Medtronic Aquamantys: troubleshooting and error codes
The ServiceForge knowledge base holds 4 documented symptoms for the Medtronic Aquamantys, with 4 ranked causes, 10 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Output power measurement
Measure output power at different resistance levels and adjust generator to 100W
Ranked causes
- 1. Output power requires calibration (unknown)
Measure output power at initial resistance
- Test point
- Generator output
- Expected
- Lower than previous readings at lower resistance
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pump flow rate verification
Procedure to verify flow rate accuracy by rotating pump head and counting revolutions.
Ranked causes
- 1. Flow rate calibration required (unknown)
Close and lock pump head
- Test point
- pump head lever
- Expected
- locked in down position toward rear of pump generator
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Output power does not decrease with increased resistance high voltage
Power measurement at higher resistance is not lower than at lower resistance, indicating non-linear power output behavior
Ranked causes
- 1. Generator power output not properly calibrated or functioning (unknown)
Measure output power at 150Ω resistance
- Test point
- Generator output
- Expected
- Power reading lower than previous measurement at lower resistance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify pump flow rate accuracy
Pump head rotation check to confirm flow rate is within specification
Ranked causes
- 1. Flow rate requires verification during preventive maintenance (unknown)
Close pump head
- Test point
- pump head lever
- Expected
- locked down toward rear of pump generator
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI