Medtronic FT10: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Medtronic FT10, with 34 ranked causes, 27 diagnostic checks with test points and expected values, and 125 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Energy platform does not respond when turned on
Unit does not power on or respond to power controls
Ranked causes
- 1. Disconnected power cord or faulty wall outlet (common)
- 2. Faulty power cord (common)
- 3. Fuse drawer open or fuses blown (occasional)
- 4. Internal component malfunction (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Energy platform is on and instrument activated, but no output delivered patient safety
Unit powers up and foot pedal/hand switch activated but no energy output to instrument
Ranked causes
- 1. Malfunctioning foot pedal or hand switching instrument (common)
- 2. Power set too low or turned off (common)
- 3. Alarm condition exists (occasional)
- 4. Internal component malfunction (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal neuromuscular stimulation patient safety
Patient experiences unexpected muscle stimulation during procedure; requires immediate surgery stop
Ranked causes
- 1. Metal-to-metal sparking (common)
- 2. Use of COAG mode with high power settings (occasional)
- 3. Abnormal 50-60 Hz leakage currents (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous monitor interference caution
Persistent interference affecting monitor display or function
Ranked causes
- 1. Malfunctioning monitor (common)
- 2. Faulty chassis-to-ground connections (occasional)
- 3. Electrical equipment grounded to different objects (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Interference with other devices when energy platform is activated caution
Other equipment experiences interference only when platform energy output is active
Ranked causes
- 1. Metal-to-metal sparking (common)
- 2. High settings used for FULGURATION (common)
- 3. Electrically inconsistent ground wires in operating room (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker interference patient safety
Pacemaker malfunction or interference during electrosurgery procedure
Ranked causes
- 1. Intermittent connections or metal-to-metal sparking (common)
- 2. Current path passing too close to pacemaker (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System is on, but did not complete self-test caution
Unit powers on but self-test sequence fails to complete
Ranked causes
- 1. Software malfunction (common)
- 2. Internal component malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen button does not respond to touch
Touch interface does not register user input on buttons
Ranked causes
- 1. Touchscreen is out of calibration (common)
- 2. Touchscreen is broken (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: High-frequency leakage current patient safety
Verify leakage current meets safety limits after calibration or annually before clinical use
Ranked causes
- 1. Leakage current exceeds allowable limits (unknown)
Monopolar 1 leakage current at 120 W SPRAY mode
- Test point
- Monopolar 1 left tine to equipotential ground lug through current transformer
- Expected
- not exceed 100 mA
- Tools
- 200 Ω 250 W 1% load, current transformer, true RMS meter
15 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Foot pedal stuck or pressed
Mono 1 foot pedal is being held in pressed position continuously
Ranked causes
- 1. Foot pedal stuck or jammed (common)
Foot pedal in released position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handswitch stuck or pressed
Mono 1 handswitch is being held in pressed position, blocking other operations
Ranked causes
- 1. Handswitch mechanically stuck or jammed (common)
Handswitch in released position
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
High-frequency leakage current exceeds limits patient safety
Measured leakage current during any mode exceeds safety threshold, indicating potential electrical hazard
Ranked causes
- 1. Leakage current out of specification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal Cardiac Defibrillator (ICD) activation patient safety
Patient's ICD triggered unintentionally during energy platform use
Ranked causes
- 1. ICD is activated by energy platform output (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM alarm condition - monopolar energy disabled patient safety
REM indicator on front panel illuminates red, REM symbol appears and enlarges on touchscreen, alarm tone sounds twice, RF energy disabled
Ranked causes
- 1. Incorrect impedance sensed in REM Polyhesive patient return electrode (common)
Inspect return-electrode plug and cord
- Test point
- REM Polyhesive electrode connector and cable
- Expected
- No visible cracks, breaks or damage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
REM pad not detected patient safety
REM pad contact quality issue preventing proper patient grounding connection
Ranked causes
- 1. REM pad not connected (common)
REM pad physically connected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not detect bipolar instrument
Bipolar instrument not recognized by energy platform
Ranked causes
- 1. Bipolar connector not firmly inserted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not detect monopolar instrument
Monopolar instrument not recognized by energy platform
Ranked causes
- 1. Monopolar connector not firmly inserted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not detect tissue-fusion instrument
LigaSmart instrument not recognized by energy platform
Ranked causes
- 1. LigaSmart connector not firmly inserted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unintentional screen input detected caution
Touchscreen is registering unintended touches or input
Ranked causes
- 1. Touchscreen contamination or malfunction (common)
Touchscreen surface clean and dry
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E101 | FPGA Failure | System initialization firmware error | critical |
E102 | CPLD Failure | Complex programmable logic device error | critical |
E103 | FPGA Failure | System initialization firmware error | critical |
E104 | Boot Sector Failure | System boot memory corruption | critical |
E105 | RAM Test Failure | Memory system error during self-test | critical |
E106 | Processor Exception | CPU exception during operation | critical |
E107 | Unexpected Interrupt | Undefined interrupt signal received | critical |
E201 | Storage Failure | Data storage system error | critical |
E202 | RTOS Failure | Real-time operating system failure | critical |
E203 | OS Failure | Operating system failure | critical |
E204 | Assert Error | Software assertion failure | critical |
E205 | Null Pointer Error | Software null pointer dereference | critical |
E206 | Database Error | Data storage or retrieval failure | critical |
E207 | Memory Allocation Error | Insufficient system memory | critical |
E208 | Stack Overflow | Software stack overflow | critical |
E209 | DMA Failure | Direct memory access error | critical |
E210 | Thread Limit Error | System error condition exceeded | critical |
E211 | Internal Messaging Error | Internal communication failure | critical |
E212 | PCBA Failure | Controller PCB malfunction | critical |
E213 | SR PCBA Failure | Steering Relay PCB malfunction | critical |
E214 | RF PCBA Failure | RF PCB malfunction | critical |
E215 | Software State Error | Software state machine error | critical |
E216 | Low Voltage Rail Failure | Power supply low voltage rail fault | critical |
E217 | High Voltage Rail Failure | Power supply high voltage rail fault | critical |
E218 | Display Driver Error | Display driver circuit failure | error |
E219 | Audio Circuit Failure | Audio circuit malfunction | error |
E220 | Multi-tasking Error | Task scheduling or execution failure | critical |
E221 | Processor Timebase Error | Processor timing reference failure | critical |
E222 | Processor Timebase Error | Processor timing reference failure | critical |
E223 | FPGA Timebase Error | FPGA timing reference failure | critical |
E224 | System Timebase Error | System timing reference failure | critical |
E225 | Processor Watchdog Error | Processor watchdog timeout | critical |
E226 | FPCA Comm Error | Communication failure with FPGA | critical |
E227 | REM Circuit Failure | Patient return electrode circuit fault | critical |
E228 | Relay Feedback Failure | Steering relay feedback error | critical |
E229 | Connector Diagnostic Failure | Cable connection fault detected | error |
E230 | Critical Data Failure | System data integrity error | critical |
E231 | Security Access Error | Authentication or authorization failure | warning |
E232 | Invalid System Configuration | System configuration mismatch detected | warning |
E233 | Field Replaceable Unit Updated | FRU change detected during boot | warning |
E234 | Processor Comm Failure | Processor communication error | critical |
E235 | REM Pad Contact Quality | Patient return electrode pad poor contact | warning |
E236 | Processor Watchdog Error | Processor watchdog timeout | critical |
E238 | Software Failure | Software execution error | critical |
E239 | Software Failure | Software execution error | critical |
E240 | Thread Check Failure | Thread synchronization error | critical |
E241 | RF Data Error | RF signal data corruption | critical |
E242 | RF Data Timing Error | RF data timing violation | critical |
E243 | RF Data Delay Error | RF data delay out of tolerance | critical |
E244 | RF Interrupt Error | RF interrupt handling failure | critical |
E246 | HWC Control Selection Error | Handpiece control selection failure | critical |
E247 | PS PCBA Failure | Power supply PCB failure | critical |
E248 | FP PCBA Failure | Front panel PCB failure | critical |
E301 | Network Diagnostic Issue | Network communication error detected | error |
E302 | Mono 1 Port Comm Issue | Monopolar 1 port communication failure | error |
E303 | Mono 2 Port Comm Issue | Monopolar 2 port communication failure | error |
E304 | Bipolar Port Comm Issue | Bipolar port communication failure | error |
E305 | LigaSure Port Comm Issue | LigaSure port communication failure | error |
E306 | Instrument Scanner Issue | Instrument identification system failure | error |
E307 | Unsupported Command | System received unsupported command | warning |
E308 | Software Timeout Issue | Software operation timeout exceeded | critical |
E309 | Activation Timing Issue | Activation sequence timing error | error |
E310 | Unintentional Screen Input | Touchscreen detecting unintended input | unknown |
E312 | Absolute Integrated Error | System self-test absolute integration failure | error |
E313 | Voltage Sensor Compare Error | Voltage sensor comparison mismatch | error |
E314 | Current Sensor Compare Error | Current sensor comparison mismatch | error |
E315 | Power Sensor Compare Error | Power sensor comparison mismatch | error |
E316 | Power Compare Error | Power calculation comparison mismatch | error |
E317 | PVI Curve Compare Error | Impedance curve comparison mismatch | error |
E318 | Voltage Sensor Clip Error | Voltage sensor clipping detected | error |
E319 | Current Sensor Clip Error | Current sensor clipping detected | error |
E320 | Check Instrument | Mono 1 handswitch stuck in pressed state | unknown |
E321 | Check Foot Pedal | Mono 1 foot pedal stuck in pressed state | unknown |
E322 | Mono 2 handswitch pressed and stuck | Handswitch activation detected when not desired | error |
E323 | Mono 2 foot-pedal pressed and stuck | Foot-pedal activation detected when not desired | error |
E324 | Bipolar handswitch pressed and stuck | Handswitch activation detected when not desired | error |
E325 | Bipolar foot-pedal pressed and stuck | Foot-pedal activation detected when not desired | error |
E326 | LigaSure handswitch pressed and stuck | Handswitch activation detected when not desired | error |
E327 | LigaSure foot-pedal pressed and stuck | Foot-pedal activation detected when not desired | error |
E328 | Unintentional input panel button error | Front panel button pressed when not intended | warning |
E329 | Invalid foot-pedal input detected | Foot-pedal sends invalid signal to controller | error |
E330 | Prolonged activation detected | System detects sustained button/pedal activation | warning |
E331 | Invalid instrument connected | Instrument does not match system configuration | error |
E332 | Unit overtemperature detected | Internal temperature exceeds safe operating limit | critical |
E333 | Cooling fan issue detected | Cooling fan malfunction or excessive noise | critical |
E334 | REM automated check error | Remote (REM) self-test fails at startup | warning |
E335 | Driver communication issue | Communication error between driver circuits | critical |
E336 | ADC synchronization issue | Analog-to-digital converter sync error during startup | warning |
E337 | Real time clock issue | Real-time clock fails at startup or during operation | warning |
E339 | RF control limit check issue | RF (radio frequency) power control limits exceeded or failed | warning |
E340 | Configuration mismatch - invalid country code | Unit country configuration does not match operational environment | warning |
E341 | Panel communication issue | Communication fails between main controller and display panel | warning |
E342 | Foot-pedal communication issue | Communication fails with foot-pedal controller | warning |
E343 | REM communication issue | Communication fails with remote (REM) controller | warning |
E344 | REM communication issue (duplicate) | Communication fails with remote controller | warning |
E345 | REM communication issue (duplicate) | Communication fails with remote controller | warning |
E346 | REM communication issue (duplicate) | Communication fails with remote controller | warning |
E347 | REM data integrity issue | Remote controller data validation fails | warning |
E348 | Software assert issue | Software assertion fails during runtime | critical |
E349 | Thread check issue | Software thread execution fails or deadlock detected | critical |
E350 | Return electrode overtemperature | Return electrode exceeds safe temperature limit | unknown |
E351 | Processor watchdog error | Processor watchdog timer expires indicating hung processor | critical |
E352 | Software update issue | Firmware or software update fails | critical |
E353 | Touchscreen calibration reset required | Touchscreen calibration lost or invalid | warning |
E354 | SWC target variance issue | Software/firmware CRC or target validation fails | warning |
E355 | Foot-pedal not supported for this activation | Attempted foot-pedal activation when not available | info |
E401 | Invalid pairing code during authorization | Pairing code validation fails | error |
E402 | Incomplete seal cycle - operator technique | Seal cycle did not complete properly | warning |
E403 | Incomplete seal cycle - tissue issue | Seal cycle did not complete due to tissue condition | warning |
E404 | Incomplete seal cycle - insufficient grip | Tissue not gripped firmly enough during seal | warning |
E405 | Incomplete seal cycle - jaw contamination | Seal cycle failed due to wet jaws | warning |
E407 | Instrument not ready - dual modality instrument missing plug | Dual modality instrument requires second plug connection | info |
E408 | Instrument not ready - dual modality instrument missing plug | Dual modality instrument requires second plug connection | info |
E411 | Invalid activation attempt - dual modality missing plug | Attempted activation with only one modality plug | info |
E412 | Invalid activation attempt - dual modality missing plug | Attempted activation with only one modality plug | info |
E413 | Invalid instrument combination - incompatible instrument | Two incompatible instruments connected simultaneously | warning |
E414 | Invalid instrument combination - incompatible instrument | Two incompatible instruments connected simultaneously | warning |
E415 | Bipolar Resection feature not enabled | Attempted use of disabled feature | info |
E416 | Unknown instrument detected | Connected instrument not recognized by system | error |
E419 | Incomplete setup - electrode tips stuck together | Forceps tips stuck or not fully separated before use | warning |
E420 | Usage limit - reusable device already used | Reusable device previously used and not recertified | warning |
E421 | Usage limit - reusable device already used | Reusable device previously used and not recertified | warning |
E422 | Incomplete setup - cable issue or electrode tips stuck | Cable insertion problem or tips not separated | warning |
E424 | Activation time exceeded | RF activation held longer than safe limit | warning |
E425 | Contact time exceeded | Tissue contact sustained beyond safety threshold | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI