Medtronic FT10: troubleshooting and error codes

The ServiceForge knowledge base holds 19 documented symptoms for the Medtronic FT10, with 34 ranked causes, 27 diagnostic checks with test points and expected values, and 125 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Energy platform does not respond when turned on

Unit does not power on or respond to power controls

Ranked causes

  1. 1. Disconnected power cord or faulty wall outlet (common)
  2. 2. Faulty power cord (common)
  3. 3. Fuse drawer open or fuses blown (occasional)
  4. 4. Internal component malfunction (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Energy platform is on and instrument activated, but no output delivered patient safety

Unit powers up and foot pedal/hand switch activated but no energy output to instrument

Ranked causes

  1. 1. Malfunctioning foot pedal or hand switching instrument (common)
  2. 2. Power set too low or turned off (common)
  3. 3. Alarm condition exists (occasional)
  4. 4. Internal component malfunction (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormal neuromuscular stimulation patient safety

Patient experiences unexpected muscle stimulation during procedure; requires immediate surgery stop

Ranked causes

  1. 1. Metal-to-metal sparking (common)
  2. 2. Use of COAG mode with high power settings (occasional)
  3. 3. Abnormal 50-60 Hz leakage currents (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous monitor interference caution

Persistent interference affecting monitor display or function

Ranked causes

  1. 1. Malfunctioning monitor (common)
  2. 2. Faulty chassis-to-ground connections (occasional)
  3. 3. Electrical equipment grounded to different objects (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Interference with other devices when energy platform is activated caution

Other equipment experiences interference only when platform energy output is active

Ranked causes

  1. 1. Metal-to-metal sparking (common)
  2. 2. High settings used for FULGURATION (common)
  3. 3. Electrically inconsistent ground wires in operating room (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pacemaker interference patient safety

Pacemaker malfunction or interference during electrosurgery procedure

Ranked causes

  1. 1. Intermittent connections or metal-to-metal sparking (common)
  2. 2. Current path passing too close to pacemaker (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System is on, but did not complete self-test caution

Unit powers on but self-test sequence fails to complete

Ranked causes

  1. 1. Software malfunction (common)
  2. 2. Internal component malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touchscreen button does not respond to touch

Touch interface does not register user input on buttons

Ranked causes

  1. 1. Touchscreen is out of calibration (common)
  2. 2. Touchscreen is broken (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: High-frequency leakage current patient safety

Verify leakage current meets safety limits after calibration or annually before clinical use

Ranked causes

  1. 1. Leakage current exceeds allowable limits (unknown)
First check

Monopolar 1 leakage current at 120 W SPRAY mode

Test point
Monopolar 1 left tine to equipotential ground lug through current transformer
Expected
not exceed 100 mA
Tools
200 Ω 250 W 1% load, current transformer, true RMS meter

15 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Foot pedal stuck or pressed

Mono 1 foot pedal is being held in pressed position continuously

Ranked causes

  1. 1. Foot pedal stuck or jammed (common)
First check

Foot pedal in released position

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Handswitch stuck or pressed

Mono 1 handswitch is being held in pressed position, blocking other operations

Ranked causes

  1. 1. Handswitch mechanically stuck or jammed (common)
First check

Handswitch in released position

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High-frequency leakage current exceeds limits patient safety

Measured leakage current during any mode exceeds safety threshold, indicating potential electrical hazard

Ranked causes

  1. 1. Leakage current out of specification (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal Cardiac Defibrillator (ICD) activation patient safety

Patient's ICD triggered unintentionally during energy platform use

Ranked causes

  1. 1. ICD is activated by energy platform output (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM alarm condition - monopolar energy disabled patient safety

REM indicator on front panel illuminates red, REM symbol appears and enlarges on touchscreen, alarm tone sounds twice, RF energy disabled

Ranked causes

  1. 1. Incorrect impedance sensed in REM Polyhesive patient return electrode (common)
First check

Inspect return-electrode plug and cord

Test point
REM Polyhesive electrode connector and cable
Expected
No visible cracks, breaks or damage

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

REM pad not detected patient safety

REM pad contact quality issue preventing proper patient grounding connection

Ranked causes

  1. 1. REM pad not connected (common)
First check

REM pad physically connected

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not detect bipolar instrument

Bipolar instrument not recognized by energy platform

Ranked causes

  1. 1. Bipolar connector not firmly inserted (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not detect monopolar instrument

Monopolar instrument not recognized by energy platform

Ranked causes

  1. 1. Monopolar connector not firmly inserted (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

System does not detect tissue-fusion instrument

LigaSmart instrument not recognized by energy platform

Ranked causes

  1. 1. LigaSmart connector not firmly inserted (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unintentional screen input detected caution

Touchscreen is registering unintended touches or input

Ranked causes

  1. 1. Touchscreen contamination or malfunction (common)
First check

Touchscreen surface clean and dry

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
E101FPGA FailureSystem initialization firmware errorcritical
E102CPLD FailureComplex programmable logic device errorcritical
E103FPGA FailureSystem initialization firmware errorcritical
E104Boot Sector FailureSystem boot memory corruptioncritical
E105RAM Test FailureMemory system error during self-testcritical
E106Processor ExceptionCPU exception during operationcritical
E107Unexpected InterruptUndefined interrupt signal receivedcritical
E201Storage FailureData storage system errorcritical
E202RTOS FailureReal-time operating system failurecritical
E203OS FailureOperating system failurecritical
E204Assert ErrorSoftware assertion failurecritical
E205Null Pointer ErrorSoftware null pointer dereferencecritical
E206Database ErrorData storage or retrieval failurecritical
E207Memory Allocation ErrorInsufficient system memorycritical
E208Stack OverflowSoftware stack overflowcritical
E209DMA FailureDirect memory access errorcritical
E210Thread Limit ErrorSystem error condition exceededcritical
E211Internal Messaging ErrorInternal communication failurecritical
E212PCBA FailureController PCB malfunctioncritical
E213SR PCBA FailureSteering Relay PCB malfunctioncritical
E214RF PCBA FailureRF PCB malfunctioncritical
E215Software State ErrorSoftware state machine errorcritical
E216Low Voltage Rail FailurePower supply low voltage rail faultcritical
E217High Voltage Rail FailurePower supply high voltage rail faultcritical
E218Display Driver ErrorDisplay driver circuit failureerror
E219Audio Circuit FailureAudio circuit malfunctionerror
E220Multi-tasking ErrorTask scheduling or execution failurecritical
E221Processor Timebase ErrorProcessor timing reference failurecritical
E222Processor Timebase ErrorProcessor timing reference failurecritical
E223FPGA Timebase ErrorFPGA timing reference failurecritical
E224System Timebase ErrorSystem timing reference failurecritical
E225Processor Watchdog ErrorProcessor watchdog timeoutcritical
E226FPCA Comm ErrorCommunication failure with FPGAcritical
E227REM Circuit FailurePatient return electrode circuit faultcritical
E228Relay Feedback FailureSteering relay feedback errorcritical
E229Connector Diagnostic FailureCable connection fault detectederror
E230Critical Data FailureSystem data integrity errorcritical
E231Security Access ErrorAuthentication or authorization failurewarning
E232Invalid System ConfigurationSystem configuration mismatch detectedwarning
E233Field Replaceable Unit UpdatedFRU change detected during bootwarning
E234Processor Comm FailureProcessor communication errorcritical
E235REM Pad Contact QualityPatient return electrode pad poor contactwarning
E236Processor Watchdog ErrorProcessor watchdog timeoutcritical
E238Software FailureSoftware execution errorcritical
E239Software FailureSoftware execution errorcritical
E240Thread Check FailureThread synchronization errorcritical
E241RF Data ErrorRF signal data corruptioncritical
E242RF Data Timing ErrorRF data timing violationcritical
E243RF Data Delay ErrorRF data delay out of tolerancecritical
E244RF Interrupt ErrorRF interrupt handling failurecritical
E246HWC Control Selection ErrorHandpiece control selection failurecritical
E247PS PCBA FailurePower supply PCB failurecritical
E248FP PCBA FailureFront panel PCB failurecritical
E301Network Diagnostic IssueNetwork communication error detectederror
E302Mono 1 Port Comm IssueMonopolar 1 port communication failureerror
E303Mono 2 Port Comm IssueMonopolar 2 port communication failureerror
E304Bipolar Port Comm IssueBipolar port communication failureerror
E305LigaSure Port Comm IssueLigaSure port communication failureerror
E306Instrument Scanner IssueInstrument identification system failureerror
E307Unsupported CommandSystem received unsupported commandwarning
E308Software Timeout IssueSoftware operation timeout exceededcritical
E309Activation Timing IssueActivation sequence timing errorerror
E310Unintentional Screen InputTouchscreen detecting unintended inputunknown
E312Absolute Integrated ErrorSystem self-test absolute integration failureerror
E313Voltage Sensor Compare ErrorVoltage sensor comparison mismatcherror
E314Current Sensor Compare ErrorCurrent sensor comparison mismatcherror
E315Power Sensor Compare ErrorPower sensor comparison mismatcherror
E316Power Compare ErrorPower calculation comparison mismatcherror
E317PVI Curve Compare ErrorImpedance curve comparison mismatcherror
E318Voltage Sensor Clip ErrorVoltage sensor clipping detectederror
E319Current Sensor Clip ErrorCurrent sensor clipping detectederror
E320Check InstrumentMono 1 handswitch stuck in pressed stateunknown
E321Check Foot PedalMono 1 foot pedal stuck in pressed stateunknown
E322Mono 2 handswitch pressed and stuckHandswitch activation detected when not desirederror
E323Mono 2 foot-pedal pressed and stuckFoot-pedal activation detected when not desirederror
E324Bipolar handswitch pressed and stuckHandswitch activation detected when not desirederror
E325Bipolar foot-pedal pressed and stuckFoot-pedal activation detected when not desirederror
E326LigaSure handswitch pressed and stuckHandswitch activation detected when not desirederror
E327LigaSure foot-pedal pressed and stuckFoot-pedal activation detected when not desirederror
E328Unintentional input panel button errorFront panel button pressed when not intendedwarning
E329Invalid foot-pedal input detectedFoot-pedal sends invalid signal to controllererror
E330Prolonged activation detectedSystem detects sustained button/pedal activationwarning
E331Invalid instrument connectedInstrument does not match system configurationerror
E332Unit overtemperature detectedInternal temperature exceeds safe operating limitcritical
E333Cooling fan issue detectedCooling fan malfunction or excessive noisecritical
E334REM automated check errorRemote (REM) self-test fails at startupwarning
E335Driver communication issueCommunication error between driver circuitscritical
E336ADC synchronization issueAnalog-to-digital converter sync error during startupwarning
E337Real time clock issueReal-time clock fails at startup or during operationwarning
E339RF control limit check issueRF (radio frequency) power control limits exceeded or failedwarning
E340Configuration mismatch - invalid country codeUnit country configuration does not match operational environmentwarning
E341Panel communication issueCommunication fails between main controller and display panelwarning
E342Foot-pedal communication issueCommunication fails with foot-pedal controllerwarning
E343REM communication issueCommunication fails with remote (REM) controllerwarning
E344REM communication issue (duplicate)Communication fails with remote controllerwarning
E345REM communication issue (duplicate)Communication fails with remote controllerwarning
E346REM communication issue (duplicate)Communication fails with remote controllerwarning
E347REM data integrity issueRemote controller data validation failswarning
E348Software assert issueSoftware assertion fails during runtimecritical
E349Thread check issueSoftware thread execution fails or deadlock detectedcritical
E350Return electrode overtemperatureReturn electrode exceeds safe temperature limitunknown
E351Processor watchdog errorProcessor watchdog timer expires indicating hung processorcritical
E352Software update issueFirmware or software update failscritical
E353Touchscreen calibration reset requiredTouchscreen calibration lost or invalidwarning
E354SWC target variance issueSoftware/firmware CRC or target validation failswarning
E355Foot-pedal not supported for this activationAttempted foot-pedal activation when not availableinfo
E401Invalid pairing code during authorizationPairing code validation failserror
E402Incomplete seal cycle - operator techniqueSeal cycle did not complete properlywarning
E403Incomplete seal cycle - tissue issueSeal cycle did not complete due to tissue conditionwarning
E404Incomplete seal cycle - insufficient gripTissue not gripped firmly enough during sealwarning
E405Incomplete seal cycle - jaw contaminationSeal cycle failed due to wet jawswarning
E407Instrument not ready - dual modality instrument missing plugDual modality instrument requires second plug connectioninfo
E408Instrument not ready - dual modality instrument missing plugDual modality instrument requires second plug connectioninfo
E411Invalid activation attempt - dual modality missing plugAttempted activation with only one modality pluginfo
E412Invalid activation attempt - dual modality missing plugAttempted activation with only one modality pluginfo
E413Invalid instrument combination - incompatible instrumentTwo incompatible instruments connected simultaneouslywarning
E414Invalid instrument combination - incompatible instrumentTwo incompatible instruments connected simultaneouslywarning
E415Bipolar Resection feature not enabledAttempted use of disabled featureinfo
E416Unknown instrument detectedConnected instrument not recognized by systemerror
E419Incomplete setup - electrode tips stuck togetherForceps tips stuck or not fully separated before usewarning
E420Usage limit - reusable device already usedReusable device previously used and not recertifiedwarning
E421Usage limit - reusable device already usedReusable device previously used and not recertifiedwarning
E422Incomplete setup - cable issue or electrode tips stuckCable insertion problem or tips not separatedwarning
E424Activation time exceededRF activation held longer than safe limitwarning
E425Contact time exceededTissue contact sustained beyond safety thresholdwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.