Medtronic 5348: troubleshooting and error codes
The ServiceForge knowledge base holds 181 documented symptoms for the Medtronic 5348, with 242 ranked causes, 179 diagnostic checks with test points and expected values, and 1 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit produces no pacing output patient safety
Device fails to deliver pacing stimulus to patient
Ranked causes
- 1. Random component failure (rare)
- 2. Battery depletion (occasional)
- 3. Device mishandling (occasional)
- 4. Improper lead or cable connection (common)
- 5. Lead displacement (occasional)
- 6. Lead or cable fracture (occasional)
Attempt to correct loss of control by turning device off then on
- Test point
- Power switch
- Expected
- Device powers off and back on
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
No output patient safety
Pacemaker fails to deliver pacing stimulus to patient
Ranked causes
- 1. Random component failure (occasional)
- 2. Battery depletion (common)
- 3. Mishandling (occasional)
- 4. Improper connection or displacement of leads or cables (occasional)
Turn device off then on
- Test point
- power control
- Expected
- device powers on
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
No output from pacemaker patient safety
Pacemaker fails to deliver pacing stimulus to the heart
Ranked causes
- 1. Random component failure (occasional)
- 2. Battery depletion (common)
- 3. Improper lead or cable connection (occasional)
- 4. Lead or cable fracture (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rapid Atrial Pacing (RAP) not functioning correctly patient safety
RAP controls unresponsive, rate not displayed, or device does not deliver RAP burst when HOLD TO DELIVER key is pressed
Ranked causes
- 1. Lead not in contact with atrium (common)
- 2. RAP Standby not enabled (common)
- 3. RAP Standby timeout occurred (occasional)
- 4. Device operating in base level pacing mode (occasional)
Verify lead placement in atrium
- Test point
- Atrial lead position
- Expected
- Lead in contact with atrium only, not ventricle
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit produces no output patient safety
Pacemaker fails to deliver pacing stimulus to patient
Ranked causes
- 1. Random component failure (occasional)
- 2. Battery depletion (common)
- 3. Device malfunction from mishandling (occasional)
- 4. Improper connection or displacement of leads or cables (occasional)
Power cycle device
- Test point
- Device power switch
- Expected
- Device turns off then on
Device senses no cardiac activity patient safety
Pacemaker fails to detect patient's intrinsic cardiac rhythm
Ranked causes
- 1. Random component failure (rare)
- 2. Improper lead connection or displacement (common)
- 3. Lead fracture (occasional)
Erratic or no pacing output patient safety
Device fails to deliver consistent pacing pulses or no output occurs
Ranked causes
- 1. Battery with incorrect physical dimensions (common)
- 2. Battery contacts contaminated (common)
- 3. Battery drawer not fully latched (occasional)
Verify battery model and physical dimensions
- Test point
- Battery compartment
- Expected
- Battery dimensions match device specifications
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead fracture or disconnection patient safety
Pacing lead is broken or disconnected from device or patient
Ranked causes
- 1. Lead fracture (occasional)
- 2. Lead displacement (occasional)
- 3. Improper connection of lead (occasional)
No output patient safety
Pacemaker fails to deliver pacing stimulus to the patient.
Ranked causes
- 1. Random component failure (occasional)
- 2. Battery depletion (common)
- 3. Lead or cable disconnection or fracture (occasional)
Verify battery presence and condition
- Test point
- Battery compartment
- Expected
- Battery installed
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker interference from electromagnetic sources patient safety
Operation of medical telemetry, communication transmitters, or MRI equipment near pacemaker may cause interference
Ranked causes
- 1. Medical telemetry equipment within one meter of pacemaker (occasional)
- 2. Cellular phones or walkie talkies operated near pacemaker (occasional)
- 3. MRI equipment operation with active pacemaker (rare)
Rapid Atrial Pacing (RAP) burst delivery patient safety
Operator presses HOLD TO DELIVER key to initiate rapid atrial pacing output; device waits up to two base pacing cycles then delivers asynchronous pulses at selected rate while key is held
Ranked causes
- 1. RAP Standby not enabled (common)
- 2. Lead not in contact with atrium (common)
- 3. RAP Standby timeout (occasional)
Verify RAP Standby is active
- Test point
- Display screen
- Expected
- Nominal RAP rate of 320 min-1 (ppm) displayed
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery depleted during operation patient safety
Pacemaker ceases functioning or exhibits reduced output due to low battery capacity; low battery indicator appears or device stops after insufficient duration of use
Ranked causes
- 1. Battery not replaced at start of patient use or after low indicator appears (occasional)
- 2. Alkaline battery used beyond seven-day continuous use limit (common)
Check battery status
- Test point
- Pacemaker battery indicator or status display
- Expected
- Battery indicator shows adequate charge; no low battery warning
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement needed patient safety
Low battery indicator appears during device operation; battery status check indicates depletion
Ranked causes
- 1. Battery depleted during continuous use (common)
- 2. Wrong battery type installed (occasional)
Verify battery status indicator
- Test point
- Device display or indicator light
- Expected
- Low battery indicator present or battery status depleted
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Connection or cable failure caution
Loose, disconnected, or damaged cables and connectors prevent proper device operation
Ranked causes
- 1. Loose or disconnected connector (common)
- 2. Cable damage or degradation (occasional)
Visually inspect all connectors and cables
- Test point
- All device connections
- Expected
- All connectors fully seated; no loose or corroded pins; cable sheathing intact
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Crystal rate deviation detected patient safety
PACE and SENSE indicators illuminate for less than one second; system self-test performed; device may turn off if rate remains incorrect
Ranked causes
- 1. Crystal rate deviation from appropriate value (rare)
- 2. Crystal rate remains incorrect after restart (rare)
Observe PACE and SENSE indicators during rate-runaway protection self-test
- Test point
- PACE and SENSE LED indicators
- Expected
- Indicators illuminate for less than one second
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not pace or paces erratically patient safety
No pacing output or irregular pacing after battery installation
Ranked causes
- 1. Battery has different physical dimensions (occasional)
- 2. Battery contacts contaminated (occasional)
Verify battery type and physical dimensions match device specification
- Test point
- Battery drawer
- Expected
- Recommended battery installed with correct dimensions
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device fails self-test and remains powered on patient safety
LEDs remain illuminated after self-test failure; device does not turn off automatically
Ranked causes
- 1. OFF key self-test failure (unknown)
- 2. Self-test failure for unspecified reason (unknown)
Verify OFF key operation and self-test completion
- Test point
- OFF key
- Expected
- Device powers off immediately upon failed self-test
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device inhibits or reverts to asynchronous pacing during electromagnetic interference patient safety
Pacemaker in demand mode may stop pacing (inhibit completely) or switch to fixed-rate asynchronous operation at the RATE dial setting when exposed to strong external electromagnetic fields
Ranked causes
- 1. Electrosurgical equipment in vicinity (common)
- 2. Diathermy equipment in vicinity (common)
Verify device mode setting
- Test point
- RATE dial / mode selector
- Expected
- Asynchronous mode when strong EMI present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device inhibits or reverts to asynchronous pacing during EMI exposure patient safety
Pacemaker responds to external electromagnetic interference by inhibiting completely or reverting to fixed-rate asynchronous operation at the RATE dial setting
Ranked causes
- 1. Electromagnetic interference (EMI) from electrosurgical equipment (common)
- 2. Diathermy equipment in vicinity of patient (common)
Verify pacemaker mode setting
- Test point
- MODE selector or display
- Expected
- Set to asynchronous mode if EMI present
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device malfunction after opening caution
Fluid sealing compromised or ESD barrier degraded after improper opening/resealing
Ranked causes
- 1. Seam opened and improperly resealed (rare)
- 2. ESD barrier label broken (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EMI from communication transmitters or medical equipment patient safety
Interference from cellular phones, walkie-talkies, emergency vehicle transmitters, or MRI equipment when operated within one meter of pacemaker
Ranked causes
- 1. Communication or telemetry transmitter operated within one meter of pacemaker (occasional)
- 2. MRI equipment operation near pacemaker (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exposed pins contacting conductive or wet surfaces patient safety
Risk of electrical shorting or contamination when exposed connector pins contact blood, body fluids, or conductive materials
Ranked causes
- 1. Unprotected exposed connector pins (common)
- 2. Blood or body fluid contamination during connection (occasional)
Inspect exposed connector pins
- Test point
- Connector pins and wires
- Expected
- Dry, clean, covered or protected from contact
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid contamination of device caution
Liquid ingress into device causing malfunction or corrosion of internal components
Ranked causes
- 1. Device exposed to liquid without protection (occasional)
- 2. Blood or body fluid contamination of receptacles (occasional)
Inspect for visible fluid or contamination
- Test point
- All external surfaces, receptacles and seams
- Expected
- Dry, clean, no visible fluid or residue
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect lead polarity configuration patient safety
Pacing threshold values affected by improper polarity connection; reduced pacing efficacy
Ranked causes
- 1. Bipolar lead pins inserted in wrong receptacles (occasional)
- 2. Unipolar system connected incorrectly (occasional)
Identify lead system type
- Test point
- Lead connector
- Expected
- System clearly identified as bipolar or unipolar
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Increased or decreased pacing rate patient safety
Pacemaker rate is higher or lower than programmed setting
Ranked causes
- 1. Random component failure (occasional)
- 2. Loss of control of rate (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Increased or decreased rate patient safety
Pacing rate does not match programmed value.
Ranked causes
- 1. Random component failure (occasional)
- 2. Loss of control due to component malfunction (occasional)
Measure pacing rate using ECG or pulse
- Test point
- Patient cardiac rhythm
- Expected
- Rate matches programmed value
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead system configuration mismatch patient safety
Polarity markings on pacing leads, cables or pacemaker terminals are not matched; operator may observe no pacing or reverse polarity effects
Ranked causes
- 1. Pacing electrode not connected to negative terminal or indifferent lead to positive terminal (occasional)
- 2. Polarity markings on components do not align (occasional)
Verify pacing electrode terminal connection
- Test point
- Pacemaker negative (-) terminal
- Expected
- Pacing electrode connected to this terminal
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of control of device parameters patient safety
Technician cannot adjust rate, output, sensitivity or power settings
Ranked causes
- 1. Random component failure (occasional)
- 2. Device firmware or logic failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery indicator during operation caution
Battery voltage drops below safe operating threshold, indicated by low battery indicator light or display.
Ranked causes
- 1. Battery depleted after extended use (common)
- 2. New patient not provided with fresh battery at start (occasional)
Check battery status
- Test point
- Low battery indicator and battery voltage
- Expected
- No low battery indicator; battery status normal
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sensing patient safety
Pacemaker fails to detect intrinsic cardiac activity.
Ranked causes
- 1. Random component failure (occasional)
- 2. Lead displacement or fracture (occasional)
Check lead attachment and position
- Test point
- Pacing lead contact
- Expected
- Secure contact with cardiac tissue
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sensing patient safety
Pacemaker fails to detect intrinsic cardiac activity
Ranked causes
- 1. Random component failure (occasional)
- 2. Improper connection or displacement of leads or cables (occasional)
Operational error after improper sterilization caution
Device damage from gamma irradiation or autoclave sterilization causing malfunction
Ranked causes
- 1. Device sterilized by gamma irradiation (rare)
- 2. Device steam-sterilized (autoclave) (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker inhibits or reverts to asynchronous mode patient safety
Device stops pacing on demand or switches to fixed-rate pacing; occurs in presence of strong electromagnetic interference (EMI)
Ranked causes
- 1. Electromagnetic interference (EMI) from electrosurgical equipment (common)
- 2. Electromagnetic interference (EMI) from other sources (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacemaker visual inspection reveals damage patient safety
Device shows signs of mechanical damage, corrosion, or contamination during visual inspection
Ranked causes
- 1. Battery corrosion or contamination (occasional)
- 2. Mechanical device damage from impact or mishandling (rare)
Visual inspection of device housing
- Test point
- Device exterior surface
- Expected
- No visible cracks, dents, or mechanical damage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacing lead connector not seating in patient cable patient safety
Pacing lead connector pins do not insert fully into patient cable receptacles during connection procedure.
Ranked causes
- 1. Connector knobs not loosened sufficiently (common)
- 2. Pacing lead connector damaged or worn (occasional)
Visually inspect pacing lead connectors
- Test point
- Lead connector pins and patient cable receptacles
- Expected
- No visible wear or damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacing lead displacement or fracture patient safety
Loss of electrical connection between pacemaker and heart due to lead problems.
Ranked causes
- 1. Improper initial connection of lead (occasional)
- 2. Mechanical displacement of lead (occasional)
Verify lead attachment at pacemaker connector
- Test point
- Connector junction
- Expected
- Lead is securely attached
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacing rate abnormal patient safety
Pacemaker paces at increased or decreased rate compared to setting
Ranked causes
- 1. Random component failure (occasional)
- 2. Loss of control of rate setting (rare)
Patient cable connector not seating properly patient safety
Patient cable connector plug does not click fully into Model 5348 receptacle, indicating poor connection to pacemaker.
Ranked causes
- 1. Connector not fully inserted (common)
- 2. Patient cable damaged or worn (occasional)
Visually inspect patient cable connector and receptacle
- Test point
- Model 5433A or 5433V patient cable connector
- Expected
- No visible wear or damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
RAP burst not delivered patient safety
HOLD TO DELIVER key pressed but device does not deliver output pulses at selected RAP rate
Ranked causes
- 1. RAP Standby not active or timed out (common)
- 2. Lead not in atrial contact (common)
Confirm RAP Standby active
- Test point
- RAP display and controls
- Expected
- RAP rate displayed, all RAP keys functional
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rapid Atrial Pacing (RAP) procedure execution patient safety
Enable RAP Standby mode, adjust pacing rate, deliver RAP bursts, and return to basic pacing
Ranked causes
- 1. RAP Standby mode enabled and configured (common)
- 2. Return to basic pacing operation (common)
Verify lead contact location
- Test point
- Atrial lead position
- Expected
- Lead in contact with atrium, not ventricle
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unintended sensing of far field signals patient safety
Pacemaker senses inappropriate cardiac signals on wrong chamber (R/T waves on atrial lead or P waves on ventricular lead)
Ranked causes
- 1. Sensitivity setting too low on wrong lead (occasional)
- 2. Bipolar lead configuration not used (occasional)
Verify sensitivity setting relative to threshold
- Test point
- Sensitivity dial
- Expected
- At least one-half the mV value of sensitivity threshold
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit performs no sensing patient safety
Pacemaker fails to detect intrinsic cardiac activity
Ranked causes
- 1. Random component failure (occasional)
- 2. Improper connection or displacement of leads or cables (occasional)
Visual damage or contamination detected caution
Physical damage to device body, corroded battery connections, or contamination in battery compartment
Ranked causes
- 1. Mechanical damage from handling or accident (occasional)
- 2. Battery corrosion or contamination (occasional)
Inspect device for mechanical damage
- Test point
- Device exterior surfaces
- Expected
- No cracks, dents, or physical deformation visible
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abrupt cessation of pacing stimuli patient safety
Pacing stops suddenly, potentially causing asystole before intrinsic rhythm restarts
Ranked causes
- 1. Immediate termination of pacing without gradual reduction (occasional)
Confirm demand mode available
- Test point
- Device mode selector
- Expected
- Demand mode selected before termination
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adjust RAP Rate patient safety
Modify rapid atrial pacing rate during RAP Standby using increase/decrease keys
Ranked causes
- 1. RAP rate adjustment needed (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
All indicator lights remain on after self-test failure patient safety
Device fails self-test, all indicator lights stay on, no pacing or sensing occurs
Ranked causes
- 1. Self-test failure - cause unspecified (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Asystole after pacing termination patient safety
Abrupt cessation of pacing stimuli may cause periods without cardiac output before intrinsic rhythm resumes
Ranked causes
- 1. Sudden cessation of pacing without gradual reduction (common)
Verify demand mode operation during pacing rate reduction
- Test point
- Pacemaker mode selector
- Expected
- Demand mode selected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Asystole after pacing termination patient safety
Abrupt cessation of pacing stimulus results in periods of asystole before intrinsic rhythm is established
Ranked causes
- 1. Abrupt termination of pacing stimulus (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Asystole after stopping pacing patient safety
Abrupt cessation of pacing stimuli causes periods without heartbeat before patient's own rhythm resumes.
Ranked causes
- 1. Abrupt termination of pacing without gradual rate reduction (common)
Confirm gradual rate reduction protocol is followed
- Test point
- Pacing rate dial
- Expected
- Rate decreases incrementally using demand mode before full termination
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery depleted before 7 days during continuous use patient safety
Alkaline battery requires replacement before the recommended 7-day interval during continuous pacemaker operation.
Ranked causes
- 1. Incorrect or non-recommended battery type installed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery depleted during continuous use patient safety
Alkaline battery reaches end of life before expected replacement interval
Ranked causes
- 1. Alkaline battery life exhausted after seven days continuous use (common)
Monitor battery age during continuous use
- Test point
- Operating time since last battery installation
- Expected
- Less than seven days continuous use
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery drawer not latching properly patient safety
Battery drawer fails to latch securely; device may not indicate a problem despite improper closure
Ranked causes
- 1. Battery drawer release buttons not pressed simultaneously or fully (common)
Verify battery drawer closure
- Test point
- Battery drawer
- Expected
- Drawer in closed position and neither release button pressed into device
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Determine pacing mode patient safety
Setting the device to demand (synchronous) or asynchronous pacing mode based on lead placement and sensitivity adjustment.
Ranked causes
- 1. Mode selection via lead placement and sensitivity setting (unknown)
Measure patient's sensitivity threshold
- Test point
- Intrinsic cardiac signal
- Expected
- Patient sensitivity threshold in mV
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Rapid Atrial Pacing rate adjustment patient safety
Procedure to set desired RAP rate within Standby mode
Ranked causes
- 1. Rate adjustment required (common)
Increase RAP rate
- Test point
- RAP rate display
- Expected
- Rate increases with ∧ key held
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Select asynchronous pacing mode patient safety
Configure pacemaker for asynchronous pacing by adjusting sensitivity dial
Ranked causes
- 1. Mode selection via sensitivity dial to ASYNC position (common)
Turn SENSITIVITY dial fully counterclockwise
- Test point
- SENSITIVITY dial
- Expected
- ASYNC position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Select demand pacing mode patient safety
Configure pacemaker for demand (synchronous) pacing by setting sensitivity threshold
Ranked causes
- 1. Mode selection via sensitivity dial adjustment (common)
Determine patient sensitivity threshold
- Test point
- SENSITIVITY dial
- Expected
- Value between 0.5 and 20 mV
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current leakage to myocardium from improperly grounded line-powered equipment high voltage
Alternating current leakage through implanted lead pathway to heart muscle may cause fibrillation
Ranked causes
- 1. Line-powered equipment not properly grounded near patient (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillation paddle placement risk high voltage
Paddles placed near pacemaker or lead system during defibrillation
Ranked causes
- 1. Paddle proximity to pacemaker or leads (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillation/cardioversion risk to pacemaker patient safety
Patient requires defibrillation or cardioversion while temporary pacemaker is active; risk of myocardial damage from current flow across lead system electrodes
Ranked causes
- 1. Pacemaker and lead system left connected during defibrillation/cardioversion (common)
Verify pacemaker is disconnected from lead system
- Test point
- Patient cable connector
- Expected
- Pacemaker physically disconnected from lead system
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Deliver RAP Burst patient safety
Initiate rapid atrial pacing burst by pressing and holding delivery key; device waits up to two base level pacing cycles then delivers asynchronous pulses at selected rate
Ranked causes
- 1. RAP burst delivery needed (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Device damaged by improper sterilization
- Device does not power on
- Device does not turn on
- Device fails self-test
- Device fails self-test
- Device fails self-test and does not turn off
- Device fails self-test but remains on
- Device fails to operate
- Device has mechanical damage
- Device inhibits or reverts to asynchronous pacing during EMI exposure
- Device malfunction due to battery depletion
- Device not operating due to missing battery
- Device physically damaged
- Device reverts to asynchronous pacing
- Device reverts to asynchronous pacing
- Device self-test on startup
- Device shuts off unexpectedly
- Device turns off unexpectedly
- Electromagnetic interference susceptibility
- Electrostatic discharge damage
- Electrostatic discharge damage
- Electrostatic discharge during lead handling
- Electrostatic discharge exposure to lead system
- Enable RAP Standby State
- Equipment may inhibit or revert to asynchronous pacing during EMI exposure
- Erratic operation due to high humidity exposure
- Erratic operation due to temperature instability
- Erratic or no pacing during operation
- False indicator light signals
- False indicator light signals
- Fluid contamination of device
- Front panel controls not responding
- High humidity degrades device operation
- High-rate atrial pacing causes rapid ventricular conduction
- High-rate atrial pacing causes rapid ventricular conduction
- High-rate atrial pacing may cause rapid ventricular conduction
- High-rate burst pacing risk
- High-rate conduction risk during RAP
- Inappropriate pacing inhibition
- Inappropriate pacing inhibition due to sensitivity setting
- Inappropriate sensing of far field signals
- Inappropriate sensing of far field signals
- Increased electromagnetic interference susceptibility
- Increased or decreased output amplitude
- Increased or decreased output amplitude or pulse width
- Increased or decreased output pulse width
- Increased or decreased rate
- Increased or decreased rate, output pulse width, or output amplitude
- Loose pacing lead connection
- Loss of control of output
- Loss of control of power
- Loss of control of rate
- Loss of control of rate, output, sensitivity or power
- Loss of control of rate, output, sensitivity or power
- Loss of control of sensitivity
- Loss of device power
- Loss of pacing due to abrupt termination
- LOW BATT indicator continuous illumination
- LOW BATT indicator flashing
- LOW BATT indicator flashing on startup
- LOW BATT. indicator continuous illumination
- LOW BATT. indicator flashing
- Low battery condition during operation
- Low battery during operation
- Low battery during pacemaker operation
- Low battery indicator appears
- Low battery indicator appears
- Low battery indicator appears during operation
- Measurement: sensing threshold
- Measurement: stimulation threshold
- Myocardial damage during defibrillation
- OFF key fails self-test
- OFF key self-test failure
- Pacemaker damage from defibrillation discharge
- Pacemaker fails self-test
- Pacemaker functional test fails self-test
- Pacemaker inhibition or asynchronous reversion during EMI exposure
- Pacemaker pulse width test measurement
- Pacemaker rate test measurement
- Pacemaker sensitivity to line-powered equipment leakage current
- Pacing lead connection instability
- Pacing lead connection procedure
- Pacing lead connector pins not seating properly
- Paddle placement during defibrillation near pacemaker
- Patient at risk during defibrillation with pacemaker connected
- Patient cable connection sequence error
- Patient cable connection unclear or loose
- Periods of asystole on pacing termination
- Physical damage to device
- Potential myocardial damage during defibrillation with implanted leads
- Preventive maintenance performed
- RAP controls unresponsive after power cycle
- RAP Standby auto-disable
- RAP Standby feature inactive
- RAP Standby state auto-disable timeout
- Rate and Sensitivity dial adjustment during RAP delivery
- Rate-runaway detected
- Reversion to asynchronous pacing
- Reversion to asynchronous pacing
- Reversion window event detected during low battery
- Risk from high-rate burst pacing
- Risk of ectopic stimulus delivery during sensing adjustment
- Risk of fibrillation from line-powered equipment leakage
- Risk of myocardial damage during defibrillation
- Risk of ventricular fibrillation during electrosurgery
- Safety cable or heartwire receptacle contaminated
- Safety markings unclear or missing
- Self-test failure with all indicator lights on
- Static discharge to patient or device
- Static electricity hazard during lead handling
- Temperature instability affects operation
- Threshold variation in bipolar lead connections
- Unable to disconnect patient cable from pacemaker
- Unauthorized access to device settings
- Unauthorized changes to pacemaker settings
- Unintended high-rate ventricular conduction
- Unipolar lead system polarity error
- Unit does not turn on
- Ventricular fibrillation risk during electrosurgery
- Ventricular fibrillation risk from electrosurgery
- Ventricular fibrillation risk from electrosurgery near implanted lead
121 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
LOW BATT | Battery voltage below operational threshold | Battery voltage drops below 7.2 volts | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.