Compression therapy systems · AV Impulse System family

Medtronic 6060: troubleshooting and error codes

The ServiceForge knowledge base holds 10 documented symptoms for the Medtronic 6060, with 21 ranked causes, 29 diagnostic checks with test points and expected values, and 12 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No air output from patient hoses patient safety

No air heard issuing from patient outlet hoses when controller is impulsing

Ranked causes

  1. 1. Compressor not delivering air (common)
  2. 2. Impulse control valves not operating (common)
  3. 3. Valve electrical connections loose (occasional)
  4. 4. Impulse control valves clogged or damaged (occasional)
First check

Confirm controller is impulsing

Test point
LCD display and patient outlets
Expected
LCD shows impulsing, air heard from patient outlets

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compressor not operating

Compressor motor does not run or produce air pressure

Ranked causes

  1. 1. Controller in standby mode (common)
  2. 2. Compressor power supply incorrect line voltage (common)
  3. 3. Compressor hardware failure (occasional)
First check

Verify controller is not in standby mode

Test point
Controller display and buttons
Expected
LCD shows active mode, not standby

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD entirely blank

Display shows no characters or segments

Ranked causes

  1. 1. Power supply output voltages incorrect (common)
  2. 2. EPROM not properly seated in socket (occasional)
  3. 3. Processor PCB assembly faulty (occasional)
First check

Measure power supply output voltages

Test point
Power supply output
Expected
Conform to table in Section 3.1

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Impulse control valves not clicking during startup

No clicking sounds heard when controller switched ON and OFF

Ranked causes

  1. 1. Valve electrical connections loose (common)
  2. 2. Impulse control valves defective (occasional)
First check

Inspect valve electrical connections

Test point
Valve driver PCB and valve coil connectors
Expected
Connections secure

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD display not backlit or partially lit

Liquid crystal display lacks backlighting or shows incomplete illumination

Ranked causes

  1. 1. Power supply voltage incorrect (common)
  2. 2. Processor PCB failure (occasional)
First check

Verify power supply output voltages

Expected
+5V and +24V outputs within ±5%, ±5%
Tools
Voltmeter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No air impulse from hose outlets patient safety

Controller shows impulsing but no air is heard from patient outlet hoses

Ranked causes

  1. 1. Compressor not producing adequate air (common)
  2. 2. Impulse valves not operating (electrical or mechanical) (occasional)
First check

Confirm controller shows impulsing on LCD

Test point
LCD display
Expected
Display indicates impulsing mode active

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Valve activation clicks not heard during startup

Controller powered ON-OFF but no audible clicking from impulse control valves during initialization sequence

Ranked causes

  1. 1. Valve electrical connections loose or disconnected (common)
  2. 2. Valve mechanism clogged or stuck (occasional)
First check

Verify valve driver PCB assembly connections are secure

Test point
Valve driver PCB connectors
Expected
All connectors fully seated and locked

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fan assembly requires replacement caution

Fan assembly needs to be replaced during maintenance or due to failure; may be indicated by fan noise, reduced airflow, or diagnostic findings.

Ranked causes

  1. 1. Fan assembly failure or wear (unknown)
First check

Disconnect fan lead from power supply

Test point
PL3 on power supply assembly
Expected
Fan lead disconnected

9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fill or vent valve piston failure

Valve piston malfunction requiring replacement, typically identified during controller functionality testing

Ranked causes

  1. 1. Worn or damaged fill or vent valve piston (unknown)
First check

Visually inspect fill or vent valve for damage or malfunction

Test point
Patient left fill and vent valves on reservoir
Expected
Valve piston intact and functional

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LCD elements incomplete

Display shows partial or incomplete LCD segments

Ranked causes

  1. 1. Dirty zebra elastomer or contact surfaces (common)
First check

Inspect display for incomplete LCD segments

Test point
LCD display
Expected
All segments visible and clear

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
E01RAM errorRAM memory test failurecritical
E02ADC errorADC output not zero at switch ONerror
E03NVR errorNon-volatile RAM test failurecritical
E04Membrane switch panel errorButtons depressed at switch ON or defective panelerror
E05Left fill valve faultLeft fill valve not closing properlyerror
E06Left vent valve faultLeft vent valve not closing properlyerror
E07Right fill valve faultRight fill valve not closing properlyerror
E08Right vent valve faultRight vent valve not closing properlyerror
E09Hardware version mismatchSoftware and hardware versions incompatiblewarning
E10High pressure trip line errorHigh pressure trip line not zero at switch ONerror
E11NVR setup failureSoftware version failed to update Non-Volatile RAMerror
E12Over Temperature CutoutTemperature exceeds 113°F (45°C)critical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI