
When any hand control or console button is pressed (up/down/auto-return/programmed position), nothing happens - actuator motor does not run or hum
Ranked causes
- 1. Power cord not plugged into facility wall outlet (common)
- 2. CONTROL PANEL POWER switch is OFF (common)
- 3. Facility circuit breaker providing power to table is tripped (common)
- 4. Wire connections loose (occasional)
- 5. 0.15 amp transformer primary fuse on interface board is blown (occasional)
- 6. Interface board is malfunctioning (occasional)
- 7. PC logic board is malfunctioning (occasional)
Verify power cord is plugged in
- Test point
- facility wall outlet
- Expected
- power cord connected
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
BACK UP and BACK DOWN functions do not work caution
When BACK UP and BACK DOWN buttons are pressed, table will not move; all other functions work
Ranked causes
- 1. BACK capacitor is weak or blown (common)
- 2. Thermal overload switch in back actuator is activated (occasional)
- 3. 5 amp BACK fuse is blown (occasional)
- 4. Back actuator assembly is malfunctioning (occasional)
- 5. Wire connections loose to back actuator (occasional)
- 6. Interface board is malfunctioning (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table up and down functions do not work
TABLE UP and TABLE DOWN buttons pressed but table does not move; all other functions operate normally
Ranked causes
- 1. Base capacitor weak or blown (common)
- 2. Thermal overload switch in base actuator activated (common)
- 3. 5 amp TABLE fuse blown on interface board (common)
- 4. Base actuator assembly malfunctioning (occasional)
- 5. Wire connections loose (common)
- 6. Interface board malfunctioning (occasional)
Perform continuity check on 5 amp TABLE fuse
- Test point
- 5 amp TABLE fuse on interface board
- Expected
- Continuity present (not blown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
TILT UP and TILT DOWN functions do not work caution
When TILT UP and TILT DOWN buttons are pressed, table will not move; all other functions work
Ranked causes
- 1. Tilt capacitor is weak or blown (common)
- 2. Thermal overload switch in tilt actuator is activated (occasional)
- 3. 5 amp TILT fuse is blown (occasional)
- 4. Tilt actuator assembly is malfunctioning (occasional)
- 5. Wire connections loose to tilt actuator (occasional)
- 6. Interface board is malfunctioning (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Programming function does not work
Table position cannot be programmed into memory; when PROGRAMMED POSITION button is pressed, table does not move to proper stored position
Ranked causes
- 1. Receiver and transmitter not aligned or malfunctioning; signal lost (common)
- 2. RF cables shorted or open (occasional)
- 3. Program panel malfunctioning; switch membrane malfunctioning (occasional)
- 4. Program panel interface board malfunctioning (occasional)
- 5. PC logic board malfunctioning (occasional)
Check L.E.D. on PC logic board for function (D1-Base, D2-Foot, D3-Back, D4-Tilt)
- Test point
- L.E.D. on PC logic board
- Expected
- L.E.D. not illuminated
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table moves fine for light patient but slowly or not for heavy patient patient safety
Table operates normally with light patients but will not move or moves slowly with very heavy patients
Ranked causes
- 1. Low voltage supplied to table (common)
- 2. Gas spring weak (common)
- 3. Table overloaded with patient exceeding weight capacity (common)
- 4. Capacitor for suspect function weak (occasional)
Check voltage at wall receptacle
- Test point
- Wall receptacle
- Expected
- 115 +/- 5 VAC, +/- 5 VAC
- Tools
- Multimeter
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto return function does not work
When AUTO RETURN START button is pressed, nothing happens
Ranked causes
- 1. Program panel malfunctioning; switch membrane malfunctioning (occasional)
- 2. Program panel interface board malfunctioning (occasional)
- 3. Base down limit switch malfunctioning; stuck open (occasional)
Perform continuity check on base down limit switch
- Test point
- Base down limit switch
- Expected
- Closed
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back up or down function does not work asymmetrically
One of back up/down functions operates properly but the other does not; all other functions work
Ranked causes
- 1. Wire connections loose to back actuator assembly (common)
- 2. Interface board malfunctioning; relay for up or down function malfunctioning (occasional)
- 3. Back actuator assembly malfunctioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive sideways play of table top patient safety
Table top is not stable and can be moved from side to side
Ranked causes
- 1. Chain assemblies loose (common)
- 2. Base slide assembly worn or deformed (occasional)
- 3. Table base not resting firmly on floor (common)
Check tension of chain assemblies
- Test point
- Chain assemblies
- Expected
- Proper tension
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot up or down function does not work asymmetrically
One of foot up/down functions operates properly but the other does not; all other functions work
Ranked causes
- 1. Wire connections loose to foot actuator assembly (common)
- 2. Interface board malfunctioning; relay for up or down function malfunctioning (occasional)
- 3. Foot actuator assembly malfunctioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more functions cannot be initiated from console caution
Some functions may be initiated from console but at least one may not; all functions can be initiated from hand control
Ranked causes
- 1. Program panel switch membrane is malfunctioning (common)
- 2. Program panel interface board is malfunctioning (common)
- 3. Wire connections loose (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more functions cannot be initiated from hand control caution
Some functions may be initiated from hand control but at least one may not; all functions can be initiated from console
Ranked causes
- 1. Hand control panel switch membrane is malfunctioning (common)
- 2. Hand control board is malfunctioning (common)
- 3. Hand control coil cord has bad connection or break (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table up or down function does not work asymmetrically
One of table up/down functions operates properly but the other does not; all other functions work
Ranked causes
- 1. Wire connections loose to base actuator (common)
- 2. Interface board malfunctioning; relay for up or down function malfunctioning (occasional)
- 3. Base actuator assembly malfunctioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tilt up or down function does not work asymmetrically
One of tilt up/down functions operates properly but the other does not; all other functions work
Ranked causes
- 1. Wire connections loose to tilt actuator assembly (common)
- 2. Interface board malfunctioning; relay for up or down function malfunctioning (occasional)
- 3. Tilt actuator assembly malfunctioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Base actuator continues running after table lowered all the way
After base actuator has lowered table top all the way down, the base actuator continues to run
Ranked causes
- 1. Base down limit switch out of adjustment (common)
- 2. Base down limit switch malfunctioning; stuck closed (occasional)
Check adjustment of base down limit switch
- Test point
- Base down limit switch position
- Expected
- Properly adjusted to stop motor at bottom
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more functions cannot be initiated from foot control caution
Some functions may be initiated from foot control but at least one may not; all functions can be initiated from hand control and console
Ranked causes
- 1. Foot switch for non-operable function is malfunctioning (common)
- 2. Wire connections loose (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Function drifts by itself
Base, tilt, back, or foot function drifts by itself; function operates properly otherwise
Ranked causes
- 1. Motor actuator brake malfunctioning (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Headrest drifts downward while supporting patient head patient safety
Headrest drifts downward while supporting patient head
Ranked causes
- 1. Headrest out of adjustment (common)
Check adjustment of headrest
- Test point
- Headrest mechanism
- Expected
- Properly adjusted to prevent drift
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Headrest locking handles pop out of locked position
Headrest locking handles pop out of locked position
Ranked causes
- 1. Headrest handles handle stops need adjusted (common)
Check adjustment of headrest handles handle stops
- Test point
- Headrest handles handle stops
- Expected
- Stops positioned to keep handles locked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stirrups do not lock into one of three positions properly patient safety
Stirrups are malfunctioning; do not lock into one of the three positions properly
Ranked causes
- 1. Stirrup index pin damaged or needs replacement (common)
Check for damaged parts on stirrup assembly
- Test point
- Stirrup index pin and associated components
- Expected
- No visible damage; pin intact and functioning
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table does not operate from hand control only caution
Some hand control or console button is pressed but nothing happens; all functions work from foot control
Ranked causes
- 1. PC logic board is malfunctioning (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Table up does not work but table down does
TABLE UP function does not operate; TABLE DOWN function operates normally
Ranked causes
- 1. Gas spring weak (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Whirling or squeaking noise from actuator assembly
Whirling or squeaking noise is heard when an actuator assembly is being run
Ranked causes
- 1. Foreign matter on ball screw threads and/or lack of lubricant (common)
Check for foreign matter and lubricant on ball screw threads
- Test point
- Ball screw threads on actuator
- Expected
- Clean and well lubricated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI