Medtronic Puritan Bennett 980: troubleshooting and error codes

The ServiceForge knowledge base holds 80 documented symptoms for the Medtronic Puritan Bennett 980, with 108 ranked causes, 134 diagnostic checks with test points and expected values, and 369 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Proximal flow system alert patient safety

Proximal pressure or flow subsystem alert triggered during SST diagnostics

Ranked causes

  1. 1. SOL2 malfunction or failure (common)
  2. 2. Improper tubing connection from SOL2 to proximal flow system (common)
  3. 3. R4 resistor missing (rare)
  4. 4. Improper connections at proximal flow accumulator (occasional)
  5. 5. Options host PCBA failure (rare)
  6. 6. Proximal flow PCBA failure (rare)
First check

Wait for proximal unit to enter idle mode

Expected
Prox in idle mode

11 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor accumulator pressure too low

EST diagnostic code FE2309 reported; compressor fails to build or maintain target accumulator pressure.

Ranked causes

  1. 1. Kinked or leaking pressure transducer tubing (common)
  2. 2. Improperly connected compressor-to-BDU interconnect hose (common)
  3. 3. Restricted inlet muffler (occasional)
  4. 4. Faulty compressor interface PCBA (occasional)
  5. 5. Faulty compressor assembly (rare)
First check

Inspect pressure transducer tubing for kinks or leaks

Test point
Pressure transducer tubing and connections
Expected
Tubing clear, no kinks, no leaks visible

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Proximal flow sensor alert AS0707

Purge suspended, user prompted to remove inspiratory filter and connect tubing.

Ranked causes

  1. 1. Proximal flow sensor tubing connection improper (common)
  2. 2. Proximal flow PCBA connection loose (occasional)
  3. 3. Proximal flow PCBA failure (rare)
  4. 4. Options host PCBA failure (rare)
First check

Run proximal autozero test

Test point
Proximal pressure sensor
Expected
Autozero test passes

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Short self test (SST) fails

VBS resistance/compliance calculation or leak detection fails; common causes include inadequate gas sources, leaks, or occlusions

Ranked causes

  1. 1. Inadequate air or oxygen gas source pressure (common)
  2. 2. Leaks in breathing system (common)
  3. 3. Occlusions in system (common)
  4. 4. Ventilator not at operating temperature (occasional)
First check

Warm up ventilator to operating temperature

Test point
Service mode
Expected
15 minutes elapsed since entering Service mode

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SST fails or produces inaccurate results

Short Self Test does not complete successfully or returns incorrect resistance, compliance, or leak detection values

Ranked causes

  1. 1. Inadequate air or oxygen gas source pressure (common)
  2. 2. Leaks in breathing system (common)
  3. 3. Occlusions in breathing system (common)
  4. 4. Ventilator not at operating temperature (occasional)
First check

Verify air and oxygen source pressures

Test point
air and oxygen inlets
Expected
241 kPa to 599 kPa (35 psi to 87 psi), 241–599 kPa

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor failed to establish stable flow

EST diagnostic code FE2505 reported; compressor unable to produce consistent, stable airflow output.

Ranked causes

  1. 1. Damaged o-rings on pneumatic interface manifold (common)
  2. 2. Faulty compressor assembly (occasional)
  3. 3. Faulty compressor interface PCBA (occasional)
First check

Inspect three o-rings on pneumatic interface manifold

Test point
Pneumatic interface manifold, three o-ring positions
Expected
All o-rings intact, no visible damage or extrusion

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor Performance error: Target pressure not reached

EST diagnostic code FE2504 reported; compressor cannot achieve target pressure during test.

Ranked causes

  1. 1. Leaks or occlusions (common)
  2. 2. Faulty compressor interface PCBA (occasional)
  3. 3. Faulty compressor assembly (occasional)
First check

Inspect system for leaks or occlusions

Test point
All pressure hoses and connections
Expected
No visible leaks, no obstructions in lines

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor unable to hold constant accumulator pressure

EST diagnostic code FE2405 reported; compressor builds pressure but cannot maintain it, indicating leakage or regulation failure.

Ranked causes

  1. 1. Leaks or occlusions in system (common)
  2. 2. Compressor motor not turning on (occasional)
  3. 3. Faulty compressor interface PCBA (occasional)
First check

Inspect system for leaks or occlusions

Test point
All pressure hoses and connections
Expected
No visible leaks, no obstructions in lines

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power not detected during compressor performance test

EST diagnostic code FE2524 reported; compressor performance test fails because mains AC power is not being detected or supplied to the ventilator.

Ranked causes

  1. 1. No AC power connected (common)
  2. 2. AC mains circuit breaker tripped (common)
First check

Verify AC power present at wall outlet

Test point
AC mains outlet
Expected
AC voltage present (nominal 120V or 240V depending on region)
Tools
DMM or power tester

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm silence key LED failure AE0808

Audio paused key LED on bezel does not illuminate during off-screen key test.

Ranked causes

  1. 1. Alarm silence key LED not functioning (occasional)
  2. 2. Ribbon cable between user interface PCBA and touch screen loose (common)
First check

Press audio paused key

Test point
Alarm silence key on bezel
Expected
LED illuminates when pressed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Atmospheric Pressure

Barometric pressure reading must be entered from external barometer; calibration may fail if incorrect pressure value provided or sensor drift occurs

Ranked causes

  1. 1. Incorrect barometric pressure value entered (common)
  2. 2. Sensor calibration drift or failure (rare)
First check

Obtain accurate barometric pressure reading from external barometer

Test point
External barometer
Expected
Stabilized reading in mmHg

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen lock key LED failure AE0807

Screen lock key LED on bezel does not illuminate during off-screen key test.

Ranked causes

  1. 1. Screen lock key LED not functioning (occasional)
  2. 2. Ribbon cable between user interface PCBA and touch screen loose (common)
First check

Press screen lock key

Test point
Screen lock key on bezel
Expected
LED illuminates when pressed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power not detected high voltage

Flow sensor or circuit pressure test failure; AC mains power not detected

Ranked causes

  1. 1. AC power not connected or circuit breaker tripped (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Atmospheric Pressure

Atmospheric pressure sensor calibration verification

Ranked causes

  1. 1. Calibration required as part of annual PM (unknown)
First check

Atmospheric Pressure calibration

Test point
Service Mode
Expected
Pass

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Atmospheric pressure transducer

Atmospheric pressure transducer requires calibration after annual or 10,000-hour preventive maintenance

Ranked causes

  1. 1. Routine preventive maintenance (common)
First check

Perform ventilator warm-up cycle

Test point
Ventilator startup
Expected
Warm-up completed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery cover slider latch replacement

Replace battery cover slider latch and reattach extension spring

Ranked causes

  1. 1. Battery cover slider latch worn or damaged (occasional)
First check

Remove screws from battery cover slider latch retainer

Test point
Two screws on slider latch retainer
Expected
Screws removed while retaining slider latch

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery door panel alignment

Align battery door panel to BDU assembly and tighten mounting screws

Ranked causes

  1. 1. Battery door panel misalignment or loose screws (common)
First check

Verify battery door panel screws are seated properly

Test point
Top two screws of battery panel door assembly
Expected
Screws fully tightened and flush

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Component Calibration patient safety

Perform component calibrations as specified in Section 4.5.2. Results must be documented on the PVR form.

Ranked causes

  1. 1. Component calibration procedure required (common)
First check

Confirm component calibration completion

Test point
PVR form
Expected
Component calibration section completed and signed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Exhalation Valve

Exhalation valve calibration verification

Ranked causes

  1. 1. Calibration required as part of annual PM (unknown)
First check

Exhalation Valve calibration

Test point
Service Mode
Expected
Pass

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Exhalation Valve

Exhalation valve calibration to build valve table; requires gold standard test circuit, proper warm-up time, and EST leak test passage.

Ranked causes

  1. 1. Exhalation valve requires calibration (common)
First check

Allow ventilator warm-up time if not at operating temperature from recent usage

Test point
Ventilator operating temperature
Expected
At least 15 minutes warm-up in Service mode

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Exhalation Valve Calibration patient safety

Calibration procedure for exhalation valve on Puritan Bennett 980 ventilator using on-screen prompts in Service mode

Ranked causes

  1. 1. System leak or improper test circuit attachment (common)
First check

Ensure exhalation filter condensate vial is secured tightly

Test point
Exhalation filter condensate vial
Expected
Secured tightly

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Extended Self Test (EST) patient safety

Routine EST procedure to verify ventilator function before clinical use. Must complete all tests in sequence without interruption using Run ALL button. All tests must pass with no failures.

Ranked causes

  1. 1. Routine EST execution required before clinical use (common)
First check

Ensure adult exhalation filter is installed before running EST

Test point
Exhalation filter connection
Expected
Adult exhalation filter installed

1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow Sensor

Flow sensor calibration verification

Ranked causes

  1. 1. Calibration required as part of annual PM (unknown)
First check

Flow Sensor calibration

Test point
Service Mode
Expected
Pass

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Oxygen sensor

Oxygen sensor requires calibration after maintenance, replacement, or extended self-test

Ranked causes

  1. 1. Routine maintenance or oxygen sensor replacement (common)
First check

Perform ventilator warm-up cycle

Test point
Ventilator startup
Expected
Warm-up completed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Oxygen Sensor

Oxygen sensor calibration verification

Ranked causes

  1. 1. Calibration required as part of annual PM (unknown)
First check

Oxygen Sensor calibration

Test point
Service Mode
Expected
Pass / Fail / Limited Use

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Oxygen Sensor Calibration patient safety

Calibration of oxygen sensor performed in normal ventilation mode after ventilator reaches operating temperature

Ranked causes

  1. 1. Oxygen sensor needs calibration (unknown)
First check

Allow ventilator to reach operating temperature

Test point
Ventilator warm-up time
Expected
At least 15 minutes in Service mode

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Oxygen sensor installation patient safety

Step-by-step procedure to install a replacement oxygen sensor with proper seating and cable reconnection

Ranked causes

  1. 1. Standard replacement procedure (common)
First check

Check o-ring presence

Test point
Oxygen sensor
Expected
O-ring present

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PTS 2000 preparation patient safety

Setup procedure for PTS 2000 sensor in ventilator circuit prior to performance verification or calibration checks.

Ranked causes

  1. 1. PTS 2000 must be integrated into patient circuit for measurement (common)
First check

Install PTS 2000 sensor in circuit

Test point
Patient circuit connection
Expected
PTS 2000 connected in series with two longest sections per Figure 11

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: SST temperature requirement

Ventilator warm-up procedure to ensure accurate SST results

Ranked causes

  1. 1. Cold ventilator from recent usage (unknown)
First check

Allow warm-up time

Test point
Service mode
Expected
15 minutes elapsed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch screen

Touch screen requires calibration after annual or 10,000-hour preventive maintenance

Ranked causes

  1. 1. Routine preventive maintenance (common)
First check

Perform ventilator warm-up cycle

Test point
Ventilator startup
Expected
Warm-up completed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch Screen

Touch screen calibration verification

Ranked causes

  1. 1. Calibration required as part of annual PM (unknown)
First check

Touch Screen calibration

Test point
Service Mode
Expected
Pass

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Ventilator date, time and settings after coin battery replacement

After replacing coin batteries, ventilator date/time and programmed settings must be verified to persist across power cycles to ensure correct operation.

Ranked causes

  1. 1. Coin batteries replaced, requiring date/time reset (common)
First check

Power on ventilator and enter date/time configuration

Test point
Configuration icon (wrench) and Date/Time Change tab
Expected
Date/Time Change tab accessible and editable

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Ventilator Performance Verification patient safety

Complete electrical safety testing and component calibration before performance verification. Document results on PVR form with software/firmware versions, serial numbers, and safety test measurements.

Ranked causes

  1. 1. Verification procedure setup (common)
First check

Ground resistance measurement

Test point
Ground integrity
Expected
<0.20 Ω, <0.20

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Ventilator Setup for PTS 2000 patient safety

Warm up and configure ventilator with exhalation filter, inspiratory filter, and PTS 2000 sensor in patient circuit; connect gas and power supplies; verify indicators and battery; perform EST, SST and calibrations.

Ranked causes

  1. 1. Ventilator requires warm-up and configuration before calibration (common)
First check

Warm up ventilator in Service mode

Test point
Service mode indicator
Expected
Minimum 15 minutes elapsed

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

EST test failure or diagnostic code patient safety

EST test fails to complete or returns a diagnostic code indicating a problem detected during testing.

Ranked causes

  1. 1. One or more EST tests failed or diagnostic code returned (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

EV performance unable to establish stable flow

EV performance test failure; unable to establish stable exhalation flow

Ranked causes

  1. 1. EVQ fault, fouled, or flow sensor miscalibration (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation auto zero solenoid not operational

Circuit pressure test failure; exhalation pressure offset solenoid not functioning

Ranked causes

  1. 1. Leaks, disconnects, occlusions, or solenoid/PCBA fault (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation filter unable to establish stable flow

Exhalation filter test failure; unable to establish stable flow with filter installed

Ranked causes

  1. 1. Exhaust filter fault or EVQ issue (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation flow sensor out of range

Flow sensor cross-check failure; exhalation flow sensor reading out of acceptable range

Ranked causes

  1. 1. Patient circuit leaks, improper EVQ seating, or faulty EVQ (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve not calibrated

EV performance test failure; exhalation valve requires calibration

Ranked causes

  1. 1. Exhalation valve calibration not completed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve pressure out of range

EV performance test failure; measured pressure outside acceptable range

Ranked causes

  1. 1. Exhaust filter occluded, o-ring missing, EVQ fault, or F5 filter issue (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Expiratory zero offset out of range

Circuit pressure test failure; expiratory zero pressure offset exceeds acceptable limits

Ranked causes

  1. 1. Pressure sensor calibration drift (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Extended self test (EST) fails

EST verifies ventilator operational subsystem integrity; failure indicates possible system component or functional issue

Ranked causes

  1. 1. SST not passing (indicates underlying system problem) (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow sensor calibration failure

Calibration procedure fails to complete successfully

Ranked causes

  1. 1. System leak present (common)
First check

Inspect exhalation filter for leaks

Test point
Exhalation filter collector vial and connections
Expected
No leaks detected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ground isolation resistance too low high voltage

Ground isolation resistance measurement fails to meet minimum threshold during electrical safety testing

Ranked causes

  1. 1. Faulty assembly, subassembly, cable or harness causing low isolation resistance (unknown)
First check

Measure ground isolation resistance

Test point
DB9 connector female mating face
Expected
> 975 Ω or > 0.975 kΩ, > specified value
Tools
DMM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory and exhalation pressure difference outside limits

Circuit pressure test failure; inspiratory and exhalation pressures differ beyond acceptable range

Ranked causes

  1. 1. Leaks, disconnects, occlusions, or pressure sensor fault (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory and safety valve pressure difference outside limits

Circuit pressure test failure; inspiratory and safety valve pressures differ beyond tolerance

Ranked causes

  1. 1. Leaks, disconnects, occlusions, or pressure sensor fault (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory auto zero solenoid not operational

Circuit pressure test failure; inspiratory pressure offset solenoid not functioning

Ranked causes

  1. 1. Leaks, disconnects, occlusions, or solenoid/PCBA fault (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory flow sensor out of range

Flow sensor cross-check failure; inspiratory flow sensor reading out of acceptable range

Ranked causes

  1. 1. Inadequate source gas supply (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory zero offset out of range

Circuit pressure test failure; inspiratory zero pressure offset exceeds acceptable limits

Ranked causes

  1. 1. Pressure sensor calibration drift (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

LVDT temperature out of range

EV performance test failure; motor temperature sensor reading out of range

Ranked causes

  1. 1. EVQ or exhalation valve assembly fault (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Mix flow sensor out of range

Flow sensor cross-check failure; mix flow sensor reading out of acceptable range

Ranked causes

  1. 1. Inadequate source gas supply (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No air supply connected patient safety

SST detects no air input to the system during flow sensor cross check

Ranked causes

  1. 1. Air supply line not connected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No gas connected

Flow sensor or circuit pressure test failure; no gas supply detected

Ranked causes

  1. 1. Air or oxygen supply not connected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No oxygen supply connected patient safety

SST detects no oxygen input to the system during flow sensor cross check

Ranked causes

  1. 1. Oxygen supply line not connected (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen sensor requires replacement patient safety

Oxygen sensor has reached recommended one-year replacement interval or shows signs of malfunction; expiration date on new sensor packaging must not be past installation date

Ranked causes

  1. 1. Scheduled maintenance - one year from installation (common)
First check

Verify expiration date on new oxygen sensor packaging

Test point
Sensor packaging
Expected
Installation date must be before or on expiration date

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Piezo alarm cable incorrectly connected

Alarm cables connected to wrong polarity terminals during BDU top cap assembly replacement

Ranked causes

  1. 1. Incorrect polarity connection during reassembly (occasional)
First check

Verify piezo alarm cable polarity matches label markings

Test point
Piezo alarm cable terminals
Expected
Positive (+) connected to red terminal, negative (-) connected to black terminal

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Post coin battery replacement system verification tests

Series of mandatory tests to be performed after coin battery replacement before returning ventilator to patient use.

Ranked causes

  1. 1. Coin battery replacement completed, requiring verification (common)
First check

Perform ventilator warm-up cycle

Test point
Service mode operation
Expected
Ventilator running continuously for 15 minutes without errors

12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PSOL current out of range

Flow sensor cross-check failure; proportional solenoid current measurement out of acceptable range

Ranked causes

  1. 1. Inadequate source gas or patient circuit leaks (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SST proximal purge error

Alert displayed during SST proximal purge procedure test

Ranked causes

  1. 1. Options host PCBA or proximal flow PCBA fault (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • SST system leak
  • System leak detected
  • System leak failure
  • Target pressure not achieved
  • Test: Adult Short Self Test (SST)
  • Test: Earth Current Leakage (Normal Polarity, Closed Neutral)
  • Test: Earth Current Leakage (Normal Polarity, Open Neutral)
  • Test: Earth Current Leakage (Reverse Polarity, Closed Neutral)
  • Test: Earth Current Leakage (Reverse Polarity, Open Neutral)
  • Test: Extended Self Test (EST)
  • Test: External Ports
  • Test: Ground Isolation Resistance
  • Test: Ground Resistance
  • Test: Neonatal Short Self Test (SST)
  • Test: Oxygen Sensor Calibration Verification
  • Test: Pediatric Short Self Test (SST)
  • Test: Ventilator Inop
  • Unable to establish pressure in circuit leak test
  • Unable to establish stable flow
  • Vent inop test

20 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
AE0602Delivery pressure solenoid leak exceeds alert limitEST Leak Test on delivery PSOLerror
AE0605Test circuit not connected to ventilatorEST Leak Test start without test circuit connectionerror
AE0607Delivery PSOL leak alertForward leakage on delivery PSOL exceeds alert limit during EST leak testwarning
AE0807Screen lock key LED failureOff-screen key LED on screen lock key does not illuminate during testerror
AE0808Alarm silence key LED failureOff-screen key LED on audio paused key does not illuminate during testerror
AE0809Key-panel LED backlight brightness not changingOff-screen Key Test executionwarning
AE1301Delivery PSOL current feedback reading out of rangeEST Delivery PSOL Test executionerror
AE1501Exhalation valve loopback current out of rangeEST Exhalation Valve Loopback Test executionerror
AE1604Exhalation valve pressure measurement out of rangeEST EV Pressure Accuracy & Seal Testerror
AE1901Serial port communication malfunctionCommunication loopback testerror
AE2015Ventilator primary battery charge below acceptable levelVentilator Battery EST during startuperror
AE2100Compressor battery not installed; test skippedCompressor Battery ESTinfo
AE2115Compressor primary battery charge below acceptable levelCompressor Battery EST during startuperror
AE2300Compressor not detected; EST skippedCompressor ESTinfo
AE2301Compressor not detected; EST skippedCompressor ESTinfo
AE2307Compressor motor speed outside acceptable rangeCompressor performance monitoringerror
AE2400Compressor leak test not performed; compressor not detectedCompressor Leak ESTinfo
AE2401Compressor leak test not performed; compressor not detectedCompressor Leak ESTinfo
AE2500Compressor Performance EST not performed; compressor not detectedCompressor Performance ESTinfo
AE2501Compressor Performance EST not performed; compressor not detectedCompressor Performance ESTinfo
AE2506Compressor unable to maintain required accumulator pressureCompressor Performance ESTerror
AE2507Compressor unable to maintain required flow rateCompressor Performance ESTerror
AS0022No oxygen supply connectedSST flow sensor cross check detects no oxygen during testcritical
AS0023No air supply connectedSST flow sensor cross check detects no air during testcritical
AS0305System Leak AlertSST Leak test detects pressure drop over timewarning
AS0405Occluded expiratory filterSST Exhalation Filter test calculates excessive pressure dropwarning
AS0408Occluded expiratory compartmentSST Exhalation Filter test verifies pressure drop out of rangewarning
AS0411Low expiratory filter ΔPSST Exhalation Filter test shows low pressure dropwarning
AS0503Occluded inspiratory limbSST Circuit Resistance test detects excessive inspiratory limb pressure dropwarning
AS0506Occluded expiratory limbSST Circuit Resistance test detects excessive expiratory limb pressure dropwarning
AS0507Unable to reach min peak flowSST Circuit Resistance test cannot achieve minimum peak flowwarning
AS0511Unable to reach min peak flowSST Circuit Resistance test characterizes total resistance and failswarning
AS0513Insp limb resistance lowSST Circuit Resistance test shows low inspiratory limb resistancewarning
AS0514Exp limb resistance lowSST Circuit Resistance test shows low expiratory limb resistancewarning
AS0605Excessive complianceSST Circuit Compliance test detects high circuit compliancewarning
AS0607Compliance lowSST Circuit Compliance test detects low circuit compliancewarning
AS0609Excessive complianceSST Circuit Compliance test detects high circuit compliancewarning
AS0611Compliance lowSST Circuit Compliance test detects low circuit compliancewarning
AS0700SST: NOT INSTALLEDVentilator does not detect proximal flow module at startupinfo
AS0701SST Prox Comm. ErrorSST Prox system recognition test failserror
AS0703Demo Mode Square Wave FailedSST Prox proximal flow system communications test failserror
AS0704P-atm Cross Check FailedSST Prox barometric pressure readings between ventilator and prox module differerror
AS0705Autozero and Leak Test FailedSST Prox checks pneumatic path for functionalityerror
AS0705Proximal flow system alertProximal flow subsystem malfunction or out of specification condition detectederror
AS0707Pressure Cross Check FailedSST Prox pressure cross check of proximal flow system failserror
AS0707Inspiratory filter disconnection or proximal sensor tubing issueFilter removed or tubing disconnection detectedwarning
AS0707Proximal flow sensor alertProximal flow sensor tubing or connection problem detected during system self-testwarning
AS0708Flow Sensor Cross Check FailedSST Prox flow sensor cross check of proximal flow system failserror
AS0710Operator Skipped Prox TestProximal flow sensor connected but operator declines to run prox SSTinfo
AS0711Purge ErrorSST Prox purge procedure test failserror
AS0712SST proximal purge errorProximal purge procedure test failurewarning
AS0713SST proximal purge errorProximal purge procedure test failurewarning
AS0714SST proximal purge errorProximal purge procedure test failurewarning
AS0715SST proximal purge errorProximal purge procedure test failurewarning
AS0805System leak alertLeak test detects circuit leakagewarning
DXXXXXAssertion code (undefined content)Unknown assertion errorwarning
FE0001Air pressure supply solenoid leak rate exceeds specificationMix Leak Test on air supply circuitcritical
FE0002Oxygen supply solenoid leak rate exceeds specificationMix Leak Test on O2 supply circuitcritical
FE0004Mix circuit accumulator unable to reach required pressureMix circuit accumulator pressure testcritical
FE0005Mix circuit accumulator pressure drop exceeds acceptable limitMix circuit leak test - pressure hold phasecritical
FE0006Mix circuit unable to pressurize accumulatorMix circuit accumulator pressurization testcritical
FE0007Air inlet pressure exceeds limit after disconnectionMix circuit air inlet pressure verificationerror
FE0008Air side mix check valve reverse flow detectedMix circuit air check valve testerror
FE0009O2 side mix check valve reverse flow detectedMix circuit O2 check valve testerror
FE0010Oxygen inlet pressure exceeds limitMix circuit O2 inlet pressure verificationerror
FE0011Mix circuit unable to reach target accumulator pressure using O2Mix circuit O2-side accumulator pressurization testerror
FE0012Mix circuit exceeded target accumulator pressure using O2Mix circuit O2-side accumulator pressurization testcritical
FE0013Mix circuit unable to reach target accumulator pressure using airMix circuit air-side accumulator pressurization testerror
FE0014Mix circuit exceeded target accumulator pressure using airMix circuit air-side accumulator pressurization testcritical
FE0015Oxygen inlet pressure below acceptable limitMix circuit O2 inlet pressure testwarning
FE0016Inspiratory flow rate exceeds specificationMix circuit flow measurement testerror
FE0017Options gas supply accumulator pressure unstableMix circuit options gas supply testerror
FE0018Option Gas Supply solenoid leak rate exceeds specificationMix circuit options gas solenoid testerror
FE0021No air or oxygen supply connected to ventilatorMix circuit gas supply detection testcritical
FE0024AC power not detected by ventilatorMix circuit AC power detection testcritical
FE0101Air supply PSOL has excessive internal leakageMix PSOL lift-off testcritical
FE0102O2 PSOL has excessive leakMix PSOL liftoff testerror
FE0103Air PSOL current out of rangeMix PSOL liftoff testerror
FE0104O2 PSOL current out of rangeMix PSOL liftoff testerror
FE0107Air PSOL maximum liftoff current exceededMix PSOL liftoff testerror
FE0108O2 PSOL maximum liftoff current exceededMix PSOL liftoff testerror
FE0109Air PSOL liftoff current out of rangeMix PSOL liftoff testerror
FE0110O2 PSOL liftoff current out of rangeMix PSOL liftoff testerror
FE0111Mix air PSOL loopback current out of rangeMix PSOL loopback testerror
FE0121No gas connectedGas connection testerror
FE0124AC power not detectedAC connection testerror
FE0201Unable to stabilize accumulator pressurePressure sensor testerror
FE0203Accumulator pressure out of rangePressure sensor testerror
FE0204Overpressure switch failed to activatePressure switch testerror
FE0205Overpressure switch failed to deactivatePressure switch testerror
FE0221No gas connectedMix accumulator testerror
FE0224AC power not detectedMix accumulator testerror
FE0301Inspiratory zero offset out of rangePressure transducer auto-zero testerror
FE0302Expiratory zero offset out of rangePressure transducer auto-zero testerror
FE0303Target pressure not achievedPressure cross check testerror
FE0304Inspiratory and safety valve pressure sensor difference outside test limitsPressure cross check testerror
FE0305Inspiratory and exhalation pressure sensor difference outside test limitsPressure cross check testerror
FE0306Inspiratory auto zero solenoid not operationalAuto-zero solenoid testerror
FE0307Exhalation auto zero solenoid not operationalAuto-zero solenoid testerror
FE0321No gas connectedCircuit pressure testerror
FE0324AC power not detectedCircuit pressure testerror
FE0401Air inlet pressure sensor zero offset out of rangeZero offset pressure testerror
FE0402O2 inlet pressure exceededZero offset pressure testerror
FE0404Accumulator pressure sensor zero offset out of rangeZero offset pressure testerror
FE0405Inspiratory pressure sensor zero offset out of rangeZero offset pressure testerror
FE0406Exhalation pressure sensor zero offset out of rangeZero offset pressure testerror
FE0409Air mix flow sensor zero offset out of rangeFlow offset testerror
FE0410O2 mix flow exceededFlow offset testerror
FE0412Inspiratory flow sensor zero offset out of rangeFlow offset testerror
FE0413Expiratory flow sensor zero offset out of rangeFlow offset testerror
FE0414Air still connectedGas connection testwarning
FE0415O2 still connectedGas connection testwarning
FE0424AC power not detectedAC power connection testerror
FE0501Unable to establish stable flowFlow stability testerror
FE0502Inspiratory flow sensor out of rangeFlow sensor cross check testerror
FE0503Exhalation flow sensor out of rangeFlow sensor cross check testerror
FE0504Mix flow sensor out of rangeFlow sensor cross check testerror
FE0505PSOL current out of rangeEST Flow Sensor Cross Check test failureerror
FE0521No gas connectedEST Flow Sensor Cross Check test failureerror
FE0524AC power not detectedEST Flow Sensor Cross Check test failureerror
FE0601Delivery PSOL leak exceeds failure limitEST Leak Test failureerror
FE0603Unable to establish pressure during leak testEST Leak Test and check valve reverse flow testerror
FE0604Excessive leak during leak testEST Leak Test failureerror
FE0606Inspiratory check valve reverse flow too lowEST Leak Test check valve reverse flow testerror
FE0607Inspiratory check valve reverse flow too highEST Leak Test check valve reverse flow testerror
FE0608Unable to establish accumulator pressureEST Leak Test delivery PSOL leak testerror
FE0609Unable to establish stable flowEST Leak Test delivery PSOL leak testerror
FE0610Safety valve failed to openEST Leak Test delivery PSOL leak testcritical
FE0611Safety valve cracking pressure out of rangeEST Leak Test delivery PSOL leak testerror
FE0621No gas connected during leak testEST Leak Test gas connection testerror
FE0624AC power not detected during leak testEST Leak Test AC connection testerror
FE0701No media driver foundEST Internal Storage test failureerror
FE0702File open errorEST Internal Storage test failureerror
FE0703File write errorEST Internal Storage test failureerror
FE0704File read errorEST Internal Storage test failureerror
FE0705File cleanup errorEST Internal Storage test failureerror
FE0801Screen brightness key failedEST Off-screen Key Test failurewarning
FE0802Screen lock key failedEST Off-screen Key Test failurewarning
FE0803Manual inspiration key failedEST Off-screen Key Test failurewarning
FE0804Alarm reset key failedEST Off-screen Key Test failurewarning
FE0805Alarm volume key failedEST Off-screen Key Test failurewarning
FE0806Alarm silence key failedEST Off-screen Key Test failurewarning
FE0810Inspiratory pause key failedEST Off-screen Key Test failurewarning
FE0811Expiratory pause key failedEST Off-screen Key Test failurewarning
FE0901Rotary knob malfunctionEST Rotary Knob Test failureerror
FE1001Status display to BD communication failedEST Status Display test failure at start-uperror
FE1002BD to status display communication failedEST Status Display test failureerror
FE1005Status display failedEST Status Display test failureerror
FE1101Primary Audio Alarm TestActive Alarm Indicator failederror
FE1102Primary Audio Alarm TestLatched Alarm Indicator failederror
FE1103Primary Audio Alarm TestAlternative gas Indicator failederror
FE1104Primary Audio Alarm TestLow Alarm sound failederror
FE1105Primary Audio Alarm TestMedium Alarm sound failederror
FE1106Primary Audio Alarm TestHigh Alarm sound failederror
FE1201BD AudioAlarm cable malfunctionerror
FE1202BD AudioPower fail cap OORerror
FE1203BD AudioPower-Fail cap not dischargingerror
FE1208BD AudioPiezo alarm failureerror
FE1321EST Delivery PSOL TestNo Gas Connectederror
FE1324EST Delivery PSOL TestAC power not detectederror
FE1401Safety SystemUnable to establish stable flowcritical
FE1402Safety SystemSafety Valve Occludedcritical
FE1403Safety SystemSafety Valve Loopback current OORcritical
FE1421Safety SystemNo Gas Connectederror
FE1424Safety SystemAC power not detectederror
FE1521EST Exhalation Valve Loopback TestNo Gas Connectederror
FE1524EST Exhalation Valve Loopback TestAC power not detectederror
FE1601EST EV Pressure Accuracy & Seal TestExhalation valve calibration requirederror
FE1602EST EV Pressure Accuracy & Seal TestUnable to establish flowerror
FE1603EST EV Pressure Accuracy & Seal TestExhalation Flow too Higherror
FE1604EST EV Pressure Accuracy & Seal TestExhalation valve pressure OORerror
FE1605EST EV Pressure Accuracy & Seal TestLVDT Temp OORerror
FE1621EST EV Pressure Accuracy & Seal TestNo Gas Connectederror
FE1624EST EV Pressure Accuracy & Seal TestAC power not detectederror
FE1701EST EV Calibration Table TestExhalation Valve Not Calibratederror
FE1702EST EV Calibration Table TestUnable to establish stable flowerror
FE1703EST EV Calibration Table TestPressure OORerror
FE1705EST EV Calibration Table TestLVDT temp OORerror
FE1721EST EV Calibration Table TestNo Gas Connectederror
FE1724EST EV Calibration Table TestAC power not detectederror
FE1801EST EV Velocity Transducer TestUnable to establish flowerror
FE1802EST EV Velocity Transducer TestExhalation velocity transducer OORerror
FE1803EST EV Velocity Transducer TestUnable to stabilize position: Reseat the exhalation valve bodyerror
FE1821EST EV Velocity Transducer TestNo Gas Connectederror
FE1824EST EV Velocity Transducer TestAC power not detectederror
FE1902CommunicationGUI to BD ping failed on Ether1error
FE2001Ventilator BatteryCard cage fan malfunctionerror
FE2002Ventilator BatteryVentilator primary battery not installedcritical
FE2003Ventilator BatteryVentilator primary battery malfunctioncritical
FE2004Ventilator BatteryVentilator primary battery temperature OORcritical
FE2010Ventilator BatteryBoard temperature OORerror
FE2013Ventilator BatteryAC power still detectedwarning
FE2014Ventilator BatteryVentilator extended battery installedwarning
FE2016Ventilator primary battery voltage out of rangeBattery voltage test detects out-of-spec voltageerror
FE2017Ventilator primary battery triad voltages out of rangeBattery voltage test detects out-of-spec triad voltageserror
FE2018Ventilator primary battery current out of rangeBattery current test detects out-of-spec currenterror
FE2024AC power not detectedAC power detection test failserror
FE2102Compressor primary battery not installedBattery installation test failserror
FE2103Compressor primary battery malfunctionBattery function test failserror
FE2104Compressor primary battery temperature out of rangeBattery temperature test detects out-of-spec temperatureerror
FE2113AC power still detected during battery discharge testUser fails to disconnect AC power when prompted during battery discharge testerror
FE2114Compressor extended battery installedExtended battery status test detects extended battery presenceerror
FE2116Compressor primary battery voltage out of rangeBattery voltage test detects out-of-spec voltageerror
FE2117Compressor primary battery triad voltages out of rangeBattery voltage test detects out-of-spec triad voltageserror
FE2118Compressor primary battery current out of rangeBattery current test detects out-of-spec currenterror
FE2124AC power not detectedAC power detection test failserror
FE2201Unable to enable mix backup ventilationMix backup ventilation enable test fails within 50 secondserror
FE2202Mix backup ventilation failed to reach target pressureMix backup ventilation pressure test shows failure to reach targeterror
FE2203No air delivery in mix backup ventilationMix backup ventilation test detects zero air flowerror
FE2204No O2 delivery in mix backup ventilationMix backup ventilation test detects zero oxygen flowerror
FE2205O2 delivery when air expectedMix backup ventilation test detects oxygen in air channelerror
FE2206Air delivery when O2 expectedMix backup ventilation test detects air in O2 channelerror
FE2207Unable to establish stable flowInspiratory backup ventilation test fails to stabilize flowerror
FE2208Delivery flow is higher than expectedInspiratory backup ventilation test detects excessive flowerror
FE2209Unable to enable inspiratory backup ventilationInspiratory backup ventilation enable test failserror
FE2210Unable to establish stable pressureInspiratory backup ventilation test fails to stabilize pressureerror
FE2211Unable to establish stable flowInspiratory backup ventilation test fails to stabilize flowerror
FE2212Unable to enable exhalation backup ventilationExhalation backup ventilation enable test failserror
FE2213Exhalation backup ventilation current out of rangeExhalation backup ventilation current test detects out-of-spec currenterror
FE2214Exhalation valve current out of rangeExhalation backup ventilation current test detects out-of-spec currenterror
FE2215Pressure out of rangeExhalation backup ventilation pressure test detects out-of-spec pressureerror
FE2216Unable to reach minimum pressureExhalation backup ventilation test fails to reach minimum pressureerror
FE2221No gas connectedGas detection test fails to detect air or oxygenerror
FE2224AC power not detectedAC power detection test failserror
FE2302Wall air pressure detectedGas detection test detects wall air during compressor operationerror
FE2303Unable to drain accumulatorCompressor accumulator pressure test fails to drainerror
FE2304Pressure sensor zero offset out of rangeCompressor sensor offset test detects out-of-spec zero offseterror
FE2305Fan speed out of rangeFan speed test detects out-of-spec speederror
FE2306Motor speed at zeroCompressor motor speed test detects zero rpmerror
FE2308Pressure exceeded maximum valueCompressor accumulator pressure test detects over-pressure conditionerror
FE2309Compressor accumulator pressure test failureAccumulator pressure below expected range during diagnostic testerror
FE2309Target pressure not reachedCompressor accumulator pressure test shows pressure below targeterror
FE2310Pressure sensor out of rangeCompressor accumulator pressure test detects sensor out-of-range conditionerror
FE2311Unable to establish pressureCompressor accumulator pressure test fails to establish pressureerror
FE2312Pressure exceeded maximum valueCompressor accumulator pressure test detects over-pressure conditionerror
FE2324AC power not detectedAC power detection test failserror
FE2402Compressor leak - wall air pressure detectedGas detection test detects wall air during leak testerror
FE2403Compressor leak - unable to drain accumulatorAccumulator pressure test fails to drainerror
FE2404Compressor leak - target pressure not reachedAccumulator pressure test shows pressure below targeterror
FE2405Compressor leak - excessive leakAccumulator pressure test shows excessive leak rateerror
FE2424AC power not detectedAC power detection test failserror
FE2502Compressor performance - wall air pressure detectedGas detection test detects wall air during performance testerror
FE2503Compressor performance - unable to drain accumulatorAccumulator pressure test fails to drainerror
FE2504Compressor Performance test: Target pressure not reachedCompressor cannot achieve target pressure during diagnostic testerror
FE2505Compressor Performance test: Unable to establish stable flowCompressor fails to produce consistent, stable airflow during diagnosticerror
FE2524AC power not detected during compressor performance testCompressor performance test initiated without mains AC power availableerror
FS0001Unable to establish stable inspiratory flowFlow sensor cross-check test failureerror
FS0002Inspiratory flow sensor out of rangeFlow accuracy test detects sensor reading outside limitserror
FS0003Exhalation flow sensor out of rangeFlow accuracy test detects sensor reading outside limitserror
FS0004Mix flow sensor out of rangeFlow accuracy test detects sensor reading outside limitserror
FS0005PSOL current out of rangePSOL current accuracy test failureerror
FS0021No gas connectedGas connection test failureerror
FS0024AC power not detectedAC power connection test failurecritical
FS0101Exhalation valve not calibratedEV calibration test failurewarning
FS0102Unable to establish stable flow with EVQ installedFlow stability test failure during EV performance testerror
FS0103Exhalation valve pressure out of rangePressure accuracy test during EV performanceerror
FS0105LVDT temperature out of rangeMotor temperature test failure during EV performanceerror
FS0121No gas connectedGas connection test failure during EV performanceerror
FS0124AC power not detectedAC power connection test failure during EV performancecritical
FS0201Inspiratory zero offset out of rangePressure offset test failure during circuit pressure checkerror
FS0202Expiratory zero offset out of rangePressure offset test failure during circuit pressure checkerror
FS0203Target pressure not achievedPressure stability test failure during circuit pressure checkerror
FS0204Inspiratory and safety valve pressure difference outside limitsSV pressure accuracy test failureerror
FS0205Inspiratory and exhalation pressure difference outside limitsPressure sensor cross-check test failureerror
FS0206Inspiratory auto zero solenoid not operationalInspiratory pressure offset test failureerror
FS0207Exhalation auto zero solenoid not operationalExhalation pressure offset test failureerror
FS0221No gas connectedGas connection test failure during circuit pressure checkerror
FS0224AC power not detectedPower detection test failure during circuit pressure checkcritical
FS0303Unable to establish pressure in circuit leak testCircuit leak test failure during SST leak checkerror
FS0304System leak failureCircuit leak test detects excessive leakageerror
FS0321No gas connectedGas connection test failure during leak checkerror
FS0324AC power not detectedPower connection test failure during leak checkcritical
FS0401Unable to establish stable flow with exhalation filterFlow stability test failure during exhalation filter checkerror
FS0403Occluded expiratory compartmentExhalation filter resistance test detects excessive resistanceerror
FS0404Occluded expiratory filterExhalation filter resistance test detects blockageerror
FS0406Patient circuit not disconnectedCircuit connection test during setupwarning
FS0407Patient circuit not reconnectedCircuit connection test after servicewarning
FS0410Low expiratory filter pressure dropExhalation filter resistance test shows low resistanceerror
FS0421No gas connectedGas connection test shows no air and oxygen suppliescritical
FS0424AC power not detectedAC connection test failscritical
FS0501Unable to establish stable flowFlow stability test cannot achieve stable flowerror
FS0502Occluded inspiratory limbInspiratory limb pressure test detects high pressureerror
FS0504Unable to reach minimum peak flowFlow stability test cannot reach required flowerror
FS0505Occluded expiratory limbExpiratory limb pressure test detects high pressureerror
FS0508Inspiratory limb resistance lowInspiratory limb resistance test shows low resistancewarning
FS0509Expiratory limb resistance lowExpiratory limb resistance test shows low resistancewarning
FS0510Unable to reach minimum peak flowFlow stability test cannot reach minimum flowerror
FS0512Wye not blockedPressure build test shows low pressure in inspiratory limbwarning
FS0521No gas connectedGas connection test shows no air and oxygen suppliescritical
FS0524AC power not detectedAC connection test shows AC power not connectedcritical
FS0601Unable to pressurize circuitCircuit compliance test cannot reach required pressure pointserror
FS0602Unable to pressurize circuitCircuit compliance test cannot pressurizeerror
FS0603Compliance calculation failurePressure build test detects wye not blocked causing leakwarning
FS0604Excessive complianceCircuit compliance test shows abnormally high compliancewarning
FS0606Compliance lowCircuit compliance test shows low complianceerror
FS0608Excessive complianceCircuit compliance test shows abnormally high complianceerror
FS0610Compliance lowCircuit compliance test shows low complianceerror
FS0621No gas connectedGas connection test shows air and O2 supplies not detectedcritical
FS0624AC power not detectedPower connection test shows AC power not connectedcritical
FS0709Gas not connectedGas connection test for proximal flow sensorerror
FS0721No gas connectedGas connection test shows no gas detectedcritical
FS0724AC power not detectedPower connection test shows AC power not connectedcritical
FS0803Unable to establish pressureCircuit pressure test cannot build pressureerror
FS0804System leak detected during circuit leak testSST circuit leak test failureerror
FS0821Gas supply not connectedSST gas connection test failureerror
FS0824AC power not detectedSST AC power connection test failureerror
KB0004Expiratory motor current out of rangeExpiratory motor current measurement outside acceptable rangeerror
KP0013/XP0013Watchdog timer testPOST watchdog timer testinfo
KP0112Ventilator inoperative test failedCPU or fundamental ventilator function failure detected during POSTcritical
KP0115Digital-to-analog converter (DAC) test failedDAC malfunction detected for specific subsystem during POSTerror
KP0116Analog devices test failed in subsystemAnalog-to-digital converter or analog circuit failure in identified subsystemerror
KP0120Safe state system test failedFailure in safe shutdown or safety valve system during POSTcritical
KP0132Power distribution subsystem test failedDC power delivery circuit failure or insufficient voltage during POSTerror
KP0133Compressor subsystem test failedCompressor circuit or power supply failure detected during POSTerror
KP0134Mix subsystem testPOST diagnostic test initiatederror
KP0135Inspiratory subsystem testPOST diagnostic test initiatederror
KP0136Exhalation Subsystem testPOST diagnostic test initiatederror
LB0002Exhalation flow OORFlow measurement out of rangeerror
LB0005Ventilator 24 V bus power out of range24V power supply out of specificationcritical
LB0011Exhalation flow temp OORThermistor reading out of rangeerror
LB0021Atmospheric pressure out of rangeAtmospheric pressure sensor reading outside acceptable rangewarning
LB0022O2 sensor OOROxygen sensor measurement out of rangeerror
LB0023O2 sensor OOR resetO2 sensor out-of-range condition clearedinfo
LB0039BD time of day clock failedReal time clock failure detectedwarning
LB0052Loss of GUI communicationsGUI communication link losterror
LB0070Task monitorTask monitoring eventinfo
LB0088Proximal errorProximal flow sensor reading errorerror
LB0090Mix inlet pressure OORGas inlet pressure out of specificationwarning
LB0094Status display reported errorsStatus display communication errorerror
LB0095Loss of status display communicationStatus display communication losserror
LB0099Mix air flow OORAir flow measurement out of rangeerror
LB0100Mix O2 flow OOROxygen flow measurement out of rangeerror
LB0115CO2 sensor faultCO2 sensor reading errorerror
LB0116CO2 sensor zeroing initiatedCO2 sensor calibration procedure activeinfo
LB0121PI 3.3V scaled out of rangePI 3.3V scaled reading out of acceptable rangewarning
LB0125PE 3.3V reference scaled out of rangePE 3.3V reference scaled reading out of acceptable rangewarning
LB0126Mix 5V supply out of rangeMix circuit 5V supply voltage out of acceptable rangeerror
LB0127Mix 5V supply out of rangeMix circuit 5V supply voltage out of acceptable rangeerror
LB0129Mix 12V supply out of rangeMix circuit 12V supply voltage out of acceptable rangeerror
LB0130Mix 24V supply out of rangeMix circuit 24V supply voltage out of acceptable rangeerror
LB0145Ventilator 24 V main supply power out of rangeMain 24V power out of specificationcritical
LB0182High air inlet pressureAir supply pressure exceeds limitwarning
LB0183High O2 inlet pressureO2 supply pressure exceeds limitwarning
LB0194Inspiratory 24V supply out of rangeInspiratory 24V power supply voltage out of acceptable rangeerror
LB0232BD NOVRAM operational time values out of rangeCPU operational hours out of rangewarning
LB0234BD from GUI Operational time values out of rangeBD time values outside expected rangewarning
LB0236Operational times values out of rangeSystem operational hours out of rangewarning
LP0125Primary battery not installed or not detected by ventilatorBattery not present or not properly seated during power-on self testerror
LP0127Expiratory flow sensor cannot be read by ventilatorExhalation valve flow sensor (EVQ) electrical failure or connection issueerror
UXXXXXAssertion code (undefined content)Unknown assertion errorinfo
XB0080Init loss of BD communicationBD communication lost during initializationinfo
XP0097SAAS self test startSelf-test initiatederror
ZB0035Touch screen blockedTouch screen input disabledinfo
ZB0036Touch screen resumedTouch screen input re-enabledinfo
ZB0040GUI time of day clock failedGUI real time clock failure detectedwarning
ZB0053Loss of BD communicationsBD communication link losterror
ZB0071GUI drops a waveform packetWaveform data packet loss detectedinfo
ZB0103USB failureUSB interface failure detectedcritical
ZB0117UI board communications errorUI board communication issueinfo
ZB0118Video lock errorUI to BDU communication errorerror
ZB0119UI heartbeat errorUI board heartbeat signal lostinfo
ZB0178Offscreen keys driver eventKey input off screen detectedinfo
ZB0179Logs media failureLog storage medium failureinfo
ZB0180Logs initialization failureLog system initialization errorinfo
ZB0181Logs transaction failureLog write transaction failedinfo
ZB0191Unable to save config files to USBUSB save operation failedinfo
ZB0192Unable to load config files from USBUSB load operation failedinfo
ZB0231USB operational time error detectedUSB time tracking errorinfo
ZB0235GUI from BD operational time values out of rangeGUI time values outside expected rangewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI