Medtronic 5388: troubleshooting and error codes
The ServiceForge knowledge base holds 57 documented symptoms for the Medtronic 5388, with 66 ranked causes, 115 diagnostic checks with test points and expected values, and 5 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Pacing lead system connection error patient safety
Incorrect or incomplete connection of pacing leads to patient cable receptacles, potentially resulting in no pacing or sensing in one or both chambers
Ranked causes
- 1. Bipolar lead system polarity reversed or incorrectly positioned (common)
- 2. Unipolar single chamber lead not connected to negative receptacle or indifferent electrode not in positive receptacle (occasional)
- 3. Unipolar dual chamber system missing jumper cable connection to indifferent electrode (common)
- 4. Lead connector pins loose or not fully inserted into cable receptacles (occasional)
Visually inspect lead connector pins for visible signs of wear or damage
- Test point
- Lead connectors
- Expected
- No visible wear, damage, bending, or corrosion
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker block point reached during rate adjustment patient safety
Device prevents further adjustment in the conflicting direction and displays warning message 'Pacemaker block point reached. Adjust RATE or parameters on Menu 2.' when user tries to set rate-dependent parameters to values that violate timing rules.
Ranked causes
- 1. RATE set higher than UPPER RATE (common)
- 2. A-V INTERVAL + 180 ms exceeds UPPER RATE interval (DVI mode) (occasional)
- 3. TARP (SAV Interval + PVARP) exceeds UPPER RATE interval (occasional)
Verify RATE versus UPPER RATE relationship
- Test point
- Menu 2 RATE and UPPER RATE settings
- Expected
- RATE ≤ UPPER RATE
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configure Atrial Tracking patient safety
Enable or disable atrial tracking between DDD and DDI modes via Menu 1 setting.
Ranked causes
- 1. A TRACKING set to OFF for DDI mode (common)
- 2. A TRACKING set to ON for DDD mode (common)
Verify atrial tracking status
- Test point
- Menu 1 A TRACKING indicator
- Expected
- ON or OFF as selected
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of ventricular capture patient safety
ECG shows no heart response after pacing pulse; heart does not contract and pump blood
Ranked causes
- 1. Stimulation threshold not properly set (common)
- 2. OUTPUT set too low (insufficient safety margin) (common)
Verify RATE setting for stimulation threshold finding
- Test point
- RATE dial
- Expected
- At least 10 ppm above patient's intrinsic rate
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test failure - error code 0004
Model 5388 fails self-test with error 0004; typically caused by key press during self-test startup sequence
Ranked causes
- 1. Key pressed during power-on self-test sequence (common)
- 2. Key stuck in pressed position (occasional)
Remove and re-insert battery to initiate fresh self-test
- Test point
- Power source
- Expected
- Device powers on and begins 4-second self-test sequence
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Undersensing detected patient safety
Pacemaker does not detect intrinsic heart activity and paces on or between beats, competitive pacing risk
Ranked causes
- 1. Sensing threshold not properly established (common)
- 2. SENSITIVITY set too high (insufficient margin) (common)
Verify RATE setting relative to patient intrinsic rate
- Test point
- RATE dial
- Expected
- At least 10 ppm below patient's intrinsic rate for sensing threshold procedure
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adjust RAP rate or atrial output during delivery patient safety
Operator can modify pacing parameters while SELECT key is held and RAP is active
Ranked causes
- 1. Need to modify parameters during active RAP delivery (occasional)
Maintain SELECT key pressed
- Test point
- SELECT key state
- Expected
- Key held down continuously
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Asynchronous pacing causes tachyarrhythmia risk patient safety
Device set to asynchronous (non-sensing) mode without inhibition of pacing during intrinsic activity
Ranked causes
- 1. Asynchronous pacing mode active with sensing disabled in one or both chambers (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Atrial rate not within tolerance patient safety
Measured atrial pulse interval outside specified range when set to target pacing rate
Ranked causes
- 1. Rate calibration drift or incorrect output setting (unknown)
Measure atrial pulse interval at 30 ppm setting
- Test point
- atrial output
- Expected
- 1900-2100 ms, +/- 5%
- Tools
- digital counter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Atrial rate out of specification patient safety
Measured atrial pulse interval does not fall within specified tolerance when set to designated rate in AAI mode
Ranked causes
- 1. Rate control adjustment incorrect or device malfunction (unknown)
Measure atrial pulse interval at 30 ppm setting
- Test point
- Atrial output
- Expected
- 1900-2100 ms, +/- 5%
- Tools
- Digital counter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Atrial rate out of tolerance
Measured atrial pulse interval does not fall within specified tolerance range when AAI mode is set
Ranked causes
- 1. Rate control adjustment incorrect (common)
Verify atrial pulse interval at 30 ppm setting
- Test point
- Atrial output
- Expected
- 1900-2100 ms, +/- 5%
- Tools
- Digital counter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Atrial rate pulse intervals out of tolerance patient safety
Measured atrial pulse interval does not match specified tolerance when set to a defined rate (30, 80, 120, or 200 ppm) in AAI mode.
Ranked causes
- 1. Rate control parameter not set correctly (common)
Measure atrial pulse interval using digital counter
- Test point
- Atrial output
- Expected
- 30 ppm: 1900–2100 ms; 80 ppm: 712–788 ms; 120 ppm: 475–525 ms; 200 ppm: 285–315 ms, +/- 5%
- Tools
- Digital counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Atrial refractory period not functioning patient safety
Atrial sense light behavior incorrect during refractory period testing; squarewaves within refractory window should not trigger sensing.
Ranked causes
- 1. ATR REFRACTORY setting incorrect or device malfunction (unknown)
Verify signal generator square-wave amplitude across atrial load resistor
- Test point
- 500 ohm atrial load resistor
- Expected
- 20 mV, as measured
- Tools
- Signal generator, voltage measurement device
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable disconnection from pacemaker caution
Need to disconnect patient cable from Model 5388 pacemaker device
Ranked causes
- 1. Planned disconnection or cable replacement needed (unknown)
Locate connector release button on patient cable plug
- Test point
- Patient cable plug connector
- Expected
- Release button clearly visible and accessible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cables not connecting to device patient safety
Surgical cables (Model 5832 or 5433 series) cannot be properly inserted into the atrial or ventricular sockets at the top of the device.
Ranked causes
- 1. Connector not aligned with correct socket (A for atrium, V for ventricle) (common)
Verify cable model matches device compatibility
- Test point
- Cable connector and device socket
- Expected
- Model 5832 or Model 5433A/5433V cable
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Atrial current output patient safety
Verify atrial output amplitude at mid-width point of pulse meets specified range for three output levels: 0.2 mA, 5.0 mA, 20.0 mA
Ranked causes
- 1. Device output requires calibration verification (unknown)
Measure atrial output amplitude at 0.2 mA setting
- Test point
- atrial load terminals
- Expected
- 0.1-0.3 mA, +/- 0.1 mA
- Tools
- oscilloscope, 500 ohm resistor
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: DC leakage (device off) patient safety
Verify DC leakage across atrial and ventricular load resistors when device is powered off
Ranked causes
- 1. DC leakage must be verified when device is off (unknown)
Measure DC voltage across atrial load resistor (device off)
- Test point
- atrial load terminals
- Expected
- 10 uA Max (5 mV across 500 ohms)
- Tools
- DC voltmeter or oscilloscope, 500 ohm resistor
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Mode switching parameter retention patient safety
Verify that OUTPUT, SENSITIVITY, INTERVAL, and RATE parameters are retained or reset appropriately when switching pacing modes.
Ranked causes
- 1. Parameter retention rules during mode change (common)
Confirm OUTPUT settings retained or reset
- Test point
- Upper OUTPUT dial
- Expected
- A OUTPUT retained if mode has atrial pacing; V OUTPUT nominal if newly added
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pulse width patient safety
Procedure to verify atrial and ventricular pulse width output is correct using time interval measurement.
Ranked causes
- 1. Pulse width calibration required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deliver rapid atrial pacing (RAP) patient safety
LED flashes during RAP pulse delivery; pacing begins at displayed RAP rate with no ventricular support during A OUTPUT
Ranked causes
- 1. Operator procedure to initiate RAP burst (unknown)
Verify atrial lead connectivity
- Test point
- atrial channel connection
- Expected
- leads in contact with atrium
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device fails power-on self-test patient safety
Device remains on but does not pace after ON key pressed; self-test lasts ~4 seconds
Ranked causes
- 1. Internal hardware or firmware failure detected during self-test (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device replacement after battery removal patient safety
Pacing maintained for 15 seconds after battery removal at 70 ppm and nominal outputs
Ranked causes
- 1. Battery removal during operation (rare)
Confirm pacing continues after battery removal
- Test point
- device output
- Expected
- pacing continues for 15 seconds typical at 70 ppm nominal outputs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device turned on during lead connection setup high voltage
Model 5388 is powered on and operating while lead system is being connected to device cables.
Ranked causes
- 1. Device powered on before lead system connection completed (common)
Verify device power state
- Test point
- Device power indicator or display
- Expected
- Device powered OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dial adjustment attempt while parameters locked triggers display indication
Turning upper screen dials while locked causes backlight to illuminate, padlock and flashing padlock icons appear in center of upper screen, lower screen displays message.
Ranked causes
- 1. User attempted to adjust locked parameters (common)
Verify padlock and flashing padlock icons appear in center of upper screen
- Test point
- Upper screen display center
- Expected
- Padlock [ ] and flashing [ ] icons visible in center
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Emergency temporary pacing connection setup patient safety
Temporary pacing leads or heartwires must be connected directly to device sockets without standard surgical cables.
Ranked causes
- 1. Standard surgical cables unavailable; emergency procedure required (rare)
Locate emergency connection sockets
- Test point
- Top end of device, four smaller sockets
- Expected
- Two sockets for atrium, two for ventricle visible and accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive current drain when device off
Device draws more than specified off-state current, indicating potential battery drain issue
Ranked causes
- 1. Internal battery drain or leakage (unknown)
Measure device current draw with power off
- Test point
- Battery contacts via special fixture
- Expected
- no more than 150 uA
- Tools
- Analog milliammeter, battery simulator fixture
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive current drain when device on
Device draws more than specified on-state current after 5 minutes with backlight off
Ranked causes
- 1. Internal power consumption exceeding specification (unknown)
Measure device current draw 5 minutes after power on with backlight off
- Test point
- Battery contacts via special fixture
- Expected
- no more than 3.0 mA
- Tools
- Analog milliammeter, battery simulator fixture
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insufficient current draw with backlight and menu on
Device current draw does not reach minimum specified level when backlight and menu are enabled
Ranked causes
- 1. Backlight or display circuit malfunction (unknown)
Measure device current draw with backlight and menu on
- Test point
- Battery contacts via special fixture
- Expected
- at least 20.0 mA
- Tools
- Analog milliammeter, battery simulator fixture
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery indicator appears patient safety
Low battery indicator appears during device operation
Ranked causes
- 1. Battery depleted (common)
Visual confirmation of low battery indicator
- Test point
- device display
- Expected
- low battery indicator appears
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery indicator displays during initialization
Low battery warning appears on display during screen initialization; does not indicate actual low battery unless backlight fails to activate
Ranked causes
- 1. Low battery indicator false positive during startup (common)
Verify backlight activation during normal operation
- Test point
- Display backlight
- Expected
- Backlight activates normally; low battery indicator disappears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery voltage at power-on patient safety
Battery indicator appears and backlights do not turn on when ON key pressed
Ranked causes
- 1. Battery voltage insufficient for normal operation (common)
Verify battery indicator presence on display
- Test point
- Lower display screen
- Expected
- Battery indicator symbol visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse width measurement required
Verification of pacing pulse width at atrial and ventricular outputs
Ranked causes
- 1. Pulse width verification procedure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse width out of specification patient safety
Measured atrial or ventricular output pulse width does not meet specification
Ranked causes
- 1. Pulse width circuit malfunction or component drift (unknown)
Measure atrial output pulse width
- Test point
- Atrial output
- Expected
- As specified in device documentation
- Tools
- Digital counter or oscilloscope at 0.2 ms/division
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse width out of tolerance patient safety
Measured pulse width does not match specification when tested at the half-peak point.
Ranked causes
- 1. Pulse width control not set correctly (unknown)
Measure pulse width from positive to negative half-peak point
- Test point
- Output pulse (atrial and/or ventricular)
- Expected
- As specified in pacemaker documentation (not provided in this section)
- Tools
- Digital counter or oscilloscope (0.2 ms/division)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
RAP rate out of tolerance patient safety
Rapid Atrial Pacing rate measured outside 350 ppm tolerance range
Ranked causes
- 1. RAP rate control setting incorrect (common)
Verify RAP rate is 350 ppm +/- 5%
- Test point
- Atrial Pace LED output
- Expected
- 163-180 ms pulse interval, +/- 5%
- Tools
- Digital counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rapid Atrial Pacing (RAP) rate not within tolerance patient safety
Measured atrial pacing rate during RAP mode outside specified range; A Pace LED may not flash at correct rate or V Pace LED continues flashing
Ranked causes
- 1. RAP mode not properly activated or rate drift (unknown)
Verify RAP mode selection and LED indication
- Test point
- A Pace LED, V Pace LED
- Expected
- A Pace LED flashing at RAP rate; V Pace LED stopped
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rapid Atrial Pacing (RAP) rate out of specification patient safety
RAP mode atrial rate does not fall within 350 ppm +/- 5% (163-180 ms pulse interval) when RAP is activated
Ranked causes
- 1. RAP rate circuit malfunction or calibration error (unknown)
Measure atrial pulse interval in RAP mode at 350 ppm setting
- Test point
- Atrial output
- Expected
- 163-180 ms, +/- 5%
- Tools
- Digital counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rapid Atrial Pacing (RAP) rate out of tolerance patient safety
Measured atrial pacing rate does not match 350 ppm ±5% (163–180 ms) when RAP mode is activated.
Ranked causes
- 1. RAP rate parameter not set correctly (common)
Verify atrial rate during RAP activation
- Test point
- Atrial output (A Pace LED)
- Expected
- 350 ppm +/- 5% (163–180 ms); A Pace LED flashes at RAP rate; V Pace LED stops flashing, +/- 5%
- Tools
- Digital counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rapid atrial pacing rate out of specification patient safety
RAP mode delivers atrial pacing at rate outside 350 ppm tolerance when activated
Ranked causes
- 1. RAP mode rate control malfunction (unknown)
Verify RAP mode pulse interval within tolerance
- Test point
- Atrial output
- Expected
- 163-180 ms, +/- 5%
- Tools
- Digital counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rapid atrial pacing rate out of tolerance patient safety
RAP rate measured on counter does not fall within 350 ppm +/- 5% when RAP mode is selected and Select Key is held
Ranked causes
- 1. RAP mode rate control miscalibration (unknown)
Measure atrial rate during RAP delivery
- Test point
- Atrial pace output
- Expected
- 350 ppm +/- 5% (163-180 ms), +/- 5%
- Tools
- Digital counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rate-dependent parameter mismatch on mode change patient safety
During a mode change, if one or more manually set rate-dependent parameters in the previous mode will cause improper pacing in the new mode, the device automatically changes the parameter to its automatic, rate-dependent setting.
Ranked causes
- 1. Manually set rate-dependent parameter incompatible with new mode (common)
Confirm parameter auto-adjustment after mode change
- Test point
- Menu 2 rate-dependent parameter values
- Expected
- Parameters set to automatic or mode-compatible values
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Refractory period test setup patient safety
Verification that pacemaker correctly inhibits sensing during refractory period and resumes sensing outside refractory period using delayed pulse injection
Ranked causes
- 1. Refractory period timing requires validation with dual-pulse sensing test (common)
Atrial first sensing pulse response (in refractory period)
- Test point
- Pacemaker sensing input
- Expected
- No detection, output pulse present ms, ±1 ms
- Tools
- Oscilloscope, signal generator
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resume pacing at upper screen settings after RAP patient safety
Return device to normal pacing operation after RAP burst completion
Ranked causes
- 1. RAP burst delivery complete; need to return to standard pacing (common)
Release SELECT key to terminate RAP
- Test point
- SELECT key
- Expected
- Key released; pacing resumes at upper screen rate/output settings
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SELECT key stuck - menu advance failure
SELECT key stuck; MENU key press will not advance menus; device does not respond to menu navigation
Ranked causes
- 1. SELECT key stuck mechanically during menu operation (rare)
Press MENU key and observe menu advance
- Test point
- MENU key and menu display
- Expected
- MENU press advances to next menu screen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Select pacing mode patient safety
Configure Medtronic 5388 temporary pacemaker to single-chamber (AOO, VOO, AAI, VVI) or dual-chamber (DDD, DDI, DVI, DOO) mode using OUTPUT and SENSITIVITY dials and MENU selections.
Ranked causes
- 1. Select via Dial-A-Mode menu (common)
Verify pacing mode activated
- Test point
- Menu M display
- Expected
- Selected pacing mode underlined and activated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-test completion and normal operation
Normal power-on self-test sequence: 4-second initialization, PACE and SENSE LEDs flash, sensing begins after completion
Ranked causes
- 1. Normal startup sequence verification (common)
Verify power-on initiates self-test sequence
- Test point
- Device power and LED indicators
- Expected
- Device powered on; self-test active for approximately 4 seconds; PACE and SENSE LEDs light sequentially
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self-test failure - PACE or SENSE LED remains on patient safety
Device fails self-test; one PACE or SENSE LED remains on continuously with no output pulses; failure code displayed
Ranked causes
- 1. Device hardware or firmware self-test failure (rare)
Observe which LED remains illuminated
- Test point
- PACE or SENSE LED
- Expected
- LED turns off upon successful self-test completion
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Set sensitivity safety margin patient safety
Configure sensitivity to provide at least 2:1 safety margin; set to at least one-half to one-third the sensing threshold value
Ranked causes
- 1. Lead maturation and drug therapy affect threshold over time (common)
Measure atrial or ventricular sensing threshold per protocol
- Test point
- A SENSITIVITY or V SENSITIVITY parameter
- Expected
- Threshold value in mV determined by decreasing sensitivity until sensing lost
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to disconnect cable from device caution
Cable cannot be removed from the device socket after insertion.
Ranked causes
- 1. Connector release button not pressed before removal attempt (common)
Locate connector release button on cable
- Test point
- Cable connector body
- Expected
- Release button present and accessible
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Upper dials locked during adjustment attempt
Upper dials do not turn and padlock icon appears on screen; backlights turn off
Ranked causes
- 1. Automatic lock activated after 60 seconds of inactivity or manual lock applied (common)
Verify padlock icon on middle of screen
- Test point
- Display screen center
- Expected
- Padlock icon and flashing key icon visible when locked
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Upper screen parameters locked unintentionally
Turning upper screen dials does not change RATE, A OUTPUT, or V OUTPUT values; padlock icon appears in upper right corner; backlight turns off.
Ranked causes
- 1. Lock feature activated after 60 seconds of inactivity or lock key pressed (common)
Verify padlock icon is displayed in upper right corner of upper screen
- Test point
- Upper screen display
- Expected
- Padlock [ ] icon visible in upper right corner
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventricular rate not within tolerance patient safety
Measured ventricular pulse interval outside specified range when set to target pacing rate
Ranked causes
- 1. Rate calibration drift or incorrect output setting (unknown)
Measure ventricular pulse interval at 30 ppm setting
- Test point
- ventricular output
- Expected
- 1900-2100 ms, +/- 5%
- Tools
- digital counter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventricular rate out of specification patient safety
Measured ventricular pulse interval does not fall within specified tolerance when set to designated rate in VVI mode
Ranked causes
- 1. Rate control adjustment incorrect or device malfunction (unknown)
Measure ventricular pulse interval at 30 ppm setting
- Test point
- Ventricular output
- Expected
- 1900-2100 ms, +/- 5%
- Tools
- Digital counter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventricular rate out of tolerance
Measured ventricular pulse interval does not fall within specified tolerance range when VVI mode is set
Ranked causes
- 1. Rate control adjustment incorrect (common)
Verify ventricular pulse interval at 30 ppm setting
- Test point
- Ventricular output
- Expected
- 1900-2100 ms, +/- 5%
- Tools
- Digital counter
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventricular rate pulse intervals out of tolerance patient safety
Measured ventricular pulse interval does not match specified tolerance when set to a defined rate (30, 80, 120, or 200 ppm) in VVI mode.
Ranked causes
- 1. Rate control parameter not set correctly (common)
Measure ventricular pulse interval using digital counter
- Test point
- Ventricular output
- Expected
- 30 ppm: 1900–2100 ms; 80 ppm: 712–788 ms; 120 ppm: 475–525 ms; 200 ppm: 285–315 ms, +/- 5%
- Tools
- Digital counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventricular refractory period not functioning patient safety
Ventricular sense light behavior incorrect during refractory period testing; squarewaves within refractory window should not trigger sensing.
Ranked causes
- 1. VEN REFRACTORY setting incorrect or device malfunction (unknown)
Verify signal generator square-wave amplitude across atrial load resistor
- Test point
- 500 ohm atrial load resistor
- Expected
- 20 mV, as measured
- Tools
- Signal generator, voltage measurement device
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventricular refractory period test patient safety
Verify that ventricular sensing respects the refractory period; input during refractory should not trigger ventricular sense light.
Ranked causes
- 1. Ventricular refractory period malfunction (unknown)
Apply first squarewave input 220 ms after pacemaker output using switch position 1
- Test point
- ventricular input
- Expected
- Ventricular sense light should not come on mV, as measured across 500 ohm load
- Tools
- Signal generator, 500 ohm ventricular load resistor, switch position 1
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0004 | Key stuck or key press during self-test | Key pressed while self-test is in process | error |
0004 | Power-on self-test failure caused by key press | User presses key while self-test in progress (first 4 seconds after power-on) | error |
Pacemaker block point reached | Device has reached a timing constraint limit and cannot proceed with the requested adjustment | User attempts to set rate-dependent parameters to values that violate timing rules or set RATE higher than UPPER RATE | warning |
RATE cannot exceed UPPER RATE | Attempted RATE adjustment exceeds the configured UPPER RATE limit | User attempts to set RATE higher than UPPER RATE | warning |
Timing Violation warning | Rate-dependent parameter adjustment would exceed safe timing constraints | RATE adjustment would conflict with UPPER RATE PVARP or A-V INTERVAL limits | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI