Medtronic Puritan Bennett 980: troubleshooting and error codes
The ServiceForge knowledge base holds 80 documented symptoms for the Medtronic Puritan Bennett 980, with 108 ranked causes, 134 diagnostic checks with test points and expected values, and 369 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Proximal flow system alert patient safety
Proximal pressure or flow subsystem alert triggered during SST diagnostics
Ranked causes
- 1. SOL2 malfunction or failure (common)
- 2. Improper tubing connection from SOL2 to proximal flow system (common)
- 3. R4 resistor missing (rare)
- 4. Improper connections at proximal flow accumulator (occasional)
- 5. Options host PCBA failure (rare)
- 6. Proximal flow PCBA failure (rare)
Wait for proximal unit to enter idle mode
- Expected
- Prox in idle mode
11 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor accumulator pressure too low
EST diagnostic code FE2309 reported; compressor fails to build or maintain target accumulator pressure.
Ranked causes
- 1. Kinked or leaking pressure transducer tubing (common)
- 2. Improperly connected compressor-to-BDU interconnect hose (common)
- 3. Restricted inlet muffler (occasional)
- 4. Faulty compressor interface PCBA (occasional)
- 5. Faulty compressor assembly (rare)
Inspect pressure transducer tubing for kinks or leaks
- Test point
- Pressure transducer tubing and connections
- Expected
- Tubing clear, no kinks, no leaks visible
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proximal flow sensor alert AS0707
Purge suspended, user prompted to remove inspiratory filter and connect tubing.
Ranked causes
- 1. Proximal flow sensor tubing connection improper (common)
- 2. Proximal flow PCBA connection loose (occasional)
- 3. Proximal flow PCBA failure (rare)
- 4. Options host PCBA failure (rare)
Run proximal autozero test
- Test point
- Proximal pressure sensor
- Expected
- Autozero test passes
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Short self test (SST) fails
VBS resistance/compliance calculation or leak detection fails; common causes include inadequate gas sources, leaks, or occlusions
Ranked causes
- 1. Inadequate air or oxygen gas source pressure (common)
- 2. Leaks in breathing system (common)
- 3. Occlusions in system (common)
- 4. Ventilator not at operating temperature (occasional)
Warm up ventilator to operating temperature
- Test point
- Service mode
- Expected
- 15 minutes elapsed since entering Service mode
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SST fails or produces inaccurate results
Short Self Test does not complete successfully or returns incorrect resistance, compliance, or leak detection values
Ranked causes
- 1. Inadequate air or oxygen gas source pressure (common)
- 2. Leaks in breathing system (common)
- 3. Occlusions in breathing system (common)
- 4. Ventilator not at operating temperature (occasional)
Verify air and oxygen source pressures
- Test point
- air and oxygen inlets
- Expected
- 241 kPa to 599 kPa (35 psi to 87 psi), 241–599 kPa
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor failed to establish stable flow
EST diagnostic code FE2505 reported; compressor unable to produce consistent, stable airflow output.
Ranked causes
- 1. Damaged o-rings on pneumatic interface manifold (common)
- 2. Faulty compressor assembly (occasional)
- 3. Faulty compressor interface PCBA (occasional)
Inspect three o-rings on pneumatic interface manifold
- Test point
- Pneumatic interface manifold, three o-ring positions
- Expected
- All o-rings intact, no visible damage or extrusion
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor Performance error: Target pressure not reached
EST diagnostic code FE2504 reported; compressor cannot achieve target pressure during test.
Ranked causes
- 1. Leaks or occlusions (common)
- 2. Faulty compressor interface PCBA (occasional)
- 3. Faulty compressor assembly (occasional)
Inspect system for leaks or occlusions
- Test point
- All pressure hoses and connections
- Expected
- No visible leaks, no obstructions in lines
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor unable to hold constant accumulator pressure
EST diagnostic code FE2405 reported; compressor builds pressure but cannot maintain it, indicating leakage or regulation failure.
Ranked causes
- 1. Leaks or occlusions in system (common)
- 2. Compressor motor not turning on (occasional)
- 3. Faulty compressor interface PCBA (occasional)
Inspect system for leaks or occlusions
- Test point
- All pressure hoses and connections
- Expected
- No visible leaks, no obstructions in lines
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power not detected during compressor performance test
EST diagnostic code FE2524 reported; compressor performance test fails because mains AC power is not being detected or supplied to the ventilator.
Ranked causes
- 1. No AC power connected (common)
- 2. AC mains circuit breaker tripped (common)
Verify AC power present at wall outlet
- Test point
- AC mains outlet
- Expected
- AC voltage present (nominal 120V or 240V depending on region)
- Tools
- DMM or power tester
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm silence key LED failure AE0808
Audio paused key LED on bezel does not illuminate during off-screen key test.
Ranked causes
- 1. Alarm silence key LED not functioning (occasional)
- 2. Ribbon cable between user interface PCBA and touch screen loose (common)
Press audio paused key
- Test point
- Alarm silence key on bezel
- Expected
- LED illuminates when pressed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Atmospheric Pressure
Barometric pressure reading must be entered from external barometer; calibration may fail if incorrect pressure value provided or sensor drift occurs
Ranked causes
- 1. Incorrect barometric pressure value entered (common)
- 2. Sensor calibration drift or failure (rare)
Obtain accurate barometric pressure reading from external barometer
- Test point
- External barometer
- Expected
- Stabilized reading in mmHg
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen lock key LED failure AE0807
Screen lock key LED on bezel does not illuminate during off-screen key test.
Ranked causes
- 1. Screen lock key LED not functioning (occasional)
- 2. Ribbon cable between user interface PCBA and touch screen loose (common)
Press screen lock key
- Test point
- Screen lock key on bezel
- Expected
- LED illuminates when pressed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power not detected high voltage
Flow sensor or circuit pressure test failure; AC mains power not detected
Ranked causes
- 1. AC power not connected or circuit breaker tripped (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Atmospheric Pressure
Atmospheric pressure sensor calibration verification
Ranked causes
- 1. Calibration required as part of annual PM (unknown)
Atmospheric Pressure calibration
- Test point
- Service Mode
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Atmospheric pressure transducer
Atmospheric pressure transducer requires calibration after annual or 10,000-hour preventive maintenance
Ranked causes
- 1. Routine preventive maintenance (common)
Perform ventilator warm-up cycle
- Test point
- Ventilator startup
- Expected
- Warm-up completed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery cover slider latch replacement
Replace battery cover slider latch and reattach extension spring
Ranked causes
- 1. Battery cover slider latch worn or damaged (occasional)
Remove screws from battery cover slider latch retainer
- Test point
- Two screws on slider latch retainer
- Expected
- Screws removed while retaining slider latch
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery door panel alignment
Align battery door panel to BDU assembly and tighten mounting screws
Ranked causes
- 1. Battery door panel misalignment or loose screws (common)
Verify battery door panel screws are seated properly
- Test point
- Top two screws of battery panel door assembly
- Expected
- Screws fully tightened and flush
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Component Calibration patient safety
Perform component calibrations as specified in Section 4.5.2. Results must be documented on the PVR form.
