Medtronic Battery Support 2: troubleshooting and error codes

The ServiceForge knowledge base holds 30 documented symptoms for the Medtronic Battery Support 2, with 37 ranked causes, 80 diagnostic checks with test points and expected values, and 29 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Performance Inspection Procedure (PIP) setup verification patient safety

Ensure proper equipment calibration, workstation power grounding, and personnel qualifications before performing Performance Inspection Procedure on Battery Support System 2

Ranked causes

  1. 1. Test equipment lacking current calibration (common)
  2. 2. Workstation power not grounded (occasional)
  3. 3. Personnel insufficiently qualified (occasional)
First check

Inspect calibration labels on all PIP test equipment

Test point
Equipment being used for PIP
Expected
Current calibration label present, issued by certified facility

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage measurement test failure or DISCARD LED lights

During voltage measurement test, DISCARD LED illuminates instead of A-READY LED, indicating device or power supply configuration error

Ranked causes

  1. 1. Reversed polarity connection between power supply and device (common)
  2. 2. Device failure or fault (occasional)
  3. 3. Incorrect dc power supply setting (common)
First check

Verify polarity of power supply connection

Test point
Device positive (+) battery pin and power supply positive (+) terminal
Expected
Positive connected to positive, negative to negative

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

A-READY LED does not light within 15 seconds patient safety

During I2C bus test, the A-READY LED fails to illuminate within the expected 15-second window.

Ranked causes

  1. 1. Battery I2C communication failure (common)
  2. 2. Device I2C bus communication failure (occasional)
First check

Verify A-READY LED lights within 15 seconds after pressing ENTER

Test point
A-READY LED
Expected
LED lights within 15 seconds

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Event codes stored in log

Device displays event codes in Event Log; codes may indicate transient EMI, ESD, or errors

Ranked causes

  1. 1. Transient electromagnetic interference (EMI) (occasional)
  2. 2. Electrostatic discharge (ESD) (occasional)
First check

Unplug Battery Support System 2 from ac line power

Test point
ac line power connection
Expected
Device unpowered

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage Measurement Test failure patient safety

Device fails to measure terminal voltage of installed batteries at discrete levels; DISCARD LED lights instead of A-READY LED within 15 seconds

Ranked causes

  1. 1. Reversed polarity connection between power supply and device (common)
  2. 2. Device failure or incorrect dc power supply setting (occasional)
First check

Verify polarity connection between power supply and device

Test point
Device positive (+) battery pin and power supply positive (+) terminal
Expected
Power supply positive terminal connected to device positive pin

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery connector pins bent, broken, or corroded patient safety

Battery connector pins exhibit visible damage including bending, breaks, corrosion, or wear

Ranked causes

  1. 1. Physical damage to connector pins (occasional)
First check

Visually examine each leaf on connector pins

Test point
battery connector pin leaves
Expected
No cracks or breaks visible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Keypad Test patient safety

Verify front panel keypad responsiveness and LED illumination during Performance Inspection Mode

Ranked causes

  1. 1. Keypad not properly calibrated or keys not responding (unknown)
First check

Enter Performance Inspection Mode Keypad Test passcode

Test point
Front panel keys
Expected
Device momentarily illuminates all three ready legends, then continuously flashes all three charge LEDs (Test Number Entry Mode)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: LIFEPAK SLA I2C Test patient safety

Performance inspection procedure for battery support system battery cells using test mode

Ranked causes

  1. 1. Battery Well C requires testing (common)
First check

Test battery Well C shelf life

Test point
Well C test number 123-A
Expected
As per shelf life specification

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Low/Medium Current Discharge Test patient safety

Performance inspection procedure for Battery Support System 2 current discharge testing across three battery wells

Ranked causes

  1. 1. Well A, B, and C discharge validation required (unknown)
First check

Execute Low Current Discharge Test for Well A

Test point
Well A
Expected
Test number 221 (A-CONDITION, B-CONDITION, C-CONDITION)

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Performance inspection mode tests patient safety

Verify device diagnostic tests: Power-On LED sequence, Keypad input, Fan operation, Battery I2C communication, Thermistor limits, Charge current levels, Discharge current levels

Ranked causes

  1. 1. Verification of system diagnostic capability (common)
First check

Power-On LED Test sequence execution

Test point
Front panel LEDs
Expected
All LEDs light in predetermined sequence
Tools
Visual inspection

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Performance Inspection Procedure patient safety

Verify Battery Support System 2 functionality using specified test equipment

Ranked causes

  1. 1. Equipment setup and validation for PIP (unknown)
First check

Verify LIFEPARK NiCd battery availability

Test point
Test equipment inventory
Expected
Medtronic MIN 3009376 NiCd battery with fuel gauge present

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Performance Inspection Procedure (PIP) patient safety

Equipment verification and workstation setup for PIP testing of Battery Support System 2

Ranked causes

  1. 1. Test equipment not properly calibrated (common)
First check

Verify test equipment has current calibration label

Test point
Test equipment
Expected
Current calibration label from certified calibration facility

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PIP Test Equipment Requirements

Selection and verification of appropriate test equipment for Performance Inspection Procedure

Ranked causes

  1. 1. Incorrect or non-equivalent test equipment used (common)
First check

Verify test equipment meets specifications

Test point
Test equipment
Expected
Equipment listed in PIP Test Equipment Requirements or equivalent specifications

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PIP Voltage Measurement high voltage

Performance Inspection Procedure voltage measurement test for battery support wells A, B, C using high and low voltage test numbers

Ranked causes

  1. 1. Routine calibration and verification procedure (unknown)
First check

Connect device to ac line power

Test point
ac line connection
Expected
Device powered and ready for test entry

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Thermistor Test patient safety

Performance Inspection Procedure test for thermistor function across three battery wells using LIFEPAK NiCd batteries

Ranked causes

  1. 1. Thermistor malfunction or battery failure (unknown)
First check

Install LIFEPAK NiCd battery in Well A

Test point
Well A
Expected
Battery installed

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chassis leakage current exceeds acceptable limits patient safety

Safety analyzer measures excessive leakage current from battery wells during performance inspection procedure

Ranked causes

  1. 1. Leakage current fault in battery well circuitry (unknown)
First check

Measure chassis leakage current at battery well negative pin (normal mode)

Test point
Battery Well A negative (-) pin
Expected
less than 100 µa, <100 µa
Tools
Safety analyzer

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Chassis leakage current exceeds limit high voltage

Safety analyzer detects excessive leakage current from chassis to ground during performance inspection procedure

Ranked causes

  1. 1. Chassis insulation degradation or fault (unknown)
First check

Measure chassis leakage current (grounded analyzer)

Test point
Well A negative battery pin
Expected
less than 100 µa, < 100 µa
Tools
Safety analyzer

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chassis leakage current out of specification high voltage

Measured leakage current exceeds acceptable limits during performance inspection procedure

Ranked causes

  1. 1. Chassis leakage current exceeds 100 µa (normal mode) or 300/500 µa (lifted ground mode) (unknown)
First check

Measure chassis leakage current in normal mode

Test point
Analyzer chassis clip to Well A negative battery pin
Expected
less than 100 µa, <100 µa
Tools
Safety analyzer

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Depleted battery insufficient for discharge test

Battery cannot supply required discharge currents to complete Low Current Discharge Test

Ranked causes

  1. 1. Battery depleted below minimum charge level (occasional)
First check

Verify battery charge state before discharge test

Test point
Test battery
Expected
Battery charged to operational level; able to supply required test currents

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Event code logged patient safety

Battery Support System 2 logs event codes when internal program or process fails to execute properly. Event codes have three digits derived from front panel LEDs, with first digit from battery Well A LEDs, second from Well B LEDs, third from Well C LEDs.

Ranked causes

  1. 1. Internal program or process failure (unknown)
First check

Derive event code from front panel LEDs

Test point
Well A, Well B, Well C LED indicators
Expected
Three-digit code where CHARGE=3, CONDITION=2, SHELF LIFE=1 for each well

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Event Log display and clear

Access and manage first five entries of the Event Log during performance inspection mode

Ranked causes

  1. 1. Review device event history (common)
First check

Event Log display access

Test point
Performance inspection mode menu
Expected
First five entries accessible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Event Log is empty

No numerical codes are displayed in the Event Log and all three battery well LEDs remain off

Ranked causes

  1. 1. No events recorded (unknown)

Passcode entry timeout

Device DISCARD LEDs illuminate when passcode not entered within time limit

Ranked causes

  1. 1. Passcode entry exceeds 15 second window (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Reversed polarity connection detected caution

DISCARD LED illuminates during test setup, indicating polarity error in test connections

Ranked causes

  1. 1. Reversed polarity in test setup connections (common)
First check

Verify polarity of all test setup connections

Test point
Test battery terminals and lead connections
Expected
Positive to positive, negative to negative connections

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SERVICE legend illuminates steadily

Front panel SERVICE legend illuminates steadily during power-up self test, indicating self-test failure detected

Ranked causes

  1. 1. Power-up self-test failure detected (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

SERVICE legend steady after self-test

Front panel SERVICE legend illuminates steadily, indicating power-up self-test failure

Ranked causes

  1. 1. Power-up self-test failure (unknown)

Smart battery pogo pins bent, broken, or corroded patient safety

Smart battery pogo pins show signs of damage or loss of spring function

Ranked causes

  1. 1. Physical damage or wear to pogo pins (occasional)
First check

Verify pogo pin free range of motion and spring action

Test point
smart battery pogo pins
Expected
Pin moves freely with visible spring action

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

View event codes from Battery Support System 2

Access stored event codes in the Event Log via power-up self-test mode to diagnose transient EMI or ESD events

Ranked causes

  1. 1. Transient EMI or ESD detected (occasional)
First check

Unplug Battery Support System 2 from ac line power

Test point
Power inlet
Expected
Device unplugged

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage measurement test at high level

Voltage measurement test procedure for 15.6 V battery well measurement

Ranked causes

  1. 1. Standard calibration/measurement procedure (unknown)
First check

Set power supply output to high voltage level

Test point
DC power supply output
Expected
15.6 ±0.1 V, ±0.1 V
Tools
DMM

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Voltage measurement test at low level

Voltage measurement test procedure for 4.0 V battery well measurement

Ranked causes

  1. 1. Standard calibration/measurement procedure (unknown)
First check

Set power supply output to low voltage level

Test point
DC power supply output
Expected
4.0 ±0.1 V, ±0.1 V
Tools
DMM

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
101Shelf Life event - Position ASelf test or inspection mode triggers shelf life status checkwarning
102Condition event - Position CSelf test or inspection mode triggers condition checkwarning
102Shelf Life event with condition changeSelf-test completion; keypad entry A-SHELF LIFE, B-None, C-CONDITIONunknown
103Charge event - Position CSelf test or inspection mode triggers charge status checkwarning
103Shelf Life event with charge changeSelf-test completion; keypad entry A-SHELF LIFE, B-None, C-CHARGEunknown
111Shelf Life event - Multiple positionsSelf test or inspection mode triggers comprehensive shelf life checkwarning
111Process stack overflowSoftware process stack grows beyond 20 levelscritical
112Process stack underflowSoftware process stack underflowserror
112Shelf Life and condition change eventSelf-test completion; keypad entry A-SHELF LIFE, B-SHELF LIFE, C-CONDITIONunknown
112Condition event - Multiple positionsSelf test or inspection mode triggers comprehensive condition checkwarning
113Event Log clear procedureoperator scrolls to code 113 and presses ENTER in Field Test Modeinfo
113Process not found in interruptSoftware process not found during interrupt handlingerror
121Serial buffer overflowSerial communication buffer overflowserror
122Open fuse or bad driverFuse open or driver circuit failure detectederror
132HC11 CONFIG errorMicrocontroller configuration errorerror
133System temperature faultSystem temperature exceeds safe limitserror
211RAM failureRandom access memory failure detectedcritical
212ROM failureRead-only memory failure detectedcritical
213EEPROM failureElectrically erasable programmable read-only memory failurecritical
221Erroneous Op-CodeInvalid operation code detected during executioncritical
222DAC failureDigital-to-analog converter failure detectederror
223Watchdog timeoutWatchdog timer expires without resetcritical
231Erroneous IRQInvalid interrupt request detectederror
232Discharge current not metBattery discharge current below specified levelerror
233Charge current not metBattery charge current below specified levelerror
DISCARDLED indication that appears during thermistor testThermistor test initiated with battery installedwarning
DISCARDThermistor test failure or battery failure indicationA-READY LED does not illuminate within 15 seconds or DISCARD LED illuminateserror
DISCARD LEDDevice has detected an error condition during voltage measurement testReversed polarity connection between power supply and device or device failureerror
DISCARD LEDReversed polarity connection detected in test setupReversed polarity connection anywhere in the test setupwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI