Medtronic PULSAR: troubleshooting and error codes

The ServiceForge knowledge base holds 5 documented symptoms for the Medtronic PULSAR, with 5 ranked causes, 17 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Cut 3 output power (PULSAR II) high voltage

Verification that CUT mode 3 on PULSAR II outputs 6 Watts at 100Ω impedance within ±20% tolerance per IEC 60601-2-2

Ranked causes

  1. 1. Routine calibration check per IEC 60601-2-2 standard (common)
First check

Set ESU load to 100 Ohms

Test point
ESU load control
Expected
100 Ω selected

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Cut 4 output power (PULSAR II) high voltage

Verification that CUT mode 4 on PULSAR II outputs 10 Watts at 100Ω impedance within ±20% tolerance

Ranked causes

  1. 1. Routine calibration check per IEC 60601-2-2 standard (common)
First check

Set ESU load to 100 Ohms

Test point
ESU load control
Expected
100 Ω selected

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Cut 5 output power (PULSAR II) high voltage

Verification that CUT mode 5 on PULSAR II outputs 20 Watts at 100Ω impedance within ±20% tolerance

Ranked causes

  1. 1. Routine calibration check per IEC 60601-2-2 standard (common)
First check

Set ESU load to 100 Ohms

Test point
ESU load control
Expected
100 Ω selected

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Load resistance not matching generator model

Incorrect load resistance used for output power measurement; affects measurement accuracy

Ranked causes

  1. 1. Wrong load resistance selected for generator model (common)
First check

Verify load resistance matches generator model

Test point
test load
Expected
100 ohm for Pulsar II or 40 ohm for Pulsar I
Tools
ohmmeter

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Verify RF output power accuracy high voltage

Measurement of radio frequency energy output across CUT and COAG modes at various power levels to confirm generator compliance with IEC 60601-2-2 standard

Ranked causes

  1. 1. Routine preventive maintenance calibration required (common)
First check

Connect PEAK PlasmaBlade device to generator under test

Test point
Generator RF output connector
Expected
Device connected and recognized

6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI