Medtronic 6060: troubleshooting and error codes
The ServiceForge knowledge base holds 10 documented symptoms for the Medtronic 6060, with 21 ranked causes, 29 diagnostic checks with test points and expected values, and 12 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No air output from patient hoses patient safety
No air heard issuing from patient outlet hoses when controller is impulsing
Ranked causes
- 1. Compressor not delivering air (common)
- 2. Impulse control valves not operating (common)
- 3. Valve electrical connections loose (occasional)
- 4. Impulse control valves clogged or damaged (occasional)
Confirm controller is impulsing
- Test point
- LCD display and patient outlets
- Expected
- LCD shows impulsing, air heard from patient outlets
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor not operating
Compressor motor does not run or produce air pressure
Ranked causes
- 1. Controller in standby mode (common)
- 2. Compressor power supply incorrect line voltage (common)
- 3. Compressor hardware failure (occasional)
Verify controller is not in standby mode
- Test point
- Controller display and buttons
- Expected
- LCD shows active mode, not standby
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD entirely blank
Display shows no characters or segments
Ranked causes
- 1. Power supply output voltages incorrect (common)
- 2. EPROM not properly seated in socket (occasional)
- 3. Processor PCB assembly faulty (occasional)
Measure power supply output voltages
- Test point
- Power supply output
- Expected
- Conform to table in Section 3.1
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Impulse control valves not clicking during startup
No clicking sounds heard when controller switched ON and OFF
Ranked causes
- 1. Valve electrical connections loose (common)
- 2. Impulse control valves defective (occasional)
Inspect valve electrical connections
- Test point
- Valve driver PCB and valve coil connectors
- Expected
- Connections secure
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display not backlit or partially lit
Liquid crystal display lacks backlighting or shows incomplete illumination
Ranked causes
- 1. Power supply voltage incorrect (common)
- 2. Processor PCB failure (occasional)
Verify power supply output voltages
- Expected
- +5V and +24V outputs within ±5%, ±5%
- Tools
- Voltmeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No air impulse from hose outlets patient safety
Controller shows impulsing but no air is heard from patient outlet hoses
Ranked causes
- 1. Compressor not producing adequate air (common)
- 2. Impulse valves not operating (electrical or mechanical) (occasional)
Confirm controller shows impulsing on LCD
- Test point
- LCD display
- Expected
- Display indicates impulsing mode active
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Valve activation clicks not heard during startup
Controller powered ON-OFF but no audible clicking from impulse control valves during initialization sequence
Ranked causes
- 1. Valve electrical connections loose or disconnected (common)
- 2. Valve mechanism clogged or stuck (occasional)
Verify valve driver PCB assembly connections are secure
- Test point
- Valve driver PCB connectors
- Expected
- All connectors fully seated and locked
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fan assembly requires replacement caution
Fan assembly needs to be replaced during maintenance or due to failure; may be indicated by fan noise, reduced airflow, or diagnostic findings.
Ranked causes
- 1. Fan assembly failure or wear (unknown)
Disconnect fan lead from power supply
- Test point
- PL3 on power supply assembly
- Expected
- Fan lead disconnected
9 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fill or vent valve piston failure
Valve piston malfunction requiring replacement, typically identified during controller functionality testing
Ranked causes
- 1. Worn or damaged fill or vent valve piston (unknown)
Visually inspect fill or vent valve for damage or malfunction
- Test point
- Patient left fill and vent valves on reservoir
- Expected
- Valve piston intact and functional
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD elements incomplete
Display shows partial or incomplete LCD segments
Ranked causes
- 1. Dirty zebra elastomer or contact surfaces (common)
Inspect display for incomplete LCD segments
- Test point
- LCD display
- Expected
- All segments visible and clear
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E01 | RAM error | RAM memory test failure | critical |
E02 | ADC error | ADC output not zero at switch ON | error |
E03 | NVR error | Non-volatile RAM test failure | critical |
E04 | Membrane switch panel error | Buttons depressed at switch ON or defective panel | error |
E05 | Left fill valve fault | Left fill valve not closing properly | error |
E06 | Left vent valve fault | Left vent valve not closing properly | error |
E07 | Right fill valve fault | Right fill valve not closing properly | error |
E08 | Right vent valve fault | Right vent valve not closing properly | error |
E09 | Hardware version mismatch | Software and hardware versions incompatible | warning |
E10 | High pressure trip line error | High pressure trip line not zero at switch ON | error |
E11 | NVR setup failure | Software version failed to update Non-Volatile RAM | error |
E12 | Over Temperature Cutout | Temperature exceeds 113°F (45°C) | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.