Ranked causes
- 1. Component calibration procedure required (common)
Confirm component calibration completion
- Test point
- PVR form
- Expected
- Component calibration section completed and signed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation Valve
Exhalation valve calibration verification
Ranked causes
- 1. Calibration required as part of annual PM (unknown)
Exhalation Valve calibration
- Test point
- Service Mode
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation Valve
Exhalation valve calibration to build valve table; requires gold standard test circuit, proper warm-up time, and EST leak test passage.
Ranked causes
- 1. Exhalation valve requires calibration (common)
Allow ventilator warm-up time if not at operating temperature from recent usage
- Test point
- Ventilator operating temperature
- Expected
- At least 15 minutes warm-up in Service mode
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Exhalation Valve Calibration patient safety
Calibration procedure for exhalation valve on Puritan Bennett 980 ventilator using on-screen prompts in Service mode
Ranked causes
- 1. System leak or improper test circuit attachment (common)
Ensure exhalation filter condensate vial is secured tightly
- Test point
- Exhalation filter condensate vial
- Expected
- Secured tightly
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Extended Self Test (EST) patient safety
Routine EST procedure to verify ventilator function before clinical use. Must complete all tests in sequence without interruption using Run ALL button. All tests must pass with no failures.
Ranked causes
- 1. Routine EST execution required before clinical use (common)
Ensure adult exhalation filter is installed before running EST
- Test point
- Exhalation filter connection
- Expected
- Adult exhalation filter installed
1 more check with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow Sensor
Flow sensor calibration verification
Ranked causes
- 1. Calibration required as part of annual PM (unknown)
Flow Sensor calibration
- Test point
- Service Mode
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen sensor
Oxygen sensor requires calibration after maintenance, replacement, or extended self-test
Ranked causes
- 1. Routine maintenance or oxygen sensor replacement (common)
Perform ventilator warm-up cycle
- Test point
- Ventilator startup
- Expected
- Warm-up completed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Sensor
Oxygen sensor calibration verification
Ranked causes
- 1. Calibration required as part of annual PM (unknown)
Oxygen Sensor calibration
- Test point
- Service Mode
- Expected
- Pass / Fail / Limited Use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen Sensor Calibration patient safety
Calibration of oxygen sensor performed in normal ventilation mode after ventilator reaches operating temperature
Ranked causes
- 1. Oxygen sensor needs calibration (unknown)
Allow ventilator to reach operating temperature
- Test point
- Ventilator warm-up time
- Expected
- At least 15 minutes in Service mode
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oxygen sensor installation patient safety
Step-by-step procedure to install a replacement oxygen sensor with proper seating and cable reconnection
Ranked causes
- 1. Standard replacement procedure (common)
Check o-ring presence
- Test point
- Oxygen sensor
- Expected
- O-ring present
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PTS 2000 preparation patient safety
Setup procedure for PTS 2000 sensor in ventilator circuit prior to performance verification or calibration checks.
Ranked causes
- 1. PTS 2000 must be integrated into patient circuit for measurement (common)
Install PTS 2000 sensor in circuit
- Test point
- Patient circuit connection
- Expected
- PTS 2000 connected in series with two longest sections per Figure 11
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SST temperature requirement
Ventilator warm-up procedure to ensure accurate SST results
Ranked causes
- 1. Cold ventilator from recent usage (unknown)
Allow warm-up time
- Test point
- Service mode
- Expected
- 15 minutes elapsed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen
Touch screen requires calibration after annual or 10,000-hour preventive maintenance
Ranked causes
- 1. Routine preventive maintenance (common)
Perform ventilator warm-up cycle
- Test point
- Ventilator startup
- Expected
- Warm-up completed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch Screen
Touch screen calibration verification
Ranked causes
- 1. Calibration required as part of annual PM (unknown)
Touch Screen calibration
- Test point
- Service Mode
- Expected
- Pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ventilator date, time and settings after coin battery replacement
After replacing coin batteries, ventilator date/time and programmed settings must be verified to persist across power cycles to ensure correct operation.
Ranked causes
- 1. Coin batteries replaced, requiring date/time reset (common)
Power on ventilator and enter date/time configuration
- Test point
- Configuration icon (wrench) and Date/Time Change tab
- Expected
- Date/Time Change tab accessible and editable
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ventilator Performance Verification patient safety
Complete electrical safety testing and component calibration before performance verification. Document results on PVR form with software/firmware versions, serial numbers, and safety test measurements.
Ranked causes
- 1. Verification procedure setup (common)
Ground resistance measurement
- Test point
- Ground integrity
- Expected
- <0.20 Ω, <0.20
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Ventilator Setup for PTS 2000 patient safety
Warm up and configure ventilator with exhalation filter, inspiratory filter, and PTS 2000 sensor in patient circuit; connect gas and power supplies; verify indicators and battery; perform EST, SST and calibrations.
Ranked causes
- 1. Ventilator requires warm-up and configuration before calibration (common)
Warm up ventilator in Service mode
- Test point
- Service mode indicator
- Expected
- Minimum 15 minutes elapsed
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EST test failure or diagnostic code patient safety
EST test fails to complete or returns a diagnostic code indicating a problem detected during testing.
Ranked causes
- 1. One or more EST tests failed or diagnostic code returned (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
EV performance unable to establish stable flow
EV performance test failure; unable to establish stable exhalation flow
Ranked causes
- 1. EVQ fault, fouled, or flow sensor miscalibration (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation auto zero solenoid not operational
Circuit pressure test failure; exhalation pressure offset solenoid not functioning
Ranked causes
- 1. Leaks, disconnects, occlusions, or solenoid/PCBA fault (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation filter unable to establish stable flow
Exhalation filter test failure; unable to establish stable flow with filter installed
Ranked causes
- 1. Exhaust filter fault or EVQ issue (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation flow sensor out of range
Flow sensor cross-check failure; exhalation flow sensor reading out of acceptable range
Ranked causes
- 1. Patient circuit leaks, improper EVQ seating, or faulty EVQ (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation valve not calibrated
EV performance test failure; exhalation valve requires calibration
Ranked causes
- 1. Exhalation valve calibration not completed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation valve pressure out of range
EV performance test failure; measured pressure outside acceptable range
Ranked causes
- 1. Exhaust filter occluded, o-ring missing, EVQ fault, or F5 filter issue (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Expiratory zero offset out of range
Circuit pressure test failure; expiratory zero pressure offset exceeds acceptable limits
Ranked causes
- 1. Pressure sensor calibration drift (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Extended self test (EST) fails
EST verifies ventilator operational subsystem integrity; failure indicates possible system component or functional issue
Ranked causes
- 1. SST not passing (indicates underlying system problem) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor calibration failure
Calibration procedure fails to complete successfully
Ranked causes
- 1. System leak present (common)
Inspect exhalation filter for leaks
- Test point
- Exhalation filter collector vial and connections
- Expected
- No leaks detected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ground isolation resistance too low high voltage
Ground isolation resistance measurement fails to meet minimum threshold during electrical safety testing
Ranked causes
- 1. Faulty assembly, subassembly, cable or harness causing low isolation resistance (unknown)
Measure ground isolation resistance
- Test point
- DB9 connector female mating face
- Expected
- > 975 Ω or > 0.975 kΩ, > specified value
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory and exhalation pressure difference outside limits
Circuit pressure test failure; inspiratory and exhalation pressures differ beyond acceptable range
Ranked causes
- 1. Leaks, disconnects, occlusions, or pressure sensor fault (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory and safety valve pressure difference outside limits
Circuit pressure test failure; inspiratory and safety valve pressures differ beyond tolerance
Ranked causes
- 1. Leaks, disconnects, occlusions, or pressure sensor fault (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory auto zero solenoid not operational
Circuit pressure test failure; inspiratory pressure offset solenoid not functioning
Ranked causes
- 1. Leaks, disconnects, occlusions, or solenoid/PCBA fault (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory flow sensor out of range
Flow sensor cross-check failure; inspiratory flow sensor reading out of acceptable range
Ranked causes
- 1. Inadequate source gas supply (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory zero offset out of range
Circuit pressure test failure; inspiratory zero pressure offset exceeds acceptable limits
Ranked causes
- 1. Pressure sensor calibration drift (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LVDT temperature out of range
EV performance test failure; motor temperature sensor reading out of range
Ranked causes
- 1. EVQ or exhalation valve assembly fault (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mix flow sensor out of range
Flow sensor cross-check failure; mix flow sensor reading out of acceptable range
Ranked causes
- 1. Inadequate source gas supply (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No air supply connected patient safety
SST detects no air input to the system during flow sensor cross check
Ranked causes
- 1. Air supply line not connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No gas connected
Flow sensor or circuit pressure test failure; no gas supply detected
Ranked causes
- 1. Air or oxygen supply not connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No oxygen supply connected patient safety
SST detects no oxygen input to the system during flow sensor cross check
Ranked causes
- 1. Oxygen supply line not connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen sensor requires replacement patient safety
Oxygen sensor has reached recommended one-year replacement interval or shows signs of malfunction; expiration date on new sensor packaging must not be past installation date
Ranked causes
- 1. Scheduled maintenance - one year from installation (common)
Verify expiration date on new oxygen sensor packaging
- Test point
- Sensor packaging
- Expected
- Installation date must be before or on expiration date
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Piezo alarm cable incorrectly connected
Alarm cables connected to wrong polarity terminals during BDU top cap assembly replacement
Ranked causes
- 1. Incorrect polarity connection during reassembly (occasional)
Verify piezo alarm cable polarity matches label markings
- Test point
- Piezo alarm cable terminals
- Expected
- Positive (+) connected to red terminal, negative (-) connected to black terminal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Post coin battery replacement system verification tests
Series of mandatory tests to be performed after coin battery replacement before returning ventilator to patient use.
Ranked causes
- 1. Coin battery replacement completed, requiring verification (common)
Perform ventilator warm-up cycle
- Test point
- Service mode operation
- Expected
- Ventilator running continuously for 15 minutes without errors
12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PSOL current out of range
Flow sensor cross-check failure; proportional solenoid current measurement out of acceptable range
Ranked causes
- 1. Inadequate source gas or patient circuit leaks (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SST proximal purge error
Alert displayed during SST proximal purge procedure test
Ranked causes
- 1. Options host PCBA or proximal flow PCBA fault (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- SST system leak
- System leak detected
- System leak failure
- Target pressure not achieved
- Test: Adult Short Self Test (SST)
- Test: Earth Current Leakage (Normal Polarity, Closed Neutral)
- Test: Earth Current Leakage (Normal Polarity, Open Neutral)
- Test: Earth Current Leakage (Reverse Polarity, Closed Neutral)
- Test: Earth Current Leakage (Reverse Polarity, Open Neutral)
- Test: Extended Self Test (EST)
- Test: External Ports
- Test: Ground Isolation Resistance
- Test: Ground Resistance
- Test: Neonatal Short Self Test (SST)
- Test: Oxygen Sensor Calibration Verification
- Test: Pediatric Short Self Test (SST)
- Test: Ventilator Inop
- Unable to establish pressure in circuit leak test
- Unable to establish stable flow
- Vent inop test
20 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AE0602 | Delivery pressure solenoid leak exceeds alert limit | EST Leak Test on delivery PSOL | error |
AE0605 | Test circuit not connected to ventilator | EST Leak Test start without test circuit connection | error |
AE0607 | Delivery PSOL leak alert | Forward leakage on delivery PSOL exceeds alert limit during EST leak test | warning |
AE0807 | Screen lock key LED failure | Off-screen key LED on screen lock key does not illuminate during test | error |
AE0808 | Alarm silence key LED failure | Off-screen key LED on audio paused key does not illuminate during test | error |
AE0809 | Key-panel LED backlight brightness not changing | Off-screen Key Test execution | warning |
AE1301 | Delivery PSOL current feedback reading out of range | EST Delivery PSOL Test execution | error |
AE1501 | Exhalation valve loopback current out of range | EST Exhalation Valve Loopback Test execution | error |
AE1604 | Exhalation valve pressure measurement out of range | EST EV Pressure Accuracy & Seal Test | error |
AE1901 | Serial port communication malfunction | Communication loopback test | error |
AE2015 | Ventilator primary battery charge below acceptable level | Ventilator Battery EST during startup | error |
AE2100 | Compressor battery not installed; test skipped | Compressor Battery EST | info |
AE2115 | Compressor primary battery charge below acceptable level | Compressor Battery EST during startup | error |
AE2300 | Compressor not detected; EST skipped | Compressor EST | info |
AE2301 | Compressor not detected; EST skipped | Compressor EST | info |
AE2307 | Compressor motor speed outside acceptable range | Compressor performance monitoring | error |
AE2400 | Compressor leak test not performed; compressor not detected | Compressor Leak EST | info |
AE2401 | Compressor leak test not performed; compressor not detected | Compressor Leak EST | info |
AE2500 | Compressor Performance EST not performed; compressor not detected | Compressor Performance EST | info |
AE2501 | Compressor Performance EST not performed; compressor not detected | Compressor Performance EST | info |
AE2506 | Compressor unable to maintain required accumulator pressure | Compressor Performance EST | error |
AE2507 | Compressor unable to maintain required flow rate | Compressor Performance EST | error |
AS0022 | No oxygen supply connected | SST flow sensor cross check detects no oxygen during test | critical |
AS0023 | No air supply connected | SST flow sensor cross check detects no air during test | critical |
AS0305 | System Leak Alert | SST Leak test detects pressure drop over time | warning |
AS0405 | Occluded expiratory filter | SST Exhalation Filter test calculates excessive pressure drop | warning |
AS0408 | Occluded expiratory compartment | SST Exhalation Filter test verifies pressure drop out of range | warning |
AS0411 | Low expiratory filter ΔP | SST Exhalation Filter test shows low pressure drop | warning |
AS0503 | Occluded inspiratory limb | SST Circuit Resistance test detects excessive inspiratory limb pressure drop | warning |
AS0506 | Occluded expiratory limb | SST Circuit Resistance test detects excessive expiratory limb pressure drop | warning |
AS0507 | Unable to reach min peak flow | SST Circuit Resistance test cannot achieve minimum peak flow | warning |
AS0511 | Unable to reach min peak flow | SST Circuit Resistance test characterizes total resistance and fails | warning |
AS0513 | Insp limb resistance low | SST Circuit Resistance test shows low inspiratory limb resistance | warning |
AS0514 | Exp limb resistance low | SST Circuit Resistance test shows low expiratory limb resistance | warning |
AS0605 | Excessive compliance | SST Circuit Compliance test detects high circuit compliance | warning |
AS0607 | Compliance low | SST Circuit Compliance test detects low circuit compliance | warning |
AS0609 | Excessive compliance | SST Circuit Compliance test detects high circuit compliance | warning |
AS0611 | Compliance low | SST Circuit Compliance test detects low circuit compliance | warning |
AS0700 | SST: NOT INSTALLED | Ventilator does not detect proximal flow module at startup | info |
AS0701 | SST Prox Comm. Error | SST Prox system recognition test fails | error |
AS0703 | Demo Mode Square Wave Failed | SST Prox proximal flow system communications test fails | error |
AS0704 | P-atm Cross Check Failed | SST Prox barometric pressure readings between ventilator and prox module differ | error |
AS0705 | Autozero and Leak Test Failed | SST Prox checks pneumatic path for functionality | error |
AS0705 | Proximal flow system alert | Proximal flow subsystem malfunction or out of specification condition detected | error |
AS0707 | Pressure Cross Check Failed | SST Prox pressure cross check of proximal flow system fails | error |
AS0707 | Inspiratory filter disconnection or proximal sensor tubing issue | Filter removed or tubing disconnection detected | warning |
AS0707 | Proximal flow sensor alert | Proximal flow sensor tubing or connection problem detected during system self-test | warning |
AS0708 | Flow Sensor Cross Check Failed | SST Prox flow sensor cross check of proximal flow system fails | error |
AS0710 | Operator Skipped Prox Test | Proximal flow sensor connected but operator declines to run prox SST | info |
AS0711 | Purge Error | SST Prox purge procedure test fails | error |
AS0712 | SST proximal purge error | Proximal purge procedure test failure | warning |
AS0713 | SST proximal purge error | Proximal purge procedure test failure | warning |
AS0714 | SST proximal purge error | Proximal purge procedure test failure | warning |
AS0715 | SST proximal purge error | Proximal purge procedure test failure | warning |
AS0805 | System leak alert | Leak test detects circuit leakage | warning |
DXXXXX | Assertion code (undefined content) | Unknown assertion error | warning |
FE0001 | Air pressure supply solenoid leak rate exceeds specification | Mix Leak Test on air supply circuit | critical |
FE0002 | Oxygen supply solenoid leak rate exceeds specification | Mix Leak Test on O2 supply circuit | critical |
FE0004 | Mix circuit accumulator unable to reach required pressure | Mix circuit accumulator pressure test | critical |
FE0005 | Mix circuit accumulator pressure drop exceeds acceptable limit | Mix circuit leak test - pressure hold phase | critical |
FE0006 | Mix circuit unable to pressurize accumulator | Mix circuit accumulator pressurization test | critical |
FE0007 | Air inlet pressure exceeds limit after disconnection | Mix circuit air inlet pressure verification | error |
FE0008 | Air side mix check valve reverse flow detected | Mix circuit air check valve test | error |
FE0009 | O2 side mix check valve reverse flow detected | Mix circuit O2 check valve test | error |
FE0010 | Oxygen inlet pressure exceeds limit | Mix circuit O2 inlet pressure verification | error |
FE0011 | Mix circuit unable to reach target accumulator pressure using O2 | Mix circuit O2-side accumulator pressurization test | error |
FE0012 | Mix circuit exceeded target accumulator pressure using O2 | Mix circuit O2-side accumulator pressurization test | critical |
FE0013 | Mix circuit unable to reach target accumulator pressure using air | Mix circuit air-side accumulator pressurization test | error |
FE0014 | Mix circuit exceeded target accumulator pressure using air | Mix circuit air-side accumulator pressurization test | critical |
FE0015 | Oxygen inlet pressure below acceptable limit | Mix circuit O2 inlet pressure test | warning |
FE0016 | Inspiratory flow rate exceeds specification | Mix circuit flow measurement test | error |
FE0017 | Options gas supply accumulator pressure unstable | Mix circuit options gas supply test | error |
FE0018 | Option Gas Supply solenoid leak rate exceeds specification | Mix circuit options gas solenoid test | error |
FE0021 | No air or oxygen supply connected to ventilator | Mix circuit gas supply detection test | critical |
FE0024 | AC power not detected by ventilator | Mix circuit AC power detection test | critical |
FE0101 | Air supply PSOL has excessive internal leakage | Mix PSOL lift-off test | critical |
FE0102 | O2 PSOL has excessive leak | Mix PSOL liftoff test | error |
FE0103 | Air PSOL current out of range | Mix PSOL liftoff test | error |
FE0104 | O2 PSOL current out of range | Mix PSOL liftoff test | error |
FE0107 | Air PSOL maximum liftoff current exceeded | Mix PSOL liftoff test | error |
FE0108 | O2 PSOL maximum liftoff current exceeded | Mix PSOL liftoff test | error |
FE0109 | Air PSOL liftoff current out of range | Mix PSOL liftoff test | error |
FE0110 | O2 PSOL liftoff current out of range | Mix PSOL liftoff test | error |
FE0111 | Mix air PSOL loopback current out of range | Mix PSOL loopback test | error |
FE0121 | No gas connected | Gas connection test | error |
FE0124 | AC power not detected | AC connection test | error |
FE0201 | Unable to stabilize accumulator pressure | Pressure sensor test | error |
FE0203 | Accumulator pressure out of range | Pressure sensor test | error |
FE0204 | Overpressure switch failed to activate | Pressure switch test | error |
FE0205 | Overpressure switch failed to deactivate | Pressure switch test | error |
FE0221 | No gas connected | Mix accumulator test | error |
FE0224 | AC power not detected | Mix accumulator test | error |
FE0301 | Inspiratory zero offset out of range | Pressure transducer auto-zero test | error |
FE0302 | Expiratory zero offset out of range | Pressure transducer auto-zero test | error |
FE0303 | Target pressure not achieved | Pressure cross check test | error |
FE0304 | Inspiratory and safety valve pressure sensor difference outside test limits | Pressure cross check test | error |
FE0305 | Inspiratory and exhalation pressure sensor difference outside test limits | Pressure cross check test | error |
FE0306 | Inspiratory auto zero solenoid not operational | Auto-zero solenoid test | error |
FE0307 | Exhalation auto zero solenoid not operational | Auto-zero solenoid test | error |
FE0321 | No gas connected | Circuit pressure test | error |
FE0324 | AC power not detected | Circuit pressure test | error |
FE0401 | Air inlet pressure sensor zero offset out of range | Zero offset pressure test | error |
FE0402 | O2 inlet pressure exceeded | Zero offset pressure test | error |
FE0404 | Accumulator pressure sensor zero offset out of range | Zero offset pressure test | error |
FE0405 | Inspiratory pressure sensor zero offset out of range | Zero offset pressure test | error |
FE0406 | Exhalation pressure sensor zero offset out of range | Zero offset pressure test | error |
FE0409 | Air mix flow sensor zero offset out of range | Flow offset test | error |
FE0410 | O2 mix flow exceeded | Flow offset test | error |
FE0412 | Inspiratory flow sensor zero offset out of range | Flow offset test | error |
FE0413 | Expiratory flow sensor zero offset out of range | Flow offset test | error |
FE0414 | Air still connected | Gas connection test | warning |
FE0415 | O2 still connected | Gas connection test | warning |
FE0424 | AC power not detected | AC power connection test | error |
FE0501 | Unable to establish stable flow | Flow stability test | error |
FE0502 | Inspiratory flow sensor out of range | Flow sensor cross check test | error |
FE0503 | Exhalation flow sensor out of range | Flow sensor cross check test | error |
FE0504 | Mix flow sensor out of range | Flow sensor cross check test | error |
FE0505 | PSOL current out of range | EST Flow Sensor Cross Check test failure | error |
FE0521 | No gas connected | EST Flow Sensor Cross Check test failure | error |
FE0524 | AC power not detected | EST Flow Sensor Cross Check test failure | error |
FE0601 | Delivery PSOL leak exceeds failure limit | EST Leak Test failure | error |
FE0603 | Unable to establish pressure during leak test | EST Leak Test and check valve reverse flow test | error |
FE0604 | Excessive leak during leak test | EST Leak Test failure | error |
FE0606 | Inspiratory check valve reverse flow too low | EST Leak Test check valve reverse flow test | error |
FE0607 | Inspiratory check valve reverse flow too high | EST Leak Test check valve reverse flow test | error |
FE0608 | Unable to establish accumulator pressure | EST Leak Test delivery PSOL leak test | error |
FE0609 | Unable to establish stable flow | EST Leak Test delivery PSOL leak test | error |
FE0610 | Safety valve failed to open | EST Leak Test delivery PSOL leak test | critical |
FE0611 | Safety valve cracking pressure out of range | EST Leak Test delivery PSOL leak test | error |
FE0621 | No gas connected during leak test | EST Leak Test gas connection test | error |
FE0624 | AC power not detected during leak test | EST Leak Test AC connection test | error |
FE0701 | No media driver found | EST Internal Storage test failure | error |
FE0702 | File open error | EST Internal Storage test failure | error |
FE0703 | File write error | EST Internal Storage test failure | error |
FE0704 | File read error | EST Internal Storage test failure | error |
FE0705 | File cleanup error | EST Internal Storage test failure | error |
FE0801 | Screen brightness key failed | EST Off-screen Key Test failure | warning |
FE0802 | Screen lock key failed | EST Off-screen Key Test failure | warning |
FE0803 | Manual inspiration key failed | EST Off-screen Key Test failure | warning |
FE0804 | Alarm reset key failed | EST Off-screen Key Test failure | warning |
FE0805 | Alarm volume key failed | EST Off-screen Key Test failure | warning |
FE0806 | Alarm silence key failed | EST Off-screen Key Test failure | warning |
FE0810 | Inspiratory pause key failed | EST Off-screen Key Test failure | warning |
FE0811 | Expiratory pause key failed | EST Off-screen Key Test failure | warning |
FE0901 | Rotary knob malfunction | EST Rotary Knob Test failure | error |
FE1001 | Status display to BD communication failed | EST Status Display test failure at start-up | error |
FE1002 | BD to status display communication failed | EST Status Display test failure | error |
FE1005 | Status display failed | EST Status Display test failure | error |
FE1101 | Primary Audio Alarm Test | Active Alarm Indicator failed | error |
FE1102 | Primary Audio Alarm Test | Latched Alarm Indicator failed | error |
FE1103 | Primary Audio Alarm Test | Alternative gas Indicator failed | error |
FE1104 | Primary Audio Alarm Test | Low Alarm sound failed | error |
FE1105 | Primary Audio Alarm Test | Medium Alarm sound failed | error |
FE1106 | Primary Audio Alarm Test | High Alarm sound failed | error |
FE1201 | BD Audio | Alarm cable malfunction | error |
FE1202 | BD Audio | Power fail cap OOR | error |
FE1203 | BD Audio | Power-Fail cap not discharging | error |
FE1208 | BD Audio | Piezo alarm failure | error |
FE1321 | EST Delivery PSOL Test | No Gas Connected | error |
FE1324 | EST Delivery PSOL Test | AC power not detected | error |
FE1401 | Safety System | Unable to establish stable flow | critical |
FE1402 | Safety System | Safety Valve Occluded | critical |
FE1403 | Safety System | Safety Valve Loopback current OOR | critical |
FE1421 | Safety System | No Gas Connected | error |
FE1424 | Safety System | AC power not detected | error |
FE1521 | EST Exhalation Valve Loopback Test | No Gas Connected | error |
FE1524 | EST Exhalation Valve Loopback Test | AC power not detected | error |
FE1601 | EST EV Pressure Accuracy & Seal Test | Exhalation valve calibration required | error |
FE1602 | EST EV Pressure Accuracy & Seal Test | Unable to establish flow | error |
FE1603 | EST EV Pressure Accuracy & Seal Test | Exhalation Flow too High | error |
FE1604 | EST EV Pressure Accuracy & Seal Test | Exhalation valve pressure OOR | error |
FE1605 | EST EV Pressure Accuracy & Seal Test | LVDT Temp OOR | error |
FE1621 | EST EV Pressure Accuracy & Seal Test | No Gas Connected | error |
FE1624 | EST EV Pressure Accuracy & Seal Test | AC power not detected | error |
FE1701 | EST EV Calibration Table Test | Exhalation Valve Not Calibrated | error |
FE1702 | EST EV Calibration Table Test | Unable to establish stable flow | error |
FE1703 | EST EV Calibration Table Test | Pressure OOR | error |
FE1705 | EST EV Calibration Table Test | LVDT temp OOR | error |
FE1721 | EST EV Calibration Table Test | No Gas Connected | error |
FE1724 | EST EV Calibration Table Test | AC power not detected | error |
FE1801 | EST EV Velocity Transducer Test | Unable to establish flow | error |
FE1802 | EST EV Velocity Transducer Test | Exhalation velocity transducer OOR | error |
FE1803 | EST EV Velocity Transducer Test | Unable to stabilize position: Reseat the exhalation valve body | error |
FE1821 | EST EV Velocity Transducer Test | No Gas Connected | error |
FE1824 | EST EV Velocity Transducer Test | AC power not detected | error |
FE1902 | Communication | GUI to BD ping failed on Ether1 | error |
FE2001 | Ventilator Battery | Card cage fan malfunction | error |
FE2002 | Ventilator Battery | Ventilator primary battery not installed | critical |
FE2003 | Ventilator Battery | Ventilator primary battery malfunction | critical |
FE2004 | Ventilator Battery | Ventilator primary battery temperature OOR | critical |
FE2010 | Ventilator Battery | Board temperature OOR | error |
FE2013 | Ventilator Battery | AC power still detected | warning |
FE2014 | Ventilator Battery | Ventilator extended battery installed | warning |
FE2016 | Ventilator primary battery voltage out of range | Battery voltage test detects out-of-spec voltage | error |
FE2017 | Ventilator primary battery triad voltages out of range | Battery voltage test detects out-of-spec triad voltages | error |
FE2018 | Ventilator primary battery current out of range | Battery current test detects out-of-spec current | error |
FE2024 | AC power not detected | AC power detection test fails | error |
FE2102 | Compressor primary battery not installed | Battery installation test fails | error |
FE2103 | Compressor primary battery malfunction | Battery function test fails | error |
FE2104 | Compressor primary battery temperature out of range | Battery temperature test detects out-of-spec temperature | error |
FE2113 | AC power still detected during battery discharge test | User fails to disconnect AC power when prompted during battery discharge test | error |
FE2114 | Compressor extended battery installed | Extended battery status test detects extended battery presence | error |
FE2116 | Compressor primary battery voltage out of range | Battery voltage test detects out-of-spec voltage | error |
FE2117 | Compressor primary battery triad voltages out of range | Battery voltage test detects out-of-spec triad voltages | error |
FE2118 | Compressor primary battery current out of range | Battery current test detects out-of-spec current | error |
FE2124 | AC power not detected | AC power detection test fails | error |
FE2201 | Unable to enable mix backup ventilation | Mix backup ventilation enable test fails within 50 seconds | error |
FE2202 | Mix backup ventilation failed to reach target pressure | Mix backup ventilation pressure test shows failure to reach target | error |
FE2203 | No air delivery in mix backup ventilation | Mix backup ventilation test detects zero air flow | error |
FE2204 | No O2 delivery in mix backup ventilation | Mix backup ventilation test detects zero oxygen flow | error |
FE2205 | O2 delivery when air expected | Mix backup ventilation test detects oxygen in air channel | error |
FE2206 | Air delivery when O2 expected | Mix backup ventilation test detects air in O2 channel | error |
FE2207 | Unable to establish stable flow | Inspiratory backup ventilation test fails to stabilize flow | error |
FE2208 | Delivery flow is higher than expected | Inspiratory backup ventilation test detects excessive flow | error |
FE2209 | Unable to enable inspiratory backup ventilation | Inspiratory backup ventilation enable test fails | error |
FE2210 | Unable to establish stable pressure | Inspiratory backup ventilation test fails to stabilize pressure | error |
FE2211 | Unable to establish stable flow | Inspiratory backup ventilation test fails to stabilize flow | error |
FE2212 | Unable to enable exhalation backup ventilation | Exhalation backup ventilation enable test fails | error |
FE2213 | Exhalation backup ventilation current out of range | Exhalation backup ventilation current test detects out-of-spec current | error |
FE2214 | Exhalation valve current out of range | Exhalation backup ventilation current test detects out-of-spec current | error |
FE2215 | Pressure out of range | Exhalation backup ventilation pressure test detects out-of-spec pressure | error |
FE2216 | Unable to reach minimum pressure | Exhalation backup ventilation test fails to reach minimum pressure | error |
FE2221 | No gas connected | Gas detection test fails to detect air or oxygen | error |
FE2224 | AC power not detected | AC power detection test fails | error |
FE2302 | Wall air pressure detected | Gas detection test detects wall air during compressor operation | error |
FE2303 | Unable to drain accumulator | Compressor accumulator pressure test fails to drain | error |
FE2304 | Pressure sensor zero offset out of range | Compressor sensor offset test detects out-of-spec zero offset | error |
FE2305 | Fan speed out of range | Fan speed test detects out-of-spec speed | error |
FE2306 | Motor speed at zero | Compressor motor speed test detects zero rpm | error |
FE2308 | Pressure exceeded maximum value | Compressor accumulator pressure test detects over-pressure condition | error |
FE2309 | Compressor accumulator pressure test failure | Accumulator pressure below expected range during diagnostic test | error |
FE2309 | Target pressure not reached | Compressor accumulator pressure test shows pressure below target | error |
FE2310 | Pressure sensor out of range | Compressor accumulator pressure test detects sensor out-of-range condition | error |
FE2311 | Unable to establish pressure | Compressor accumulator pressure test fails to establish pressure | error |
FE2312 | Pressure exceeded maximum value | Compressor accumulator pressure test detects over-pressure condition | error |
FE2324 | AC power not detected | AC power detection test fails | error |
FE2402 | Compressor leak - wall air pressure detected | Gas detection test detects wall air during leak test | error |
FE2403 | Compressor leak - unable to drain accumulator | Accumulator pressure test fails to drain | error |
FE2404 | Compressor leak - target pressure not reached | Accumulator pressure test shows pressure below target | error |
FE2405 | Compressor leak - excessive leak | Accumulator pressure test shows excessive leak rate | error |
FE2424 | AC power not detected | AC power detection test fails | error |
FE2502 | Compressor performance - wall air pressure detected | Gas detection test detects wall air during performance test | error |
FE2503 | Compressor performance - unable to drain accumulator | Accumulator pressure test fails to drain | error |
FE2504 | Compressor Performance test: Target pressure not reached | Compressor cannot achieve target pressure during diagnostic test | error |
FE2505 | Compressor Performance test: Unable to establish stable flow | Compressor fails to produce consistent, stable airflow during diagnostic | error |
FE2524 | AC power not detected during compressor performance test | Compressor performance test initiated without mains AC power available | error |
FS0001 | Unable to establish stable inspiratory flow | Flow sensor cross-check test failure | error |
FS0002 | Inspiratory flow sensor out of range | Flow accuracy test detects sensor reading outside limits | error |
FS0003 | Exhalation flow sensor out of range | Flow accuracy test detects sensor reading outside limits | error |
FS0004 | Mix flow sensor out of range | Flow accuracy test detects sensor reading outside limits | error |
FS0005 | PSOL current out of range | PSOL current accuracy test failure | error |
FS0021 | No gas connected | Gas connection test failure | error |
FS0024 | AC power not detected | AC power connection test failure | critical |
FS0101 | Exhalation valve not calibrated | EV calibration test failure | warning |
FS0102 | Unable to establish stable flow with EVQ installed | Flow stability test failure during EV performance test | error |
FS0103 | Exhalation valve pressure out of range | Pressure accuracy test during EV performance | error |
FS0105 | LVDT temperature out of range | Motor temperature test failure during EV performance | error |
FS0121 | No gas connected | Gas connection test failure during EV performance | error |
FS0124 | AC power not detected | AC power connection test failure during EV performance | critical |
FS0201 | Inspiratory zero offset out of range | Pressure offset test failure during circuit pressure check | error |
FS0202 | Expiratory zero offset out of range | Pressure offset test failure during circuit pressure check | error |
FS0203 | Target pressure not achieved | Pressure stability test failure during circuit pressure check | error |
FS0204 | Inspiratory and safety valve pressure difference outside limits | SV pressure accuracy test failure | error |
FS0205 | Inspiratory and exhalation pressure difference outside limits | Pressure sensor cross-check test failure | error |
FS0206 | Inspiratory auto zero solenoid not operational | Inspiratory pressure offset test failure | error |
FS0207 | Exhalation auto zero solenoid not operational | Exhalation pressure offset test failure | error |
FS0221 | No gas connected | Gas connection test failure during circuit pressure check | error |
FS0224 | AC power not detected | Power detection test failure during circuit pressure check | critical |
FS0303 | Unable to establish pressure in circuit leak test | Circuit leak test failure during SST leak check | error |
FS0304 | System leak failure | Circuit leak test detects excessive leakage | error |
FS0321 | No gas connected | Gas connection test failure during leak check | error |
FS0324 | AC power not detected | Power connection test failure during leak check | critical |
FS0401 | Unable to establish stable flow with exhalation filter | Flow stability test failure during exhalation filter check | error |
FS0403 | Occluded expiratory compartment | Exhalation filter resistance test detects excessive resistance | error |
FS0404 | Occluded expiratory filter | Exhalation filter resistance test detects blockage | error |
FS0406 | Patient circuit not disconnected | Circuit connection test during setup | warning |
FS0407 | Patient circuit not reconnected | Circuit connection test after service | warning |
FS0410 | Low expiratory filter pressure drop | Exhalation filter resistance test shows low resistance | error |
FS0421 | No gas connected | Gas connection test shows no air and oxygen supplies | critical |
FS0424 | AC power not detected | AC connection test fails | critical |
FS0501 | Unable to establish stable flow | Flow stability test cannot achieve stable flow | error |
FS0502 | Occluded inspiratory limb | Inspiratory limb pressure test detects high pressure | error |
FS0504 | Unable to reach minimum peak flow | Flow stability test cannot reach required flow | error |
FS0505 | Occluded expiratory limb | Expiratory limb pressure test detects high pressure | error |
FS0508 | Inspiratory limb resistance low | Inspiratory limb resistance test shows low resistance | warning |
FS0509 | Expiratory limb resistance low | Expiratory limb resistance test shows low resistance | warning |
FS0510 | Unable to reach minimum peak flow | Flow stability test cannot reach minimum flow | error |
FS0512 | Wye not blocked | Pressure build test shows low pressure in inspiratory limb | warning |
FS0521 | No gas connected | Gas connection test shows no air and oxygen supplies | critical |
FS0524 | AC power not detected | AC connection test shows AC power not connected | critical |
FS0601 | Unable to pressurize circuit | Circuit compliance test cannot reach required pressure points | error |
FS0602 | Unable to pressurize circuit | Circuit compliance test cannot pressurize | error |
FS0603 | Compliance calculation failure | Pressure build test detects wye not blocked causing leak | warning |
FS0604 | Excessive compliance | Circuit compliance test shows abnormally high compliance | warning |
FS0606 | Compliance low | Circuit compliance test shows low compliance | error |
FS0608 | Excessive compliance | Circuit compliance test shows abnormally high compliance | error |
FS0610 | Compliance low | Circuit compliance test shows low compliance | error |
FS0621 | No gas connected | Gas connection test shows air and O2 supplies not detected | critical |
FS0624 | AC power not detected | Power connection test shows AC power not connected | critical |
FS0709 | Gas not connected | Gas connection test for proximal flow sensor | error |
FS0721 | No gas connected | Gas connection test shows no gas detected | critical |
FS0724 | AC power not detected | Power connection test shows AC power not connected | critical |
FS0803 | Unable to establish pressure | Circuit pressure test cannot build pressure | error |
FS0804 | System leak detected during circuit leak test | SST circuit leak test failure | error |
FS0821 | Gas supply not connected | SST gas connection test failure | error |
FS0824 | AC power not detected | SST AC power connection test failure | error |
KB0004 | Expiratory motor current out of range | Expiratory motor current measurement outside acceptable range | error |
KP0013/XP0013 | Watchdog timer test | POST watchdog timer test | info |
KP0112 | Ventilator inoperative test failed | CPU or fundamental ventilator function failure detected during POST | critical |
KP0115 | Digital-to-analog converter (DAC) test failed | DAC malfunction detected for specific subsystem during POST | error |
KP0116 | Analog devices test failed in subsystem | Analog-to-digital converter or analog circuit failure in identified subsystem | error |
KP0120 | Safe state system test failed | Failure in safe shutdown or safety valve system during POST | critical |
KP0132 | Power distribution subsystem test failed | DC power delivery circuit failure or insufficient voltage during POST | error |
KP0133 | Compressor subsystem test failed | Compressor circuit or power supply failure detected during POST | error |
KP0134 | Mix subsystem test | POST diagnostic test initiated | error |
KP0135 | Inspiratory subsystem test | POST diagnostic test initiated | error |
KP0136 | Exhalation Subsystem test | POST diagnostic test initiated | error |
LB0002 | Exhalation flow OOR | Flow measurement out of range | error |
LB0005 | Ventilator 24 V bus power out of range | 24V power supply out of specification | critical |
LB0011 | Exhalation flow temp OOR | Thermistor reading out of range | error |
LB0021 | Atmospheric pressure out of range | Atmospheric pressure sensor reading outside acceptable range | warning |
LB0022 | O2 sensor OOR | Oxygen sensor measurement out of range | error |
LB0023 | O2 sensor OOR reset | O2 sensor out-of-range condition cleared | info |
LB0039 | BD time of day clock failed | Real time clock failure detected | warning |
LB0052 | Loss of GUI communications | GUI communication link lost | error |
LB0070 | Task monitor | Task monitoring event | info |
LB0088 | Proximal error | Proximal flow sensor reading error | error |
LB0090 | Mix inlet pressure OOR | Gas inlet pressure out of specification | warning |
LB0094 | Status display reported errors | Status display communication error | error |
LB0095 | Loss of status display communication | Status display communication loss | error |
LB0099 | Mix air flow OOR | Air flow measurement out of range | error |
LB0100 | Mix O2 flow OOR | Oxygen flow measurement out of range | error |
LB0115 | CO2 sensor fault | CO2 sensor reading error | error |
LB0116 | CO2 sensor zeroing initiated | CO2 sensor calibration procedure active | info |
LB0121 | PI 3.3V scaled out of range | PI 3.3V scaled reading out of acceptable range | warning |
LB0125 | PE 3.3V reference scaled out of range | PE 3.3V reference scaled reading out of acceptable range | warning |
LB0126 | Mix 5V supply out of range | Mix circuit 5V supply voltage out of acceptable range | error |
LB0127 | Mix 5V supply out of range | Mix circuit 5V supply voltage out of acceptable range | error |
LB0129 | Mix 12V supply out of range | Mix circuit 12V supply voltage out of acceptable range | error |
LB0130 | Mix 24V supply out of range | Mix circuit 24V supply voltage out of acceptable range | error |
LB0145 | Ventilator 24 V main supply power out of range | Main 24V power out of specification | critical |
LB0182 | High air inlet pressure | Air supply pressure exceeds limit | warning |
LB0183 | High O2 inlet pressure | O2 supply pressure exceeds limit | warning |
LB0194 | Inspiratory 24V supply out of range | Inspiratory 24V power supply voltage out of acceptable range | error |
LB0232 | BD NOVRAM operational time values out of range | CPU operational hours out of range | warning |
LB0234 | BD from GUI Operational time values out of range | BD time values outside expected range | warning |
LB0236 | Operational times values out of range | System operational hours out of range | warning |
LP0125 | Primary battery not installed or not detected by ventilator | Battery not present or not properly seated during power-on self test | error |
LP0127 | Expiratory flow sensor cannot be read by ventilator | Exhalation valve flow sensor (EVQ) electrical failure or connection issue | error |
UXXXXX | Assertion code (undefined content) | Unknown assertion error | info |
XB0080 | Init loss of BD communication | BD communication lost during initialization | info |
XP0097 | SAAS self test start | Self-test initiated | error |
ZB0035 | Touch screen blocked | Touch screen input disabled | info |
ZB0036 | Touch screen resumed | Touch screen input re-enabled | info |
ZB0040 | GUI time of day clock failed | GUI real time clock failure detected | warning |
ZB0053 | Loss of BD communications | BD communication link lost | error |
ZB0071 | GUI drops a waveform packet | Waveform data packet loss detected | info |
ZB0103 | USB failure | USB interface failure detected | critical |
ZB0117 | UI board communications error | UI board communication issue | info |
ZB0118 | Video lock error | UI to BDU communication error | error |
ZB0119 | UI heartbeat error | UI board heartbeat signal lost | info |
ZB0178 | Offscreen keys driver event | Key input off screen detected | info |
ZB0179 | Logs media failure | Log storage medium failure | info |
ZB0180 | Logs initialization failure | Log system initialization error | info |
ZB0181 | Logs transaction failure | Log write transaction failed | info |
ZB0191 | Unable to save config files to USB | USB save operation failed | info |
ZB0192 | Unable to load config files from USB | USB load operation failed | info |
ZB0231 | USB operational time error detected | USB time tracking error | info |
ZB0235 | GUI from BD operational time values out of range | GUI time values outside expected range | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI