Medrad Stellant: troubleshooting and error codes
The ServiceForge knowledge base holds 44 documented symptoms for the Medrad Stellant, with 64 ranked causes, 56 diagnostic checks with test points and expected values, and 41 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Docking sensor A error (DOCK SENSA prefix) patient safety
System has detected Piston A docking sensor closed with no syringe attached on side A
Ranked causes
- 1. Syringe not installed correctly (common)
- 2. Dirty syringe interface area (common)
- 3. Faulty size Sensor A (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Docking sensor B error (DOCK SENSB prefix) patient safety
System has detected Piston B docking sensor closed with no syringe attached on side B
Ranked causes
- 1. Syringe not installed correctly (common)
- 2. Dirty syringe interface area (common)
- 3. Faulty size Sensor B (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Encoder A error (ENCA prefix)
System has found problem with Encoder on motor A or associated circuitry
Ranked causes
- 1. P4 connector not connected (common)
- 2. Faulty Power Drive card (occasional)
- 3. Faulty motor A (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Encoder B error (ENCB prefix)
System has found problem with Encoder on motor B or associated circuitry (Dual only)
Ranked causes
- 1. P4 connector not connected correctly (common)
- 2. Faulty Power Drive card (occasional)
- 3. Faulty motor B (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injector not ready for scanning patient safety
Injector icon not flashing and highlighted in yellow, indicating system is not ready to execute imaging procedure.
Ranked causes
- 1. Protocol not locked before arming (common)
- 2. Air present in fluid pathway (common)
- 3. Injector not armed (common)
Verify Protocol Lock button has been pressed
- Test point
- Protocol Lock icon in lower left corner of screen
- Expected
- Icon highlighted in yellow
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System unable to arm or injection interrupted patient safety
Injection system fails to arm or stops during injection; error codes displayed on control panel
Ranked causes
- 1. Operator-induced condition (common)
- 2. System malfunction (occasional)
- 3. Voltage transient or noisy electrical environment (occasional)
Perform power reset
- Test point
- Power cord and power source
- Expected
- System resets after one minute power removal and reapplication
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
White screen displaying MEDRAD
Display shows white screen with MEDRAD text, indicates communication or software problem
Ranked causes
- 1. Software mismatched between Display and Head (common)
- 2. Faulty Head extension cable (occasional)
- 3. Faulty Servo/CPU card (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
White screen with MEDRAD displayed
Display shows white screen with MEDRAD text, indicates communication or software problem
Ranked causes
- 1. Software mismatched between Display and Head (unknown)
- 2. Faulty Head extension cable (unknown)
- 3. Servo/CPU card failure (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration error (CAL prefix)
System has found error with system calibration values missing or corrupt
Ranked causes
- 1. Calibration values missing or corrupt (common)
- 2. Faulty Servo/CPU card (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heat maintainer not warm or fault indicator illuminated caution
Heat maintainer does not reach operating temperature or fault indicator light is on while attached to head
Ranked causes
- 1. Heat maintainer heater element failure (occasional)
- 2. Heat maintainer fault sensor triggered (occasional)
Touch heat maintainer while attached to head
- Test point
- Heat maintainer surface
- Expected
- Heat maintainer is warm
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Piston does not auto-dock with syringe plunger caution
Piston fails to automatically dock and advance to full forward position when empty syringe is installed
Ranked causes
- 1. Auto Dock setting disabled or malfunctioning (occasional)
- 2. Syringe not seated properly (occasional)
Install empty syringe in holder
- Test point
- Syringe seat on injector head
- Expected
- Syringe fully seated
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Piston forward/reverse not responding patient safety
Pistons do not advance or retract when forward/reverse controls are activated
Ranked causes
- 1. Motor or drive mechanism failure, or ENABLE key not pressed (common)
- 2. One piston not responding while other piston functions normally (occasional)
Press ENABLE key and verify it activates control mode
- Test point
- ENABLE key
- Expected
- ENABLE key activates when pressed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arm lights not functioning caution
Indicator lights on injector head do not illuminate during operation
Ranked causes
- 1. LED bulb failure or electrical connection issue (occasional)
Verify arm lights illuminate on injector head
- Test point
- Injector head arm lights
- Expected
- Lights illuminate when powered
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arm lights on injector head not functioning caution
Indicator lights on the injector head arm do not illuminate
Ranked causes
- 1. LED or light circuit failure (unknown)
Observe arm lights on injector head
- Test point
- Injector head arm light indicators
- Expected
- Lights illuminate during operation
Audible tones not functioning caution
Display Control Unit or injector head does not emit audible tones during operation or diagnostics
Ranked causes
- 1. Speaker or audio amplifier malfunction (rare)
Listen for audible tones from Display Control Unit and injector head
- Test point
- Display Control Unit and injector head speakers
- Expected
- Audible tones present during operation and diagnostics
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto Advance not enabled caution
Auto Advance feature is disabled when it should be enabled
Ranked causes
- 1. Auto Advance setting disabled in system configuration (unknown)
Verify Auto Advance setting is enabled
- Test point
- System settings menu
- Expected
- Auto Advance enabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto Dock not enabled caution
Auto Dock feature is disabled when it should be enabled for automatic piston docking
Ranked causes
- 1. Auto Dock setting disabled in system configuration (unknown)
Verify Auto Dock setting is enabled
- Test point
- System settings menu
- Expected
- Auto Dock enabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto functions disabled or not responding patient safety
Auto Dock, Auto Advance, or Auto Retract functions are not enabled or do not activate
Ranked causes
- 1. Auto function settings not enabled in system configuration (common)
Verify Auto Dock function is enabled
- Test point
- Auto Dock setting in main menu
- Expected
- Auto Dock is enabled
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto Retract not enabled caution
Auto Retract feature is disabled when it should be enabled
Ranked causes
- 1. Auto Retract setting disabled in system configuration (unknown)
Verify Auto Retract setting is enabled
- Test point
- System settings menu
- Expected
- Auto Retract enabled
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Component corrosion or damage from water exposure caution
Excessive water or cleaning solution contact causing component damage or corrosion
Ranked causes
- 1. Excessive water or cleaning solution exposure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control button malfunction patient safety
Forward, Reverse, Start, Stop, or Fill buttons do not respond to user input
Ranked causes
- 1. Button switch failure or electrical disconnect (occasional)
Test Forward and Reverse buttons
- Test point
- Forward/Reverse button
- Expected
- Button responds and injector advances/retracts
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Controls not responding (Forward, Reverse, Start, Stop, Fill) caution
One or more system controls do not function when operated during checkout
Ranked causes
- 1. Control input not registering (unknown)
Test Forward control functionality
- Test point
- Forward button/control
- Expected
- System advances forward
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Display brightness cannot be adjusted caution
Display brightness controls (up/down keys) do not brighten or darken screen
Ranked causes
- 1. Brightness control button failure or backlight circuit malfunction (occasional)
Press Display Brightness Up key multiple times
- Test point
- Brightness Up key on rear of Display
- Expected
- Screen brightness increases to fullest extent
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display Brightness controls not responding caution
Display Brightness Up or Down keys do not adjust screen brightness
Ranked causes
- 1. Brightness control circuit malfunction (unknown)
Press Display Brightness Up key and observe screen
- Test point
- Display Control Unit brightness up button
- Expected
- Screen brightens to fullest extent
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Electrical fire hazard from incorrect fuse high voltage
Incorrect fuse type installed, creating fire risk
Ranked causes
- 1. Incorrect fuse type installed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrical shock from worn or damaged cabling high voltage
Worn, damaged, or exposed cables create shock hazard during operation or service
Ranked causes
- 1. Worn or damaged cables (occasional)
Visually inspect all cabling for wear or damage
- Test point
- All cables connected to ISI 800 Module
- Expected
- No visible wear, cracks, or damage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Explosion hazard from flammable presence patient safety
Use of injection system in environment with flammable anesthetics or other flammables
Ranked causes
- 1. Flammables present in area (anesthetics, etc.) (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Injection does not complete normally or HOLD feature fails caution
Injection sequence does not finish as programmed or HOLD feature does not activate for specified duration
Ranked causes
- 1. Injection control software or motor drive malfunction (unknown)
Arm system in single mode with empty syringe
- Test point
- System mode setting
- Expected
- System in single mode
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Injection stops due to scanner communication loss patient safety
Injection halts or does not complete normally when communication link between scanner and injector fails during coupled operation.
Ranked causes
- 1. Communication link failure between scanner and injector (occasional)
Monitor communication during injection
- Test point
- ISI Communication icon at bottom left of Monitor
- Expected
- Icon remains highlighted with yellow background throughout injection
Injector fails to arm or disarms unexpectedly patient safety
Injector cannot be armed or automatically disarms during operation when ISI 800 Module is configured.
Ranked causes
- 1. ISI 800 Module communication failure with scanner (common)
Verify ISI Communication icon status
- Test point
- ISI Communication icon at bottom left of Monitor
- Expected
- Icon highlighted with yellow background or no communication problem indicated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient injury from improper accessories patient safety
Use of non-approved accessories or options incompatible with ISI 800 Module
Ranked causes
- 1. Use of non-approved accessories or options (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Piston does not advance to full forward position after injection caution
Piston fails to fully advance forward position after injection sequence completes
Ranked causes
- 1. Piston motor or drive circuit malfunction (unknown)
Command piston to full forward position
- Test point
- Piston position indicator
- Expected
- Piston advances to full forward position
Piston does not auto-retract after syringe removal caution
Piston fails to automatically retract after syringe is removed and advance is complete
Ranked causes
- 1. Auto Retract setting disabled or malfunctioning (occasional)
Remove syringe after piston reaches full forward position
- Test point
- Piston position / Auto Retract mechanism
- Expected
- Piston automatically retracts after syringe removal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Piston does not fully retract caution
Piston fails to return to full reverse position when controlled manually
Ranked causes
- 1. Piston drive motor or mechanical binding (unknown)
Manually command piston to full reverse position
- Test point
- Piston position indicator
- Expected
- Piston reaches full reverse position
Pistons do not respond to forward/reverse controls caution
One or both pistons fail to advance or retract when control buttons are pressed
Ranked causes
- 1. Piston motor or control circuit failure (unknown)
Activate ENABLE key
- Test point
- ENABLE key on Display Control Unit
- Expected
- Piston motors enabled
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Safety screen does not display after power-up caution
System powers on but Safety screen does not appear after diagnostics complete
Ranked causes
- 1. Diagnostics did not complete properly (unknown)
Verify Safety screen displays after system diagnostics
- Test point
- Display screen
- Expected
- Safety screen appears after diagnostics complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System damage from incorrect voltage supply caution
Unit connected to outlet with voltage or frequency that does not match system nameplate rating
Ranked causes
- 1. Voltage or frequency mismatch with outlet (occasional)
Verify voltage and frequency match nameplate rating
- Test point
- Serial tag on back of unit and electrical outlet
- Expected
- Voltage and frequency on serial tag match outlet specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System does not power up properly patient safety
Safety screen not displayed after system diagnostics, or system fails to respond to controls after power-up
Ranked causes
- 1. System diagnostics failure or initialization incomplete (unknown)
Verify Safety screen appears after power-up diagnostics
- Test point
- Display Control Unit screen
- Expected
- Safety screen displayed after system diagnostics
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System labels missing or illegible patient safety
Safety and warning labels on the system are not present or cannot be read clearly
Ranked causes
- 1. Labels damaged, faded, or removed (unknown)
Visually inspect all system safety and warning labels
- Test point
- All visible areas of injector head and Display Control Unit
- Expected
- All labels present and legible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System malfunction during operation patient safety
System does not operate correctly or displays uncorrectable fault message
Ranked causes
- 1. System malfunction (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System malfunction from missed preventive maintenance patient safety
System not calibrated or functioning properly due to lack of regular maintenance
Ranked causes
- 1. Failure to perform regular preventive maintenance (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System malfunction occurs during patient injection patient safety
System malfunction message appears on display during or after patient injection procedure
Ranked causes
- 1. Electrical or mechanical failure within injector (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Type 3 system malfunction message displayed patient safety
System displays Type 3 malfunction message requiring power removal. Message may recur if underlying condition not corrected. Code number shown in lower left corner of dialog box.
Ranked causes
- 1. Unknown system malfunction requiring diagnostic action (unknown)
Immediately remove system from patient
- Test point
- System connection to patient
- Expected
- System disconnected from patient
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails after installation patient safety
ISI 800 Module may not function properly if installation is incorrect
Ranked causes
- 1. Improper installation by untrained personnel (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
CAL | Calibration error | System found error with calibration values missing or corrupt | error |
CAL | Calibration error with missing or corrupt system calibration values | System detects missing or corrupt calibration values | error |
DOCK SENSA | Docking sensor A error | Piston A docking sensor closed with no syringe attached on side A | warning |
DOCK SENSA | Piston A docking sensor closed with no syringe attached | Docking sensor A detects closed state without syringe | error |
DOCK SENSB | Docking sensor B error | Piston B docking sensor closed with no syringe attached on side B | warning |
DOCK SENSB | Piston B docking sensor closed with no syringe attached | Docking sensor B detects closed state without syringe | error |
ENCA | Encoder A error | Problem with Encoder on motor A or associated circuitry | error |
ENCA | Encoder A problem on motor A or associated circuitry | Error detected during encoder or motor A circuit test | error |
ENCB | Encoder B error | Problem with Encoder on motor B or associated circuitry | error |
ENCB | Encoder B problem on motor B or associated circuitry (Dual only) | Error detected during encoder or motor B circuit test | error |
FRA | Flex Ring problem on side A | Error detected in Flex Ring A system | error |
FRB | Flex Ring problem on side B | Error detected in Flex Ring B system | error |
HOVER | Head or DCU overlay switch card problem | Error detected in head, switch card or associated circuitry during diagnostics | error |
HSW | Handswitch problem or associated circuitry failure | Error detected in handswitch system | error |
HUB | Ethernet HUB problem or associated circuitry failure | Error detected in Ethernet HUB or circuit | error |
ICD | Display Interface Card problem or associated circuitry failure | Error detected in Display Interface Card or circuit | error |
ICE | Expansion Base Interface Card problem or associated circuitry failure | Error detected in Base Interface Card or circuit | error |
ICS | Head Interface Card problem or associated circuitry failure | Error detected in Head Interface Card(s) or circuit | error |
MOTA | Motor A problem or associated circuitry during H Bridge test | Error detected during motor A or H Bridge circuit test | error |
MOTB | Motor B problem or associated circuitry during H Bridge test (Dual only) | Error detected during motor B or H Bridge circuit test | error |
PCMCIA | PCMCIA card problem or associated circuitry failure | Error detected in PCMCIA card or circuit | warning |
POTA | Potentiometer A problem or associated circuitry failure | Error detected in potentiometer A or circuit during test | error |
POTB | Potentiometer B problem or associated circuitry failure (Dual only) | Error detected in potentiometer B or circuit during test | error |
POWR | Power Drive Card problem or associated circuitry failure | Error detected in power card or circuit | error |
PRSSRA | Pressure Transducer A problem or associated circuitry failure | Error detected in pressure transducer A or circuit | error |
PRSSRB | Pressure Transducer B problem or associated circuitry failure (Dual only) | Error detected in pressure transducer B or circuit | error |
RATE | Over rate injection condition detected | System detects injection rate exceeds limits | error |
SBC | Single Board Computer problem or associated circuitry failure | Error detected in SBC or circuit | error |
SER | Servo/CPU Card problem, may be electrical transient or hardware failure | Error detected in Servo/CPU card or Head card | error |
SSD | Start Switch (Display) problem or connection failure | Error detected in Display start switch or circuit | error |
SSH | Start Switch (Head) problem or connection failure | Error detected in Head start switch or circuit | error |
SW | Internal software problem, may be electrical transient or corruption | Software error detected during operation or self-test | error |
SWMTCH | Software mismatch between Head and Display | Software version mismatch detected between Head and Display CPUs | error |
SWSBC | CPU software problem on display CPU card | Software error detected on display CPU card | error |
SWSER | Servo software problem on Head CPU card | Software error detected on Head CPU card | error |
SYR SENA | Size Sensor A problem or associated circuitry failure | Error detected in syringe size sensor on A side or circuit | error |
SYR SENB | Size Sensor B problem or associated circuitry failure (Dual only) | Error detected in syringe size sensor on B side or circuit | error |
TCHSCR | Touch screen problem or associated circuitry failure | Error detected in touch screen or circuit during diagnostics | error |
Type 3 | System malfunction message requiring power removal | Various system malfunctions depending on error category and subcategory | critical |
UOM | Unintended opposite motion, DSP routing timeout | DSP software routine did not report within allowed time | error |
VOL | Over volume injection condition detected | System detects injection volume exceeds limits | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.